Orthopaedic Proceedings
Vol. 93-B, Issue SUPP_IV | Pages 469 - 469
1 Nov 2011
Ball S Hulst J Wu G LeDuff M Amstutz H
Full Access

Recently, monoblock cups have increased in popularity for hip resurfacing and large femoral head total hips. However, there have been no studies specifically evaluating the durability of this type of cup. The purpose of this study was to define the mid-term survivorship of cobalt-chrome alloy, monoblock acetabular components.

A retrospective radiographic review of 426 consecutive hip resurfacings using the ConserveÒPlus prosthesis was performed with specific attention to the acetabular component.

Radiographs were analyzed for cup position, the presence of radiolucencies, cup migration, bead shedding, osteolysis and stress remodeling of the pelvic bone. Kaplan-Meier (KM) survival estimates were calculated using revision for aseptic loosening of the acetabular component as the end point.

Average follow-up was 8.6 years (range 5.4 to 12.3). Mean abduction angle and anteversion angle were 46.6° (± 6.8°), and 21.6° (± 8.6°), respectively. Radiolucent gaps behind the cup from incomplete seating were visible in zone 2 in 16% of cases. These were typically 1 to 2 mm in size and radiographically filled in all but 2 cases. No cups with early lucencies went on to fail. Late radiolucencies developed in zone 1 in 8 cups (1.9%), in zone 2 in 8 cups (1.9%), and in zone 3 in 19 cups (4.5%). Radiolucencies in multiple zones were seen in 6 cups (1.4%).

Small amounts of socket migration (2mm or less) were suspected in 3 cups (0.7%) but each of these has remained stable. There were no cases of bead shedding. Small osteolytic lesions were suspected in 12 hips (2.8%). There were 2 revisions for aseptic loosening of the cup at 5 and 8 years, and one revision for protrusion of the cup through the medial wall 4 days after surgery.

Additionally, one cup at 9 years follow-up is believed to be loose but has yet to be revised. The KM survival estimate was 99.6% at 5 years (95% C.I. 98.4% to 99.9%) and 98.7% at 10 years (95% C.I. 94.5% to 99.7%).

In conclusion, this study demonstrates excellent mid-term survivorship of a cobaltchrome alloy monoblock acetabular component, which matches that of conventional titanium implants. Small early gaps seen behind the cup from incomplete seating do not appear to effect cup survivorship as long as a good peripheral press-fit is obtained. Osteolysis with this prosthesis is rare but does occur.