Aims. The aim of this audit was to assess and improve the completeness and accuracy of the
The National Joint Registry for England, Wales and Northern Ireland
(NJR) has extended its scope to report on hospital, surgeon and
implant performance. Data linkage of the NJR to the London Implant
Retrieval Centre (LIRC) has previously evaluated data quality for
hip primary procedures, but did not assess revision records. We analysed metal-on-metal hip revision procedures performed
between 2003 and 2013. A total of 69 929 revision procedures from
the NJR and 929 revised pairs of components from the LIRC were included.Aims
Methods
Recent publications have drawn attention to the fact that some brands of joint replacement may contain variants which perform significantly worse (or better) than their ‘siblings’. As a result, the
Aims. Despite the increasing numbers of ankle arthroplasties, there are limited studies on their survival and comparisons between different implants. The primary aim of this study was to determine the failure rates of primary ankle arthroplasties commonly used in the UK. Methods. A data linkage study combined
Aims. The Exeter V40 femoral stem is the most implanted stem in the
Aims. The aim of this study was to review the provision of total elbow arthroplasties (TEAs) in England, including the incidence, the characteristics of the patients and the service providers, the types of implant, and the outcomes. Methods. We analyzed the primary TEAs recorded in the
Aims. This study evaluates the association between consultant and hospital volume and the risk of re-revision and 90-day mortality following first-time revision of primary hip arthroplasty for aseptic loosening. Methods. We conducted a cohort study of first-time, single-stage revision hip arthroplasties (RHAs) performed for aseptic loosening and recorded in the
Aims. Metal and ceramic humeral head bearing surfaces are available choices in anatomical shoulder arthroplasties. Wear studies have shown superior performance of ceramic heads, however comparison of clinical outcomes according to bearing surface in total shoulder arthroplasty (TSA) and hemiarthroplasty (HA) is limited. This study aimed to compare the rates of revision and reoperation following metal and ceramic humeral head TSA and HA using data from the
Aims. When a total ankle arthroplasty (TAA) fails, it can be converted to a fusion or a revision arthroplasty. Despite the increasing numbers of TAAs being undertaken, there is little information in the literature about the management of patients undergoing fusion following a failed TAA. The primary aim of this study was to analyze the survival of fusions following a failed TAA using a large dataset from the
Aims. Our main aim was to describe the trend in the comorbidities of patients undergoing elective total hip arthroplasties (THAs) and knee arthroplasties (KAs) between 1 January 2005 and 31 December 2018 in England. Methods. We combined data from the
Aims. The number of revision total ankle arthroplasties (TAAs) which are undertaken is increasing. Few studies have reported the survival after this procedure. The primary aim of this study was to analyze the survival of revision ankle arthroplasties using large datasets. Secondary aims were to summarize the demographics of the patients, the indications for revision TAA, further operations, and predictors of survival. Methods. The study combined data from the
Aims. The aim of this study was to identify variables associated with time to revision, demographic details associated with revision indication, and type of prosthesis employed, and to describe the survival of hinge knee arthroplasty (HKA) when used for first-time knee revision surgery and factors that were associated with re-revision. Methods. Patient demographic details, BMI, American Society of Anesthesiologists (ASA) grade, indication for revision, surgical approach, surgeon grade, implant type (fixed and rotating), time of revision from primary implantation, and re-revision if undertaken were obtained from the
Aims. A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database. Methods. Data from the
Aims. Total hip arthroplasties (THAs) are performed by surgeons at various stages in training with varying levels of supervision, but we do not know if this is safe practice with comparable outcomes to consultant-performed THA. Our aim was to examine the association between surgeon grade, the senior supervision of trainees, and the risk of revision following THA. Methods. We performed an observational study using
Aims. The aim of this study was to capture 12-month outcomes from a representative multicentre cohort of patients undergoing total ankle arthroplasty (TAA), describe the pattern of patient-reported outcome measures (PROMs) at 12 months, and identify predictors of these outcome measures. Methods. Patients listed for a primary TAA at 19 NHS hospitals between February 2016 and October 2017 were eligible. PROMs data were collected preoperatively and at six and 12 months including: Manchester-Oxford Foot and Ankle Questionnaire (MOXFQ (foot and ankle)) and the EuroQol five-dimension five-level questionnaire (EQ-5D-5L). Radiological pre- and postoperative data included Kellgren-Lawrence score and implant position measurement. This was supplemented by data from the
