Aims. Computer hexapod assisted orthopaedic surgery (CHAOS), is a method
to achieve the intra-operative correction of long bone deformities
using a hexapod external fixator before definitive internal fixation
with minimally invasive stabilisation techniques. The aims of this study were to determine the reliability of this
method in a consecutive case series of patients undergoing femoral
deformity correction, with a minimum six-month follow-up, to assess
the complications and to define the ideal group of patients for
whom this treatment is appropriate. Patients and Methods. The medical records and radiographs of all patients who underwent
CHAOS for femoral
The STRYDE nail is an evolution of the PRECICE Intramedullary Limb Lengthening System, with unique features regarding its composition. It is designed for load bearing throughout treatment in order to improve patient experience and outcomes and allow for simultaneous bilateral lower limb lengthening. The literature published to date is limited regarding outcomes and potential problems. We report on our early experience and raise awareness for the potential of adverse effects from this device. This is a retrospective review of prospective data collected on all patients treated in our institution using this implant. We report the demographics, nail accuracy, reliability, consolidation index, and cases where concerning clinical and radiological findings were encountered. There were 14 STRYDE nails implanted in nine patients (three male and six female) between June 2019 and September 2020. Mean age at surgery was 33 years (14 to 65). Five patients underwent bilateral lengthening (two femoral and three tibial) and four patients unilateral femoral lengthening for multiple aetiologies.Aims
Methods
We report the results of using a combination of fixator-assisted nailing with lengthening over an intramedullary nail in patients with tibial
Aims. As an alternative to external fixators, intramedullary lengthening nails (ILNs) can be employed for distraction osteogenesis. While previous studies have demonstrated that typical complications of external devices, such as soft-tissue tethering, and pin site infection can be avoided with ILNs, there is a lack of studies that exclusively investigated tibial distraction osteogenesis with motorized ILNs inserted via an antegrade approach. Methods. A total of 58 patients (median age 17 years (interquartile range (IQR) 15 to 21)) treated by unilateral tibial distraction osteogenesis for a median leg length discrepancy of 41 mm (IQR 34 to 53), and nine patients with disproportionate short stature treated by bilateral simultaneous tibial distraction osteogenesis, with magnetically controlled motorized ILNs inserted via an antegrade approach, were retrospectively analyzed. The median follow-up was 37 months (IQR 30 to 51). Outcome measurements were accuracy, precision, reliability, bone healing, complications, and patient-reported outcome assessed by the Limb Deformity-Scoliosis Research Society Score (LD-SRS-30). Results. A median tibial distraction of 44 mm (IQR 31 to 49) was achieved with a mean distraction index of 0.5 mm/day (standard deviation 0.13) and median consolidation index of 41.2 days/cm (IQR 34 to 51). Accuracy, precision, and reliability were 91%, 92%, and 97%, respectively. New temporary range of motion limitations occurred in 51% of segments (34/67). Distraction-related equinus
Aims. Double-level lengthening, bone transport, and bifocal compression-distraction are commonly undertaken using Ilizarov or other fixators. We performed double-level fixator-assisted nailing, mainly for the correction of
Distraction osteogenesis with intramedullary lengthening devices has undergone rapid development in the past decade with implant enhancement. In this first single-centre matched-pair analysis we focus on the comparison of treatment with the PRECICE and STRYDE intramedullary lengthening devices and aim to clarify any clinical and radiological differences. A single-centre 2:1 matched-pair retrospective analysis of 42 patients treated with the STRYDE and 82 patients treated with the PRECICE nail between May 2013 and November 2020 was conducted. Clinical and lengthening parameters were compared while focusing radiological assessment on osseous alterations related to the nail’s telescopic junction and locking bolts at four different stages.Aims
Methods
