Advertisement for orthosearch.org.uk
Results 1 - 50 of 168
Results per page:
The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 23 - 26
1 Nov 2014
Cooper HJ Della Valle CJ

Dislocation remains among the most common complications of, and reasons for, revision of both primary and revision total hip replacements (THR). Hence, there is great interest in maximising stability to prevent this complication. Head size has been recognised to have a strong influence on the risk of dislocation post-operatively. As femoral head size increases, stability is augmented, secondary to an increase in impingement-free range of movement. Larger head sizes also greatly increase the ‘jump distance’ required for the head to dislocate in an appropriately positioned cup. Level-one studies support the use of larger diameter heads as they decrease the risk of dislocation following primary and revision THR. Highly cross-linked polyethylene has allowed us to increase femoral head size, without a marked increase in wear. However, the thin polyethylene liners necessary to accommodate larger heads may increase the risk of liner fracture and larger heads have also been implicated in causing soft-tissue impingement resulting in groin pain. Larger diameter heads also impart larger forces on the femoral trunnion, which may contribute to corrosion, metal release, and adverse local tissue reactions. Alternative large bearings including large ceramic heads and dual mobility bearings may mitigate some of these risks, and several of these devices have been used with clinical success. Cite this article: Bone Joint J 2014;96-B(11 Suppl A):23–6


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 52 - 54
1 Nov 2012
Rodriguez JA Rathod PA

Large femoral heads have been used with increasing frequency over the last decade. The prime reason is likely the effect of large heads on stability. The larger head neck ratio, combined with the increased jump distance of larger heads result in a greater arc of impingement free motion, and greater resistance to dislocation in a provocative position. Multiple studies have demonstrated clear clinical efficacy in diminishing dislocation rates with the use of large femoral heads. With crosslinked polyethylene, wear has been shown to be equivalent between larger and smaller heads. However, the stability advantages of increasing diameter beyond 38 mm have not been clearly demonstrated. More importantly, recent data implicates large heads in the increasing prevalence of groin pain and psoas impingement. There are clear benefits with larger femoral head diameters, but the advantages of diameters beyond 38 mm have not yet been demonstrated clinically


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 2 | Pages 202 - 208
1 Feb 2010
Berton C Girard J Krantz N Migaud H

Implantation of a large-diameter femoral head prosthesis with a metal-on-metal bearing surface reduces the risk of dislocation, increases the range of movement, minimises the risk of impingement and, in theory, results in little wear.

Between February 2004 and March 2007 we implanted 100 consecutive total hip replacements with a metal-on-metal bearing and a large femoral head into 92 patients. There were 51 men and 41 women with a mean age of 50 years (18 to 70) at the time of surgery.

Outcome was assessed using the Western Ontario McMaster University osteoarthritis index and the Harris hip score as well as the Devane activity score. These all improved significantly (p < 0.0001). At the last follow-up there were no cases of dislocation, no impingement, a good range of movement and no osteolysis, but seven revisions, two for infection and five for aseptic loosening. The probability of groin pain increased if the other acetabular component inclination exceeded 50° (p = 0.0007). At 4.8 years of follow-up, the projected survival of the Durom acetabular component, with revision for any reason, was 92.4% (sd 2.8) (95% confidence interval 89.6 to 95.2).

The design of the component made it difficult both to orientate and seat, which when combined with a poor porous coating, produced unpredictable fixation and a low survival at five years.


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1633 - 1640
1 Oct 2021
Lex JR Evans S Parry MC Jeys L Stevenson JD

Aims. Proximal femoral endoprosthetic replacements (PFEPRs) are the most common reconstruction option for osseous defects following primary and metastatic tumour resection. This study aimed to compare the rate of implant failure between PFEPRs with monopolar and bipolar hemiarthroplasties and acetabular arthroplasties, and determine the optimum articulation for revision PFEPRs. Methods. This is a retrospective review of 233 patients who underwent PFEPR. The mean age was 54.7 years (SD 18.2), and 99 (42.5%) were male. There were 90 patients with primary bone tumours (38.6%), 122 with metastatic bone disease (52.4%), and 21 with haematological malignancy (9.0%). A total of 128 patients had monopolar (54.9%), 74 had bipolar hemiarthroplasty heads (31.8%), and 31 underwent acetabular arthroplasty (13.3%). Results. At a mean 74.4 months follow-up, the overall revision rate was 15.0%. Primary malignancy (p < 0.001) and age < 50 years (p < 0.001) were risk factors for revision. The risks of death and implant failure were similar in patients with primary disease (p = 0.872), but the risk of death was significantly greater for patients who had metastatic bone disease (p < 0.001). Acetabular-related implant failures comprised 74.3% of revisions; however, no difference between hemiarthroplasty or arthroplasty groups (p = 0.209), or between monopolar or bipolar hemiarthroplasties (p = 0.307), was observed. There was greater radiological wear in patients with longer follow-up and primary bone malignancy. Re-revision rates following a revision PFEPR was 34.3%, with dual-mobility bearings having the lowest rate of instability and re-revision (15.4%). Conclusion. Hemiarthroplasty and arthroplasty PFEPRs carry the same risk of revision in the medium term, and is primarily due to acetabular complications. There is no difference in revision rates or erosion between monopolar and bipolar hemiarthroplasties. The main causes of failure were acetabular wear in the hemiarthroplasty group and instability in the arthroplasty group. These risks should be balanced and patient prognosis considered when contemplating the bearing choice. Dual-mobility, constrained bearings, or large diameter heads (> 32 mm) are recommended in all revision PFEPRs. Cite this article: Bone Joint J 2021;103-B(10):1633–1640


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 105 - 111
1 Jul 2020
Engh, Jr. CA McAsey CJ Cororaton AD Ho H Hopper, Jr. RH

Aims. The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. Methods. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083 metal-on-metal (MOM), 90 ceramic-on-ceramic (COC), and 81 surface arthroplasties (SAs). With the exception of the COC, all other groups used cobalt-chromium (CoCr) femoral heads. The mean follow-up was 10 (0.008 to 35) years. Descriptive statistics with revisions per 100 component years (re/100 yr) and survival analysis with revision for any reason as the endpoint were used to compare bearing surfaces. Results. XLPE liners demonstrated a lower cumulative incidence of revision at 15 years compared to the CPE-GA and CPE-GP groups owing to the absence of wear-related revisions (4% for XLPE vs 18%, p = 0.02, and 15%, p = 0.003, respectively). Revisions for adverse local tissue reactions occurred exclusively among the MOM (0.8 re/100 year) and SA groups (0.1 re/100 year). The revision rate for instability was lower among hips with 36 mm and larger head sizes compared to smaller head sizes (0.2% vs 2%, p < 0.001). Conclusion. The introduction of XLPE has eliminated wear-related revisions through 15-year follow-up compared to CPE-GP and CPE-GA. Dislocation incidence has been reduced with the introduction of larger diameter heads but remains a persistent concern. The potential for adverse local tissue reactions with MOM requires continued follow-up. Cite this article: Bone Joint J 2020;102-B(7 Supple B):105–111


