The STRYDE nail is an evolution of the PRECICE Intramedullary Limb Lengthening System, with unique features regarding its composition. It is designed for load bearing throughout treatment in order to improve patient experience and outcomes and allow for simultaneous bilateral lower limb lengthening. The literature published to date is limited regarding outcomes and potential problems. We report on our early experience and raise awareness for the potential of adverse effects from this device. This is a retrospective review of prospective data collected on all patients treated in our institution using this implant. We report the demographics, nail accuracy, reliability, consolidation index, and cases where concerning clinical and radiological findings were encountered. There were 14 STRYDE nails implanted in nine patients (three male and six female) between June 2019 and September 2020. Mean age at surgery was 33 years (14 to 65). Five patients underwent bilateral lengthening (two femoral and three tibial) and four patients unilateral femoral lengthening for multiple aetiologies.Aims
Methods
Conventional growing rods are the most commonly
used distraction-based devices in the treatment of progressive early-onset
scoliosis. This technique requires repeated lengthenings with the
patient anaesthetised in the operating theatre. We describe the
outcomes and complications of using a non-invasive magnetically
controlled growing rod (MCGR) in children with early-onset scoliosis.
Lengthening is performed on an outpatient basis using an external remote
control with the patient awake. Between November 2009 and March 2011, 34 children with a mean
age of eight years (5 to 12) underwent treatment. The mean length
of follow-up was 15 months (12 to 18). In total, 22 children were
treated with dual rod constructs and 12 with a single rod. The mean
number of distractions per patient was 4.8 (3 to 6). The mean pre-operative
Cobb angle was 69° (46° to 108°); this was corrected to a mean 47°
(28° to 91°) post-operatively. The mean Cobb angle at final review
was 41° (27° to 86°). The mean pre-operative distance from T1 to
S1 was 304 mm (243 to 380) and increased to 335 mm (253 to 400)
in the immediate post-operative period. At final review the mean distance
from T1 to S1 had increased to 348 mm (260 to 420). Two patients developed a superficial wound infection and a further
two patients in the single rod group developed a loss of distraction.
In the dual rod group, one patient had pull-out of a hook and one
developed prominent metalwork. Two patients had a rod breakage;
one patient in the single rod group and one patient in the dual
rod group. Our
A review is presented of
The Acclaim total elbow replacement is a modular system which allows implantation in both unlinked and linked modes. The results of the use of this implant in primary total elbow replacement in 36 patients, operated on between July 2000 and August 2002, are presented at a mean follow-up of 36 months (24 to 49). Only one patient did not have good relief of pain, but all had improved movement and function. No implant showed clinical or radiological loosening, although one had a lucent area in three of seven humeral zones. The short-term results of the Acclaim total elbow replacement are encouraging. However, 11 patients (30.5%) suffered an intra-operative fracture of the humeral condyle. This did not affect the outcome, or the requirement for further surgery, except in one case where the fracture failed to unite. This problem has hopefully been addressed by redesigning the humeral resection guide. Other complications included three cases of ulnar neuropathy (8.3%) and one of deep infection (2.8%).
Autologous chondrocyte implantation (ACI) has been used most commonly as a treatment for cartilage defects in the knee and there are few studies of its use in other joints. We describe ten patients with an osteochondral lesion of the talus who underwent ACI using cartilage taken from the knee and were prospectively reviewed with a mean follow-up of 23 months. In nine patients the satisfaction score was ‘pleased’ or ‘extremely pleased’, which was sustained at four years. The mean Mazur ankle score increased by 23 points at a mean follow-up of 23 months. The Lysholm knee score returned to the pre-operative level at one year in three patients, with the remaining seven showing a reduction of 15% at 12 months, suggesting donor-site morbidity. Nine patients underwent arthroscopic examination at one year and all were shown to have filled defects and stable cartilage. Biopsies taken from graft sites showed mostly fibrocartilage with some hyaline cartilage. The short-term results of ACI for osteochondral lesions of the talus are good despite some morbidity at the donor site.
The aim of this study was to compare the incidence of aseptic
loosening after the use of a cemented acetabular component and a
Trabecular Metal (TM) acetabular component (Zimmer Inc., Warsaw,
Indiana) at acetabular revision with bone impaction grafting. A total of 42 patients were included in the study. Patients were
randomised to receive an all- polyethylene cemented acetabular component
(n = 19) or a TM component (n = 23). Radiostereometric analysis
and conventional radiographic examinations were performed regularly
up to two years post-operatively or until further revision.Aims
Patients and Methods
We report the two- to four-year results following the insertion of the Leeds-Keio prosthetic ligament for chronic anterior cruciate deficiency. Virtually all the 20 patients were less disabled by instability, but objective results were good or excellent in only two-thirds and under anaesthesia the pivot shift sign was still positive in half. Arthroscopic and histological assessment in 16 patients failed to show the development of a functional neoligament, and the common appearance of a synovitic reaction to polyester particles gave concern.
