Aims. The Intraosseous Transcutaneous Amputation Prosthesis (ITAP)
may improve quality of life for amputees by avoiding soft-tissue
complications associated with socket prostheses and by improving
sensory feedback and function. It relies on the formation of a seal
between the soft tissues and the implant and currently has a flange
with drilled holes to promote dermal attachment. Despite this, infection
remains a significant risk. This study explored alternative strategies
to enhance soft-tissue integration. Materials and Methods. The effect of ITAP pins with a fully porous titanium alloy flange
with interconnected pores on soft-tissue integration was investigated.
The flanges were coated with fibronectin-functionalised hydroxyapatite
and silver
This study explores data quality in operation type and fracture classification recorded as part of a large research study and a national audit with an independent review. At 17 centres, an expert surgeon reviewed a randomly selected subset of cases from their centre with regard to fracture classification using the AO system and type of operation performed. Agreement for these variables was then compared with the data collected during conduct of the World Hip Trauma Evaluation (WHiTE) cohort study. Both types of surgery and fracture classification were collapsed to identify the level of detail of reporting that achieved meaningful agreement. In the National Hip Fracture Database (NHFD), the types of operation and fracture classification were explored to identify the proportion of “highly improbable” combinations.Aims
Patients and Methods
Periprosthetic femoral fractures (PFF) following total hip arthroplasty
(THA) are devastating complications that are associated with functional
limitations and increased overall mortality. Although cementless
implants have been associated with an increased risk of PFF, the
precise contribution of implant geometry and design on the risk
of both intra-operative and post-operative PFF remains poorly investigated.
A systematic review was performed to aggregate all of the PFF literature
with specific attention to the femoral implant used. A systematic search strategy of several journal databases and
recent proceedings from the American Academy of Orthopaedic Surgeons
was performed. Clinical articles were included for analysis if sufficient
implant description was provided. All articles were reviewed by
two reviewers. A review of fundamental investigations of implant
load-to-failure was performed, with the intent of identifying similar
conclusions from the clinical and fundamental literature.Aims
Patients and Methods
Our aim in this pilot study was to evaluate the fixation of, the bone remodelling around, and the clinical outcome after surgery of a new, uncemented, fully hydroxyapatite-coated, collared and tapered femoral component, designed specifically for elderly patients with a fracture of the femoral neck. We enrolled 50 patients, of at least 70 years of age, with an acute displaced fracture of the femoral neck in this prospective single-series study. They received a total hip replacement using the new component and were followed up regularly for two years. Fixation was evaluated by radiostereometric analysis and bone remodelling by dual-energy x-ray absorptiometry. Hip function and the health-related quality of life were assessed using the Harris hip score and the EuroQol-5D. Up to six weeks post-operatively there was a mean subsidence of 0.2 mm (−2.1 to +0.5) and a retroversion of a mean of 1.2° (−8.2° to +1.5°). No component migrated after three months. The patients had a continuous loss of peri-prosthetic bone which amounted to a mean of 16% (−49% to +10%) at two years. The mean Harris hip score was 82 (51 to 100) after two years. The two-year results from this pilot study indicate that this new, uncemented femoral component can be used for elderly patients with osteoporotic fractures of the femoral neck.
Using data from the Norwegian Hip Fracture Register,
8639 cemented and 2477 uncemented primary hemiarthroplasties for
displaced fractures of the femoral neck in patients aged > 70 years
were included in a prospective observational study. A total of 218
re-operations were performed after cemented and 128 after uncemented
procedures. Survival of the hemiarthroplasties was calculated using
the Kaplan-Meier method and hazard rate ratios (HRR) for revision
were calculated using Cox regression analyses. At five years the
implant survival was 97% (95% confidence interval (CI) 97 to 97)
for cemented and 91% (95% CI 87 to 94) for uncemented hemiarthroplasties.
Uncemented hemiarthroplasties had a 2.1 times increased risk of
revision compared with cemented prostheses (95% confidence interval
1.7 to 2.6, p < 0.001). The increased risk was mainly caused
by revisions for peri-prosthetic fracture (HRR = 17), aseptic loosening
(HRR = 17), haematoma formation (HRR = 5.3), superficial infection
(HRR = 4.6) and dislocation (HRR = 1.8). More intra-operative complications,
including intra-operative death, were reported for the cemented
hemiarthroplasties. However, in a time-dependent analysis, the HRR
for re-operation in both groups increased as follow-up increased. This study showed that the risk for revision was higher for uncemented
than for cemented hemiarthroplasties.
We undertook a prospective randomised controlled trial involving 400 patients with a displaced intracapsular fracture of the hip to determine whether there was any difference in outcome between treatment with a cemented Thompson hemiarthroplasty and an uncemented Austin-Moore prosthesis. The surviving patients were followed up for between two and five years by a nurse blinded to the type of prosthesis used. The mean age of the patients was 83 years (61 to 104) and 308 (77%) were women. The degree of residual pain was less in those treated with a cemented prosthesis (p <
0.0001) three months after surgery. Regaining mobility was better in those treated with a cemented implant (p = 0.005) at six months after operation. No statistically significant difference was found between the two groups with regard to mortality, implant-related complications, re-operations or post-operative medical complications. The use of a cemented Thompson hemiarthroplasty resulted in less pain and less deterioration in mobility than an uncemented Austin-Moore prosthesis with no increase in complications.