Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_10 | Pages 5 - 5
1 Aug 2021
Thompson J Corbett J Bye D Jones A Tissingh E Nolan J
Full Access

The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems.

A systematic literature search was performed adhering to PRISMA criteria. In parallel we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003– June 2020.

There are 25 reported cases of such prosthetic fractures confined to small case series and case reports within the literature. We report an additional 19 cases to the literature (mean age 66.3 ± 11.7 years; 12 female [63%]; body mass index 32.9 ± 5.9 kg/m2). The mean time from index procedure to fracture was 7.8 years (2.5–16.3, ±3.6). Exeter V40 stem fracture incidence was 0.27%. Incidence was significantly higher in 125 mm length stems compared to ≥150 mm length stems (1.26% vs 0.13%, respectively, p <0.001) and revision arthroplasty (1.209% vs 0.149%, p <0.001). When comparing different stem length cohorts, 125-mm short-stem were associated with stem body fractures (92% vs 29%, p = 0.0095), earlier time to fracture (6.2 vs 11.0 years, p = 0.0018), younger patient age at time of fracture (62.7 vs 72.6 years old, p = 0.037) and female sex (75% vs 43%, p = 0.33).

This case series in conjunction with the systematic review provides evidence stem morphology plays a role in femoral implant fracture. This complication remains rare, although we report a significantly higher incidence at up to 17 years follow-up than in the literature. As femoral geometries remain the same, increasing BMIs in THR patients should raise concern. Short 125 mm length Exeter V40 stems undoubtedly have a role in restoring anatomy and biomechanics in smaller femoral geometries, although the surgeon has to appreciate the higher risk of stem fracture and the associated predisposing factors which may necessitate meticulous surgical technique and planning.


Bone & Joint Open
Vol. 2, Issue 6 | Pages 443 - 456
28 Jun 2021
Thompson JW Corbett J Bye D Jones A Tissingh EK Nolan J

Aims

The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems.

Methods

A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 6 - 6
1 Jun 2016
Ebreo D Felgate H Martinez-Estrada F Donell S Nolan J Clark I
Full Access

Introduction

The biological pathways responsible for adverse reactions to metal debris (ARMD) are unknown. Necrotic and inflammatory changes in response to Co-Cr nanoparticles in periprosthetic tissues may involve both a cytotoxic response and a type IV delayed hypersensitivity response.

Our aim was to establish whether differences in biological cascade activation exists in tissues of patients with end-stage OA compared to those with aseptic loosening of a metal on polyethylene (MoP) THR and those with ARMD from metal-on-metal (MoM) THR.

Patients & Methods

A microarray experiment (Illumina HT12-v4) was performed to identify the range of differential gene expression between 24 patients across 3 phenotypes: Primary OA (n=8), revision for aseptic loosening of MoP THR (n=8) and ARMD associated with MoM THR (n=8).

Results were validated using Taqman Low Density Array (TLDA) selecting the top 36 genes in terms of fold-change (FC)>2 and a significant difference (p<0.05) on ANOVA.

Pathways of cellular interaction were explored using Ingenuity IPA software.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 8 | Pages 1025 - 1030
1 Aug 2009
Ollivere B Darrah C Barker T Nolan J Porteous MJ

The rate and mode of early failure in 463 Birmingham hip resurfacings in a two-centre, multisurgeon series were examined. Of the 463 patients two have died and three were lost to follow-up. The mean radiological and clinical follow-up was for 43 months (6 to 90).

We have revised 13 resurfacings (2.8%) including seven for pain, three for fracture, two for dislocation and another for sepsis. Of these, nine had macroscopic and histological evidence of metallosis. The survival at five years was 95.8% (95% confidence interval (CI) 94.1 to 96.8) for revision for all causes and 96.9% (95% CI 95.5 to 98.3) for metallosis.

The rate of metallosis related revision was 3.1% at five years. Risk factors for metallosis were female gender, a small femoral component, a high abduction angle and obesity. We do not advocate the use of the Birmingham Hip resurfacing procedure in patients with these risk factors.