We describe the experience with the first consecutive 230 Birmingham hip resurfacings at our centre. At a mean follow-up of three years (25 to 52 months) survivorship was 99.14% with revision in one patient for a loose acetabular component and one death from unrelated causes. One patient developed a fracture of the femoral neck at six weeks which united unremarkably after a period of non-weight-bearing. The Harris hip score improved from a mean of 62.54 (8 to 92) to 97.74 (61 to 100). The mean flexion improved from 91.52° (25 to 140) to 110.41° (80 to 145). Most patients (97%) considered the outcome to be good or excellent. Our preliminary experience with this implant is encouraging and the results are superior to the earlier generation of resurfacings for the same length of follow-up.
This is a retrospective review of the results of the Acclaim total elbow replacement in 11 older patients aged ≥ 65 years with primary osteoarthritis of the elbow, with a mean follow-up of 57.6 months (30 to 86.4). Significant reductions in pain and improvement in range of movement and function were recorded. Radiological review revealed two patients with 1 mm lucencies in a single zone, and one patient with 1 mm lucencies in two zones. No components required revision. There were no deep infections, dislocations or mechanical failures. Complications included one intra-operative medial condylar fracture and one post-operative transient ulnar neuropathy, which resolved. This study demonstrates that the Acclaim prosthesis provides good symptomatic relief and improvement of function in patients with primary osteoarthritis, with low rates of loosening or other complications. This prosthesis can therefore be considered for patients aged ≥ 65 years with primary osteoarthritis of the elbow.
We have carried out a retrospective review of 11 Souter-Strathclyde primary total elbow arthroplasties in ten patients with osteoarthritis, over a period of nine years. The diagnosis was primary osteoarthritis in nine elbows and post-traumatic arthritis in two. The mean follow-up was 68 months (15 to 117). Although no patient was symptomatic, radiological review revealed evidence of loosening affecting three humeral and two ulnar components, one of which subsequently failed and was revised at 97 months. There were no dislocations, deep infections or mechanical failures. Complications included two superficial wound infections and two neurapraxias of the ulnar nerve which resolved. This study shows that the unlinked Souter-Strathclyde total elbow arthroplasty can be considered for patients with osteoarthritis and gives good symptomatic relief and improvement in function.
We reviewed our initial seven-year experience
with a non-invasive extendible prosthesis in 34 children with primary bone
tumours. The distal femur was replaced in 25 cases, total femur
in five, proximal femur in one and proximal tibia in three. The
mean follow-up was 44Â months (15 to 86) and 27 patients (79%) remain
alive. The prostheses were lengthened by an electromagnetic induction
mechanism in an outpatient setting and a mean extension of 32 mm
(4 to 80) was achieved without anaesthesia. There were lengthening
complications in two children: failed lengthening in one and the
formation of scar tissue in the other. Deep infection developed
in six patients (18%) and local recurrence in three. A total of
11 patients required further surgery to the leg. Amputation was
necessary in five patients (20%) and a two-stage revision in another.
There were no cases of loosening, but two patients had implant breakage
and required revision. The mean Musculoskeletal Tumor Society functional
score was 85% (60% to 100%) at last known follow-up. These early
results demonstrate that the non-invasive extendible prosthesis
allows successful lengthening without surgical intervention, but
the high incidence of infection is a cause for concern.
Fifteen patients with rheumatoid arthritis had 19 elbow arthroplasties with the Wadsworth type of surface replacement prosthesis. After a follow-up period averaging 30 months, 11 patients with 15 elbow arthroplasties were entirely satisfied with their freedom from pain and range of movement. Radiologically, however, the humeral component was loose in 10 of the 19 elbows and the ulnar component in 5. Two patients had reoperation, one to remove a prosthesis for early deep infection and one to exchange a humeral component which was loose. The risk of mechanical loosening is reduced by accurate positioning of the humeral component, but there is a high potential for failure. Changes in the design of the prosthesis and better instrumentation for alignment of the components are desirable. Prosthetic replacement of the elbow should still be regarded as experimental.
