The objective of this study was to compare simulated range of motion (ROM) for reverse total shoulder arthroplasty (rTSA) with and without adjustment for scapulothoracic orientation in a global reference system. We hypothesized that values for simulated ROM in preoperative planning software with and without adjustment for scapulothoracic orientation would be significantly different. A statistical shape model of the entire humerus and scapula was fitted into ten shoulder CT scans randomly selected from 162 patients who underwent rTSA. Six shoulder surgeons independently planned a rTSA in each model using prototype development software with the ability to adjust for scapulothoracic orientation, the starting position of the humerus, as well as kinematic planes in a global reference system simulating previously described posture types A, B, and C. ROM with and without posture adjustment was calculated and compared in all movement planes.Aims
Methods
The aim of this study was to assess and compare active rotation of the forearm in normal subjects after the application of a short-arm cast (SAC) in the semisupination position and a long-arm cast (LAC) in the neutral position. A clinical study was also conducted to compare the functional outcomes of using a SAC in the semisupination position with those of using a LAC in the neutral position in patients who underwent arthroscopic triangular fibrocartilage complex (TFCC) foveal repair. A total of 40 healthy right-handed volunteers were recruited. Active pronation and supination of the forearm were measured in each subject using a goniometer. In the retrospective clinical study, 40 patients who underwent arthroscopic foveal repair were included. The wrist was immobilized postoperatively using a SAC in the semisupination position (approximately 45°) in 16 patients and a LAC in 24. Clinical outcomes were assessed using grip strength and patient-reported outcomes. The degree of disability caused by cast immobilization was also evaluated when the cast was removed.Aims
Methods
The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed.Aims
Materials and Methods
Aims
Patients and Methods
Scapular notching is a frequently observed radiographic phenomenon
in reverse shoulder arthroplasty (RSA), signifying impingement of
components. The purposes of this study were to evaluate the effect
of glenoid component size and glenosphere type on impingement-free
range of movement (ROM) for extension and internal and external
rotation in a virtual RSA model, and to determine the optimal configuration
to reduce the incidence of friction-type scapular notching. Preoperative CT scans obtained in 21 patients (three male, 18
female) with primary osteoarthritis were analyzed using modelling
software. Two concurrent factors were tested for impingement-free
ROM and translation of the centre of rotation: glenosphere diameter
(36 mm Aims
Materials and Methods
The objective of this study was to investigate bone healing after
internal fixation of displaced femoral neck fractures (FNFs) with
the Dynamic Locking Blade Plate (DLBP) in a young patient population
treated by various orthopaedic (trauma) surgeons. We present a multicentre prospective case series with a follow-up
of one year. All patients aged ≤ 60 years with a displaced FNF treated
with the DLBP between 1st August 2010 and December 2014 were included.
Patients with pathological fractures, concomitant fractures of the
lower limb, symptomatic arthritis, local infection or inflammation,
inadequate local tissue coverage, or any mental or neuromuscular
disorder were excluded. Primary outcome measure was failure in fracture
healing due to nonunion, avascular necrosis, or implant failure
requiring revision surgery.Aims
Patients and Methods
Computer hexapod assisted orthopaedic surgery (CHAOS), is a method
to achieve the intra-operative correction of long bone deformities
using a hexapod external fixator before definitive internal fixation
with minimally invasive stabilisation techniques. The aims of this study were to determine the reliability of this
method in a consecutive case series of patients undergoing femoral
deformity correction, with a minimum six-month follow-up, to assess
the complications and to define the ideal group of patients for
whom this treatment is appropriate. The medical records and radiographs of all patients who underwent
CHAOS for femoral deformity at our institution between 2005 and
2011 were retrospectively reviewed. Records were available for all
55 consecutive procedures undertaken in 49 patients with a mean
age of 35.6 years (10.9 to 75.3) at the time of surgery.Aims
Patients and Methods
The interaction between the lumbosacral spine
and the pelvis is dynamically related to positional change, and
may be complicated by co-existing pathology. This review summarises
the current literature examining the effect of sagittal spinal deformity
on pelvic and acetabular orientation during total hip arthroplasty
(THA) and provides recommendations to aid in placement of the acetabular
component for patients with co-existing spinal pathology or long
spinal fusions. Pre-operatively, patients can be divided into four
categories based on the flexibility and sagittal balance of the
spine. Using this information as a guide, placement of the acetabular
component can be optimal based on the type and significance of co-existing
spinal deformity. Cite this article:
The use of hinged implants in primary total knee
replacement (TKR) should be restricted to selected indications and mainly
for elderly patients. Potential indications for a rotating hinge
or pure hinge implant in primary TKR include: collateral ligament
insufficiency, severe varus or valgus deformity (>
20°) with necessary
relevant soft-tissue release, relevant bone loss including insertions
of collateral ligaments, gross flexion-extension gap imbalance,
ankylosis, or hyperlaxity. Although data reported in the literature
are inconsistent, clinical results depend on implant design, proper
technical use, and adequate indications. We present our experience
with a specific implant type that we have used for over 30 years
and which has given our elderly patients good mid-term results.
