Proximal femoral endoprosthetic replacements (PFEPRs) are the most common reconstruction option for osseous defects following primary and metastatic tumour resection. This study aimed to compare the rate of implant failure between PFEPRs with monopolar and bipolar hemiarthroplasties and acetabular arthroplasties, and determine the optimum articulation for revision PFEPRs. This is a retrospective review of 233 patients who underwent PFEPR. The mean age was 54.7 years (SD 18.2), and 99 (42.5%) were male. There were 90 patients with primary bone tumours (38.6%), 122 with metastatic bone disease (52.4%), and 21 with haematological malignancy (9.0%). A total of 128 patients had monopolar (54.9%), 74 had bipolar hemiarthroplasty heads (31.8%), and 31 underwent acetabular arthroplasty (13.3%).Aims
Methods
Endoprosthetic reconstruction following distal femur tumour resection has been widely advocated. In this paper, we present the design of an uncemented endoprosthesis system featuring a short, curved stem, with the goal of enhancing long-term survivorship and functional outcomes. This study involved patients who underwent implantation of an uncemented distal femoral endoprosthesis with a short and curved stem between 2014 and 2019. Functional outcomes were assessed using the 1993 version of the Musculoskeletal Tumour Society (MSTS-93) score. Additionally, we quantified five types of complications and assessed osseointegration radiologically. The survivorship of the endoprosthesis was evaluated according to two endpoints. A total of 134 patients with a median age of 26 years (IQR 16 to 41) were included in our study. The median follow-up time was 61 months (IQR 56 to 76), and the median functional MSTS-93 was 83% (IQR 73 to 91) postoperatively.Aims
Methods
Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model. We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the bone-implant fixation); or amputation.Aims
Methods
We aimed to examine the long-term mechanical survivorship, describe the modes of all-cause failure, and identify risk factors for mechanical failure of all-polyethylene tibial components in endoprosthetic reconstruction. This is a retrospective database review of consecutive endoprosthetic reconstructions performed for oncological indications between 1980 and 2019. Patients with all-polyethylene tibial components were isolated and analyzed for revision for mechanical failure. Outcomes included survival of the all-polyethylene tibial component, revision surgery categorized according to the Henderson Failure Mode Classification, and complications and functional outcome, as assessed by the Musculoskeletal Tumor Society (MSTS) score at the final follow-up.Aims
Methods
The purpose of this study was to report the long-term results of extendable endoprostheses of the humerus in children after the resection of a bone sarcoma. A total of 35 consecutive patients treated with extendable endoprosthetic replacement of the humerus in children were included. There were 17 boys and 18 girls in the series with a median age at the time of initial surgery of nine years (interquartile range (IQR) 7 to 11).Aims
Methods
The aim of this study was to describe the use of 3D-printed sacral endoprostheses to reconstruct the pelvic ring and re-establish spinopelvic stability after total We retrospectively reviewed 32 patients who underwent TES in our hospital between January 2015 and December 2017. We divided the patients into three groups on the basis of the method of reconstruction: an endoprosthesis group (n = 10); a combined reconstruction group (n = 14), who underwent non-endoprosthetic combined reconstruction, including anterior spinal column fixation; and a spinopelvic fixation (SPF) group (n = 8), who underwent only SPF. Spinopelvic stability, implant survival (IS), intraoperative haemorrhage rate, and perioperative complication rate in the endoprosthesis group were documented and compared with those of other two groups.Aims
Patients and Methods
Aseptic loosening is a major cause of failure in cemented endoprosthetic reconstructions. This paper presents the long-term outcomes of a custom-designed cross-pin fixation construct designed to minimize rotational stress and subsequent aseptic loosening in selected patients. The paper will also examine the long-term survivorship and modes of failure when using this technique. A review of 658 consecutive, prospectively collected cemented endoprosthetic reconstructions for oncological diagnoses at a single centre between 1980 and 2017 was performed. A total of 51 patients were identified with 56 endoprosthetic implants with cross-pin fixation, 21 of which were implanted following primary resection of tumour. Locations included distal femoral (n = 36), proximal femoral (n = 7), intercalary (n = 6), proximal humeral (n = 3), proximal tibial (n = 3), and distal humeral (n = 1).Aims
Patients and Methods
Aims
Patients and Methods
Following the resection of an extensive amount of bone in the
treatment of a tumour, the residual segment may be insufficient
to accept a standard length intramedullary cemented stem. Short-stemmed
endoprostheses conceivably have an increased risk of aseptic loosening.
