Orthopaedic surgeons are currently faced with an overwhelming number of choices surrounding total knee arthroplasty (TKA), not only with the latest technologies and prostheses, but also fundamental decisions on alignment philosophies. From ‘mechanical’ to ‘adjusted mechanical’ to ‘restricted kinematic’ to ‘unrestricted kinematic’ — and how constitutional alignment relates to these — there is potential for ambiguity when thinking about and discussing such concepts. This annotation summarizes the various alignment strategies currently employed in TKA. It provides a clear framework and consistent language that will assist surgeons to compare confidently and contrast the concepts, while also discussing the latest opinions about alignment in TKA. Finally, it provides suggestions for applying consistent nomenclature to future research, especially as we explore the implications of 3D alignment patterns on patient outcomes. Cite this article:
Aims. A comprehensive classification for coronal lower limb alignment with predictive capabilities for knee balance would be beneficial in total knee arthroplasty (TKA). This paper describes the Coronal Plane Alignment of the Knee (CPAK) classification and examines its utility in preoperative soft tissue balance prediction, comparing kinematic alignment (KA) to mechanical alignment (MA). Methods. A radiological analysis of 500 healthy and 500 osteoarthritic (OA) knees was used to assess the applicability of the CPAK classification. CPAK comprises nine phenotypes based on the arithmetic HKA (aHKA) that estimates constitutional limb alignment and joint line obliquity (JLO). Intraoperative balance was compared within each phenotype in a cohort of 138 computer-assisted TKAs randomized to KA or MA. Primary outcomes included descriptive analyses of healthy and OA groups per CPAK type, and comparison of balance at 10° of flexion within each type. Secondary outcomes assessed balance at 45° and 90° and bone recuts required to achieve final knee balance within each CPAK type. Results. There was similar frequency distribution between healthy and arthritic groups across all CPAK types. The most common categories were Type II (39.2% healthy vs 32.2% OA), Type I (26.4% healthy vs 19.4% OA) and Type V (15.4% healthy vs 14.6% OA). CPAK Types VII, VIII, and IX were rare in both populations. Across all CPAK types, a greater proportion of KA TKAs achieved optimal balance compared to MA. This effect was largest, and statistically significant, in CPAK Types I (100% KA vs 15% MA; p < 0.001), Type II (78% KA vs 46% MA; p = 0.018). and Type IV (89% KA vs 0% MA; p < 0.001). Conclusion. CPAK is a pragmatic, comprehensive classification for coronal
This prospective randomised controlled trial was designed to
evaluate the outcome of both the MRI- and CT-based patient-specific
matched guides (PSG) from the same manufacturer. A total of 137 knees in 137 patients (50 men, 87 women) were
included, 67 in the MRI- and 70 in the CT-based PSG group. Their
mean age was 68.4 years (47.0 to 88.9). Outcome was expressed as
the biomechanical limb alignment (centre hip-knee-ankle: HKA-axis)
achieved post-operatively, the position of the individual components
within 3° of the pre-operatively planned alignment, correct planned
implant size and operative data (e.g. operating time and blood loss).Aims
Patients and Methods
Aims. The Coronal Plane Alignment of the Knee (CPAK) classification has been developed to predict individual variations in inherent
Aims. Porous metaphyseal cones can be used for fixation in revision total knee arthroplasty (rTKA) and complex TKAs. This metaphyseal fixation has led to some surgeons using shorter cemented stems instead of diaphyseal engaging cementless stems with a potential benefit of ease of obtaining proper alignment without being beholden to the diaphysis. The purpose of this study was to evaluate short term clinical and radiographic outcomes of a series of TKA cases performed using 3D-printed metaphyseal cones. Methods. A retrospective review of 86 rTKAs and nine complex primary TKAs, with an average age of 63.2 years (SD 8.2) and BMI of 34.0 kg/m. 2. (SD 8.7), in which metaphyseal cones were used for both femoral and tibial fixation were compared for their
Aims. The aim of this study was to compare robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) in order to determine the changes in the anatomy of the
The restoration of
As advancements in total knee arthroplasty progress at an exciting pace, two areas are of special interest, as they directly impact implant design and surgical decision making. Knee morphometry considers the three-dimensional shape of the articulating surfaces within the knee joint, and knee phenotyping provides the ability to categorize alignment into practical groupings that can be used in both clinical and research settings. This annotation discusses the details of these concepts, and the ways in which they are helping us better understand the individual subtleties of each patient’s knee. Cite this article:
