We assessed the orientation of the acetabular
component in 1070 primary total hip arthroplasties with hard-on-soft, small
diameter bearings, aiming to determine the size and site of the
target zone that optimises outcome. Outcome measures included complications,
dislocations, revisions and ΔOHS (the difference between the Oxford
Hip Scores pre-operatively and five years post-operatively). A wide
scatter of orientation was observed (2 This study demonstrated that with traditional technology surgeons
can only reliably achieve a target zone of ±15°. As the optimal
zone to diminish the risk of dislocation is also ±15°, surgeons
should be able to achieve this. This is the first study to demonstrate
that optimal orientation of the acetabular component improves the
functional outcome. However, the target zone is small (± 5°) and
cannot, with current technology, be consistently achieved. Cite this article:
Aims. There has been limited literature regarding outcomes of acetabular rim syndrome (ARS) with persistent acetabular os in the setting of acetabular dysplasia. The purpose of this study was to characterize a cohort of adolescent and young adult patients with ARS with persistent os and compare their radiological and clinical outcomes to patients with acetabular dysplasia without an os. Methods. We reviewed a prospective database of patients undergoing periacetabular osteotomy (PAO) for symptomatic acetabular dysplasia between January 1999 and December 2021 to identify hips with preoperative os acetabuli, defined as a closed triradiate cartilage but persistence of a superolateral os acetabulum. A total of 14 hips in 12 patients with persistent os acetabuli (ARS cohort) were compared to 50 randomly selected ‘control’ hips without persistent os acetabuli. Preoperative and postoperative radiographs were measured for markers of dysplasia: lateral centre-edge angle, anterior centre-edge angle,
Aims. The aims of this study were to compare clinically relevant measurements of hip dysplasia on radiographs taken in the supine and standing position, and to compare Hip2Norm software and Picture Archiving and Communication System (PACS)-derived digital radiological measurements. Methods. Preoperative supine and standing radiographs of 36 consecutive patients (43 hips) who underwent periacetabular osteotomy surgery were retrospectively analyzed from a single-centre, two-surgeon cohort. Anterior coverage (AC), posterior coverage (PC), lateral centre-edge angle (LCEA),
Aims. Femoral component anteversion is an important factor in the success of total hip arthroplasty (THA). This retrospective study aimed to investigate the accuracy of femoral component anteversion with the Mako THA system and software using the Exeter cemented femoral component, compared to the Accolade II cementless femoral component. Methods. We reviewed the data of 30 hips from 24 patients who underwent THA using the posterior approach with Exeter femoral components, and 30 hips from 24 patients with Accolade II components. Both groups did not differ significantly in age, sex, BMI, bone quality, or disease. Two weeks postoperatively, CT images were obtained to measure acetabular and femoral component anteversion. Results. The mean difference in femoral component anteversion between intraoperative and postoperative CT measurements (system accuracy of component anteversion) was 0.8° (SD 1.8°) in the Exeter group and 2.1° (SD 2.3°) in the Accolade II group, respectively (p = 0.020). The mean difference in anteversion between the plan and the postoperative CT measurements (clinical accuracy of femoral component anteversion) was 1.2° (SD 3.6°) in the Exeter group, and 4.2° (SD 3.9°) in the Accolade II group (p = 0.003). No significant differences were found in
Aims. The aims of this study were to examine the rate at which the
positioning of the acetabular component, leg length discrepancy
and femoral offset are outside an acceptable range in total hip
arthroplasties (THAs) which either do or do not involve the use
of intra-operative digital imaging. Patients and Methods. A retrospective case-control study was undertaken with 50 patients
before and 50 patients after the integration of an intra-operative
digital imaging system in THA. The demographics of the two groups
were comparable for body mass index, age, laterality and the indication
for surgery. The digital imaging group had more men than the group without.
Surgical data and radiographic parameters, including the inclination
and anteversion of the acetabular component, leg length discrepancy,
and the difference in femoral offset compared with the contralateral
hip were collected and compared, as well as the incidence of altering
the position of a component based on the intra-operative image. Results. Digital imaging took a mean of five minutes (2.3 to 14.6) to
perform. Intra-operative changes with the use of digital imaging
were made for 43 patients (86%), most commonly to adjust leg length
and femoral offset. There was a decrease in the incidence of outliers
when using intra-operative imaging compared with not using it in
regard to leg length discrepancy (20% versus 52%,
p = 0.001) and femoral offset inequality (18% versus 44%,
p = 0.004). There was also a difference in the incidence of outliers
in
Aims. Vitamin E-diffused, highly crosslinked polyethylene (VEPE) and porous titanium-coated (PTC) shells were introduced in total hip arthroplasty (THA) to reduce the risk of aseptic loosening. The purpose of this study was: 1) to compare the wear properties of VEPE to moderately crosslinked polyethylene; 2) to assess the stability of PTC shells; and 3) to report their clinical outcomes at seven years. Patients and Methods. A total of 89 patients were enrolled into a prospective study. All patients received a PTC shell and were randomized to receive a VEPE liner (n = 44) or a moderately crosslinked polyethylene (ModXLPE) liner (n = 45). Radiostereometric analysis (RSA) was used to measure polyethylene wear and component migration. Differences in wear were assessed while adjusting for body mass index, activity level,
Aims. Optimal exposure through the direct anterior approach (DAA) for total hip arthroplasty (THA) conducted on a regular operating theatre table is achieved with a standardized capsular releasing sequence in which the anterior capsule can be preserved or resected. We hypothesized that clinical outcomes and implant positioning would not be different in case a capsular sparing (CS) technique would be compared to capsular resection (CR). Methods. In this prospective trial, 219 hips in 190 patients were randomized to either the CS (n = 104) or CR (n = 115) cohort. In the CS cohort, a medial based anterior flap was created and sutured back in place at the end of the procedure. The anterior capsule was resected in the CR cohort. Primary outcome was defined as the difference in patient-reported outcome measures (PROMs) after one year. PROMs (Harris Hip Score (HHS), Hip disability and Osteoarthritis Outcome Score (HOOS), and Short Form 36 Item Health Survey (SF-36)) were collected preoperatively and one year postoperatively. Radiological parameters were analyzed to assess implant positioning and implant ingrowth. Adverse events were monitored. Results. At one year, there was no difference in HSS (p = 0.728), HOOS (Activity Daily Life, p = 0.347; Pain, p = 0.982; Quality of Life, p = 0.653; Sport, p = 0.994; Symptom, p = 0.459), or SF-36 (p = 0.338).
We present a new CT-based method which measures cover of the femoral head in both normal and dysplastic hips and allows assessment of
We measured the orientation of the acetabular and femoral components in 45 patients (33 men, 12 women) with a mean age of 53.4 years (30 to 74) who had undergone revision of metal-on-metal hip resurfacings. Three-dimensional CT was used to measure the inclination and version of the acetabular component, femoral version and the horizontal femoral offset, and the linear wear of the removed acetabular components was measured using a roundness machine. We found that acetabular version and combined version of the acetabular and femoral components were weakly positively correlated with the rate of wear. The
Aims. The purpose of this study was to compare the clinical, radiological, and patient-reported outcome measures (PROMs) in the first 100 consecutive patients undergoing total hip arthroplasty (THA) via a direct superior approach (DSA) with a matched group of patients undergoing THA by the same surgeon, using a posterolateral approach (PLA). Methods. This was a retrospective single surgeon study comparing the first 100 consecutive DSA THA patients with a matched group of patients using a standard PLA. Case notes were examined for patient demographics, length of hospital stay, operating time, intra- and postoperative complications, pain score, satisfaction score, and Oxford Hip Score (OHS). Leg length discrepancy and component positioning were measured from postoperative plain radiographs. Results. The DSA patients had a shorter length of hospital stay (mean 2.09 days (SD 1.20) DSA vs 2.74 days (SD 1.17) PLA; p < 0.001) and shorter time to discharge from the inpatient physiotherapy teams (mean 1.44 days (SD 1.17) DSA vs 1.93 days (SD 0.96) PLA; p < 0.001). There were no differences in operating time (p = 0.505), pain levels up to postoperative day 1 (p = 0.106 to p =0.242), OHS (p = 0.594 to p = 0.815), satisfaction levels (p = 0.066 to p = 0.299), stem alignment (p = 0.240),
We report the clinical and radiological outcome of 86 revisions of cemented hip arthroplasties using JRI-Furlong hydroxyapatite-ceramic-coated acetabular and femoral components. The acetabular component was revised in 62 hips and the femoral component in all hips. The mean follow-up was 12.6 years and no patient was lost to follow-up. The mean age of the patients was 71.2 years. The mean Harris hip and Oxford scores were 82 (59 to 96) and 23.4 (14 to 40), respectively. The mean Charnley modification of the Merle d’Aubigné and Postel score was 5 (3 to 6) for pain, 4.9 (3 to 6) for movement and 4.4 (3 to 6) for mobility. Migration of the acetabular component was seen in two hips and the mean
Between December 2004 and June 2006, 136 patients (156 total hip replacements), were sent from the waiting list of the Cardiff Vale NHS Trust to the NHS Treatment Centre, Weston-super-Mare, in an attempt to reduce the waiting time for total hip replacement. Because of concerns about their outcome, each patient was contacted and invited to attend a review appointment with a consultant specialising in hip and revision hip replacement. A total of 98 patients (113 hips) were reviewed after a mean of 23 months (11 to 30). There were 104 cemented hips, seven hybrid and two cementless. An
Aims. The aim of this study was to evaluate the accuracy of implant placement when using robotic assistance during total hip arthroplasty (THA). Patients and Methods. A total of 20 patients underwent a planned THA using preoperative CT scans and robotic-assisted software. There were nine men and 11 women (n = 20 hips) with a mean age of 60.8 years (. sd. 6.0). Pelvic and femoral bone models were constructed by segmenting both preoperative and postoperative CT scan images. The preoperative anatomical landmarks using the robotic-assisted system were matched to the postoperative 3D reconstructions of the pelvis. Acetabular and femoral component positions as measured intraoperatively and postoperatively were evaluated and compared. Results. The system reported accurate values for reconstruction of the hip when compared to those measured postoperatively using CT. The mean deviation from the executed overall hip length and offset were 1.6 mm (. sd. 2.9) and 0.5 mm (. sd. 3.0), respectively. Mean combined anteversion was similar and correlated between intraoperative measurements and postoperative CT measurements (32.5°, . sd. 5.9° versus 32.2°, . sd. 6.4°; respectively; R. 2. = 0.65; p < 0.001). There was a significant correlation between mean intraoperative (40.4°, . sd. 2.1°)
The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate. A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.Aims
Methods
The primary objective of this study was to develop a validated classification system for assessing iatrogenic bone trauma and soft-tissue injury during total hip arthroplasty (THA). The secondary objective was to compare macroscopic bone trauma and soft-tissues injury in conventional THA (CO THA) versus robotic arm-assisted THA (RO THA) using this classification system. This study included 30 CO THAs versus 30 RO THAs performed by a single surgeon. Intraoperative photographs of the osseous acetabulum and periacetabular soft-tissues were obtained prior to implantation of the acetabular component, which were used to develop the proposed classification system. Interobserver and intraobserver variabilities of the proposed classification system were assessed.Aims