The Bone & Joint Journal
Vol. 102-B, Issue 7 | Pages 832 - 837
1 Jul 2020
Dover C Kuiper JH Craig P Shaylor P

Aims. We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following metal-on-polyethylene uncemented total hip arthroplasty. Further data have been collected, to see whether these associations have altered with time and to determine the long-term implications for these patients and our practice. Methods. Patients from our previous study who underwent Trident-Accolade primary total hip arthroplasties using a metal-on-polyethylene bearing in 2009 were reviewed. Patients were invited to have their cobalt and chromium levels retested, and were provided an Oxford Hip Score. Serum ion levels were then compared between groups (28 mm, 36 mm, and 40 mm heads) and over time. Results. Metal ion levels were repeated in 33 patients. When comparing the results of serum metal ion levels over time, regardless of head size, there was a significant increase in both cobalt and chromium levels (p < 0.001). Two patients with larger head sizes had undergone revision arthroplasty with evidence of trunnion damage at surgery. Two patients within the 40 mm subgroup had metal ion levels above the MHRA (Medicines and Healthcare Products Regulatory Agency) threshold for detailed investigation. The increase in cobalt and chromium, when comparing the 36 mm and 40 mm groups with those of the 28 mm group, was not significant (36 mm vs 28 mm; p = 0.092/p = 0.191; 40 mm vs 28 mm; p = 0.200/p = 0.091, respectively). There was no difference, between femoral head sizes, when comparing outcome as measured by the Oxford Hip Score. Conclusion. This study shows an increase in cobalt and chromium levels over time for all modular femoral head sizes in patients with metal-on-polyethylene bearings, with two patients demonstrating ion levels above the MHRA threshold for failure, and a further two patients requiring revision surgery. These results may have clinical implications regarding longer term follow-up of patients and future implant choice, particularly among younger patients. Cite this article: Bone Joint J 2020;102-B(7):832–837


The Bone & Joint Journal
Vol. 101-B, Issue 12 | Pages 1557 - 1562
1 Dec 2019
Tillman R Tsuda Y Puthiya Veettil M Young PS Sree D Fujiwara T Abudu A

Aims. The aim of this study was to present the long-term surgical outcomes, complications, implant survival, and causes of implant failure in patients treated with the modified Harrington procedure using antegrade large diameter pins. Patients and Methods. A cohort of 50 consecutive patients who underwent the modified Harrington procedure for periacetabular metastasis or haematological malignancy between January 1996 and April 2018 were studied. The median follow-up time for all survivors was 3.2 years (interquartile range 0.9 to 7.6 years). Results. The five-year overall survival rate was 33% for all the patients. However, implant survival rates were 100% and 46% at five and ten years, respectively. Eight patients survived beyond five years. There was no immediate perioperative mortality or complications. A total of 15 late complications occurred in 11 patients (22%). Five patients (10%) required further surgery to treat complications. The most frequent complication was pin breakage without evidence of acetabular loosening (6%). Two patients (4%) underwent revision for aseptic loosening at 6.5 and 8.9 years after surgery. Ambulatory status and pain level were improved in 83% and 89%, respectively. Conclusion. The modified Harrington procedure for acetabular destruction has low complication rates, good functional outcome, and improved pain relief in selected patients. Cite this article: Bone Joint J 2019;101-B:1557–1562


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1304 - 1312
1 Oct 2017
Langton DJ Sidaginamale RP Joyce TJ Meek RD Bowsher JG Deehan D Nargol AVF Holland JP

Aims. We sought to determine whether cobalt-chromium alloy (CoCr) femoral stem tapers (trunnions) wear more than titanium (Ti) alloy stem tapers (trunnions) when used in a large diameter (LD) metal-on-metal (MoM) hip arthroplasty system. Patients and Methods. We performed explant analysis using validated methodology to determine the volumetric material loss at the taper surfaces of explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy (n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs with a rough male taper surface and a nominal included angle close to 5.666° were included. Multiple regression modelling was undertaken using taper angle, taper roughness, bearing diameter (horizontal lever arm) as independent variables. Material loss was mapped using a coordinate measuring machine, profilometry and scanning electron microscopy. Results. After adjustment for other factors, CoCr stem tapers were found to have significantly greater volumetric material loss than the equivalent Ti stem tapers. Conclusion. When taper junction damage is identified during revision of a LD MoM hip, it should be suspected that a male taper composed of a standard CoCr alloy has sustained significant changes to the taper cone geometry which are likely to be more extensive than those affecting a Ti alloy stem. Cite this article: Bone Joint J 2017;99-B:1304–12


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 579 - 584
1 May 2016
Osman K Panagiotidou AP Khan M Blunn G Haddad FS

There is increasing global awareness of adverse reactions to metal debris and elevated serum metal ion concentrations following the use of second generation metal-on-metal total hip arthroplasties. The high incidence of these complications can be largely attributed to corrosion at the head-neck interface. Severe corrosion of the taper is identified most commonly in association with larger diameter femoral heads. However, there is emerging evidence of varying levels of corrosion observed in retrieved components with smaller diameter femoral heads. This same mechanism of galvanic and mechanically-assisted crevice corrosion has been observed in metal-on-polyethylene and ceramic components, suggesting an inherent biomechanical problem with current designs of the head-neck interface. We provide a review of the fundamental questions and answers clinicians and researchers must understand regarding corrosion of the taper, and its relevance to current orthopaedic practice. Cite this article: Bone Joint J 2016;98-B:579–84


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 660 - 665
1 May 2016
Jung HJ Song JH Kekatpure AL Adikrishna A Hong HP Lee WJ Chun JM Jeon IH

