This study reports the mid-term results of a large-bearing hybrid
This study aims to assess the relationship between history of pseudotumour formation secondary to metal-on-metal (MoM) implants and periprosthetic joint infection (PJI) rate, as well as establish ESR and CRP thresholds that are suggestive of infection in these patients. We hypothesized that patients with a pseudotumour were at increased risk of infection. A total of 1,171 total hip arthroplasty (THA) patients with MoM articulations from August 2000 to March 2014 were retrospectively identified. Of those, 328 patients underwent metal artefact reduction sequence MRI and had minimum two years’ clinical follow-up, and met our inclusion criteria. Data collected included demographic details, surgical indication, laterality, implants used, history of pseudotumour, and their corresponding preoperative ESR (mm/hr) and CRP (mg/dl) levels. Multivariate logistic regression modelling was used to evaluate PJI and history of pseudotumour, and receiver operating characteristic curves were created to assess the diagnostic capabilities of ESR and CRP to determine the presence of infection in patients undergoing revision surgery.Aims
Methods
We investigated the incidence of soft-tissue
lesions after small head
In a randomised study, 28 patients with a mean age of 62.2 years (32 to 81) with osteoarthritis or avascular necrosis of the hip received either a ceramic-on-ceramic or a
The present study investigated the five-year interval changes in pseudotumours and measured serum metal ions at long-term follow-up of a previous report of 28 mm diameter metal-on-metal (MoM) total hip arthroplasty (THA). A total of 72 patients (mean age 46.6 years (37 to 55); 43 men, 29 women; 91 hips) who underwent cementless primary MoM THA with a 28 mm modular head were included. The mean follow-up duration was 20.3 years (18 to 24). All patients had CT scans at a mean 15.1 years (13 to 19) after the index operation and subsequent follow-up at a mean of 20.2 years (18 to 24). Pseudotumour volume, type of mass, and new-onset pseudotumours were evaluated using CT scanning. Clinical outcomes were assessed by Harris Hip Score (HHS) and the presence of groin pain. Serum metal ion (cobalt (Co) and chromium (Cr)) levels were measured at the latest follow-up.Aims
Patients and Methods
A total of 397 hips were randomised to receive
Metasul metal-on-metal (MoM), metal-on-conventional polyethylene (MoP)
or ceramic-on-polyethylene (CoP) bearings using a cemented triple-tapered
polished femoral component (MS-30). There were 129 MoM hips in 123
patients (39 male and 84 female, mean age 63.3 years (40.7 to 72.9)),
137 MoP hips in 127 patients (39 male and 88 female, mean age 62.8
years (24.5 to 72.7)) and 131 CoP hips in 124 patients (51 male
and 73 female, mean age 63.9 years (30.6 to 73.8)). All acetabular
components were cemented Weber polyethylene components with the
appropriate inlay for the MoM articulation. Clinical evaluation
was undertaken using the Harris hip score (HHS) and radiological
assessments were made at two, five and seven years. The HHS and radiological
analysis were available for 341 hips after seven years. The MoM
group had the lowest mean HHS (p = 0.124), a higher rate of revision
(p <
0.001) and a higher incidence of radiolucent lines in unrevised
hips (p <
0.001). In all, 12 revisions had been performed in
12 patients: eight in the MoM group (four for infection, four for
aseptic loosening, three in the MoP group (one each of infection,
dislocation and pain) and one in the CoP group (infection). Our findings reveal no advantage to the MoM bearing and identified
a higher revision rate and a greater incidence of radiolucent lines
than with the other articulations. We recommend that patients with
a 28 mm Metasul MoM bearing be followed carefully. Cite this article:
A total of 219 hips in 192 patients aged between
18 and 65 years were randomised to 28-mm metal-on-metal uncemented
total hip replacements (THRs, 107 hips) or hybrid hip resurfacing
(HR, 112 hips). At a mean follow-up of eight years (6.6 to 9.3)
there was no significant difference between the THR and HR groups