The aim of this study was to investigate the distribution of phenotypes in Asian patients with end-stage osteoarthritis (OA) and assess whether the phenotype affected the clinical outcome and survival of mechanically aligned total knee arthroplasty (TKA). We also compared the survival of the group in which the phenotype unintentionally remained unchanged with those in which it was corrected to neutral. The study involved 945 TKAs, which were performed in 641 patients with primary OA, between January 2000 and January 2009. These were classified into 12 phenotypes based on the combined assessment of four categories of the arithmetic hip-knee-ankle angle and three categories of actual joint line obliquity. The rates of survival were analyzed using Kaplan-Meier methods and the log-rank test. The Hospital for Special Surgery score and survival of each phenotype were compared with those of the reference phenotype with neutral alignment and a parallel joint line. We also compared long-term survival between the unchanged phenotype group and the corrected to neutral alignment-parallel joint line group in patients with Type IV-b (mild to moderate varus alignment-parallel joint line) phenotype.Aims
Methods
The optimal management of the tibial slope in
achieving a high flexion angle in posterior-stabilised (PS) total
knee replacement (TKR) is not well understood, and most studies
evaluating the posterior tibial slope have been conducted on cruciate-retaining
TKRs. We analysed pre- and post-operative tibial slope differences,
pre- and post-operative coronal
We describe injuries to the posterior root of the medial meniscus in patients with spontaneous osteonecrosis of the medial compartment of the knee. We identified 30 consecutive patients with spontaneous osteonecrosis of the medial femoral condyle. The radiographs and MR imaging were reviewed. We found tears of the posterior root of the medial meniscus in 24 patients (80%). Of these, 15 were complete and nine were partial. Complete tears were associated with >
3 mm of meniscal extrusion. Neither the presence of a root tear nor the volume of the osteonecrotic lesion were associated with age, body mass index (BMI), gender, side affected, or
Anterior cruciate ligament (ACL) rupture commonly leads to post-traumatic osteoarthritis, regardless of surgical reconstruction. This study uses standing MRI to investigate changes in contact area, contact centroid location, and tibiofemoral alignment between ACL-injured knees and healthy controls, to examine the effect of ACL reconstruction on these parameters. An upright, open MRI was used to directly measure tibiofemoral contact area, centroid location, and alignment in 18 individuals with unilateral ACL rupture within the last five years. Eight participants had been treated nonoperatively and ten had ACL reconstruction performed within one year of injury. All participants were high-functioning and had returned to sport or recreational activities. Healthy contralateral knees served as controls. Participants were imaged in a standing posture with knees fully extended.Aims
Methods
The objective of this study was to compare the two-year migration pattern and clinical outcomes of a total knee arthroplasty (TKA) with an asymmetrical tibial design (Persona PS) and a well-proven TKA with a symmetrical tibial design (NexGen LPS). A randomized controlled radiostereometric analysis (RSA) trial was conducted including 75 cemented posterior-stabilized TKAs. Implant migration was measured with RSA. Maximum total point motion (MTPM), translations, rotations, clinical outcomes, and patient-reported outcome measures (PROMs) were assessed at one week postoperatively and at three, six, 12, and 24 months postoperatively.Aims
Methods
In the last decade, interest in partial knee arthroplasties and bicruciate retaining total knee arthroplasties has increased. In addition, patient-related outcomes and functional results such as range of movement and ambulation may be more promising with less invasive procedures such as bicompartmental arthroplasty (BCA). The purpose of this study is to evaluate clinical and radiological outcomes after a third-generation patellofemoral arthroplasty (PFA) combined with a medial or lateral unicompartmental knee arthroplasty (UKA) at mid- to long-term follow-up. A total of 57 procedures were performed. In 45 cases, a PFA was associated with a medial UKA and, in 12, with a lateral UKA. Patients were followed with validated patient-reported outcome measures (Oxford Knee Score (OKS), EuroQol five-dimension questionnaire (EQ-5D), EuroQoL Visual Analogue Scale (EQ-VAS)), the Knee Society Score (KSS), the Forgotten Joint Score (FJS), and radiological analysis.Aims
Methods
