Aims. We compared the clinical and radiological outcomes of using a
polyetheretherketone cage with (TiPEEK) and without a titanium coating
(PEEK) for instrumented transforaminal lumbar interbody fusion (TLIF). Materials and Methods. We conducted a randomised clinical pilot trial of 40 patients
who were scheduled to undergo a TLIF procedure at one or two levels
between L2 and L5. The Oswestry disability index (ODI), EuroQoL-5D,
and back and leg pain were determined pre-operatively, and at three,
six, and 12 months post-operatively. Fusion rates were assessed
by thin slice CT at three months and by functional radiography at
12 months. Results. At final follow-up, one patient in each group had been lost to
follow-up. Two patients in each of the PEEK and TiPEEK groups were
revised for pseudarthrosis (p = 1.00). The rate of complete or partial
fusion at three months was 91.7% in both groups. Overall, there
were no significant differences in ODI or in radiological outcomes
between the groups. Conclusion. Favourable results with identical clinical outcomes and a high
rate of fusion was seen in both groups. The titanium
Periprosthetic joint infection (PJI) is one of the most dreaded complications after arthroplasty surgery; thus numerous approaches have been undertaken to equip metal surfaces with antibacterial properties. Due to its antimicrobial effects, silver is a promising
Aims. The aim of this study was to develop a single-layer hybrid organic-inorganic sol-gel
A silver-containing hydroxyapatite (Ag-HA)
The success of long-term transcutaneous implants
depends on dermal attachment to prevent downgrowth of the epithelium
and infection. Hydroxyapatite (HA)
Four uncemented Symax hip stems were extracted at three weeks and nine, 13 and 32 months, respectively, for reasons other than loosening. The reasons for implant removal were infection in two cases, recurrent dislocation in one and acetabular fracture in one. They were analysed to assess the effect and behaviour of an electrochemically deposited, completely resorbable biomimetic BONIT-hydroxyapatite (HA)
We inserted two hydroxyapatite (HA)-coated implants with crystallinities of either 50% (HA-50%) or 75% (HA-75%) bilaterally into the medial femoral condyles of the knees of 16 dogs. The implants were allocated to two groups with implantation periods of 16 and 32 weeks. They were weight-bearing and subjected to controlled micromovement of 250 μm during each gait cycle. After 16 weeks, mechanical fixation of the HA-50% implants was increased threefold as compared with the HA-75% implants. After 32 weeks there was no difference between HA-50% and HA-75%. Fixation of HA-75% increased from 16 to 32 weeks whereas that of HA-50% was unchanged. HA-50% implants had 100% more bone ingrowth than HA-75% implants after 16 weeks. More HA
Aims. The Intraosseous Transcutaneous Amputation Prosthesis (ITAP)
may improve quality of life for amputees by avoiding soft-tissue
complications associated with socket prostheses and by improving
sensory feedback and function. It relies on the formation of a seal
between the soft tissues and the implant and currently has a flange
with drilled holes to promote dermal attachment. Despite this, infection
remains a significant risk. This study explored alternative strategies
to enhance soft-tissue integration. Materials and Methods. The effect of ITAP pins with a fully porous titanium alloy flange
with interconnected pores on soft-tissue integration was investigated.
The flanges were coated with fibronectin-functionalised hydroxyapatite
and silver
The aim of this study was to evaluate whether
coating titanium discs with selenium in the form of sodium selenite decreased
bacterial adhesion of Staphylococcus aureus and Staph.
