Aims. This combined clinical and in vitro study aimed to determine the incidence of liner malseating in modular dual mobility (MDM) constructs in primary total hip arthroplasties (THAs) from a large volume arthroplasty centre, and determine whether malseating increases the potential for fretting and corrosion at the modular metal interface in malseated MDM constructs using a simulated corrosion chamber. Methods. For the clinical
There have been several studies examining the
association between the morphological characteristics seen in acetabular
dysplasia and the incidence of the osteoarthritis (OA). However, most studies focus mainly on acetabular morphological
analysis, and few studies have scrutinised the effect of femoral
morphology. In this study we enrolled 36 patients with bilateral
acetabular dysplasia and early or mid-stage OA in one hip and no
OA in the contralateral hip. CT scans were performed from the iliac
crest to 2 cm inferior to the tibial tuberosity, and the morphological
characteristics of both acetabulum and femur were studied. . In addition, 200 hips in 100 healthy volunteer Chinese adults
formed a control group. The results showed that the dysplastic group
with OA had a significantly larger femoral neck anteversion and
a significantly shorter abductor lever
Achieving accurate implant positioning and restoring native hip biomechanics are key surgeon-controlled technical objectives in total hip arthroplasty (THA). The primary objective of this study was to compare the reproducibility of the planned preoperative centre of hip rotation (COR) in patients undergoing robotic arm-assisted THA versus conventional THA. This prospective randomized controlled trial (RCT) included 60 patients with symptomatic hip osteoarthritis undergoing conventional THA (CO THA) versus robotic arm-assisted THA (RO THA). Patients in both arms underwent pre- and postoperative CT scans, and a patient-specific plan was created using the robotic software. The COR, combined offset, acetabular orientation, and leg length discrepancy were measured on the pre- and postoperative CT scanogram at six weeks following surgery.Aims
Methods
The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022.Aims
Methods
The aim of the study was to investigate whether the primary stability of press-fit acetabular components can be improved by altering the impaction procedure. Three impaction procedures were used to implant acetabular components into human cadaveric acetabula using a powered impaction device. An impaction frequency of 1 Hz until complete component seating served as reference. Overimpaction was simulated by adding ten strokes after complete component seating. High-frequency implantation was performed at 6 Hz. The lever-out moment of the acetabular components was used as measure for primary stability. Permanent bone deformation was assessed by comparison of double micro-CT (µCT) measurements before and after impaction. Acetabular component deformation and impaction forces were recorded, and the extent of bone-implant contact was determined from 3D laser scans.Aims
Methods
The primary outcome was investigating differences in wear, as measured by femoral head penetration, between cross-linked vitamin E-diffused polyethylene (vE-PE) and cross-linked polyethylene (XLPE) acetabular component liners and between 32 and 36 mm head sizes at the ten-year follow-up. Secondary outcomes included acetabular component migration and patient-reported outcome measures (PROMs) such as the EuroQol five-dimension questionnaire, 36-Item Short-Form Health Survey, Harris Hip Score, and University of California, Los Angeles Activity Scale (UCLA). A single-blinded, multi-arm, 2 × 2 factorial randomized controlled trial was undertaken. Patients were recruited between May 2009 and April 2011. Radiostereometric analyses (RSAs) were performed from baseline to ten years. Of the 220 eligible patients, 116 underwent randomization, and 82 remained at the ten-year follow-up. Eligible patients were randomized into one of four interventions: vE-PE acetabular liner with either 32 or 36 mm femoral head, and XLPE acetabular liner with either 32 or 36 mm femoral head. Parameters were otherwise identical except for acetabular liner material and femoral head size.Aims
Methods
Developmental dysplasia of the hip (DDH) describes a pathological relationship between the femoral head and acetabulum. Periacetabular osteotomy (PAO) may be used to treat this condition. The aim of this study was to evaluate the results of PAO in adolescents and adults with persistent DDH. Patients were divided into four groups: A, adolescents who had not undergone surgery for DDH in childhood (25 hips); B, adolescents who had undergone surgery for DDH in childhood (20 hips); C, adults with DDH who had not undergone previous surgery (80 hips); and D, a control group of patients with healthy hips (70 hips). The radiological evaluation of digital anteroposterior views of hips included the Wiberg angle (centre-edge angle (CEA)), femoral head cover (FHC), medialization, distalization, and the ilioischial angle. Clinical assessment involved the Harris Hip Score (HHS) and gluteal muscle performance assessment.Aims
