Aims. We compared the clinical and radiological outcomes of using a
polyetheretherketone cage with (TiPEEK) and without a titanium coating
(PEEK) for instrumented transforaminal lumbar interbody fusion (TLIF). Materials and Methods. We conducted a randomised clinical pilot trial of 40 patients
who were scheduled to undergo a TLIF procedure at one or two levels
between L2 and L5. The Oswestry disability index (ODI), EuroQoL-5D,
and back and leg pain were determined pre-operatively, and at three,
six, and 12 months post-operatively. Fusion rates were assessed
by thin slice CT at three months and by functional radiography at
12 months. Results. At final follow-up, one patient in each group had been lost to
follow-up. Two patients in each of the PEEK and TiPEEK groups were
revised for pseudarthrosis (p = 1.00). The rate of complete or partial
fusion at three months was 91.7% in both groups. Overall, there
were no significant differences in ODI or in radiological outcomes
between the groups. Conclusion. Favourable results with identical clinical outcomes and a high
rate of fusion was seen in both groups. The titanium
Periprosthetic joint infection (PJI) is one of the most dreaded complications after arthroplasty surgery; thus numerous approaches have been undertaken to equip metal surfaces with antibacterial properties. Due to its antimicrobial effects, silver is a promising
Aims. The aim of this study was to develop a single-layer hybrid organic-inorganic sol-gel
A silver-containing hydroxyapatite (Ag-HA)
The success of long-term transcutaneous implants
depends on dermal attachment to prevent downgrowth of the epithelium
and infection. Hydroxyapatite (HA)
Four uncemented Symax hip stems were extracted at three weeks and nine, 13 and 32 months, respectively, for reasons other than loosening. The reasons for implant removal were infection in two cases, recurrent dislocation in one and acetabular fracture in one. They were analysed to assess the effect and behaviour of an electrochemically deposited, completely resorbable biomimetic BONIT-hydroxyapatite (HA)
Aims. The Intraosseous Transcutaneous Amputation Prosthesis (ITAP)
may improve quality of life for amputees by avoiding soft-tissue
complications associated with socket prostheses and by improving
sensory feedback and function. It relies on the formation of a seal
between the soft tissues and the implant and currently has a flange
with drilled holes to promote dermal attachment. Despite this, infection
remains a significant risk. This study explored alternative strategies
to enhance soft-tissue integration. Materials and Methods. The effect of ITAP pins with a fully porous titanium alloy flange
with interconnected pores on soft-tissue integration was investigated.
The flanges were coated with fibronectin-functionalised hydroxyapatite
and silver
We inserted two hydroxyapatite (HA)-coated implants with crystallinities of either 50% (HA-50%) or 75% (HA-75%) bilaterally into the medial femoral condyles of the knees of 16 dogs. The implants were allocated to two groups with implantation periods of 16 and 32 weeks. They were weight-bearing and subjected to controlled micromovement of 250 μm during each gait cycle. After 16 weeks, mechanical fixation of the HA-50% implants was increased threefold as compared with the HA-75% implants. After 32 weeks there was no difference between HA-50% and HA-75%. Fixation of HA-75% increased from 16 to 32 weeks whereas that of HA-50% was unchanged. HA-50% implants had 100% more bone ingrowth than HA-75% implants after 16 weeks. More HA
The aim of this study was to evaluate whether
coating titanium discs with selenium in the form of sodium selenite decreased
bacterial adhesion of Staphylococcus aureus and Staph.
epidermidis and impeded osteoblastic cell growth. In order to evaluate bacterial adhesion, sterile titanium discs
were coated with increasing concentrations of selenium and incubated
with bacterial solutions of Staph. aureus (ATCC
29213) and Staph. epidermidis (DSM 3269) and stained
with Safranin-O. The effect of selenium on osteoblastic cell growth
was also observed. The adherence of MG-63 cells on the coated discs
was detected by staining with Safranin-O. The proportion of covered
area was calculated with imaging software. The tested Staph. aureus strain showed a significantly
reduced attachment on titanium discs with 0.5% (p = 0.011) and 0.2%
(p = 0.02) selenium
We report the radiological and clinical outcome of a press-fit (SLF) acetabular component at two to three years in two groups of patients having primary total hip replacement. In 69 the implant was coated with hydroxyapatite (HA) and in 40 it was uncoated. The stability of the cup was assessed by measurement of proximal migration and change in the angle of inclination. The clinical results in the two groups did not differ significantly, and the mean proximal linear wear was similar in both. Fewer radiolucent lines (RLLs) were seen on the radiographs of cups coated with HA. The mean proximal migration was studied by calculating regression lines for each patient using migration measurements: for the SLF+HA group the mean slope was 0.06 mm/year and for the SLF–HA group 0.20 mm/year (p = 0.22). The change in the angle of inclination during follow-up was also consistently smaller in HA–coated cups. Using regression methods the SLF+HA group had a mean slope of 0.08°/year and the SLF-HA group 0.44°/year (p = 0.023). Partial HA
In a single-blind, randomised series of knee replacements in 116 patients, we used radiostereometric analysis (RSA) to measure micromotion in three types of tibial implant fixation for two years after knee replacement. We compared hydroxyapatite-augmented porous
We carried out a histological study of a proximally hydroxyapatite (HA)-coated femoral component, retrieved after 9.5 years of good function. The HA
The outcome at ten years of 100 Freeman hip stems (Finsbury Orthopaedics, Leatherhead, United Kingdom) retaining the neck with a proximal hydroxyapatite
We investigated the effects of hydroxyapatite (HA)
