Bone & Joint Research
Vol. 10, Issue 12 | Pages 780 - 789
1 Dec 2021
Eslam Pour A Lazennec JY Patel KP Anjaria MP Beaulé PE Schwarzkopf R

Aims

In computer simulations, the shape of the range of motion (ROM) of a stem with a cylindrical neck design will be a perfect cone. However, many modern stems have rectangular/oval-shaped necks. We hypothesized that the rectangular/oval stem neck will affect the shape of the ROM and the prosthetic impingement.

Methods

Total hip arthroplasty (THA) motion while standing and sitting was simulated using a MATLAB model (one stem with a cylindrical neck and one stem with a rectangular neck). The primary predictor was the geometry of the neck (cylindrical vs rectangular) and the main outcome was the shape of ROM based on the prosthetic impingement between the neck and the liner. The secondary outcome was the difference in the ROM provided by each neck geometry and the effect of the pelvic tilt on this ROM. Multiple regression was used to analyze the data.


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 808 - 816
1 Jul 2019
Eftekhary N Shimmin A Lazennec JY Buckland A Schwarzkopf R Dorr LD Mayman D Padgett D Vigdorchik J

There remains confusion in the literature with regard to the spinopelvic relationship, and its contribution to ideal acetabular component position. Critical assessment of the literature has been limited by use of conflicting terminology and definitions of new concepts that further confuse the topic. In 2017, the concept of a Hip-Spine Workgroup was created with the first meeting held at the American Academy of Orthopedic Surgeons Annual Meeting in 2018. The goal of this workgroup was to first help standardize terminology across the literature so that as a topic, multiple groups could produce literature that is immediately understandable and applicable. This consensus review from the Hip-Spine Workgroup aims to simplify the spinopelvic relationship, offer hip surgeons a concise summary of available literature, and select common terminology approved by both hip surgeons and spine surgeons for future research.

Cite this article: Bone Joint J 2019;101-B:808–816.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 42 - 42
1 Apr 2019
Kim YW Lazennec JY Hani J Pour AE
Full Access

Background

Postural change after total hip arthroplasty (THA) is still a matter of discussion. Previous studies have mainly concentrated on the pelvic motions. We report the postoperative changes of the global sagittal posture using pelvic, spinal and lower extremities parameters.

Methods

139 patients (primary THA, without previous spinal or lower extremity surgery) were included. We measured pelvic parameters [SS: Sacral Slope, PI: Pelvic Incidence, PT: Pelvic Tilt, APP angle: Anterior Pelvic Plane angle] and the global posture parameters (SVA: Sagittal Vertical Angle, GSA: Global Sagittal Angle, TPA: T1 pelvic angle). Patients were categorized into low PI group <45°, 45°< medium PI <65° and high PI >65°.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 53 - 53
1 Apr 2019
Lazennec JY Kim YW Hani J Pour AE
Full Access

Introduction

Spatial orientation of the pelvis in the sagittal plane is a key parameter for hip function. Pelvic extension (or retroversion) and pelvic flexion(or anteversion) are currently assessed using Sacral Slope (SS) evaluation (respectively SS decrease and SS increase). Pelvic retroversion may be a risk situation for THA patients. But the magnitude of SS is dependant on the magnitude of pelvic incidence (PI) and may fail to discriminate pelvic position due to patient's anatomy and the potential adaptation mechanisms: a high PI patient has a higher SS but this situation can hide an associated pelvic extension due to compensatory mechanisms of the pelvic area. A low PI patient has a lower SS with less adaptation possibilities in case of THA especially in aging patients. The individual relative pelvic version (RPV) is defined as the difference between « measured SS » (SSm) minus the « normal SS »(SSn) described for the standard population. The aim of the study was to evaluate RPV in standing and sitting position with a special interest for high and low PI patients.

