Nonagenarians (aged 90 to 99 years) have experienced the fastest percent decile population growth in the USA recently, with a consequent increase in the prevalence of nonagenarians living with joint arthroplasties. As such, the number of revision total hip arthroplasties (THAs) and total knee arthroplasties (TKAs) in nonagenarians is expected to increase. We aimed to determine the mortality rate, implant survivorship, and complications of nonagenarians undergoing aseptic revision THAs and revision TKAs. Our institutional total joint registry was used to identify 96 nonagenarians who underwent 97 aseptic revisions (78 hips and 19 knees) between 1997 and 2018. The most common indications were aseptic loosening and periprosthetic fracture for both revision THAs and revision TKAs. Mean age at revision was 92 years (90 to 98), mean BMI was 27 kg/m2 (16 to 47), and 67% (n = 65) were female. Mean time between primary and revision was 18 years (SD 9). Kaplan-Meier survival was used for patient mortality, and compared to age- and sex-matched control populations. Reoperation risk was assessed using cumulative incidence with death as a competing risk. Mean follow-up was five years.Aims
Methods
Knowledge on total knee arthroplasties (TKAs) in patients with a history of poliomyelitis is limited. This study compared implant survivorship and clinical outcomes among affected and unaffected limbs in patients with sequelae of poliomyelitis undergoing TKAs. A retrospective review of our total joint registry identified 94 patients with post-polio syndrome undergoing 116 primary TKAs between January 2000 and December 2019. The mean age was 70 years (33 to 86) with 56% males (n = 65) and a mean BMI of 31 kg/m2 (18 to 49). Rotating hinge TKAs were used in 14 of 63 affected limbs (22%), but not in any of the 53 unaffected limbs. Kaplan-Meier survivorship analyses were completed. The mean follow-up was eight years (2 to 19).Aims
Methods
This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant. A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life.Aims
Methods
Aims.
Metaphyseal cones with cemented stems are frequently used in revision total knee arthroplasty (TKA). However, if the diaphysis has been previously violated, the resultant sclerotic canal can impair cemented stem fixation, which is vital for bone ingrowth into the cone, and long-term fixation. We report the outcomes of our solution to this problem, in which impaction grafting and a cemented stem in the diaphysis is combined with an uncemented metaphyseal cone, for revision TKA in patients with severely compromised bone. A metaphyseal cone was combined with diaphyseal impaction grafting and cemented stems for 35 revision TKAs. There were two patients with follow-up of less than two years who were excluded, leaving 33 procedures in 32 patients in the study. The mean age of the patients at the time of revision TKA was 67 years (32 to 87); 20 (60%) were male. Patients had undergone a mean of four (1 to 13) previous knee arthroplasty procedures. The indications for revision were aseptic loosening (80%) and two-stage reimplantation for prosthetic joint infection (PJI; 20%). The mean follow-up was four years (2 to 11).Aims
Methods
Metaphyseal fixation during revision total knee arthroplasty (TKA) is important, but potentially difficult when using historical designs of cone. Material and manufacturing innovations have improved the size and shape of the cones which are available, and simplified the required bone preparation. In a large series, we assessed the implant survivorship, radiological results, and clinical outcomes of new porous 3D-printed titanium metaphyseal cones featuring a reamer-based system. We reviewed 142 revision TKAs in 139 patients using 202 cones (134 tibial, 68 femoral) which were undertaken between 2015 and 2016. A total of 60 involved tibial and femoral cones. Most cones (149 of 202; 74%) were used for Type 2B or 3 bone loss. The mean age of the patients was 66 years (44 to 88), and 76 (55 %) were female. The mean body mass index (BMI) was 34 kg/m2 (18 to 60). The patients had a mean of 2.4 (1 to 8) previous operations on the knee, and 68 (48%) had a history of prosthetic infection. The mean follow-up was 2.4 years (2 to 3.6).Aims
Methods
Varus-valgus constrained (VVC) implants are often used during revision total knee arthroplasty (TKA) to gain coronal plane stability. However, the increased mechanical torque applied to the bone-cement interface theoretically increases the risk of aseptic loosening. We assessed mid-term survivorship, complications, and clinical outcomes of a fixed-bearing VVC device in revision TKAs. A total of 416 consecutive revision TKAs (398 patients) were performed at our institution using a single fixed-bearing VVC TKA from 2007 to 2015. Mean age was 64 years (33 to 88) with 50% male (199). Index revision TKA diagnoses were: instability (n = 122, 29%), aseptic loosening (n = 105, 25%), and prosthetic joint infection (PJI) (n = 97, 23%). All devices were cemented on the epiphyseal surfaces. Femoral stems were used in 97% (n = 402) of cases, tibial stems in 95% (n = 394) of cases; all were cemented. In total, 93% (n = 389) of cases required a stemmed femoral and tibial component. Femoral cones were used in 29%, and tibial cones in 40%. Survivorship was assessed via competing risk analysis; clinical outcomes were determined using Knee Society Scores (KSSs) and range of movement (ROM). Mean follow-up was four years (2 to 10).Aims