Aims. The main objective of this study was to examine whether the Oxford Shoulder Score (OSS) demonstrated floor or ceiling effects when used to measure outcomes following shoulder arthroplasty in a large national cohort. Secondary objectives were to assess its pain and function subscales, and to identify independent predictors for patients achieving a postoperative ceiling score following shoulder arthroplasty. Methods. Secondary database analysis of the
Aims. The aim of this study was to determine if uncemented acetabular polyethylene (PE) liner geometry, and lip size, influenced the risk of revision for instability or loosening. Methods. A total of 202,511 primary total hip arthroplasties (THAs) with uncemented acetabular components were identified from the
Aims. To determine if primary cemented acetabular component geometry (long posterior wall (LPW), hooded, or offset reorientating) influences the risk of revision total hip arthroplasty (THA) for instability or loosening. Methods. The
Aims. The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs). Methods. We undertook a registry study combining the
Aims. To determine mortality risk after first revision total hip arthroplasty (THA) for periprosthetic femoral fracture (PFF), and to compare this to mortality risk after primary and first revision THA for other common indications. Methods. The study cohort consisted of THAs recorded in the
Aims. The aim of this study was to estimate the 90-day risk of revision for periprosthetic femoral fracture associated with design features of cementless femoral stems, and to investigate the effect of a collar on this risk using a biomechanical in vitro model. Materials and Methods. A total of 337 647 primary total hip arthroplasties (THAs) from the United Kingdom
Aims. Ceramic-on-ceramic (CoC) bearings in total hip arthroplasty (THA)
are commonly used, but concerns exist regarding ceramic fracture.
This study aims to report the risk of revision for fracture of modern
CoC bearings and identify factors that might influence this risk,
using data from the
Unicompartmental knee arthroplasty (UKA) has
advantages over total knee arthroplasty but
Cite this article:
The importance of registries has been brought into focus by recent UK national reports focusing on implant (Cumberlege) and surgeon (Paterson) performance. National arthroplasty registries provide real-time, real-world information about implant, hospital, and surgeon performance and allow case identification in the event of product recall or adverse surgical outcomes. They are a valuable resource for research and service improvement given the volume of data recorded and the longitunidal nature of data collection. This review discusses the current value of registry data as it relates to both clinical practice and research. Cite this article:
Periprosthetic joint infection (PJI) remains an extremely challenging complication. We have focused on this issue more over the last decade than previously, but there are still many unanswered questions. We now have a workable definition that everyone should align to, but we need to continue to focus on identifying the organisms involved. Surgical strategies are evolving and care is becoming more patient-centred. There are some good studies under way. There are, however, still numerous problems to resolve, and the challenge of PJI remains a major one for the orthopaedic community. This annotation provides some up-to-date thoughts about where we are, and the way forward. There is still scope for plenty of research in this area. Cite this article:
The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs). At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays.Aims
Methods
Unicompartmental knee arthroplasty (UKA) has
numerous advantages over total knee arthroplasty (TKA) and one disadvantage,
the higher revision rate. The best way to minimize the revision
rate is for surgeons to use UKA for at least 20% of their knee arthroplasties.
To achieve this, they need to learn and apply the appropriate indications
and techniques. This would decrease the revision rate and increase
the number of UKAs which were implanted, which would save money
and patients would benefit from improved outcomes over their lifetime. Cite this article:
Aims. This study describes the variation in the annual volumes of revision hip arthroplasty (RHA) undertaken by consultant surgeons nationally, and the rate of accrual of RHA and corresponding primary hip arthroplasty (PHA) volume for new consultants entering practice. Methods.
The debate whether to use cemented or uncemented
components in primary total hip replacement (THR) has not yet been
considered with reference to the cost implications to the National
Health Service. We obtained the number of cemented and uncemented components
implanted in 2009 from the National Joint Registry for England and
Wales. The cost of each component was established. The initial financial
saving if all were cemented was then calculated. Subsequently the
five-year rates of revision for each type of component were reviewed
and the predicted number of revisions at five years for the actual
components used was compared with the predicted number of revisions
for a cemented THR. This was then multiplied by the mean cost of
revision surgery to provide an indication of the savings over the
first five years if all primary THRs were cemented. The saving at primary THR was calculated to be £10 million with
an additional saving during the first five years of between £5 million
and £8.5 million. The use of cemented components in routine primary
THR in the NHS as a whole can be justified on a financial level
but we recognise individual patient factors must be considered when deciding
which components to use.