Methods
A significant reduction in wear at five and ten years was previously reported when comparing Durasul highly cross-linked polyethylene with nitrogen-sterilized Sulene polyethylene in total hip arthroplasty (THA). We investigated whether the improvement observed at the earlier follow-up continued, resulting in decreased osteolysis and revision surgery rates over the second decade. Between January 1999 and December 2001, 90 patients underwent surgery using the same acetabular and femoral components with a 28 mm metallic femoral head and either a Durasul or Sulene liner. A total of 66 hips of this prospective randomized study were available for a minimum follow-up of 20 years. The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Dorr method on digitized radiographs with a software package.Aims
Methods
This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA). Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up.Aims
Methods
In this in vitro study of the
hip joint we examined which soft tissues act as primary and secondary
passive rotational restraints when the hip joint is functionally
loaded. A total of nine cadaveric left hips were mounted in a testing
rig that allowed the application of forces, torques and rotations
in all six degrees of freedom. The hip was rotated throughout a
complete range of movement (ROM) and the contributions of the iliofemoral
(medial and lateral arms), pubofemoral and ischiofemoral ligaments
and the ligamentum teres to rotational restraint was determined
by resecting a ligament and measuring the reduced torque required
to achieve the same angular position as before resection. The contribution
from the acetabular labrum was also measured. Each of the capsular
ligaments acted as the primary hip rotation restraint somewhere
within the complete ROM, and the ligamentum teres acted as a secondary
restraint in high flexion, adduction and external rotation. The
iliofemoral lateral
The purpose of this study is to evaluate early outcomes with the use of a smartphone-based exercise and educational care management system after total hip arthroplasty (THA) and demonstrate decreased use of in-person physiotherapy (PT). A multicentre, prospective randomized controlled trial was conducted to evaluate a smartphone-based care platform for primary THA. Patients randomized to the control group (198) received the institution’s standard of care. Those randomized to the treatment group (167) were provided with a smartwatch and smartphone application. PT use, THA complications, readmissions, emergency department/urgent care visits, and physician office visits were evaluated. Outcome scores include the Hip disability and Osteoarthritis Outcome Score (HOOS, JR), health-related quality-of-life EuroQol five-dimension five-level score (EQ-5D-5L), single leg stance (SLS) test, and the Timed Up and Go (TUG) test.Aims
Methods
We analysed one surgeon’s attempt to reconstruct the hip in 66 patients (84 hips) with chronic dislocation and to restore the height of the centre of rotation above the transverse teardrop line, the bodyweight lever
A typical pattern of blood loss associated with total hip arthroplasty (THA) is 200 ml intraoperatively and 1.3 l in the first 48 postoperative hours. Tranexamic acid (TXA) is most commonly given as a single preoperative dose only and is often withheld from patients with a history of thromboembolic disease as they are perceived to be “high-risk” with respect to postoperative venous thromboembolism (VTE). The TRanexamic ACid for 24 hours trial (TRAC-24) aimed to identify if an additional 24-hour postoperative TXA regime could further reduce blood loss beyond a once-only dose at the time of surgery, without excluding these high-risk patients. TRAC-24 was a prospective, phase IV, single centre, open label, parallel group, randomized controlled trial (RCT) involving patients undergoing primary unilateral elective THA. The primary outcome measure was the indirect calculated blood loss (IBL) at 48 hours. The patients were randomized into three groups. Group 1 received 1 g intravenous (IV) TXA at the time of surgery and an additional oral regime for 24 hours postoperatively, group 2 only received the intraoperative dose, and group 3 did not receive any TXA.Aims
Methods
The current study aimed to compare robotic arm-assisted (RA-THA), computer-assisted (CA-THA), and manual (M-THA) total hip arthroplasty regarding in-hospital metrics including length of stay (LOS), discharge disposition, in-hospital complications, and cost of RA-THA versus M-THA and CA-THA versus M-THA, as well as trends in use and uptake over a ten-year period, and future projections of uptake and use of RA-THA and CA-THA. The National Inpatient Sample was queried for primary THAs (2008 to 2017) which were categorized into RA-THA, CA-THA, and M-THA. Past and projected use, demographic characteristics distribution, income, type of insurance, location, and healthcare setting were compared among the three cohorts. In-hospital complications, LOS, discharge disposition, and in-hospital costs were compared between propensity score-matched cohorts of M-THA versus RA-THA and M-THA versus CA-THA to adjust for baseline characteristics and comorbidities.Aims
