Distraction osteogenesis with intramedullary lengthening devices has undergone rapid development in the past decade with implant enhancement. In this first single-centre matched-pair analysis we focus on the comparison of treatment with the PRECICE and STRYDE intramedullary lengthening devices and aim to clarify any clinical and radiological differences. A single-centre 2:1 matched-pair retrospective analysis of 42 patients treated with the STRYDE and 82 patients treated with the PRECICE nail between May 2013 and November 2020 was conducted. Clinical and lengthening parameters were compared while focusing radiological assessment on osseous alterations related to the nail’s telescopic junction and locking bolts at four different stages.Aims
Methods
Limb-lengthening nails have largely replaced external fixation in limb-lengthening and reconstructive surgery. However, the adverse events and high prevalence of radiological changes recently noted with the STRYDE lengthening nail have raised concerns about the use of internal lengthening nails. The aim of this study was to compare the prevalence of radiological bone abnormalities between STRYDE, PRECICE, and FITBONE nails prior to nail removal. This was a retrospective case series from three centres. Patients were included if they had either of the three limb-lengthening nails (STYDE, PRECICE, or FITBONE) removed. Standard orthogonal radiographs immediately prior to nail removal were examined for bone abnormalities at the junction of the telescoping nail parts.Aims
Methods
The Precice intramedullary limb-lengthening system has demonstrated significant benefits over external fixation lengthening methods, leading to a paradigm shift in limb lengthening. This study compares outcomes following antegrade and retrograde femoral lengthening in both adolescent and adult patients. A retrospective review of prospectively collected data was undertaken of a consecutive series of 107 femoral lengthening operations in 92 patients. In total, 73 antegrade nails and 34 retrograde nails were inserted. Outcome was assessed by the regenerate healing index (HI), hip and knee range of movement (ROM), and the presence of any complications.Aims
Patients and Methods
Aims. Patients undergoing femoral lengthening by external fixation
tolerate treatment less well when compared to tibial lengthening.
Lengthening of the femur with an intramedullary device may have
advantages. Patients and Methods. We reviewed all cases of simple femoral lengthening performed
at our unit from 2009 to 2014. Cases of nonunions, concurrent deformities,
congenital limb deficiencies and lengthening with an unstable hip
were excluded, leaving 33 cases (in 22 patients; 11 patients had
bilateral procedures) for review. Healing index, implant tolerance
and complications were compared. Results. In 20 cases (15 patients) the