Aims. The treatment of septic arthritis of the shoulder is challenging. The infection frequently recurs and the clinical outcome can be very poor. We aimed to review the outcomes following the use of continuous negative pressure after open debridement with a large diameter drain in patients with septic arthritis of the shoulder. Patients and Methods. A total of 68 consecutive patients with septic arthritis of the shoulder underwent arthrotomy, irrigation and debridement. A small diameter suction drain was placed in the glenohumeral joint and a large diameter drain was placed in the subacromial space with continuous negative pressure of 15 cm H. 2. O. All patients received a standardised protocol of antibiotics for a mean of 5.1 weeks (two to 11.1). Results. Negative pressure was maintained for a mean of 24 days (14 to 32). A total of 67 patients (98.5%) were cured without further treatment being required. At a mean follow-up of 14 months (three to 72), the mean forward flexion was 123° (80° to 140°) and the mean external rotation was 28°(10° to 40°) in those with a rotator cuff tear, and 125° (85° to 145°) and 35° (15° to 45°) in those without a rotator cuff tear. Conclusion. Continuous negative pressure, following open arthrotomy, irrigation and debridement, was effective in treating septic arthritis of the shoulder. The rate of recurrence was significantly lower than with conventional treatment involving arthroscopic or open debridement reported in the literature. Functional outcomes, even in patients with rotator cuff tears, were excellent. Take home message: Continuous negative pressure is effective in treating septic arthritis of the shoulder. Cite this article: Bone Joint J 2016;98-B:660–5


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 736 - 740
1 Jun 2016
Sassoon AA Barrack RL

The use of large-diameter metal-on-metal (MoM) components in total hip arthroplasty (THA) is associated with an increased risk of early failure due to adverse local tissue reaction to metal debris (ARMD) in response to the release of metal ions from the bearing couple and/or head-neck taper corrosion. The aim of this paper was to present a review of the incidence and natural history of ARMD and the forms of treatment, with a focus on the need for and extent of resection or debulking of the pseudotumour. An illustrative case report is presented of a patient with an intra-pelvic pseudotumour associated with a large diameter MoM THA, which was treated successfully with revision of the bearing surface to a dual mobility couple and retention of the well-fixed acetabular and femoral components. The pseudotumour was left in situ. Resolution of the intra-pelvic mass and normalisation of metal ion levels was observed seven months post-operatively. Cite this article: Bone Joint J 2016;98-B:736–40


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1600 - 1608
1 Dec 2018
Bouaicha S Ernstbrunner L Jud L Meyer DC Snedeker JG Bachmann E

Aims. In patients with a rotator cuff tear, tear pattern and tendon involvement are known risk factors for the development of pseudoparalysis of the shoulder. It remains unclear, however, why similar tears often have very different functional consequences. The present study hypothesizes that individual shoulder anatomy, specifically the moment arms (MAs) of the rotator cuff (RC) and the deltoid muscle, as well as their relative recruitment during shoulder abduction, plays a central role in pseudoparalysis. Materials and Methods. Biomechanical and clinical analyses of the pseudoparalytic shoulder were conducted based on the ratio of the RC/deltoid MAs, which were used to define a novel anatomical descriptor called the Shoulder Abduction Moment (SAM) index. The SAM index is the ratio of the radii of two concentric spheres based on the centre of rotation of the joint. One sphere captures the humeral head (numerator) and the other the deltoid origin of the acromion (denominator). A computational rigid body simulation was used to establish the functional link between the SAM index and a potential predisposition for pseudoparalysis. A retrospective radiological validation study based on these measures was also undertaken using two cohorts with and without pseudoparalysis and massive RC tears. Results. Decreased RC activity and improved glenohumeral stability was predicted by simulations of SAM indices with larger diameters of the humeral head, being consequently beneficial for joint stability. Clinical investigation of the SAM index showed significant risk of pseudoparalysis in patients with massive tears and a SAM < 0.77 (odds ratio (OR) 11). Conclusion. The SAM index, which represents individual biomechanical characteristics of shoulder morphology, plays a determinant role in the presence or absence of pseudoparalysis in shoulders with massive RC tears


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 917 - 924
1 Jul 2016
Whittaker RK Hothi HS Meswania JM Berber R Blunn GW Skinner JA Hart AJ

Aims. Surgeons have commonly used modular femoral heads and stems from different manufacturers, although this is not recommended by orthopaedic companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head taper junction in two groups of retrieved hips; those with mixed manufacturers (MM) and those from the same manufacturer (SM). . Materials and Methods. We identified 151 retrieved hips with large-diameter cobalt-chromium heads; 51 of two designs that had been paired with stems from different manufacturers (MM) and 100 of seven designs paired with stems from the same manufacturer (SM). We determined the severity of corrosion with the Goldberg corrosion score and the volume of material loss at the head/stem junction. We used multivariable statistical analysis to determine if there was a significant difference between the two groups. . Results. We found no significant difference in the corrosion scores of the two groups. The median rate of material loss at the head/stem junction for the MM and SM groups were 0.39 mm. 3. /year (0.00 to 4.73) and 0.46 mm. 3. /year (0.00 to 6.71) respectively; this difference was not significant after controlling for confounding factors (p = 0.06). . Conclusion. The use of stems with heads of another manufacturer does not appear to affect the amount of metal lost from the surfaces between these two components at total hip arthroplasty. Other surgical, implant and patient factors should be considered when determining the mechanisms of failure of large diameter metal-on-metal hip arthroplasties. Cite this article: Bone Joint J 2016;98-B:917–24


The Bone & Joint Journal
Vol. 100-B, Issue 11 | Pages 1434 - 1441
1 Nov 2018
Blakeney WG Beaulieu Y Puliero B Lavigne M Roy A Massé V Vendittoli P

Aims. This study reports the mid-term results of total hip arthroplasty (THA) performed using a monoblock acetabular component with a large-diameter head (LDH) ceramic-on-ceramic (CoC) bearing. Patients and Methods. Of the 276 hips (246 patients) included in this study, 264 (96%) were reviewed at a mean of 67 months (48 to 79) postoperatively. Procedures were performed with a mini posterior approach. Clinical and radiological outcomes were recorded at regular intervals. A noise assessment questionnaire was completed at last follow-up. Results. There were four re-operations (1%) including one early revision for insufficient primary fixation (0.4%). No hip dislocation was reported. The mean University of California, Los Angeles (UCLA) activity score, 12-Item Short-Form Health Survey (SF-12) Mental Component Summary (MCS) score, SF-12 Physical Component Summary (PCS) score, Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) score, and Forgotten Joint Score (FJS) were 6.6 (2 to 10), 52.8 (25.5 to 65.7), 53.0 (27.2 to 66.5), 7.7 (0 to 63), and 88.5 (23 to 100), respectively. No signs of loosening or osteolysis were observed on radiological review. The incidence of squeaking was 23% (n = 51/225). Squeaking was significantly associated with larger head diameter (p < 0.001), younger age (p < 0.001), higher SF-12 PCS (p < 0.001), and UCLA scores (p < 0.001). Squeaking did not affect patient satisfaction, with 100% of the squeaking hips satisfied with the surgery. Conclusion. LDH CoC THAs have demonstrated excellent functional outcomes at medium-term follow-up, with very low revision rate and no dislocations. The high incidence of squeaking did not affect patient satisfaction or function. LDH CoC with a monoblock acetabular component has the potential to provide long term implant survivorship with unrestricted activity, while avoiding implant impingement, liner fracture at insertion, and hip instability. Cite this article: Bone Joint J 2018;100-B:1434–41


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 555 - 564
1 Jun 2024
Leal J Holland CT Cochrane NH Seyler TM Jiranek WA Wellman SS Bolognesi MP Ryan SP

Aims

This study aims to assess the relationship between history of pseudotumour formation secondary to metal-on-metal (MoM) implants and periprosthetic joint infection (PJI) rate, as well as establish ESR and CRP thresholds that are suggestive of infection in these patients. We hypothesized that patients with a pseudotumour were at increased risk of infection.