In previous studies, we have demonstrated a fibrocartilaginous membrane around hydroxyapatite-coated implants subjected to micromovement in contrast to the fibrous connective tissue which predominates around similarly loaded titanium alloy implants. In the present study, in mature dogs, we investigated the effect of immobilising titanium (Ti)- or hydroxyapatite (HA)-coated implants already surrounded by a movement-induced fibrous membrane and compared the results with those of similar implants in which continuous micromovement was allowed to continue. The implants were inserted in the medial femoral condyles of 14 dogs and subjected to 150 microns movements during each gait cycle. After four weeks (when a fibrous membrane had developed), half the implants were immobilised to prevent further micromovement. The dogs were killed at 16 weeks and the results were evaluated by push-out tests and histological analysis. The continuously loaded Ti-coated implants were surrounded by a fibrous membrane, whereas bridges of new bone anchored the HA-coated implants. The immobilised implants were surrounded by bone irrespective of the type of
We have studied the beneficial effects of a hydroxyapatite (HA)
We report the histological findings in post-mortem specimens obtained ten days, 17 days and seven weeks after implantation of hydroxyapatite-coated femoral components of hip arthroplasties. There was early deposition of woven bone on the hydroxyapatite ceramic, identical to that deposited on surviving cancellous trabeculae. The space between these deposits became bridged from both sides by new trabeculae, and there was no evidence of an inflammatory reaction or of fibrous tissue formation. The use of an hydroxyapatite
We studied two groups of femoral hip prostheses: 43 TiAlV ridged press-fit stems, and 26 with similar stems coated with hydroxyapatite on the proximal half. At one year, radiological measurement showed a mean downward migration of 0.99 mm for the TialV prostheses and 0.12 mm for the HA-coated prostheses (p = 0.0002). Hydroxyapatite
We describe the survival at ten years of 100 femoral components of the Freeman hip prosthesis. It is proximally hydroxyapatite (HA)-coated and was fixed without cement. Radiological assessment identified radiolucent lines (RLLs) and lytic lesions and was used to measure migration. The criterion of failure was revision or impending revision for aseptic femoral loosening. No femoral components were revised or are awaiting revision for aseptic loosening, giving 100% survival at ten years (95% confidence interval, 95.7 to 100), although 59 were at risk at ten years. Two components were revised for fracture of a ceramic head with damage to the trunnion. Although well fixed in each, for survival analysis we evaluated the hip as if the patient had died. Twelve acetabular components were revised and at each operation the femoral component was found to be well fixed, was not disturbed and remained in the survival analysis. Three patients were lost to follow-up, and 12 died with well-functioning prostheses. Radiologically, all except one of the components appeared to be well fixed with no RLLs and no lytic lesions at the latest follow-up. The mean vertical migration was 0.4 mm at one year, 0.8 mm at two years and 1.4 mm at ten years. One component had migrated 7.6 mm at ten years (2.1 mm in year 1) and developed RLLs in Gruen zones I and II. The symptoms, however, were only minor and revision was not indicated. Our study has shown that proximal HA
One concern about the fixation of HA-coated implants is the possible disintegration of the surface, with the migration of HA granules into the joint space, producing third-body wear. We report a study of six revisions of HA-coated polyethylene RM cups at 9 to 14 years after successful primary arthroplasty. In all six hips, we found HA granules embedded in the articulating surface of the polyethylene, with abrasive wear of the cup and the metal femoral head. The cup had loosened in four hips and three showed severe osteolysis of the proximal femur. Third-body wear due to HA particles from implant
In 33 dogs we implanted femoral stems made either of carbon composite, some coated with hydroxyapatite, or of titanium alloy with a porous
We measured the insertion and extraction torque forces in a randomised study of 76 external fixation screws in 19 patients treated by hemicallotasis for osteoarthritis of the medial side of the knee. The patients were randomised to have either standard tapered screws (Orthofix 6/5 mm) or the same screws with hydroxyapatite (HA)
We evaluated the effects of a serum protein
In rabbits and goats, test implants with a porous surface of two layers of Tl-6A;-4V beads were examined at intervals for bond strength with bone. Half of the implants were coated with hydroxyapatite by plasma spray. The bonding strength with bone in the coated specimens was about four times greater than that of the uncoated specimens at two weeks, and twice as strong at six weeks. Twelve weeks after implantation, the strengths were similar. The hydroxyapatite
We have studied the effect of hydroxyapatite (HA)
This prospective study compares the outcome of
157 hydroxyapatite (HA)-coated tibial components with 164 cemented
components in the ROCC Rotating Platform total knee replacement
in 291 patients. The mean follow-up was 7.6 years (5.2 to 11). There
were two revisions for loosening: one for an HA-coated and one for
a cemented tibial component. Radiological evaluation demonstrated
no radiolucent lines with the HA-coated femoral components. A total
of three HA-coated tibial components exhibited radiolucent lines
at three months post-operatively and these disappeared after three
further months of protected weight-bearing. With HA-coated components
the operating time was shorter (p <
0.006) and the radiological
assessment of the tibial interface was more stable (p <
0.01).
Using revision for aseptic loosening of the tibial component as
the end point, the survival rates at nine years was identical for
both groups at 99.1%. Our results suggest that HA-coated components perform at least
as well as the same design with cemented components and compare
favourably with those of series describing cemented or porous-coated
knee replacements, suggesting that fixation of both components with
hydroxyapatite is a reliable option in primary total knee replacement.