Materials and Methods

96 patients without THA (reference group) and 96 THA patients were included. Pelvic parameters (SS and PI) were measured on standing and sitting EOS images. RPV standing (SSm-SSn) was calculated using the formula SSm – (9 + 0.59 × PI) according to previous publications. SSn in sitting position was calculated according to PI using linear regression: RPV sitting was calculated using the formula RPV = SS – (3,54+ 0,38 × PI). Three subgroups were defined according to pelvic incidence (PI): low PI <45°, 45°<normal PI<65° or high PI>65°.


Introduction

Literature describes pelvic rotation on lateral X rays from standing to sitting position. EOS full body lateral images provide additional information about the global posture. The projection of the vertical line from C7 (C7 VL) is used to evaluate the spine balance. C7 VL can also measure pelvic sagittal translation (PST) by its horizontal distance to the hip center (HC). This study evaluates the impact of a THA implantation on pelvic rotation and sagittal translation.

Materials and Method

Lumbo-pelvic parameters of 120 patients have been retrospectively assessed pre and post- operatively on both standing and sitting acquisitions (primary unilateral THA without complication). PST is zero when C7VL goes through the center of the femoral heads and positive when C7VL is posterior to the hips' center (negative if anterior). Three subgroups were defined according to pelvic incidence (PI): low PI <45°, 45°<normal PI<65° or high PI>65°.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 38 - 38
1 Apr 2019
Lazennec JY Rakover JP Rousseau MA
Full Access

INTRODUCTION

Lumbar total disc replacement (TDR) is an alternative treatment to avoid fusion related adverse events, specifically adjacent segment disease. New generation of elastomeric non-articulating devices have been developed to more effectively replicate the shock absorption and flexural stiffness of native disc. This study reports 5 years clinical and radiographic outcomes, range of motion and position of the center of rotation after a viscoelastic TDR.

Material and methods

This prospective observational cohort study included 61 consecutive patients with monosegmental TDR. We selected patients with intermediate functional activity according to Baecke score. Hybrid constructs had been excluded. Only cases with complete clinical and radiological follow-up at 3, 6, 12, 24 and 60 months were included. Mean age at the time of surgery was 42.8 +7.7 years-old (27–60) and mean BMI was 24.2 kg/m² +3.4 (18–33). TDR level was L5-S1 in 39 cases and L4-L5 in 22 cases.

The clinical evaluation was based on Visual Analog Scale (VAS) for pain, Oswestry Disability Index (ODI) score, Short Form-36 (SF36) including physical component summary (PCS) and mental component summary (MCS) and General Health Questionnaire GHQ28. The radiological outcomes were range of motion and position of the center of rotation at the index and the adjacent levels and the adjacent disc height changes.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 46 - 46
1 Apr 2019
Kim YW Girinon F Lazennec JY Skalli W
Full Access

Introduction

Stand to sit pelvis kinematics is commonly considered as a rotation around the bicoxofemoral axis. However, abnormal kinematics could occur for patients with musculoskeletal disorders affecting the hip-spine complex. The aim of this study is to perform a quantitative analysis of the stand to sit pelvis kinematics using 3D reconstruction from bi-planar x-rays.

Materials and Methods

Thirty healthy volunteers as a control group (C), 30 patients with hip pathology (Hip) and 30 patients with spine pathology (Spine) were evaluated. All subjects underwent standing and sitting full-body bi-planar x-rays. 3D reconstruction was performed in each configuration and then translated such as the middle of the line joining the center of each acetabulum corresponds to the origin. Rigid registration quantified the finite helical axis (FHA) describing the transition between standing and sitting with two specific parameters. The orientation angle (OA) is the signed 3D angle between FHA and bicoxofemoral axis and the rotation angle (RA) represents the signed angle around FHA. Pelvic incidence, sacral slope and pelvic tilt were also measured. After checking normality of distribution, parameters were compared statistically between the 3 groups (p<0.05).


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 96 - 96
1 Apr 2019
Lazennec JY Rischke B Rakover JP Ricart O Rousseau MA
Full Access

Introduction

The viscoelastic cervical disk prosthesis CP-ESP is an innovative one-piece deformable but cohesive interbody spacer. It is an evolution of the LP-ESP lumbar disk implanted since 2006. The implant provides 6 full degrees of freedom including shock absorption. The design allows a limitation for rotation and translation with resistance to motion (elastic return property) aimed at avoiding overload of the posterior facets. The rotation center can vary freely during motion. It thus differs substantially from current prostheses.

