Between 1982 and 1997, 403 consecutive patients (522 elbows) with rheumatoid arthritis underwent Souter-Strathclyde total elbow replacement. By the end of 2007, there had been 66 revisions for aseptic loosening in 60 patients. The mean time of follow-up was 10.6 years (0 to 25) The survival rates at five-, ten, 15 and 19 years were 96% (95%, confidence interval (CI) 95 to 98), 89% (95% CI 86 to 92), 83% (95% CI 78 to 87), and 77% (95% CI 69 to 85), respectively. The small and medium-sized short-stemmed primary humeral components had a 5.6-fold and 3.6-fold risk of revision for aseptic loosening respectively, compared to the medium-sized long-stemmed component. The small and medium-sized all-polyethylene ulnar components had respectively a 28.2-fold and 8.4-fold risk of revision for aseptic loosening, compared to the metal-backed ulnar components. The use of retentive ulnar components was not associated with an increased risk of aseptic loosening compared to non-retentive implants.
We analysed the outcome of the Coonrad-Morrey total elbow replacement used for fracture of the distal humerus in elderly patients with no evidence of inflammatory arthritis and compared the results for early One patient in the early group developed chronic regional pain syndrome and another type 4 aseptic loosening. Two elbows in the early group also showed type 1 radiological loosening. Two patients in the delayed group had an infection, two an ulnar nerve palsy, one developed heterotopic ossification and one type 4 aseptic loosening. Two elbows in this group also showed type 1 radiological loosening. The Kaplan-Meier survivorship analysis for the early and delayed treatment groups was 93% at 88 months and 76% at 84 months, respectively. No significant difference was found between the two groups.
We divided 309 patients with an inflammatory arthritis who had undergone primary elbow replacement using the Souter-Strathclyde implant into two groups according to their age. The mean follow-up in the older group (mean age 64 years) was 7.3 years while in the younger patients (mean age 42 years) it was 12 years. Survivorship for three different failure end-points (revision, revision because of aseptic loosening of the humeral component, and gross loosening of the humeral implant), was compared in both groups. Our findings showed that there was no significant difference in the incidence of loosening when young rheumatoid patients were compared with an older age group.