Advertisement for orthosearch.org.uk
Results 1 - 11 of 11
Results per page:
The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 565 - 572
1 May 2019
Teeter MG Marsh JD Howard JL Yuan X Vasarhelyi EM McCalden RW Naudie DDR

Aims

The purpose of the present study was to compare patient-specific instrumentation (PSI) and conventional surgical instrumentation (CSI) for total knee arthroplasty (TKA) in terms of early implant migration, alignment, surgical resources, patient outcomes, and costs.

Patients and Methods

The study was a prospective, randomized controlled trial of 50 patients undergoing TKA. There were 25 patients in each of the PSI and CSI groups. There were 12 male patients in the PSI group and seven male patients in the CSI group. The patients had a mean age of 69.0 years (sd 8.4) in the PSI group and 69.4 years (sd 8.4) in the CSI group. All patients received the same TKA implant. Intraoperative surgical resources and any surgical waste generated were recorded. Patients underwent radiostereometric analysis (RSA) studies to measure femoral and tibial component migration over two years. Outcome measures were recorded pre- and postoperatively. Overall costs were calculated for each group.


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 891 - 897
1 Jul 2018
Teeter MG Lanting BA Naudie DD McCalden RW Howard JL MacDonald SJ

Aims

The aim of this study was to determine whether there is a difference in the rate of wear between acetabular components positioned within and outside the ‘safe zones’ of anteversion and inclination angle.

Patients and Methods

We reviewed 100 hips in 94 patients who had undergone primary total hip arthroplasty (THA) at least ten years previously. Patients all had the same type of acetabular component with a bearing couple which consisted of a 28 mm cobalt-chromium head on a highly crosslinked polyethylene (HXLPE) liner. A supine radiostereometric analysis (RSA) examination was carried out which acquired anteroposterior (AP) and lateral paired images. Acetabular component anteversion and inclination angles were measured as well as total femoral head penetration, which was divided by the length of implantation to determine the rate of polyethylene wear.


The Bone & Joint Journal
Vol. 100-B, Issue 2 | Pages 170 - 175
1 Feb 2018
Lam Tin Cheung K Lanting BA McCalden RW Yuan X MacDonald SJ Naudie DD Teeter MG

Aims

The aim of this study was to evaluate the long-term inducible displacement of cemented tibial components ten years after total knee arthroplasty (TKA).

Patients and Methods

A total of 15 patients from a previously reported prospective trial of fixation using radiostereometric analysis (RSA) were examined at a mean of 11 years (10 to 11) postoperatively. Longitudinal supine RSA examinations were acquired at one week, one year, and two years postoperatively and at final follow-up. Weight-bearing RSA examinations were also undertaken with the operated lower limb in neutral and in maximum internal rotation positions. Maximum total point motion (MTPM) was calculated for the longitudinal and inducible displacement examinations (supine versus standing, standing versus internal rotation, and supine versus standing with internal rotation).


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 616 - 621
1 May 2016
Teeter MG Thoren J Yuan X McCalden RW MacDonald SJ Lanting BA Naudie DDR

Aims

The purpose of the present study was to examine the long-term fixation of a cemented fixed-bearing polished titanium tibial baseplate (Genesis ll).

Patients and Methods

Patients enrolled in a previous two-year prospective trial (n = 35) were recalled at ten years. Available patients (n = 15) underwent radiostereometric analysis (RSA) imaging in a supine position using a conventional RSA protocol. Migration of the tibial component in all planes was compared between initial and ten-year follow-up. Outcome scores including the Knee Society Score, Western Ontario and McMaster Universities Arthritis Index, 12-item Short Form Health Survey, Forgotten Joint Score, and University of California, Los Angeles Activity Score were recorded.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 28 - 32
1 Jan 2016
Hanna SA Somerville L McCalden RW Naudie DD MacDonald SJ

Aims

The purpose of this study was to compare the long-term results of primary total hip arthroplasty (THA) in young patients using either a conventional (CPE) or a highly cross-linked (HXLPE) polyethylene liner in terms of functional outcome, incidence of osteolysis, radiological wear and rate of revision.