We reviewed 40 cementless Porous Coated Anatomic knee replacements in 34 consecutive patients. The average follow-up was 12.9 months (range 6 to 36 months). At review, loose beads were identified in 23 knees, in 11 on the femoral side and in 16 on the tibial side (in five knees on both sides). One loose bead was adjacent to the patella. In four knees there were intra-articular beads. Radiolucent lines at the interface between bone and prosthesis were seen in 19 knees, in six on the femoral side and in 16 on the tibial side (in three knees on both sides). In relation to the tibial component loose beads were associated with a radiolucent line in all except one case. Progressive loosening with an increasing number of loose beads was seen in three knees. At this early stage of follow-up there was no correlation between the clinical results and the presence of loose beads.
Aims. BoneMaster is a thin electrochemically applied hydroxyapatite (HA)
Aims. This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant. Methods. A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life. Results. No differences in mean age (69 years, 40 to 86), mean age adapted Charlson Comorbidity Index (3.1 (SD 1.4)), mean BMI (29.2 kg/m. 2. (SD 5.8)), or number of other implants were observed between groups. Significant improvements in outcome scores and pain levels were achieved for all groups, and there was no difference in quality of life (12-Item Short-Form Health Survey questionnaire (SF-12)). Mean cobalt and chromium ion levels were significantly higher for the GenuX versus the BPKS hinged TKA (GenuX vs BPKS: cobalt: 16.3 vs 9.4 µg/l; chromium: 9.5 vs 5.2 µg/l). The TiN-coated implants did not appear to confer improvement in the metal ion levels. Metal ion concentrations above 7 µg/l were detected in 81%(29/36) of GenuX patients versus 41% (11/27) in the BPKS group. No GenuX patients had normal levels under 2 µg/l, versus 22% of BPKS patients. No significant reduction in outcome scores was observed regardless of the metal ion levels, whereas higher work-related activity was correlated with higher chromium concentrations. Conclusion. Hinged TKA, using MoM hinges, resulted in critically high cobalt and chromium ion concentrations. The BPKS hinged TKA showed significantly lower metal ion concentrations compared with the GenuX TKA. No benefits were observed using TiN
Aims. The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Materials and Methods. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed. Results. In 104 implants, a decrease in the odds of loosening and an increase in metallosis and tissue staining in the Duofix group relative to Porocoat group was detected (p = 0.028). There was an increased presence of embedded metallic debris in the Duofix group (p < 0.001). Decreased tissue ingrowth was associated with the Duofix surface (p < 0.001). The attached beads had reduced microhardness, indicative of adverse thermal processing, which resulted in bead shedding, particulate debris, and metallosis. Conclusion. Hydroxyapatite
Aims. The purpose of this study was to evaluate the biological fixation of a 3D printed porous implant, with and without different hydroxyapatite (HA)
The clinical use of hydroxyapatite (HA)
Resurfacing of the humeral head is commonly used
within the UK to treat osteoarthritis (OA) of the shoulder. We present
the results of a small prospective randomised study of this procedure
using the Global CAP prosthesis with two different
This study describes 146 primary total knee replacements, either fully or partially coated with hydroxyapatite of which 74 knees in 68 patients were available for clinical and radiological assessment at a mean of 11.2 years (10 to 15). The global failure rate was 1.37% and survival rate with mechanical failure as the end-point was 98.14%. Radiological assessment indicated intimate contact between bone and the hydroxyapatite
The aim of this study was to evaluate the survival of a collarless, straight, hydroxyapatite-coated femoral stem in total hip arthroplasty (THA) at a minimum follow-up of 20 years. We reviewed the results of 165 THAs using the Omnifit HA system in 138 patients, performed between August 1993 and December 1999. The mean age of the patients at the time of surgery was 46 years (20 to 77). Avascular necrosis was the most common indication for THA, followed by ankylosing spondylitis and primary osteoarthritis. The mean follow-up was 22 years (20 to 31). At 20 and 25 years, 113 THAs in 91 patients and 63 THAs in 55 patients were available for review, respectively, while others died or were lost to follow-up. Kaplan-Meier analysis was performed to evaluate the survival of the stem. Radiographs were reviewed regularly, and the stability of the stem was evaluated using the Engh classification.Aims
Methods
The aim of this study was to compare the pattern of initial fixation and changes in periprosthetic bone mineral density (BMD) between patients who underwent total hip arthroplasty (THA) using a traditional fully hydroxyapatite (HA)-coated stem (T-HA group) and those with a newly introduced fully HA-coated stem (N-HA group). The study included 36 patients with T-HA stems and 30 with N-HA stems. Dual-energy X-ray absorptiometry was used to measure the change in periprosthetic BMD, one and two years postoperatively. The 3D contact between the stem and femoral cortical bone was evaluated using a density-mapping system, and clinical assessment, including patient-reported outcome measurements, was recorded.Aims
Methods
The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design. This was a retrospective review of 707 primary TKAs in 590 patients from a single institution, using a cementless, metal-backed patellar component with a mean follow-up of 6.9 years (2 to 12). A total of 409 TKAs were performed in 338 females and 298 TKAs in 252 males. The mean age of the patients was 63 years (34 to 87) and their mean BMI was 34.3 kg/m2 (18.8 to 64.5). The patients were chosen to undergo a cementless procedure based on age and preoperative radiological and intraoperative bone quality. Outcome was assessed using the Knee Society knee and function scores and range of motion (ROM), complications, and revisions.Aims
Methods
The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface. Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up.Aims
Methods
Implant-associated infection is a major source
of morbidity in orthopaedic surgery. There has been extensive research
into the development of materials that prevent biofilm formation,
and hence, reduce the risk of infection. Silver nanoparticle technology
is receiving much interest in the field of orthopaedics for its
antimicrobial properties, and the results of studies to date are