Methods
We aimed to examine the long-term mechanical survivorship, describe the modes of all-cause failure, and identify risk factors for mechanical failure of all-polyethylene tibial components in endoprosthetic reconstruction. This is a retrospective database review of consecutive endoprosthetic reconstructions performed for oncological indications between 1980 and 2019. Patients with all-polyethylene tibial components were isolated and analyzed for revision for mechanical failure. Outcomes included survival of the all-polyethylene tibial component, revision surgery categorized according to the Henderson Failure Mode Classification, and complications and functional outcome, as assessed by the Musculoskeletal Tumor Society (MSTS) score at the final follow-up.Aims
Methods
There is little literature about total knee arthroplasty (TKA) after distal femoral osteotomy (DFO). Consequently, the purpose of this study was to analyze the outcomes of TKA after DFO, with particular emphasis on: survivorship free from aseptic loosening, revision, or any re-operation; complications; radiological results; and clinical outcome. We retrospectively reviewed 29 patients (17 women, 12 men) from our total joint registry who had undergone 31 cemented TKAs after a DFO between 2000 and 2012. Their mean age at TKA was 51 years (22 to 76) and their mean body mass index 32 kg/m2 (20 to 45). The mean time between DFO and TKA was ten years (2 to 20). The mean follow-up from TKA was ten years (2 to 16). The prostheses were posterior-stabilized in 77%, varus-valgus constraint (VVC) in 13%, and cruciate-retaining in 10%. While no patient had metaphyseal fixation (e.g. cones or sleeves), 16% needed a femoral stem.Aims
Patients and Methods
The aim of this study was to analyse the gait
pattern, muscle force and functional outcome of patients who had undergone
replacement of the proximal tibia for tumour and alloplastic reconstruction
of the extensor mechanism using the patellar-loop technique. Between February 1998 and December 2009, we carried out wide
local excision of a primary sarcoma of the proximal tibia, proximal
tibial replacement and reconstruction of the extensor mechanism
using the patellar-loop technique in 18 patients. Of these, nine
were available for evaluation after a mean of 11.6 years (0.5 to
21.6). The strength of the knee extensors was measured using an
Isobex machine and gait analysis was undertaken in our gait assessment
laboratory. Functional outcome was assessed using the American Knee
Society (AKS) and Musculoskeletal Tumor Society (MSTS) scores. The gait pattern of the patients differed in ground contact time,
flexion heel strike, maximal flexion loading response and total
sagittal plane excursion. The mean maximum active flexion was 91°
(30° to 110°). The overall mean extensor lag was 1° (0° to 5°).
The mean extensor muscle strength was 25.8% (8.3% to 90.3%) of that
in the non-operated leg (p <
0.001). The mean functional scores
were 68.7% (43.4% to 83.3%) (MSTS) and 71.1 (30 to 90) (AKS functional
score). In summary, the results show that reconstruction of the extensor
mechanism using this technique gives good biomechanical and functional
results. The patients’ gait pattern is close to normal, except for
a somewhat stiff knee gait pattern. The strength of the extensor
mechanism is reduced, but sufficient for walking. Cite this article:
Instability after total knee replacement (TKR)
accounts for 10% to 22% of revision procedures. All patients who
present for evaluation of instability require a thorough history to
be taken and physical examination, as well as appropriate imaging.
Deep periprosthetic infection must be ruled out by laboratory testing
and an aspiration of the knee must be carried out. The three main
categories of instability include flexion instability, extension instability
(symmetric and asymmetric), and genu recurvatum. Most recently,
the aetiologies contributing to, and surgical manoeuvres required
to correct, flexion instability have been elucidated. While implant
design and patient-related factors may certainly contribute to the
aetiology, surgical technique is also a significant factor in all
forms of post-operative instability. Cite this article:
The use of hinged implants in primary total knee
replacement (TKR) should be restricted to selected indications and mainly
for elderly patients. Potential indications for a rotating hinge
or pure hinge implant in primary TKR include: collateral ligament
insufficiency, severe varus or valgus deformity (>
20°) with necessary
relevant soft-tissue release, relevant bone loss including insertions
of collateral ligaments, gross flexion-extension gap imbalance,
ankylosis, or hyperlaxity. Although data reported in the literature
are inconsistent, clinical results depend on implant design, proper
technical use, and adequate indications. We present our experience
with a specific implant type that we have used for over 30 years
and which has given our elderly patients good mid-term results.