Methods
The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years. The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.Aims
Methods
Modular dual-mobility (DM) articulations are increasingly used during total hip arthroplasty (THA). However, concerns remain regarding the metal liner modularity. This study aims to correlate metal artifact reduction sequence (MARS)-MRI abnormalities with serum metal ion levels in patients with DM articulations. A total of 45 patients (50 hips) with a modular DM articulation were included with mean follow-up of 3.7 years (SD 1.2). Enrolled patients with an asymptomatic, primary THA and DM articulation with over two years’ follow-up underwent MARS-MRI. Each patient had serum cobalt, chromium, and titanium levels drawn. Patient satisfaction, Oxford Hip Score, and Forgotten Joint Score-12 (FJS-12) were collected. Each MARS-MRI was independently reviewed by fellowship-trained musculoskeletal radiologists blinded to serum ion levels.Aims
Methods
Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct. We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m2 (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97).Aims
Methods
The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface. Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up.Aims
Methods
The primary outcome was investigating differences in wear, as measured by femoral head penetration, between cross-linked vitamin E-diffused polyethylene (vE-PE) and cross-linked polyethylene (XLPE) acetabular component liners and between 32 and 36 mm head sizes at the ten-year follow-up. Secondary outcomes included acetabular component migration and patient-reported outcome measures (PROMs) such as the EuroQol five-dimension questionnaire, 36-Item Short-Form Health Survey, Harris Hip Score, and University of California, Los Angeles Activity Scale (UCLA). A single-blinded, multi-arm, 2 × 2 factorial randomized controlled trial was undertaken. Patients were recruited between May 2009 and April 2011. Radiostereometric analyses (RSAs) were performed from baseline to ten years. Of the 220 eligible patients, 116 underwent randomization, and 82 remained at the ten-year follow-up. Eligible patients were randomized into one of four interventions: vE-PE acetabular liner with either 32 or 36 mm femoral head, and XLPE acetabular liner with either 32 or 36 mm femoral head. Parameters were otherwise identical except for acetabular liner material and femoral head size.Aims
Methods
The primary purpose of this study of metal-on-metal
(MoM) hip resurfacing was to compare the effect of using a cementless
or cemented femoral component on the subsequent bone mineral density
(BMD) of the femoral neck. . This was a single-centre, prospective, double-blinded control
trial which randomised 120 patients (105 men and 15 women) with
a mean age of 49.4 years (21 to 68) to receive either a cemented
or cementless femoral component. Follow-up was to two years. Outcome
measures included total and six-point region-of-interest BMD of
the femoral neck, radiological measurements of
Achieving accurate implant positioning and restoring native hip biomechanics are key surgeon-controlled technical objectives in total hip arthroplasty (THA). The primary objective of this study was to compare the reproducibility of the planned preoperative centre of hip rotation (COR) in patients undergoing robotic arm-assisted THA versus conventional THA. This prospective randomized controlled trial (RCT) included 60 patients with symptomatic hip osteoarthritis undergoing conventional THA (CO THA) versus robotic arm-assisted THA (RO THA). Patients in both arms underwent pre- and postoperative CT scans, and a patient-specific plan was created using the robotic software. The COR, combined offset, acetabular orientation, and leg length discrepancy were measured on the pre- and postoperative CT scanogram at six weeks following surgery.Aims
Methods
There is little information in the literature about the use of dual-mobility (DM) bearings in preventing re-dislocation in revision total hip arthroplasty (THA). The aim of this study was to compare the use of DM bearings, standard bearings, and constrained liners in revision THA for recurrent dislocation, and to identify risk factors for re-dislocation. We reviewed 86 consecutive revision THAs performed for dislocation between August 2012 and July 2019. A total of 38 revisions (44.2%) involved a DM bearing, while 39 (45.3%) and nine (10.5%) involved a standard bearing and a constrained liner, respectively. Rates of re-dislocation, re-revision for dislocation, and overall re-revision were compared. Radiographs were assessed for the positioning of the acetabular component, the restoration of the centre of rotation, leg length, and offset. Risk factors for re-dislocation were determined by Cox regression analysis. The modified Harris Hip Scores (mHHSs) were recorded. The mean age of the patients at the time of revision was 70 years (43 to 88); 54 were female (62.8%). The mean follow-up was 5.0 years (2.0 to 8.75).Aims
Methods
Implantation of a large-diameter femoral head prosthesis with a metal-on-metal bearing surface reduces the risk of dislocation, increases the range of movement, minimises the risk of impingement and, in theory, results in little wear. Between February 2004 and March 2007 we implanted 100 consecutive total hip replacements with a metal-on-metal bearing and a large femoral head into 92 patients. There were 51 men and 41 women with a mean age of 50 years (18 to 70) at the time of surgery. Outcome was assessed using the Western Ontario McMaster University osteoarthritis index and the Harris hip score as well as the Devane activity score. These all improved significantly (p <
0.0001). At the last follow-up there were no cases of dislocation, no impingement, a good range of movement and no osteolysis, but seven revisions, two for infection and five for aseptic loosening. The probability of groin pain increased if the other
Literature surrounding artificial intelligence (AI)-related applications for hip and knee arthroplasty has proliferated. However, meaningful advances that fundamentally transform the practice and delivery of joint arthroplasty are yet to be realized, despite the broad range of applications as we continue to search for meaningful and appropriate use of AI. AI literature in hip and knee arthroplasty between 2018 and 2021 regarding image-based analyses, value-based care, remote patient monitoring, and augmented reality was reviewed. Concerns surrounding meaningful use and appropriate methodological approaches of AI in joint arthroplasty research are summarized. Of the 233 AI-related orthopaedics articles published, 178 (76%) constituted original research, while the rest consisted of editorials or reviews. A total of 52% of original AI-related research concerns hip and knee arthroplasty (n = 92), and a narrative review is described. Three studies were externally validated. Pitfalls surrounding present-day research include conflating vernacular (“AI/machine learning”), repackaging limited registry data, prematurely releasing internally validated prediction models, appraising model architecture instead of inputted data, withholding code, and evaluating studies using antiquated regression-based guidelines. While AI has been applied to a variety of hip and knee arthroplasty applications with limited clinical impact, the future remains promising if the question is meaningful, the methodology is rigorous and transparent, the data are rich, and the model is externally validated. Simple checkpoints for meaningful AI adoption include ensuring applications focus on: administrative support over clinical evaluation and management; necessity of the advanced model; and the novelty of the question being answered. Cite this article:
We investigated factors that were thought to be associated with an increased incidence of squeaking of ceramic-on-ceramic total hip replacements. Between June 1997 and December 2008 the three senior authors implanted 2406 primary total hip replacements with a ceramic-on-ceramic bearing surface. The mean follow-up was 10.6 years. The diagnosis was primary osteoarthritis in each case, and no patient had undergone previous surgery to the hip. We identified 74 squeaking hips (73 patients) giving an incidence of 3.1% at a mean follow-up of 9.5 years (4.1 to 13.3). Taller, heavier and younger patients were significantly more likely to have hips that squeaked. Squeaking hips had a significantly higher range of post-operative internal (p = 0.001) and external rotation (p = 0.003) compared with silent hips. Patients with squeaking hips had significantly higher activity levels (p = 0.009). A squeaking hip was not associated with a significant difference in patient satisfaction (p = 0.24) or Harris hip score (p = 0.34). Four implant position factors enabled good prediction of squeaking. These were high
The outcome of total hip replacement (THR) is potentially affected by the body mass index (BMI) of the patient. We studied the outcome of 2026 consecutive primary cementless THRs performed for osteoarthritis. The mean follow-up was 6.3 years (0 to 11.71) and no patient was lost to follow-up for survival analysis. The patients were divided into two groups according to their BMI as follows: non-obese (BMI <
30 kg/m. 2. ) and obese (BMI ≥ 30 kg/m. 2. ). The obese patient undergoing surgery was found to be significantly younger (p <
0.001). The log-rank test for equality of survival showed no difference in the mid-term survival (p = 0.552) with an estimated survival at 11 years of 95.2% (95% CI 92.5 to 98.0) in the non-obese and 96.7% (95% CI 94.9 to 98.5) in the obese groups. The clinical and radiological outcome was determined in a case-matched study performed on 134 obese individuals closely matched with 134 non-obese controls. The non-obese group was found to have a significantly higher post-operative Harris hip score (p <
0.001) and an increased range of movement, but overall satisfaction with surgery was comparable with that of the obese patients. Radiological analysis of the acetabular and femoral components showed no significant differences with regard to radiolucent lines, osteolysis, ingrowth of the femoral component, the
We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by mode of failure as either ‘unexplained pain’ or ‘defined causes’. Using this classification, chromium and cobalt ion levels were present over a wider range in synovial fluid and not as strongly correlated with blood ion levels as previously reported. There was no significant difference between metal ion concentrations and manufacturer of the implant, nor femoral head size below or above 50 mm. There was a moderately positive correlation between metal ion levels and
There is an increased risk of dislocation of the hip after the resection of a periacetabular tumour and endoprosthetic reconstruction of the defect in the hemipelvis. The aim of this study was to determine the rate and timing of dislocation and to identify its risk factors. To determine the dislocation rate, we conducted a retrospective single-institution study of 441 patients with a periacetabular tumour who had undergone a standard modular hemipelvic endoprosthetic reconstruction between 2003 and 2019. After excluding ineligible patients, 420 patients were enrolled. Patient-specific, resection-specific, and reconstruction-specific variables were studied using univariate and multivariate analyses.Aims
Methods
A retrospective study was conducted to investigate
the changes in metal ion levels in a consecutive series of Birmingham
Hip Resurfacings (BHRs) at a minimum ten-year follow-up. We reviewed
250 BHRs implanted in 232 patients between 1998 and 2001. Implant
survival, clinical outcome (Harris hip score), radiographs and serum chromium
(Cr) and cobalt (Co) ion levels were assessed. Of 232 patients, 18 were dead (five bilateral BHRs), 15 lost
to follow-up and ten had been revised. The remaining 202 BHRs in
190 patients (136 men and 54 women; mean age at surgery 50.5 years
(17 to 76)) were evaluated at a minimum follow-up of ten years (mean
10.8 years (10 to 13.6)). The overall implant survival at 13.2 years
was 92.4% (95% confidence interval 90.8 to 94.0). The mean Harris
hip score was 97.7 (median 100; 65 to 100). Median and mean ion
levels were low for unilateral resurfacings (Cr: median 1.3 µg/l,
mean
1.95 µg/l (<
0.5 to 16.2); Co: median 1.0 µg/l, mean 1.62 µg/l
(<
0.5 to 17.3)) and bilateral resurfacings (Cr: median 3.2 µg/l,
mean 3.46 µg/l (<
0.5 to 10.0); Co: median 2.3 µg/l, mean 2.66
µg/l (<
0.5 to 9.5)). In 80 unilateral BHRs with sequential ion
measurements, Cr and Co levels were found to decrease significantly
(p <
0.001) from the initial assessment at a median of six years
(4 to 8) to the last assessment at a median of 11 years (9 to 13),
with a mean reduction of 1.24 µg/l for Cr and 0.88 µg/l for Co.