Methods

We included all patients between the ages of 45 and 65 years who, between January 2000 and December 2001, had undergone a primary THA for osteoarthritis at our hospital using a CPE or HXLPE acetabular liner and a 28 mm cobalt-chrome femoral head.

From a total of 160 patients, 158 (177 hips) were available for review (CPE 89; XLPE 88). The mean age, body mass index (BMI) and follow-up in each group were: CPE: 56.8 years (46 to 65); 30.7 kg/m2 (19 to 58); 13.2 years (2.1 to 14.7) and HXLPE: 55.6 years (45 to 65); BMI: 30 kg/m2 (18 to 51); 13.1 years (5.7 to 14.4).


The Bone & Joint Journal
Vol. 97-B, Issue 12 | Pages 1640 - 1644
1 Dec 2015
Dzaja I Vasarhelyi EM Lanting BA Naudie DD Howard JL Somerville L McCalden RW MacDonald SJ

The purpose of this study was to compare clinical outcomes of total knee arthroplasty (TKA) after manipulation under anaesthesia (MUA) for post-operative stiffness with a matched cohort of TKA patients who did not requre MUA.

In total 72 patients (mean age 59.8 years, 42 to 83) who underwent MUA following TKA were identified from our prospective database and compared with a matched cohort of patients who had undergone TKA without subsequent MUA. Patients were evaluated for range of movement (ROM) and clinical outcome scores (Western Ontario and McMaster Universities Arthritis Index, Short-Form Health Survey, and Knee Society Clinical Rating System) at a mean follow-up of 36.4 months (12 to 120). MUA took place at a mean of nine weeks (5 to 18) after TKA. In patients who required MUA, mean flexion deformity improved from 10° (0° to 25°) to 4.4° (0° to 15°) (p < 0.001), and mean range of flexion improved from 79.8° (65° to 95°) to 116° (80° to 130°) (p < 0.001). There were no statistically significant differences in ROM or functional outcome scores at three months, one year, or two years between those who required MUA and those who did not. There were no complications associated with manipulation.

At most recent follow-up, patients requiring MUA achieved equivalent ROM and clinical outcome scores when compared with a matched control group. While other studies have focused on ROM after manipulation, the current study adds to current literature by supplementing this with functional outcome scores.

Cite this article: Bone Joint J 2015;97-B:1640–4.


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 595 - 602
1 May 2015
McCalden RW Korczak A Somerville L Yuan X Naudie DD

This was a randomised controlled trial studying the safety of a new short metaphyseal fixation (SMF) stem. We hypothesised that it would have similar early clinical results and micromovement to those of a standard-length tapered Synergy metaphyseal fixation stem. Using radiostereometric analysis (RSA) we compared the two stems in 43 patients. A short metaphyseal fixation stem was used in 22 patients and a Synergy stem in 21 patients. No difference was found in the clinical outcomes pre- or post-operatively between groups. RSA showed no significant differences two years post-operatively in mean micromovement between the two stems (except for varus/valgus tilt at p = 0.05) (subsidence 0.94 mm (sd 1.71) vs 0.32 mm (sd 0.45), p = 0.66; rotation 0.96° (sd 1.49) vs 1.41° (sd 2.95), p = 0.88; and total migration 1.09 mm (sd 1.74) vs 0.73 mm (sd 0.72), p = 0.51). A few stems (four SMF and three Synergy) had initial migration > 1.0 mm but stabilised by three to six months, with the exception of one SMF stem which required revision three years post-operatively. For most stems, total micromovement was very low at two years (subsidence < 0.5 mm, rotation < 1.0°, total migration < 0.5 mm), which was consistent with osseous ingrowth. The small sample makes it difficult to confirm the universal applicability of or elucidate the potential contraindications to the use of this particular new design of stem.