Aims. To determine the trajectories of patient reported pain and functional disability over five years following total hip arthroplasty (THA) or total knee arthroplasty (TKA). Methods. A prospective, longitudinal cohort sub-study within the
Aims. Debate remains whether the patella should be resurfaced during total knee replacement (TKR). For non-resurfaced TKRs, we estimated what the revision rate would have been if the patella had been resurfaced, and examined the risk of re-revision following secondary patellar resurfacing. Methods. A retrospective observational study of the
Aims. We report the medium-term outcomes of a consecutive series of 118 Zenith total ankle arthroplasties (TAAs) from a single, non-designer centre. Methods. Between December 2010 and May 2016, 118 consecutive Zenith prostheses were implanted in 114 patients. Demographic, clinical, and patient-reported outcome measures (PROMs) data were collected. The endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan-Meier survival curves were generated with 95% confidence intervals (CIs) and the rate of failure calculated for each year. Results. Eight patients (ten ankles) died during follow-up, but none required revision. Of the surviving 106 patients (108 ankles: rheumatoid arthritis (RA), n = 15; osteoarthritis (OA), n = 93), 38 were women and 68 were men, with a mean age of 68.2 years (48 to 86) at the time of surgery. Mean follow-up was 5.1 years (2.1 to 9.0). A total of ten implants failed (8.5%), thus requiring revision. The implant survival at seven years, using revision as an endpoint, was 88.2% (95% CI 100% to 72.9%). There was a mean improvement in Manchester-Oxford Foot and Ankle Questionnaire (MOXFQ) from 85.0 to 32.8 and visual analogue scale (VAS) scores from 7.0 to 3.2, and overall satisfaction was 89%. The three commonest complications were malleolar fracture (14.4%, n = 17), wound healing (13.6%, n = 16), and superficial infection (12.7%, n = 15). The commonest reason for revision was aseptic loosening. No patients in our study were revised for deep infection. Conclusion. Our results show that Zenith survival rates are comparable with those in the literature for other implants and in the
Aims. The rate of dislocation when traditional single bearing implants are used in revision total hip arthroplasty (THA) has been reported to be between 8% and 10%. The use of dual mobility bearings can reduce this risk to between 0.5% and 2%. Dual mobility bearings are more expensive, and it is not clear if the additional clinical benefits constitute value for money for the payers. We aimed to estimate the cost-effectiveness of dual mobility compared with single bearings for patients undergoing revision THA. Methods. We developed a Markov model to estimate the expected cost and benefits of dual mobility compared with single bearing implants in patients undergoing revision THA. The rates of revision and further revision were calculated from the
Aims. The aim of this study was to investigate whether the use of antibiotic-loaded bone cement influenced the risk of revision surgery after primary total hip arthroplasty (THA) for osteoarthritis. Methods. The study involved data collected by the
Aims. We present the results, in terms of survival, clinical outcome, and radiological appearance at 20 years, in a cohort of 225 cemented Exeter Universal femoral components (Stryker, Newbury, UK) implanted in 207 patients, at a district general hospital. Methods. All patients in this study had a total hip arthroplasty (THA) using an Exeter Universal femoral component with a cemented (n = 215) or cementless (n = 10) acetabular component. Clinical and radiological data were collected prospectively at one year, five years, and every five years thereafter. Patients lost to radiological and clinical follow-up (five) were cross-referenced with
Aims. We compared implant and patient survival following intraoperative periprosthetic femoral fractures (IOPFFs) during primary total hip arthroplasty (THA) with matched controls. Patients and Methods. This retrospective cohort study compared 4831 hips with IOPFF and 48 154 propensity score matched primary THAs without IOPFF implanted between 2004 and 2016, which had been recorded on a
Aims. The aim of this study was to describe temporal trends and survivorship of total hip arthroplasty (THA) in very young patients, aged ≤ 20 years. Patients and Methods. A descriptive observational study was undertaken using data from the
Aims. The aim of this study was to determine whether patients with
metal-on-metal (MoM) arthroplasties of the hip have an increased
risk of cardiac failure compared with those with alternative types
of arthroplasties (non-MoM). Patients and Methods. A linkage study between the
Aim. There has been a significant reduction in unicompartmental knee arthroplasty (UKA) procedures recorded in Australia. This follows several
Aims. Antibiotic-loaded bone cements (ALBCs) may offer early protection against the formation of bacterial biofilm after joint arthroplasty. Use in hip arthroplasty is widely accepted, but there is a lack of evidence in total knee arthroplasty (TKA). The objective of this study was to evaluate the use of ALBC in a large population of TKA patients. Materials and Methods. Data from the
Aims. The aim of this study was to identify the effect of the manufacturing characteristics of polyethylene acetabular liners on the survival of cementless and hybrid total hip arthroplasty (THA). Methods. Prospective cohort study using linked
Aims. We performed a systematic review of the current literature regarding
the outcomes of unconstrained metacarpophalangeal joint (MCPJ) arthroplasty. Materials and Methods. We initially identified 1305 studies, and 406 were found to be
duplicates. After exclusion criteria were applied, seven studies
were included. Outcomes extracted included pre- and post-operative
pain visual analogue scores, range of movement (ROM), strength of
pinch and grip, satisfaction and patient reported outcome measures
(PROMs). Clinical and radiological complications were recorded.