Alternative alignment concepts, including kinematic and restricted kinematic, have been introduced to help improve clinical outcomes following total knee arthroplasty (TKA). The purpose of this study was to evaluate the clinical results, along with patient satisfaction, following TKA using the concept of restricted kinematic alignment. A total of 121 consecutive TKAs performed between 11 February 2018 to 11 June 2019 with preoperative varus deformity were reviewed at minimum one-year follow-up. Three knees were excluded due to severe preoperative varus deformity greater than 15°, and a further three due to requiring revision surgery, leaving 109 patients and 115 knees to undergo primary TKA using the concept of restricted kinematic alignment with advanced technology. Patients were stratified into three groups based on the preoperative limb varus deformity: Group A with 1° to 5° varus (43 knees); Group B between 6° and 10° varus (56 knees); and Group C with varus greater than 10° (16 knees). This study group was compared with a matched cohort of 115 TKAs and 115 patients using a neutral mechanical alignment target with manual instruments performed from 24 October 2016 to 14 January 2019.Aims
Methods
Patients with a deformity of the hindfoot present a particular challenge when performing total knee arthroplasty (TKA). The literature contains little information about the relationship between TKA and hindfoot alignment. This systematic review aimed to determine from both clinical and radiological studies whether TKA would alter a preoperative hindfoot deformity and whether the outcome of TKA is affected by the presence of a postoperative hindfoot deformity. A systematic literature search was performed in the databases PubMed, EMBASE, Cochrane Library, and Web of Science. Search terms consisted of “total knee arthroplasty/replacement” combined with “hindfoot/ankle alignment”. Inclusion criteria were all English language studies analyzing the association between TKA and the alignment of the hindfoot, including the clinical or radiological outcomes. Exclusion criteria consisted of TKA performed with a concomitant extra-articular osteotomy and case reports or expert opinions. An assessment of quality was conducted using the modified Methodological Index for Non-Randomized Studies (MINORS). The review was performed according to the Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) guidelines and registered in the PROSPERO database (CRD42019106980).Aims
Methods
The results of kinematic total knee arthroplasty (KTKA) have been reported in terms of limb and component alignment parameters but not in terms of gap laxities and differentials. In kinematic alignment (KA), balance should reflect the asymmetrical balance of the normal knee, not the classic rectangular flexion and extension gaps sought with gap-balanced mechanical axis total knee arthroplasty (MATKA). This paper aims to address the following questions: 1) what factors determine coronal joint congruence as measured on standing radiographs?; 2) is flexion gap asymmetry produced with KA?; 3) does lateral flexion gap laxity affect outcomes?; 4) is lateral flexion gap laxity associated with lateral extension gap laxity?; and 5) can consistent ligament balance be produced without releases? A total of 192 KTKAs completed by a single surgeon using a computer-assisted technique were followed for a mean of 3.5 years (2 to 5). There were 116 male patients (60%) and 76 female patients (40%) with a mean age of 65 years (48 to 88). Outcome measures included intraoperative gap laxity measurements and component positions, as well as joint angles from postoperative three-foot standing radiographs. Patient-reported outcome measures (PROMs) were analyzed in terms of alignment and balance: EuroQol (EQ)-5D visual analogue scale (VAS), Knee Injury and Osteoarthritis Outcome Score (KOOS), KOOS Joint Replacement (JR), and Oxford Knee Score (OKS).Aims
Patients and Methods
We compared lower limb coronal alignment measurements
obtained pre- and post-operatively with long-leg radiographs and
computer navigation in patients undergoing primary total knee replacement
(TKR). A series of 185 patients had their pre- and post-implant
radiological and computer-navigation system measurements of coronal alignment
compared using the Bland-Altman method. The study included 81 men
and 104 women with a mean age of 68.5 years (32 to 87) and a mean
body mass index of 31.7 kg/m. 2. (19 to 49). Pre-implant
Bland–Altman limits of agreement were -9.4° to 8.6° with a repeatability
coefficient of 9.0°. The Bland–Altman plot showed a tendency for the
radiological measurement to indicate a higher level of pre-operative
deformity than the corresponding navigation measurement. Post-implant
limits of agreement were -5.0° to 5.4° with a repeatability coefficient
of 5.2°. The tendency for valgus knees to have greater deformity
on the radiograph was still seen, but was weaker for varus