The introduction of a trabecular tantalum rod
has been proposed for the management of early-stage osteonecrosis of
the femoral head but serves as a single-point of support of the
necrotic lesion. We describe a technique using two or three 4.2
mm (or later 4.7 mm) tantalum pegs for the prevention of collapse
of the necrotic lesion. We prospectively studied 21 patients (26
hips) with non-traumatic osteonecrosis of the femoral head treated
in this manner. Of these, 21 patients (24 hips) were available for
radiological and clinical evaluation at a mean follow-up of 46 months
(18 to 67). Radiological assessment showed that only eight hips
deteriorated according to the Association Research Circulation Osseous
classification, and four hips according to the Classification of
the Japanese Investigation Committee of Health and Welfare. Functional
improvement was obtained with an improvement in the mean Harris
hip score from 65.2 (33.67 to 95) to 88.1 (51.72 to 100), the mean
Merle D’Aubigné-Postel score from 13 (6 to 18) to 16 (11 to 18),
a mean visual analogue score for pain from 5.2 (0 to 9.5) to 2.6
(0 to 7), and the mean Short-Form 36 score from 80.4 (56.8 to 107.1)
to 92.4 (67.5 to 115.7). Of these 24 hips followed for a minimum
of 18Â months, three were considered as failures at the final follow-up,
having required total hip replacement. One of the hips without full
follow-up was also considered to be a failure. In more than two-thirds
of the surviving hips a satisfactory clinical outcome was achieved
with promising radiological findings. The estimated mean implant
survival was 60 months (95% confidence interval 53.7 to 66.3).
We treated 39 knees with chronic deficiency of the anterior cruciate ligament by reconstruction using the ABC carbon and polyester prosthetic ligament; 31 (79.5%) were reviewed at an average follow-up of 34 months. There had been four complete failures requiring revision. The remaining 27 were studied in detail. On the Lysholm rating, only 11 knees (41%) had good results with a score of over 76. The mean anterior drawer movement was reduced from 7.6 mm before operation to 5.8 mm at review. The mean difference from the opposite uninjured knee was 3.9 mm before operation, 1 mm (in 21 patients) at mean follow-up of 7.4 months and 2.5 mm (in 27 patients) at 34 months, indicating progressive loss of effect. In our opinion the results are unsatisfactory: we do not recommend the use of this prosthetic ligament.
We studied 24 children (40 feet) to demonstrate that a physiotherapist-delivered Ponseti service is as successful as a medically-led programme in obtaining correction of an idiopathic congenital talipes equinovarus deformity. The median Pirani score at the start of treatment was 5.5 (mean 4.75; 2 to 6). A Pirani score of ≥5 predicted the need for tenotomy (p <
0.01). Of the 40 feet studied, 39 (97.5%) achieved correction of deformity. The remaining foot required surgical correction. A total of 25 (62.5%) of the feet underwent an Achilles tenotomy, which was performed by a surgeon in the physiotherapy clinic. There was full compliance with the foot abduction orthoses in 36 (90%) feet. Continuity of care was assured, as one practitioner was responsible for all patient contact. This was rated highly by the patient satisfaction survey. We believe that the Ponseti technique is suitable for use by non-medical personnel, but a holistic approach and good continuity of care are essential to the success of the programme.
This paper describes the preliminary results of a proximal tibial endoprosthesis which spares the knee joint and enables retention of the natural articulation by replacing part of the tibial metaphysis and diaphysis. In eight patients who had a primary malignant bone tumour of the proximal tibia, the distal stem, which had a hydroxyapatite-coated collar to improve fixation, was cemented into the medullary canal. The proximal end had hydroxyapatite-coated extracortical plates which were secured to the remaining proximal tibial metaphysis using cortical screws. The mean age of the patients at operation was 28.9 years (8 to 43) and the mean follow-up was for 35 months (4 to 48). The mean Musculoskeletal Tumour Society score was 79% (57% to 90%), the mean Oxford Knee score was 40 points of 48 (36 to 46) and the mean knee flexion was 112° (100° to 120°). In one patient, revision to a below-knee amputation through the prosthesis was required because of recurrence of the tumour. Another patient sustained a periprosthetic fracture which healed with a painful malunion. This was revised to a further endoprosthesis which replaced the knee. In the remaining six patients the prosthesis allowed preservation of the knee joint with good function and no early evidence of loosening. Further follow-up is required to assess the longevity of these prostheses.