This study reports the clinical results of a prospective observational study series of 89 patients who are representative of the current use of the ESP implant since 2012.

The radiological results are focused on the evolution of the mean center of rotation (MCR) as an additional information to the range of motion (ROM) for the evaluation of the quality of spine movement.

Materials and Methods

89 patients (33 males, mean age 45 years [28–60], 107 implants) were included for an open, prospective and non-randomized study between October 2012 and December 2015.

One level patients were at C3C4 (3), C4C5 (3), C5C6 (41) C6C7 (24) C7T1 (1)

Two levels patients were C4C5/C5C6 (3), C5C6/C6C7 (12), C6C7/C7D1 (1) and 3 levels C4C5/C5C6/C6C7 (1)


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 139 - 139
1 Jan 2016
Lazennec JY Brusson A Ricart O Rakover JP
Full Access

Introduction

The viscoelastic cervical disk prosthesis ESP is an innovative one-piece deformable but cohesive interbody spacer. It is an evolution of the LP ESP lumbar disk implanted since 2006. CP ESP provides 6 full degrees of freedom about the 3 axes including shock absorption. The prosthesis geometry allows limited rotation and translation with resistance to motion (elastic return property) aimed at avoiding overload of the posterior facets. The rotation center can vary freely during motion (figure 1)

It thus differs substantially from current prostheses. This study reports the results of a prospective series of 49 patients who are representative of the current use of the ESP implant since 2012.

Material and methods

The surgeries were performed by 3 senior surgeons. There were 34 women and 15 men in this group. The average age was 44±7 (32–59). The implantation was single level in 78 % of cases. 55 CP ESP prostheses were analyzed.

Clinical data and X-rays were collected at the preoperative time and at 3, 6, and 12 months post-op.

The functional results were measured using Neck and Arm VAS, NDI, SF-36, (physical componentPCS and mental component MCS). The analysis was performed by a single observer who was independent from the selection of patients and from the surgical procedure.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 141 - 141
1 Jan 2016
Lazennec JY Brusson A Rousseau M Clarke I Pour AE
Full Access

Introduction

The assessment of leg length is essential for planning the correction of deformities and for the compensation of length discrepancy, especially after hip or knee arthroplasty. CT scan measures the “anatomical” lengths but does not evaluate the “functional” length experienced by the patients in standing position. Functional length integrates frontal orientation, flexion or hyperextension. EOS system provides simultaneously AP and lateral measures in standing position and thus provides anatomical and functional evaluations of the lower limb lengths.

The objective of this study was to measure 2D and 3D anatomical and functional lengths, to verify whether these measures are different and to evaluate the parameters significantly influencing these potential differences

Material and Methods

70 patients without previous surgery of the lower limbs (140 lower extremities) were evaluated on EOS images obtained in bipodal standing position according to a previously described protocol.

We used the following definitions:

anatomical femoral length between the center of the femoral head (A) and center of the trochlea (B)

anatomical tibial length between the center tibial spine (intercondylar eminence) (C) and the center of the ankle joint (D)

functional length is AD

global anatomical length is AB + CD

Other parameters measured are HKA, HKS, femoral and tibial mechanical angles (FMA, TMA), angles of flexion or hyperextension of the knee, femoral and tibial torsion, femoro-tibial torsion in the knee, and cumulative torsional index (CTI). All 2D et3D measures were evaluated and compared for their repeatability.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 142 - 142
1 Jan 2016
Lazennec JY Brusson A Pour AE Rousseau M
Full Access

Introduction

The gold standard for knee surgery is the restoration of the so-called «neutral mechanical alignment ». Recent literature as pointed out the patients with «constitutional varus »; in these cases, restoring neutral alignment could be abnormal and even undesirable. The same situation can be observed in patients with «constitutional valgus alignment ». To date, these outliers cases have only been explored focusing on the lower limb; the influence of the pelvic morphotype has not been studied. Intuitively, the pelvic width could be a significant factor. The EOS low dose imaging technique provides full body standing X-rays to evaluate the global anatomy of the patient. This work explores the influence of the pelvic parameters on the frontal knee alignment.