Robotic-assisted unicompartmental knee arthroplasty (UKA) promises accurate implant placement with the potential of improved survival and functional outcomes. The aim of this study was to present the current evidence for robotic-assisted UKA and describe the outcome in terms of implant positioning, range of movement (ROM), function and survival, and the types of robot and implants that are currently used. A search of PubMed and Medline was performed in October 2018 in line with the Preferred Reporting Items for Systematic Review and Meta-Analysis statement. Search terms included “robotic”, “knee”, and “surgery”. The criteria for inclusion was any study describing the use of robotic UKA and reporting implant positioning, ROM, function, and survival for clinical, cadaveric, or dry bone studies.Aims
Materials and Methods
The use of vitamin E-infused highly crosslinked polyethylene (HXLPE) in total knee prostheses is controversial. In this paper we have compared the clinical and radiological results between conventional polyethylene and vitamin E-infused HXLPE inserts in total knee arthroplasty (TKA). The study included 200 knees (175 patients) that underwent TKA using the same total knee prostheses. In all, 100 knees (77 patients) had a vitamin E-infused HXLPE insert (study group) and 100 knees (98 patients) had a conventional polyethylene insert (control group). There were no significant differences in age, sex, diagnosis, preoperative knee range of movement (ROM), and preoperative Knee Society Score (KSS) between the two groups. Clinical and radiological results were evaluated at two years postoperatively.Aims
Patients and Methods
The aim of this study was to assess the effect of posterior cruciate ligament (PCL) resection on flexion-extension gaps, mediolateral soft-tissue laxity, fixed flexion deformity (FFD), and limb alignment during posterior-stabilized (PS) total knee arthroplasty (TKA). This prospective study included 110 patients with symptomatic osteoarthritis of the knee undergoing primary robot-assisted PS TKA. All operations were performed by a single surgeon using a standard medial parapatellar approach. Optical motion capture technology with fixed femoral and tibial registration pins was used to assess gaps before and after PCL resection in extension and 90° knee flexion. Measurements were made after excision of the anterior cruciate ligament and prior to bone resection. There were 54 men (49.1%) and 56 women (50.9%) with a mean age of 68 years (Aims
Patients and Methods
The role of modular tibial implants in total knee replacement is not fully defined. We performed a prospective randomised controlled clinical trial using radiostereophotogrammetric analysis to compare the performance of an all-polyethylene tibia with a metal-backed cruciate-retaining condylar design, PFC-∑ total knee replacement for up to 24 months. There were 51 patients who were randomised into two treatment groups. There were 10 subsequent withdrawals, leaving 21 all-polyethylene and 20 metal-backed tibial implants. No patient was lost to follow-up. There were no significant demographic differences between the groups. At two years one metal-backed implant showed migration >
1 mm, but no polyethylene implant reached this level. There was a significant increase in the SF-12 and Oxford knee scores after operation in both groups. In an uncomplicated primary total knee replacement the all-polyethylene PFC-∑ tibial prosthesis showed no statistical difference in migration from that of the metal-backed counterpart. There was no difference in the clinical results as assessed by the SF-12, the Oxford
To assess the effect of high tibial and distal femoral osteotomies
(HTO and DFO) on the pressure characteristics of the ankle joint. Varus and valgus malalignment of the knee was simulated in human
cadaver full-length legs. Testing included four measurements: baseline
malalignment, 5° and 10° re-aligning osteotomy, and control baseline
malalignment. For HTO, testing was rerun with the subtalar joint
fixed. In order to represent half body weight, a 300 N force was applied
onto the femoral head. Intra-articular sensors captured ankle pressure.Aims
Materials and Methods
In arthritis of the varus knee, a high tibial
osteotomy (HTO) redistributes load from the diseased medial compartment
to the unaffected lateral compartment. We report the outcome of 36 patients (33 men and three women)
with 42 varus, arthritic knees who underwent HTO and dynamic correction
using a Garches external fixator until they felt that normal alignment
had been restored. The mean age of the patients was 54.11 years
(34 to 68). Normal alignment was achieved at a mean 5.5 weeks (3
to 10) post-operatively. Radiographs, gait analysis and visual analogue
scores for pain were measured pre- and post-operatively, at one
year and at medium-term follow-up (mean six years; 2 to 10). Failure
was defined as conversion to knee arthroplasty. Pre-operative gait analysis divided the 42 knees into two equal
groups with high (17 patients) or low (19 patients) adductor moments.
