We undertook a retrospective study investigating
the accuracy and safety of percutaneous pedicle screws placed under
fluoroscopic guidance in the lumbosacral junction and lumbar spine.
The CT scans of patients were chosen from two centres: European
patients from University Medical Center Hamburg-Eppendorf, Germany,
and Asian patients from the University of Malaya, Malaysia. Screw
perforations were classified into grades 0, 1, 2 and 3. A total
of 880 percutaneous pedicle screws from 203 patients were analysed:
614 screws from 144 European patients and 266 screws from 59 Asian
patients. The mean age of the patients was 58.8 years (16 to 91)
and there were 103 men and 100 women. The total rate of perforation
was 9.9% (87 screws) with 7.4% grade 1, 2.0% grade 2 and 0.5% grade
3 perforations. The rate of perforation in Europeans was 10.4% and
in Asians was 8.6%, with no significant difference between the two
(p = 0.42). The rate of perforation was the highest in S1 (19.4%)
followed by L5 (14.9%). The accuracy and safety of percutaneous
pedicle screw placement are comparable to those cited in the literature
for the open method of pedicle screw placement. Greater caution
must be taken during the insertion of L5 and S1 percutaneous pedicle
screws owing to their more angulated pedicles, the anatomical variations
in their vertebral bodies and the morphology of the spinal canal
at this location. Cite this article:
Percutaneous placement of pedicle screws is a
well-established technique, however, no studies have compared percutaneous
and open placement of screws in the thoracic spine. The aim of this
cadaveric study was to compare the accuracy and safety of these
techniques at the thoracic spinal level. A total of 288 screws were
inserted in 16 (eight cadavers, 144 screws in percutaneous and eight
cadavers, 144 screws in open). Pedicle perforations and fractures
were documented subsequent to wide laminectomy followed by skeletalisation
of the vertebrae. The perforations were classified as grade 0: no
perforation, grade 1: <
2 mm perforation, grade 2: 2 mm to 4
mm perforation and grade 3: >
4 mm perforation. In the percutaneous
group, the perforation rate was 11.1% with 15 (10.4%) grade 1 and
one (0.7%) grade 2 perforations. In the open group, the perforation
rate was 8.3% (12 screws) and all were grade 1. This difference
was not significant (p = 0.45). There were 19 (13.2%) pedicle fractures
in the percutaneous group and 21 (14.6%) in the open group (p =
0.73). In summary, the safety of percutaneous fluoroscopy-guided
pedicle screw placement in the thoracic spine between T4 and T12
is similar to that of the conventional open technique. Cite this article:
Aims. We compared the accuracy, operating time and radiation exposure
of the introduction of iliosacral screws using O-arm/Stealth Navigation
and standard fluoroscopy. Materials and Methods. Iliosacral screws were introduced percutaneously into the first
sacral body (S1) of ten human cadavers, four men and six women.
The mean age was 77 years (58 to 85). Screws were introduced using
a standard technique into the left side of S1 using C-Arm fluoroscopy
and then into the right side using O-Arm/Stealth Navigation. The
radiation was measured on the surgeon by dosimeters placed under
a lead thyroid shield and apron, on a finger, a hat and on the cadavers. Results. There were no neuroforaminal
We prospectively studied the use of intercostal EMG monitoring as an indicator of the accuracy of the placement of pedicle screws in the thoracic spine. We investigated 95 thoracic pedicles in 17 patients. Before insertion of the screw, the surgeon recorded his assessment of the integrity of the pedicle track. We then stimulated the track using a K-wire pedicle probe connected to a constant current stimulator. A compound muscle action potential (CMAP) was recorded from the appropriate intercostal or abdominal muscles. Postoperative CT was performed to establish the position of the screw. The stimulus intensity required to evoke a muscle response was correlated with the position of the screw on the CT scan. There were eight unrecognised
Between December 2004 and June 2006, 136 patients (156 total hip replacements), were sent from the waiting list of the Cardiff Vale NHS Trust to the NHS Treatment Centre, Weston-super-Mare, in an attempt to reduce the waiting time for total hip replacement. Because of concerns about their outcome, each patient was contacted and invited to attend a review appointment with a consultant specialising in hip and revision hip replacement. A total of 98 patients (113 hips) were reviewed after a mean of 23 months (11 to 30). There were 104 cemented hips, seven hybrid and two cementless. An acetabular inclination of >