Methods

A total of 1,171 total hip arthroplasty (THA) patients with MoM articulations from August 2000 to March 2014 were retrospectively identified. Of those, 328 patients underwent metal artefact reduction sequence MRI and had minimum two years’ clinical follow-up, and met our inclusion criteria. Data collected included demographic details, surgical indication, laterality, implants used, history of pseudotumour, and their corresponding preoperative ESR (mm/hr) and CRP (mg/dl) levels. Multivariate logistic regression modelling was used to evaluate PJI and history of pseudotumour, and receiver operating characteristic curves were created to assess the diagnostic capabilities of ESR and CRP to determine the presence of infection in patients undergoing revision surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 31 - 34
1 Jan 2006
Loughead JM Starks I Chesney D Matthews JNS McCaskie AW Holland JP

Resurfacing arthroplasty of the hip is being performed more frequently in the United Kingdom. The majority of these patients are younger than 55 years of age, and in this group the key benefits include conservation of femoral bone stock and the potential reduction in the rate of dislocation afforded by the larger resurfacing head. Early aseptic loosening is well recognised in patients younger than 55 years of age, and proponents of resurfacing believe that the improved wear characteristics of the metal-on-metal bearing may improve the long-term survival of this implant. There has been some concern, however, that resurfacing may not be conservative of acetabular bone. We compared a series of 33 consecutive patients who had a hybrid total hip arthroplasty with an uncemented acetabular component and a cemented femoral implant, with 35 patients undergoing a Birmingham hip resurfacing arthroplasty. We compared the diameter of the implanted acetabulum in both groups and, because they were not directly comparable, we corrected for patient size by measuring the diameter of the contralateral femoral head. The data were analysed using unpaired t-tests and analysis of covariance. There was a significantly larger acetabulum in the Birmingham arthroplasty group (mean diameter 56.6 mm vs 52.0 mm; p < 0.001). However, this group had a significantly larger femoral head diameter on the contralateral side (p = 0.03). Analysis of covariance revealed a significant difference between the mean size of the acetabular component implanted in the two operations. The greatest difference in the size of acetabulum was in those patients with a larger diameter of the femoral head. This study shows that more bone is removed from the acetabulum in hip resurfacing than during hybrid total hip arthroplasty, a difference which is most marked in larger patients


The Bone & Joint Journal
Vol. 105-B, Issue 8 | Pages 912 - 919
1 Aug 2023
Cunningham LJ Walton M Bale S Trail IA

Aims

Reverse total shoulder arthroplasty (rTSA) can be used in complex cases when the glenoid requires reconstruction. In this study, a baseplate with composite bone autograft and a central trabecular titanium peg was implanted, and its migration was assessed for two years postoperatively using radiostereometric analysis (RSA).

Methods

A total of 14 patients who underwent a rTSA with an autograft consented to participate. Of these, 11 had a primary rTSA using humeral head autograft and three had a revision rTSA with autograft harvested from the iliac crest. The mean age of the patients was 66 years (39 to 81). Tantalum beads were implanted in the scapula around the glenoid. RSA imaging (stereographic radiographs) was undertaken immediately postoperatively and at three, six, 12, and 24 months. Analysis was completed using model-based RSA software. Outcomes were collected preoperatively and at two years postoperatively, including the Oxford Shoulder Score, the American Shoulder and Elbow Score, and a visual analogue score for pain. A Constant score was also obtained for the assessment of strength and range of motion.


The Bone & Joint Journal
Vol. 105-B, Issue 9 | Pages 946 - 952
1 Sep 2023
Dhawan R Young DA Van Eemeren A Shimmin A

Aims

The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit.

Methods

Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis.


The Bone & Joint Journal
Vol. 107-B, Issue 1 | Pages 50 - 57
1 Jan 2025
Hussein Y Iljazi A Sørensen MS Overgaard S Petersen MM

Aims

Dislocation is a major concern following total hip arthroplasty (THA) for osteoarthritis (OA). Both dual-mobility components and standard acetabular components with large femoral heads are used to reduce the risk of dislocation. We investigated whether dual-mobility components are superior to standard components in reducing the two-year dislocation and revision risk in a propensity-matched sample from the Danish Hip Arthroplasty Register (DHR).

Methods

This population-based cohort study analyzed data from the DHR and the Danish National Patient Register. We included all patients undergoing primary THA for OA from January 2010 to December 2019 with either dual-mobility or standard acetabular components with metal-on-polyethylene or ceramic-on-polyethylene articulations with a 36 mm femoral head. The samples were propensity score-matched on patient and implant characteristics. The primary outcome was the difference in the absolute risk of dislocation within two years, with a secondary outcome of the difference in the absolute risk of revision surgery of any cause within the same timeframe. The cumulative incidence of dislocation was calculated using the Aalen-Johansen estimator, while the difference in absolute risk was estimated using absolute risk regression (ARR).


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 29 - 34
1 Jan 2023
Fransen BL Bengoa FJ Neufeld ME Sheridan GA Garbuz DS Howard LC

Aims

Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up.

Methods

We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 393 - 395
1 Mar 2007
Briant-Evans TW Norton MR Fern ED

We describe two cases of fracture of Corin Taper-Fit stems used for cement-in-cement revision of congenital dysplasia of the hip. Both prostheses were implanted in patients in their 50s, with high offsets (+7.5 mm and +3.5 mm), one with a large diameter (48 mm) head and one with a constrained acetabular component. Fracture of the stems took place at nine months and three years post-operatively following low-demand activity. Both fractures occurred at the most medial of the two stem introducer holes in the neck of the prosthesis, a design feature that is unique to the Taper-Fit stem. We would urge caution in the use of these particular stems for cement-in-cement revisions


The Bone & Joint Journal
Vol. 104-B, Issue 4 | Pages 504 - 509
1 Apr 2022
Kennedy IW Farhan-Alanie OM Young D Kelly MP Young PS

Aims

The aim of this study was to assess the clinical and radiological outcomes of an antiprotrusio acetabular cage (APC) when used in the surgical treatment of periacetabular bone metastases.