Three female patients had a >
2.5 µg/l increase of Co ions, associated with
head sizes ≤ 50 mm, clinical symptoms and osteolysis. Overall, there
was no significant difference in change of ion levels between genders
(Cr, p = 0.845; Co, p = 0.310) or component sizes (Cr, p = 0.505;
Co, p = 0.370). Higher
Rotational acetabular osteotomy (RAO) has been reported to be effective in improving symptoms and preventing osteoarthritis (OA) progression in patients with mild to severe develomental dysplasia of the hip (DDH). However, some patients develop secondary OA even when the preoperative joint space is normal; determining who will progress to OA is difficult. We evaluated whether the preoperative cartilage condition may predict OA progression following surgery using T2 mapping MRI. We reviewed 61 hips with early-stage OA in 61 patients who underwent RAO for DDH. They underwent preoperative and five-year postoperative radiological analysis of the hip. Those with a joint space narrowing of more than 1 mm were considered to have 'OA progression'. Preoperative assessment of articular cartilage was also performed using 3T MRI with the T2 mapping technique. The region of interest was defined as the weightbearing portion of the acetabulum and femoral head.Aims
Methods
The aims of this systematic review were to assess the learning curve of semi-active robotic arm-assisted total hip arthroplasty (rTHA), and to compare the accuracy, patient-reported functional outcomes, complications, and survivorship between rTHA and manual total hip arthroplasty (mTHA). Searches of PubMed, Medline, and Google Scholar were performed in April 2020 in line with the Preferred Reporting Items for Systematic Review and Meta-Analysis statement. Search terms included “robotic”, “hip”, and “arthroplasty”. The criteria for inclusion were published clinical research articles reporting the learning curve for rTHA (robotic arm-assisted only) and those comparing the implantation accuracy, functional outcomes, survivorship, or complications with mTHA.Aims
Methods
The complex relationship between acetabular component position and spinopelvic mobility in patients following total hip arthroplasty (THA) renders it difficult to optimize acetabular component positioning. Mobility of the normal lumbar spine during postural changes results in alterations in pelvic tilt (PT) to maintain the sagittal balance in each posture and, as a consequence, markedly changes the functional component anteversion (FCA). This study aimed to investigate the in vivo association of lumbar degenerative disc disease (DDD) with the PT angle and with FCA during postural changes in THA patients. A total of 50 patients with unilateral THA underwent CT imaging for radiological evaluation of presence and severity of lumbar DDD. In all, 18 patients with lumbar DDD were compared to 32 patients without lumbar DDD. In vivo PT and FCA, and the magnitudes of changes (ΔPT; ΔFCA) during supine, standing, swing-phase, and stance-phase positions were measured using a validated dual fluoroscopic imaging system.Aims
Methods
Vitamin E-infused highly crosslinked polyethylene (VEPE) has been introduced into total hip arthroplasty (THA) with the aim of further improving the wear characteristics of moderately and highly crosslinked polyethylenes (ModXLPE and HXLPE). There are few studies analyzing the outcomes of vitamin E-infused components in cemented arthroplasty, though early acetabular component migration has been reported. The aim of this study was to measure five-year polyethylene wear and acetabular component stability of a cemented VEPE acetabular component compared with a ModXLPE cemented acetabular component. In a prospective randomized controlled trial (RCT), we assessed polyethylene wear and acetabular component stability (primary outcome) with radiostereometric analysis (RSA) in 68 patients with reverse hybrid THA at five years follow-up. Patients were randomized to either a VEPE or a ModXLPE cemented acetabular component.Aims
Methods
Highly cross-linked polyethylene (HXLPE) has greatly improved the durability of total hip arthroplasty (THA) in young patients because of its improved wear characteristics. Few studies have followed this population into the second decade, and therefore the purpose of this investigation was to evaluate the clinical outcome for THA patients 50 years of age and younger at a minimum of 15 years postoperatively. The second purpose was to evaluate the radiological findings secondary to wear or mechanical failure of the implant. Between October 1999 and December 2005, 105 THAs were performed in 95 patients (53 female, 42 male) aged 50 years and younger (mean 42 years (20 to 50)). There were 87 patients (96 hips) that were followed for a minimum of 15 years (mean 17.3 years (15 to 21)) for analysis. Posterior approach was used with cementless fixation with a median head size of 28 mm. HXLPE was the acetabular bearing for all hips. Radiographs were evaluated for polyethylene wear, radiolucent lines, and osteolysis.Aims
Methods
The aims of this study were to determine the change in the sagittal alignment of the pelvis and the associated impact on acetabular component position at one-year follow-up after total hip arthroplasty (THA). This study represents the one-year follow-up of a previous short-term study at our institution. Using the patient population from our prior study, the radiological pelvic ratio was assessed in 91 patients undergoing THA, of whom 50 were available for follow-up of at least one year (median 1.5; interquartile range (IQR) 1.1 to 2.0). Anteroposterior radiographs of the pelvis were obtained in the standing position preoperatively and at one year postoperatively. Pelvic ratio was defined as the ratio between the vertical distance from the inferior sacroiliac (SI) joints to the superior pubic symphysis and the horizontal distance between the inferior SI joints. Apparent acetabular component position changes were determined from the change in pelvic ratio. A change of at least 5° was considered clinically meaningful.Aims
Methods
Use of the direct anterior approach (DAA) for total hip arthroplasty (THA) has increased in recent years due to proposed benefits, including a lower risk of dislocation and improved early functional recovery. This study investigates the dislocation rate in a non-selective, consecutive cohort undergoing THA via the DAA without any exclusion or bias in patient selection based on habitus, deformity, age, sex, or fixation method. We retrospectively reviewed all patients undergoing THA via the DAA between 2011 and 2017 at our institution. Primary outcome was dislocation at minimum two-year follow-up. Patients were stratified by demographic details and risk factors for dislocation, and an in-depth analysis of dislocations was performed.Aims
Methods
The orientation of the acetabular component can
influence both the short- and long-term outcomes of total hip replacement
(THR). We performed a prospective, randomised, controlled trial
of two groups, comprising of 40 patients each, in order to compare
freehand introduction of the component with introduction using the transverse
acetabular ligament (TAL) as a reference for anteversion. Anteversion
and inclination were measured on pelvic radiographs. With respect to anteversion, in the freehand group 22.5% of the
components were outside the safe zone versus 0%
in the transverse acetabular ligament group (p = 0.002). The mean
angle of anteversion in the freehand group was 21° (2° to 35°) which
was significantly higher compared with 17° (2° to 25°) in the TAL
group (p = 0.004). There was a significant difference comparing
the variations of both groups (p = 0.008). With respect to inclination, in the freehand group 37.5% of the
components were outside the safe zone versus 20%
in the TAL group (p = 0.14). There was no significant difference
regarding the accuracy or variation of the angle of inclination
when comparing the two groups. . The transverse acetabular ligament may be used to obtain the
appropriate anteversion when introducing the acetabular component
during THR, but not
The hypothesis of this study was that thigh circumference, distinct from body mass index (BMI), may be associated with the positioning of components when undertaking total hip arthroplasty (THA) using the direct anterior approach (DAA), and that an increased circumference might increase the technical difficulty. We performed a retrospective review of prospectively collected data involving 155 consecutive THAs among 148 patients undertaken using the DAA at an academic medical centre by a single fellowship-trained surgeon. Preoperatively, thigh circumference was measured at 10 cm, 20 cm, and 30 cm distal to the anterior superior iliac spine, in quartiles. Two blinded reviewers assessed the inclination and anteversion of the acetabular component, radiological leg-length discrepancy, and femoral offset. The radiological outcomes were considered as continuous and binary outcome variables based on Lewinnek’s ‘safe zone’.Aims
Patients and Methods
We present the clinical results and survivorship
of consecutive 100 Birmingham Hip Resurfacings in 90 patients at
a minimum follow-up of ten years. All procedures were carried out
by an independent surgeon who commenced a prospective study in 1998.