55° was seen in 18 (16%). Radiolucent lines around the acetabular component were seen in 76 (67%). The femoral component was in more than 4° of varus in 47 (42%). The medial floor had been
The aim of this study was to assess the accuracy of pedicle screw placement, as well as intraoperative factors, radiation exposure, and complication rates in adult patients with degenerative disorders of the thoracic and lumbar spines who have undergone robotic-navigated spinal surgery using a contemporary system. The authors reviewed the prospectively collected data on 196 adult patients who had pedicle screws implanted with robot-navigated assistance (RNA) using the Mazor X Stealth system between June 2019 and March 2022. Pedicle screws were implanted by one experienced spinal surgeon after completion of a learning period. The accuracy of pedicle screw placement was determined using intraoperative 3D fluoroscopy.Aims
Methods
In the UK, the agricultural, military, and construction sectors have stringent rules about the use of hearing protection due to the risk of noise-induced hearing loss. Orthopaedic staff may also be at risk due to the use of power tools. The UK Health and Safety Executive (HSE) have clear standards as to what are deemed acceptable occupational levels of noise on A-weighted and C-weighted scales. The aims of this review were to assess the current evidence on the testing of exposure to noise in orthopaedic operating theatres to see if it exceeds these regulations. A search of PubMed and EMBASE databases was conducted using PRISMA guidelines. The review was registered prospectively in PROSPERO. Studies which assessed the exposure to noise for orthopaedic staff in operating theatres were included. Data about the exposure to noise were extracted from these studies and compared with the A-weighted and C-weighted acceptable levels described in the HSE regulations.Aims
Methods
Conventional amputation prostheses rely on the attachment of the socket to the stump, which may lead to soft-tissue complications. Intraosseous transcutaneous amputation prostheses (ITAPs) allow direct loading of the skeleton, but their success is limited by infection resulting from
The aim of this study was to reassess the rate of neurological, psoas-related, and abdominal complications associated with L4-L5 lateral lumbar interbody fusion (LLIF) undertaken using a standardized preoperative assessment and surgical technique. This was a multicentre retrospective study involving consecutively enrolled patients who underwent L4-L5 LLIF by seven surgeons at seven institutions in three countries over a five-year period. The demographic details of the patients and the details of the surgery, reoperations and complications, including femoral and non-femoral neuropraxia, thigh pain, weakness of hip flexion, and abdominal complications, were analyzed. Neurological and psoas-related complications attributed to LLIF or posterior instrumentation and persistent symptoms were recorded at one year postoperatively.Aims
Methods
The aim of this study was to examine the implant accuracy of custom-made partial pelvis replacements (PPRs) in revision total hip arthroplasty (rTHA). Custom-made implants offer an option to achieve a reconstruction in cases with severe acetabular bone loss. By analyzing implant deviation in CT and radiograph imaging and correlating early clinical complications, we aimed to optimize the usage of custom-made implants. A consecutive series of 45 (2014 to 2019) PPRs for Paprosky III defects at rTHA were analyzed comparing the preoperative planning CT scans used to manufacture the implants with postoperative CT scans and radiographs. The anteversion (AV), inclination (IC), deviation from the preoperatively planned implant position, and deviation of the centre of rotation (COR) were explored. Early postoperative complications were recorded, and factors for malpositioning were sought. The mean follow-up was 30 months (SD 19; 6 to 74), with four patients lost to follow-up.Aims
Methods
The use of autograft bone is the best option
when undertaking a procedure that requires bone graft because it
is osteogenic, osteoconductive and osseo-inductive. Pain, morbidity
and complications associated with harvesting iliac or non-iliac
sites occur in between 6% and 30% of cases. An alternative source
of graft with possibly a lower morbidity is the intramedullary canal.