Extra-cortical plates have been added to minimise this risk by supplementing
fixation. The aim of this study was to investigate the survivorship
of short-stemmed endoprostheses and extra-cortical plates. The study involved 37 patients who underwent limb salvage surgery
for a primary neoplasm of bone between 1998 and 2013. Endoprosthetic
replacement involved the proximal humerus in nine, the proximal
femur in nine, the distal femur in 13 and the proximal tibia in
six patients. There were 12 primary (32%) and 25 revision procedures (68%).
Implant survivorship was compared with matched controls. The amount
of bone that was resected was >
70% of its length and statistically
greater than the standard control group at each anatomical site.Aims
Patients and Methods
Aims. The use of a noninvasive growing endoprosthesis in the management
of primary bone tumours in children is well established. However,
the efficacy of such a prosthesis in those requiring a revision
procedure has yet to be established. The aim of this series was
to present our results using extendable prostheses for the revision
of previous endoprostheses. Patients and Methods. All patients who had a noninvasive growing endoprosthesis inserted
at the time of a revision procedure were identified from our database.
A total of 21 patients (seven female patients, 14 male) with a mean
age of 20.4 years (10 to 41) at the time of revision were included.
The indications for revision were mechanical failure, trauma or infection
with a residual leg-length discrepancy. The mean follow-up was 70
months (17 to 128). The mean shortening prior to revision was 44 mm
(10 to 100). Lengthening was performed in all but one patient with
a mean lengthening of 51 mm (5 to 140). Results. The mean residual leg length discrepancy at final follow-up of
15 mm (1 to 35). Two patients developed a deep periprosthetic infection,
of whom one required amputation to eradicate the infection; the
other required two-stage
The aim of this study was to establish what happens to patients
in the long term after endoprosthetic replacement for a primary
malignant tumour of bone. We conducted a retrospective analysis of a prospectively maintained
database to identify all patients who had undergone an endoprosthetic
replacement more than 25 years ago and who were still alive. Their
outcomes were investigated with reference to their complications and
need for further surgery. A total of 230 patients were identified.
Their mean age at diagnosis was 20.7 years (five to 62). The most
common diagnosis was osteosarcoma (132). The most common site was
the distal femur (102). Aims
Patients and Methods
We reviewed our initial seven-year experience
with a non-invasive extendible prosthesis in 34 children with primary bone
tumours. The distal femur was replaced in 25 cases, total femur
in five, proximal femur in one and proximal tibia in three. The
mean follow-up was 44 months (15 to 86) and 27 patients (79%) remain
alive. The prostheses were lengthened by an electromagnetic induction
mechanism in an outpatient setting and a mean extension of 32 mm
(4 to 80) was achieved without anaesthesia. There were lengthening
complications in two children: failed lengthening in one and the
formation of scar tissue in the other. Deep infection developed
in six patients (18%) and local recurrence in three. A total of
11 patients required further surgery to the leg. Amputation was
necessary in five patients (20%) and a two-stage revision in another.
There were no cases of loosening, but two patients had implant breakage
and required revision. The mean Musculoskeletal Tumor Society functional
score was 85% (60% to 100%) at last known follow-up. These early
results demonstrate that the non-invasive extendible prosthesis
allows successful lengthening without surgical intervention, but
the high incidence of infection is a cause for concern.
We report our early experience with a new peri-acetabular reconstruction endoprosthesis used for pelvic reconstruction after tumour resection. The outcome of 21 patients who underwent limb salvage following type II pelvic resection and reconstruction using the peri-acetabular reconstruction prosthesis between 2000 and 2006 was retrospectively reviewed. This prosthesis was designed to use the remaining part of the ilium to support a horizontally placed acetabular component secured with internal fixation and bone cement. Into this device a constrained acetabular liner is positioned which is articulates with a conventional femoral component to which a modular extension and modular head are attached. The mean follow-up was 20.5 months (1 to 77). The most common complications were deep infection, superficial wound infections, and dislocation. The mean musculoskeletal tumor society functional outcome score for the survivors was 20.1(11 to 27). We recommend the use of the peri-acetabular reconstruction prosthesis for reconstruction of large defects after type II pelvic resection, as this design has a greater inherent stability over other available prostheses.