Secure fixation of displaced proximal fractures of the humerus is a challenging problem. A total of 32 patients with acutely displaced three- or four-part proximal fractures of the humerus were treated by open reduction and internal fixation using the proximal humeral internal locking system (PHILOS) plate. There were 23 women and nine men with a mean age of 59.9 years (18 to 87). Data were collected prospectively and the outcomes were assessed using the Constant score. The mean follow-up was for 11 months (3 to 24). In 31 patients (97%) the fracture united clinically and radiologically at a mean of 10 weeks (8 to 24). The mean Constant score at final review was 66.5 (30 to 92). There was no significant difference in outcome when comparing patients aged more than 60 years (18 patients) with those aged less than 60 years (14 patients) ( This plate provides an alternative method of fixation for fractures of the proximal humerus. It provides a stable fixation in young patients with good-quality bone sufficient to permit early mobilisation. Failure of the screws to maintain fixation in the elderly remains a problem.
Aims. There is a lack of long-term data for minimally invasive acromioclavicular (AC) joint repair. Furthermore, it is not clear if good
We carried out a prospective investigation into
the radiological outcomes of uncemented Oxford medial compartment
unicondylar replacement in 220 consecutive patients (231Â knees)
performed in a single centre with a minimum two-year follow-up.
The functional outcomes using the mean Oxford knee score and the
mean high-activity arthroplasty score were significantly improved
over the pre-operative scores (p <
0.001). There were 196Â patients
with a two-year radiological examination performed under fluoroscopic
guidance, aiming to provide images acceptable for analysis of the
bone–implant interface. Of the six tibial zones examined on each
knee on the anteroposterior radiograph, only three had a partial
radiolucent line. All were in the medial aspect of the tibial base plate
(zone 1) and all measured <
1Â mm. All of these patients were
asymptomatic. There were no radiolucent lines seen around the femoral
component or on the lateral view. There was one revision for loosening
at one year due to initial inadequate seating of the tibial component.
These results confirm that the early uncemented Oxford medial unicompartmental
compartmental knee replacements were reliable and the incidence
of radiolucent lines was significantly decreased compared with the
reported results of cemented versions of this implant. These independent
results confirm those of the designing centre.
We report the kinematic and early clinical results
of a patient- and observer-blinded randomised controlled trial in which
CT scans were used to compare potential impingement-free range of
movement (ROM) and acetabular component cover between patients treated
with either the navigated ‘femur-first’ total hip arthroplasty (THA) method
(n = 66; male/female 29/37, mean age 62.5 years; 50 to 74) or conventional
THA (n = 69; male/female 35/34, mean age 62.9 years; 50 to 75).
The Hip Osteoarthritis Outcome Score, the Harris hip score, the
Euro-Qol-5D and the Mancuso THA patient expectations score were
assessed at six weeks, six months and one year after surgery. A
total of 48 of the patients (84%) in the navigated ‘femur-first’
group and 43 (65%) in the conventional group reached all the desirable
potential ROM boundaries without prosthetic impingement for activities
of daily living (ADL) in flexion, extension, abduction, adduction
and rotation (p = 0.016). Acetabular component cover and surface
contact with the host bone were >
87% in both groups. There was
a significant difference between the navigated and the conventional
groups’ Harris hip scores six weeks after surgery (p = 0.010). There
were no significant differences with respect to any clinical outcome
at six months and one year of follow-up. The navigated ‘femur-first’
technique improves the potential ROM for ADL without prosthetic
impingement, although there was no observed clinical difference
between the two treatment groups. Cite this article:
Aims. Dual-mobility acetabular components (DMCs) have improved total hip arthroplasty (THA) stability in femoral neck fractures (FNFs). In osteoarthritis, the direct anterior approach (DAA) has been promoted for improving
Aims. The aim of this study was to determine whether
Reverse total shoulder arthroplasty (rTSA) can be used in complex cases when the glenoid requires reconstruction. In this study, a baseplate with composite bone autograft and a central trabecular titanium peg was implanted, and its migration was assessed for two years postoperatively using radiostereometric analysis (RSA). A total of 14 patients who underwent a rTSA with an autograft consented to participate. Of these, 11 had a primary rTSA using humeral head autograft and three had a revision rTSA with autograft harvested from the iliac crest. The mean age of the patients was 66 years (39 to 81). Tantalum beads were implanted in the scapula around the glenoid. RSA imaging (stereographic radiographs) was undertaken immediately postoperatively and at three, six, 12, and 24 months. Analysis was completed using model-based RSA software. Outcomes were collected preoperatively and at two years postoperatively, including the Oxford Shoulder Score, the American Shoulder and Elbow Score, and a visual analogue score for pain. A Constant score was also obtained for the assessment of strength and range of motion.Aims
