We report the long-term outcome of a modified second-generation
We have evaluated the effect of vacuum aspiration of the iliac wing on the osseointegration of cement into the acetabulum. We entered a total of 40 patients undergoing primary total hip arthroplasty into two consecutive study groups. Group 1 underwent acetabular cement pressurisation for 60 seconds before insertion of the acetabular component. Group 2 had the same pressurisation with simultaneous vacuum suction of the ilium using an iliac-wing aspirator. Standard post-operative radiographs were reviewed blindly to assess the penetration of cement into the iliac wing. Penetration was significantly greater in the group with aspiration of the iliac wing.
To assess the effect of improved methods of femoral cementing on the loosening rates in young patients, we reviewed 50 'second-generation' cemented hip arthroplasties in 44 patients aged 50 years or less. The femoral stems were all collared and rectangular in cross-section with rounded corners. The cement was delivered by a gun into a medullary canal occluded distally with a cement plug. A clinical and radiographic review was undertaken at an average of 12 years (10 to 14.8) and no patient was lost to follow-up. No femoral component was revised for aseptic loosening, and only one stem was definitely loose by radiographic criteria. By contrast, 11 patients had undergone revision for symptomatic aseptic loosening of the acetabular component and 11 more had radiographic signs of acetabular loosening.
Pathological fractures of the humerus are associated with pain, morbidity, loss of function and a diminished quality of life. We report our experience of stabilising these fractures using polymethylmethacrylate and non-locking plates. We undertook a retrospective review over 20 years of patients treated at a tertiary musculoskeletal oncology centre. Those who had undergone surgery for an impending or completed pathological humeral fracture with a diagnosis of metastatic disease or myeloma were identified from our database. There were 63 patients (43 men, 20 women) in the series with a mean age of 63 years (39 to 87). All had undergone intralesional curettage of the tumour followed by fixation with intramedullary polymethylmethacrylate and plating. Complications occurred in 14 patients (22.2%) and seven (11.1%) required re-operation. At the latest follow-up, 47 patients (74.6%) were deceased and 16 (25.4%) were living with a mean follow-up of 75 months (1 to 184). A total of 54 (86%) patients had no or mild pain and 50 (80%) required no or minimal assistance with activities of daily living. Of the 16 living patients none had pain and all could perform activities of daily living without assistance. Intralesional resection of the tumour, filling of the cavity with cement, and plate stabilisation of the pathological fracture gives immediate rigidity and allows an early return of function without the need for bony union. The patient’s local disease burden is reduced, which may alleviate tumour-related pain and slow the progression of the disease. The cemented-plate technique provides a reliable option for the treatment of pathological fractures of the humerus.
Aims. To achieve the functional benefits of the direct anterior (DA) approach and the fixation benefits of cemented replacement, this study combined the two techniques posing the following questions: does the limited access of the DA approach adversely affect the
Aims. Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line
The aim of this retrospective study was to assess the incidence of early periprosthetic femoral fracture (PFF) associated with Charnley-Kerboull (CK) femoral components cemented according to the ‘French paradox’ principles through the Hueter anterior approach (HAA) in patients older than 70 years. From a prospectively collected database, all short CK femoral components implanted consecutively from January 2018 to May 2022 through the HAA in patients older than 70 years were included. Exclusion criteria were age below 70 years, use of cementless femoral component, and approaches other than the HAA. A total of 416 short CK prostheses used by 25 surgeons with various levels of experience were included. All patients had a minimum of one-year follow-up, with a mean of 2.6 years (SD 1.1). The mean age was 77.4 years (70 to 95) and the mean BMI was 25.3 kg/m2 (18.4 to 43). Femoral anatomy was classified according to Dorr. The measured parameters included canal flare index, morphological cortical index, canal-calcar ratio, ilium-ischial ratio, and anterior superior iliac spine to greater trochanter (GT) distance.Aims
Methods
We present the medium-term results of hybrid total hip arthroplasties using pre-coated stems with a second-generation
Obtaining solid implant fixation is crucial in revision total knee arthroplasty (rTKA) to avoid aseptic loosening, a major reason for re-revision. This study aims to validate a novel grading system that quantifies implant fixation across three anatomical zones (epiphysis, metaphysis, diaphysis). Based on pre-, intra-, and postoperative assessments, the novel grading system allocates a quantitative score (0, 0.5, or 1 point) for the quality of fixation achieved in each anatomical zone. The criteria used by the algorithm to assign the score include the bone quality, the size of the bone defect, and the type of fixation used. A consecutive cohort of 245 patients undergoing rTKA from 2012 to 2018 were evaluated using the current novel scoring system and followed prospectively. In addition, 100 first-time revision cases were assessed radiologically from the original cohort and graded by three observers to evaluate the intra- and inter-rater reliability of the novel radiological grading system.Aims
Methods
This multicentre retrospective observational study’s aims were to investigate whether there are differences in the occurrence of radiolucent lines (RLLs) following total knee arthroplasty (TKA) between the conventional Attune baseplate and its successor, the novel Attune S+, independent from other potentially influencing factors; and whether tibial baseplate design and presence of RLLs are associated with differing risk of revision. A total of 780 patients (39% male; median age 70.7 years (IQR 62.0 to 77.2)) underwent cemented TKA using the Attune Knee System) at five centres, and with the latest radiograph available for the evaluation of RLL at between six and 36 months from surgery. Univariate and multivariate logistic regression models were performed to assess associations between patient and implant-associated factors on the presence of tibial and femoral RLLs. Differences in revision risk depending on RLLs and tibial baseplate design were investigated with the log-rank test.Aims
Methods
We have investigated the mid-term outcome of total shoulder replacement using a keeled cemented glenoid component and a modern cementing technique with regard to the causes of failure and loosening of the components. Between 1997 and 2003 we performed 96 total shoulder replacements on 88 patients, 24 men and 64 women with a mean age of 69.7 years (31 to 82). The minimum follow-up was five years and at the time of review 87 shoulders (77 patients) were examined at a mean follow-up of 89.1 months (60 to 127). Cumulative survival curves were generated with re-operations (accomplished and planned), survivorship of the proshesis, loosening of the glenoid (defined as tilt >
5° or subsidence >
5 mm), the presence of radiolucent lines and a Constant score of <
30 as the endpoints. There were two re-operations not involving revision of the implants and the survival rate of the prosthesis was 100.0% for the follow-up period, with an absolute Constant score of >
30 as the endpoint the survival rate was 98%. Radiological glenoid loosening was 9% after five years, and 33% after nine years. There was an incidence of 8% of radiolucent lines in more than three of six zones in the immediate post-operative period, of 37.0% after the first year which increased to 87.0% after nine years. There was no correlation between the score of Boileau and the total Constant score at the latest follow-up, but there was correlation between glenoid loosening and pain (p = 0.001). We found that total shoulder replacement had an excellent mid-term survivorship and clinical outcome. The surgical and
Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by radiostereometric analysis (RSA), clinical outcomes, patient-reported outcome measures (PROMs), and radiological outcomes. A total of 74 primary TKAs were included in this single-blind RCT. RSA examinations were performed, and PROMs and clinical outcomes were collected immediate postoperatively, and at three, six, 12, 24, and 60 months’ follow-up. Radiolucent lines (RLLs) were measured in standard anteroposterior radiographs at six weeks, and 12 and 60 months postoperatively.Aims
Methods
Large bone defects resulting from osteolysis, fractures, osteomyelitis, or metastases pose significant challenges in acetabular reconstruction for total hip arthroplasty. This study aimed to evaluate the survival and radiological outcomes of an acetabular reconstruction technique in patients at high risk of reconstruction failure (i.e. periprosthetic joint infection (PJI), poor bone stock, immunosuppressed patients), referred to as Hip Reconstruction In Situ with Screws and Cement (HiRISC). This involves a polyethylene liner embedded in cement-filled bone defects reinforced with screws and/or plates for enhanced fixation. A retrospective chart review of 59 consecutive acetabular reconstructions was performed by four surgeons in a single institution from 18 October 2018 to 5 January 2023. Cases were classified based on the Paprosky classification, excluding type 1 cases (n = 26) and including types 2 or 3 for analysis (n = 33). Radiological loosening was evaluated by an orthopaedic surgeon who was not the operating surgeon, by comparing the immediate postoperative radiographs with the ones at latest follow-up. Mean follow-up was 557 days (SD 441; 31 to 1,707).Aims
