The purpose of this study was to evaluate the
risk of late displacement after the treatment of distal radial fractures with
a locking volar plate, and to investigate the clinical and radiological
factors that might correlate with re-displacement. From March 2007
to October 2009, 120 of an original cohort of 132 female patients
with unstable fractures of the distal radius were treated with a
volar locking plate, and were studied over a follow-up period of
six months. In the immediate post-operative and final follow-up
radiographs, late displacement was evaluated as judged by ulnar
variance, radial inclination, and dorsal angulation. We also analysed
the correlation of a variety of clinical and radiological factors
with re-displacement. Ulnar variance was significantly overcorrected
(p <
0.001) while radial inclination and dorsal angulation were
undercorrected when compared statistically (p <
0.001) with the unaffected
side in the immediate post-operative stage. During follow-up, radial
shortening and dorsal angulation progressed statistically, but none
had a value beyond the acceptable range. Bone mineral density measured
at the proximal femur and the position of the screws in the subchondral
region, correlated with slight progressive radial shortening, which
was not clinically relevant. Volar locking plating of distal radial fractures is a reliable
form of treatment without substantial late displacement. Cite this article:
This study was performed to review the safety and outcome of total shoulder replacements in patients who are ≥ 80 years of age. A total of 50 total shoulder replacements in 44 patients at a mean age of 82 years (80 to 89) were studied. Their health and shoulder status, the operation and post-operative course were analysed, including pain, movement, patient satisfaction, medical and surgical complications, radiographs, the need for revision surgery, and implant and patient survival. A total of 27 patients had an ASA classification of III or IV and medical abnormalities were common. Of the 13 shoulders with bony deficiency of the glenoid, nine required grafting. The duration of hospital stay was prolonged and blood transfusions were common. There were no peri-operative deaths. The mean follow-up was for 5.5 years (2 to 12). Pain was significantly reduced (p <
0.001) and movement improved in active elevation and both external and internal rotation (p <
0.001). Using the Neer scale for assessing outcome, 40 (80%) shoulders had an excellent or satisfactory result. There were medical or surgical complications in 17 cases. Four shoulders developed radiological evidence of loosened glenoid components, and three of these had a poor outcome. Three other shoulders required revision, two for instability. By the time of this review 39 of the patients had died from unrelated causes at a mean of 7.5 years (0.8 to 16.4) after surgery. Total shoulder replacement is a relatively effective treatment in this elderly group of patients. However, there is a requirement for more intense patient care in the peri-operative period, and non-fatal medical or surgical complications are common. Most of these
We analysed the outcome of the Coonrad-Morrey total elbow replacement used for fracture of the distal humerus in
We present six patients with chronic dislocation of the elbow who were treated by primary semiconstrained total elbow arthroplasty. All were women with a mean age of 65 years (51 to 76), the mean interval between dislocation and surgery was 17 weeks (5 to 52) and the mean follow-up 58 months (24 to 123). The most dramatic improvement was in function. The mean American Shoulder and Elbow Surgeon score was 5.2 times better (p <
0.001) and the mean total range of movement increased from 33° to 121° (p <
0.001) after operation. Three patients developed wear of polyethylene. One required revision for a periprosthetic fracture, and another required a bushing exchange. Primary semiconstrained elbow arthroplasty provides significant, predictable functional improvement. Potential solutions for wear of polyethylene include a different operative technique or design of implant. Despite the high incidence of such wear, total elbow arthroplasty should be considered as a viable treatment option for chronic dislocation of the elbow in
We reviewed 80 shoulders (77 patients) at a mean follow-up of 44 months after insertion of a Grammont inverted shoulder prosthesis. Three implants had failed and had been revised. The mean Constant score had increased from 22.6 points pre-operatively to 65.6 points at review. In 96% of these shoulders there was no or only minimal pain. The mean active forward elevation increased from 73° to 138°. The integrity of teres minor is essential for the recovery of external rotation and significantly influenced the Constant score. Five cases of aseptic loosening of the glenoid and seven of dissociation of the glenoid component were noted. This study confirms the promising early results obtained with the inverted prosthesis in the treatment of a cuff-tear arthropathy. It should be considered in the treatment of osteoarthritis with a massive tear of the cuff but should be reserved for