regarding rate of revision (4.0% (4 of 99) Cite this article:
Peri-articular soft-tissue masses or ‘pseudotumours’
can occur after large-diameter metal-on-metal (MoM) resurfacing
of the hip and conventional total hip replacement (THR). Our aim
was to assess the incidence of pseudotumour formation and to identify
risk factors for their formation in a prospective cohort study. A total of 119 patients who underwent 120 MoM THRs with large-diameter
femoral heads between January 2005 and November 2007 were included
in the study. Outcome scores, serum metal ion levels, radiographs
and CT scans were obtained. Patients with symptoms or an identified
pseudotumour were offered MRI and an ultrasound-guided biopsy. There were 108 patients (109 hips) eligible for evaluation by
CT scan at a mean follow-up of 3.6 years (2.5 to 4.5); 42 patients
(39%) were diagnosed with a pseudotumour. The hips of 13 patients
(12%) were revised to a polyethylene acetabular component with small-diameter
metal head. Patients with elevated serum metal ion levels had a
four times increased risk of developing a pseudotumour. This study shows a substantially higher incidence of pseudotumour
formation and subsequent revisions in patients with MoM THRs than
previously reported. Because most revision cases were identified
only after an intensive screening protocol, we recommend close monitoring
of patients with MoM THR.
Patients from a randomised trial on resurfacing
hip arthroplasty (RHA) (n = 36, 19 males; median age 57 years, 24
to 65) comparing a conventional 28 mm metal-on-metal total hip arthroplasty
(MoM THA) (n = 28, 17 males; median age 59 years, 37 to 65) and
a matched control group of asymptomatic patients with a 32 mm ceramic-on-polyethylene
(CoP) THA (n = 33, 18 males; median age 63 years, 38 to 71) were
cross-sectionally screened with metal artefact reducing sequence-MRI
(MARS-MRI) for pseudotumour formation at a median of 55 months (23
to 72) post-operatively. MRIs were scored by consensus according
to three different classification systems for pseudotumour formation. Clinical scores were available for all patients and metal ion
levels for MoM bearing patients. Periprosthetic lesions with a median volume of 16 mL (1.5 to
35.9) were diagnosed in six patients in the RHA group (17%), one
in the MoM THA group (4%) and six in the CoP group (18%). The classification
systems revealed no clear differences between the groups. Solid
lesions (n = 3) were exclusively encountered in the RHA group. Two patients
in the RHA group and one in the MoM THA group underwent a revision
for pseudotumour formation. There was no statistically significant
relationship between clinical scoring, metal ion levels and periprosthetic
lesions in any of the groups. Periprosthetic fluid collections are seen on MARS-MRI after conventional
CoP THA and RHA and may reflect a soft-tissue collection or effusion. Currently available MRI classification systems seem to score
these collections as pseudotumours, causing an-overestimatation
of the incidence of pseudotumours. Cite this article:
Modern metal-on-metal bearings produce less wear debris and osteolysis, but have the potential adverse effect of release of ions. Improved ceramic-on-ceramic bearings have the lowest wear of all, but the corrosion process has not been analysed. Our aim was to measure the serum ion release (ng/ml) in 23 patients having stable hip prostheses with a ceramic-on-ceramic coupling (group A) and to compare it with the release in 42 patients with a metal-on-metal bearing (group B) in the medium term. Reference values were obtained from a population of 47 healthy subjects (group C). The concentrations of chromium, cobalt, aluminium and titanium were measured. There was a significant increase of cobalt, chromium and aluminium levels (p <
0.05) in group B compared with groups A and C. Group A did not differ significantly from the control group. Despite the apparent advantage of a metal-on-metal coupling, especially in younger patients with a long life expectancy, a major concern arises regarding the extent and duration of ion exposure. For this reason, the low corrosion level in a ceramic-on-ceramic coupling could be advantageous.