encouraging. Antimicrobial effects have been seen when silver nanoparticles are
used in trauma implants, tumour prostheses, bone cement, and also
when combined with hydroxyapatite
Reverse total shoulder arthroplasty (rTSA) can be used in complex cases when the glenoid requires reconstruction. In this study, a baseplate with composite bone autograft and a central trabecular titanium peg was implanted, and its migration was assessed for two years postoperatively using radiostereometric analysis (RSA). A total of 14 patients who underwent a rTSA with an autograft consented to participate. Of these, 11 had a primary rTSA using humeral head autograft and three had a revision rTSA with autograft harvested from the iliac crest. The mean age of the patients was 66 years (39 to 81). Tantalum beads were implanted in the scapula around the glenoid. RSA imaging (stereographic radiographs) was undertaken immediately postoperatively and at three, six, 12, and 24 months. Analysis was completed using model-based RSA software. Outcomes were collected preoperatively and at two years postoperatively, including the Oxford Shoulder Score, the American Shoulder and Elbow Score, and a visual analogue score for pain. A Constant score was also obtained for the assessment of strength and range of motion.Aims
Methods
We have updated our previous randomised controlled trial comparing release of chromium (Cr) and cobalt (Co) ions and included levels of titanium (Ti) ions. We have compared the findings from 28 mm metal-on-metal total hip replacement, performed using titanium CLS/Spotorno femoral components and titanium AlloFit acetabular components with Metasul bearings, with Durom hip resurfacing using a Metasul articulation or bearing and a titanium plasma-sprayed
Endoprosthetic reconstruction following distal femur tumour resection has been widely advocated. In this paper, we present the design of an uncemented endoprosthesis system featuring a short, curved stem, with the goal of enhancing long-term survivorship and functional outcomes. This study involved patients who underwent implantation of an uncemented distal femoral endoprosthesis with a short and curved stem between 2014 and 2019. Functional outcomes were assessed using the 1993 version of the Musculoskeletal Tumour Society (MSTS-93) score. Additionally, we quantified five types of complications and assessed osseointegration radiologically. The survivorship of the endoprosthesis was evaluated according to two endpoints. A total of 134 patients with a median age of 26 years (IQR 16 to 41) were included in our study. The median follow-up time was 61 months (IQR 56 to 76), and the median functional MSTS-93 was 83% (IQR 73 to 91) postoperatively.Aims
Methods
The aim of this study was to analyze how proximal radial neck resorption (PRNR) starts and progresses radiologically in two types of press-fit radial head arthroplasties (RHAs), and to investigate its clinical relevance. A total of 97 patients with RHA were analyzed: 56 received a bipolar RHA (Group 1) while 41 received an anatomical implant (Group 2). Radiographs were performed postoperatively and after three, six, nine, and 12 weeks, six, nine, 12, 18, and 24 months, and annually thereafter. PRNR was measured in all radiographs in the four radial neck quadrants. The Mayo Elbow Performance Score (MEPS), the abbreviated version of the Disabilities of the Arm, Shoulder, and Hand questionnaire (QuickDASH), and the patient-assessed American Shoulder and Elbow Surgeons score - Elbow (pASES-E) were used for the clinical assessment. Radiological signs of implant loosening were investigated.Aims
Methods
We investigated the implant-bone interface around one design of femoral stem, proximally coated with either a plasma-sprayed porous
The benefit of a dual-mobility acetabular component (DMC) for primary total hip arthroplasties (THAs) is controversial. This study aimed to compare the dislocation and complication rates when using a DMC compared to single-mobility (SM) acetabular component in primary elective THA using data collected at a single centre, and compare the revision rates and survival outcomes in these two groups. Between 2010 and 2019, 2,075 primary THAs using either a cementless DM or SM acetabular component were included. Indications for DMC were patients aged older than 70 years or with high risk of dislocation. All other patients received a SM acetabular component. Exclusion criteria were cemented implants, patients treated for femoral neck fracture, and follow-up of less than one year. In total, 1,940 THAs were analyzed: 1,149 DMC (59.2%) and 791 SM (40.8%). The mean age was 73 years (SD 9.2) in the DMC group and 57 years (SD 12) in the SM group. Complications and revisions have been analyzed retrospectively.Aims
Methods
The primary aim of this study was to compare the migration of the femoral and tibial components of the cementless rotating platform Attune and Low Contact Stress (LCS) total knee arthroplasty (TKA) designs, two years postoperatively, using radiostereometric analysis (RSA) in order to assess the risk of the development of aseptic loosening. A secondary aim was to compare clinical and patient-reported outcome measures (PROMs) between the designs. A total of 61 TKAs were analyzed in this randomized clinical RSA trial. RSA examinations were performed one day and three, six, 12, and 24 months postoperatively. The maximal total point motion (MPTM), translations, and rotations of the components were analyzed. PROMs and clinical data were collected preoperatively and at six weeks and three, six, 12, and 24 months postoperatively. Linear mixed effect modelling was used for statistical analyses.Aims
Methods
Aims. Hydroxyapatite
Aims. Pelvic reconstruction after the resection of a tumour around
the acetabulum is a challenging procedure due to the complex anatomy
and biomechanics. Several pelvic endoprostheses have been introduced,
but the rates of complication remain high. Our aim was to review
the use of a stemmed acetabular pedestal cup in the management of
these patients. Patients and Methods. The study involved 48 patients who underwent periacetabular reconstruction
using a stemmed pedestal cup (Schoellner cup; Zimmer Biomet Inc.,
Warsaw, Indiana) between 2000 and 2013. The indications for treatment included
a primary bone tumour in 27 patients and metastatic disease in 21
patients. The mean age of the patients at the time of surgery was
52 years (16 to 83). Results. At a median follow-up of 6.6 years (95% confidence interval 4.6
to 8.2), local control was achieved in all patients; 19 patients
had died (16 of disease). Complications occurred in 19 patients
(40%), of which deep infection was the most common, affecting eight
patients (17%). Seven patients (15%) had a dislocation of the hip.