Cite this article: Bone Joint J 2015; 97-B:595–602.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 321 - 325
1 Mar 2011
McCalden RW Charron KD MacDonald SJ Bourne RB Naudie DD

We evaluated the outcome of primary total hip replacement (THR) in 3290 patients with the primary diagnosis of osteoarthritis at a minimum follow-up of two years. They were stratified into categories of body mass index (BMI) based on the World Health Organisation classification of obesity. Statistical analysis was carried out to determine if there was a difference in the post-operative Western Ontario and McMaster Universities osteoarthritis index, the Harris hip score and the Short-Form-12 outcome based on the BMI.

While the pre- and post-operative scores were lower for the group classified as morbidly obese, the overall change in outcome scores suggested an equal if not greater improvement compared with the non-morbidly obese patients. The overall survivorship and rate of complications were similar in the BMI groups although there was a slightly higher rate of revision for sepsis in the morbidly obese group.

Morbid obesity does not affect the post-operative outcome after THR, with the possible exception of a marginally increased rate of infection. Therefore withholding surgery based on the BMI is not justified.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 409 - 413
1 Mar 2011
McCalden RW Charron KD Davidson RD Teeter MG Holdsworth DW

We present a case of early retrieval of an Oxinium femoral head and corresponding polyethylene liner where there was significant surface damage to the head and polyethylene. The implants were retrieved at the time of revision surgery to correct leg-length discrepancy just 48 hours after the primary hip replacement. Appropriate analysis of the retrieved femoral head demonstrated loss of the Oxinium layer with exposure of the underlying substrate and transfer of titanium from the acetabular shell at the time of a reduction of the index total hip replacement. In addition, the level of damage to the polyethylene was extensive despite only 48 hours in situ.

The purpose of this report is to highlight the care that is required at the time of reduction, especially with these hard femoral counter-faces such as Oxinium. To our knowledge, the damage occurring at the time of reduction has not been previously reported following the retrieval of an otherwise well-functioning hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 7 | Pages 935 - 940
1 Jul 2010
McCalden RW Charron KD Yuan X Bourne RB Naudie DD MacDonald SJ

This was a safety study where the hypothesis was that the newer-design CPCS femoral stem would demonstrate similar early clinical results and micromovement to the well-established Exeter stem. Both are collarless, tapered, polished cemented stems, the only difference being a slight lateral to medial taper with the CPCS stem. A total of 34 patients were enrolled in a single-blinded randomised controlled trial in which 17 patients received a dedicated radiostereometric CPCS stem and 17 a radiostereometric Exeter stem. No difference was found in any of the outcome measures pre-operatively or post-operatively between groups. At two years, the mean subsidence for the CPCS stem was nearly half that seen for the Exeter stem (0.77 mm (−0.943 to 1.77) and 1.25 mm (0.719 to 1.625), respectively; p = 0.032). In contrast, the mean internal rotation of the CPCS stem was approximately twice that of the Exeter (1.61° (−1.07° to 4.33°) and 0.59° (0.97° to 1.64°), respectively; p = 0.048). Other migration patterns were not significantly different between the stems. The subtle differences in designs may explain the different patterns of migration.

Comparable migration with the Exeter stem suggests that the CPCS design will perform well in the long term.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 10 | Pages 1333 - 1336
1 Oct 2005
Busch CA Charles MN Haydon CM Bourne RB Rorabeck CH MacDonald SJ McCalden RW

We identified five (2.3%) fractures of the stem in a series of 219 revision procedures using a cementless, cylindrical, extensively porous-coated, distally-fixed femoral stem. Factors relating to the patients, the implant and the operations were compared with those with intact stems. Finite-element analysis was performed on two of the fractured implants.

Factors associated with fracture of the stem were poor proximal bone support (type III–type IV; p = 0.001), a body mass index > 30; (p = 0.014), a smaller diameter of stem (< 13.5 mm; p = 0.007) and the use of an extended trochanteric osteotomy (ETO 4/5: p = 0.028). Finite-element analysis showed that the highest stresses on the stem occurred adjacent to the site of the fracture. The use of a strut graft wired over an extended trochanteric osteotomy in patients lacking proximal femoral cortical support decreased the stresses on the stem by 48%.

We recommend the use of a strut allograft in conjunction with an extended trochanteric osteotomy in patients with poor proximal femoral bone stock.