Methods

This retrospective cohort study using a prospectively collected database involved 56 patients who underwent acetabular reconstruction for periacetabular bone metastases or haematological malignancy using a single APC between January 2009 and 2020. The mean follow-up was 20 months (1 to 143). The primary outcome measure was implant survival. Postoperative radiographs were analyzed for loosening and failure. Patient and implant survival were assessed using a competing risk analysis. Secondary parameters included primary malignancy, oncological treatment, surgical factors, length of stay in hospital, and postoperative complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 6 | Pages 913 - 917
1 Aug 2003
Clarke MT Lee PTH Arora A Villar RN

Metal-on-metal (MOM) bearings for hip arthroplasty are increasing in popularity. Concern remains, however, regarding the potential toxicological effects of the metal ions which these bearings release. The serum levels of cobalt and chromium in 22 patients who had undergone MOM resurfacing arthroplasty were compared with a matched group of 22 patients who had undergone 28 mm MOM total hip arthroplasty (THA). At a median of 16 months (7 to 56) after resurfacing arthroplasty, we found the median serum levels of cobalt and chromium to be 38 nmol/l (14 to 44) and 53 nmol/l (23 to 165) respectively. These were significantly greater than the levels after 28 mm MOM THA which were 22 nmol/l (15 to 87, p = 0.021) and 19 nmol/l (2 to 58, p < 0.001) respectively. Since the upper limit for normal patients without implants is typically 5 nmol/l, both groups had significantly raised levels of metal ions. MOM bearings of large diameter, however, result in a greater systemic exposure of cobalt and chromium ions than bearings of small diameter. This may be of relevance for potential long-term side-effects. It is not known to what extent this difference is due to corrosion of the surfaces of the component or of the wear particles produced


The Journal of Bone & Joint Surgery British Volume
Vol. 44-B, Issue 3 | Pages 503 - 519
1 Aug 1962
Little K Kelly M Courts A

1. The appearance of decalcified bone matrix in the electron microscope is described. 2. In the matrix two types of collagen fibril have been distinguished. Differences observed are in solubility, x-ray diffraction pattern and appearance. In infant bone the form which appears as fine fibrils predominates. In adult bone the form which appears as tubular fibrils of larger diameter predominates. 3. In bones from elderly subjects the chemical reaction employed to convert collagen into eucollagen sometimes hydrolyses fatty acid esters, and lines due to the free fatty acid are found on the x-ray diffraction patterns of the insoluble residue after citrate extraction. 4. In ancient bones and fossils the stable tubular form of collagen survives, but not the fine fibrils. 5. When decalcified, the matrix in osteoporotic bones loses its architecture and fibrillar form. Under conditions in which only a small fraction is dissolved from normal bone most of the collagen in osteoporotic bone disperses in citric acid. The insoluble residue then gives a modified x-ray diffraction pattern. 6. Evidence has been produced to suggest that the immediate cause of many forms of osteoporosis is some local factor affecting the osteocytes, rather than a general chemical effect


The Bone & Joint Journal
Vol. 104-B, Issue 5 | Pages 598 - 603
1 May 2022
Siljander MP Gausden EB Wooster BM Karczewski D Sierra RJ Trousdale RT Abdel MP

Aims

The aim of this study was to evaluate the incidence of liner malseating in two commonly used dual-mobility (DM) designs. Secondary aims included determining the risk of dislocation, survival, and clinical outcomes.

Methods

We retrospectively identified 256 primary total hip arthroplasties (THAs) that included a DM component (144 Stryker MDM and 112 Zimmer-Biomet G7) in 233 patients, performed between January 2012 and December 2019. Postoperative radiographs were reviewed independently for malseating of the liner by five reviewers. The mean age of the patients at the time of THA was 66 years (18 to 93), 166 (65%) were female, and the mean BMI was 30 kg/m2 (17 to 57). The mean follow-up was 3.5 years (2.0 to 9.2).


The Bone & Joint Journal
Vol. 104-B, Issue 2 | Pages 206 - 211
1 Feb 2022
Bloch BV White JJE Matar HE Berber R Manktelow ARJ

Aims

Total hip arthroplasty (THA) is a very successful and cost-effective operation, yet debate continues about the optimum fixation philosophy in different age groups. The concept of the 'cementless paradox' and the UK 'Getting it Right First Time' initiative encourage increased use of cemented fixation due to purported lower revision rates, especially in elderly patients, and decreased cost.

Methods

In a high-volume, tertiary referral centre, we identified 10,112 THAs from a prospectively collected database, including 1,699 cemented THAs, 5,782 hybrid THAs, and 2,631 cementless THAs. The endpoint was revision for any reason. Secondary analysis included examination of implant survivorship in patients aged over 70 years, over 75 years, and over 80 years at primary THA.


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 8 - 11
1 Jan 2022
Wright-Chisem J Elbuluk AM Mayman DJ Jerabek SA Sculco PK Vigdorchik JM

Dislocation following total hip arthroplasty (THA) is a well-known and potentially devastating complication. Clinicians have used many strategies in attempts to prevent dislocation since the introduction of THA. While the importance of postoperative care cannot be ignored, particular emphasis has been placed on preoperative planning in the prevention of dislocation. The strategies have progressed from more traditional approaches, including modular implants, the size of the femoral head, and augmentation of the offset, to newer concepts, including patient-specific component positioning combined with computer navigation, robotics, and the use of dual-mobility implants. As clinicians continue to pursue improved outcomes and reduced complications, these concepts will lay the foundation for future innovation in THA and ultimately improved outcomes.

Cite this article: Bone Joint J 2022;104-B(1):8–11.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 73 - 77
1 Jul 2021
Lawrie CM Barrack RL Nunley RM

Aims

Dual mobility (DM) implants have been shown to reduce the dislocation rate after total hip arthroplasty (THA), but there remain concerns about the use of cobalt chrome liners inserted into titanium shells. The aim of this study was to assess the clinical outcomes, metal ion levels, and periprosthetic femoral bone mineral density (BMD) at mid-term follow-up in young, active patients receiving a modular DM THA.

Methods

This was a prospective study involving patients aged < 65 years, with a BMI of < 35 kg/m2, and University of California, Los Angeles activity score of > 6 who underwent primary THA with a modular cobalt chrome acetabular liner, highly cross-linked polyethylene mobile bearing, and a cementless titanium femoral stem. Patient-reported outcome measures, whole blood metal ion levels (μg/l), and periprosthetic femoral BMD were measured at baseline and at one, two, and five years postoperatively. The results two years postoperatively for this cohort have been previously reported.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 66 - 72
1 Jul 2021
Hernandez NM Hinton ZW Wu CJ Lachiewicz PF Ryan SP Wellman SS

Aims

Modular dual mobility (MDM) acetabular components are often used with the aim of reducing the risk of dislocation in revision total hip arthroplasty (THA). There is, however, little information in the literature about its use in this context. The aim of this study, therefore, was to evaluate the outcomes in a cohort of patients in whom MDM components were used at revision THA, with a mean follow-up of more than five years.