Over a seven-year period we treated a consecutive series of 58 patients, 20 men and 38 women with a mean age of 66 years (21 to 87) who had an acute complex anterior fracture-dislocation of the proximal humerus. Two patterns of injury are proposed for study based upon a prospective assessment of the pattern of soft-tissue and bony injury and the degree of devascularisation of the humeral head. In 23 patients, the head had retained capsular attachments and arterial back-bleeding (type-I injury), whereas in 35 patients the head was devoid of significant soft-tissue attachments with no active arterial bleeding (type-II injury). Following treatment by open reduction and internal fixation, only two of 23 patients with type-I injuries developed radiological evidence of osteonecrosis of the humeral head, compared with four of seven patients with type-II injuries. A policy of primary treatment by open reduction and internal fixation of type-I injuries is justified, whereas most
In light of the growing number of elderly osteopenic
patients with distal humeral fractures, we discuss the history of
their management and current trends. Under most circumstances operative
fixation and early mobilisation is the treatment of choice, as it
gives the best results. The relative indications for and results
of total elbow replacement
We evaluated the biomechanical properties of two different methods of fixation for unstable fractures of the proximal humerus. Biomechanical testing of the two groups, locking plate alone (LP), and locking plate with a fibular strut graft (LPSG), was performed using seven pairs of human cadaveric humeri. Cyclical loads between 10 N and 80 N at 5 Hz were applied for 1 000 000 cycles. Immediately after cycling, an increasing axial load was applied at a rate of displacement of 5 mm/min. The displacement of the construct, maximum failure load, stiffness and mode of failure were compared. The displacement was significantly less in the LPSG group than in the LP group (p = 0.031). All maximum failure loads and measures of stiffness in the LPSG group were significantly higher than those in the LP group (p = 0.024 and p = 0.035, respectively). In the LP group, varus collapse and plate bending were seen. In the LPSG group, the humeral head cut out and the fibular strut grafts fractured. No broken plates or screws were seen in either group. We conclude that strut graft augmentation significantly increases both the maximum failure load and the initial stiffness of this construct compared with a locking plate alone.
We have previously described the short-term outcome of the use of reverse shoulder arthroplasty in the treatment of acute complex proximal humeral fractures in the elderly. We now report the clinical and radiological outcome of 36 fractures at a mean of 6.6 years (1 to 16). Previously, at a mean follow-up of 6 years (1 to 12) the mean Constant score was 58.5; this was reduced to 53 points with the further follow-up. A total of 23 patients (63%) had radiological evidence of loosening of the glenoid component. Nevertheless, only one patient had aseptic loosening of the baseplate at 12 years’ follow-up. The reduction in the mean Constant score with longer follow-up and the further development of scapular notching is worrying. New developments in design, bearing surfaces and surgical technique, and further follow-up, will determine whether reverse shoulder arthroplasty has a place in the management of complex proximal humeral fractures in the elderly.
This study reports the clinical and sonographic
outcome of arthroscopic rotator cuff repair in patients aged ≥ 70 years
and aimed to determine factors associated with re-tear. A total
of 69 consecutive repairs were performed in 68 patients with a mean
age of 77 years (70 to 86). Constant-Murley scores were collected
pre-operatively and at one year post-operatively. The integrity
of the repair was assessed using ultrasound. Re-tear was detected
in 20 of 62 patients (32%) assessed with ultrasound. Age at operation We conclude that arthroscopic rotator cuff repair in patients
aged ≥ 70 years is a successful procedure. The gender and age of
the patient are important factors to consider when planning management. Cite this article:
We have developed an illustrated questionnaire, the Hand20, comprising 20 short and easy-to-understand questions to assess disorders of the upper limb. We have examined the usefulness of this questionnaire by comparing reliability, validity, responsiveness and the level of missing data with those of the Disabilities of the Arm, Shoulder and Hand (DASH) questionnaire. A series of 431 patients with disorders of the upper limb completed the Hand20 and the Japanese version of the DASH (DASH-JSSH) questionnaire. The norms for Hand20 scores were determined in another cross-sectional study. Most patients had no difficulty in completing the Hand20 questionnaire, whereas the DASH-JSSH had a significantly higher rate of missing data. The standard score for the Hand20 was smaller than the reported norms for the DASH. Our study showed that the Hand20 questionnaire provided validation comparable with that of the DASH-JSSH. Explanatory illustrations and short questions which were easy-to-understand led to better rates of response and fewer missing data, even in elderly individuals with cognitive deterioration.
