The purpose of the present study was to compare patient-specific instrumentation (PSI) and conventional surgical instrumentation (CSI) for total knee arthroplasty (TKA) in terms of early implant migration, alignment, surgical resources, patient outcomes, and costs. The study was a prospective, randomized controlled trial of 50 patients undergoing TKA. There were 25 patients in each of the PSI and CSI groups. There were 12 male patients in the PSI group and seven male patients in the CSI group. The patients had a mean age of 69.0 years (Aims
Patients and Methods
The aim of this study was to determine whether there is a difference
in the rate of wear between acetabular components positioned within
and outside the ‘safe zones’ of anteversion and inclination angle. We reviewed 100 hips in 94 patients who had undergone primary
total hip arthroplasty (THA) at least ten years previously. Patients
all had the same type of acetabular component with a bearing couple
which consisted of a 28 mm cobalt-chromium head on a highly crosslinked
polyethylene (HXLPE) liner. A supine radiostereometric analysis
(RSA) examination was carried out which acquired anteroposterior
(AP) and lateral paired images. Acetabular component anteversion
and inclination angles were measured as well as total femoral head
penetration, which was divided by the length of implantation to
determine the rate of polyethylene wear.Aims
Patients and Methods
The aim of this study was to evaluate the long-term inducible
displacement of cemented tibial components ten years after total
knee arthroplasty (TKA). A total of 15 patients from a previously reported prospective
trial of fixation using radiostereometric analysis (RSA) were examined
at a mean of 11 years (10 to 11) postoperatively. Longitudinal supine
RSA examinations were acquired at one week, one year, and two years
postoperatively and at final follow-up. Weight-bearing RSA examinations
were also undertaken with the operated lower limb in neutral and
in maximum internal rotation positions. Maximum total point motion
(MTPM) was calculated for the longitudinal and inducible displacement examinations
(supine Aims
Patients and Methods
The purpose of the present study was to examine the long-term
fixation of a cemented fixed-bearing polished titanium tibial baseplate
(Genesis ll). Patients enrolled in a previous two-year prospective trial (n
= 35) were recalled at ten years. Available patients (n = 15) underwent
radiostereometric analysis (RSA) imaging in a supine position using
a conventional RSA protocol. Migration of the tibial component in
all planes was compared between initial and ten-year follow-up.
Outcome scores including the Knee Society Score, Western Ontario
and McMaster Universities Arthritis Index, 12-item Short Form Health
Survey, Forgotten Joint Score, and University of California, Los
Angeles Activity Score were recorded.Aims
Patients and Methods
The purpose of this study was to compare the long-term results
of primary total hip arthroplasty (THA) in young patients using
either a conventional (CPE) or a highly cross-linked (HXLPE) polyethylene
liner in terms of functional outcome, incidence of osteolysis, radiological
wear and rate of revision. We included all patients between the ages of 45 and 65 years
who, between January 2000 and December 2001, had undergone a primary
THA for osteoarthritis at our hospital using a CPE or HXLPE acetabular
liner and a 28 mm cobalt-chrome femoral head. From a total of 160 patients, 158 (177 hips) were available for
review (CPE 89; XLPE 88). The mean age, body mass index (BMI) and
follow-up in each group were: CPE: 56.8 years (46 to 65); 30.7 kg/m2 (19
to 58); 13.2 years (2.1 to 14.7) and HXLPE: 55.6 years (45 to 65);
BMI: 30 kg/m2 (18 to 51); 13.1 years (5.7 to 14.4).Aims
Methods
The purpose of this study was to compare clinical
outcomes of total knee arthroplasty (TKA) after manipulation under
anaesthesia (MUA) for post-operative stiffness with a matched cohort
of TKA patients who did not requre MUA. In total 72 patients (mean age 59.8 years, 42 to 83) who underwent
MUA following TKA were identified from our prospective database
and compared with a matched cohort of patients who had undergone
TKA without subsequent MUA. Patients were evaluated for range of
movement (ROM) and clinical outcome scores (Western Ontario and
McMaster Universities Arthritis Index, Short-Form Health Survey,
and Knee Society Clinical Rating System) at a mean follow-up of
36.4 months (12 to 120). MUA took place at a mean of nine weeks
(5 to 18) after TKA. In patients who required MUA, mean flexion
deformity improved from 10° (0° to 25°) to 4.4° (0° to 15°) (p <
0.001),
and mean range of flexion improved from 79.8° (65° to 95°) to 116°
(80° to 130°) (p <
0.001). There were no statistically significant
differences in ROM or functional outcome scores at three months,
one year, or two years between those who required MUA and those
who did not. There were no complications associated with manipulation At most recent follow-up, patients requiring MUA achieved equivalent
ROM and clinical outcome scores when compared with a matched control
group. While other studies have focused on ROM after manipulation,
