Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 117 - 117
1 Feb 2020
Wankier Z Sinclair S Drew A Taylor C Kubiak E Agarwal J
Full Access

Introduction

Direct skeletal attachment of prosthetic limbs, commonly known as osseointegration (“OI”), is being investigated by our team with the goal of safely introducing this technology into the United States for human use. OI technology allows for anchorage of prosthetic devices directly to bone using an intramedullary stem. For OI to be effective and secure, bone ingrowth and remodeling around the implant must be achieved. Physicians need an effective way to measure bone remodeling in order to make informed decisions on prescribed loading. This work describes methodology that was developed that utilizes computed tomography (CT) imaging as a tool for analyzing bone remodeling around an osseointegrated implant.

Method

A subject implanted with a new Percutaneous Osseointegrated Prosthesis (POP) (DJO Surgical, Austin, TX) had CTs taken of their residual femur at 6-weeks and 12-months post-op in a FDA Early Feasibility Study with Institutional Review Board approval. Three-dimensional models of the femur were created from dicom files of the CT slices using Mimics (v21.0, Materialise, Leuven, Belgium). Each scan was segmented into four objects: cortical bone, medullary cavity, total volume (cortical bone plus the medullary cavity) and endoprosthetic stem (Fig. 1).

Following segmentation, models were uploaded to 3-Matic Research (v13.0, Materialise, Leuven, Blegium) in STL format for alignment to a common world coordinate system (Fig. 2). A common origin was set by taking the average distance between planes of the femoral head and the greater trochanter. Once aligned to the coordinate system, biomechanical length (BML) was calculated from the proximal origin to the distal end of the amputated femur.

BML and STLs of the aligned medullary cavity and femur volume were entered into custom Matlab code designed to measure cortical and medullary morphology in transverse cross sections of the femur. Morphology data from 6-weeks and 12-month time points were compared in order to determine if bone remodeling around the POP implant could be detected using these methods.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 178 - 178
1 Sep 2012
Sinclair S Konz G Dawson J Bloebaum R
Full Access

Background

Synthetic interbody spinal fusion devices are used to restore and maintain disc height and ensure proper vertebral alignment. These devices are often filled with autograft bone to facilitate bone bridging through the device while providing mechanical stability. Nonporous polyetheretherketone (PEEK) devices are widely used clinically for such procedures.1Trabecular Metal devices are an alternative, fabricated from porous tantalum. It was hypothesized that the porous Trabecular Metal device would better maintain autograft viability through the center of the device, the ‘graft hole’ (GH).

Methods

Twenty-five goats underwent anterior cervical discectomy and fusion using a Trabecular Metal or PEEK device for 6, 12 or 26 weeks. The GH of each device was filled with autograft bone morsels harvested from the animal at implantation. Fluorochrome labeling oxytetracycline was administered to the animals and used to determine bone viability in the device regions. Following necropsy, the vertebral segments were embedded in poly(methyl methacrylate) sectioned and analyzed using fluorescence and backscatter electron (BSE) imaging. The percent of bone tissue present within the GH was measured as a volume percent using BSE images (Fig. 1).