Methods
Loosening of components after total knee arthroplasty (TKA) can be associated with the development of radiolucent lines (RLLs). The aim of this study was to assess the rate of formation of RLLs in the cemented original design of the ATTUNE TKA and their relationship to loosening. A systematic search was undertaken using the Cochrane methodology in three online databases: MEDLINE, Embase, and CINAHL. Studies were screened against predetermined criteria, and data were extracted. Available National Joint Registries in the Network of Orthopaedic Registries of Europe were also screened. A random effects model meta-analysis was undertaken.Aims
Methods
The aim of this study was to evaluate the survival of a collarless, straight, hydroxyapatite-coated femoral stem in total hip arthroplasty (THA) at a minimum follow-up of 20 years. We reviewed the results of 165 THAs using the Omnifit HA system in 138 patients, performed between August 1993 and December 1999. The mean age of the patients at the time of surgery was 46 years (20 to 77). Avascular necrosis was the most common indication for THA, followed by ankylosing spondylitis and primary osteoarthritis. The mean follow-up was 22 years (20 to 31). At 20 and 25 years, 113 THAs in 91 patients and 63 THAs in 55 patients were available for review, respectively, while others died or were lost to follow-up. Kaplan-Meier analysis was performed to evaluate the survival of the stem. Radiographs were reviewed regularly, and the stability of the stem was evaluated using the Engh classification.Aims
Methods
Polished taper-slip (PTS) cemented stems have an excellent clinical track record and are the most common stem type used in primary total hip arthroplasty (THA) in the UK. Due to low rates of aseptic loosening, they have largely replaced more traditional composite beam (CB) cemented stems. However, there is now emerging evidence from multiple joint registries that PTS stems are associated with higher rates of postoperative periprosthetic femoral fracture (PFF) compared to their CB stem counterparts. The risk of both intraoperative and postoperative PFF remains greater with uncemented stems compared to either of these cemented stem subtypes. PFF continues to be a devastating complication following primary THA and is associated with high complication and mortality rates. Recent efforts have focused on identifying implant-related risk factors for PFF in order to guide preventative strategies, and therefore the purpose of this article is to present the current evidence on the effect of cemented femoral stem design on the risk of PFF. Cite this article:
The aim of this study was to report the long-term follow-up of cemented short Exeter femoral components when used in primary total hip arthroplasty (THA). We included all primary 394 THAs with a cemented short Exeter femoral component (≤ 125 mm) used in our tertiary referral centre between October 1993 and December 2021. A total of 83 patients (21%) were male. The median age of the patients at the time of surgery was 42 years (interquartile range (IQR) 30 to 55). The main indication for THA was a childhood hip disease (202; 51%). The median follow-up was 6.7 years (IQR 3.1 to 11.0). Kaplan-Meier survival analyses were performed to determine the rates of survival with femoral revision for any indication, for septic loosening, for fracture of the femoral component and for aseptic loosening as endpoints. The indications for revision were evaluated. Fractures of the femoral component were described in detail.Aims
Methods
Osteoporosis can determine surgical strategy for total hip arthroplasty (THA), and perioperative fracture risk. The aims of this study were to use hip CT to measure femoral bone mineral density (BMD) using CT X-ray absorptiometry (CTXA), determine if systematic evaluation of preoperative femoral BMD with CTXA would improve identification of osteopenia and osteoporosis compared with available preoperative dual-energy X-ray absorptiometry (DXA) analysis, and determine if improved recognition of low BMD would affect the use of cemented stem fixation. Retrospective chart review of a single-surgeon database identified 78 patients with CTXA performed prior to robotic-assisted THA (raTHA) (Group 1). Group 1 was age- and sex-matched to 78 raTHAs that had a preoperative hip CT but did not have CTXA analysis (Group 2). Clinical demographics, femoral fixation method, CTXA, and DXA data were recorded. Demographic data were similar for both groups.Aims
Methods
Despite higher rates of revision after total hip arthroplasty (THA) being reported for uncemented stems in patients aged > 75 years, they are frequently used in this age group. Increased mortality after cemented fixation is often used as a justification, but recent data do not confirm this association. The aim of this study was to investigate the influence of the design of the stem and the type of fixation on the rate of revision and immediate postoperative mortality, focusing on the age and sex of the patients. A total of 333,144 patients with primary osteoarthritis (OA) of the hip who underwent elective THA between November 2012 and September 2022, using uncemented acetabular components without reconstruction shells, from the German arthroplasty registry were included in the study. The revision rates three years postoperatively for four types of stem (uncemented, uncemented with collar, uncemented short, and cemented) were compared within four age groups: < 60 years (Young), between 61 and 70 years (Mid-I), between 71 and 80 years (Mid-II), and aged > 80 years (Old). A noninferiority analysis was performed on the most frequently used designs of stem.Aims
Methods
Refobacin Bone Cement R and Palacos Overall, 75 patients were included in the study and 71 were available at two years postoperatively. Prior to surgery, they were randomized to one of the three combinations studied: Palacos cement with use of the Optivac mixing system, Refobacin with use of the Optivac system, and Refobacin with use of the Optipac system. Cemented MS30 stems and cemented Exceed acetabular components were used in all hips. Postoperative radiographs were used to assess the quality of the cement mantle according to Barrack et al, and the position and migration of the femoral stem. Harris Hip Score, Oxford Hip Score, Forgotten Joint Score, and University of California, Los Angeles Activity Scale were collected.Aims
Methods
Arthroplasties of the elbow, including total elbow arthroplasty, radial head arthroplasty, distal humeral hemiarthroplasty, and radiocapitellar arthroplasty, are rarely undertaken. This scoping review aims to outline the current research in this area to inform the development of future research. A scoping review was undertaken adhering to the Joanna Briggs Institute guidelines using Medline, Embase, CENTRAL, and trial registries, limited to studies published between 1 January 1990 and 7 February 2021. Endnote software was used for screening and selection, and included randomized trials, non-randomized controlled trials, prospective and retrospective cohort studies, case-control studies, analytical cross-sectional studies, and case series of ten or more patients reporting the clinical outcomes of elbow arthroplasty. The results are presented as the number of types of studies, sample size, length of follow-up, clinical outcome domains and instruments used, sources of funding, and a narrative review.Aims
Methods
The aim of this study is to report the long-term outcomes of instrumented femoral revisions with impaction allograft bone grafting (IBG) using the X-change femoral revision system at 30 years after introduction of the technique. We updated the outcomes of our previous study, based on 208 consecutive revisions using IBG and the X-change femoral revision system in combination with a cemented polished stem, performed in our tertiary care institute between 1991 and 2007. Kaplan-Meier survival analyses were used to determine the survival rate of the revisions with endpoint revision for any reason and aseptic loosening. Secondary outcomes were radiological loosening and patient-reported outcome measures.Aims
Methods
The aim of this study was to explore the relationship between reason for revision total hip arthroplasty (rTHA) and outcomes in terms of patient-reported outcome measures (PROMs). We reviewed a prospective cohort of 647 patients undergoing full or partial rTHA at a single high-volume centre with a minimum of two years’ follow-up. The reasons for revision were classified as: infection; aseptic loosening; dislocation; structural failure; and painful THA for other reasons. PROMs (modified Oxford Hip Score (mOHS), EuroQol five-dimension three-level health questionnaire (EQ-5D-3L) score, and visual analogue scales for pain during rest and activity), complication rates, and failure rates were compared among the groups.Aims
Methods
The optimum
We performed 83 consecutive cemented revision total hip arthroplasties in 77 patients between 1977 and 1983 using improved
We aimed to evaluate the long-term outcome of highly cross-linked polyethylene (HXLPE) cemented acetabular components and assess whether any radiolucent lines (RLLs) which arose were progressive. We retrospectively reviewed 170 patients who underwent 187 total hip arthroplasties at two hospitals with a minimum follow-up of ten years. All interventions were performed using the same combination of HXLPE cemented acetabular components with femoral stems made of titanium alloy. Kaplan-Meier survival analysis was performed for the primary endpoint of acetabular component revision surgery for any reason and secondary endpoint of the appearance of RLLs. RLLs that had appeared once were observed over time. We statistically assessed potential relationships between RLLs and a number of factors, including the technique of femoral head autografting and the Japanese Orthopaedic Association score.Aims
Methods