Aseptic loosening was found in three patients (6%). Two (4%) underwent
hindquarter amputation for non-oncological reasons. The risk of
revision, with death being treated as a competing event, was 28%
at one year, 39% at five years and 48% at ten years post-operatively.
The mean Musculoskeletal Tumour Society Score at final follow-up
was 71% (27% to 93%). Conclusion. This type of reconstruction is a satisfactory option for the
treatment of patients with a periacetabular tumour. There remains,
however, a high rate of complication, which may be reduced by future
modifications of the device such as silver
Aims. The optimal method of tibial component fixation remains uncertain
in total knee arthroplasty (TKA). Hydroxyapatite
We present the 5.6- to 7.6-year results of our first 118 hydroxyapatite (HA)-coated total hip replacements in patients under 66 years of age. The titanium femoral component has proximal HA
Periprosthetic joint infection (PJI) remains an extremely challenging complication. We have focused on this issue more over the last decade than previously, but there are still many unanswered questions. We now have a workable definition that everyone should align to, but we need to continue to focus on identifying the organisms involved. Surgical strategies are evolving and care is becoming more patient-centred. There are some good studies under way. There are, however, still numerous problems to resolve, and the challenge of PJI remains a major one for the orthopaedic community. This annotation provides some up-to-date thoughts about where we are, and the way forward. There is still scope for plenty of research in this area. Cite this article:
The ‘jumbo’ acetabular component is now commonly
used in acetabular revision surgery where there is extensive bone
loss. It offers high surface contact, permits weight bearing over
a large area of the pelvis, the need for bone grafting is reduced
and it is usually possible to restore centre of rotation of the
hip. Disadvantages of its use include a technique in which bone
structure may not be restored, a risk of excessive posterior bone
loss during reaming, an obligation to employ screw fixation, limited
bone ingrowth with late failure and high hip centre, leading to increased
risk of dislocation. Contraindications include unaddressed pelvic
dissociation, inability to implant the component with a rim fit,
and an inability to achieve screw fixation. Use in acetabulae with
<
50% bone stock has also been questioned. Published results
have been encouraging in the first decade, with late failures predominantly because
of polyethylene wear and aseptic loosening. Dislocation is the most
common complication of jumbo acetabular revisions, with an incidence
of approximately 10%, and often mandates revision. Based on published results,
a hemispherical component with an enhanced porous
Femoral component anteversion is an important factor in the success of total hip arthroplasty (THA). This retrospective study aimed to investigate the accuracy of femoral component anteversion with the Mako THA system and software using the Exeter cemented femoral component, compared to the Accolade II cementless femoral component. We reviewed the data of 30 hips from 24 patients who underwent THA using the posterior approach with Exeter femoral components, and 30 hips from 24 patients with Accolade II components. Both groups did not differ significantly in age, sex, BMI, bone quality, or disease. Two weeks postoperatively, CT images were obtained to measure acetabular and femoral component anteversion.Aims
Methods
Implant failure has become more common as the number of primary total ankle arthroplasties (TAAs) performed has increased. Although revision arthroplasty has gained attention for functional preservation, the long-term results remain unclear. This study aimed to assess the long-term outcomes of revision TAA using a mobile-bearing prosthesis in a considerably large cohort; the risk factors for failure were also determined. This single-centre retrospective cohort study included 116 patients (117 ankles) who underwent revision TAA for failed primary TAA between July 2000 and March 2010. Survival analysis and risk factor assessment were performed, and clinical performance and patient satisfaction were evaluated preoperatively and at last follow-up.Aims
Methods
Delayed postoperative inoculation of orthopaedic implants with persistent wound drainage or bacterial seeding of a haematoma can result in periprosthetic joint infection (PJI). The aim of this in vivo study was to compare the efficacy of vancomycin powder with vancomycin-eluting calcium sulphate beads in preventing PJI due to delayed inoculation. A mouse model of PJI of the knee was used. Mice were randomized into groups with intervention at the time of surgery (postoperative day (POD) 0): a sterile control (SC; n = 6); infected control (IC; n = 15); systemic vancomycin (SV; n = 9); vancomycin powder (VP; n = 21); and vancomycin bead (VB; n = 19) groups. Delayed inoculation was introduced during an arthrotomy on POD 7 with 1 × 105 colony-forming units (CFUs) of a bioluminescent strain of Aims
Methods
Dual-mobility (DM) components are increasingly used to prevent and treat dislocation after total hip arthroplasty (THA). Intraprosthetic dissociation (IPD) is a rare complication of DM that is believed to have decreased with contemporary implants. This study aimed to report incidence, treatment, and outcomes of contemporary DM IPD. A total of 1,453 DM components were implanted at a single academic institution between January 2010 and December 2021: 695 in primary and 758 in revision THA. Of these, 49 presented with a dislocation of the large DM head and five presented with an IPD. At the time of closed reduction of the large DM dislocation, six additional IPDs occurred. The mean age was 64 years (SD 9.6), 54.5% were female (n = 6), and mean follow-up was 4.2 years (SD 1.8). Of the 11 IPDs, seven had a history of instability, five had abductor insufficiency, four had prior lumbar fusion, and two were conversions for failed fracture management.Aims