Material and methods

– We included 170 patients (340 lower extremities). 2 operators performed measurements once per patient on AP X-rays.

The classical anatomical parameters were:

Femoral mechanical angle (FMA)

Tibial mechanical angle (TMA)

Hip knee shaft angle (HKS)

Hip knee ankle angle (HKA)

Femoral and tibial lengths

The morphotype was evaluated by:

the distances between the center of two femoral heads (FHD), between knees (KD) and between ankles (AD)

the medial neck-shaft angle (MNSA)

the femoral offset

The horizontal distance between the limb mechanical axis (line passing from center of the femoral head to the center of the ankle) and the center of the knee was called the intrinsic mechanical axis deviation (IMAD) (fig 1). The horizontal distance between the pelvic mechanical axis (line from the center of the sacral plate to the center of the ankle) and the center of the knee was called the global mechanical axis deviation (GMAD) (fig 2).

Inter-Operator Reliability was calculated with Intra-class Correlation Coefficient (ICC) and Inter-Reader Agreement was assessed with Bland-Altman test.

A relationship between IMAD and GMAD to the other parameters was assessed using Pearson's correlation coefficient.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_2 | Pages 140 - 140
1 Jan 2016
Lazennec JY Brusson A Rousseau M Clarke I Pour AE
Full Access

Introduction

Coronal misalignment of the lower limbs is closely related to the onset and progression of osteoarthritis. In cases of severe genu varus or valgus, evaluating this alignment can assist in choosing specific surgical strategies. Furthermore, restoring satisfactory alignment after total knee replacement promotes longevity of the implant and better functional results. Knee coronal alignment is typically evaluated with the Hip-Knee-Ankle (HKA) angle. It is generally measured on standing AP long-leg radiographs (LLR). However, patient positioning influences the accuracy of this 2D measurement. A new 3D method to measure coronal lower limb alignment using low-dose EOS images has recently been developed and validated. The goal of this study was to evaluate the relevance of this technique when determining knee coronal alignment in a referral population, and more specifically to evaluate how the HKA angle measured with this 3D method differs from conventional 2D methods.

Materials and methods

70 patients (140 lower extremities) were studied for 2D and 3D lower limb alignment measurements. Each patient received AP monoplane and biplane acquisition of their entire lower extremities on the EOS system according the classical protocols for LLR. For each patient, the HKA angle was measured on this AP X-ray with a 2D viewer. The biplane acquisition was used to perform stereoradiographic 3D modeling. Valgus angulation was considered positive, varus angulation negative. Student's T-test was used to determine if there was a bias in the HKA angle measurement between these two methods and to assess the effect of flexion/hyperextension, femoral rotation and tibial rotation on the 2D measurements. One operator did measurements 2 times.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 265 - 265
1 Dec 2013
Clarke I Lazennec JY Brusson A Burgett M Donaldson T
Full Access

This study of retrieved 28 mm Metasul™ (cemented) revealed for the first time adverse wear effects created by impingement-subluxation in MOM. The 10 cases selected (with femoral stems) had annual follow-up 3–11 years. (1) Unequivocal evidence here shows that all heads routinely subluxed from the Metasul liners.

Femoral stems revealed well-demarcated notches (DN) on necks and trunnions (Fig. 1a: n = 6), shallow cosmetic blemishes (Fig. 1b CB: n = 4), and abrasion by cement (Fig. 1b: PMMA). As demonstrated by EOS radiographs, impingement locations varied with implant positioning, pelvic mobility and patient functionality – both anterior and posterior notching (Fig. 1). The first impingement notch occurred with head located (Fig. 2a), whereas the head had subluxed from the cup at 2nd notch (Fig. 2b). The model demonstrated that patients gained 20° motion by such head-subluxation manoeuvres. It was surprising that there was no collateral damage evident on the liners. Even with severe notching of Ti6Al4V and CoCr stems, the cup rims generally appeared well-polished.