The aim of this study was to determine the outcome of all primary total hip arthroplasties (THAs) and their subsequent revision procedures in patients aged under 50 years performed at our institution. All 1,049 primary THAs which were undertaken in 860 patients aged under 50 years between 1988 and 2018 in our tertiary care institution were included. We used cemented implants in both primary and revision surgery. Impaction bone grafting was used in patients with acetabular or femoral bone defects. Kaplan-Meier analyses were used to determine the survival of primary and revision THA with the endpoint of revision for any reason, and of revision for aseptic loosening.Aims
Methods
We examined the roles of methylmethacrylate (MMA) monomer and
We studied prospectively the long-term results of the Charnley Elite-Plus femoral stem in 184 consecutive young patients (194 hips). There were 130 men and 54 women with a mean age of 49.1 years (21 to 60). The predominant diagnosis was osteonecrosis of the femoral head (63.6%, 117 patients). Clinical and radiological evaluation was undertaken at each follow-up. The mean follow-up was 11.2 years (10 to 12). The mean pre-operative Harris hip score was 43.4 (12 to 49) which improved to 91 (59 to 100) at the final follow-up. The survival of the femoral stem at 12 years was 99% with revision as the end-point. The mean annual linear wear of the polyethylene liner was 0.17 mm (0.13 to 0.22). The prevalence of acetabular osteolysis was 10.8% (21 hips) and osteolysis of the calcar femorale 12.9% (25 hips). A third-generation
A complete cement mantle is important for the longevity of a total hip replacement. In the minimally-invasive direct anterior approach used at the Innsbruck University hospital, the femoral component has to be inserted into the femoral canal by an angulated movement. In a cadaver study, the quality and the extent of the cement mantle surrounding 13 Exeter femoral components implanted straight through a standard anterolateral transgluteal approach were compared with those of 13 similar femoral components implanted in an angulated fashion through a direct anterior approach. A third-generation
We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary
We conducted a longitudinal study including patients with the same type of primary hybrid total hip replacement and evaluated patient activity and femoral osteolysis at either five or ten years post-operatively. Activity was measured using the University of California, Los Angeles scale. The primary outcome was the radiological assessment of femoral osteolysis. Secondary outcomes were revision of the femoral component for aseptic loosening and the patients’ quality of life. Of 503 hip replacements in 433 patients with a mean age of 67.7 years (30 to 91), 241 (48%) were seen at five and 262 (52%) at ten years post-operatively. Osteolytic lesions were identified in nine of 166 total hip replacements (5.4%) in patients with low activity, 21 of 279 (7.5%) with moderate activity, and 14 of 58 (24.1%) patients with high activity. The risk of osteolysis increased with participation in a greater number of sporting activities. In multivariate logistic regression adjusting for age, gender, body mass index and the inclination angle of the acetabular component, the adjusted odds ratio for osteolysis comparing high vs moderate activity was 3.6 (95% confidence interval 1.6 to 8.3). Stratification for the
To establish our early clinical results of a new total knee arthroplasty (TKA) tibial component introduced in 2013 and compare it to other designs in use at our hospital during the same period. This is a retrospective study of 166 (154 patients) consecutive cemented, fixed bearing, posterior-stabilized (PS) TKAs (ATTUNE) at one hospital performed by five surgeons. These were compared with a reference cohort of 511 knees (470 patients) of other designs (seven manufacturers) performed at the same hospital by the same surgeons. There were no significant differences in age, sex, BMI, or follow-up times between the two cohorts. The primary outcome was revision performed or pending.Aims
Methods
Uncemented metal acetabular components show good osseointegration, but material stiffness causes stress shielding and retroacetabular bone loss. Cemented monoblock polyethylene components load more physiologically; however, the cement bone interface can suffer fibrous encapsulation and loosening. It was hypothesized that an uncemented titanium-sintered monoblock polyethylene component may offer the optimum combination of osseointegration and anatomical loading. A total of 38 patients were prospectively enrolled and received an uncemented monoblock polyethylene acetabular (pressfit) component. This single cohort was then retrospectively compared with previously reported randomized cohorts of cemented monoblock (cemented) and trabecular metal (trabecular) acetabular implants. The primary outcome measure was periprosthetic bone density using dual-energy x-ray absorptiometry over two years. Secondary outcomes included radiological and clinical analysis.Aims
Methods
The aims of this study were to develop an in vivo model of periprosthetic joint infection (PJI) in cemented hip hemiarthroplasty, and to monitor infection and biofilm formation in real-time. Sprague-Dawley rats underwent cemented hip hemiarthroplasty via the posterior approach with pre- and postoperative gait assessments. Infection with Aims
Methods
The aim of this study was to radiologically evaluate the quality of cement mantle and alignment achieved with a polished tapered cemented femoral stem inserted through the anterior approach and compared with the posterior approach. A comparative retrospective study of 115 consecutive hybrid total hip arthroplasties or cemented hemiarthroplasties in 110 patients, performed through anterior (n = 58) or posterior approach (n = 57) using a collarless polished taper-slip femoral stem, was conducted. Cement mantle quality and thickness were assessed in both planes. Radiological outcomes were compared between groups.Aims
Methods
Various surgical techniques have been described for total hip arthroplasty (THA) in patients with Crowe type III dislocated hips, who have a large acetabular bone defect. The aim of this study was to evaluate the long-term clinical results of patients in whom anatomical reconstruction of the acetabulum was performed using a cemented acetabular component and autologous bone graft from the femoral neck. A total of 22 patients with Crowe type III dislocated hips underwent 28 THAs using bone graft from the femoral neck between 1979 and 2000. A Charnley cemented acetabular component was placed at the level of the true acetabulum after preparation with bone grafting. All patients were female with a mean age at the time of surgery of 54 years (35 to 68). A total of 18 patients (21 THAs) were followed for a mean of 27.2 years (20 to 33) after the operation.Aims
Methods
The excellent long-term results for the first-generation Charnley stem may not apply to later versions with flanges. It seems possible that the early design functioned as a taper-slip system, as accepted in the Exeter prosthesis. Comparison with the requirements for the alternative composite-beam system for the femoral component shows considerable differences that have important implications. These include design, surface finish,
For many designs of total knee arthroplasty (TKA) it remains unclear whether cemented or uncemented fixation provides optimal long-term survival. The main limitation in most studies is a retrospective or non-comparative study design. The same is true for comparative trials looking only at the survival rate as extensive sample sizes are needed to detect true differences in fixation and durability. Studies using radiostereometric analysis (RSA) techniques have shown to be highly predictive in detecting late occurring aseptic loosening at an early stage. To investigate the difference in predicted long-term survival between cemented, uncemented, and hybrid fixation of TKA, we performed a randomized controlled trial using RSA. A total of 105 patients were randomized into three groups (cemented, uncemented, and hybrid fixation of the ACS Mobile Bearing (ACS MB) knee system, implantcast). RSA examinations were performed on the first day after surgery and at scheduled follow-up visits at three months, six months, one year, and two years postoperatively. Patient-reported outcome measures (PROMs) were obtained preoperatively and after two years follow-up. Patients and follow-up investigators were blinded for the result of randomization.Aims
Methods
This study aimed to compare the effect of antibiotic-loaded bone cement (ALBC) versus plain bone cement (PBC) on revision rates for periprosthetic joint infection (PJI) and all-cause revisions following primary elective total hip arthroplasty (THA) and total knee arthroplasty (TKA). MEDLINE, Embase, Web of Science, and Cochrane databases were systematically searched for studies comparing ALBC versus PBC, reporting on revision rates for PJI or all-cause revision following primary elective THA or TKA. A random-effects meta-analysis was performed. The study was registered in the International Prospective Register of Systematic Reviews (PROSPERO ID CRD42018107691).Aims
Methods
We have reviewed 97 consecutive primary hip replacements with a cemented femoral component and a porous-ingrowth acetabular component at a minimum five-year follow-up (average 6.5 years). The average Harris hip score was 93, and 85 hips had no pain or only slight pain. There had been no deterioration in the results since the two-year follow-up. The hybrid hip is successful for up to eight years and appears to be suitable for many patients. Long-term femoral fixation has been shown to improve with second-generation