After correction, a statistically significant (p <
0.001, At final follow-up, after a mean of 15.9 years (12 to 20), there
was a survivorship of 59% (95% CI 59.6 to 68.9) irrespective of
adductor moment group, with a mean time to conversion to knee arthroplasty
of 9.5 years (3 to 18; 95% confidence interval ± 2.5). HTO remains a useful option in the medium-term for the treatment
of medial compartment osteoarthritis of the knee but does not last
in the long-term. Cite this article:
We wished to examine the effectiveness of tibial lengthening
using a two ring Ilizarov frame in skeletally immature patients.
This is a potentially biomechanically unstable construct which risks
the loss of axial control. We retrospectively reviewed a consecutive series of 24 boys and
26 girls, with a mean age of 8.6 years (4 to 14), who underwent
52 tibial lengthening procedures with a mean follow-up of 4.3 years
(4.0 to 16.9). Tibial alignment was measured before and after treatment
using joint orientation lines from the knee and a calculation of
the oblique plane axis.Aims
Patients and Methods
Many aspects of total knee arthroplasty have
changed since its inception. Modern prosthetic design, better fixation techniques,
improved polyethylene wear characteristics and rehabilitation, have
all contributed to a large change in revision rates. Arthroplasty
patients now expect longevity of their prostheses and demand functional
improvement to match. This has led to a re-examination of the long-held
belief that mechanical alignment is instrumental to a successful
outcome and a focus on restoring healthy joint kinematics. A combination
of kinematic restoration and uncemented, adaptable fixation may
hold the key to future advances. Cite this article:
Total knee replacement (TKR) smart tibial trials
have load-bearing sensors which will show quantitative compartment
pressure values and femoral-tibial tracking patterns. Without smart
trials, surgeons rely on feel and visual estimation of imbalance
to determine if the knee is optimally balanced. Corrective soft-tissue
releases are performed with minimal feedback as to what and how
much should be released. The smart tibial trials demonstrate graphically
where and how much imbalance is present, so that incremental releases
can be performed. The smart tibial trials now also incorporate accelerometers
which demonstrate the axial alignment. This now allows the surgeon
the option to perform a slight recut of the tibia or femur to provide
soft-tissue balance without performing soft-tissue releases. Using
a smart tibial trial to assist with soft-tissue releases or bone
re-cuts, improved patient outcomes have been demonstrated at one
year in a multicentre study of 135 patients (135 knees). Cite this article:
We have previously reported the short-term radiological
results of a randomised controlled trial comparing kinematically
aligned total knee replacement (TKR) and mechanically aligned TKR,
along with early pain and function scores. In this study we report
the two-year clinical results from this trial. A total of 88 patients
(88 knees) were randomly allocated to undergo either kinematically
aligned TKR using patient-specific guides, or mechanically aligned
TKR using conventional instruments. They were analysed on an intention-to-treat
basis. The patients and the clinical evaluator were blinded to the
method of alignment. At a minimum of two years, all outcomes were better for the kinematically
aligned group, as determined by the mean Oxford knee score (40 (15
to 48) In this study, the use of a kinematic alignment technique performed
with patient-specific guides provided better pain relief and restored
better function and range of movement than the mechanical alignment
technique performed with conventional instruments. Cite this article:
Analysis of the morphology of the distal femur, and by extension
of the femoral components in total knee arthroplasty (TKA), has
largely been related to the aspect ratio, which represents the width
of the femur. Little is known about variations in trapezoidicity
(i.e. whether the femur is more rectangular or more trapezoidal).
