The Oxford mobile-bearing unicompartmental knee
replacement (UKR) is an effective and safe treatment for osteoarthritis
of the medial compartment. The results in the lateral compartment
have been disappointing due to a high early rate of dislocation
of the bearing. A series using a newly designed domed tibial component
is reported. The first 50 consecutive domed lateral Oxford UKRs in 50 patients
with a mean follow-up of three years (2.0 to 4.3) were included.
Clinical scores were obtained prospectively and Kaplan-Meier survival
analysis was performed for different endpoints. Radiological variables
related to the position and alignment of the components were measured. One patient died and none was lost to follow-up. The cumulative
incidence of dislocation was 6.2% (95% confidence interval (CI)
2.0 to 17.9) at three years. Survival using revision for any reason
and aseptic revision was 94% (95% CI 82 to 98) and 96% (95% CI 85
to 99) at three years, respectively. Outcome scores, visual analogue
scale for pain and maximum knee flexion showed a significant improvement
(p <
0.001). The mean Oxford knee score was 43 (. sd. 5.3),
the mean Objective American Knee Society score was 91 (. sd. 13.9)
and the mean Functional American Knee Society score was 90 (. sd. 17.5).
The mean maximum flexion was 127° (90° to 145°). Significant elevation
of the lateral joint line as measured by the proximal tibial varus
angle (p = 0.04) was evident in the dislocation group when compared
with the non-dislocation group. Clinical results are excellent and short-term survival has improved
when compared with earlier series. The risk of dislocation remains
higher using a mobile-bearing UKR in the lateral compartment when
compared with the medial compartment. Patients should be informed
about this complication. To avoid dislocations, care must be taken
not to elevate the lateral joint line.