We undertook a cemental unipolar proximal femoral endoprosthetic replacement in 131 patients with a mean age of 50 years (2 to 84). Primary malignant tumours were present in 54 patients and 67 had metastatic disease. In addition, eight patients had either lymphoma or myeloma and two had non-oncological disorders. The mean follow-up was 27 months (0 to 180). An acetabular revision was required later in 14 patients, 12 of whom had been under the age of 21 years at the time of insertion of their original prosthesis. The risk of acetabular revision in patients over 21 years of age was 8% at five years compared with 36% in those aged under 21 years. All the unipolar hips in this younger age group required revision within 11 years of the initial operation. We conclude that unipolar replacement should not be used in younger patients and should be avoided in patients with a life expectancy of more than five years.
We reviewed 25 patients in whom a MUTARS megaprosthesis with a conical fluted stem had been implanted. There were three types of stem: a standard stem was used in 17 cases (three in the proximal femur, nine in the distal femur and five proximal tibia), a custom-made proximal femoral stem in four cases and a custom-made distal femoral stem in four cases. The mean age of the patients was 40.1 years (17 to 70) and the mean follow-up was for 2.5 years (0.9 to 7.4). At follow-up two patients had died from their disease: one was alive with disease and 22 were disease-free. One of 23 prostheses had been removed for infection and another revised to a cemented stem. The mean Musculoskeletal Tumor Society score was 24.9 (12 to 30) and the mean Karnofsky index was 82% (60% to 100%). There was no radiological evidence of loosening or subsidence. Stem stress shielding was seen in 11 patients and was marked in five of these. There were five complications, rupture of the extensor mechanism of the knee after extra-articular resection in two patients, deep venous thrombosis in one, septic loosening in one, and dislocation of the hip in one. The survival rate after seven years was 87% (95% confidence interval (CI) 83 to 91) for the patients and 95% (95% CI 91 to 99) for the megaprosthesis. A longer follow-up is needed to confirm these encouraging results.
Between 1966 and 2001, 1254 patients underwent excision of a bone tumour with endoprosthetic replacement. All patients who had radiotherapy were identified. Their clinical details were retrieved from their records. A total of 63 patients (5%) had received adjunctive radiotherapy, 29 pre-operatively and 34 post-operatively. The mean post-operative Musculoskeletal Tumor Society scores of irradiated patients were significantly lower (log-rank test, p = 0.009). The infection rate in the group who had not been irradiated was 9.8% (117 of 1191), compared with 20.7% (6 of 29) in those who had pre-operative radiotherapy and 35.3% (12 of 34) in those who radiotherapy post-operatively. The infection-free survival rate at ten years was 85.5% for patients without radiotherapy, 74.1% for those who had pre-operative radiotherapy and 44.8% for those who had post-operative radiotherapy (log-rank test, p <
0.001). The ten-year limb salvage rate was 89% for those who did not have radiotherapy and 76% for those who did (log-rank test, p = 0.02). Radiotherapy increased the risk of revision (log-rank test, p = 0.015). A total of ten amputations were necessary to control infection, of which nine were successful. Radiotherapy may be necessary for the treatment of a bone sarcoma but increases the risk of deep infection for which amputation may be the only solution.
We have investigated whether improvements in design have altered the outcome for patients undergoing endoprosthetic replacement of the proximal tibia following resection of a tumour. Survival of the implant and ‘servicing’ procedures have been documented using a prospective database. A total of 194 patients underwent a proximal tibial replacement, with 95 having a fixed-hinge design and 99 a rotating-hinge with a hydroxyapatite collar; their median age was 21.5 years (10 to 74). At a mean follow-up of 14.7 years (5 to 29), 115 patients remain alive. The risk of revision for any reason in the fixed-hinge group was 32% at five years, 61% at ten years and 75% at 15 and 20 years, and in the rotating-hinge group 12% at five years, 25% at ten years and 30% at 15 years. Aseptic loosening was the most common reason for revision in the fixed-hinge knees, fracture of the implant in the early design of rotating hinges and infection in the current version. The risk of revision for aseptic loosening in the fixed-hinge knees was 46% at ten years. This was reduced to 3% in the rotating-hinge knee with a hydroxyapatite collar. The cemented, rotating hinge design currently offers the best chance of long-term survival of the prosthesis.