Aims. The Precice intramedullary limb-lengthening system has demonstrated significant benefits over external fixation lengthening methods, leading to a paradigm shift in limb lengthening. This study compares outcomes following antegrade and retrograde femoral lengthening in both adolescent and adult patients. Patients and Methods. A retrospective review of prospectively collected data was undertaken of a consecutive series of 107 femoral lengthening operations in 92 patients. In total, 73 antegrade nails and 34 retrograde nails were inserted. Outcome was assessed by the regenerate healing index (HI), hip and knee range of movement (ROM), and the presence of any complications. Results. The mean lengthening was 4.65 cm (1.5 to 8) in the antegrade group and 4.64 cm (1.6 to 8) in the retrograde group. Of the 107 lengthenings, 100 had sufficient datapoints to calculate the mean HI. This was 31.6 days/cm (15 to 108). There was a trend toward a lower (better) HI with an antegrade nail and better outcomes in adolescent patients, but these were not statistically significant. Hip and knee ROM was maintained and/or improved following commencement of femoral lengthening in 44 patients (60%) of antegrade nails and 13 patients (38%) of retrograde nails. In female patients, loss of movement occurred both earlier and following less total length achieved. Minor implant complications included locking bolt migration and in one patient
The aetiologies of common degenerative spine, hip, and knee pathologies are still not completely understood. Mechanical theories have suggested that those diseases are related to sagittal pelvic morphology and spinopelvic-femoral dynamics. The link between the most widely used parameter for sagittal pelvic morphology, pelvic incidence (PI), and the onset of degenerative lumbar, hip, and knee pathologies has not been studied in a large-scale setting. A total of 421 patients from the Cohort Hip and Cohort Knee (CHECK) database, a population-based observational cohort, with hip and knee complaints < 6 months, aged between 45 and 65 years old, and with lateral lumbar, hip, and knee radiographs available, were included. Sagittal spinopelvic parameters and pathologies (spondylolisthesis and degenerative disc disease (DDD)) were measured at eight-year follow-up and characteristics of hip and knee osteoarthritis (OA) at baseline and eight-year follow-up. Epidemiology of the degenerative disorders and clinical outcome scores (hip and knee pain and Western Ontario and McMaster Universities Osteoarthritis Index) were compared between low PI (< 50°), normal PI (50° to 60°), and high PI (> 60°) using generalized estimating equations.Aims
Methods
Adult presentation of neglected congenital muscular torticollis is rare. We report 12 patients with this condition who underwent a modified Ferkel’s release comprising a bipolar release of sternocleidomastoid with Z-lengthening. They had a mean age of 24 years (17 to 31) and were followed up for a minimum of two years. Post-operatively a cervical collar was applied for three weeks with intermittent supervised active assisted exercises for six weeks. Outcome was assessed using a modified Lee score and a Cheng and Tang score. The mean pre-operative rotational deficit was 8.25° (0° to 15°) and mean lateral flexion deficit was 20.42° (15° to 30°), which improved after treatment to a mean of 1.67° (0° to 5°) and 7.0° (4° to 14°) after treatment, respectively. According to the modified Lee scoring system, six patients had excellent results, two had good results and four had fair results, and using the Cheng and Tang score, eight patients had excellent results and four had good results. Surgical management of adult patients with neglected congenital muscular torticollis using a modified Ferkel’s bipolar release gives excellent results. The range of neck movement and head tilt improved in all 12 patients and cosmesis improved in 11, despite the long-standing nature of the
A flexed knee gait is common in patients with bilateral spastic
cerebral palsy and occurs with increased age. There is a risk for
the recurrence of a flexed knee gait when treated in childhood,
and the aim of this study was to investigate whether multilevel
procedures might also be undertaken in adulthood. At a mean of 22.9 months (standard deviation 12.9), after single
event multi level surgery, 3D gait analysis was undertaken pre-