Methods
The advent of modular porous metal augments has ushered in a new form of treatment for acetabular bone loss. The function of an augment can be seen as reducing the size of a defect or reconstituting the anterosuperior/posteroinferior columns and/or allowing supplementary fixation. Depending on the function of the augment, the surgeon can decide on the sequence of introduction of the hemispherical shell, before or after the augment. Augments should always, however, be used with cement to form a unit with the acetabular component. Given their versatility, augments also allow the use of a hemispherical shell in a position that restores the centre of rotation and biomechanics of the hip. Progressive shedding or the appearance of metal debris is a particular finding with augments and, with other radiological signs of failure, should be recognized on serial radiographs. Mid- to long-term outcomes in studies reporting the use of augments with hemispherical shells in revision total hip arthroplasty have shown rates of survival of > 90%. However, a higher risk of failure has been reported when augments have been used for patients with chronic pelvic discontinuity. Cite this article:
The aim of the LightFix Trial was to evaluate the clinical outcomes for one year after the treatment of impending and completed pathological fractures of the humerus using the IlluminOss System (IS), and to analyze the performance of this device. A total of 81 patients with an impending or completed pathological fracture were enrolled in a multicentre, open label single cohort study and treated with IS. Inclusion criteria were visual analogue scale (VAS) Pain Scores > 60 mm/100 mm and Mirels’ Score ≥ 8. VAS pain, Musculoskeletal Tumor Society (MSTS) Upper Limb Function, and The European Organization for Research and Treatment of Cancer QoL Group Bone Metastases Module (QLQ-BM22) scores were all normalized to 100, and radiographs were obtained at baseline and at 14, 30, 90, 180, and 360 days postoperatively.Aims
Methods
The aim of this study was to compare the longer-term outcomes of operatively and nonoperatively managed patients treated with a removable brace (fixed-angle removable orthosis) or a plaster cast immobilization for an acute ankle fracture. This is a secondary analysis of a multicentre randomized controlled trial comparing adults with an acute ankle fracture, initially managed either by operative or nonoperative care. Patients were randomly allocated to receive either a cast immobilization or a fixed-angle removable orthosis (removable brace). Data were collected on baseline characteristics, ankle function, quality of life, and complications. The Olerud-Molander Ankle Score (OMAS) was the primary outcome which was used to measure the participant’s ankle function. The primary endpoint was at 16 weeks, with longer-term follow-up at 24 weeks and two years.Aims
Methods
This study aimed to analyze the accuracy and errors associated with 3D-printed, patient-specific resection guides (3DP-PSRGs) used for bone tumour resection. We retrospectively reviewed 29 bone tumour resections that used 3DP-PSRGs based on 3D CT and 3D MRI. We evaluated the resection amount errors and resection margin errors relative to the preoperative plans. Guide-fitting errors and guide distortion were evaluated intraoperatively and one month postoperatively, respectively. We categorized each of these error types into three grades (grade 1, < 1 mm; grade 2, 1 to 3 mm; and grade 3, > 3 mm) to evaluate the overall accuracy.Aims
Methods
This study aimed to gather insights from elbow experts using the Delphi method to evaluate the influence of patient characteristics and fracture morphology on the choice between operative and nonoperative treatment for coronoid fractures. A three-round electronic (e-)modified Delphi survey study was performed between March and December 2023. A total of 55 elbow surgeons from Asia, Australia, Europe, and North America participated, with 48 completing all questionnaires (87%). The panellists evaluated the factors identified as important in literature for treatment decision-making, using a Likert scale ranging from "strongly influences me to recommend nonoperative treatment" (1) to "strongly influences me to recommend operative treatment" (5). Factors achieving Likert scores ≤ 2.0 or ≥ 4.0 were deemed influential for treatment recommendation. Stable consensus is defined as an agreement of ≥ 80% in the second and third rounds.Aims
Methods