The Motec cementless modular metal-on-metal ball-and-socket
wrist arthroplasty was implanted in 16 wrists with scaphoid nonunion
advanced collapse (SNAC; grades 3 or 4) and 14 wrists with scapholunate
advanced collapse (SLAC) in 30 patients (20 men) with severe (grades
3 or 4) post-traumatic osteoarthritis of the wrist. The mean age of
the patients was 52 years (31 to 71). All prostheses integrated
well radiologically. At a mean follow-up of 3.2 years (1.1 to 6.1)
no luxation or implant breakage occurred. Two wrists were converted
to an arthrodesis for persistent pain. Loosening occurred in one
further wrist at five years post-operatively. The remainder demonstrated close
bone–implant contact. The clinical results were good, with markedly
decreased Disabilities of the Arm Shoulder and Hand (DASH) and pain
scores, and increased movement and grip strength. No patient used
analgesics and most had returned to work. Good short-term function was achieved using this wrist arthroplasty
in a high-demand group of patients with post-traumatic osteoarthritis.
Dislocation of the shoulder may occur during
seizures in epileptics and other patients who have convulsions. Following
the initial injury, recurrent instability is common owing to a tendency
to develop large bony abnormalities of the humeral head and glenoid
and a susceptibility to further seizures. Assessment is difficult
and diagnosis may be missed, resulting in chronic locked dislocations
with protracted morbidity. Many patients have medical comorbidities,
and successful treatment requires a multidisciplinary approach addressing
the underlying seizure disorder in addition to the shoulder pathology.
The use of bony augmentation procedures may have improved the outcomes
after surgical intervention, but currently there is no evidence-based
consensus to guide treatment. This review outlines the epidemiology
and pathoanatomy of seizure-related instability, summarising the
currently-favoured options for treatment, and their results.
The aim of this study was to compare a third-generation
cementing procedure for glenoid components with a new technique
for cement pressurisation. In 20 pairs of scapulae, 20 keeled and
20 pegged glenoid components were implanted using either a third-generation
cementing technique (group 1) or a new pressuriser (group 2). Cement penetration
was measured by three-dimensional (3D) analysis of micro-CT scans.
The mean 3D depth of penetration of the cement was significantly
greater in group 2 (p <
0.001). The mean thickness of the cement
mantle for keeled glenoids was 2.50 mm (2.0 to 3.3) in group 1 and
5.18 mm (4.4 to 6.1) in group 2, and for pegged glenoids it was 1.72 mm
(0.9 to 2.3) in group 1 and 5.63 mm (3.6 to 6.4) in group 2. A cement
mantle <
2 mm was detected less frequently in group 2 (p <
0.001). Using the cement pressuriser the proportion of cement mantles
<
2 mm was significantly reduced compared with the third-generation
cementing technique.
We describe the results of Copeland surface replacement shoulder arthroplasty using the mark III prosthesis in patients over 80 years of age. End-stage arthritis of the shoulder is a source of significant pain and debilitating functional loss in the elderly. An arthroplasty offers good relief of pain and may allow the patient to maintain independence. The risk-benefit ratio of shoulder replacement may be felt to be too high in an elderly age group, but there is no published evidence to support this theory. We have assessed whether the procedure was as reliable and safe as previously seen in a younger cohort of patients. Between 1993 and 2003, 213 Copeland surface replacement arthroplasty procedures were performed in our unit, of which 29 (13.6%) were undertaken in patients over the age of 80. This group of patients was followed up for a mean of 4.5 years (2.1 to 9.3). Their mean age was 84.3 years (81 to 93), the mean operating time was 40 minutes (30 to 45) and the mean in-patient stay was five days (2 to 21). There were no peri-operative deaths or significant complications. The mean Constant score adjusted for age and gender, improved from 15.1% to 77%. Copeland surface replacement shoulder arthroplasty may be performed with minimal morbidity and rapid rehabilitation in the elderly.