and post-operatively for 37 adult patients with bilateral cerebral
palsy and a fixed knee gait.Aims
Patients and Methods
Single-event multilevel surgery (SEMLS) has been used as an effective
intervention in children with bilateral spastic cerebral palsy (BSCP)
for 30 years. To date there is no evidence for SEMLS in adults with
BSCP and the intervention remains focus of debate. This study analysed the short-term outcome (mean 1.7 years, standard
deviation 0.9) of 97 ambulatory adults with BSCP who performed three-dimensional
gait analysis before and after SEMLS at one institution. Aims
Methods
External fixation is widely used in orthopaedic
and trauma surgery. Infections around pin or wire sites, which are usually
localised, non-invasive, and are easily managed, are common. Occasionally,
more serious invasive complications such as necrotising fasciitis
(NF) and toxic shock syndrome (TSS) may occur. We retrospectively reviewed all patients who underwent external
fixation between 1997 and 2012 in our limb lengthening and reconstruction
programme. A total of eight patients (seven female and one male)
with a mean age of 20 years (5 to 45) in which pin/wire track infections
became limb- or life-threatening were identified. Of these, four
were due to TSS and four to NF. Their management is described. A
satisfactory outcome was obtained with early diagnosis and aggressive
medical and surgical treatment. Clinicians caring for patients who have external fixation and
in whom infection has developed should be aware of the possibility
of these more serious complications. Early diagnosis and aggressive
treatment are required in order to obtain a satisfactory outcome. Cite this article:
This article presents an overview of mycetoma
and offers guidelines for orthopaedic surgeons who may be involved in
the care of patients with this condition. Cite this article:
There has been an in increase in the availability
of effective biological agents for the treatment of rheumatoid arthritis
as well as a shift towards early diagnosis and management of the
inflammatory process. This article explores the impact this may
have on the place of orthopaedic surgery in the management of patients
with rheumatoid arthritis. Cite this article:
We report our experience using a biodegradable
calcium sulphate antibiotic carrier containing tobramycin in the surgical
management of patients with chronic osteomyelitis. The patients
were reviewed to determine the rate of recurrent infection, the
filling of bony defects, and any problems with wound healing. A
total of 193 patients (195 cases) with a mean age of 46.1 years
(16.1 to 82.0) underwent surgery. According to the Cierny–Mader
classification of osteomyelitis there were 12 type I, 1 type II,
144 type III and 38 type IV cases. The mean follow-up was 3.7 years (1.3
to 7.1) with recurrent infection occurring in 18 cases (9.2%) at
a mean of 10.3 months post-operatively (1 to 25.0). After further
treatment the infection resolved in 191 cases (97.9%). Prolonged
wound ooze (longer than two weeks post-operatively) occurred in
30 cases (15.4%) in which there were no recurrent infection. Radiographic
assessment at final follow-up showed no filling of the defect with
bone in 67 (36.6%), partial filling in 108 (59.0%) and complete filling
in eight (4.4%). A fracture occurred in nine (4.6%) of the treated
osteomyelitic segments at a mean of 1.9 years (0.4 to 4.9) after
operation. We conclude that Osteoset T is helpful in the management of patients
with chronic osteomyelitis, but the filling of the defect in bone
is variable. Prolonged wound ooze is usually self-limiting and not
associated with recurrent infection. Cite this article:
Recent recommendations by the National Institute
for Health and Care Excellence (NICE) suggest that all patients undergoing
elective orthopaedic surgery should be assessed for the risk of
venous thromboembolism (VTE). Little is known about the incidence of symptomatic VTE after
elective external fixation. We studied a consecutive series of adult
patients who had undergone elective Ilizarov surgery without routine
pharmacological prophylaxis to establish the incidence of symptomatic
VTE. A review of a prospectively maintained database of consecutive
patients who were treated between October 1998 and February 2011
identified 457 frames in 442 adults whose mean age was 42.6 years
(16.0 to 84.6). There were 425 lower limb and 32 upper limb frames.