Fatigue fractures which originate at stress-concentrating voids located at the implant-cement interface are a potential cause of septic loosening of cemented femoral components. Heating of the component to 44°C is known to reduce the porosity of the cement-prosthesis interface. The temperature of the cement-bone interface was recorded intra-operatively as 32.3°C. A simulated femoral model was devised to study the effect of heating of the component on the implant-cement interface. Heating of the implant and vacuum mixing have a synergistic effect on the porosity of the implant-cement interface, and heating also reverses the gradients of microhardness in the mantle. Heating of the implant also reduces porosity at the interface depending on the temperature. A minimum difference in temperature between the implant and the bone of 3°C was required to produce this effect. The optimal difference was 7°C, representing a balance between maximal reduction of porosity and an increased risk of thermal injury. Using contemporary
We selected 20 matched pairs of patients who had had total hip arthroplasty by the same surgeon using the same
We have compared the interface morphology at the stem-cement interface of standard Charnley stems with a satin finish (Ra = 0.75 μm) with identical stems which had been grit-blasted over their proximal third (Ra = 5.3 μm) to promote a proximal bond. The stems were cemented into cadaver femora using conventional contemporary
Vitamin E-infused highly crosslinked polyethylene (VEPE) has been introduced into total hip arthroplasty (THA) with the aim of further improving the wear characteristics of moderately and highly crosslinked polyethylenes (ModXLPE and HXLPE). There are few studies analyzing the outcomes of vitamin E-infused components in cemented arthroplasty, though early acetabular component migration has been reported. The aim of this study was to measure five-year polyethylene wear and acetabular component stability of a cemented VEPE acetabular component compared with a ModXLPE cemented acetabular component. In a prospective randomized controlled trial (RCT), we assessed polyethylene wear and acetabular component stability (primary outcome) with radiostereometric analysis (RSA) in 68 patients with reverse hybrid THA at five years follow-up. Patients were randomized to either a VEPE or a ModXLPE cemented acetabular component.Aims
Methods
Hip reconstruction after resection of a periacetabular chondrosarcoma is complex and associated with a high rate of complications. Previous reports have compared no reconstruction with historical techniques that are no longer used. The aim of this study was to compare the results of tantalum acetabular reconstruction to both historical techniques and no reconstruction. We reviewed 66 patients (45 males and 21 females) with a mean age of 53 years (24 to 81) who had undergone acetabular resection for chondrosarcoma. A total of 36 patients (54%) underwent acetabular reconstruction, most commonly with a saddle prosthesis (n = 13; 36%) or a tantalum total hip arthroplasty (THA) (n = 10; 28%). Mean follow-up was nine years (SD 4).Aims
Methods
We present the results, in terms of survival, clinical outcome, and radiological appearance at 20 years, in a cohort of 225 cemented Exeter Universal femoral components (Stryker, Newbury, UK) implanted in 207 patients, at a district general hospital. All patients in this study had a total hip arthroplasty (THA) using an Exeter Universal femoral component with a cemented (n = 215) or cementless (n = 10) acetabular component. Clinical and radiological data were collected prospectively at one year, five years, and every five years thereafter. Patients lost to radiological and clinical follow-up (five) were cross-referenced with National Joint Registry (NJR) data and general practitioner (GP) records to assess whether they had undergone revision for any reason.Aims
Methods
We reviewed the records of the long-term outcome of 208 Charnley and 982 Stanmore total hip replacements (THR) performed by or under the supervision of one surgeon from 1973 to 1987. The Stanmore implant had a better survival rate before revision at 14 years (86% to 79%, p = 0.004), but the difference only became apparent at ten years. The later Stanmore implants did better than the early ones (97% to 92% at ten years, p = 0.005), the improvement coinciding with the introduction of a new
We studied 108 patients (116 hips) who were followed for a minimum of six years (73 to 89 months) after primary total hip arthroplasty using an uncemented porous-coated anatomic hip. The average age of the patients at operation was 48.4 years, and the diagnosis was avascular necrosis of the femoral head in 46 hips, neglected femoral neck fracture in 27, osteoarthritis secondary to childhood pyogenic arthritis in 24 and to childhood tuberculous arthritis in five, and miscellaneous in 14. The average preoperative Harris hip score was 55, which improved to 91 at latest follow-up. All patients with loose femoral components or disabling thigh pain had received prostheses which were undersized in the coronal or the sagittal plane, or in both. No patient with a satisfactory fit in both coronal and sagittal planes had loosening of the femoral component or disabling thigh pain. Three acetabular components showed aseptic loosening and 20 showed excessive wear (5 to 11 mm) of the polyethylene liner. Excessive wear was related to young age, but not to body-weight, gender, primary diagnosis, hip score, or range of hip movement. There was a high incidence of osteolysis (38 of 116 hips, 33%). We recommend careful preoperative planning, with the use of a contemporary
During the COVID-19 pandemic, many patients continue to require urgent surgery for hip fractures. However, the impact of COVID-19 on perioperative outcomes in these high-risk patients remains unknown. The objectives of this study were to establish the effects of COVID-19 on perioperative morbidity and mortality, and determine any risk factors for increased mortality in patients with COVID-19 undergoing hip fracture surgery. This multicentre cohort study included 340 COVID-19-negative patients versus 82 COVID-19-positive patients undergoing surgical treatment for hip fractures across nine NHS hospitals in Greater London, UK. Patients in both treatment groups were comparable for age, sex, body mass index, fracture configuration, and type of surgery performed. Predefined perioperative outcomes were recorded within a 30-day postoperative period. Univariate and multivariate analysis were used to identify risk factors associated with increased risk of mortality.Aims
Methods
Total knee arthroplasty is an established treatment for knee osteoarthritis with excellent long-term results, but there remains controversy about the role of uncemented prostheses. We present the long-term results of a randomized trial comparing an uncemented tantalum metal tibial component with a conventional cemented component of the same implant design. Patients under the age of 70 years with symptomatic osteoarthritis of the knee were randomized to receive either an uncemented tantalum metal tibial monoblock component or a standard cemented modular component. The mean age at time of recruitment to the study was 63 years (50 to 70), 46 (51.1%) knees were in male patients, and the mean body mass index was 30.4 kg/m2 (21 to 36). The same cruciate retaining total knee system was used in both groups. All patients received an uncemented femoral component and no patients had their patella resurfaced. Patient outcomes were assessed preoperatively and postoperatively using the modified Oxford Knee Score, Knee Society Score, and 12-Item Short-Form Health Survey questionnaire (SF-12) score. Radiographs were analyzed using the American Knee Society Radiograph Evaluation score. Operative complications, reoperations, or revision surgery were recorded. A total of 90 knees were randomized and at last review 77 knees were assessed. In all, 11 patients had died and two were lost to follow-up.Aims
Methods
The aims of this study were to validate the outcome of total elbow arthroplasty (TEA) in patients with rheumatoid arthritis (RA), and to identify factors that affect the outcome. We searched PubMed, MEDLINE, Cochrane Reviews, and Embase from between January 2003 and March 2019. The primary aim was to determine the implant failure rate, the mode of failure, and risk factors predisposing to failure. A secondary aim was to identify the overall complication rate, associated risk factors, and clinical performance. A meta-regression analysis was completed to identify the association between each parameter with the outcome.Aims
Methods
Aseptic loosening of the tibial component is a frequent cause of failure in primary total knee arthroplasty (TKA). Management options include an isolated tibial revision or full component revision. A full component revision is frequently selected by surgeons unfamiliar with the existing implant or who simply wish to “start again”. This option adds morbidity compared with an isolated tibial revision. While isolated tibial revision has a lower morbidity, it is technically more challenging due to difficulties with exposure and maintaining prosthetic stability. This study was designed to compare these two reconstructive options. Patients undergoing revision TKA for isolated aseptic tibial loosening between 2012 and 2017 were identified. Those with revision implants or revised for infection, instability, osteolysis, or femoral component loosening were excluded. A total of 164 patients were included; 88 had an isolated tibial revision and 76 had revision of both components despite only having a loose tibial component. The demographics and clinical and radiological outcomes were recorded.Aims
Methods
To compare long-term survival of all-cemented and hybrid total hip arthroplasty (THA) using the Exeter Universal stem. Details of 1,086 THAs performed between 1999 and 2005 using the Exeter stem and either a cemented (632) or uncemented acetabular component (454) were collected from local records and the New Zealand Joint Registry. A competing risks regression survival analysis was performed with death as the competing risk with adjustments made for age, sex, approach, and bearing.Aims
Methods