Patients were assessed clinically using the Western Ontario and
McMaster Universities osteoarthritis index, Short-Form 36, Harris
hip score and University of California, Los Angeles activity score. Radiological
analysis was performed by an independent observer and blood metal
ion levels concentrations were measured at ten years post-operatively
in 62 patients. The median
This study reports the mid-term results of total hip arthroplasty (THA) performed using a monoblock acetabular component with a large-diameter head (LDH) ceramic-on-ceramic (CoC) bearing. Of the 276 hips (246 patients) included in this study, 264 (96%) were reviewed at a mean of 67 months (48 to 79) postoperatively. Procedures were performed with a mini posterior approach. Clinical and radiological outcomes were recorded at regular intervals. A noise assessment questionnaire was completed at last follow-up.Aims
Patients and Methods
The primary objective of this study was to compare accuracy in restoring the native centre of hip rotation in patients undergoing conventional manual total hip arthroplasty (THA) This prospective cohort study included 50 patients undergoing conventional manual THA and 25 patients receiving robotic-arm assisted THA. Patients undergoing conventional manual THA and robotic-arm assisted THA were well matched for age (mean age, 69.4 years (Objectives
Materials and Methods
While previously underappreciated, factors related to the spine contribute substantially to the risk of dislocation following total hip arthroplasty (THA). These factors must be taken into consideration during preoperative planning for revision THA due to recurrent instability. We developed a protocol to assess the functional position of the spine, the significance of these findings, and how to address different pathologies at the time of revision THA. Prospectively collected data on 111 patients undergoing revision THA for recurrent instability from January 2014 to January 2017 at two institutions were included (protocol group) and matched 1:1 to 111 revisions specifically performed for instability not using this protocol (control group). Mean follow-up was 2.8 years. Protocol patients underwent standardized preoperative imaging including supine and standing anteroposterior (AP) pelvis and lateral radiographs. Each case was scored according to the Hip-Spine Classification in Revision THA.Aims
Patients and Methods
Our first aim was to determine whether there are significant
changes in the level of metal ions in the blood at mid-term follow-up,
in patients with an Articular Surface Replacement (ASR) arthroplasty.
Secondly, we sought to identify risk factors for any increases. The study involved 435 patients who underwent unilateral, metal-on-metal
(MoM) hip resurfacing (HRA) or total hip arthroplasty (THA). These
patients all had one measurement of the level of metal ions in the
blood before seven years had passed post-operatively (early evaluation)
and one after seven years had passed post-operatively (mid-term evaluation).
Changes in ion levels were tested using a Wilcoxon signed-rank test.
We identified subgroups at the highest risk of increase using a
multivariable linear logistic regression model.Aims
Patients and Methods
The hip’s capsular ligaments passively restrain extreme range of movement (ROM) by wrapping around the native femoral head/neck. We determined the effect of hip resurfacing arthroplasty (HRA), dual-mobility total hip arthroplasty (DM-THA), conventional THA, and surgical approach on ligament function. Eight paired cadaveric hip joints were skeletonized but retained the hip capsule. Capsular ROM restraint during controlled internal rotation (IR) and external rotation (ER) was measured before and after HRA, DM-THA, and conventional THA, with a posterior (right hips) and anterior capsulotomy (left hips).Aims
Materials and Methods
Concurrent hip and spine pathologies can alter the biomechanics of spinopelvic mobility in primary total hip arthroplasty (THA). This study examines how differences in pelvic orientation of patients with spine fusions can increase the risk of dislocation risk after THA. We identified 84 patients (97 THAs) between 1998 and 2015 who had undergone spinal fusion prior to primary THA. Patients were stratified into three groups depending on the length of lumbar fusion and whether or not the sacrum was involved. Mean age was 71 years (40 to 87) and 54 patients (56%) were female. The mean body mass index (BMI) was 30 kg/m2 (19 to 45). Mean follow-up was six years (2 to 17). Patients were 1:2 matched to patients with primary THAs without spine fusion. Hazard ratios (HR) were calculated.Aims
Patients and Methods
Computer-based applications are increasingly being used by orthopaedic surgeons in their clinical practice. With the integration of technology in surgery, augmented reality (AR) may become an important tool for surgeons in the future. By superimposing a digital image on a user’s view of the physical world, this technology shows great promise in orthopaedics. The aim of this review is to investigate the current and potential uses of AR in orthopaedics. A systematic review of the PubMed, MEDLINE, and Embase databases up to January 2019 using the keywords ‘orthopaedic’ OR ‘orthopedic AND augmented reality’ was performed by two independent reviewers.Aims
Materials and Methods
The primary aim of this independent prospective randomised trial
was to compare serum metal ion levels for ceramic-on-metal (CoM)
and metal-on-metal (MoM) bearing surfaces in total hip arthroplasty
(THA). Our one-year results demonstrated elevation in metal ion
levels above baseline with no significant difference between the
CoM and MoM groups. This paper reviews the five-year data. The implants used in each patient differed only in respect to
the type of femoral head (ceramic or metal). At five-year follow-up
of the 83 enrolled patients, data from 67 (36 CoM, 31 MoM) was available
for comparison.Aims
Patients and Methods
The aims of this study were to review the surgical technique for a combined femoral head reduction osteotomy (FHRO) and periacetabular osteotomy (PAO), and to report the short-term clinical and radiological results of a combined FHRO/PAO for the treatment of selected severe femoral head deformities. Between 2011 and 2016, six female patients were treated with a combined FHRO and PAO. The mean patient age was 13.6 years (12.6 to 15.7). Clinical data, including patient demographics and patient-reported outcome scores, were collected prospectively. Radiologicalally, hip morphology was assessed evaluating the Tönnis angle, the lateral centre to edge angle, the medial offset distance, the extrusion index, and the alpha angle.Aims
Patients and Methods
Aims
Patients and Methods
In order to prevent dislocation of the hip after total hip arthroplasty
(THA), patients have to adhere to precautions in the early post-operative
period. The hypothesis of this study was that a protocol with minimal
precautions after primary THA using the posterolateral approach
would not increase the short-term (less than three months) risk
of dislocation. We prospectively monitored a group of unselected patients undergoing
primary THA managed with standard precautions (n = 109, median age
68.9 years; interquartile range (IQR) 61.2 to 77.3) and a group
who were managed with fewer precautions (n = 108, median age 67.2
years; IQR 59.8 to 73.2). There were no significant differences between
the groups in relation to predisposing risk factors. The diameter
of the femoral head ranged from 28 mm to 36 mm; meticulous soft-tissue
repair was undertaken in all patients. The medical records were
reviewed and all patients were contacted three months post-operatively
to confirm whether they had experienced a dislocation. Aims
Patients and Methods
This study compared multiple sclerosis (MS) patients who underwent
primary total hip arthroplasty (THA) with a matched cohort. Specifically,
we evaluated: 1) implant survivorship; 2) functional outcomes (modified
Harris Hip Scores (mHHS), Hip Disability and Osteoarthritis Outcome
Score, Joint Replacement (HOOS JR), and modified Multiple Sclerosis
Impact Scale (mMSIS) scores (with the MS cohort also evaluated based
on the disease phenotype)); 3) physical therapy duration and return
to function; 4) radiographic outcomes; and 5) complications. We reviewed our institution’s database to identify MS patients
who underwent THA between January 2008 and June 2016. A total of
34 MS patients (41 hips) were matched in a 1:2 ratio to a cohort
of THA patients who did not have MS, based on age, body mass index
(BMI), and Charlson/Deyo score. Patient records were reviewed for complications,
and their functional outcomes and radiographs were reviewed at their
most recent follow-up.Aims
Patients and Methods
Posterior tilt of the pelvis with sitting provides biological
acetabular opening. Our goal was to study the post-operative interaction
of skeletal mobility and sagittal acetabular component position. This was a radiographic study of 160 hips (151 patients) who
prospectively had lateral spinopelvic hip radiographs for skeletal
and implant measurements. Intra-operative acetabular component position
was determined according to the pre-operative spinal mobility. Sagittal
implant measurements of ante-inclination and sacral acetabular angle were
used as surrogate measurements for the risk of impingement, and
intra-operative acetabular component angles were compared with these.Aims
Materials and Methods
The purpose of this study was to determine patient-reported
outcomes of patients with mild to moderate developmental dysplasia
of the hip (DDH) and femoroacetabular impingement (FAI) undergoing
arthroscopy of the hip in the treatment of chondrolabral pathology.
A total of 28 patients with a centre-edge angle between 15° and
19° were identified from an institutional database. Their mean age
was 34 years (18 to 53), with 12 female and 16 male patients. All
underwent labral treatment and concomitant correction of FAI. There
were nine reoperations, with two patients requiring revision arthroscopy,
two requiring periacetabular osteotomy and five needing total hip arthroplasty. Patients who required further major surgery were more likely
to be older, male, and to have more severe DDH with a larger alpha
angle and decreased joint space. At a mean follow-up of 42 months (24 to 89), the mean modified
Harris hip score improved from 59 (20 to 98) to 82 (45 to 100; p
<
0.001). The mean Western Ontario and McMaster Universities
Osteoarthritis Index score improved from 30 (1 to 61) to 16 (0 to
43; p <
0.001). Median patient satisfaction was 9.0/10 (1 to
10). Patients reported excellent improvement in function following
arthroscopy of the hip. This study shows that with proper patient selection, arthroscopy
of the hip can be successful in the young patient with mild to moderate
DDH and FAI. Cite this article:
The aim of this study was to compare early functional and health
related quality of life outcomes (HRQoL) in patients who have undergone
total hip arthroplasty (THA) using a bone conserving short stem
femoral component and those in whom a conventional length uncemented
component was used. Outcome was assessed using a validated performance
based outcome instrument as well as patient reported outcome measures
(PROMs). We prospectively analysed 33 patients whose THA involved a contemporary
proximally porous coated tapered short stem femoral component and
53 patients with a standard conventional femoral component, at a
minimum follow-up of two years. The mean follow-up was 31.4 months
(24 to 39). Patients with poor proximal femoral bone quality were
excluded. The mean age of the patients was 66.6 years (59 to 77)
and the mean body mass index was 30.2 kg/m2 (24.1 to
41.0). Outcome was assessed using the Oxford Hip Score (OHS) and
the University College Hospital (UCH) hip score which is a validated
performance based instrument. HRQoL was assessed using the EuroQol
5D (EQ-5D).Aims
Patients and Methods
Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head
taper junction in two groups of retrieved hips; those with mixed
manufacturers (MM) and those from the same manufacturer (SM). We identified 151 retrieved hips with large-diameter cobalt-chromium
heads; 51 of two designs that had been paired with stems from different
manufacturers (MM) and 100 of seven designs paired with stems from
the same manufacturer (SM). We determined the severity of corrosion
with the Goldberg corrosion score and the volume of material loss
at the head/stem junction. We used multivariable statistical analysis
to determine if there was a significant difference between the two
groups. Aims
Materials and Methods
We carried out a further study of the long-term results of the
cemented Exeter femoral component in patients under the age of 40
with a mean follow-up of 13.6 years (10 to 20). We reviewed our original cohort of 104 cemented Exeter stems
in 78 consecutive patients with a mean age of 31 years (16 to 39).