This study aimed to quantify additional morphological characteristics
of the distal femur and identify anatomical features associated
with higher risks of over- or under-sizing of components in TKA. We analysed the shape of 114 arthritic knees at the time of primary
TKA using the pre-operative CT scans. The aspect ratio and trapezoidicity
ratio were quantified, and the post-operative prosthetic overhang
was calculated. We compared the morphological characteristics with
those of 12 TKA models.Aims
Methods
Primary total knee arthroplasty (TKA) is a reliable
procedure with reproducible long-term results. Nevertheless, there
are conditions related to the type of patient or local conditions
of the knee that can make it a difficult procedure. The most common
scenarios that make it difficult are discussed in this review. These
include patients with many previous operations and incisions, and
those with severe coronal deformities, genu recurvatum, a stiff knee,
extra-articular deformities and those who have previously undergone
osteotomy around the knee and those with chronic dislocation of
the patella. Each condition is analysed according to the characteristics of
the patient, the pre-operative planning and the reported outcomes. When approaching the difficult primary TKA surgeons should use
a systematic approach, which begins with the review of the existing
literature for each specific clinical situation. Cite this article:
Smart trials are total knee tibial trial liners
with load bearing and alignment sensors that will graphically show quantitative
compartment load-bearing forces and component track patterns. These
values will demonstrate asymmetrical ligament balancing and misalignments
with the medial retinaculum temporarily closed. Currently surgeons
use feel and visual estimation of imbalance to assess soft-tissue
balancing and tracking with the medial retinaculum open, which results
in lower medial compartment loads and a wider anteroposterior tibial
tracking pattern. The sensor trial will aid the total knee replacement
surgeon in performing soft-tissue balancing by providing quantitative
visual feedback of changes in forces while performing the releases
incrementally. Initial experience using a smart tibial trial is
presented.
Treatment for osteoarthritis (OA) has traditionally
focused on joint replacement for end-stage disease. An increasing number
of surgical and pharmaceutical strategies for disease prevention
have now been proposed. However, these require the ability to identify
OA at a stage when it is potentially reversible, and detect small
changes in cartilage structure and function to enable treatment
efficacy to be evaluated within an acceptable timeframe. This has
not been possible using conventional imaging techniques but recent
advances in musculoskeletal imaging have been significant. In this
review we discuss the role of different imaging modalities in the
diagnosis of the earliest changes of OA. The increasing number of
MRI sequences that are able to non-invasively detect biochemical
changes in cartilage that precede structural damage may offer a
great advance in the diagnosis and treatment of this debilitating
condition. Cite this article:
Patient-specific cutting guides (PSCGs) are designed
to improve the accuracy of alignment of total knee replacement (TKR).
We compared the accuracy of limb alignment and component positioning
after TKR performed using PSCGs or conventional instrumentation.
A total of 80 patients were randomised to undergo TKR with either of
the different forms of instrumentation, and radiological outcomes
and peri-operative factors such as operating time were assessed.
No significant difference was observed between the groups in terms
of tibiofemoral angle or femoral component alignment. Although the
tibial component in the PSCGs group was measurably closer to neutral
alignment than in the conventional group, the size of the difference
was very small (89.8° ( Cite this article:
The aim of this study was to examine the loading
of the other joints of the lower limb in patients with unilateral osteoarthritis
(OA) of the knee. We recruited 20 patients with no other symptoms
or deformity in the lower limbs from a consecutive cohort of patients
awaiting knee replacement. Gait analysis and electromyographic recordings were
performed to determine moments at both knees and hips, and contraction
patterns in the medial and lateral quadriceps and hamstrings bilaterally.
The speed of gait was reduced in the group with OA compared with
the controls, but there were only minor differences in stance times
between the limbs. Patients with OA of the knee had significant
increases in adduction moment impulse at both knees and the contralateral
hip (adjusted p-values: affected knee: p <
0.01, unaffected knee
p = 0.048, contralateral hip p = 0.03), and significantly increased
muscular co-contraction bilaterally compared with controls (all
comparisons for co-contraction, p <
0.01). The other major weight-bearing joints are at risk from abnormal
biomechanics in patients with unilateral OA of the knee. Cite this article:
The Oxford mobile-bearing unicompartmental knee
replacement (UKR) is an effective and safe treatment for osteoarthritis
of the medial compartment. The results in the lateral compartment
have been disappointing due to a high early rate of dislocation
of the bearing. A series using a newly designed domed tibial component
is reported. The first 50 consecutive domed lateral Oxford UKRs in 50 patients
with a mean follow-up of three years (2.0 to 4.3) were included.