Complex displaced osteoporotic acetabular fractures in the elderly are associated with high levels of morbidity and mortality. Surgical options include either open reduction and internal fixation alone, or combined with total hip arthroplasty (THA). There remains a cohort of severely comorbid patients who are deemed unfit for extensive surgical reconstruction and are treated conservatively. We describe the results of a coned hemipelvis reconstruction and THA inserted via a posterior approach to the hip as the primary treatment for this severely high-risk cohort. We have prospectively monitored a series of 22 cases (21 patients) with a mean follow-up of 32 months (13 to 59).Aims
Methods
The aim of this study was to compare the actual cost of a cemented and cementless total knee arthroplasty (TKA) procedure. The cost of operative time, implants, cement, and cementing accessories were included in the overall cost of the TKA procedure. Operative time was determined from a previously published study comparing cemented and cementless implants of the same design. The cost of operative time, implants, cement, and cementing accessories was determined from market and institutional data.Aims
Materials and Methods
Vancouver type B periprosthetic femoral fractures (PFF) are challenging complications after total hip arthroplasty (THA), and some treatment controversies remain. The objectives of this study were: to evaluate the short-to-mid-term clinical outcomes after treatment of Vancouver type B PFF and to compare postoperative outcome in subgroups according to classifications and treatments; to report the clinical outcomes after conservative treatment; and to identify risk factors for postoperative complications in Vancouver type B PFF. A total of 97 consecutive PPFs (49 males and 48 females) were included with a mean age of 66 years (standard deviation (SD) 14.9). Of these, 86 patients were treated with surgery and 11 were treated conservatively. All living patients had a minimum two-year follow-up. Patient demographics details, fracture healing, functional scores, and complications were assessed. Clinical outcomes between internal fixation and revisions in patients with or without a stable femoral component were compared. Conservatively treated PPFs were evaluated in terms of mortality and healing status. A logistic regression analysis was performed to identify risk factors for complications.Aims
Methods
In previous studies, we identified multiple factors influencing the survivorship of hip resurfacing arthroplasties (HRAs), such as initial anatomical conditions and surgical technique. In addition, the University of California, Los Angeles (UCLA) activity score presents a ceiling effect, so a better quantification of activity is important to determine which activities may be advisable or detrimental to the recovered patient. We aimed to determine the effect of specific groups of sporting activities on the survivorship free of aseptic failure of a large series of HRA. A total of 661 patients (806 hips) representing 77% of a consecutive series of patients treated with metal-on-metal hybrid HRA answered a survey to determine the types and amounts of sporting activities they regularly participated in. There were 462 male patients (70%) and 199 female patients (30%). Their mean age at the time of surgery was 51.9 years (14 to 78). Their mean body mass index (BMI) was 26.5 kg/m2 (16.7 to 46.5). Activities were regrouped into 17 categories based on general analogies between these activities. Scores for typical frequency and duration of the sessions were used to quantify the patients’ overall time spent engaging in sporting activities. Impact and cycle scores were computed. Multivariable models were used.Aims
Patients and Methods
Radiostereometric analysis (RSA) studies of vitamin E-doped, highly crosslinked polyethylene (VEPE) liners show low head penetration rates in cementless acetabular components. There is, however, currently no data on cemented VEPE acetabular components in total hip arthroplasty (THA). The aim of this study was to evaluate the safety of a new cemented VEPE component, compared with a conventional polyethylene (PE) component regarding migration, head penetration, and clinical results. We enrolled 42 patients (21 male, 21 female) with osteoarthritis and a mean age of 67 years (Aims
Patients and Methods
To our knowledge, no study has compared the long-term results of cemented and hybrid total hip arthroplasty (THA) in patients with osteoarthritis (OA) secondary to congenital hip disease (CHD). This is a demanding procedure that may require special techniques and implants. Our aim was to compare the long-term outcome of cemented low-friction arthroplasty (LFA) and hybrid THA performed by one surgeon. Between January 1989 and December 1997, 58 hips (44 patients; one man, 43 woman; mean age 56.6 years (25 to 77)) with OA secondary to CHD were treated with a cemented Charnley LFA (group A), and 55 hips (39 patients; two men, 37 women; mean age 49.1 years (27 to 70)) were treated with a hybrid THA (group B), by the senior author (GH). The clinical outcome and survivorship were compared.Aims
Patients and Methods
This study presents the long-term survivorship, risk factors for prosthesis survival, and an assessment of the long-term effects of changes in surgical technique in a large series of patients treated by metal-on-metal (MoM) hip resurfacing arthroplasty (HRA). Between November 1996 and January 2012, 1074 patients (1321 hips) underwent HRA using the Conserve Plus Hip Resurfacing System. There were 787 men (73%) and 287 women (27%) with a mean age of 51 years (14 to 83). The underlying pathology was osteoarthritis (OA) in 1003 (75.9%), developmental dysplasia of the hip (DDH) in 136 (10.3%), avascular necrosis in 98 (7.4%), and other conditions, including inflammatory arthritis, in 84 (6.4%).Aims
Patients and Methods
Aims. Loosening of the tibial component after total knee arthroplasty (TKA) is a common indication for revision. Increasing the strength of the initial tibial implant/cement interface is desirable. There is little information about the surgical techniques that lead to the highest strength. We investigated the effects of eight variables on the strength of the initial tibial baseplate/cement interface. Materials and Methods. A total of 48 tibial trays were cemented into acrylic holders using cement from two manufacturers, at three different times (early, normal, and late) using two
The reasons for failure of a hemirthroplasty (HA) when used to
treat a proximal humeral fracture include displaced or necrotic
tuberosities, insufficient metaphyseal bone-stock, and rotator cuff
tears. Reverse total shoulder arthroplasty (rTSA) is often the only
remaining form of treatment in these patients. The aim of this study
was to evaluate the clinical outcome after conversions from a failed
HA to rTSA. A total of 35 patients, in whom a HA, as treatment for a fracture
of the proximal humerus, had failed, underwent conversion to a rTSA.
A total of 28 were available for follow-up at a mean of 61 months
(37 to 91), having been initially reviewed at a mean of 20 months
(12 to 36) postoperatively. Having a convertible design, the humeral
stem could be preserved in nine patients. The stem was removed in
the other 19 patients and a conventional rTSA was implanted. At
final follow-up, patients were assessed using the American Shoulder
and Elbow Surgeons (ASES) score, the Constant Score, and plain radiographs.Aims
Material and Methods
We carried out a further study of the long-term results of the
cemented Exeter femoral component in patients under the age of 40
with a mean follow-up of 13.6 years (10 to 20). We reviewed our original cohort of 104 cemented Exeter stems
in 78 consecutive patients with a mean age of 31 years (16 to 39).
Only one patient was lost to radiological follow-up.Aims
Patients and Methods
We report the incidence of radiolucent lines (RLLs) using two
flanged acetabular components at total hip arthroplasty (THA) and
the effect of the Rim Cutter. We performed a retrospective review of 300 hips in 292 patients
who underwent primary cemented THA. A contemporary flanged acetabular
component was used with (group 1) and without (group 2) the use
of the Rim Cutter and the Rimfit acetabular component was used with
the Rim Cutter (group 3). RLLs and clinical outcomes were evaluated
immediately post-operatively and at five years post-operatively.Aims
Patients and Methods
The aim of this study was to compare the incidence of aseptic
loosening after the use of a cemented acetabular component and a
Trabecular Metal (TM) acetabular component (Zimmer Inc., Warsaw,
Indiana) at acetabular revision with bone impaction grafting. A total of 42 patients were included in the study. Patients were
randomised to receive an all- polyethylene cemented acetabular component
(n = 19) or a TM component (n = 23). Radiostereometric analysis
and conventional radiographic examinations were performed regularly
up to two years post-operatively or until further revision.Aims
Patients and Methods
We present the clinical and radiological results at a minimum
follow-up of five years for patients who have undergone multiple
cement-in-cement revisions of their femoral component at revision
total hip arthroplasty (THA). We reviewed the outcome on a consecutive series of 24 patients
(10 men, 14 women) (51 procedures) who underwent more than one cement-in-cement
revision of the same femoral component. The mean age of the patients was
67.5 years (36 to 92) at final follow-up. Function was assessed using the original Harris hip score (HHS),
Oxford Hip Score (OHS) and the Merle D’Aubigné Postel score (MDP).Aims
Patients and Methods
Hip resurfacing arthroplasty (HRA) is an alternative to conventional
total hip arthroplasty for patients with osteonecrosis (ON) of the
femoral head. Our aim was to report the long-term outcome of HRA,
which is not currently known. Long-term survivorship, clinical scores and radiographic results
for 82 patients (99 hips) treated with HRA for ON over a period
of 18 years were reviewed retrospectively. The mean age of the 67
men and 15 women at the time of surgery was 40.8 years (14 to 64).