Methods

Using the database of a single academic centre, 126 revision THAs in 117 patients using a single design of an MDM acetabular component were retrospectively reviewed. A total of 94 revision THAs in 88 patients with a mean follow-up of 5.5 years were included in the study. Survivorship was analyzed with the endpoints of dislocation, reoperation for dislocation, acetabular revision for aseptic loosening, and acetabular revision for any reason. The secondary endpoints were surgical complications and the radiological outcome.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 165 - 170
1 Jun 2021
Larson DJ Rosenberg JH Lawlor MA Garvin KL Hartman CW Lyden E Konigsberg BS

Aims

Stemmed tibial components are frequently used in revision total knee arthroplasty (TKA). The purpose of this study was to evaluate patient satisfaction, overall pain, and diaphyseal tibial pain in patients who underwent revision TKA with cemented or uncemented stemmed tibial components.

Methods

This is a retrospective cohort study involving 110 patients with revision TKA with cemented versus uncemented stemmed tibial components. Patients who underwent revision TKA with stemmed tibial components over a 15-year period at a single institution with at least two-year follow-up were assessed. Pain was evaluated through postal surveys. There were 63 patients with cemented tibial stems and 47 with uncemented stems. Radiographs and Knee Society Scores were used to evaluate for objective findings associated with pain or patient dissatisfaction. Postal surveys were analyzed using Fisher’s exact test and the independent-samples t-test. Logistic regression was used to adjust for age, sex, and preoperative bone loss.


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 479 - 485
1 Mar 2021
Nugent M Young SW Frampton CM Hooper GJ

Aims

Joint registries typically use revision of an implant as an endpoint and report survival rates after a defined number of years. However, reporting lifetime risk of revision may be more meaningful, especially in younger patients. We aimed to assess lifetime risk of revision for patients in defined age groups at the time of primary surgery.

Methods

The New Zealand Joint Registry (NZJR) was used to obtain rates and causes of revision for all primary total hip arthroplasties (THAs) performed between January 1999 and December 2016. The NZJR is linked to the New Zealand Registry of Births, Deaths and Marriages to obtain complete and accurate data. Patients were stratified by age at primary surgery, and lifetime risk of revision calculated according to age, sex, and American Society of Anesthesiologists (ASA) classification. The most common causes for revision were also analyzed for each age group.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 305 - 308
1 Feb 2021
Howell M Rae FJ Khan A Holt G

Aims

Iliopsoas pathology is a relatively uncommon cause of pain following total hip arthroplasty (THA), typically presenting with symptoms of groin pain on active flexion and/or extension of the hip. A variety of conservative and surgical treatment options have been reported. In this retrospective cohort study, we report the incidence of iliopsoas pathology and treatment outcomes.

Methods

A retrospective review of 1,000 patients who underwent THA over a five-year period was conducted, to determine the incidence of patients diagnosed with iliopsoas pathology. Outcome following non-surgical and surgical management was assessed.


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 123 - 130
1 Jan 2021
Lapner P Pollock JW Laneuville O Uhthoff HK Zhang T Sheikh A McIlquham K Trudel G

Aims

Despite recent advances in arthroscopic rotator cuff repair, re-tear rates remain high. New methods to improve healing rates following rotator cuff repair must be sought. Our primary objective was to determine if adjunctive bone marrow stimulation with channelling five to seven days prior to arthroscopic cuff repair would lead to higher Western Ontario Rotator Cuff (WORC) scores at 24 months postoperatively compared with no channelling.

Methods

A prospective, randomized controlled trial was conducted in patients undergoing arthroscopic rotator cuff repair. Patients were randomized to receive either a percutaneous bone channelling of the rotator cuff footprint or a sham procedure under ultrasound guidance five to seven days prior to index surgery. Outcome measures included the WORC, American Shoulder and Elbow Surgeons (ASES), and Constant scores, strength, ultrasound-determined healing rates, and adverse events.


Aims

Surgical treatment of hip fracture is challenging; the bone is porotic and fixation failure can be catastrophic. Novel implants are available which may yield superior clinical outcomes. This study compared the clinical effectiveness of the novel X-Bolt Hip System (XHS) with the sliding hip screw (SHS) for the treatment of fragility hip fractures.

Methods

We conducted a multicentre, superiority, randomized controlled trial. Patients aged 60 years and older with a trochanteric hip fracture were recruited in ten acute UK NHS hospitals. Participants were randomly allocated to fixation of their fracture with XHS or SHS. A total of 1,128 participants were randomized with 564 participants allocated to each group. Participants and outcome assessors were blind to treatment allocation. The primary outcome was the EuroQol five-dimension five-level health status (EQ-5D-5L) utility at four months. The minimum clinically important difference in utility was pre-specified at 0.075. Secondary outcomes were EQ-5D-5L utility at 12 months, mortality, residential status, mobility, revision surgery, and radiological measures.


The Bone & Joint Journal
Vol. 102-B, Issue 12 | Pages 1752 - 1759
1 Dec 2020
Tsuda Y Tsoi K Stevenson JD Laitinen M Ferguson PC Wunder JS Griffin AM van de Sande MAJ van Praag V Leithner A Fujiwara T Yasunaga H Matsui H Parry MC Jeys LM

Aims

Our aim was to develop and validate nomograms that would predict the cumulative incidence of sarcoma-specific death (CISSD) and disease progression (CIDP) in patients with localized high-grade primary central and dedifferentiated chondrosarcoma.

Methods

The study population consisted of 391 patients from two international sarcoma centres (development cohort) who had undergone definitive surgery for a localized high-grade (histological grade II or III) conventional primary central chondrosarcoma or dedifferentiated chondrosarcoma. Disease progression captured the first event of either metastasis or local recurrence. An independent cohort of 221 patients from three additional hospitals was used for external validation. Two nomograms were internally and externally validated for discrimination (c-index) and calibration plot.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 52 - 56
1 Jul 2020
Elkins JM Dennis DA Kleeman-Forsthuber L Yang CC Miner TM Jennings JM

Aims

Of growing concern in arthroplasty is the emergence of atypical infections, particularly Cutibacterium (formerly Propionibacterium) sp. infections. Currently, the dermal colonization rate of Cutibacterium about the hip is unknown. Therefore, the aim of this study was to investigate colonization rates of Cutibacterium sp. at locations approximating anterior and posterolateral approaches to the hip joint.