Femoral heads revealed macro-stripe damage on articular surfaces (Fig. 3), as did cups. Basal and polar macro-stripes on heads were always located at hip impingement positions. The equatorial stripes were formed at main-wear zone boundaries. Thus equatorial stripes were likely created by some form of rim-impact damage (micro-separation) or by local ingress of 3rd-body wear particles under the cup rim.

Micro-grooving was evident within these macro-size stripes and frequently featured large raised lips (Fig. 3), interpreted as signs of adverse 3rd-body wear mechanisms, and rarely described.(2) It would appear that large metal particulates were released during MOM impingement-subluxation manoeuvres and circulated the hip joint to producing severe 3rd-body abrasion. Gradual decomposition of such large debris to nano-sized particulates under joint loading would then produce the often-referenced ‘self polishing’ effect of CoCr. EDS studies revealed metal smears on the CoCr surfaces containing the elements of titanium alloy (Ti, Al, V). This was further evidence of impingement-subluxation manoeuvres.(1, 3)

In-vivo cup wear patterns also appeared much larger than those produced in MOM simulators. Such differences likely reflected head-subluxation in vivo, whereby heads unconstrained by the subluxation maneuver were free to orbit up and even cross cup rims, i.e. “edge wear”. This appears to be the first study detailing the adverse wear mechanisms in MOM bearings. There are two limitations to our retrieval study, a) these wear results may not be representative for all MOM designs, and b) it is unknown whether such results have relevance to MOM cases continuing successfully.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 152 - 152
1 Dec 2013
Pour AE Lazennec JY Brusson A Rousseau M Clarke I
Full Access

Introduction

Accurate evaluation of femoral offset is difficult with conventional anteroposterior (AP) X-rays. Routine CT imaging is costly and exposes patients to a significant dose of radiation. The EOS® imaging system is an innovative slot-scanning radiography system that makes possible the acquisition of simultaneous and orthogonal AP and lateral images of the patient in standing position. These 2-dimensional (2D) images are equivalent to standard plane X-rays. Three-dimension (3D) reconstructions are obtained from these paired images according to a validated protocol. This prospective study explores for the first time the value of the EOS® imaging system for comparing measurements of femoral offset obtained from 2D images and 3D reconstructions.

Materials and Methods

Following our standard protocol, we included a series of 100 patients with unilateral total hip arthroplasty (THA). The 2D offset was measured on the AP view with the same protocol as for standard X-rays. The 3D offset was calculated from the reconstructions based on the orthogonal AP and lateral views. Reproducibility and repeatability studies were conducted for each measurement. We compared the 2D and 3D offsets for both hips (with and without THA).


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 389 - 389
1 Dec 2013
Lazennec JY Brusson A Rakover JP Rousseau M
Full Access

Introduction

The viscoelastic lumbar disk prosthesis ESP is an innovative one-piece deformable but cohesive interbody spacer; it provides 6 full degrees of freedom about the 3 axes including shock absorption. The prosthesis geometry allows limited rotation and translation with resistance to motion (elastic return property) aimed at avoiding overload of the posterior facets. The rotation center can vary freely during motion. It thus differs substantially from current prostheses. This study reports the results of a prospective series of 120 patients who are representative of the current use of the ESP implant since 2006.

Material and methods

The surgeries were performed by 2 senior surgeons. There were 73 women and 47 men in this group. The average age was 42 (27–60). The average body mass index was 24.2 kg/m2 (18–33). The implantation was single level in 89% of cases. 134 ESP prostheses were analyzed.

Clinical data and X-rays were collected at the preoperative time and at 3, 6, 12, 24, and 36 months post-op.