Patients were resurfaced regardless of the size of the osteonecrotic
lesion.Aims
Patients and Methods
The widely used and well-proven Palacos R (a.k.a. Refobacin Palacos
R) bone cement is no longer commercially available and was superseded
by Refobacin bone cement R and Palacos R + G in 2005. However, the
performance of these newly introduced bone cements have not been
tested in a phased evidence-based manner, including roentgen stereophotogrammetric
analysis (RSA). In this blinded, randomised, clinical RSA study, the migration
of the Stanmore femoral component was compared between Refobacin
bone cement R and Palacos R + G in 62 consecutive total hip arthroplasties.
The primary outcome measure was femoral component migration measured
using RSA and secondary outcomes were Harris hip score (HHS), Hip
disability and Osteoarthritis Outcome Score (HOOS), EuroQol 5D (EQ-5D)
and Short Form 36 (SF-36).Aims
Patients and Methods
The primary purpose of this study of metal-on-metal
(MoM) hip resurfacing was to compare the effect of using a cementless
or cemented femoral component on the subsequent bone mineral density
(BMD) of the femoral neck. This was a single-centre, prospective, double-blinded control
trial which randomised 120 patients (105 men and 15 women) with
a mean age of 49.4 years (21 to 68) to receive either a cemented
or cementless femoral component. Follow-up was to two years. Outcome
measures included total and six-point region-of-interest BMD of
the femoral neck, radiological measurements of acetabular inclination,
neck-shaft and stem-shaft angles, and functional outcome scores
including the Harris hip score, the Western Ontario and McMaster
Universities Osteoarthritis Index and the University of California
at Los Angeles activity scale. In total, 17 patients were lost to follow-up leaving 103 patients
at two years. There were no revisions in the cementless group and
three revisions (5%) in the cemented group (two because of hip pain
and one for pseudotumour). The total BMD was significantly higher in the cementless group
at six months (p <
0.001) and one year (p = 0.01) than in the
cemented group, although there was a loss of statistical significance
in the difference at two years (p = 0.155). All patient outcomes improved significantly: there were no significant
differences between the two groups. The results show better preservation of femoral neck BMD with
a cementless femoral component after two years of follow-up. Further
investigation is needed to establish whether this translates into
improved survivorship. Cite this article:
Revision total elbow arthroplasty (TEA) is often challenging.
The aim of this study was to report on the clinical and radiological
results of revision arthroplasty of the elbow with the Latitude
TEA. Between 2006 and 2010 we used the Latitude TEA for revision in
18 consecutive elbows (17 patients); mean age 53 years (28 to 80);
14 women. A Kudo TEA was revised in 15 elbows and a Souter-Strathclyde
TEA in three. Stability, range of movement (ROM), visual analogue score (VAS)
for pain and functional scores, Elbow Functional Assessment Scale
(EFAS), the Functional Rating Index of Broberg and Morrey (FRIBM)
and the Modified Andrews’ Elbow Scoring System (MAESS) were assessed
pre-operatively and at each post-operative follow-up visit (six,
12 months and biennially thereafter). Radiographs were analysed
for loosening, fractures and dislocation. The mean follow-up was
59 months (26 to 89).Aims
Patients and Methods
This review summarises the evidence for the treatment
of displaced fractures of the femoral neck in elderly patients.
Results from randomised clinical trials and national register studies are
presented when available. The advantages of arthroplasty compared with internal fixation
are supported by several studies. A number of studies contribute
to the discussions of total hip arthroplasty (THA) The direct lateral approach reduces the risk of dislocation compared
with the posterior approach. Cemented implants lower the risk of
periprosthetic fracture and its subsequent morbidity and mortality.
As the risk of peri-operative death related to bone cement can be reduced
by adequate measures, cemented implants are recommended in fracture
cases. Take home message: There remains a great variation in the surgical
management of patients with a hip fracture, and an evidence-based
approach should improve the outcomes for this vulnerable patient
group. Cite this article:
The highly cross-linked polyethylene Exeter RimFit flanged cemented
acetabular component was introduced in the United Kingdom in 2010.
This study aimed to examine the rates of emergence of radiolucent
lines observed when the Rimfit acetabular component was implanted
at total hip arthroplasty (THA) using two different techniques: firstly,
the ‘rimcutter’ technique in which the flange sits on a pre-prepared
acetabular rim; and secondly, the ‘trimmed flange’ technique in
which the flange is trimmed and the acetabular component is seated
inside the rim of the acetabulum. The radiographs of 150 THAs (75 ‘rimcutter’, 75 ‘trimmed flange’)
involving this component were evaluated to assess for radiolucencies
at the cement/bone interface by three observers. Aims
Patients and Methods
We describe our technique and rationale using
hybrid fixation for primary total hip arthroplasty (THA) at the Hospital
for Special Surgery. Modern uncemented acetabular components have
few screw holes, or no holes, polished inner surfaces, improved
locking mechanisms, and maximised thickness and shell-liner conformity. Uncemented
sockets can be combined with highly cross-linked polyethylene liners,
which have demonstrated very low wear and osteolysis rates after
ten to 15 years of implantation. The results of cement fixation
with a smooth or polished surface finished stem have been excellent,
virtually eliminating complications seen with cementless fixation
like peri-operative femoral fractures and thigh pain. Although mid-term
results of modern cementless stems are encouraging, the long-term
data do not show reduced revision rates for cementless stems compared
with cemented smooth stems. In this paper we review the conduct
of a hybrid THA, with emphasis on pre-operative planning, surgical
technique, hypotensive epidural anaesthesia, and intra-operative
physiology. Cite this article:
This prospective study reports the 15-year survival and ten-year
functional outcome of a consecutive series of 1000 minimally invasive
Phase 3 Oxford medial UKAs (818 patients, 393 men, 48%, 425 women,
52%, mean age 66 years; 32 to 88). These were implanted by two surgeons
involved with the design of the prosthesis to treat anteromedial
osteoarthritis and spontaneous osteonecrosis of the knee, which
are recommended indications. Patients were prospectively identified
and followed up independently for a mean of 10.3 years (5.3 to 16.6). At ten years, the mean Oxford Knee Score was 40 (standard deviation
( This is the only large series of minimally invasive UKAs with
15-year survival data. The results support the continued use of
minimally invasive UKA for the recommended indications. Cite this article:
Aseptic loosening of the femoral component is
an important indication for revision surgery in unicompartmental knee
replacement (UKR). A new design of femoral component with an additional
peg was introduced for the cemented Oxford UKR to increase its stability.
The purpose of this study was to compare the primary stability of
the two designs of component. Medial Oxford UKR was performed in 12 pairs of human cadaver
knees. In each pair, one knee received the single peg and one received
the twin peg design. Three dimensional micromotion and subsidence
of the component in relation to the bone was measured under cyclical
loading at flexion of 40° and 70° using an optical measuring system.
Wilcoxon matched pairs signed-rank test was performed to detect
differences between the two groups. There was no significant difference in the relative micromotion
(p = 0.791 and 0.380, respectively) and subsidence (p = 0.301 and
0.176, respectively) of the component between the two groups at
both angles of flexion. Both designs of component offered good strength
of fixation in this cadaver study. Cite this article:
The purpose of the study was to investigate whether closed intramedullary
(IM) nailing with percutaneous cement augmentation is better than
conventional closed nailing at relieving pain and suppressing tumours
in patients with metastases of the femur and humerus. A total of 43 patients (27 men, 16 women, mean age 63.7 years,
standard deviation (Aims
Patients and Methods
A common situation presenting to the orthopaedic
surgeon today is a worn acetabular liner with substantial acetabular
and pelvic osteolysis. The surgeon has many options for dealing
with osteolytic defects. These include allograft, calcium based
substitutes, demineralised bone matrix, or combinations of these
options with or without addition of platelet rich plasma. To date
there are no clinical studies to determine the efficacy of using
bone-stimulating materials in osteolytic defects at the time of
revision surgery and there are surprisingly few studies demonstrating
the clinical efficacy of these treatment options. Even when radiographs
appear to demonstrate incorporation of graft material CT studies
have shown that incorporation is incomplete. The surgeon, in choosing
a graft material for a surgical procedure must take into account
the efficacy, safety, cost and convenience of that material. Cite this article:
Metaphyseal bone loss is common with revision
total knee replacement (RTKR). Using the Anderson Orthopaedic Research
Institute (AORI) classification, type 2-B and type 3 defects usually
require large metal blocks, bulk structural allograft or highly
porous metal cones. Tibial and femoral trabecular metal metaphyseal
cones are a unique solution for large bone defects. These cones
substitute for bone loss, improve metaphyseal fixation, help correct
malalignment, restore the joint line and may permit use of a shorter
stem. The technique for insertion involves sculpturing of the remaining
bone with a high speed burr and rasp, followed by press-fit of the
cone into the metaphysis. The fixation and osteoconductive properties
of the porous cone outer surface allow ingrowth and encourage long-term
biological fixation. The revision knee component is then cemented
into the porous cone inner surface, which provides superior fixation
compared with cementing into native but deficient metaphyseal bone.
