The acetabular component is a monobloc hemispherical cup manufactured from Ultra High Molecular Weight Polyethylene (UHMWPE), with a pure titanium particle coated surface. With heat and pressure, the particles are blasted into the polyethylene surface. The coating promote osseointegration. Stability of the cup is provided by 2 anchoring pegs on the weight bearing part on its outer surface. The inclination of pegs and holes diverge by 5 degrees providing a press-fit effect that increases the rigidity of the primary fixation and this is supplemented by screws inserted through the periphery of the cup. Tow cups designs are available, a full-profile and a bevelled cup. In all cases the bevelled cup was used. The purpose of this study was to assess the clinical performance with 15 years of this cup.
The diagnosis was osteoarthritis, rheumatic arthritis, femur neck fracture, developmental dysplasia of the hip (DDH). 1034 patients were contacted by telephone. Out of this patients group 539 patients (678 cups) were evaluated by clinical examination (HHS), radiographic investigation and social evaluation by the WOMAC and NHS score. 451 patients who had died unrelated to the operation. Lost of follow up were 65 patients (69 Cups) and 34 patients (48 cups) had to be revised.
In a prospective study, 93 unselected consecutive uncemented hip arthroplasties were performed in 80 patients using the titanium-coated RM acetabular component and the CLS femoral component. The mean age of the patients at operation was 52 years (28 to 81). None were lost to follow-up. In the 23 patients who had died (26 hips) only one acetabular component had been revised. In the 57 living patients (67 hips), 13 such revisions had been performed. Of the 14 revisions, seven were for osteolysis, five for loosening and two for infection. Survival analysis of this implant showed a total probability of survival of 83% (95% confidence interval 73 to 90), with all revisions as the endpoint, and a probability of 94% (95% confidence interval 87 to 98) with revision for aseptic loosening as the endpoint, indicating reliable long-term fixation of the titanium-coated RM acetabular component.