The functional results were measured using VAS, GHQ 28, ODI, SF-36, (physical component PCS and mental component MCS. The analysis was performed by a single observer who was independent from the selection of patients and from the surgical procedure.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 117 - 117
1 Dec 2013
Clarke I Lazennec JY Brusson A Burgett M Donaldson T
Full Access

Controversy exists over the role of fretting-corrosion in modular junctions of large-diameter metal-on-metal (MOM) heads given the many design plus alloy mix-and-match variations. Overall data was also scant regarding a) fitting stem trunnions to head tapers, b) role of taper angles, c) role of smooth vs threaded trunnion junctions, d) role of head neck-lengths and e) role of head diameters. While the “12: 14” taper has been used with small CoCr heads for 40 years, we could not find retrieval analyses on this European ‘gold-standard’. We therefore selected 10 femoral stems with 28 mm modular heads for analysis (3–8 years follow-up). Unique to this study were the threaded taper profiles on both stems and heads (Fig. 1).

Six stems were cemented Ti6Al4V (Alize, FH-Orthopedics, France) with 12/14 taper angle defined as 5° 42′. These represented Ti64: CoCr combinations from 2 vendors. The other four were CoCr stems including the CoCrMo (Protasul-2) and CoNiCrMo (Protasul-10) alloys (cemented and HA-coated; Sulzer, Switzerland). These CoCr: CoCr combinations from one vendor had “12/14” stem-taper defined as 5° 38′. Anatomical positioning of Metasul heads (Sulzer, Switzerland) was identified by main-wear zone maps. Femoral heads were then bi-valved in horizontal plane for direct imaging by interferometry (WLI) and SEM. Visual corrosion mapping (3) was recorded digitally in 4 anatomical views. Quantitative analysis used 1 to 5 taper zones with 6-replicate measurements per zone (Fig. 1).

The WLI and SEM studies showed that non-contacting taper zones inside CoCr heads (Fig. 2) were threaded with pitch of 70 μm (PV: peak-valley depth = 5–7 μm). The non-contact zones on Sulzer stems had 130 μm pitch (PV = 4–8 μm) whereas Alize stems had 210 μm pitch (PV = 10–12 μm). Threads on both stem types were much coarser than CoCr heads; Ti64 stem threads were much coarser than CoCr stems. In contact zones, the Metasul threads had flattened (avg. roughness = 0.45 μm Ra). With CoCr stems there was little difference. Difference in pitch of stem-threads vs head-threads indicated there was no imprinting onto head tapers. Nor were there statistically significant differences evident in the contact zones along CoCr or Ti64 tapers. Small damaged areas (Fig. 3: arrows) may have been due to alternatively; initial machining, surgical impaction, in-vivo cold-welds, fretting, corrosion, or from surgical-removal. The as labeled “corrosion damage” was well within the “mild” grade for all implants.(3) Thus even with this considerable variety of design and material parameters, we were satisfied that these gold-standard taper junctions with threaded interfaces had performed very well with 28 mm MOM at 3–8 years follow-up.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 484 - 484
1 Dec 2013
Pour AE Lazennec JY Brusson A Rousseau M
Full Access

Introduction

The position and orientation of the lower extremities are fundamental for planning and follow-up imaging after arthroplasty and lower extremity osteotomy. But no studies have reported the reproducibility of measurements over time in the same patient, and experience shows variability of the results depending on the protocols for patient positioning. This study explores the reproducibility of measurements in the lower extremity with the patients in “comfortable standing position” by the EOS® imaging system.

Materials and Methods

Two whole-body acquisitions were performed in each of 40 patients who were evaluated for a spine pathology. The average interval between acquisitions was 15 months (4–35 months). Patients did not have severe spine pathology and did not undergo any surgery between acquisitions. The “comfortable standing position” is achieved without imposing on the patient any specific position of the lower limbs and pelvis. All the measurements were performed and compared in both 2- and 3-dimensional images. Distances between the centers of the femoral heads and between the centers of the knees and ankles were measured from the front. The profile is shown by the flexion angle between the axis of the femur (center of the femoral head and the top of the line Blumensaat) and the axis of the tibia.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 391 - 391
1 Dec 2013
Lazennec JY Brusson A Ebramzadeh E Clarke I
Full Access

Unlike conventional radiographic methods, the newly introduced EOS system provides simultaneously-synchronized anteroposterior (AP) and true-lateral (LAT) x-ray images. EOS offers considerable potential for calculating parameters such as true femoral and acetabular angular positioning, impingement sites, and also for measuring wear in polyethylene cups. In this study we used THA wear-simulation fixtures to assess 3D-wear in polyethylene cups using EOS algorithms.