epidermidis and impeded osteoblastic cell growth. In order to evaluate bacterial adhesion, sterile titanium discs
were coated with increasing concentrations of selenium and incubated
with bacterial solutions of Staph. aureus (ATCC
29213) and Staph. epidermidis (DSM 3269) and stained
with Safranin-O. The effect of selenium on osteoblastic cell growth
was also observed. The adherence of MG-63 cells on the coated discs
was detected by staining with Safranin-O. The proportion of covered
area was calculated with imaging software. The tested Staph. aureus strain showed a significantly
reduced attachment on titanium discs with 0.5% (p = 0.011) and 0.2%
(p = 0.02) selenium
We report the radiological and clinical outcome of a press-fit (SLF) acetabular component at two to three years in two groups of patients having primary total hip replacement. In 69 the implant was coated with hydroxyapatite (HA) and in 40 it was uncoated. The stability of the cup was assessed by measurement of proximal migration and change in the angle of inclination. The clinical results in the two groups did not differ significantly, and the mean proximal linear wear was similar in both. Fewer radiolucent lines (RLLs) were seen on the radiographs of cups coated with HA. The mean proximal migration was studied by calculating regression lines for each patient using migration measurements: for the SLF+HA group the mean slope was 0.06 mm/year and for the SLF–HA group 0.20 mm/year (p = 0.22). The change in the angle of inclination during follow-up was also consistently smaller in HA–coated cups. Using regression methods the SLF+HA group had a mean slope of 0.08°/year and the SLF-HA group 0.44°/year (p = 0.023). Partial HA
In a single-blind, randomised series of knee replacements in 116 patients, we used radiostereometric analysis (RSA) to measure micromotion in three types of tibial implant fixation for two years after knee replacement. We compared hydroxyapatite-augmented porous
We carried out a histological study of a proximally hydroxyapatite (HA)-coated femoral component, retrieved after 9.5 years of good function. The HA
The outcome at ten years of 100 Freeman hip stems (Finsbury Orthopaedics, Leatherhead, United Kingdom) retaining the neck with a proximal hydroxyapatite
We investigated the effects of hydroxyapatite (HA)
In previous studies, we have demonstrated a fibrocartilaginous membrane around hydroxyapatite-coated implants subjected to micromovement in contrast to the fibrous connective tissue which predominates around similarly loaded titanium alloy implants. In the present study, in mature dogs, we investigated the effect of immobilising titanium (Ti)- or hydroxyapatite (HA)-coated implants already surrounded by a movement-induced fibrous membrane and compared the results with those of similar implants in which continuous micromovement was allowed to continue. The implants were inserted in the medial femoral condyles of 14 dogs and subjected to 150 microns movements during each gait cycle. After four weeks (when a fibrous membrane had developed), half the implants were immobilised to prevent further micromovement. The dogs were killed at 16 weeks and the results were evaluated by push-out tests and histological analysis. The continuously loaded Ti-coated implants were surrounded by a fibrous membrane, whereas bridges of new bone anchored the HA-coated implants. The immobilised implants were surrounded by bone irrespective of the type of
We have studied the beneficial effects of a hydroxyapatite (HA)
We report the histological findings in post-mortem specimens obtained ten days, 17 days and seven weeks after implantation of hydroxyapatite-coated femoral components of hip arthroplasties. There was early deposition of woven bone on the hydroxyapatite ceramic, identical to that deposited on surviving cancellous trabeculae. The space between these deposits became bridged from both sides by new trabeculae, and there was no evidence of an inflammatory reaction or of fibrous tissue formation. The use of an hydroxyapatite
We studied two groups of femoral hip prostheses: 43 TiAlV ridged press-fit stems, and 26 with similar stems coated with hydroxyapatite on the proximal half. At one year, radiological measurement showed a mean downward migration of 0.99 mm for the TialV prostheses and 0.12 mm for the HA-coated prostheses (p = 0.0002). Hydroxyapatite
We describe the survival at ten years of 100 femoral components of the Freeman hip prosthesis. It is proximally hydroxyapatite (HA)-coated and was fixed without cement. Radiological assessment identified radiolucent lines (RLLs) and lytic lesions and was used to measure migration. The criterion of failure was revision or impending revision for aseptic femoral loosening. No femoral components were revised or are awaiting revision for aseptic loosening, giving 100% survival at ten years (95% confidence interval, 95.7 to 100), although 59 were at risk at ten years. Two components were revised for fracture of a ceramic head with damage to the trunnion. Although well fixed in each, for survival analysis we evaluated the hip as if the patient had died. Twelve acetabular components were revised and at each operation the femoral component was found to be well fixed, was not disturbed and remained in the survival analysis. Three patients were lost to follow-up, and 12 died with well-functioning prostheses. Radiologically, all except one of the components appeared to be well fixed with no RLLs and no lytic lesions at the latest follow-up. The mean vertical migration was 0.4 mm at one year, 0.8 mm at two years and 1.4 mm at ten years. One component had migrated 7.6 mm at ten years (2.1 mm in year 1) and developed RLLs in Gruen zones I and II. The symptoms, however, were only minor and revision was not indicated. Our study has shown that proximal HA