The advantages of the cone compared with allograft include: technical
ease, biological fixation, no resorption, and possibly a lower risk
of infection. The disadvantages include: difficult extraction and
relatively short-term follow-up. Several studies using cones report
promising short-term results for the reconstruction of large bone
defects in RTKR. Cite this article:
Previous studies of failure mechanisms leading
to revision total knee replacement (TKR) performed between 1986 and
2000 determined that many failed early, with a disproportionate
amount accounted for by infection and implant-associated factors
including wear, loosening and instability. Since then, efforts have
been made to improve implant performance and instruct surgeons in
best practice. Recently our centre participated in a multi-centre evaluation
of 844 revision TKRs from 2010 to 2011. The purpose was to report
a detailed analysis of failure mechanisms over time and to see if
failure modes have changed over the past 10 to 15 years. Aseptic
loosening was the predominant mechanism of failure (31.2%), followed
by instability (18.7%), infection (16.2%), polyethylene wear (10.0%),
arthrofibrosis (6.9%) and malalignment (6.6%). The mean time to
failure was 5.9 years (ten days to 31 years), 35.3% of all revisions
occurred at less than two years, and 60.2% in the first five years.
With improvements in implant and polyethylene manufacture, polyethylene
wear is no longer a leading cause of failure. Early mechanisms of
failure are primarily technical errors. In addition to improving
implant longevity, industry and surgeons must work together to decrease
these technical errors. All reports on failure of TKR contain patients
with unexplained pain who not infrequently have unmet expectations.
Surgeons must work to achieve realistic patient expectations pre-operatively,
and therefore, improve patient satisfaction post-operatively. Cite this article:
We investigated whether the indentation of bone
cement spacers used in revision of infected joint arthroplasty with a
MacDonald dissector increased the elution of antibiotic The fluid sampled at 72 hours from the indented discs containing
0.17 g gentamicin (0.88% w/w) contained a mean of 113 mcg/ml (90.12
to 143.5) compared with 44.5 mcg/ml (44.02 to 44.90) in the fluid
sampled from the plain discs (p = 0.012). In discs containing 0.33
g gentamicin (1.75% w/w), the concentration eluted from the indented discs
at 72 hours was a mean of 316 mcg/ml (223 to 421) compared with
a mean of 118 mcg/ml (100 to 140) from the plain discs (p <
0.001). At two weeks, these significant differences persisted. At nine
weeks the indented discs eluted a greater concentration for all
gentamicin doses, but the difference was only significant for the
discs containing 0.17 g (0.88% w/w, p = 0.006). However if the area
under the curve is taken as a measure of the total antibiotic eluted,
the indented discs eluted more gentamicin than the plain discs for
the 0.17 g (0.88% w/w, p = 0.031), the 0.25 g (1.41% w/w, p <
0.001) and the 0.33 g (1.75% w/w, p <
0.001) discs. When preparing antibiotic spacers for use in staged revision
arthroplasty surgery we recommend indenting the spacer with a MacDonald
dissector to increase the elution of antibiotic. Cite this article:
Previous classification systems of failure of
limb salvage focused primarily on endoprosthetic failures and lacked sufficient
depth for the effective study of the causes of failure. In order
to address these inadequacies, the International Society of Limb
Salvage (ISOLS) formed a committee to recommend revisions of the
previous systems. The purpose of this study was to report on their
recommendations. The modifications were prepared using an earlier,
evidence-based model with subclassification based on the existing
medical literature. Subclassification for all five primary types
of failure of limb salvage following endoprosthetic reconstruction
were formulated and a complementary system was derived for the failure
of biological reconstruction. An additional classification of failure
in paediatric patients was also described. Limb salvage surgery presents a complex array of potential mechanisms
of failure, and a complete and precise classification of types of
failure is required. Earlier classification systems lacked specificity,
and the evidence-based system outlined here is designed to correct
these weaknesses and to provide a means of reporting failures of
limb salvage in order to allow the interpretation of outcome following
reconstructive surgery. Cite this article:
Aseptic loosening of the acetabular component continues to be the most common indication for revision of total hip replacements in younger patients. Early in the evolution of the cemented hip, arthroplasty surgeons switched from removal to retention of the acetabular subchondral bone plate, theorising that unfavourable mechanical forces were the cause of loosening at the bone-cement interface. It is now known that the cause of aseptic loosening is probably biological rather than mechanical and removing the subchondral bone plate may enhance biological fixation of cement to bone. With this in mind, perhaps it is time to revive removal of the subchondral bone as a standard part of acetabular preparation.
Revision total hip replacement (THR) for young
patients is challenging because of technical complexity and the potential
need for subsequent further revisions. We have assessed the survivorship,
functional outcome and complications of this procedure in patients
aged <
50 years through a large longitudinal series with consistent treatment
algorithms. Of 132 consecutive patients (181 hips) who underwent
revision THR, 102 patients (151 hips) with a mean age of 43 years
(22 to 50) were reviewed at a mean follow-up of 11 years (2 to 26)
post-operatively. We attempted to restore bone stock with allograft
where indicated. Using further revision for any reason as an end point,
the survival of the acetabular component was 71% ( This overall perspective on the mid- to long-term results is
valuable when advising young patients on the prospects of revision
surgery at the time of primary replacement. Cite this article:
The original forged Müller straight stem (CoNiCr) has shown excellent ten- to 15-year results. We undertook a long-term survival analysis with special emphasis on radiological changes within a 20-year period of follow-up. In all, 165 primary total hip replacements, undertaken between July 1984 and June 1987 were followed prospectively. Clinical follow-up included a standardised clinical examination, and radiological assessment was based on a standardised anteroposterior radiograph of the pelvis, which was studied for the presence of osteolysis, debonding and cortical atrophy. Survival of the stem with revision for any reason was 81% (95% confidence interval (CI), 76 to 86) at 20 years and for aseptic loosening 87% (95% CI, 82 to 90). At the 20-year follow-up, 15 of the surviving 36 stems showed no radiological changes. Debonding (p = 0.005), osteolysis (p = 0.003) and linear polyethylene wear (p = 0.016) were associated with aseptic loosening, whereas cortical atrophy was not associated with failure (p = 0.008). The 20-year results of the Müller straight stem are comparable to those of other successful cemented systems with similar follow-up. Radiological changes are frequently observed, but with a low incidence of progression, and rarely result in revision. Cortical atrophy appears to be an effect of ageing and not a sign of loosening of the femoral component.
Impaction bone grafting for the reconstitution
of bone stock in revision hip surgery has been used for nearly 30 years.
Between 1995 and 2001 we used this technique in acetabular reconstruction,
in combination with a cemented component, in 304 hips in 292 patients
revised for aseptic loosening. The only additional supports used
were stainless steel meshes placed against the medial wall or laterally
around the acetabular rim to contain the graft. All Paprosky grades
of defect were included. Clinical and radiographic outcomes were
collected in surviving patients at a minimum of ten years after
the index operation. Mean follow-up was 12.4 years ( Cite this article:
The strain on clinic and surgeon resources resulting
from a rise in demand for total knee replacement (TKR) requires reconsideration
of when and how often patients need to be seen for follow-up. Surgeons
will otherwise require increased paramedical staff or need to limit
the number of TKRs they undertake. We reviewed the outcome data
of 16 414 primary TKRs undertaken at our centre to determine the
time to re-operation for any reason and for specific failure mechanisms.