Only one patient was lost to radiological follow-up.Aims
Patients and Methods
We compared the incidence of pseudotumours after
large head metal-on-metal (MoM) total hip arthroplasty (THA) with
that after conventional metal-on-polyethylene (MoP) THA and assessed
the predisposing factors to pseudotumour formation. From a previous randomised controlled trial which compared large
head (38 mm to 60 mm) cementless MoM THA with conventional head
(28 mm) cementless MoP THA, 93 patients (96 THAs: 41 MoM (21 males,
20 females, mean age of 64 years, standard deviation ( The incidence of pseudotumours, measured using a standardised
CT protocol was 22 (53.7%) after MoM THA and 12 (21.8%) after MoP
THA. Women with a MoM THA were more likely to develop a pseudotumour
than those with a MoP THA (15 Contrary to popular belief, pseudotumours occur frequently around
MoP THAs. Women with a MoM THA and an elevated cobalt level are
at greatest risk. In this study, pseudotumours had no effect on
the functional outcome after either large head MoM or conventional
MoP THA. Cite this article:
There are limited published data detailing the volumetric material loss from tapers of conventional metal-on-polyethylene (MoP) total hip arthroplasties (THAs). Our aim was to address this by comparing the taper wear rates measured in an explanted cohort of the widely used Exeter THA with those measured in a group of metal-on-metal (MoM) THAs. We examined an existing retrieval database to identify all Exeter V40 and Universal MoP THAs. Volumetric wear analysis of the taper surfaces was conducted using previously validated methodology. These values were compared with those obtained from a series of MoM THAs using non-parametric statistical methodology. A number of patient and device variables were accounted for using multiple regression modelling.Aims
Patients and Methods
In order to evaluate the relationship between acetabular and proximal femoral alignment in the initiation and evolution of osteoarthritis of the dysplastic hip, the acetabular and femoral angles were calculated geometrically from radiographs of 62 patients with pre-arthrosis and early osteoarthritis. The sum of the lateral opening angle of the acetabulum and the neck-shaft angle was defined as the lateral instability index (LII), and the sum of the anterior opening angle of the acetabulum and the anteversion angle of the femoral neck as the anterior instability index (AII). These two indices were compared in dysplastic and unaffected hips. A total of 22 unilateral hips with pre-arthrosis were followed for at least 15 years to determine whether the two indices were associated with the progression of osteoarthritis. The LII of the affected hips (197.4 (
In a double-blinded randomised controlled trial,
83 patients with primary osteoarthritis of the hip received either
a ceramic-on-metal (CoM) or metal-on-metal (MoM) total hip replacement
(THR). The implants differed only in the bearing surfaces used.
The serum levels of cobalt and chromium and functional outcome scores
were compared pre-operatively and at six and 12 months post-operatively. Data were available for 41 CoM and 36 MoM THRs (four patients
were lost to follow-up, two received incorrect implants). The baseline
characteristics of both cohorts were similar. Femoral head size
measured 36 mm in all but two patients who had 28 mm heads. The
mean serum cobalt and chromium levels increased in both groups,
with no difference noted between groups at six months (cobalt p
= 0.67, chromium p = 0.87) and 12 months (cobalt p = 0.76, chromium
p = 0.76) post-operatively. Similarly, the mean Oxford hip scores,
Western Ontario and McMaster Universities Osteoarthritis index and
University of California, Los Angeles activity scores showed comparable improvement
at 12 months. Our findings indicate that CoM and MoM couplings are associated
with an equivalent increase in serum cobalt and chromium levels,
and comparable functional outcome scores at six and 12-months follow-up.
Accurate placement of the acetabular component during total hip
arthroplasty (THA) is an important factor in the success of the
procedure. However, the reported accuracy varies greatly and is
dependent upon whether free hand or navigated techniques are used.
The aim of this study was to assess the accuracy of an instrument
system that incorporates 3D printed, patient-specific guides designed
to optimise the placement of the acetabular component. A total of 100 consecutive patients were prospectively enrolled
and the accuracy of placement of the acetabular component was measured
using post-operative CT scans.Aims
Patients and Methods
Wear of metal-on-metal bearings causes elevated levels of cobalt and chromium in blood and body fluids. Metal-on-metal bearings have two distinct wear phases. In the early phase, the wear rate is high. Later, it decreases and the bearing enters a steady-state phase. It is expected that as the wear rates decline, the level of cobalt detected in plasma will also decrease. We studied the baseline and exercise-related cobalt rise in 21 patients (13 men and eight women) with a mean age of 54 years (38 to 80) who had undergone successful hip resurfacing at a mean of 44 months (10 to 96) earlier. Our results showed that circulating baseline cobalt levels were not significantly correlated with the time since implantation (r = 0.08, p = 0.650). By contrast, the exercise-related cobalt rise was directly correlated with the inclination angle of the acetabular component (r = 0.47, p = 0.032) and inversely correlated with the time since implantation (r = −0.5, p = 0.020). Inclination of the acetabular component should be kept less than 40° to decrease the production of wear debris.
We present the ten-year data of a cohort of patients, aged between
18 and 65 years (mean age 52.7 years; 19 to 64), who underwent total
hip arthroplasty. Patients were randomised to be treated with a
cobalt-chrome (CoCr) femoral head with an ultra-high molecular weight
polyethylene (UHMWPE), highly cross-linked polyethylene (XLPE) or
ceramic-on-ceramic (CoC) bearing surface. A total of 102 hips (91 patients) were randomised into the three
groups. At ten years, 97 hips were available for radiological and
functional follow-up. Two hips (two patients) had been revised (one
with deep infection and one for periprosthetic fracture) and three
were lost to follow-up. Radiological analysis was performed using
a validated digital assessment programme to give linear, directional
and volumetric wear of the two polyethylene groups.Aims
Patients and Methods
Cemented femoral stems with force closed fixation
designs have shown good clinical results despite high early subsidence.
A new triple-tapered stem in this category (C-stem AMT) was introduced
in 2005. This study compares this new stem with an established stem
of similar design (Exeter) in terms of migration (as measured using radiostereometric
analysis), peri-prosthetic bone remodelling (measured using dual
energy x-ray densitometry, DXA), Oxford Hip Score, and plain radiographs. A total of 70 patients (70 hips) with a mean age of 66 years
(53 to 78) were followed for two years. Owing to missing data of
miscellaneous reasons, the final analysis represents data from 51
(RSA) and 65 (DXA) patients. Both stems showed a typical pattern
of migration: Subsidence and retroversion that primarily occurred
during the first three months. C-stem AMT subsided less during the
first three months (p = 0.01), before stabilising at a subsidence rate
similar to the Exeter stem from years one to two. The rate of migration
into retroversion was slightly higher for C-stem AMT during the
second year (p = 0.03). Whilst there were slight differences in
movement patterns between the stems, the C-stem AMT exhibits good
early clinical outcomes and displays a pattern of migration and
bone remodelling that predicts good clinical performance. Cite this article:
This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties: hip resurfacing and modular total hip replacement (THR). There was no significant difference in the median rate of linear wear between the groups for both acetabular (p = 0.4633) and femoral (p = 0.0872) components. There was also no significant difference in the median linear wear rates when failed hip resurfacing and modular THR hips of the same type (ASR and Birmingham hip resurfacing (BHR)) were compared. Unlike other studies of well-functioning hips, there was no significant difference in pre-revision blood metal ion levels between hip resurfacing and modular THR. Edge loading was common in both groups, but more common in the resurfacing group (67%) than in the modular group (57%). However, this was not significant (p = 0.3479). We attribute this difference to retention of the neck in resurfacing of the hip, leading to impingement-type edge loading. This was supported by visual evidence of impingement on the femur. These findings show that failed metal-on-metal hip resurfacing and modular THRs have similar component wear rates and are both associated with raised pre-revision blood levels of metal ions.
We report the kinematic and early clinical results
of a patient- and observer-blinded randomised controlled trial in which
CT scans were used to compare potential impingement-free range of
movement (ROM) and acetabular component cover between patients treated
with either the navigated ‘femur-first’ total hip arthroplasty (THA) method
(n = 66; male/female 29/37, mean age 62.5 years; 50 to 74) or conventional
THA (n = 69; male/female 35/34, mean age 62.9 years; 50 to 75).