One concern about the fixation of HA-coated implants is the possible disintegration of the surface, with the migration of HA granules into the joint space, producing third-body wear. We report a study of six revisions of HA-coated polyethylene RM cups at 9 to 14 years after successful primary arthroplasty. In all six hips, we found HA granules embedded in the articulating surface of the polyethylene, with abrasive wear of the cup and the metal femoral head. The cup had loosened in four hips and three showed severe osteolysis of the proximal femur. Third-body wear due to HA particles from implant
In 33 dogs we implanted femoral stems made either of carbon composite, some coated with hydroxyapatite, or of titanium alloy with a porous
We measured the insertion and extraction torque forces in a randomised study of 76 external fixation screws in 19 patients treated by hemicallotasis for osteoarthritis of the medial side of the knee. The patients were randomised to have either standard tapered screws (Orthofix 6/5 mm) or the same screws with hydroxyapatite (HA)
In rabbits and goats, test implants with a porous surface of two layers of Tl-6A;-4V beads were examined at intervals for bond strength with bone. Half of the implants were coated with hydroxyapatite by plasma spray. The bonding strength with bone in the coated specimens was about four times greater than that of the uncoated specimens at two weeks, and twice as strong at six weeks. Twelve weeks after implantation, the strengths were similar. The hydroxyapatite
We evaluated the effects of a serum protein
We have studied the effect of hydroxyapatite (HA)
This prospective study compares the outcome of
157 hydroxyapatite (HA)-coated tibial components with 164 cemented
components in the ROCC Rotating Platform total knee replacement
in 291 patients. The mean follow-up was 7.6 years (5.2 to 11). There
were two revisions for loosening: one for an HA-coated and one for
a cemented tibial component. Radiological evaluation demonstrated
no radiolucent lines with the HA-coated femoral components. A total
of three HA-coated tibial components exhibited radiolucent lines
at three months post-operatively and these disappeared after three
further months of protected weight-bearing. With HA-coated components
the operating time was shorter (p <
0.006) and the radiological
assessment of the tibial interface was more stable (p <
0.01).
Using revision for aseptic loosening of the tibial component as
the end point, the survival rates at nine years was identical for
both groups at 99.1%. Our results suggest that HA-coated components perform at least
as well as the same design with cemented components and compare
favourably with those of series describing cemented or porous-coated
knee replacements, suggesting that fixation of both components with
hydroxyapatite is a reliable option in primary total knee replacement.
We reviewed 40 cementless Porous Coated Anatomic knee replacements in 34 consecutive patients. The average follow-up was 12.9 months (range 6 to 36 months). At review, loose beads were identified in 23 knees, in 11 on the femoral side and in 16 on the tibial side (in five knees on both sides). One loose bead was adjacent to the patella. In four knees there were intra-articular beads. Radiolucent lines at the interface between bone and prosthesis were seen in 19 knees, in six on the femoral side and in 16 on the tibial side (in three knees on both sides). In relation to the tibial component loose beads were associated with a radiolucent line in all except one case. Progressive loosening with an increasing number of loose beads was seen in three knees. At this early stage of follow-up there was no correlation between the clinical results and the presence of loose beads.
Aims. BoneMaster is a thin electrochemically applied hydroxyapatite (HA)
Aims. This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant. Methods. A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life. Results. No differences in mean age (69 years, 40 to 86), mean age adapted Charlson Comorbidity Index (3.1 (SD 1.4)), mean BMI (29.2 kg/m. 2. (SD 5.8)), or number of other implants were observed between groups. Significant improvements in outcome scores and pain levels were achieved for all groups, and there was no difference in quality of life (12-Item Short-Form Health Survey questionnaire (SF-12)). Mean cobalt and chromium ion levels were significantly higher for the GenuX versus the BPKS hinged TKA (GenuX vs BPKS: cobalt: 16.3 vs 9.4 µg/l; chromium: 9.5 vs 5.2 µg/l). The TiN-coated implants did not appear to confer improvement in the metal ion levels. Metal ion concentrations above 7 µg/l were detected in 81%(29/36) of GenuX patients versus 41% (11/27) in the BPKS group. No GenuX patients had normal levels under 2 µg/l, versus 22% of BPKS patients. No significant reduction in outcome scores was observed regardless of the metal ion levels, whereas higher work-related activity was correlated with higher chromium concentrations. Conclusion. Hinged TKA, using MoM hinges, resulted in critically high cobalt and chromium ion concentrations. The BPKS hinged TKA showed significantly lower metal ion concentrations compared with the GenuX TKA. No benefits were observed using TiN
Aims. The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Materials and Methods. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed. Results. In 104 implants, a decrease in the odds of loosening and an increase in metallosis and tissue staining in the Duofix group relative to Porocoat group was detected (p = 0.028). There was an increased presence of embedded metallic debris in the Duofix group (p < 0.001). Decreased tissue ingrowth was associated with the Duofix surface (p < 0.001). The attached beads had reduced microhardness, indicative of adverse thermal processing, which resulted in bead shedding, particulate debris, and metallosis. Conclusion. Hydroxyapatite