Chondrosarcoma is the second most common surgically treated primary bone sarcoma. Despite a large number of scientific papers in the literature, there is still significant controversy about diagnostics, treatment of the primary tumour, subtypes, and complications. Therefore, consensus on its day-to-day treatment decisions is needed. In January 2024, the Birmingham Orthopaedic Oncology Meeting (BOOM) attempted to gain global consensus from 300 delegates from over 50 countries. The meeting focused on these critical areas and aimed to generate consensus statements based on evidence amalgamation and expert opinion from diverse geographical regions. In parallel, periprosthetic joint infection (PJI) in oncological reconstructions poses unique challenges due to factors such as adjuvant treatments, large exposures, and the complexity of surgery. The meeting debated two-stage revisions, antibiotic prophylaxis, managing acute PJI in patients undergoing chemotherapy, and defining the best strategies for wound management and allograft reconstruction. The objectives of the meeting extended beyond resolving immediate controversies. It sought to foster global collaboration among specialists attending the meeting, and to encourage future research projects to address unsolved dilemmas. By highlighting areas of disagreement and promoting collaborative research endeavours, this initiative aims to enhance treatment standards and potentially improve outcomes for patients globally. This paper sets out some of the controversies and questions that were debated in the meeting. Cite this article:
Favourable short-term outcomes have been reported following latissimus dorsi tendon transfer for patients with an irreparable subscapularis (SSC) tendon tear. The aim of this study was to investigate the long-term outcomes of this transfer in these patients. This was a retrospective study involving 30 patients with an irreparable SSC tear and those with a SSC tear combined with a reparable supraspinatus tear, who underwent a latissimus dorsi tendon transfer. Clinical scores and active range of motion (aROM), SSC-specific physical examination and the rate of return to work were assessed. Radiological assessment included recording the acromiohumeral distance (AHD), the Hamada grade of cuff tear arthropathy and the integrity of the transferred tendon. Statistical analysis compared preoperative, short-term (two years), and final follow-up at a mean of 8.7 years (7 to 10).Aims
Methods
The Vantage Total Ankle System is a fourth-generation low-profile fixed-bearing implant that has been available since 2016. We aimed to describe our early experience with this implant. This is a single-centre retrospective review of patients who underwent primary total ankle arthroplasty (TAA) with a Vantage implant between November 2017 and February 2020, with a minimum of two years’ follow-up. Four surgeons contributed patients. The primary outcome was reoperation and revision rate of the Vantage implant at two years. Secondary outcomes included radiological alignment, peri-implant complications, and pre- and postoperative patient-reported outcomes.Aims
Methods
The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface. Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up.Aims
Methods
This study aimed to evaluate rasterstereography of the spine as a diagnostic test for adolescent idiopathic soliosis (AIS), and to compare its results with those obtained using a scoliometer. Adolescents suspected of AIS and scheduled for radiographs were included. Rasterstereographic scoliosis angle (SA), maximal vertebral surface rotation (ROT), and angle of trunk rotation (ATR) with a scoliometer were evaluated. The area under the curve (AUC) from receiver operating characteristic (ROC) plots were used to describe the discriminative ability of the SA, ROT, and ATR for scoliosis, defined as a Cobb angle > 10°. Test characteristics (sensitivity and specificity) were reported for the best threshold identified using the Youden method. AUC of SA, ATR, and ROT were compared using the bootstrap test for two correlated ROC curves method.Aims
Methods
The aims of this study were first, to determine if adding fusion to a decompression of the lumbar spine for spinal stenosis decreases the rate of radiological restenosis and/or proximal adjacent level stenosis two years after surgery, and second, to evaluate the change in vertebral slip two years after surgery with and without fusion. The Swedish Spinal Stenosis Study (SSSS) was conducted between 2006 and 2012 at five public and two private hospitals. Six centres participated in this two-year MRI follow-up. We randomized 222 patients with central lumbar spinal stenosis at one or two adjacent levels into two groups, decompression alone and decompression with fusion. The presence or absence of a preoperative spondylolisthesis was noted. A new stenosis on two-year MRI was used as the primary outcome, defined as a dural sac cross-sectional area ≤ 75 mm2 at the operated level (restenosis) and/or at the level above (proximal adjacent level stenosis).Aims
Methods
The aim of this retrospective cohort study was to assess and investigate the safety and efficacy of using a distal tibial osteotomy compared to proximal osteotomy for limb lengthening in children. In this study, there were 59 consecutive tibial lengthening and deformity corrections in 57 children using a circular frame. All were performed or supervised by the senior author between January 2013 and June 2019. A total of 25 who underwent a distal tibial osteotomy were analyzed and compared to a group of 34 who had a standard proximal tibial osteotomy. For each patient, the primary diagnosis, time in frame, complications, and lengthening achieved were recorded. From these data, the frame index was calculated (days/cm) and analyzed.Aims
Methods
Aims. To establish our