The Hip Osteoarthritis Outcome Score, the Harris hip score, the
Euro-Qol-5D and the Mancuso THA patient expectations score were
assessed at six weeks, six months and one year after surgery. A
total of 48 of the patients (84%) in the navigated ‘femur-first’
group and 43 (65%) in the conventional group reached all the desirable
potential ROM boundaries without prosthetic impingement for activities
of daily living (ADL) in flexion, extension, abduction, adduction
and rotation (p = 0.016). Acetabular component cover and surface
contact with the host bone were >
87% in both groups. There was
a significant difference between the navigated and the conventional
groups’ Harris hip scores six weeks after surgery (p = 0.010). There
were no significant differences with respect to any clinical outcome
at six months and one year of follow-up. The navigated ‘femur-first’
technique improves the potential ROM for ADL without prosthetic
impingement, although there was no observed clinical difference
between the two treatment groups. Cite this article:
We undertook a retrospective cohort study to
determine clinical outcomes following the revision of metal-on-metal (MoM)
hip replacements for adverse reaction to metal debris (ARMD), and
to identify predictors of time to revision and outcomes following
revision. Between 1998 and 2012 a total of 64 MoM hips (mean age
at revision of 57.8 years; 46 (72%) female; 46 (72%) hip resurfacings
and 18 (28%) total hip replacements) were revised for ARMD at one specialist
centre. At a mean follow-up of 4.5 years (1.0 to 14.6) from revision
for ARMD there were 13 hips (20.3%) with post-operative complications
and eight (12.5%) requiring re-revision. The Kaplan–Meier five-year survival rate for ARMD revision was
87.9% (95% confidence interval 78.9 to 98.0; 19 hips at risk). Excluding
re-revisions, the median absolute Oxford hip score (OHS) following
ARMD revision using the percentage method (0% best outcome and 100%
worst outcome) was 18.8% (interquartile range (IQR) 7.8% to 48.3%),
which is equivalent to 39/48 (IQR 24.8/48 to 44.3/48) when using
the modified OHS. Histopathological response did not affect time
to revision for ARMD (p = 0.334) or the subsequent risk of re-revision
(p = 0.879). Similarly, the presence or absence of a contralateral
MoM hip bearing did not affect time to revision for ARMD (p = 0.066)
or the subsequent risk of re-revision (p = 0.178). Patients revised to MoM bearings had higher rates of re-revision
(five of 16 MoM hips re-revised; p = 0.046), but those not requiring
re-revision had good functional results (median absolute OHS 14.6%
or 41.0/48). Short-term morbidity following revision for ARMD was
comparable with previous reports. Caution should be exercised when choosing
bearing surfaces for ARMD revisions. Cite this article:
We sought to determine whether cobalt-chromium alloy (CoCr) femoral
stem tapers (trunnions) wear more than titanium (Ti) alloy stem
tapers (trunnions) when used in a large diameter (LD) metal-on-metal
(MoM) hip arthroplasty system. We performed explant analysis using validated methodology to
determine the volumetric material loss at the taper surfaces of
explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy
(n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs
with a rough male taper surface and a nominal included angle close
to 5.666° were included. Multiple regression modelling was undertaken
using taper angle, taper roughness, bearing diameter (horizontal
lever arm) as independent variables. Material loss was mapped using
a coordinate measuring machine, profilometry and scanning electron
microscopy.Aims
Patients and Methods
The primary aim of this study was to analyse the position of
the acetabular and femoral components in total hip arthroplasty
undertaken using an anterior surgical approach. In a prospective, single centre study, we used the EOS imaging
system to analyse the position of components following THA performed
via the anterior approach in 102 patients (103 hips) with a mean
age of 64.7 years (Aims
Patients and Methods
We report the outcome at a minimum of five years of 110 consecutive metal-on-metal Birmingham Hip Resurfacing arthroplasties in 98 patients. The procedures were performed between October 1999 and June 2002 by one surgeon. All patients were followed up clinically and radiologically. The mean follow-up was 71 months (60 to 93). Revision of either component was defined as failure. The mean Harris Hip score at follow-up was 96.4 (53 to 100). The mean Oxford hip score was 41.9 (16 to 57) pre-operatively and 15.4 (12 to 49) post-operatively (p <
0.001). The mean University of California Los Angeles activity score was 3.91 (1 to 10) pre-operatively and 7.5 (4 to 10) post-operatively (p <
0.001). There were four failures giving a survival at five years of 96.3% (95% confidence interval 92.8 to 99.8). When applying a new method to estimate narrowing of the femoral neck we identified a 10% thinning of the femoral neck in 16 hips (14.5%), but the relevance of this finding to the long-term outcome remains unclear. These good medium-term results from an independent centre confirm the original data from Birmingham.
There have been no large comparative studies of the blood levels of metal ions after implantation of commercially available hip resurfacing devices which have taken into account the effects of femoral size and inclination and anteversion of the acetabular component. We present the results in 90 patients with unilateral articular surface replacement (ASR) hip resurfacings (mean time to blood sampling 26 months) and 70 patients with unilateral Birmingham Hip Resurfacing (BHR) implants (mean time 47 months). The whole blood and serum chromium (Cr) and cobalt (Co) concentrations were inversely related to the size of the femoral component in both groups (p <
0.05). Cr and Co were more strongly influenced by the position of the acetabular component in the case of the ASR, with an increase in metal ions observed at inclinations >
45° and anteversion angles of <
10° and >
20°. These levels were only increased in the BHR group when the acetabular component was implanted with an inclination >
55°. A significant relationship was identified between the anteversion of the BHR acetabular component and the levels of Cr and Co (p <
0.05 for Co), with an increase observed at anteversion angles <
10° and >
20°. The median whole blood and serum Cr concentrations of the male ASR patients were significantly lower than those of the BHR men (p <
0.001). This indicates that reduced diametral clearance may equate to a reduction in metal ion concentrations in larger joints with satisfactory orientation of the acetabular component.
We report the findings of an independent review
of 230 consecutive Birmingham hip resurfacings (BHRs) in 213 patients
(230 hips) at a mean follow-up of 10.4 years (9.6 to 11.7). A total
of 11 hips underwent revision; six patients (six hips) died from
unrelated causes; and 13 patients (16 hips) were lost to follow-up.
The survival rate for the whole cohort was 94.5% (95% confidence
interval (CI) 90.1 to 96.9). The survival rate in women was 89.1%
(95% CI 79.2 to 94.4) and in men was 97.5% (95% CI 92.4 to 99.2).
Women were 1.4 times more likely to suffer failure than men. For
each millimetre increase in component size there was a 19% lower
chance of a failure. The mean Oxford hip score was 45.0 (median
47.0, 28 to 48); mean University of California, Los Angeles activity
score was 7.4 (median 8.0, 3 to 9); mean patient satisfaction score
was 1.4 (median 1.0, 0 to 9). A total of eight hips had lysis in
the femoral neck and two hips had acetabular lysis. One hip had
progressive radiological changes around the peg of the femoral component.
There was no evidence of progressive neck narrowing between five
and ten years. Our results confirm that BHR provides good functional outcome
and durability for men, at a mean follow-up of ten years. We are
now reluctant to undertake hip resurfacing in women with this implant.
Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell. Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%). At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis. Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.
Adverse reaction to wear and corrosion debris
is a cause for concern in total hip arthroplasty (THA). Modular junctions
are a potential source of such wear products and are associated
with secondary pseudotumour formation. We present a consecutive series of 17 patients treated at our
unit for this complication following metal-on-highly cross-linked
polyethylene (MoP) THA. We emphasise the risk of misdiagnosis as
infection, and present the aggregate laboratory results and pathological
findings in this series. The clinical presentation was pain, swelling or instability.
Solid, cystic and mixed soft-tissue lesions were noted on imaging
and confirmed intra-operatively. Corrosion at the head–neck junction
was noted in all cases. No bacteria were isolated on multiple pre-
and intra-operative samples yet the mean erythrocyte sedimentation
rate was 49 (9 to 100) and C-reactive protein 32 (0.6 to 106) and
stromal polymorphonuclear cell counts were noted in nine cases. Adverse soft–tissue reactions can occur in MoP THA owing to corrosion
products released from the head–neck junction. The diagnosis should
be carefully considered when investigating pain after THA. This
may avoid the misdiagnosis of periprosthetic infection with an unidentified
organism and mitigate the unnecessary management of these cases
with complete single- or two-stage exchange. Cite this article:
Orientation of the acetabular component influences
wear, range of movement and the incidence of dislocation after total
hip replacement (THR). During surgery, such orientation is often
referenced to the anterior pelvic plane (APP), but APP inclination
relative to the coronal plane (pelvic tilt) varies substantially
between individuals. In contrast, the change in pelvic tilt from
supine to standing (dPT) is small for nearly all individuals. Therefore,
in THR performed with the patient supine and the patient’s coronal
plane parallel to the operating table, we propose that freehand placement
of the acetabular component placement is reliable and reflects standing
(functional) cup position. We examined this hypothesis in 56 hips
in 56 patients (19 men) with a mean age of 61 years (29 to 80) using
three-dimensional CT pelvic reconstructions and standing lateral
pelvic radiographs. We found a low variability of acetabular component
placement, with 46 implants (82%) placed within a combined range
of 30° to 50° inclination and 5° to 25° anteversion. Changing from
the supine to the standing position (analysed in 47 patients) was associated
with an anteversion change <
10° in 45 patients (96%). dPT was
<
10° in 41 patients (87%). In conclusion, supine THR appears
to provide reliable freehand acetabular component placement. In
most patients a small reclination of the pelvis going from supine
to standing causes a small increase in anteversion of the acetabular component Cite this article:
We report the clinical and radiological outcome at ten years of 104 primary total hip replacements (100 patients) using the Metasul metal-on-metal bearing. Of these, 52 had a cemented Stuehmer-Weber polyethylene acetabular component with a Metasul bearing and 52 had an uncemented Allofit acetabular component with a Metasul liner. A total of 15 patients (16 hips) died before their follow-up at ten years and three were lost to follow-up. The study group therefore comprised 82 patients (85 hips). The mean Oxford score at ten years was 20.7 (12 to 42). Six of 85 hips required revision surgery. One was performed because of infection, one for aseptic loosening of the acetabular component and four because of unexplained pain. Histological examination showed an aseptic lymphocytic vasculitis associated lesion-type tissue response in two of these. Continued follow-up is advocated in order to monitor the long-term performance of the Metasul bearing and tissue responses to metal debris.
We evaluated the short-term functional outcome and prevalence
of bearing-specific generation of audible noise in 301 patients
(336 hips) operated on with fourth generation (Delta) medium diameter
head, ceramic-on-ceramic (CoC) total hip arthroplasties (THAs). There were 191 female (63%) and 110 male patients (37%) with
a mean age of 61 years (29 to 78) and mean follow-up of 2.1 years
(1.3 to 3.4). Patients completed three questionnaires: Oxford Hip
Score (OHS), Research and Development 36-item health survey (RAND-36)
and a noise-specific symptom questionnaire. Plain radiographs were also
analysed. A total of three hips (0.9%) were revised. Aims
Patients and Methods
One goal of total hip arthroplasty is to restore normal hip anatomy.