Because revision of implants with long cemented stems can be very
challenging, an effort should be made in the future to use shorter stems
in modular versions of hinged implants. Cite this article:
We determined the short-term clinical outcome
and migration within the bone of the humeral cementless component
of the Instrumented Bone Preserving (IBP) total elbow replacement
in a series of 16 patients. There were four men and 12 women with
a mean age at operation of 63 years (40 to 81). Migration was calculated
using radiostereometric analysis. There were no intra-operative
complications and no revisions. At two-year follow-up, all patients
showed a significant reduction in pain and functional improvement
of the elbow (both p <
0.001). Although ten components (63%)
showed movement or micromovement during the first six weeks, 14
(88%) were stable at one year post-operatively. Translation was
primarily found in the proximal direction (median 0.3 mm (interquartile range
(IQR) -0.09 to 0.8); the major
Mobile-bearing (MB) total knee replacement (TKR)
was introduced to reduce the risk of aseptic loosening and wear of
polyethylene inserts. However, no consistent clinical advantages
of mobile- over fixed-bearing (FB) TKR have been found. In this
study we evaluated whether mobile bearings have an advantage over
fixed bearings with regard to revision rates and clinical outcome
scores. Furthermore, we determined which modifying variables affected
the outcome. A systematic search of the literature was conducted to collect
clinical trials comparing MB and FB in primary TKR. The primary
outcomes were revision rates for any reason, aseptic loosening and
wear. Secondary outcomes included range of movement, Knee Society
score (KSS), Oxford knee score (OKS), Short-Form 12 (SF-12) score
and radiological parameters. Meta-regression techniques were used
to explore factors modifying the observed effect. Our search yielded 1827 publications, of which 41 studies met
our inclusion criteria, comprising over 6000 TKRs. Meta-analyses
showed no clinically relevant differences in terms of revision rates,
clinical outcome scores or patient-reported outcome measures between
MB and FB TKRs. It appears that theoretical assumptions of superiority
of MB over FB TKR are not borne out in clinical practice. Cite this article:
This prospective multicentre study was undertaken
to determine segmental movement, disc height and sagittal alignment
after total disc replacement (TDR) in the lumbosacral spine and
to assess the correlation of biomechanical properties to clinical
outcomes. A total of 173 patients with degenerative disc disease and low
back pain for more than one year were randomised to receive either
TDR or multidisciplinary rehabilitation (MDR). Segmental movement
in the sagittal plane and disc height were measured using distortion
compensated roentgen analysis (DCRA) comparing radiographs in active flexion
and extension. Correlation analysis between the range of movement
or disc height and patient-reported outcomes was performed in both
groups. After two years, no significant change in movement in the
sagittal plane was found in segments with TDR or between the two
treatment groups. It remained the same or increased slightly in
untreated segments in the TDR group and in this group there was
a significant increase in disc height in the operated segments.
There was no correlation between segmental movement or disc height
and patient-reported outcomes in either group. In this study, insertion of an intervertebral disc prosthesis
TDR did not increase movement in the sagittal plane and segmental
movement did not correlate with patient-reported outcomes. This
suggests that in the lumbar spine the movement preserving properties
of TDR are not major determinants of clinical outcomes. Cite this article:
The biomechanical function of the anteromedial
(AM) and posterolateral (PL) bundles of the anterior cruciate ligament
(ACL) remains controversial. Some studies report that the AM bundle
stabilises the knee joint in anteroposterior (AP) translation and
rotational movement (both internal and external) to the same extent
as the PL bundle. Others conclude that the PL bundle is more important
than the AM in controlling
Radiostereometric analysis (RSA) can detect early
micromovement in unstable implant designs which are likely subsequently
to have a high failure rate. In 2010, the Articular Surface Replacement
(ASR) was withdrawn because of a high failure rate. In 19 ASR femoral
components, the mean micromovement over the first two years after implantation
was 0.107 mm ( We conclude that the ASR femoral component achieves initial stability
and that early migration is not the mode of failure for this resurfacing
arthroplasty.
This study compared the outcome of total knee
replacement (TKR) in adult patients with fixed- and mobile-bearing prostheses
during the first post-operative year and at five years’ follow-up,
using gait parameters as a new objective measure. This double-blind
randomised controlled clinical trial included 55 patients with mobile-bearing (n
= 26) and fixed-bearing (n = 29) prostheses of the same design,
evaluated pre-operatively and post-operatively at six weeks, three
months, six months, one year and five years. Each participant undertook
two walking trials of 30 m and completed the EuroQol questionnaire,
Western Ontario and McMaster Universities osteoarthritis index,
Knee Society score, and visual analogue scales for pain and stiffness.
Gait analysis was performed using five miniature angular rate sensors
mounted on the trunk (sacrum), each thigh and calf. The study population
was divided into two groups according to age (≤ 70 years Improvements in most gait parameters at five years’ follow-up
were greater for fixed-bearing TKRs in older patients (>
70 years),
and greater for mobile-bearing TKRs in younger patients (≤ 70 years).
