We report the use of porous metal acetabular
revision shells in the treatment of contained bone loss. The outcomes of
53 patients with ≤ 50% acetabular bleeding host bone
contact were compared with a control group of 49 patients with >
50% to 85% bleeding host bone contact. All patients were treated
with the same type of trabecular metal acetabular revision shell.
The mean age at revision was 62.4 years (42 to 80) and the mean
follow-up of both groups was 72.4 months (60 to 102). Clinical,
radiological and functional outcomes were assessed. There were four
(7.5%) mechanical failures in the ≤ 50% host bone contact
group and no failures in the >
50% host bone contact group (p = 0.068).
Out of both groups combined there were four infections (3.9%) and
five recurrent dislocations (4.9%) with a stable acetabular component
construct that were revised to a constrained liner. Given the complexity
of the reconstructive challenge, porous metal revision acetabular
shells show acceptable failure rates at five to ten years’ follow-up
in the setting of significant contained bone defects. This favourable
outcome might be due to the improved initial stability achieved
by a high coefficient of friction between the acetabular implant
and the host bone, and the high porosity, which affords good bone
ingrowth.