The aim of this study was to compare displacement of the centre
of rotation (COR) using a standard reaming technique with a technique
in which the acetabulum was reamed immediately peripherally and
referenced off the rim. In the first cohort the acetabulum was reamed to the floor followed
by sequentially larger reamers. In the second cohort the acetabulum
was only reamed peripherally, starting with a reamer the same size
as the native femoral head. Anteroposterior pelvic radiographs were
analysed for acetabular floor depth and vertical and horizontal position
of the COR.Aims
Patients and Methods
The aim of this study was to investigate the
possible benefit of large-head metal-on-metal bearing on a stem
for primary hip replacement compared with a 28 mm diameter conventional
metal-on-polyethylene bearing in a prospective randomised controlled
trial. We investigated cemented stem behaviour between these two
different bearings using Einzel-Bild-Röntgen-Analyse, clinical and
patient reported measures (Harris hip score, Western Ontario and
McMaster Universities osteoarthritis index, Short Form-36 and satisfaction)
and whole blood metal ion levels at two years. A power study indicated
that 50 hips were needed in each group to detect subsidence of >
5 mm at two years with a
p-value of <
0.05. Significant improvement (p <
0.001) was found in the mean
clinical and patient reported outcomes at two years for both groups.
Comparison of outcomes between the groups at two years showed no
statistically significant difference for mean stem migration, clinical
and patient reported outcomes; except overall patient satisfaction which
was higher for metal-on-metal group (p = 0.05). Metal ion levels
were raised above the Medicines and Healthcare products Regulatory
Agency advised safety level (7 µg per litre) in 20% of the metal-on-metal
group and in one patient in metal-on-polyethylene group (who had
a metal-on-metal implant on the contralateral side). Two patients
in the metal-on-metal group were revised, one for pseudotumour and
one for peri-prosthetic fracture. Use of large modular heads is associated with a risk of raised
whole blood metal ion levels despite using a proven bearing from
resurfacing. The head-neck junction or excess stem micromotion are
possibly the weak links warranting further research.
Obesity is a risk factor for complications following
many orthopaedic procedures. The purpose of this study was to investigate
whether obesity was an independent risk factor increasing the rate
of complications following periacetabular osteotomy (PAO) and to
determine whether radiographic correction after PAO was affected
by obesity. We retrospectively collected demographic, clinical and radiographic
data on 280 patients (231 women; 82.5% and 49 men; 17.5%) who were
followed for a mean of 48 months (12 to 60) after PAO. A total of
65 patients (23.2%) were obese (body mass index (BMI) >
30 kg/m2).
Univariate and multivariate analysis demonstrated that BMI was an independent
risk factor associated with the severity of the complications. The
average probability of a patient developing a major complication
was 22% (95% confidence interval (CI) 11.78 to 38.21) for an obese
patient compared with 3% (95% CI 1.39 to 6.58) for a non-obese patient
The odds of a patient developing a major complication were 11 times
higher (95% CI 4.71 to 17.60, p <
0.0001) for an obese compared
with a non-obese patient. Following PAO surgery, there was no difference in radiographic
correction between obese and non-obese patients. PAO procedures
in obese patients correct the deformity effectively but are associated
with an increased rate of complications. Cite this article:
Our aim was to report survivorship data and lessons learned with
the Corail/Pinnacle cementless total hip arthroplasty (THA) system. Between August 2005 and March 2015, a total of 4802 primary cementless
Corail/Pinnacle THAs were performed in 4309 patients. In March 2016,
we reviewed these hips from a prospectively maintained database.Aims
Patients and Methods
One method of femoral head preservation following avascular necrosis
(AVN) is core decompression and insertion of a tantalum rod. However,
there may be a high failure rate associated with this procedure.
The purpose of this study was to document the clinical and radiological
outcomes following total hip arthroplasty (THA) subsequent to failed
tantalum rod insertion. A total of 37 failed tantalum rods requiring total hip arthroplasty
were identified from a prospective database. There were 21 hips
in 21 patients (12 men and nine women, mean age 37 years, 18 to
53) meeting minimum two year clinical and radiographic follow-up
whose THAs were carried out between November 2002 and April 2013
(mean time between tantalum rod implantation and conversion to a
THA was 26 months, 6 to 72). These were matched by age and gender
to individuals (12 men, nine women, mean age 40 years, 18 to 58)
receiving THA for AVN without prior tantalum rod insertion.Aims
Patients and Methods
Increased concentrations of metal ions after metal-on-metal resurfacing arthroplasty of the hip remain a concern. Although there has been no proven link to long-term health problems or early prosthetic failure, variables associated with high metal ion concentrations should be identified and, if possible, corrected. Our study provides data on metal ion levels from a series of 76 consecutive patients (76 hips) after resurfacing arthroplasty with the Articular Surface Replacement. Chromium and cobalt ion concentrations in the whole blood of patients with smaller (≤ 51 mm) femoral components were significantly higher than in those with the larger (≥ 53 mm) components (p <
0.01). Ion concentrations in the former group were significantly related to the inclination (p = 0.01) and anteversion (p = 0.01) of the acetabular component. The same relationships were not significant in the patients with larger femoral components (p = 0.61 and p = 0.49, respectively). Accurate positioning of the acetabular component intra-operatively is essential in order to reduce the concentration of metal ions in the blood after hip resurfacing arthroplasty with the Articular Surface Replacement implant.
Second-generation metal-on-metal bearings were introduced as a response to the considerable incidence of wear-induced failures associated with conventional replacements, especially in young patients. We present the results at ten years of a consecutive series of patients treated using a metal-on-metal hip resurfacing. A distinct feature of the bearings used in our series was that they had been subjected to double-heat treatments during the post-casting phase of their manufacture. In the past these bearings had not been subjected to thermal treatments, making this a unique metal-on-metal bearing which had not been used before in clinical practice. We report the outcome of 184 consecutive hips (160 patients) treated using a hybrid-fixed metal-on-metal hip resurfacing during 1996. Patients were invited for a clinicoradiological follow-up at a minimum of ten years. The Oxford hip score and anteroposterior and lateral radiographs were obtained. The mean age at operation was 54 years (21 to 75). A series of 107 consecutive hips (99 patients) who received the same prosthesis, but subjected to a single thermal treatment after being cast, between March 1994 and December 1995, were used as a control group for comparison. In the 1994 to 1995 group seven patients (seven hips) died from unrelated causes and there were four revisions (4%) for osteolysis and aseptic loosening. In the 1996 group nine patients died at a mean of 6.9 years after operation because of unrelated causes. There were 30 revisions (16%) at a mean of 7.3 years (1.2 to 10.9), one for infection at 1.2 years and 29 for osteolysis and aseptic loosening. Furthermore, in the latter group there were radiological signs of failure in 27 (24%) of the 111 surviving hips. The magnitude of the problem of osteolysis and aseptic loosening in the 1996 cohort did not become obvious until five years after the operation. Our results indicate that double-heat treatments of metal-on-metal bearings can lead to an increased incidence of wear-induced osteolysis.
We hypothesised that the removal of the subchondral
bone plate (SCBP) for cemented acetabular component fixation in
total hip arthroplasty (THA) offers advantages over retention by
improving the cement-bone interface, without jeopardising implant
stability. We have previously published two-year follow-up data
of a randomised controlled trial (RCT), in which 50 patients with
primary osteoarthritis were randomised to either retention or removal
of the SCBP. The mean age of the retention group (n = 25, 13 males)
was 70.0 years ( Cite this article:
We report the survival, radiological and functional outcomes of a single surgeon series of his first 144 consecutive Birmingham hip resurfacing procedures (130 patients) at a minimum of ten years. There were ten revisions during this time. Although no patients were lost to follow-up some did not complete the scoring assessment or undergo radiological assessment at ten years. The ten-year survival for male patients was 98.0% (95% confidence interval 95.2 to 100). The ten-year survival for the total cohort with aseptic revision as the endpoint was 95.5% (95% confidence interval 91.8 to 99.0) and including revisions for sepsis was 93.5% (95% confidence interval 89.2 to 97.6). The median modified Oxford hip score at ten years was 4.2% (interquartile range 0 to 19) and the median University of California, Los Angeles score was 7.0 (interquartile range 5.0 to 8.0). This study confirms the midterm reports that metal-on-metal hip resurfacing using the Birmingham Hip provides a durable alternative to total hip replacement, particularly in younger male patients wishing to maintain a high level of function, with low risk of revision for at least ten years.
Long-term clinical outcomes for ceramic-on-ceramic (CoC) bearings
are encouraging. However, there is a risk of squeaking. Guidelines
for the orientation of the acetabular component are defined from
static imaging, but the position of the pelvis and thus the acetabular
component during activities associated with edge-loading are likely
to be very different from those measured when the patient is supine.
We assessed the functional orientation of the acetabular component. A total of 18 patients with reproducible squeaking in their CoC
hips during deep flexion were investigated with a control group
of 36 non-squeaking CoC hips. The two groups were matched for the
type of implant, the orientation of the acetabular component when
supine, the size of the femoral head, ligament laxity, maximum hip
flexion and body mass index. Aims
Patients and Methods
We conducted a longitudinal study including patients with the same type of primary hybrid total hip replacement and evaluated patient activity and femoral osteolysis at either five or ten years post-operatively. Activity was measured using the University of California, Los Angeles scale. The primary outcome was the radiological assessment of femoral osteolysis. Secondary outcomes were revision of the femoral component for aseptic loosening and the patients’ quality of life. Of 503 hip replacements in 433 patients with a mean age of 67.7 years (30 to 91), 241 (48%) were seen at five and 262 (52%) at ten years post-operatively. Osteolytic lesions were identified in nine of 166 total hip replacements (5.4%) in patients with low activity, 21 of 279 (7.5%) with moderate activity, and 14 of 58 (24.1%) patients with high activity. The risk of osteolysis increased with participation in a greater number of sporting activities. In multivariate logistic regression adjusting for age, gender, body mass index and the inclination angle of the acetabular component, the adjusted odds ratio for osteolysis comparing high In conclusion, of patients engaged in high activity, 24% had developed femoral osteolysis five to ten years post-operatively.