A significant reduction in wear at five and ten years was previously reported when comparing Durasul highly cross-linked polyethylene with nitrogen-sterilized Sulene polyethylene in total hip arthroplasty (THA). We investigated whether the improvement observed at the earlier follow-up continued, resulting in decreased osteolysis and revision surgery rates over the second decade. Between January 1999 and December 2001, 90 patients underwent surgery using the same acetabular and femoral components with a 28 mm metallic femoral head and either a Durasul or Sulene liner. A total of 66 hips of this prospective randomized study were available for a minimum follow-up of 20 years. The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Dorr method on digitized radiographs with a software package.Aims
Methods
There has been an increasing use of early operative fixation for scaphoid fractures, despite uncertain evidence. We conducted a meta-analysis to evaluate up-to-date evidence from randomized controlled trials (RCTs), comparing the effectiveness of the operative and nonoperative treatment of undisplaced and minimally displaced (≤ 2 mm displacement) scaphoid fractures. A systematic review of seven databases was performed from the dates of their inception until the end of March 2021 to identify eligible RCTs. Reference lists of the included studies were screened. No language restrictions were applied. The primary outcome was the patient-reported outcome measure of wrist function at 12 months after injury. A meta-analysis was performed for function, pain, range of motion, grip strength, and union. Complications were reported narratively.Aims
Methods
Aims. Spinal deformity surgery carries the risk of neurological injury. Neurophysiological monitoring allows early identification of intraoperative cord injury which enables
Aims. Complex displaced osteoporotic acetabular fractures in the elderly are associated with high levels of morbidity and mortality. Surgical options include either open reduction and internal fixation alone, or combined with total hip arthroplasty (THA). There remains a cohort of severely comorbid patients who are deemed unfit for extensive surgical reconstruction and are treated conservatively. We describe the results of a coned hemipelvis reconstruction and THA inserted via a posterior approach to the hip as the primary treatment for this severely high-risk cohort. Methods. We have prospectively monitored a series of 22 cases (21 patients) with a mean follow-up of 32 months (13 to 59). Results. The mean patient age was 79 years (67 to 87), and the mean ASA score was 3.3 (3 to 5). Three patients had high-energy injuries and 18 had low-energy injuries. All cases were associated fractures (Letournel classification: anterior column posterior hemitransverse, n = 13; associated both column, n = 6; transverse posterior wall, n = 3) with medialization of the femoral head. Mean operative time was 93 minutes (61 to 135). There have been no revisions to date. Of the 21 patients, 20 were full weight-bearing on day 1 postoperatively. Mean length of hospital stay was 12 days (5 to 27). Preoperative mobility status was maintained in 13 patients. At one year, mean Merle d’Aubigné score was 13.1 (10 to 18), mean Oxford Hip Score was 38.5 (24 to 44), mean EuroQol five-dimension five-level (EQ-5D-5L) health score was 68 (30 to 92), and mean EQ-5D-5L index score was 0.68 (0.335 to 0.837); data from 14 patients. Mortality was 9.5% (2/21) at one year. There have been no thromboembolic events, deep infections, or revisions. Conclusion. The coned hemipelvis reconstruction bypasses the fracture, creating an immediately stable construct that allows immediate full weight-bearing. The posterior approach minimizes the operative time and physiological insult in this vulnerable patient population.
Aims. Although good clinical outcomes have been reported for monolithic tapered, fluted, titanium stems (TFTS),
The aim of this prospective study was to assess the long-term clinical, radiological, functional, and quality of life (QoL)-related outcome of patients treated with the synthetic Ligament Advanced Reinforcement System (LARS) device for anterior cruciate ligament (ACL) rupture. A total of 41 patients who underwent ACL reconstruction with the LARS device (mean age 39.8 years (SD 12.1 ); 32% females (n = 13)) were prospectively included between August 2001 and March 2005. MRI scans and radiographs were performed at a median follow-up of 2.0 years (interquartile range (IQR) 1.3 to 3.0; n = 40) and 12.8 years (IQR 12.1 to 13.8; n = 22). Functional and QoL-related outcome was assessed in 29 patients at a median follow-up of 12.8 years (IQR 12.0 to 14.0) and clinically reconfirmed at latest median follow-up of 16.5 years (IQR 15.5 to 17.9). International Knee Documentation Committee (IKDC) and Tegner scores were obtained pre- and postoperatively, and Lysholm score postoperatively only. At latest follow-up, range of motion, knee stability tests, 36-Item Short Form Health Survey (SF-36), and IKDC scores were ascertained. Complications and reoperations during follow-up were documented.Aims
Methods
Aims. Extendible endoprostheses have been available for more than 30
years and have become more sophisticated with time. The latest generation
is ‘non-invasive’ and can be lengthened with an external magnetic
force.