Methods
The aim of the study was to investigate whether the primary stability of press-fit acetabular components can be improved by altering the impaction procedure. Three impaction procedures were used to implant acetabular components into human cadaveric acetabula using a powered impaction device. An impaction frequency of 1 Hz until complete component seating served as reference. Overimpaction was simulated by adding ten strokes after complete component seating. High-frequency implantation was performed at 6 Hz. The lever-out moment of the acetabular components was used as measure for primary stability. Permanent bone deformation was assessed by comparison of double micro-CT (µCT) measurements before and after impaction. Acetabular component deformation and impaction forces were recorded, and the extent of bone-implant contact was determined from 3D laser scans.Aims
Methods
The Vantage Total Ankle System is a fourth-generation low-profile fixed-bearing implant that has been available since 2016. We aimed to describe our early experience with this implant. This is a single-centre retrospective review of patients who underwent primary total ankle arthroplasty (TAA) with a Vantage implant between November 2017 and February 2020, with a minimum of two years’ follow-up. Four surgeons contributed patients. The primary outcome was reoperation and revision rate of the Vantage implant at two years. Secondary outcomes included radiological alignment, peri-implant complications, and pre- and postoperative patient-reported outcomes.Aims
Methods
The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years. The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.Aims
Methods
In this study we present our experience with
four generations of uncemented total knee arthroplasty (TKA) from Smith
&
Nephew: Tricon M, Tricon LS, Tricon II and Profix, focusing
on the failure rates correlating with each design change. Beginning
in 1984, 380 Tricon M, 435 Tricon LS, 305 Tricon 2 and 588 Profix
were implanted by the senior author. The rate of revision for loosening
was 1.1% for the Tricon M, 1.1% for the Tricon LS, 0.5% for the
Tricon 2 with a HA coated tibial component, and 1.3% for the Profix
TKA. No loosening of the femoral component was seen with the Tricon
M, Tricon LS or Tricon 2, with no loosening seen of the tibial component
with the Profix TKA. Regarding revision for wear, the incidence
was 13.1% for the Tricon M, 6.6% for the Tricon LS, 2.3% for the
Tricon 2, and 0% for the Profix. These results demonstrate that
improvements in the design of uncemented components, including increased
polyethylene thickness, improved polyethylene quality, and the introduction
of hydroxyapatite
The aims of this study were to evaluate the incidence
of local argyria in patients with silver-coated megaprostheses and
to identify a possible association between argyria and elevated
levels of silver both locally and in the blood. Between 2004 and
2011, 32 megaprostheses with silver
Between March 1990 and May 1991 we performed 85 primary total hip replacements in 74 patients using the Landos Atoll hydroxyapatite (HA)-coated cup and the Corail HA-coated stem. The patients were followed up for a mean of ten years. Of the 85 cups, 26 (31%) have already been revised and a further six are radiologically unstable and awaiting revision. Two femoral stems have been revised for infection without loosening. The retrieved acetabular cups were studied by SEM and image-processing techniques to quantify the amount of residual HA on the cup. This was correlated with the clinical variables and modes of failure. The residual HA (as a percentage of the surface) on the loose cups correlated negatively with the duration of implantation (r = −0.732, p <
0.001). Six cups were stable at revision and had more residual HA
Between 2002 and 2008, 130 consecutive ankles were replaced with an hydroxyapatite (HA) and titanium-HA-coated Ankle Evolutive System total ankle prosthesis. Plain radiographs were analysed by two independent observers. Osteolytic lesions were classified by their size and location, with cavities >
10 mm in diameter considered to be ‘marked’. CT scanning was undertaken in all patients with marked osteolysis seen on the plain radiographs. Osteolytic lesions were seen on the plain films in 48 (37%) and marked lesions in 27 (21%) ankles. The risk for osteolysis was found to be 3.1 (95% confidence interval 1.6 to 5.9) times higher with implants with Ti-HA porous
We investigated whether strontium-enriched calcium
phosphate cement (Sr-CPC)-treated soft-tissue tendon graft results
in accelerated healing within the bone tunnel in reconstruction
of the anterior cruciate ligament (ACL). A total of 30 single-bundle
ACL reconstructions using tendo Achillis allograft were performed
in 15 rabbits. The graft on the tested limb was treated with Sr-CPC,
whereas that on the contralateral limb was untreated and served
as a control. At timepoints three, six, nine, 12 and 24 weeks after
surgery, three animals were killed for histological examination.
At six weeks, the graft–bone interface in the control group was
filled in with fibrovascular tissue. However, the gap in the Sr-CPC
group had already been completely filled in with new bone, and there
was evidence of the early formation of Sharpey fibres. At 24 weeks,
remodelling into a normal ACL–bone-like insertion was found in the
Sr-CPC group.