Peak risk years for failure were determined by comparing the conditional
probability of failure, the number of failures divided by the total
number of TKRs cases, for each year. The median times to failure
for the most common failure mechanisms were 4.9 years (interquartile
range (IQR) 1.7 to 10.7) for femoral and tibial loosening, 1.9 years
(IQR 0.8 to 3.9) for infection, 3.1 years (IQR 1.6 to 5.5) for tibial
collapse and 5.6 years (IQR 3.4 to 9.3) for instability. The median
time to failure for all revisions was 3.3 years (IQR 1.2 to 8.5),
with an overall revision rate of 1.7% (n = 282). Results from our
patient population suggest that patients be seen for follow-up at
six months, one year, three years, eight years, 12 years, and every
five years thereafter. Patients with higher pain in the early post-operative
period or high body mass index (≥ 41 kg/m2) should be
monitored more closely. Cite this article:
We have evaluated the in vivo migration patterns of 164 primary consecutive Charnley-Kerboull total hip replacements which were undertaken in 155 patients. The femoral preparation included removal of diaphyseal cancellous bone to obtain primary rotational stability of the stem before line-to-line cementing. We used the Ein Bild Roentgen Analyse femoral component method to assess the subsidence of the femoral component. At a mean of 17.3 years (15.1 to 18.3) 73 patients were still alive and had not been revised, eight had been revised, 66 had died and eight had been lost to follow-up. The mean subsidence of the entire series was 0.63 mm (0.0 to 1.94). When using a 1.5 mm threshold, only four stems were considered to have subsided. Our study showed that, in most cases, a highly polished double-tapered stem cemented line-to-line does not subside at least up to 18 years after implantation.
We reviewed 142 consecutive primary total hip replacements implanted into 123 patients between 1988 and 1993 using the Exeter Universal femoral stem. A total of 74 patients (88 hips) had survived for ten years or more and were reviewed at a mean of 12.7 years (10 to 17). There was no loss to follow-up. The rate of revision of the femoral component for aseptic loosening and osteolysis was 1.1% (1 stem), that for revision for any cause was 2.2% (2 stems), and for re-operation for any cause was 21.6% (19 hips). Re-operation was because of failure of the acetabular component in all but two hips. All but one femoral component subsided within the cement mantle to a mean of 1.52 mm (0 to 8.3) at the final follow-up. One further stem had subsided excessively (8 mm) and had lucent lines at the cement-stem and cement-bone interfaces. This was classified as a radiological failure and is awaiting revision. One stem was revised for deep infection and one for excessive peri-articular osteolysis. Defects of the cement mantle (Barrack grade C and D) were found in 28% of stems (25 hips), associated with increased subsidence (p = 0.01), but were not associated with endosteal lysis or failure. Peri-articular osteolysis was significantly related to the degree of polyethylene wear (p <
0.001), which was in turn associated with a younger age (p = 0.01) and male gender (p <
0.001). The use of the Exeter metal-backed acetabular component was a notable failure with 12 of 32 hips (37.5%) revised for loosening. The Harris-Galante components failed with excessive wear, osteolysis and dislocation with 15% revised (5 of 33 hips). Only one of 23 hips with a cemented Elite component (4%) was revised for loosening and osteolysis. Our findings show that the Exeter Universal stem implanted outside the originating centre has excellent medium-term results.
We report the effectiveness of revision of total
elbow replacement by re-cementing. Between 1982 and 2004, 53 elbows
in 52 patients were treated with re-cementing of a total elbow replacement
into part or all of the existing cement mantle or into the debrided
host-bone interface, without the use of structural bone augmentation
or a custom prosthesis. The original implant revision was still
We evaluated the outcome of 41 consecutive Charnley
low-friction arthroplasties (LFAs) performed by a single surgeon
in 28 patients aged ≤ 35 years at operation between 23 and 36 years
previously. There were 20 women and eight men with a mean age of
32 years (23 to 35) at surgery. Two patients (three hips) were lost
to follow-up at 12 and 17 years post-operatively, respectively,
and one patient (one hip) died at 13 years post-operatively. These patients
were excluded from the final evaluation. The survival rate of the
acetabular components was 92.7% (95% confidence interval (CI) 88.7
to 96.7) at ten years, 67.1% (95% CI 59.75 to 74.45) at 20 years
and 53.2% (95% CI 45.3 to 61.1) at 25 years. For the femoral component
the survival was 95.1% (95% CI 91.8 to 98.5) at ten years, 77.1%
(95% CI 73.9 to 80.3) at 20 years and 68.2% (95% CI 60.7 to 75.8)
at 25 years. The results indicate that the Charnley LFA remains
a reasonable choice in the treatment of young patients and can serve
for comparison with newer techniques and implants. Cite this article:
We describe the survivorship of the Exeter femoral component in a District General Hospital. Between 1994 and 1996, 230 Exeter Universal cemented femoral components were implanted in 215 patients who were reviewed at a mean of 11.2 years (10 to 13). We used one acetabular implant, the Elite Ogee component, in 218 of the 230 hips. During the period of this study 76 patients (79 hips) died. Of the remaining 139 patients (151 hips), 121 were able to attend for radiological analysis at a minimum of ten years. One patient was lost to follow-up. No femoral component was revised for aseptic loosening. Three hips were revised for deep infection and six acetabular components required revision, four for loosening and two for recurrent dislocation. Taking the ‘worst-case scenario’ including the one patient lost to follow-up, the overall survival rate was 94.4% at 13 years. Our results confirm excellent medium-term results for the Exeter Universal femoral component, implanted in a general setting. The excellent survival of this femoral component, when used in combination with the Ogee acetabular component, suggests that this is a successful pairing.
We evaluated the impact of pre-coating the tibial
component with polymethylmethacrylate (PMMA) on implant survival
in a cohort of 16 548 primary NexGen total knee replacements (TKRs)
in 14 113 patients. In 13 835 TKRs a pre-coated tray was used while
in 2713 TKRs the non-pre-coated version of the same tray was used.
All the TKRs were performed between 2001 and 2009 and were cemented.
TKRs implanted with a pre-coated tibial component had a lower cumulative
survival than those with a non-pre-coated tibial component (p =
0.01). After adjusting for diagnosis, age, gender, body mass index,
American Society of Anesthesiologists grade, femoral coupling design, surgeon
volume and hospital volume, pre-coating was an independent risk
factor for all-cause aseptic revision (hazard ratio 2.75, p = 0.006).
Revision for aseptic loosening was uncommon for both pre-coated
and non-pre-coated trays (rates of 0.12% and 0%, respectively).
Pre-coating with PMMA does not appear to be protective of revision
for this tibial tray design at short-term follow-up. Cite this article:
The ideal acetabular component is characterised by reliable, long-term fixation with physiological loading of bone and a low rate of wear. Trabecular metal is a porous construct of tantalum which promotes bony ingrowth, has a modulus of elasticity similar to that of cancellous bone, and should be an excellent material for fixation. Between 2004 and 2006, 55 patients were randomised to receive either a cemented polyethylene or a monobloc trabecular metal acetabular component with a polyethylene articular surface. We measured the peri-prosthetic bone density around the acetabular components for up to two years using dual-energy x-ray absorptiometry. We found evidence that the cemented acetabular component loaded the acetabular bone centromedially whereas the trabecular metal monobloc loaded the lateral rim and behaved like a hemispherical rigid metal component with regard to loading of the acetabular bone. We suspect that this was due to the peripheral titanium rim used for the mechanism of insertion.
We evaluated all cases involving the combined use of a subtrochanteric derotational femoral shortening osteotomy with a cemented Exeter stem performed at our institution. With severe developmental dysplasia of the hip an osteotomy is often necessary to achieve shortening and derotation of the proximal femur. Reduction can be maintained with a 3.5 mm compression plate while the implant is cemented into place. Such a plate was used to stabilise the osteotomy in all cases. Intramedullary autograft helps to prevent cement interposition at the osteotomy site and promotes healing. There were 15 female patients (18 hips) with a mean age of 51 years (33 to 75) who had a Crowe IV dysplasia of the hip and were followed up for a mean of 114 months (52 to 168). None was lost to follow-up. All clinical scores were collected prospectively. The Charnley modification of the Merle D’Aubigné-Postel scores for pain, function and range of movement showed a statistically significant improvement from a mean of 2.4 (1 to 4), 2.3 (1 to 4), 3.4 (1 to 6) to 5.2 (3 to 6), 4.4 (3 to 6), 5.2 (4 to 6), respectively. Three acetabular revisions were required for aseptic loosening; one required femoral revision for access. One osteotomy failed to unite at 14 months and was revised successfully. No other case required a femoral revision. No postoperative sciatic nerve palsy was observed. Cemented Exeter femoral components perform well in the treatment of Crowe IV dysplasia with this procedure.