The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 261 - 268
1 Mar 2023
Ruhr M Huber G Niki Y Lohner L Ondruschka B Morlock MM

Aims

The aim of the study was to investigate whether the primary stability of press-fit acetabular components can be improved by altering the impaction procedure.

Methods

Three impaction procedures were used to implant acetabular components into human cadaveric acetabula using a powered impaction device. An impaction frequency of 1 Hz until complete component seating served as reference. Overimpaction was simulated by adding ten strokes after complete component seating. High-frequency implantation was performed at 6 Hz. The lever-out moment of the acetabular components was used as measure for primary stability. Permanent bone deformation was assessed by comparison of double micro-CT (µCT) measurements before and after impaction. Acetabular component deformation and impaction forces were recorded, and the extent of bone-implant contact was determined from 3D laser scans.


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 307 - 314
1 Mar 2023
de Villeneuve Bargemon J Mathoulin C Jaloux C Levadoux M Gras M Merlini L

Aims

A conventional arthroscopic capsuloligamentous repair is a reliable surgical solution in most patients with scapholunate instability. However, this repair does not seem to be sufficient for more advanced injuries. The aim of this study was to evaluate the functional results of a wide arthroscopic dorsal capsuloligamentous repair (WADCLR) in the management of severe scapholunate instability.

Methods

This was a prospective single-centre study undertaken between March 2019 and May 2021. The primary outcome was the evaluation of the reduction of the radiological deformity and the functional outcomes after WADCLR. A secondary outcome was the evaluation of the effectiveness of this technique in patients with the most severe instability (European Wrist Arthroscopy Society (EWAS) stage 5). The patients were reviewed postoperatively at three, six, and 12 months.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1339 - 1347
1 Oct 2013
Scott CEH Eaton MJ Nutton RW Wade FA Pankaj P Evans SL

As many as 25% to 40% of unicompartmental knee replacement (UKR) revisions are performed for pain, a possible cause of which is proximal tibial strain. The aim of this study was to examine the effect of UKR implant design and material on cortical and cancellous proximal tibial strain in a synthetic bone model. Composite Sawbone tibiae were implanted with cemented UKR components of different designs, either all-polyethylene or metal-backed. The tibiae were subsequently loaded in 500 N increments to 2500 N, unloading between increments. Cortical surface strain was measured using a digital image correlation technique. Cancellous damage was measured using acoustic emission, an engineering technique that detects sonic waves (‘hits’) produced when damage occurs in material.

Anteromedial cortical surface strain showed significant differences between implants at 1500 N and 2500 N in the proximal 10 mm only (p < 0.001), with relative strain shielding in metal-backed implants. Acoustic emission showed significant differences in cancellous bone damage between implants at all loads (p = 0.001). All-polyethylene implants displayed 16.6 times the total number of cumulative acoustic emission hits as controls. All-polyethylene implants also displayed more hits than controls at all loads (p < 0.001), more than metal-backed implants at loads ≥ 1500 N (p < 0.001), and greater acoustic emission activity on unloading than controls (p = 0.01), reflecting a lack of implant stiffness. All-polyethylene implants were associated with a significant increase in damage at the microscopic level compared with metal-backed implants, even at low loads. All-polyethylene implants should be used with caution in patients who are likely to impose large loads across their knee joint.

Cite this article: Bone Joint J 2013;95-B:1339–47.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 2 | Pages 236 - 240
1 Feb 2005
Belthur MV Bradish CF Gibbons PJ

Between 1990 and 2001, 24 children aged between 15 months and 11 years presented with late orthopaedic sequelae after meningococcal septicaemia. The median time to presentation was 32 months (12 to 119) after the acute phase of the disease. The reasons for referral included angular deformity, limb-length discrepancy, joint contracture and problems with prosthetic fitting. Angular deformity with or without limb-length discrepancy was the most common presentation. Partial growth arrest was the cause of the angular deformity. Multiple growth-plate involvement occurred in 14 children. The lower limbs were affected much more often than the upper. Twenty-three children underwent operations for realignment of the mechanical axis and limb-length equalisation. In 15 patients with angular deformity around the knee the deformity recurred. As a result we recommend performing a realignment procedure with epiphysiodesis of the remaining growth plate when correcting angular deformities.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1176 - 1181
1 Aug 2010
Tayton E Evans S O’Doherty D

We implanted titanium and carbon fibre-reinforced plastic (CFRP) femoral prostheses of the same dimensions into five prosthetic femora. An abductor jig was attached and a 1 kN load applied. This was repeated with five control femora. Digital image correlation was used to give a detailed two-dimensional strain map of the medial cortex of the proximal femur. Both implants caused stress shielding around the calcar. Distally, the titanium implant showed stress shielding, whereas the CFRP prosthesis did not produce a strain pattern which was statistically different from the controls. There was a reduction in strain beyond the tip of both the implants.

This investigation indicates that use of the CFRP stem should avoid stress shielding in total hip replacement.