Aims. The purpose of this study was to evaluate the biological fixation of a 3D printed porous implant, with and without different hydroxyapatite (HA)
The clinical use of hydroxyapatite (HA)
Resurfacing of the humeral head is commonly used
within the UK to treat osteoarthritis (OA) of the shoulder. We present
the results of a small prospective randomised study of this procedure
using the Global CAP prosthesis with two different
This study describes 146 primary total knee replacements, either fully or partially coated with hydroxyapatite of which 74 knees in 68 patients were available for clinical and radiological assessment at a mean of 11.2 years (10 to 15). The global failure rate was 1.37% and survival rate with mechanical failure as the end-point was 98.14%. Radiological assessment indicated intimate contact between bone and the hydroxyapatite
The aim of this study was to evaluate the survival of a collarless, straight, hydroxyapatite-coated femoral stem in total hip arthroplasty (THA) at a minimum follow-up of 20 years. We reviewed the results of 165 THAs using the Omnifit HA system in 138 patients, performed between August 1993 and December 1999. The mean age of the patients at the time of surgery was 46 years (20 to 77). Avascular necrosis was the most common indication for THA, followed by ankylosing spondylitis and primary osteoarthritis. The mean follow-up was 22 years (20 to 31). At 20 and 25 years, 113 THAs in 91 patients and 63 THAs in 55 patients were available for review, respectively, while others died or were lost to follow-up. Kaplan-Meier analysis was performed to evaluate the survival of the stem. Radiographs were reviewed regularly, and the stability of the stem was evaluated using the Engh classification.Aims
Methods
The aim of this study was to compare the pattern of initial fixation and changes in periprosthetic bone mineral density (BMD) between patients who underwent total hip arthroplasty (THA) using a traditional fully hydroxyapatite (HA)-coated stem (T-HA group) and those with a newly introduced fully HA-coated stem (N-HA group). The study included 36 patients with T-HA stems and 30 with N-HA stems. Dual-energy X-ray absorptiometry was used to measure the change in periprosthetic BMD, one and two years postoperatively. The 3D contact between the stem and femoral cortical bone was evaluated using a density-mapping system, and clinical assessment, including patient-reported outcome measurements, was recorded.Aims
Methods
The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design. This was a retrospective review of 707 primary TKAs in 590 patients from a single institution, using a cementless, metal-backed patellar component with a mean follow-up of 6.9 years (2 to 12). A total of 409 TKAs were performed in 338 females and 298 TKAs in 252 males. The mean age of the patients was 63 years (34 to 87) and their mean BMI was 34.3 kg/m2 (18.8 to 64.5). The patients were chosen to undergo a cementless procedure based on age and preoperative radiological and intraoperative bone quality. Outcome was assessed using the Knee Society knee and function scores and range of motion (ROM), complications, and revisions.Aims
Methods
The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface. Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up.Aims
Methods
Implant-associated infection is a major source
of morbidity in orthopaedic surgery. There has been extensive research
into the development of materials that prevent biofilm formation,
and hence, reduce the risk of infection. Silver nanoparticle technology
is receiving much interest in the field of orthopaedics for its
antimicrobial properties, and the results of studies to date are
encouraging. Antimicrobial effects have been seen when silver nanoparticles are
used in trauma implants, tumour prostheses, bone cement, and also
when combined with hydroxyapatite
Reverse total shoulder arthroplasty (rTSA) can be used in complex cases when the glenoid requires reconstruction. In this study, a baseplate with composite bone autograft and a central trabecular titanium peg was implanted, and its migration was assessed for two years postoperatively using radiostereometric analysis (RSA). A total of 14 patients who underwent a rTSA with an autograft consented to participate. Of these, 11 had a primary rTSA using humeral head autograft and three had a revision rTSA with autograft harvested from the iliac crest. The mean age of the patients was 66 years (39 to 81). Tantalum beads were implanted in the scapula around the glenoid. RSA imaging (stereographic radiographs) was undertaken immediately postoperatively and at three, six, 12, and 24 months. Analysis was completed using model-based RSA software. Outcomes were collected preoperatively and at two years postoperatively, including the Oxford Shoulder Score, the American Shoulder and Elbow Score, and a visual analogue score for pain. A Constant score was also obtained for the assessment of strength and range of motion.Aims
Methods
We have updated our previous randomised controlled trial comparing release of chromium (Cr) and cobalt (Co) ions and included levels of titanium (Ti) ions. We have compared the findings from 28 mm metal-on-metal total hip replacement, performed using titanium CLS/Spotorno femoral components and titanium AlloFit acetabular components with Metasul bearings, with Durom hip resurfacing using a Metasul articulation or bearing and a titanium plasma-sprayed