The mean duration of treatment was 25.7 weeks (1.6 to 85.3). According to NICE guidelines all the patients had at least one
risk factor for VTE, 246 had two, 172 had three and 31 had four
or more. One patient (0.23%) developed a pulmonary embolus after surgery
and was later found to have an inherited thrombophilia. There were
27 deaths, all unrelated to VTE. The cost of providing VTE prophylaxis according to NICE guidelines
in this group of patients would be £89 493.40 (£195.80 per patient)
even if the cheapest recommended medication was used. The rate of symptomatic VTE after Ilizarov surgery was low despite
using no pharmacological prophylaxis. This study leads us to question
whether NICE guidelines are applicable to these patients. Cite this article:
Ten patients, who were unsuitable for limb lengthening over an intramedullary nail, underwent lengthening with a submuscular locking plate. Their mean age at operation was 18.5 years (11 to 40). After fixing a locking plate submuscularly on the proximal segment, an external fixator was applied to lengthen the bone after corticotomy. Lengthening was at 1 mm/day and on reaching the target length, three or four screws were placed in the plate in the distal segment and the external fixator was removed. All patients achieved the pre-operative target length at a mean of 4.0 cm (3.2 to 5.5). The mean duration of external fixation was 61.6 days (45 to 113) and the mean external fixation index was 15.1 days/cm (13.2 to 20.5), which was less than one-third of the mean healing index (48 days/cm (41.3 to 55). There were only minor complications. Lengthening with a submuscular locking plate can successfully permit early removal of the fixator with fewer complications and is a useful alternative in children or when nailing is difficult.
Lengthening of the humerus is now an established
technique. We compared the complications of humeral lengthening
with those of femoral lengthening and investigated whether or not
the callus formation in the humerus proceeds at a higher rate than
that in the femur. A total of 24 humeral and 24 femoral lengthenings
were performed on 12 patients with achondroplasia. We measured the
pixel value ratio (PVR) of the lengthened area on radiographs and
each radiograph was analysed for the shape, type and density of
the callus. The quality of life (QOL) of the patients after humeral
lengthening was compared with that prior to surgery. The complication
rate per segment of humerus and femur was 0.87% and 1.37%, respectively.
In the humerus the PVR was significantly higher than that of the
femur. Lower limbs were associated with an increased incidence of
concave, lateral and central callus shapes. Humeral lengthening
had a lower complication rate than lower-limb lengthening, and QOL
increased significantly after humeral lengthening. Callus formation
in the humerus during the distraction period proceeded at a significantly
higher rate than that in the femur. These findings indicate that humeral lengthening has an important
role in the management of patients with achondroplasia.
We present a retrospective review of a single-surgeon series of 30 consecutive lengthenings in 27 patients with congenital short femur using the Ilizarov technique performed between 1994 and 2005. The mean increase in length was 5.8 cm/18.65% (3.3 to 10.4, 9.7% to 48.8%), with a mean time in the frame of 223 days (75 to 363). By changing from a distal to a proximal osteotomy for lengthening, the mean range of knee movement was significantly increased from 98.1° to 124.2° (p = 0.041) and there was a trend towards a reduced requirement for quadricepsplasty, although this was not statistically significant (p = 0.07). The overall incidence of regenerate deformation or fracture requiring open reduction and internal fixation was similar in the distal and proximal osteotomy groups (56.7% and 53.8%, respectively). However, in the proximal osteotomy group, pre-placement of a Rush nail reduced this rate from 100% without a nail to 0% with a nail (p <
0.001). When comparing a distal osteotomy with a proximal one over a Rush nail for lengthening, there was a significant decrease in fracture rate from 58.8% to 0% (p = 0.043). We recommend that in this group of patients lengthening of the femur with an Ilizarov construct be carried out through a proximal osteotomy over a Rush nail. Lengthening should also be limited to a maximum of 6 cm during one treatment, or 20% of the original length of the femur, in order to reduce the risk of complications.
Limb lengthening by callus distraction and external fixation has a high rate of complications. We describe our experience using an intramedullary nail (Fitbone) which contains a motorised and programmable sliding mechanism for limb lengthening and bone transport. Between 2001 and 2004 we lengthened 13 femora and 11 tibiae in ten patients (seven men and three women) with a mean age of 32 years (21 to 47) using this nail. The indications for operation were short stature in six patients and developmental or acquired disorders in the rest. The mean lengthening achieved was 40 mm (27 to 60). The mean length of stay in hospital was seven days (5 to 9). The mean healing index was 35 days/cm (18.8 to 70.9). There were no cases of implant-related infection or malunion.