Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_17 | Pages 9 - 9
11 Oct 2024
Zace P Maas Z McIntyre R Khan Z Bailey O
Full Access

Increasing the accuracy of information provided through X-Rays maximises pre-operative planning. Aim of this project is to determine the necessity of calibration probes that would improve the accuracy of pre-operative templating. This is a retrospective study involving leg length and pelvis X-Rays performed across the NHS Lanarkshire from 01/03/2023 until 31/04/2024.

A total of 87 leg length X-Rays were identified, 18 had a calibration probe present. Leg length was measured on each and the X-Rays were calibrated against the existing probe. In 66.7% of cases there was a major leg length discrepancy of over 2cm between the pre-calibrated and post-calibrated X-Rays. Pelvic X-Rays of 80 patients that underwent total hip replacement were reviewed. Preoperative templating was compared to the implants inserted. An average of 1.94 discrepancy in the size of the acetabular implant was identified whilst in 30 cases the size of the femoral stem was incorrect by at least 1 size. Magnification of 119.7% on X-Rays was found to provide the most accurate templating. Seventy seven cases of pelvic X-Rays before and after hip hemiarthroplasty were also reviewed. The implant head was templated incorrectly in 74% of cases and the stem in 51%. It was identified that pelvic X-Ray magnification of 121.7% would provide the most accurate results.

X-Rays with no calibration probes provide inaccurate measurements leading to faulty preoperative planning. Standardised use of a calibration sphere is strongly suggested and whenever that is not available, we suggest magnification of 121%.