These findings should be confirmed by an extended age controlled
study, as the ideal choice of prosthesis might depend on the age
of the patient at the time of surgery.
Patients with skeletal dysplasia are prone to
developing advanced osteoarthritis of the knee requiring total knee replacement
(TKR) at a younger age than the general population. TKR in this
unique group of patients is a technically demanding procedure owing
to the deformity, flexion contracture, generalised hypotonia and ligamentous
laxity. We retrospectively reviewed the outcome of 11 TKRs performed
in eight patients with skeletal dysplasia at our institution using
the Stanmore Modular Individualised Lower Extremity System (SMILES)
custom-made rotating-hinge TKR. There were three men and five women
with mean age of 57 years (41 to 79). Patients were followed clinically
and radiologically for a mean of seven years (3 to 11.5). The mean
Knee Society clinical and function scores improved from 24 (14 to
36) and 20 (5 to 40) pre-operatively, respectively, to 68 (28 to
80) and 50 (22 to 74), respectively, at final follow-up. Four complications
were recorded, including a patellar fracture following a fall, a
tibial peri-prosthetic fracture, persistent anterior knee pain,
and aseptic loosening of a femoral component requiring revision.
Our results demonstrate that custom primary rotating-hinge TKR in
patients with skeletal dysplasia is effective at relieving pain,
with a satisfactory range of movement and improved function. It compensates
for bony deformity and ligament deficiency and reduces the likelihood
of corrective osteotomy. Patellofemoral joint complications are
frequent and functional outcome is worse than with primary TKR in
the general population.
We examined the morphology of mammalian hips asking whether evolution can explain the morphology of impingement in human hips. We describe two stereotypical mammalian hips, coxa recta and coxa rotunda. Coxa recta is characterised by a straight or aspherical section on the femoral head or head-neck junction. It is a sturdy hip seen mostly in runners and jumpers. Coxa rotunda has a round femoral head with ample head-neck offset, and is seen mostly in climbers and swimmers. Hominid evolution offers an explanation for the variants in hip morphology associated with impingement. The evolutionary conflict between upright gait and the birth of a large-brained fetus is expressed in the female pelvis and hip, and can explain pincer impingement in a coxa profunda. In the male hip, evolution can explain cam impingement in coxa recta as an adaptation for running.
We examined the reliability of radiological findings in predicting segmental instability in 112 patients (56 men, 56 women) with a mean age of 66.5 years (27 to 84) who had degenerative disease of the lumbar spine. They underwent intra-operative biomechanical evaluation using a new measurement system. Biomechanical instability was defined as a segment with a neutral zone >
2 mm/N. Risk factor analysis to predict instability was performed on radiographs (range of segmental movement, disc height), MRI (Thompson grade, Modic type), and on the axial CT appearance of the facet (type, opening, vacuum and the presence of osteophytes, subchondral erosion, cysts and sclerosis) using multivariate logistic regression analysis with a forward stepwise procedure. The facet type was classified as sagittally orientated, coronally orientated, anisotropic or wrapped. Stepwise multivariate regression analysis revealed that facet opening was the strongest predictor for instability (odds ratio 5.022, p = 0.009) followed by spondylolisthesis, MRI grade and subchondral sclerosis. Forward stepwise multivariate logistic regression indicated that spondylolisthesis, MRI grade, facet opening and subchondral sclerosis of the facet were risk factors. Symptoms evaluated by the Short-Form 36 and visual analogue scale showed that patients with an unstable segment were in significantly more pain than those without. Furthermore, the surgical procedures determined using the intra-operative measurement system were effective, suggesting that segmental instability influences the symptoms of lumbar degenerative disease.
Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce cementless fixation, which had been used successfully in the fixation of acetabular and tibial components. This modification appeared to have resulted in metallosis due to abrasive two-body wear. The component has subsequently been recalled and is no longer in use. The presentation, investigation, and findings at revision are described and a possible aetiology and its implications are discussed.
Components from 73 failed knee replacements (TKRs) consisting of rotating-platform, mobile-bearing and fixed-bearing implants were examined to assess the patterns of wear. The patterns were divided into low-grade (burnishing, abrasion and cold flow) and high-grade (scratching, pitting/metal embedding and delamination) to assess the severity of the wear of polyethylene. The rotating-platform group had a higher incidence of low-grade wear on the upper surface compared with the fixed-bearing group. By contrast, high-grade wear comprising scratching, pitting and third-body embedding was seen on the lower surface. Linear regression analysis showed a significant correlation of the wear scores between the upper and lower surfaces of the tibial insert (R2 = 0.29, p = 0.04) for the rotating-platform group, but no significant correlation was found for the fixed-bearing counterpart. This suggests that high-grade wear patterns on the upper surface are reduced with the rotating-platform design. However, the incidence of burnishing, pitting/third-body embedding and scratching wear patterns on the lower surface was higher compared with that in the fixed-bearing knee.