Methods

For this non-randomized non-blinded study, 101 adult patients scheduled for hip or knee surgery were recruited. For each, four 3 mm dermal punch biopsies were collected after administration of anaesthesia, but prior to antibiotics. Prebiopsy skin preparation consisted of a standardized preoperative 2% chlorhexidine skin cleansing protocol and an additional 70% isopropyl alcohol mechanical skin scrub immediately prior to biopsy collection. Two skin samples 10 cm apart were collected from a location approximating a standard direct anterior skin incision, and two samples 10 cm apart were collected from a lateral skin incision (suitable for posterior, direct-lateral, or anterolateral approaches). Samples were cultured for two weeks using a protocol optimized for Cutibacterium.


The Bone & Joint Journal
Vol. 102-B, Issue 11 | Pages 1457 - 1466
2 Nov 2020
Cha Y Yoo J Kim J Park C Ahn Y Choy W Ha Y Koo K

Aims

To evaluate the rate of dislocation following dual mobility total hip arthroplasty (DM-THA) in patients with displaced femoral neck fractures, and to compare rates of dislocation, surgical-site infection, reoperation, and one-year mortality between DM-THA and bipolar hemiarthroplasty (BHA).

Methods

Studies were selected based on the following criteria: 1) study design (retrospective cohort studies, prospective cohort studies, retrospective comparative studies, prospective comparative studies, and randomized controlled studies (RCTs)); 2) study population (patients with femoral neck fracture); 3) intervention (DM-THA or BHA); and 4) outcomes (complications during postoperative follow-up and clinical results). Pooled meta-analysis was carried out to evaluate the dislocation rate after DM-THA and to compare outcomes between DM-THA and BHA.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 27 - 32
1 Jul 2020
Heckmann N Weitzman DS Jaffri H Berry DJ Springer BD Lieberman JR

Aims

Dual mobility (DM) bearings are an attractive treatment option to obtain hip stability during challenging primary and revision total hip arthroplasty (THA) cases. The purpose of this study was to analyze data submitted to the American Joint Replacement Registry (AJRR) to characterize utilization trends of DM bearings in the USA.

Methods

All primary and revision THA procedures reported to AJRR from 2012 to 2018 were analyzed. Patients of all ages were included and subdivided into DM and traditional bearing surface cohorts. Patient demographics, geographical region, hospital size, and teaching affiliation were assessed. Associations were determined by chi-squared analysis and logistic regression was performed to assess outcome variables.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 20 - 26
1 Jul 2020
Romero J Wach A Silberberg S Chiu Y Westrich G Wright TM Padgett DE

Aims

This combined clinical and in vitro study aimed to determine the incidence of liner malseating in modular dual mobility (MDM) constructs in primary total hip arthroplasties (THAs) from a large volume arthroplasty centre, and determine whether malseating increases the potential for fretting and corrosion at the modular metal interface in malseated MDM constructs using a simulated corrosion chamber.

Methods

For the clinical arm of the study, observers independently reviewed postoperative radiographs of 551 primary THAs using MDM constructs from a single manufacturer over a three-year period, to identify the incidence of MDM liner-shell malseating. Multivariable logistic regression analysis was performed to identify risk factors including age, sex, body mass index (BMI), cup design, cup size, and the MDM case volume of the surgeon. For the in vitro arm, six pristine MDM implants with cobalt-chrome liners were tested in a simulated corrosion chamber. Three were well-seated and three were malseated with 6° of canting. The liner-shell couples underwent cyclic loading of increasing magnitudes. Fretting current was measured throughout testing and the onset of fretting load was determined by analyzing the increase in average current.


The Bone & Joint Journal
Vol. 102-B, Issue 7 | Pages 811 - 821
1 Jul 2020
You D Sepehri A Kooner S Krzyzaniak H Johal H Duffy P Schneider P Powell J

Aims

Dislocation is the most common indication for further surgery following total hip arthroplasty (THA) when undertaken in patients with a femoral neck fracture. This study aimed to assess the complication rates of THA with dual mobility components (THA-DMC) following a femoral neck fracture and to compare outcomes between THA-DMC, conventional THA, and hemiarthroplasty (HA).

Methods

We performed a systematic review of all English language articles on THA-DMC published between 2010 and 2019 in the MEDLINE, EMBASE, and Cochrane databases. After the application of rigorous inclusion and exclusion criteria, 23 studies dealing with patients who underwent treatment for a femoral neck fracture using THA-DMC were analyzed for the rate of dislocation. Secondary outcomes included reoperation, periprosthetic fracture, infection, mortality, and functional outcome. The review included 7,189 patients with a mean age of 77.8 years (66.4 to 87.6) and a mean follow-up of 30.9 months (9.0 to 68.0).


Aims

The use of frozen tumour-bearing autograft combined with a vascularized fibular graft (VFG) represents a new technique for biological reconstruction of massive bone defect. We have compared the clinical outcomes between this technique and Capanna reconstruction.

Methods

From June 2011 to January 2016 a retrospective study was carried out of patients with primary osteosarcoma of lower limbs who underwent combined biological intercalary reconstruction. Patients were categorized into two groups based on the reconstructive technique: frozen tumour-bearing autograft combined with concurrent VFG (Group 1) and the Capanna method (Group 2). Demographics, operating procedures, oncological outcomes, graft union, limb function, and postoperative complications were compared.


The Bone & Joint Journal
Vol. 102-B, Issue 8 | Pages 1003 - 1009
1 Aug 2020
Mononen H Sund R Halme J Kröger H Sirola J

Aims

There is evidence that prior lumbar fusion increases the risk of dislocation and revision after total hip arthroplasty (THA). The relationship between prior lumbar fusion and the effect of femoral head diameter on THA dislocation has not been investigated. We examined the relationship between prior lumbar fusion or discectomy and the risk of dislocation or revision after THA. We also examined the effect of femoral head component diameter on the risk of dislocation or revision.

Methods

Data used in this study were compiled from several Finnish national health registers, including the Finnish Arthroplasty Register (FAR) which was the primary source for prosthesis-related data. Other registers used in this study included the Finnish Health Care Register (HILMO), the Social Insurance Institutions (SII) registers, and Statistics Finland. The study was conducted as a prospective retrospective cohort study. Cox proportional hazards regression and Kaplan-Meier survival analysis were used for analysis.


The Bone & Joint Journal
Vol. 102-B, Issue 6 | Pages 709 - 715
1 Jun 2020
Abdelsamie KR Elhawary I Ali H Ali M EL-Shafie M Dominic Meek RM

Aims

Femoral revision component subsidence has been identified as predicting early failure in revision hip surgery. This comparative cohort study assessed the potential risk factors of subsidence in two commonly used femoral implant designs.

