1 Apr 2009
Ornstein E Linder L Ranstam J Lewold S Eisler T Torper M

We identified 1305 femoral impaction bone grafting revisions using the Exeter stem performed between 1989 and 2002 in 30 hospitals throughout Sweden. There were 1188 patients with a mean age of 71 years (29 to 94) followed up for between five and 18 years.

The participating departments reported 70 further revisions in total, of which 57 could also be identified on the Swedish National Arthroplasty Registry.

Kaplan-Meier survivorship for all causes of failure was 94.0% (95% confidence interval (CI) 92 to 96) for women and 94.7% (95% CI, 92 to 96) for men at 15 years. Survivorship at 15 years for aseptic loosening was 99.1% (95% CI 98.4 to 99.5), for infection 98.6% (95% CI 97.6 to 99.2), for subsidence 99.0% (95% CI 98.2 to 99.4) and for fracture 98.7% (95% CI 97.9 to 99.2)

Statistically significant predictors of failure were the year in which revision was conducted (p < 0.001). The number of previous revisions was slightly above the level of signifance (p = 0.056). Age, gender, the length of the stem and previous septic loosening were not predictors of failure (p = 0.213, p = 0.399, p = 0.337, p = 0.687, respectively). The difference in survivorship between high- and low-volume departments was only 3% at ten years.

We conclude that impaction bone grafting with the Exeter stem has an excellent long-term survivorship following revision arthroplasty. The technique of impaction grafting appears to be reliable, can be learned rapidly and produces a predictably low incidence of aseptic loosening.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 12 | Pages 1585 - 1593
1 Dec 2008
Henricson A Linder L Nilsson KG

We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component.

All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting migration comprising subsidence and lift-off from the tibial tray, most of the trabecular metal components showed subsidence only, probably due to the elasticity of the implant.

This pattern of subsidence is regarded as being beneficial for uncemented fixation.


Orthopaedic Proceedings
Vol. 88-B, Issue SUPP_II | Pages 211 - 211
1 May 2006
Rydholm U Andersson T Linder L Maxander P Besjakov J Montgomery F Carlsson A
Full Access

25 RA patients with their ankles fused with an intramedullary nail were compared to 35 RA patients with their ankles fused with compression screws.

24/25 nailed patients showed radiographic healing at follow-up after 3 (1–8) years, and 26/35 ankles in the compression screw group examined after 6 (1–14) years healed after the first attempt and another 5 after repeat surgery.

In the nailed group 23 patients were satisfied and 2 somewhat satisfied. In the compression screw group 20 were satisfied, 12 somewhat satisfied and 3 dissatisfied.

There were 4 deep infections (3 healed after nail extraction and antibiotics, one unhealed) in the nail group and 1 deep infection (healed after antibiotics) in the compression screw group.

Six patients in the nailed group also had a permanent plantar sensory loss.

Conclusion: Ankle fusion with retrograde intramedullary nailing seems to result in a high rate of healing and satisfied patients, but cares a substantial risk of deep infection compared to fusion with compression screws, which has a lower fusion rate fewer satisfied patients, but less complications.

RA patients with a normal subtalar joint are of course only managed by compression screws.


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 5 | Pages 693 - 696
1 Sep 1993
Ling R Timperley A Linder L

We report the histology of a femur retrieved 3.5 years after a cemented revision of a hip replacement in which impaction allografting had been used to fill two large cortical defects. The allograft chips had largely been replaced by viable cortical bone, and the interface between cement and tissue resembled that seen after primary cemented arthroplasty.


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 1 | Pages 14 - 21
1 Jan 1993
Gie G Linder L Ling R Simon J Slooff T Timperley A

We report the results of using impacted cancellous allografts and cement for fixation of the femoral component when revision arthroplasty is required in the face of lost bone stock. In 56 hips reviewed after 18 to 49 months there were few complications and a majority of satisfactory results with evidence of incorporation of the graft. Further study and review are necessary, but the use of the method appears to be justified.


The Journal of Bone & Joint Surgery British Volume
Vol. 70-B, Issue 4 | Pages 550 - 555
1 Aug 1988
Linder L Carlsson A Marsal L Bjursten L Branemark P

In an experimental clinical study, 25 implants of pure titanium were inserted into the proximal tibia of 11 volunteer patients, four with rheumatoid arthritis and seven with osteoarthritis. The implants were removed from five weeks to 24 months later and detailed histological analysis was performed. The implants generally healed with direct bone-metal contact, showing so-called osseointegration. Only one of the 21 implants which had been in place for over five months did not show osseointegration, probably because of inadequate primary contact with bone. The presence of rheumatoid disease did not prevent osseointegration, but accompanying osteoporosis seemed to be a risk factor.


The Journal of Bone & Joint Surgery British Volume
Vol. 65-B, Issue 5 | Pages 646 - 649
1 Nov 1983
Linder L Hansson H

An extremely strong mechanical bond between cement and bone was observed in three patients with cemented hip protheses who underwent revision operations. The nature of the bone--cement attachment was studied by electron microscopy. The tissue at the interface was found to be made up of viable bone alternating with areas of soft tissue containing macrophages. The important qualitative differences between this reaction and the reaction seen around inert materials such as titanium are discussed.