the current study adds to current literature by supplementing this
with functional outcome scores. Cite this article:
This was a randomised controlled trial studying
the safety of a new short metaphyseal fixation (SMF) stem. We hypothesised
that it would have similar early clinical results and micromovement
to those of a standard-length tapered Synergy metaphyseal fixation
stem. Using radiostereometric analysis (RSA) we compared the two
stems in 43 patients. A short metaphyseal fixation stem was used
in 22 patients and a Synergy stem in 21 patients. No difference
was found in the clinical outcomes pre- or post-operatively between
groups. RSA showed no significant differences two years post-operatively
in mean micromovement between the two stems (except for varus/valgus tilt
at p = 0.05) (subsidence 0.94 mm ( Cite this article:
We evaluated the outcome of primary total hip replacement (THR) in 3290 patients with the primary diagnosis of osteoarthritis at a minimum follow-up of two years. They were stratified into categories of body mass index (BMI) based on the World Health Organisation classification of obesity. Statistical analysis was carried out to determine if there was a difference in the post-operative Western Ontario and McMaster Universities osteoarthritis index, the Harris hip score and the Short-Form-12 outcome based on the BMI. While the pre- and post-operative scores were lower for the group classified as morbidly obese, the overall change in outcome scores suggested an equal if not greater improvement compared with the non-morbidly obese patients. The overall survivorship and rate of complications were similar in the BMI groups although there was a slightly higher rate of revision for sepsis in the morbidly obese group. Morbid obesity does not affect the post-operative outcome after THR, with the possible exception of a marginally increased rate of infection. Therefore withholding surgery based on the BMI is not justified.
We present a case of early retrieval of an Oxinium femoral head and corresponding polyethylene liner where there was significant surface damage to the head and polyethylene. The implants were retrieved at the time of revision surgery to correct leg-length discrepancy just 48 hours after the primary hip replacement. Appropriate analysis of the retrieved femoral head demonstrated loss of the Oxinium layer with exposure of the underlying substrate and transfer of titanium from the acetabular shell at the time of a reduction of the index total hip replacement. In addition, the level of damage to the polyethylene was extensive despite only 48 hours The purpose of this report is to highlight the care that is required at the time of reduction, especially with these hard femoral counter-faces such as Oxinium. To our knowledge, the damage occurring at the time of reduction has not been previously reported following the retrieval of an otherwise well-functioning hip replacement.
This was a safety study where the hypothesis was that the newer-design CPCS femoral stem would demonstrate similar early clinical results and micromovement to the well-established Exeter stem. Both are collarless, tapered, polished cemented stems, the only difference being a slight lateral to medial taper with the CPCS stem. A total of 34 patients were enrolled in a single-blinded randomised controlled trial in which 17 patients received a dedicated radiostereometric CPCS stem and 17 a radiostereometric Exeter stem. No difference was found in any of the outcome measures pre-operatively or post-operatively between groups. At two years, the mean subsidence for the CPCS stem was nearly half that seen for the Exeter stem (0.77 mm (−0.943 to 1.77) and 1.25 mm (0.719 to 1.625), respectively; p = 0.032). In contrast, the mean internal rotation of the CPCS stem was approximately twice that of the Exeter (1.61° (−1.07° to 4.33°) and 0.59° (0.97° to 1.64°), respectively; p = 0.048). Other migration patterns were not significantly different between the stems. The subtle differences in designs may explain the different patterns of migration. Comparable migration with the Exeter stem suggests that the CPCS design will perform well in the long term.
We identified five (2.3%) fractures of the stem in a series of 219 revision procedures using a cementless, cylindrical, extensively porous-coated, distally-fixed femoral stem. Factors relating to the patients, the implant and the operations were compared with those with intact stems. Finite-element analysis was performed on two of the fractured implants. Factors associated with fracture of the stem were poor proximal bone support (type III–type IV; p = 0.001), a body mass index >
30; (p = 0.014), a smaller diameter of stem (<
13.5 mm; p = 0.007) and the use of an extended trochanteric osteotomy (ETO 4/5: p = 0.028). Finite-element analysis showed that the highest stresses on the stem occurred adjacent to the site of the fracture. The use of a strut graft wired over an extended trochanteric osteotomy in patients lacking proximal femoral cortical support decreased the stresses on the stem by 48%. We recommend the use of a strut allograft in conjunction with an extended trochanteric osteotomy in patients with poor proximal femoral bone stock.