Material and methods

A validated phantom apparatus was used to simulate values of three-dimensional wear, controlled in the 3 directions (antero-posterior, medio-lateral, cranio-caudal) using micrometers. (Figure 1)

24 simulations of wear with controlled amplitudes and directions were imaged using the biplane EOS slot-scanning system. Wear amplitudes were between 0 and 3464 μm. Using dedicated software, wear was measured by a 2D/3D matching of 3D spheres onto the 2D frontal and lateral radiographs, allowing the determination of the 3D coordinates of both the cup and femoral head centers and thus the calculation of a 3D wear vector. (Figure 2)

Measured wear vector were compared to real wear vectors in terms of amplitude and direction.3D wear vectors were measured twice by 3 independent observers (for a total of 144 measurements) in order to evaluate intra- and inter-observer reliability.

Results

There was a strong correlation between the measured wear amplitude and the real wear amplitude (Pearson's r = 0,99). Mean error when comparing wear measurement amplitude with real wear amplitude was 356 μm (SD = 127 μm). None of the 144 measurements presented an error over 1 mm.

The accuracy of wear direction evaluation was highly correlated with wear amplitude (Spearman's rho = 0,98), the measurement of 3D wear direction presenting an accuracy better than 15° for wear amplitudes over 1,5 mm. Intra-observer errors for wear amplitude were between 138 μm and 221 μm depending on the observer. Inter-observer error for wear amplitude was 333 μm.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 390 - 390
1 Dec 2013
Lazennec JY Pour AE Brusson A Rousseau M Clarke I
Full Access

Introduction

Femoral stem anteversion after total hip arthroplasty (THA) has always been assessed using CT scan in supine position. In this study, we evaluated the anteversion of the femoral prosthesis neck in functional standing position using EOS® technology with repeatability and reproducibility of the measurements. The data obtained were compared with conventional anatomic measurements.

Materials and Methods

We measured the anteversion of the femoral prosthesis neck in 45 consecutive patients who had THA performed in nine hospitals. All measurements were obtained using the EOS® imaging system with patients in comfortable standing position. The orientation of the final vector representing the femoral neck was measured on 3-dimensional reconstructions. The anatomic femoral anteversion was calculated as in a transverse plane relative to the scanner and to the plane of the reconstructed bicondylar femoral segment (femoral prosthesis neck against the femoral condyles). Functional femoral anteversion (FFA) was measured in the horizontal plane relative to the frontal plane of the patient through the center of two femoral heads. FFA embodies true anteversion of the femoral prosthesis neck relative to the pelvis, representing the combined lower extremity anteversion.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 273 - 273
1 Sep 2012
Lazennec JY Rakover JP Aaron A Pascal Mousselard H Rousseau MA
Full Access

Introduction

Current artificial discs include 1 or 2 bearing surfaces, providing 3 or 5 degrees of freedom. The ESP® is a one-piece e implant made of silicon and polycarbonate -urethane securely fixed to titanium endplates. It allows limited rotation and translation with elastic return. This cushion without fixed rotation center achieves 6 degrees of freedom including shock absorption. This objective of this study was to evaluate the safety and efficacy of the concept in a prospective nonrandomized trial.

Material and methods

Prior to clinical implantations, the device was endurance tested at least 40 millions cycles. The polymer core weight and properties, the geometrical characteristics and cohesion of the implants remained stable.

A prospective trial was initiated in 2004 for L3L4, L4L5 and L5S1 levels. Total disc replacements have been performed in 153 lumbar levels through extra-peritoneal mini-invasive anterior approach with a minimum 2 years follow-up.