Because revision of implants with long cemented stems can be very
challenging, an effort should be made in the future to use shorter stems
in modular versions of hinged implants. Cite this article:
We evaluated the clinical results and complications
after extra-articular resection of the distal femur and/or proximal
tibia and reconstruction with a tumour endoprosthesis (MUTARS) in
59 patients (mean age 33 years (11 to 74)) with malignant bone or
soft-tissue tumours. According to a Kaplan–Meier analysis, limb
survival was 76% (95% confidence interval (CI) 64.1 to 88.5) after
a mean follow-up of 4.7 years (one month to 17 years). Peri-prosthetic infection
was the most common indication for subsequent amputation (eight
patients). Survival of the prosthesis without revision was 48% (95%
CI 34.8 to 62.0) at two years and 25% (95% CI 11.1 to 39.9) at five years
post-operatively. Failure of the prosthesis was due to deep infection
in 22 patients (37%), aseptic loosening in ten patients (17%), and
peri-prosthetic fracture in six patients (10%). Wear of the bearings
made a minor revision necessary in 12 patients (20%). The mean Musculoskeletal
Tumor Society score was 23 (10 to 29). An extensor lag >
10° was
noted in ten patients (17%). These results suggest that limb salvage after extra-articular
resection with a tumour prosthesis can achieve good functional results
in most patients, although the rates of complications and subsequent
amputation are higher than in patients treated with intra-articular
resection. Cite this article:
We report the results of revision total knee
replacement (TKR) in 26 patients with major metaphyseal osteolytic defects
using 29 trabecular metal cones in conjunction with a rotating hinged
total knee prosthesis. The osteolytic defects were types II and
III (A or B) according to the Anderson Orthopaedic Research Institute
(AORI) classification. The mean age of the patients was 72 years
(62 to 84) and there were 15 men and 11 women. In this series patients had
undergone a mean of 2.34 previous total knee arthroplasties. The
main objective was to restore anatomy along with stability and function
of the knee joint to allow immediate full weight-bearing and active
knee movement. Outcomes were measured using Knee Society scores,
Oxford knee scores, range of movement of the knee and serial radiographs.
Patients were followed for a mean of 36 months (24 to 49). The mean
Oxford knee clinical scores improved from 12.83 (10 to 15) to 35.20
(32 to 38) (p <
0.001) and mean American Knee Society scores
improved from 33.24 (13 to 36) to 81.12 (78 to 86) (p <
0.001).
No radiolucent lines suggestive of loosening were seen around the trabecular
metal cones, and by one year all the radiographs showed good osteo-integration.
There was no evidence of any collapse or implant migration. Our
early results confirm the findings of others that trabecular metal
cones offer a useful way of managing severe bone loss in revision
TKR. Cite this article:
Patients with skeletal dysplasia are prone to
developing advanced osteoarthritis of the knee requiring total knee replacement
(TKR) at a younger age than the general population. TKR in this
unique group of patients is a technically demanding procedure owing
to the deformity, flexion contracture, generalised hypotonia and ligamentous
laxity. We retrospectively reviewed the outcome of 11 TKRs performed
in eight patients with skeletal dysplasia at our institution using
the Stanmore Modular Individualised Lower Extremity System (SMILES)
custom-made
We undertook a retrospective review of 33 patients who underwent total femoral endoprosthetic replacement as limb salvage following excision of a malignant bone tumour. In 22 patients this was performed as a primary procedure following total femoral resection for malignant disease. Revision to a total femoral replacement was required in 11 patients following failed segmental endoprosthetic or allograft reconstruction. There were 33 patients with primary malignant tumours, and three had metastatic lesions. The mean age of the patients was 31 years (5 to 68). The mean follow-up was 4.2 years (9 months to 16.4 years). At five years the survival of the implants was 100%, with removal as the endpoint and 56% where the endpoint was another surgical intervention. At five years the patient survival was 32%. Complications included dislocation of the hip in six patients (18%), local recurrence in three (9%), peri-prosthetic fracture in two and infection in one. One patient subsequently developed pulmonary metastases. There were no cases of aseptic loosening or amputation. Four patients required a change of bushings. The mean Musculoskeletal Tumour Society functional outcome score was 67%, the mean Harris Hip Score was 70, and the mean Oxford Knee Score was 34. Total femoral endoprosthetic replacement can provide good functional outcome without compromising patient survival, and in selected cases provides an effective alternative to amputation.