Metallosis after shoulder replacement has not previously been described in the literature. We report a patient who developed extensive metallosis after implantation of an uncemented Nottingham shoulder replacement. He underwent a revision procedure. Examination of the retrieved prosthesis showed that the titanium porous
This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA). Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up.Aims
Methods
The aim of this study was to evaluate the performance of first-generation annealed highly cross-linked polyethylene (HXLPE) in cementless total hip arthroplasty (THA). We retrospectively evaluated 29 patients (35 hips) who underwent THA between December 2000 and February 2002. The survival rate was estimated using the Kaplan-Meier method. Hip joint function was evaluated using the Japanese Orthopaedic Association (JOA) score. Two-dimensional polyethylene wear was estimated using Martell’s Hip Analysis Suite. We calculated the wear rates between years 1 and 5, 5 and 10, 10 and 15, and 15 and final follow-up.Aims
Methods
The Cambridge Cup has been designed to replace the horseshoe-shaped articular cartilage of the acetabulum and the underlying subchondral bone. It is intended to provide physiological loading with minimal resection of healthy bone. The cup has been used in 50 women with displaced, subcapital fractures of the neck of the femur. In 24 cases, the cup was coated with hydroxyapatite. In 26, the
Periprosthetic joint infection (PJI) is a difficult complication requiring a comprehensive eradication protocol. Cure rates have essentially stalled in the last two decades, using methods of antimicrobial cement joint spacers and parenteral antimicrobial agents. Functional spacers with higher-dose antimicrobial-loaded cement and antimicrobial-loaded calcium sulphate beads have emphasized local antimicrobial delivery on the premise that high-dose local antimicrobial delivery will enhance eradication. However, with increasing antimicrobial pressures, microbiota have responded with adaptive mechanisms beyond traditional antimicrobial resistance genes. In this review we describe adaptive resistance mechanisms that are relevant to the treatment of PJI. Some mechanisms are well known, but others are new. The objective of this review is to inform clinicians of the known adaptive resistance mechanisms of microbes relevant to PJI. We also discuss the implications of these adaptive mechanisms in the future treatment of PJI. Cite this article:
We studied the effects of
Knee arthroplasty surgery is a highly effective treatment for arthritis and disorders of the knee. There are a wide variety of implant brands and types of knee arthroplasty available to surgeons. As a result of a number of highly publicized failures, arthroplasty surgery is highly regulated in the UK and many other countries through national registries, introduced to monitor implant performance, surgeons, and hospitals. With time, the options available within many brand portfolios have grown, with alternative tibial or femoral components, tibial insert materials, or shapes and patella resurfacings. In this study we have investigated the effect of the expansion of implant brand portfolios and where there may be a lack of transparency around a brand name. We also aimed to establish the potential numbers of compatible implant construct combinations. Hypothetical implant brand portfolios were proposed, and the number of compatible implant construct combinations was calculated.Aims
Methods
Implantation of a large-diameter femoral head prosthesis with a metal-on-metal bearing surface reduces the risk of dislocation, increases the range of movement, minimises the risk of impingement and, in theory, results in little wear. Between February 2004 and March 2007 we implanted 100 consecutive total hip replacements with a metal-on-metal bearing and a large femoral head into 92 patients. There were 51 men and 41 women with a mean age of 50 years (18 to 70) at the time of surgery. Outcome was assessed using the Western Ontario McMaster University osteoarthritis index and the Harris hip score as well as the Devane activity score. These all improved significantly (p <
0.0001). At the last follow-up there were no cases of dislocation, no impingement, a good range of movement and no osteolysis, but seven revisions, two for infection and five for aseptic loosening. The probability of groin pain increased if the other acetabular component inclination exceeded 50° (p = 0.0007). At 4.8 years of follow-up, the projected survival of the Durom acetabular component, with revision for any reason, was 92.4% (. sd. 2.8) (95% confidence interval 89.6 to 95.2). The design of the component made it difficult both to orientate and seat, which when combined with a poor porous
It has been hypothesized that proximal radial neck resorption (PRNR) following press-fit radial head arthroplasty (RHA) is due to stress-shielding. We compared two different press-fit stems by means of radiographs to investigate whether the shape and size of the stems are correlated with the degree of PRNR. The radiographs of 52 RHAs were analyzed both at 14 days postoperatively and after two years. A cylindrical stem and a conical stem were implanted in 22 patients (group 1) and 30 patients (group 2), respectively. The PRNR was measured in the four quadrants of the radial neck and the degree of stem filling was calculated by analyzing the ratio between the prosthetic stem diameter (PSD) and the medullary canal diameter (MCD) at the proximal portion of the stem (level A), halfway along the stem length (level B), and distally at the stem tip (level C).Aims
Methods
Stemless humeral implants have been developed to overcome stem-related complications in total shoulder arthroplasty (TSA). However, stemless implant designs may hypothetically result in less stable initial fixation, potentially affecting long-term survival. The aim of this study is to investigate early fixation and migration patterns of the stemless humeral component of the Simpliciti Shoulder System and to evaluate clinical outcomes. In this prospective cohort study, radiostereometric analysis (RSA) radiographs were obtained in 24 patients at one day, six weeks, six months, one year, and two years postoperatively. Migration was calculated using model-based RSA. Clinical outcomes were evaluated using the visual analogue scale (VAS), the Oxford Shoulder Score (OSS), the Constant-Murley Score (CMS), and the Disabilities of the Arm, Shoulder and Hand (DASH) score.Aims
Methods
Implants of solid sintered hydroxyapatite form very tight bonds with living bone, but are susceptible to fatigue failure. This problem can be overcome by using plasma-sprayed apatite
Ten acetabular cups coated with hydroxyapatite (HA) had originally been inserted in five primary and five revision total hip replacements. The thickness of the HA was 155 ± 35 μm. The cups, which were well-fixed, were retrieved, with their adherent tissue, at reoperation after 0.3 to 5.8 years because of infection (five hips), wear of polyethylene (three hips), and instability (two hips). Undecalcified sections showed a direct contact between bone and osteoid-like tissue which had formed directly onto the HA
