Advertisement for orthosearch.org.uk
Results 1 - 6 of 6
Results per page:
Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 140 - 140
1 May 2016
Pritchett J
Full Access

Background

Cross-linked polyethylene has much less wear than conventional polyethylene and can used in a more bone conserving thickness of 4 mm. We have used it for hip resurfacing since 2001.

Questions/Purposes

This study evaluated the effectiveness of a highly cross-linked polyethylene acetabular component for hip resurfacing in patients under age 50. I posed 5 questions: (1) What are the functional results, (2) What are the complications, (3) What is the 10-year implant survivorship, (4) What is the femoral head penetration into the polyethylene, and (5) What is the bone conservation?


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 139 - 139
1 May 2016
Pritchett J
Full Access

BACKGROUND

We originally performed metal-on-metal hip resurfacing using a Townley designed Vitallium Total Articular Replacement Arthroplasty (TARA) curved stemmed prosthesis. Neither the acetabular or femoral components were cemented or had porous coating. The bearing surfaces were consistently polar bearing. The surgical objectives were to preserve bone stock, maintain normal anatomy and mechanics of the hip joint and to approximate the normal stress transmission to the supporting femoral bone. The functional objectives were better sports participation, less thigh pain and limp, less perception of a leg length difference and a greater perception of a normal hip. Metal-on-metal was selected to conserve acetabular bone and avoid polyethylene associated osteolysis.

Relatively few cases were performed until the Conserve Plus and later the Birmingham Hip Resurfacing systems became available.

METHODS

We examined the results of metal-on-metal hip resurfacing in patient with at least 10 years of follow-up and an age less than 50 at the time of surgery. We did not have access to the Birmingham Prosthesis until 2006. We performed 101 TARA procedures and 397 Conserve Plus procedures for 357 patients. For the combined series the mean age was 43 and 62% of patients were male. 34 patients had a conventional total hip replacement on the contralateral side. We used both the anterolateral and posterior approaches. All acetabular components were placed without cement and all the Conserve Plus Femoral Components were cemented.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 138 - 138
1 May 2016
Pritchett J
Full Access

BACKGROUND

The most common salvage of a failed metal-on-metal hip resurfacing is to remove both the femoral and acetabular resurfacing components and perform a total hip replacement. The other choices are to perform an acetabular or femoral only revision. A one or two piece acetabular component or a polyethylene bipolar femoral component that matches the retained metal resurfacing acetabular component is used. The considerations in favor of performing a one component resurfacing revision are maintaining the natural femoral head size, limiting the surgical effort for the patient and surgeon, and bone conservation. There are often favorable cost considerations with single component revision surgery.

The reasons for femoral component revision are femoral neck fracture, femoral component loosening and an adverse reaction to metal wear debris. Performing a femoral component only revision requires a well fixed and well oriented acetabular component. Acetabular revision is most often performed for an adverse reaction to metal wear debris or loosening.

METHODS

81 acetabular revisions and 46 femoral revisions were evaluated 4 to 14 years after surgery. 83% of patients had their initial surgery at outside institutions. The mean age was 46 and 65% of patients were women. A two piece titanium backed polyethylene component was used in 44 patients and a one or two piece metal component was used in 37. A dual mobility femoral prosthesis mated to a retained metal acetabular component was used for the femoral revisions and no conversions to a metal-on-metal total hip replacement were performed. We selected polyethylene acetabular components for patients with adverse reactions to metal wear debris if their femoral component was less than 48 mm or if there was no matching metal acetabular component available for their femoral component. We used dual mobility components for femoral loosening, femoral neck fractures and adverse reactions to metal wear debris in patients with well-fixed and well oriented metal acetabular components. Dual mobility components were also used if there are any concerns about the femoral component or in some older patients. We performed one component revisions rather than conversion to total hip replacement on 88% of patients presenting with failed resurfacing prostheses.


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 2 | Pages 267 - 269
1 Mar 1995
Pritchett J

Fifty patients with complex distal radial fractures treated by primary external fixation were compared with 50 with similar fractures treated by closed medullary pinning. All the patients had Frykman type-VIII injuries. The two groups were similar in regard to demographic characteristics and the method of treatment was randomly chosen. All the fractures healed within three months. In the external fixation group 92% of fractures healed in excellent alignment as did 88% of the medullary pinning group. Both groups had similar results with respect to eventual function, range of motion, and grip strength. Complications and complaints were fewer and the estimated costs of treatment were significantly less in the medullary pinning group. More patients were satisfied with closed medullary fixation than with external fixation.


The Journal of Bone & Joint Surgery British Volume
Vol. 68-B, Issue 4 | Pages 561 - 565
1 Aug 1986
Pritchett J

Deformity of the forearm is common in patients with hereditary multiple exostoses, producing cosmetic and functional impairment in which shortening of the ulna is a significant factor. The results of ulnar lengthening in 10 forearms of eight patients are reported. Lengthening was performed by osteotomy of the shaft followed immediately by a bone graft and internal fixation, or by gradual distraction with an external fixator. In all patients the appearance was improved and the range of radial deviation at the wrist was increased. In most patients forearm movement and radial head stability were improved. Partial recurrence of the deformity was seen during the follow-up of skeletally immature patients, but in general ulnar lengthening was found to be a useful operation.


The Journal of Bone & Joint Surgery British Volume
Vol. 67-B, Issue 5 | Pages 715 - 718
1 Nov 1985
Pritchett J

Ten patients with humeral shaft fractures and no clinical or radiographic signs of healing after at least six weeks' immobilisation were treated by flexible intramedullary nailing using a closed retrograde technique. Bone grafting was not performed, and active movement was encouraged after operation. Nine fractures healed; the mean time to union was 10.5 weeks (range 6 to 22 weeks). One patient needed compression plating and bone grafting at 22 weeks, and another required re-operation for distal migration of the rods. There were no infections, nerve palsies or other complications. Stiffness of the shoulder which had developed during early treatment improved after operation.