We reviewed 25 patients who had undergone resection of a primary bone sarcoma which extended to within 5 cm of the knee with reconstruction by a combination of a free vascularised fibular graft and a massive allograft bone shell. The distal femur was affected in four patients and the proximal tibia in 21. Their mean age at the time of operation was 19.7 years (5 to 52) and the mean follow-up period 140 months (28 to 213). Three vascularised transfers failed. The mean time to union of the fibula was 5.6 months (3 to 10) and of the allograft 19.6 months (10 to 34). Full weight-bearing was allowed at a mean of 21.4 months (14 to 36). The mean functional score at final follow-up was 27.4 (18 to 30) using a modfied 30-point Musculoskeletal Tumour Society rating system. The overall limb-salvage rate was 88%. The results of our study suggest that the combined use of a vascularised fibular graft and allograft is of value as a limb-salvage procedure for intercalary reconstruction after resection of bone tumours around the knee, especially in skeletally immature patients.
We describe the application of a non-invasive extendible endoprosthetic replacement in skeletally-mature patients undergoing revision for failed joint replacement with resultant limb-length inequality after malignant or non-malignant disease. This prosthesis was developed for tumour surgery in skeletally-immature patients but has now been adapted for use in revision procedures to reconstruct the joint or facilitate an arthrodesis, replace bony defects and allow limb length to be restored gradually in the post-operative period. We record the short-term results in nine patients who have had this procedure after multiple previous reconstructive operations. In six, the initial reconstruction had been performed with either allograft or endoprosthetic replacement for neoplastic disease and in three for non-neoplastic disease. The essential components of the prosthesis are a magnetic disc, a gearbox and a drive screw which allows painless lengthening of the prosthesis using the principle of electromagnetic induction. The mean age of the patients was 37 years (18 to 68) with a mean follow-up of 34 months (12 to 62). They had previously undergone a mean of six (2 to 14) open procedures on the affected limb before revision with the non-invasive extendible endoprosthesis. The mean length gained was 56 mm (19 to 107) requiring a mean of nine (3 to 20) lengthening episodes performed in the outpatient department. There was one case of recurrent infection after revision of a previously infected implant and one fracture of the prosthesis after a fall. No amputations were performed. Planned exchange of the prosthesis was required in three patients after attainment of the maximum lengthening capacity of the implant. There was no failure of the lengthening mechanism. The Mean Musculoskeletal Tumour Society rating score was 22 of 30 available points (18 to 28). The use of a non-invasive extendible endoprosthesis in this manner provided patients with good functional results and restoration of leg-length equality, without the need for multiple open lengthening procedures.
This paper describes the preliminary results of a proximal tibial endoprosthesis which spares the knee joint and enables retention of the natural articulation by replacing part of the tibial metaphysis and diaphysis. In eight patients who had a primary malignant bone tumour of the proximal tibia, the distal stem, which had a hydroxyapatite-coated collar to improve fixation, was cemented into the medullary canal. The proximal end had hydroxyapatite-coated extracortical plates which were secured to the remaining proximal tibial metaphysis using cortical screws. The mean age of the patients at operation was 28.9 years (8 to 43) and the mean follow-up was for 35 months (4 to 48). The mean Musculoskeletal Tumour Society score was 79% (57% to 90%), the mean Oxford Knee score was 40 points of 48 (36 to 46) and the mean knee flexion was 112° (100° to 120°). In one patient, revision to a below-knee amputation through the prosthesis was required because of recurrence of the tumour. Another patient sustained a periprosthetic fracture which healed with a painful malunion. This was revised to a further endoprosthesis which replaced the knee. In the remaining six patients the prosthesis allowed preservation of the knee joint with good function and no early evidence of loosening. Further follow-up is required to assess the longevity of these prostheses.
The management of patients with a painful total knee replacement requires careful assessment and a stepwise approach in order to diagnose the underlying pathology accurately. The management should include a multidisciplinary approach to the patient’s pain as well as addressing the underlying aetiology. Pain should be treated with appropriate analgesia, according to the analgesic ladder of the World Health Organisation. Special measures should be taken to identify and to treat any neuropathic pain. There are a number of intrinsic and extrinsic causes of a painful knee replacement which should be identified and treated early. Patients with unexplained pain and without any recognised pathology should be treated conservatively since they may improve over a period of time and rarely do so after a revision operation.