In this study, 28 patients undergoing 30 arthrodesis procedures
on the hindfoot had a mean of 48 cm. 3 . (43 to 50) of bone
harvested locally from the hindfoot or the tibial shaft by antegrade or
retrograde reaming. No patient sustained a fracture of the calcaneum,
talus or tibia. There was no morbidity except for one complication
when the reamer
1. A case of spontaneous fracture of a first rib is described. 2. Its evolution from a previously intact rib through the stage of a "linear" crack to pseudarthrosis has been traced. 3. An older pseudarthrosis was present on the opposite side. 4. Alderson's observation is confirmedâthat the
A modified form of the collateral operation originally described by Dillwyn Evans has been used in severe relapsed club feet to correct the sagittally
1. Fifteen patients with spina bifida occulta are described in whom the fifth lumbar spinous process was pressing on the fibrous membrane that closes the spina bifida, or on the bony stumps of the defective lamina. 2. That this is a cause of low back pain, with or without pain referred to the lower limbs, is supported by the findings at operation and the results of treatment. 3. Eleven patients were treated by operation; in each the spinous process of the fifth lumbar vertebra pressed against the spina bifida, either directly or through the vestigial remnant of the first sacral spinous process. 4. In eight cases the fibrous membrane was adherent to the dura mater; in one case the nerve roots were adherent. 5. Excision of the fifth lumbar spinous process and the membrane across the
This study estimated trends in incidence of open fractures and the adherence to clinical standards for open fracture care in England. Longitudinal data collected by the Trauma Audit and Research Network were used to identify 38,347 patients with open fractures, and a subgroup of 12,170 with severe open fractures of the tibia, between 2008 and 2019 in England. Incidence rates per 100,000 person-years and 95% confidence intervals were calculated. Clinical care was compared with the British Orthopaedic Association Standards for Trauma and National Major Trauma Centre audit standards.Aims
Methods
The treatment of osteochondritis dissecans (OCD) in children and adolescents is determined by the stability of the lesion and the state of the overlying cartilage. MRI has been advocated as an accurate way of assessing and staging such lesions. Our aim was to determine if MRI scans accurately predicted the subsequent arthroscopic findings in adolescents with OCD of the knee. Some authors have suggested that a high signal line behind a fragment on the T2-weighted image indicates the presence of synovial fluid and is a sign of an unstable lesion. More recent reports have suggested that this high signal line is due to the presence of vascular granulation tissue and may represent a healing reaction. We were able to improve the accuracy of MRI for staging the OCD lesion from 45% to 85% by interpreting the high signal T2 line as a predictor of instability only when it was accompanied by a
The aim of this retrospective study was to compare the correction achieved using a convex pedicle screw technique and a low implant density achieved using periapical concave-sided screws and a high implant density. We hypothesized that there would be no difference in outcome between the two techniques. We retrospectively analyzed a series of 51 patients with a thoracic adolescent idiopathic scoliosis. There were 26 patients in the convex pedicle screw group who had screws implanted periapically (Group 2) and a control group of 25 patients with bilateral pedicle screws (Group 1). The patients’ charts were reviewed and pre- and postoperative radiographs evaluated. Postoperative patient-reported outcome measures (PROMs) were recorded.Aims
Methods
The use and variety of stemless humeral components in anatomical total shoulder arthroplasty (TSA) have proliferated since their advent in 2004. Early outcomes are reassuring but independent mid-term results are scarce. This independent study reports a consecutive series of 143 Eclipse stemless shoulder prostheses with a minimum five-year (5 to 10) follow-up. Outcomes of 143 procedures undertaken for all indications in 131 patients were reviewed, with subset analysis of those for osteoarthritis (OA) (n = 99). The primary outcome was the Oxford Shoulder Score (OSS) at a minimum of five years. Secondary outcomes were ranges of motion and radiological analysis of humeral radiolucency, rotator cuff failure, and glenoid loosening.Aims
Methods
Deep surgical site infection (SSI) is common after lower limb fracture. We compared the diagnosis of deep SSI using alternative methods of data collection and examined the agreement of clinical photography and in-person clinical assessment by the Centers for Disease Control and Prevention (CDC) criteria after lower limb fracture surgery. Data from two large, UK-based multicentre randomized controlled major trauma trials investigating SSI and wound healing after surgical repair of open lower limb fractures that could not be primarily closed (UK WOLLF), and surgical incisions for fractures that were primarily closed (UK WHiST), were examined. Trial interventions were standard wound care management and negative pressure wound therapy after initial surgical debridement. Wound outcomes were collected from 30 days to six weeks. We compared the level of agreement between wound photography and clinical assessment of CDC-defined SSI. We are also assessed the level of agreement between blinded independent assessors of the photographs.Aims
Methods