Aims. The ideal management of acute syndesmotic injuries in elite athletes is controversial. Among several treatment methods used to stabilize the syndesmosis and facilitate healing of the ligaments, the use of suture tape (InternalBrace) has previously been described. The purpose of this study was to analyze the functional outcome, including American Orthopaedic Foot & Ankle Society (AOFAS) scores, knee-to-wall measurements, and the time to return to play in days, of unstable syndesmotic injuries treated with the use of the InternalBrace in elite athletes. Methods. Data on a consecutive group of elite athletes who underwent isolated reconstruction of the anterior inferior tibiofibular ligament using the InternalBrace were collected prospectively. Our patient group consisted of 19 elite male athletes with a mean age of 24.5 years (17 to 52). Isolated injuries were seen in 12 patients while associated injuries were found in seven patients (fibular fracture, medial malleolus fracture, anterior talofibular ligament rupture, and posterior malleolus fracture). All patients had a minimum follow-up period of 17 months (mean 27 months (17 to 35)). Results. All patients returned to their pre-injury level of sports activities. One patient developed a delayed union of the medial malleolus. The mean return to play was 62 days (49 to 84) for isolated injuries, while the patients with concomitant injuries returned to play in a mean of 104 days (56 to 196). The AOFAS score returned to 100 postoperatively in all patients. Knee-to-wall measurements were the same as the contralateral side in 18 patients, while one patient lacked 2 cm compared to the contralateral side. Conclusion. This study suggests the use of the InternalBrace in the management of unstable syndesmotic injuries offers an alternative method of stabilization, with good short-term
Aims. To report our experience with trunnion corrosion following metal-on-polyethylene
total hip arthroplasty, in particular to report the spectrum of
presentation and determine the mean time to presentation. Patients and Methods. We report the presenting symptoms and signs, intraoperative findings,
and
The low contact stress patellofemoral replacement consists of a trochlear component and a modular patellar component which has a metal-backed mobile polyethylene bearing. We present the
Alternative alignment concepts, including kinematic and restricted kinematic, have been introduced to help improve clinical outcomes following total knee arthroplasty (TKA). The purpose of this study was to evaluate the clinical results, along with patient satisfaction, following TKA using the concept of restricted kinematic alignment. A total of 121 consecutive TKAs performed between 11 February 2018 to 11 June 2019 with preoperative varus deformity were reviewed at minimum one-year follow-up. Three knees were excluded due to severe preoperative varus deformity greater than 15°, and a further three due to requiring revision surgery, leaving 109 patients and 115 knees to undergo primary TKA using the concept of restricted kinematic alignment with advanced technology. Patients were stratified into three groups based on the preoperative limb varus deformity: Group A with 1° to 5° varus (43 knees); Group B between 6° and 10° varus (56 knees); and Group C with varus greater than 10° (16 knees). This study group was compared with a matched cohort of 115 TKAs and 115 patients using a neutral mechanical alignment target with manual instruments performed from 24 October 2016 to 14 January 2019.Aims
Methods
The purpose was to evaluate early clinical, patient-reported, and radiological outcomes of the scapholunate ligament 360° tenodesis (SL 360) technique for treatment of scapholunate (SL) instability. We studied the results of nine patients (eight males and one female with a mean age of 44.7 years (26 to 55)) who underwent the SL 360 procedure for reducible SL instability between January 2016 and June 2019, and who were identified from retrospective review of electronic medical records. Final follow-up of any kind was a mean of 33.7 months (12.0 to 51.3). Clinical, radiological, and patient-reported outcome data included visual analogue scale (VAS) for pain, Quick Disabilities of the Arm, Shoulder and Hand (QuickDASH), Mayo Wrist Score (MWS), and Patient-Rated Wrist Examination (PRWE). Means were analyzed using paired Aims
Methods
In the last decade, interest in partial knee arthroplasties and bicruciate retaining total knee arthroplasties has increased. In addition, patient-related outcomes and functional results such as range of movement and ambulation may be more promising with less invasive procedures such as bicompartmental arthroplasty (BCA). The purpose of this study is to evaluate clinical and radiological outcomes after a third-generation patellofemoral arthroplasty (PFA) combined with a medial or lateral unicompartmental knee arthroplasty (UKA) at mid- to long-term follow-up. A total of 57 procedures were performed. In 45 cases, a PFA was associated with a medial UKA and, in 12, with a lateral UKA. Patients were followed with validated patient-reported outcome measures (Oxford Knee Score (OKS), EuroQol five-dimension questionnaire (EQ-5D), EuroQoL Visual Analogue Scale (EQ-VAS)), the Knee Society Score (KSS), the Forgotten Joint Score (FJS), and radiological analysis.Aims
Methods