We undertook a retrospective analysis of 306
procedures on 233 patients, with a mean age of 12 years (1 to 21),
in order to evaluate the use of somatosensory evoked potential (SSEP)
monitoring for the early detection of nerve compromise during external
fixation procedures for limb lengthening and correction of deformity.
Significant SSEP changes were identified during 58 procedures (19%).
In 32 instances (10.5%) the changes were transient, and resolved
once the surgical cause had been removed. The remaining 26 (8.5%)
were analysed in two groups, depending on whether or not corrective
action had been performed in response to critical changes in the
SSEP recordings. In 16 cases in which no corrective action was taken,
13 (81.2%, 4.2% overall) developed a post-operative neurological
deficit, six of which were permanent and seven temporary, persisting
for five to 18 months. In the ten procedures in which corrective
action was taken, four patients (40%, 1.3% overall) had a temporary
(one to eight months) post-operative neuropathy and six had no deficit. After appropriate intervention in response to SSEP changes, the
incidence and severity of neurological deficits were significantly
reduced, with no cases of permanent neuropathy. SSEP monitoring
showed 100% sensitivity and 91% specificity for the detection of
nerve injury during external fixation. It is an excellent diagnostic
technique for identifying nerve lesions when they are still highly
reversible.
Peripheral nerve injury is an uncommon but serious
complication of hip surgery that can adversely affect the outcome.
Several studies have described the use of electromyography and intra-operative
sensory evoked potentials for early warning of nerve injury. We
assessed the results of multimodal intra-operative monitoring during
complex hip surgery. We retrospectively analysed data collected
between 2001 and 2010 from 69 patients who underwent complex hip
surgery by a single surgeon using multimodal intra-operative monitoring
from a total pool of 7894 patients who underwent hip surgery during
this period. In 24 (35%) procedures the surgeon was alerted to a
possible lesion to the sciatic and/or femoral nerve. Alerts were
observed most frequently during peri-acetabular osteotomy. The surgeon
adapted his approach based on interpretation of the neurophysiological changes.
From 69 monitored surgical procedures, there was only one true positive
case of post-operative nerve injury. There were no false positives
or false negatives, and the remaining 68 cases were all true negative.
The sensitivity for predicting post-operative nerve injury was 100%
and the specificity 100%. We conclude that it is possible and appropriate
to use this method during complex hip surgery and it is effective
for alerting the surgeon to the possibility of nerve injury.