Large-head
We have updated our previous randomised controlled trial comparing release of chromium (Cr) and cobalt (Co) ions and included levels of titanium (Ti) ions. We have compared the findings from 28 mm
Aims. Early evidence has emerged suggesting that ceramic-on-ceramic
articulations induce a different tissue reaction to ceramic-on-polyethylene
and metal-on-metal bearings. Therefore, the aim of this study was
to investigate the tissue reaction and cellular response to ceramic
total hip arthroplasty (THA) materials in vitro,
as well as the tissue reaction in capsular tissue after revision
surgery of ceramic-on-ceramic THAs. Patients and Methods. We investigated tissue collected at revision surgery from nine
ceramic-on-ceramic articulations. we compared our findings with
tissue obtained from five
Metal-on-metal hip arthroplasty gained significant
favor in the first decade of the millennium. However, the past several
years have seen increasing reports of failure, pseudotumor and other
adverse reactions. This study presents the results of a single center’s
15-year experience with
A recent report from France suggested an association between the use of cobalt-chrome (CoCr) femoral heads in total hip arthroplasties (THAs) and an increased risk of dilated cardiomyopathy and heart failure. CoCr is a commonly used material in orthopaedic implants. If the reported association is causal, the consequences would be significant given the millions of joint arthroplasties and other orthopaedic procedures in which CoCr is used annually. We examined whether CoCr-containing THAs were associated with an increased risk of all-cause mortality, heart outcomes, cancer, and neurodegenerative disorders in a large national database. Data from the National Joint Registry was linked to NHS English hospital inpatient episodes for 374,359 primary THAs with up to 14.5 years' follow-up. We excluded any patients with bilateral THAs, knee arthroplasties, indications other than osteoarthritis, aged under 55 years, and diagnosis of one or more outcome of interest before THA. Implants were grouped as either containing CoCr or not containing CoCr. The association between implant construct and the risk of all-cause mortality and incident heart failure, cancer, and neurodegenerative disorders was examined.Aims
Methods
To find out if there is an inverse association between estimated glomerular filtration rate (eGFR) and whole blood cobalt (Co) and chromium (Cr) levels in patients with metal-on-metal (MoM) hip arthroplasties and renal insufficiency, suggesting that renal insufficiency could cause accumulation of Co and Cr in blood. Out of 2,520 patients with 3,013 MoM hip arthroplasties, we identified 1,244 patients with whole blood Co, Cr, and creatinine measured within no more than a one-year interval. We analyzed the correlation of blood metal ion levels and eGFR to identify a potential trend of accumulating Co or Cr with decreasing eGFR.Aims
Methods
Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues. A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy.Aims
Methods
Surface replacement arthroplasty (SRA), compared with traditional total hip arthroplasty (THA), is more expensive and carries unique concern related to metal ions production and hypersensitivity. Additionally, SRA is a more demanding procedure with a decreased margin for error compared with THA. To justify its use, SRA must demonstrate comparable component survival and some clinical advantages. We therefore performed a systematic literature review to investigate the differences in complication rates, patient-reported outcomes, stress shielding, and hip biomechanics between SRA and THA. A systematic review of the literature was completed using MEDLINE and EMBASE search engines. Inclusion criteria were level I to level III articles that reported clinical outcomes following primary SRA compared with THA. An initial search yielded 2503 potential articles for inclusion. Exclusion criteria included review articles, level IV or level V evidence, less than one year’s follow-up, and previously reported data. In total, 27 articles with 4182 patients were available to analyze.Aims
Materials and Methods
The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083 metal-on-metal (MOM), 90 ceramic-on-ceramic (COC), and 81 surface arthroplasties (SAs). With the exception of the COC, all other groups used cobalt-chromium (CoCr) femoral heads. The mean follow-up was 10 (0.008 to 35) years. Descriptive statistics with revisions per 100 component years (re/100 yr) and survival analysis with revision for any reason as the endpoint were used to compare bearing surfaces.Aims
Methods