Clinical scores were obtained prospectively and Kaplan-Meier survival
analysis was performed for different endpoints. Radiological variables
related to the position and alignment of the components were measured. One patient died and none was lost to follow-up. The cumulative
incidence of dislocation was 6.2% (95% confidence interval (CI)
2.0 to 17.9) at three years. Survival using revision for any reason
and aseptic revision was 94% (95% CI 82 to 98) and 96% (95% CI 85
to 99) at three years, respectively. Outcome scores, visual analogue
scale for pain and maximum knee flexion showed a significant improvement
(p <
0.001). The mean Oxford knee score was 43 ( Clinical results are excellent and short-term survival has improved
when compared with earlier series. The risk of dislocation remains
higher using a mobile-bearing UKR in the lateral compartment when
compared with the medial compartment. Patients should be informed
about this complication. To avoid dislocations, care must be taken
not to elevate the lateral joint line.
The role of computer-assisted surgery in maintaining the level of the joint in primary knee joint replacement (TKR) has not been well defined. We undertook a blinded randomised controlled trial comparing joint-line maintenance, functional outcomes, and quality-of-life outcomes between patients undergoing computer-assisted and conventional TKR. A total of 115 patients were randomised (computer-assisted, n = 55; conventional, n = 60). Two years post-operatively no significant correlation was found between computer-assisted and conventional surgery in terms of maintaining the joint line. Those TKRs where the joint line was depressed post-operatively improved the least in terms of functional scores. No difference was detected in terms of quality-of-life outcomes. Change in joint line was found to be related to change in alignment. Change in alignment significantly affects change in joint line and functional scores.
Valgus high tibial osteotomy for osteoarthritis of the medial compartment of the knee can be performed using medial opening- and lateral closing-wedge techniques. The latter have been thought to offer greater initial stability. We measured and compared the stability of opening- and closing-wedge osteotomies fixed by TomoFix plates using radiostereometry in a series of 42 patients in a prospective, randomised clinical trial. There were no differences between the opening- and closing-wedge groups in the time to regain knee function and full weight-bearing. Pain and knee function were significantly improved in both groups without any differences between them. All the osteotomies united within one year. Radiostereometry showed no clinically relevant movement of bone or differences between either group. Medial opening-wedge high tibial osteotomy secured by a TomoFix plate offers equal stability to a lateral closing-wedge technique. Both give excellent initial stability and provide significantly improved knee function and reduction in pain, although the opening-wedge technique was more likely to produce the intended correction.
The purpose of this study was to determine objectively the outcome of total knee replacement in patients with ankylosed knees. There were 82 patients (99 knees) with ankylosed knees who underwent total knee replacement with a condylar constrained or a posterior stabilised prosthesis. Their mean age was 41.9 years (23 to 60) and the mean follow-up was for 8.9 years (6.6 to 14). Pre- and post-operative data included the Hospital for Special Surgery (HSS), the Knee Society (KS) and the Western Ontario and McMaster University Osteoarthritis index (WOMAC) scores. The mean HSS, KS and WOMAC scores improved from 60, 53, and 79 pre-operatively to 81, 85, and 37 at follow-up. These improvements were statistically significant (p = 0.018, 0.001 and 0.014 respectively). The mean physical, social and emotional WOMAC scores also improved significantly (p = 0.032, p = 0.023 and p <
0.001 respectively). The mean satisfaction score was 8.5 ( Total knee replacement gives good mid-term results in patients with ankylosed knees.
We performed a prospective, randomised study to compare the results and rates of complications of primary total knee replacement performed using a quadriceps-sparing technique or a standard arthrotomy in 120 patients who had bilateral total knee replacements carried out under the same anaesthetic. The clinical results, pain scales, surgical and hospital data, post-operative complications and radiological results were compared. No significant differences were found between the two groups with respect to the blood loss, knee score, function score, pain scale, range of movement or radiological findings. In contrast, the operating time (p = 0.0001) and the tourniquet time (p <
0.0001) were significantly longer in the quadriceps-sparing group, as was the rate of complications (p = 0.0468). We therefore recommend the use of a standard arthrotomy with the shortest possible skin incision for total knee replacement.
The aim of this study was to compare the results in patients having a quadriceps sparing total knee replacement (TKR) with those undergoing a standard TKR at a minimum follow-up of two years. All patients who had a TKR with a high-flex posterior-stabilised prosthesis prior to December 2002 were reviewed retrospectively. There were 57 patients available for follow-up. Those with a quadriceps sparing TKR had less pain peri-operatively with a greater degree of flexion at all the post-operative visits and at the final follow-up, but their operations took longer, with less accurate radiological alignment. There was no difference in the complications and in the Knee Society scores between the two groups at the final follow-up. Total knee replacement through a quadriceps sparing approach has some peri-operative advantages over the standard incision. At a minimum follow-up of two years the clinical results were similar to those with a standard incision, but the radiological outcomes of the quadriceps sparing group were inferior.