Pseudotumours are a rare complication of hip resurfacing. They are thought to be a response to metal debris which may be caused by edge loading due to poor orientation of the acetabular component. Our aim was to determine the optimal acetabular orientation to minimise the risk of pseudotumour formation. We matched 31 hip resurfacings revised for pseudotumour formation with 58 controls who had a satisfactory outcome from this procedure. The radiographic inclination and anteversion angles of the acetabular component were measured on anteroposterior radiographs of the pelvis using Einzel-Bild-Roentgen-Analyse software. The mean inclination angle (47°, 10° to 81°) and anteversion angle (14°, 4° to 34°) of the pseudotumour cases were the same (p = 0.8, p = 0.2) as the controls, 46° (29° to 60°) and 16° (4° to 30°) respectively, but the variation was greater. Assuming an accuracy of implantation of ± 10° about a target position, the optimal radiographic position was found to be approximately 45° of inclination and 20° of anteversion. The incidence of pseudotumours inside the zone was four times lower (p = 0.007) than outside the zone. In order to minimise the risk of pseudotumour formation we recommend that surgeons implant the acetabular component at an inclination of 45° (± 10) and anteversion of 20° (± 10) on post-operative radiographs. Because of differences between the radiographic and the operative angles, this may be best achieved by aiming for an inclination of 40° and an anteversion of 25°.
The issues surrounding raised levels of metal
ions in the blood following large head metal-on-metal total hip replacement
(THR), such as cobalt and chromium, have been well documented. Despite
the national popularity of uncemented metal-on-polyethylene (MoP)
THR using a large-diameter femoral head, few papers have reported
the levels of metal ions in the blood following this combination.
Following an isolated failure of a 44 mm Trident–Accolade uncemented
THR associated with severe wear between the femoral head and the
trunnion in the presence of markedly elevated levels of cobalt ions
in the blood, we investigated the relationship between modular femoral head
diameter and the levels of cobalt and chromium ions in the blood
following this THR. A total of 69 patients received an uncemented Trident–Accolade
MoP THR in 2009. Of these, 43 patients (23 men and 20 women, mean
age 67.0 years) were recruited and had levels of cobalt and chromium
ions in the blood measured between May and June 2012. The patients
were then divided into three groups according to the diameter of
the femoral head used: 12 patients in the 28 mm group (controls),
18 patients in the 36 mm group and 13 patients in the 40 mm group.
A total of four patients had identical bilateral prostheses in situ
at phlebotomy: one each in the 28 mm and 36 mm groups and two in
the 40 mm group. There was a significant increase in the mean levels of cobalt
ions in the blood in those with a 36 mm diameter femoral head compared
with those with a 28 mm diameter head (p = 0.013). The levels of
cobalt ions in the blood were raised in those with a 40 mm diameter
head but there was no statistically significant difference between
this group and the control group (p = 0.152). The levels of chromium
ions in the blood were normal in all patients. The clinical significance of this finding is unclear, but we
have stopped using femoral heads with a diameter of ≤ 36 mm, and
await further larger studies to clarify whether, for instance, this
issue particularly affects this combination of components. Cite this article:
The aim of this study was to establish the natural
course of unrevised asymptomatic pseudotumours after metal-on-metal
(MoM) hip resurfacing during a six- to 12-month follow-up period.
We used repeated metal artefact reduction sequence (MARS)-magnetic
resonance imaging (MRI), serum metal ion analysis and clinical examination to
study 14 unrevised hips (mean patient age 52.7 years, 46 to 68,
5 female, 7 male) with a pseudotumour and 23 hips (mean patient
age 52.8 years, 38 to 69, 7 female, 16 male) without a pseudotumour.
The mean post-operative time to the first MARS-MRI scan was 4.3 years
(2.2 to 8.3), and mean time between the first and second MARS-MRI scan
was eight months (6 to 12). At the second MRI scan, the grade of
severity of the pseudotumour had not changed in 35 hips. One new
pseudotumour (Anderson C2 score, moderate) was observed, and one
pseudotumour was downgraded from C2 (moderate) to C1 (mild). In
general, the characteristics of the pseudotumours hardly changed. Repeated MARS-MRI scans within one year in patients with asymptomatic
pseudotumours after MoM hip resurfacing showed little or no variation.
In 23 patients without pseudotumour, one new asymptomatic pseudotumour
was detected. This is the first longitudinal study on the natural history of
pseudotumours using MARS-MRI scans in hip resurfacing, and mirrors
recent results for 28 mm diameter MoM total hip replacement. Cite this article:
Large-head metal-on-metal (MoM) total hip replacements
(THR) have given rise to concern. Comparative studies of small-head
MoM THRs over a longer follow-up period are lacking. Our objective
was to compare the incidence of complications such as infection,
dislocation, revision, adverse local tissue reactions, mortality
and radiological and clinical outcomes in small-head (28 mm) MoM
and ceramic-on-polyethylene (CoP) THRs up to 12 years post-operatively. A prospective cohort study included 3341 THRs in 2714 patients.
The mean age was 69.1 years (range 24 to 98) and 1848 (55.3%) were
performed in women, with a mean follow-up of 115 months (18 to 201).
There were 883 MoM and 2458 CoP bearings. Crude incidence rates
(cases/1000 person-years) were: infection 1.3 In conclusion, we found similar results for small-head MoM and
CoP bearings up to ten years post-operatively, but after ten years
MoM THRs had a higher risk of all-cause revision. Furthermore, the
presence of an adverse response to metal debris seen in the small-head
MOM group at revision is a cause for concern. Cite this article:
In developmental dysplasia of the hip (DDH),
a bone defect is often observed superior to the acetabulum after
the reconstruction at the level of the true acetabulum during total
hip replacement (THR). However, the essential amount of uncemented
acetabular component coverage required for a satisfactory outcome
remains controversial. The purpose of this study was to assess the
stability and function of acetabular components with a lack of coverage >
30% (31% to 50%). A total of 760 DDH patients underwent THR with
acetabular reconstruction at the level of the true floor. Lack of
coverage above the acetabular component of >
30% occurred in 56
patients. Intra-operatively, autogenous morcellised bone grafts
were used to fill the uncovered portion. Other than two screws inserted through
the acetabular shell, no additional structural supports were used
in these hips. In all, four patients were lost to follow-up. Therefore,
52 patients (52 hips, 41 women and 11 men) with a mean age of 60.1
years (42 to 78) were available for this study at a mean of 4.8
years (3 to 7). There were no instances of prosthesis revision or
marked loosening during the follow-up. The Harris hip score improved
from a mean of 40.7 points ( Cite this article:
The dysplasia cup, which was devised as an adjunct to the Birmingham Hip Resurfacing system, has a hydroxyapatite-coated porous surface and two supplementary neutralisation screws to provide stable primary fixation, permit early weight-bearing, and allow incorporation of morcellised autograft without the need for structural bone grafting. A total of 110 consecutive dysplasia resurfacing arthroplasties in 103 patients (55 men and 48 women) performed between 1997 and 2000 was reviewed with a minimum follow-up of six years. The mean age at operation was 47.2 years (21 to 62) and 104 hips (94%) were Crowe grade II or III. During the mean follow-up of 7.8 years (6 to 9.6), three hips (2.7%) were converted to a total hip replacement at a mean of 3.9 years (2 months to 8.1 years), giving a cumulative survival of 95.2% at nine years (95% confidence interval 89 to 100). The revisions were due to a fracture of the femoral neck, a collapse of the femoral head and a deep infection. There was no aseptic loosening or osteolysis of the acetabular component associated with either of the revisions performed for failure of the femoral component. No patient is awaiting a revision. The median Oxford hip score in 98 patients with surviving hips at the final review was 13 and the 10th and the 90th percentiles were 12 and 23, respectively.
We report the long-term survival and functional
outcome of the Birmingham Hip Resurfacing (BHR) in patients aged <
50 years at operation, and explore the factors affecting survival.
Between 1997 and 2006, a total of 447 BHRs were implanted in 393
patients (mean age 41.5 years (14.9 to 49.9)) by one designing surgeon.
The mean follow-up was 10.1 years (5.2 to 14.7), with no loss to
follow-up. In all, 16 hips (3.6%) in 15 patients were revised, giving
an overall cumulative survival of 96.3% (95% confidence interval
(CI) 93.7 to 98.3) at ten years and 94.1% (95% CI 84.9 to 97.3)
at 14 years. Using aseptic revision as the endpoint, the survival
for men with primary osteoarthritis (n = 195) was 100% (95% CI 100
to 100) at both ten years and 14 years, and in women with primary
osteoarthritis (n = 109) it was 96.1% (95% CI 90.1 to 99.9) at ten
years and 91.2% (95% CI 68.6 to 98.7) at 14 years. Female gender
(p = 0.047) and decreasing femoral head size (p = 0.044) were significantly
associated with an increased risk of revision. The median Oxford
hip score (OHS, modified as a percentage with 100% indicating worst
outcome) at last follow-up was 4.2% (46 of 48; interquartile range
(IQR) 0% to 24%) and the median University of California, Los Angeles
(UCLA) score was 6.0 (IQR 5 to 8). Men had significantly better
OHS (p = 0.02) and UCLA scores (p = 0.01) than women. The BHR provides
excellent survival and functional results in men into the second
decade, with good results achieved in appropriately selected women. Cite this article:
We carried out a prospective randomised study designed to compare the penetration rate of acetabular polyethylene inserts of identical design but different levels of cross-linking at a minimum of four years follow-up. A total of 102 patients (102 hips) were randomised to receive either highly cross-linked Durasul, or contemporary Sulene polyethylene inserts at total hip replacement. A single blinded observer used the Martell system to assess penetration of the femoral head. At a mean follow-up of 4.9 years (4.2 to 6.1) the mean femoral head penetration rate was 0.025 mm/year (
We report an independent prospective review of the first 230 Birmingham hip resurfacings in 212 patients at a mean follow-up of five years (4 to 6). Two patients, one with a loose acetabular component and the other with suspected avascular necrosis of the femoral head, underwent revision. There were two deaths from unrelated causes and one patient was lost to follow-up. The survivorship with the worst-case scenario was 97.8% (95% confidence interval 95.8 to 99.5). The mean Harris hip score improved significantly (paired On radiological review at five years, one patient had a progressive lucent line around the acetabular component and six had progressive lucent lines around the femoral component. A total of 18 femoral components (8%) had migrated into varus and those with lucent lines present migrated a mean of 3.8° (1.02° to 6.54°) more than the rest. Superolateral notching of the femoral neck and reactive sclerosis at the tip of the peg of the femoral component were associated with the presence of lucent lines (chi-squared test, p <
0.05), but not with migration of the femoral component, and are of unknown significance. Our results with the Birmingham hip resurfacing continue to be satisfactory at a mean follow-up of five years.