We designed an in vivo study to determine if the superimposition of a microtexture on the surface of sintered titanium beads affected the extent of bone ingrowth. Cylindrical titanium intramedullary implants were coated with titanium beads to form a porous finish using commercial sintering techniques. A control group of implants was left in the as-sintered condition. The test group was etched in a boiling acidic solution to create an irregular surface over the entire porous
Higher osteoblastic bone activity is expected in aseptic loosening and painful unicompartmental knee arthroplasty (UKA). However, insights into normal bone activity patterns after medial UKAs are lacking. The aim of this study was to identify the evolution in bone activity pattern in well-functioning medial mobile-bearing UKAs. In total, 34 patients (13 female, 21 male; mean age 62 years (41 to 79); BMI 29.7 kg/m2 (23.6 to 42.1)) with 38 medial Oxford partial UKAs (20 left, 18 right; 19 cementless, 14 cemented, and five hybrid) were prospectively followed with sequential 99mTc-hydroxymethane diphosphonate single photon emission CT (SPECT)/CT preoperatively, and at one and two years postoperatively. Changes in mean osteoblastic activity were investigated using a tracer localization scheme with volumes of interest (VOIs), reported by normalized mean tracer values. A SPECT/CT registration platform additionally explored cortical tracer evolution in zones of interest identified by previous experimental research.Aims
Methods
We allocated randomly 27 patients undergoing 28 primary uncemented total hip replacements (THR) to receive prosthetic components of similar design with either plasma-sprayed titanium alloy (Ti)
The vertical migration of four configurations of a proximal femoral prosthesis, followed for up to nine years, was measured on standard radiographs. The same implant was used without cement (group 1) and with cement (group 2). The migration of both groups was linear from six months onwards. The mean migration rate and the incidence of late aseptic loosening were both greater in group 1. Survival analysis of the two groups, however, showed no statistically significant difference. In both groups, hips later destined for revision migrated more rapidly from the initial postoperative period onwards, than did the remainder. A threshold migration of 1.2 mm/year during the first two years after implantation detected hips likely to fail with a specificity of 86% and a sensitivity of 78%. This 'migration test' was applied to the results in two further groups of patients in which a modified femoral prosthesis had been implanted without hydroxyapatite
We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty. In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes.Aims
Methods
Adverse local tissue reactions associated with abnormal wear considerably slowed down the general use of metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), now limited to a few specialized centres. In this study, we provide the clinical results of 400 consecutive MoM HRAs implanted more than 20 years ago in one such centre. A total of 355 patients (400 hips) were treated with Conserve Plus HRA between November 1996 and November 2000. There were 96 female (27%) and 259 male patients (73%). Their mean age was 48.2 years (SD 10.9). The University of California, Los Angeles (UCLA) hip scores and 12-item Short Form Survey (SF-12) quality of life scores were reported. Survivorship was assessed using Kaplan-Meier analyses.Aims
Methods
One-stage revision hip arthroplasty for periprosthetic joint infection (PJI) has several advantages; however, resection of the proximal femur might be necessary to achieve higher success rates. We investigated the risk factors for resection and re-revisions, and assessed complications and subsequent re-revisions. In this single-centre, case-control study, 57 patients who underwent one-stage revision arthroplasty for PJI of the hip and required resection of the proximal femur between 2009 and 2018 were identified. The control group consisted of 57 patients undergoing one-stage revision without bony resection. Logistic regression analysis was performed to identify any correlation with resection and the risk factors for re-revisions. Rates of all-causes re-revision, reinfection, and instability were compared between groups.Aims
Methods
The aims of this study were to develop an in vivo model of periprosthetic joint infection (PJI) in cemented hip hemiarthroplasty, and to monitor infection and biofilm formation in real-time. Sprague-Dawley rats underwent cemented hip hemiarthroplasty via the posterior approach with pre- and postoperative gait assessments. Infection with Aims
Methods
We have evaluated the survivorship, outcomes, and failures of an interlocking, reconstruction-mode stem-sideplate implant used to preserve the native hip joint and achieve proximal fixation when there is little residual femur during large endoprosthetic reconstruction of the distal femur. A total of 14 patients underwent primary or revision reconstruction of a large femoral defect with a short remaining proximal femur using an interlocking, reconstruction-mode stem-sideplate for fixation after oncological distal femoral and diaphyseal resections. The implant was attached to a standard endoprosthetic reconstruction system. The implant was attached to a standard endoprosthetic reconstruction system. None of the femoral revisions were amenable to standard cemented or uncemented stem fixation. Patient and disease characteristics, surgical history, final ambulatory statusAims
Methods
Cementless total knee arthroplasty (TKA) offers the potential for strong biological fixation compared with cemented TKA where fixation is achieved by the mechanical integration of the cement. Few mid-term results are available for newer cementless TKA designs, which have used additive manufacturing (3D printing). The aim of this study was to present mid-term clinical outcomes and implant survivorship of the cementless Stryker Triathlon Tritanium TKA. This was a single institution registry review of prospectively gathered data from 341 cementless Triathlon Tritanium TKAs at four to 6.8 years follow-up. Outcomes were determined by comparing pre- and postoperative Knee Injury and Osteoarthritis Outcome Score for Joint Replacement (KOOS JR) scores, and pre- and postoperative 12-item Veterans RAND/Short Form Health Survey (VR/SF-12) scores. Aseptic loosening and revision for any reason were the endpoints which were used to determine survivorship at five years.Aims
Methods
The purpose of this study was to compare clinical results, long-term survival, and complication rates of stemless shoulder prosthesis with stemmed anatomical shoulder prostheses for treatment of osteoarthritis and to analyze radiological bone changes around the implants during follow-up. A total of 161 patients treated with either a stemmed or a stemless shoulder arthroplasty for primary osteoarthritis of the shoulder were evaluated with a mean follow-up of 118 months (102 to 158). The Constant score (CS), the Disabilities of the Arm, Shoulder and Hand (DASH) score, and active range of motion (ROM) were recorded. Radiological analysis for bone adaptations was performed by plain radiographs. A Kaplan-Meier survivorship analysis was calculated and complications were noted.Aims