Methods
We performed 52 total hip replacements in 52 patients using a cementless acetabular component combined with a circumferential osteotomy of the medial acetabular wall for the late sequelae of childhood septic arthritis of the hip. The mean age of the patients at operation was 44.5 years (22 to 66) and the mean follow-up was 7.8 years (5 to 11.8). The mean improvement in the Harris Hip Score was 29.6 points (19 to 51) at final follow-up. The mean cover of the acetabular component was 98.5% (87.8% to 100%). The medial acetabular wall was preserved with a mean thickness of 8.3 mm (1.7 to 17.4) and the mean length of abductor lever
The direct anterior approach (DAA) for total hip arthroplasty (THA) has potential advantages over other approaches and is most commonly performed with the patient in the supine position. We describe a technique for DAA THA with the patient in the lateral decubitus position and report the early clinical and radiological outcomes, the characteristics of the learning curve, and perioperative complications. All primary DAA THAs performed in the lateral position by a single surgeon over a four-year period from the surgeon’s first case using the technique were identified from a prospectively collected database. Modified Harris Hip Scores (mHHS) were collected to assess clinical outcome, and routine radiological analysis was performed. Retrospective review of the medical records identified perioperative complications, the characteristics of the learning curve, and revisions.Aims
Methods
Adverse local tissue reactions associated with abnormal wear considerably slowed down the general use of metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), now limited to a few specialized centres. In this study, we provide the clinical results of 400 consecutive MoM HRAs implanted more than 20 years ago in one such centre. A total of 355 patients (400 hips) were treated with Conserve Plus HRA between November 1996 and November 2000. There were 96 female (27%) and 259 male patients (73%). Their mean age was 48.2 years (SD 10.9). The University of California, Los Angeles (UCLA) hip scores and 12-item Short Form Survey (SF-12) quality of life scores were reported. Survivorship was assessed using Kaplan-Meier analyses.Aims
Methods
Uncemented metal acetabular components show good osseointegration, but material stiffness causes stress shielding and retroacetabular bone loss. Cemented monoblock polyethylene components load more physiologically; however, the cement bone interface can suffer fibrous encapsulation and loosening. It was hypothesized that an uncemented titanium-sintered monoblock polyethylene component may offer the optimum combination of osseointegration and anatomical loading. A total of 38 patients were prospectively enrolled and received an uncemented monoblock polyethylene acetabular (pressfit) component. This single cohort was then retrospectively compared with previously reported randomized cohorts of cemented monoblock (cemented) and trabecular metal (trabecular) acetabular implants. The primary outcome measure was periprosthetic bone density using dual-energy x-ray absorptiometry over two years. Secondary outcomes included radiological and clinical analysis.Aims
Methods
The most frequent indication for revision surgery in total hip arthroplasty (THA) is aseptic loosening. Aseptic loosening is associated with polyethylene liner wear, and wear may be reduced by using vitamin E-doped liners. The primary objective of this study was to compare proximal femoral head penetration into the liner between a) two cross-linked polyethylene (XLPE) liners (vitamin E-doped (vE-PE)) versus standard XLPE liners, and b) two modular femoral head diameters (32 mm and 36 mm). Patients scheduled for a THA were randomized to receive a vE-PE or XLPE liner with a 32 mm or 36 mm metal head (four intervention groups in a 2 × 2 factorial design). Head penetration and acetabular component migration were measured using radiostereometric analysis at baseline, three, 12, 24, and 60 months postoperatively. The Harris Hip Score, University of California, Los Angeles (UCLA) Activity Score, EuroQol five-dimension questionnaire (EQ-5D), and 36-Item Short-Form Health Survey questionnaire (SF-36) were assessed at baseline, three, 12, 36, and 60 months.Aims
Methods
Despite few good-quality studies on the subject, total hip arthroplasty (THA) is increasingly being performed for displaced intracapsular fractures of the neck of femur. We compared outcomes of all patients with displacement of these fractures treated surgically over a ten-year period in one institution. A total of 2,721 patients with intracapsular fractures of the femoral neck treated with either a cemented hemiarthroplasty or a THA at a single centre were retrospectively reviewed. The primary outcomes analyzed were readmission for any reason and revision surgery. We secondarily looked at mortality rates.Aims
Methods