A series of 100 consecutive osteoarthritic patients was randomised to undergo total knee replacement using a Miller-Galante II prosthesis, with or without a cemented polyethylene patellar component. Knee function was evaluated using the American Knee Society score, Western Ontario and McMaster University Osteoarthritis index, specific patellofemoral-related questions and radiographic evaluation until the fourth post-operative year, then via questionnaire until ten years post-operatively. A ten-point difference in the American Knee Society score between the two groups was considered a significant change in knee performance, with α and β levels of 0.05. The mean age of the patients in the resurfaced group was 71 years (53 to 88) and in the non-resurfaced group was 73 years (54 to 86). After ten years 22 patients had died, seven were suffering from dementia, three declined further participation and ten were lost to follow-up. Two patients in the non-resurfaced group subsequently had their patellae resurfaced. In the resurfaced group one patient had an arthroscopic lateral release. There was no significant difference between the two treatment groups: both had a similar deterioration of scores with time, and no further patellofemoral complications were observed in either group. We are unable to recommend routine patellar resurfacing in osteoarthritic patients undergoing total knee replacement on the basis of our findings.
We compared the alignment of 39 total knee replacements implanted using the conventional alignment guide system with 37 implanted using a CT-based navigation system, performed by a single surgeon. The knees were evaluated using full-length weight-bearing anteroposterior radiographs, lateral radiographs and CT scans. The mean hip-knee-ankle angle, coronal femoral component angle and coronal tibial component angle were 181.8° (174.2° to 188.3°), 88.5° (84.0° to 91.8°) and 89.7° (86.3° to 95.1°), respectively for the conventional group and 180.8° (178.2° to 185.1°), 89.3° (85.8° to 92.0°) and 89.9° (88.0° to 93.0°), respectively for the navigated group. The mean sagittal femoral component angle was 85.5° (80.6° to 92.8°) for the conventional group and 89.6° (85.5° to 94.0°) for the navigated group. The mean rotational femoral and tibial component angles were −0.7° (−8.8° to 9.8°) and −3.3° (−16.8° to 5.8°) for the conventional group and −0.6° (−3.5° to 3.0°) and 0.3° (−5.3° to 7.7°) for the navigated group. The ideal angles of all alignments in the navigated group were obtained at significantly higher rates than in the conventional group. Our results demonstrated significant improvements in component positioning with a CT-based navigation system, especially with respect to rotational alignment.
Bilateral sequential total knee replacement was carried out under one anaesthetic in 100 patients. One knee was replaced using a CT-free computer-assisted navigation system and the other conventionally without navigation. The two methods were compared for accuracy of orientation and alignment of the components. There were 85 women and 15 men with a mean age of 67.6 years (54 to 83). Radiological and CT imaging was carried out to determine the alignment of the components. The mean follow-up was 2.3 years (2 to 3). The operating and tourniquet times were significantly longer in the navigation group (p <
0.001). There were no significant pre- or post-operative differences between the knee scores of the two groups (p = 0.288 and p = 0.429, respectively). The results of imaging and the number of outliers for all radiological parameters were not statistically different (p = 0.109 to p = 0.920). In this series computer-assisted navigated total knee replacement did not result in more accurate orientation and alignment of the components than that achieved by conventional total knee replacement.
Between 1996 and 2003 six institutions in the United States and France contributed a consecutive series of 234 fractures of the femur in 229 children which were treated by titanium elastic nailing. Minor or major complications occurred in 80 fractures. Full information was available concerning 230 fractures, of which the outcome was excellent in 150 (65%), satisfactory in 57 (25%), and poor in 23 (10%). Poor outcomes were due to leg-length discrepancy in five fractures, unacceptable angulation in 17, and failure of fixation in one. There was a statistically significant relationship (p = 0.003) between age and outcome, and the odds ratio for poor outcome was 3.86 for children aged 11 years and older compared with those below this age. The difference between the weight of children with a poor outcome and those with an excellent or satisfactory outcome was statistically significant (54 kg