Methods

A comparative cohort study was undertaken, analyzing a consecutive series of patients following revision total hip arthroplasties using either a tapered-modular (TM) fluted titanium or a porous-coated cylindrical modular (PCM) titanium femoral component, between April 2006 and May 2018. Clinical and radiological assessment was compared for both treatment cohorts. Risk factors for subsidence were assessed and compared.


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1192 - 1198
1 Oct 2019
Sköldenberg OG Rysinska AD Chammout G Salemyr M Mukka SS Bodén H Eisler T

Aims

Radiostereometric analysis (RSA) studies of vitamin E-doped, highly crosslinked polyethylene (VEPE) liners show low head penetration rates in cementless acetabular components. There is, however, currently no data on cemented VEPE acetabular components in total hip arthroplasty (THA). The aim of this study was to evaluate the safety of a new cemented VEPE component, compared with a conventional polyethylene (PE) component regarding migration, head penetration, and clinical results.

Patients and Methods

We enrolled 42 patients (21 male, 21 female) with osteoarthritis and a mean age of 67 years (sd 5), in a double-blinded, noninferiority, randomized controlled trial. The subjects were randomized in a 1:1 ratio to receive a reverse hybrid THA with a cemented component of either argon-gas gamma-sterilized PE component (controls) or VEPE, with identical geometry. The primary endpoint was proximal implant migration of the component at two years postoperatively measured with RSA. Secondary endpoints included total migration of the component, penetration of the femoral head into the component, and patient-reported outcome measurements.


The Bone & Joint Journal
Vol. 101-B, Issue 8 | Pages 897 - 901
1 Aug 2019
Konan S Alazzawi S Yoon B Cha Y Koo K

Ceramic bearings have several desirable properties, such as resistance to wear, hardness, and biocompatibility, that favour it as an articulating surface in hip arthroplasty. However, ceramic fracture remains a concern. We have reviewed the contemporary literature, addressing the factors that can influence the incidence of ceramic bearing surface fracture.

Cite this article: Bone Joint J 2019;101-B:897–901.


The Bone & Joint Journal
Vol. 101-B, Issue 3 | Pages 317 - 324
1 Mar 2019
Moon J Kim Y Hwang K Yang J Ryu J Kim Y

Aims

The present study investigated the five-year interval changes in pseudotumours and measured serum metal ions at long-term follow-up of a previous report of 28 mm diameter metal-on-metal (MoM) total hip arthroplasty (THA).

Patients and Methods

A total of 72 patients (mean age 46.6 years (37 to 55); 43 men, 29 women; 91 hips) who underwent cementless primary MoM THA with a 28 mm modular head were included. The mean follow-up duration was 20.3 years (18 to 24). All patients had CT scans at a mean 15.1 years (13 to 19) after the index operation and subsequent follow-up at a mean of 20.2 years (18 to 24). Pseudotumour volume, type of mass, and new-onset pseudotumours were evaluated using CT scanning. Clinical outcomes were assessed by Harris Hip Score (HHS) and the presence of groin pain. Serum metal ion (cobalt (Co) and chromium (Cr)) levels were measured at the latest follow-up.


The Bone & Joint Journal
Vol. 101-B, Issue 9 | Pages 1168 - 1176
1 Sep 2019
Calder PR McKay JE Timms AJ Roskrow T Fugazzotto S Edel P Goodier WD

Aims

The Precice intramedullary limb-lengthening system has demonstrated significant benefits over external fixation lengthening methods, leading to a paradigm shift in limb lengthening. This study compares outcomes following antegrade and retrograde femoral lengthening in both adolescent and adult patients.

Patients and Methods

A retrospective review of prospectively collected data was undertaken of a consecutive series of 107 femoral lengthening operations in 92 patients. In total, 73 antegrade nails and 34 retrograde nails were inserted. Outcome was assessed by the regenerate healing index (HI), hip and knee range of movement (ROM), and the presence of any complications.


The Bone & Joint Journal
Vol. 101-B, Issue 1_Supple_A | Pages 41 - 45
1 Jan 2019
Jones CW De Martino I D’Apolito R Nocon AA Sculco PK Sculco TP

Aims

Instability continues to be a troublesome complication after total hip arthroplasty (THA). Patient-related risk factors associated with a higher dislocation risk include the preoperative diagnosis, an age of 75 years or older, high body mass index (BMI), a history of alcohol abuse, and neurodegenerative diseases. The goal of this study was to assess the dislocation rate, radiographic outcomes, and complications of patients stratified as high-risk for dislocation who received a dual mobility (DM) bearing in a primary THA at a minimum follow-up of two years.

Materials and Methods

We performed a retrospective review of a consecutive series of DM THA performed between 2010 and 2014 at our institution (Hospital for Special Surgery, New York, New York) by a single, high-volume orthopaedic surgeon employing a single prosthesis design (Anatomic Dual Mobility (ADM) Stryker, Mahwah, New Jersey). Patient medical records and radiographs were reviewed to confirm the type of implant used, to identify any preoperative risk factors for dislocation, and any complications. Radiographic analysis was performed to assess for signs of osteolysis or remodelling of the acetabulum.


The Bone & Joint Journal
Vol. 100-B, Issue 9 | Pages 1182 - 1186
1 Sep 2018
Werner BS Chaoui J Walch G

Aims

Scapular notching is a frequently observed radiographic phenomenon in reverse shoulder arthroplasty (RSA), signifying impingement of components. The purposes of this study were to evaluate the effect of glenoid component size and glenosphere type on impingement-free range of movement (ROM) for extension and internal and external rotation in a virtual RSA model, and to determine the optimal configuration to reduce the incidence of friction-type scapular notching.

Materials and Methods

Preoperative CT scans obtained in 21 patients (three male, 18 female) with primary osteoarthritis were analyzed using modelling software. Two concurrent factors were tested for impingement-free ROM and translation of the centre of rotation: glenosphere diameter (36 mm vs 39 mm) and type (centred, 2 mm inferior eccentric offset, 10° inferior tilt).


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1310 - 1319
1 Oct 2018
Langton DJ Wells SR Joyce TJ Bowsher JG Deehan D Green S Nargol AVF Holland JP

Aims

There are limited published data detailing the volumetric material loss from tapers of conventional metal-on-polyethylene (MoP) total hip arthroplasties (THAs). Our aim was to address this by comparing the taper wear rates measured in an explanted cohort of the widely used Exeter THA with those measured in a group of metal-on-metal (MoM) THAs.

Patients and Methods

We examined an existing retrieval database to identify all Exeter V40 and Universal MoP THAs. Volumetric wear analysis of the taper surfaces was conducted using previously validated methodology. These values were compared with those obtained from a series of MoM THAs using non-parametric statistical methodology. A number of patient and device variables were accounted for using multiple regression modelling.