Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_9 | Pages 30 - 30
1 May 2018
Spiegelberg B Lanting B Howard J Teeter M Naudie D
Full Access

Background

There has been a trend in the evolution of total hip arthroplasty towards increased modularity, with this increase in modularity come some potentially harmful consequences. Modularity at the neck shaft junction has been linked to corrosion, adverse reaction to metal debris and pseudotumor formation.

The aim of this retrieval study is to assess whether the surface integrity of the polyethylene (PE) liner is affected by metal wear debris in a single implant design series of THA revised for trunnionosis.

Method

A retrieval analysis of thirty dual-taper modular neck hip prostheses was performed, the mean time from implantation to revision was 2.7 years (1.02–6.2). The PE liners were analysed using a scanning electron microscope with an energy dispersive spectrometer to assess for metal particles embedded on the liner surface. Serum metal ion levels and inflammatory markers were also analysed.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 49 - 49
1 Jun 2016
Spiegelberg B Kendrick B Taylor A Stubbs D McLardy-Smith P Gundle R
Full Access

Introduction

Native hip joint infection can result in hip arthrosis as a complication requiring the need for subsequent arthroplasty. There is little evidence to support single or staged THA. We present the results of patients who have undergone total hip arthroplasty (THA) following hip joint sepsis in our institution.

Methods

Patients receiving a THA following previous hip joint sepsis between 2003 and 2015 were identified from the operative records database. Further clinical information was collected from the medical records, including timing of infection and surgery, as well as culture results. Functional outcome (Oxford Hip Score, OHS) and patient satisfaction scores (PSS) were recorded.