Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_11 | Pages 4 - 4
1 Aug 2018
Cobb J Clarke S Halewood C Wozencroft R Jeffers J Logishetty K Keane B Johal H
Full Access

We aimed to demonstrate the clinical safety of a novel anatomic cementless ceramic hip resurfacing device. Concerns around the safety of metal on metal arthroplasty have made resurfacing less attractive, while long term function continues to make the concept appealing. Biolox Delta ceramic is now used in more than 50% of all hip arthroplasties, suggesting that it's safety profile is acceptable. We wondered if a combination of these concepts might work?

The preclinical testing of anatomic hip resurfacing device developed by our group was presented last year. A twenty patient safety study was designed. Patients had to be between the ages of 18 and 70. The initial size range was restricted to femoral heads between 46 and 54, representing the common sizes of hip resurfacing. The primary outcomes were clinical safety, PROMs and radiological control. Secondary outcomes include CTRSA and metal ion levels.

20 patients were recruited, aged 30–69. 7 were women and 13 were men. There were no operative adverse events in their operations undertaken between September 2017 and February 2018. One patient had a short episode of atrial fibrillation on the second postoperative day, and no other complications. At three months the median oxford hip score had risen from 27 (range 14–38) to 46 (31–48). Cobalt and chromium levels were almost undetectable at 3 months. Fixation appeared satisfactory in all patients, with no migration detected in either component. CTRSA is in process.

The initial safety of a novel cementless ceramic resurfacing device is demonstrated by this data. The 10 year, 250 case efficacy study will continue in 5 other European centres.