The transfer of part of the ulnar nerve to the musculocutaneous nerve, first described by Oberlin, can restore flexion of the elbow following brachial plexus injury. In this study we evaluated the additional benefits and effectiveness of quantitative electrodiagnosis to select a donor fascicle. Eight patients who had undergone transfer of a simple fascicle of the ulnar nerve to the motor branch of the musculocutaneous nerve were evaluated. In two early patients electrodiagnosis had not been used. In the remaining six patients, however, all fascicles of the ulnar nerve were separated and electrodiagnosis was performed after stimulation with a commercially available electromyographic system. In these procedures, recording electrodes were placed in flexor carpi ulnaris and the first dorsal interosseous. A single fascicle in the flexor carpi ulnaris in which a high amplitude had been recorded was selected as a donor and transferred to the musculocutaneous nerve. In the two patients who had not undergone electrodiagnosis, the recovery of biceps proved insufficient for normal use. Conversely, in the six patients in whom quantitative electrodiagnosis was used, elbow flexion recovered to an M4 level. Quantitative intra-operative electrodiagnosis is an effective method of selecting a favourable donor fascicle during the Oberlin procedure. Moreover, fascicles showing a high-amplitude in reading flexor carpi ulnaris are donor nerves that can restore normal elbow flexion without intrinsic loss.
Displaced fractures of the lateral end of the clavicle in young patients have a high incidence of nonunion and a poor functional outcome after conservative management. Operative treatment is therefore usually recommended. However, current techniques may be associated with complications which require removal of the fixation device. We have evaluated the functional and radiological outcomes using a novel technique of open reduction and internal fixation. A series of 16 patients under 60 years of age with displaced fractures of the lateral end were treated by open reduction and fixation using a twin coracoclavicular endobutton technique. They were followed up for the first year after their injury. At one year the mean Constant score was 87.1 and the median Disabilities of the Arm, Shoulder and Hand score was 3.3. All fractures had united, except in one patient who developed an asymptomatic fibrous union. One patient had post-traumatic stiffness of the shoulder, which resolved with physiotherapy. None required re-operation. This technique produces good functional and radiological outcomes with a low prevalence of complications and routine implant removal is not necessary.
We reviewed the outcome of 28 patients who had been treated using the Aequalis fracture prosthesis for an acute fracture of the proximal humerus at a mean follow-up of 39.3 months (24 to 63). The mean age of the patients at the time of the fracture was 66.3 years (38 to 80). The mean Constant score was 68.2 (37 to 84) for the operated shoulder, which represented 89.5% of the mean score for the uninjured side (p <
0.001). The quality of the reconstruction as shown on the immediate post-operative radiographs was categorised into three types, anatomical, acceptable, and unacceptable, depending on the position of the tuberosities relative to the prosthetic head and the humeral shaft. Anatomical reconstruction was associated with a higher mean Constant score as well as higher mean values of anterior forward elevation, abduction and external rotation than the other types, but the differences were not statistically significant (p >
0.231). A total of 18 patients had active anterior elevation ≥150°. Their mean active abduction and external rotation were 163.6° and 31.3°, respectively. In seven of the 28 patients, the mean active anterior elevation, abduction and external rotation were 130.7°, 129.2° and 22.8°, respectively. In all, 12 patients were very satisfied with the results, 12 were satisfied, two were dissatisfied and two were disappointed; 26 reported no or only mild pain while only two had moderate pain. In five patients proximal migration of the humeral head was shown on the anteroposterior radiographs of the shoulder. No evidence of loosening was found in any prosthesis.
This study reviews the predisposing features, the clinical, and laboratory findings at the time of diagnosis and the results of single-stage revision of prosthetic replacement of the elbow for infection. Deep infection occurred in six of 305 (1.9%) primary total elbow replacements. The mean follow-up after revision was 6.8 years (6 months to 16 years) and the mean age at the time of revision was 62.7 years (56 to 74). All six cases with infection had rheumatoid arthritis and had received steroid therapy. The infective organism was
There have been few reports in the literature of total elbow arthroplasty extending beyond 10 to 15 years. We reviewed 40 patients (41 elbows) with a mean age of 56 years (19 to 83) who had undergone a Coonrad/Coonrad-Morrey elbow arthroplasty by one surgeon for various diagnoses between 1974 and 1994. Surgical selection excluded patients with previous elbow infection or who refused to accept a sedentary level of elbow activity postoperatively. Objective data were collected from charts, radiographs, clinical photographs and supplemented by the referring orthopaedic surgeons’ records and radiographs if health or distance prevented a patient from returning for final review. Subjective outcome was defined by patient satisfaction. Of the 41 elbows, 21 were functional between 10 and 14 years after operation, ten between 15 and 19 years and ten between 20 and 31 years. There were 14 complications and 13 revisions, but no cases of acute infection, or permanent removal of any implant.