The results are presented in three groups based on the design of
the arthroplasty and the aetiology (pyrocarbon-osteoarthritis (pyro-OA),
pyrocarbon-inflammatory arthritis (pyro-IA), metal-on-polyethylene
(MoP)). Results. Results show that pyrocarbon implants provide an 85% reduction
in pain, 144% increase of pinch grip and 13° improvements in ROM
for both OA and IA combined. Patients receiving MoP arthroplasties
had a reduction in pinch strength. Satisfaction rates were 91% and
92% for pyrocarbon-OA and pyrocarbon-IA groups, respectively. There were
nine failures in 87 joints (10.3%) over a mean follow-up of 5.5
years (1.0 to 14.3) for pyro-OA. There were 18 failures in 149 joints
(12.1%) over a mean period of 6.6 years (1.0 to 16.0) for pyro-IA.
Meta-analysis was not possible due to the heterogeneity of the studies
and the limited presentation of data. Conclusion. We would recommend prospective data collection for small joint
arthroplasties of the hand consisting of PROMs that would allow
clinicians to come to stronger conclusions about the impact on function
of replacing the MCPJs. A
The National Institute for Clinical Excellence (NICE) produces recommendations on appropriate treatment within the National Health Service (NHS) in England and Wales. The NICE guidelines on prophylaxis for venous thromboembolism in orthopaedic surgery recommend that all patients be offered a low molecular weight heparin (LMWH). The linked hospital episode statistics of 219 602 patients were examined to determine the rates of complications following lower limb arthroplasty for the 12-month periods prior to and following the publication of these guidelines. These were compared with data from the
The effective capture of outcome measures in
the healthcare setting can be traced back to Florence Nightingale’s
investigation of the in-patient mortality of soldiers wounded in
the Crimean war in the 1850s. Only relatively recently has the formalised collection of outcomes
data into Registries been recognised as valuable in itself. With the advent of surgeon league tables and a move towards value
based health care, individuals are being driven to collect, store
and interpret data. Following the success of the
Recent guidance recommends the use of a well-proven
cemented femoral stem for hemiarthroplasty in the management of
fractures of the femoral neck, and the Exeter Trauma Stem (ETS)
has been suggested as an example of such an implant. The design
of this stem was based on the well-proven Exeter Total Hip Replacement
stem (ETHRS). This study assessed the surface finish of the ETS
in comparison with the ETHRS. Two ETSs and two ETHRSs were examined
using a profilometer with a precision of 1 nm and compared with
an explanted Exeter Matt stem. The mean roughness average (RA) of
the ETSs was approximately ten times higher than that of the ETHRSs (0.235 μm
(0.095 to 0.452) versus 0.025 μm (0.011 to 0.059);
p <
0.001). The historical Exeter Matt stem roughness measured
a mean RA of 0.973 μm (0.658 to 1.159). The change of the polished
Exeter stem to a matt surface finish in 1976 resulted in a high
stem failure rate. We do not yet know whether the surface differences
between ETS and ETHRS will be clinically significant. We propose
the inclusion of hemiarthroplasty stems in
Aims . The primary aim of this independent prospective randomised trial
was to compare serum metal ion levels for ceramic-on-metal (CoM)
and metal-on-metal (MoM) bearing surfaces in total hip arthroplasty
(THA). Our one-year results demonstrated elevation in metal ion
levels above baseline with no significant difference between the
CoM and MoM groups. This paper reviews the five-year data. Patients and Methods. The implants used in each patient differed only in respect to
the type of femoral head (ceramic or metal). At five-year follow-up
of the 83 enrolled patients, data from 67 (36 CoM, 31 MoM) was available
for comparison. Results. The mean serum cobalt (Co) and chromium (Cr) ion levels remained
above baseline in both groups (CoM: Co 1.16 μg/l (0.41 to 14.67),
Cr 1.05 μg/l (0.16 to 12.58); MoM: Co 2.93 μg/l (0.35 to 30.29),
Cr 1.85 μg/l (0.36 to 17.00)) but the increase was significantly
less in the CoM cohort (Co difference p = 0.001, Cr difference p
= 0.002). These medium-term results, coupled with lower revision
rates from
There is widespread concern regarding the incidence of adverse soft-tissue reactions after metal-on-metal (MoM) hip replacement. Recent