Aims. Joint-preserving surgery of the hip (JPSH) has evolved considerably
and now includes a number of procedures, including arthroscopy,
surgical dislocation, and redirectional osteotomies of the femur
and acetabulum. There are a number of different factors which lead
to failure of JPSH. Consequently, it is of interest to assess the
various
Peri-prosthetic femoral fracture after total hip replacement (THR) is associated with a poor outcome and high mortality. However, little is known about its long-term incidence after uncemented THR. We retrospectively reviewed a consecutive series of 326 patients (354 hips) who had received a CLS Spotorno replacement with an uncemented, straight, collarless tapered titanium stem between January 1985 and December 1989. The mean follow-up was 17 years (15 to 20). The occurrence of peri-prosthetic femoral fracture during follow-up was noted. Kaplan-Meier survival analysis was used to estimate the cumulative incidence of fracture. At the last follow-up, 86 patients (89 hips) had died and eight patients (eight hips) had been lost to follow-up. A total of 14 fractures in 14 patients had occurred. In ten hips, the femoral component had to be revised and in four the fracture was treated by open reduction and internal fixation. The cumulative incidence of peri-prosthetic femoral fracture was 1.6% (95% confidence interval 0.7 to 3.8) at ten years and 4.5% (95% confidence interval 2.6 to 8.0) at 17 years after the primary THR. There was no association between the occurrence of fracture and gender or age at the time of the primary replacement. Our findings indicate that peri-prosthetic femoral fracture is a significant
Despite the increasing interest and subsequent published literature on hip resurfacing arthroplasty, little is known about the prevalence of its complications and in particular the less common
Aims. The aim of this study was to estimate the 90-day risk of revision for periprosthetic femoral fracture associated with design features of cementless femoral stems, and to investigate the effect of a collar on this risk using a biomechanical in vitro model. Materials and Methods. A total of 337 647 primary total hip arthroplasties (THAs) from the United Kingdom National Joint Registry (NJR) were included in a multivariable survival and regression analysis to identify the adjusted hazard of revision for periprosthetic fracture following primary THA using a cementless stem. The effect of a collar in cementless THA on this risk was evaluated in an in vitro model using paired fresh frozen cadaveric femora. Results. The prevalence of early revision for periprosthetic fracture was 0.34% (1180/337 647) and 44.0% (520/1180) occurred within 90 days of surgery. Implant risk factors included: collarless stem, non-grit-blasted finish, and triple-tapered design. In the in vitro model, a medial calcar collar consistently improved the stability and resistance to fracture. Conclusion. Analysis of features of stem design in registry data is a useful method of identifying implant characteristics that affect the risk of early periprosthetic fracture around a cementless femoral stem. A collar on the calcar reduced the risk of an early periprosthetic fracture and this was confirmed by biomechanical testing. This approach may be useful in the analysis of other uncommon
Aims. This study presents the long-term survivorship, risk factors for prosthesis survival, and an assessment of the long-term effects of changes in surgical technique in a large series of patients treated by metal-on-metal (MoM) hip resurfacing arthroplasty (HRA). Patients and Methods. Between November 1996 and January 2012, 1074 patients (1321 hips) underwent HRA using the Conserve Plus Hip Resurfacing System. There were 787 men (73%) and 287 women (27%) with a mean age of 51 years (14 to 83). The underlying pathology was osteoarthritis (OA) in 1003 (75.9%), developmental dysplasia of the hip (DDH) in 136 (10.3%), avascular necrosis in 98 (7.4%), and other conditions, including inflammatory arthritis, in 84 (6.4%). Results. The mean follow-up time was 10.5 years (1 to 20). Using revision for any reason as the endpoint, the overall survivorship at 15 years was 89.4% (95% confidence interval (CI) 86.8 to 91.4). There was a substantial increase between the first and second generation of surgical technique (86.6% vs 90.1%; p = 0.05). Men with idiopathic OA had a 15-year survivorship of 94.5% and women, 82.2% (p = 0.001); gender was not a risk factor after stratification by component size and aetiology. Using revision for excessive wear (ion levels > 7 µg/l associated with symptoms or adverse local tissue reactions) as the endpoint, the 15-year survivorship was 98.5%. Risk factors for revision for all
Aims. The Corail stem has good long-term results. After four years
of using this stem, we have detected a small group of patients who
have presented with symptomatic metaphyseal debonding. The aim of
this study was to quantify the incidence of this complication, to
delineate the characteristics of patients presenting with this complication
and to compare these patients with asymptomatic controls to determine
any important predisposing factors. Patients and Methods. Of 855 Corail collarless cementless stems implanted for osteoarthritis,
18 presented with symptomatic metaphyseal debonding. A control group
of 74 randomly selected patients was assembled. Clinical and radiological
parameters were measured and a logistic regression model was created
to evaluate factors associated with metaphyseal debonding. Results. The prevalence of this complication was 2.1% in our series. In
the multivariable model, the presence of a Dorr B-type proximal
femur was associated with metaphyseal debonding (odds ratio (OR)
10.73, 95% confidence interval (CI) 2.31 to 49.97, p = 0.002), as
was a body mass index >
25 kg/m. 2. (OR 6.85, 95% CI 1.06
to 44.28, p = 0.04). Smaller stems and the use of a polyethylene
acetabular liner appeared to be protective when compared with metal
and ceramic setting hard-on-hard bearings. Conclusion. We have described an uncommon but important
We systematically reviewed the peer-reviewed literature to relate the survival of hybrid metal-on-metal hip resurfacing arthroplasty devices to a National Institute of Clinical Excellence (NICE) benchmark for choosing a primary total hip replacement, which is a survival rate of 90% at a follow-up of ten years. A total of 29 articles (10 621 resurfaced hips) met the inclusion criteria. The mean follow-up ranged from 0.6 to 10.5 years and the survival of the implant ranged from 84% to 100%. Of the 10 621 hips, 370 were revised (3.5%), with aseptic loosening as the most frequent
We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by
Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without modular DM options. In these circumstances, a monoblock DM cup, designed for cementing, can be cemented into an uncemented acetabular shell. The goal of this study was to describe the implant survival, complications, and radiological outcomes of this construct. We identified 64 patients (65 hips) who had a single-design cemented DM cup cemented into an uncemented acetabular shell during revision THA between 2018 and 2020 at our institution. Cups were cemented into either uncemented cups designed for liner cementing (n = 48; 74%) or retained (n = 17; 26%) acetabular components. Median outer head diameter was 42 mm. Mean age was 69 years (SD 11), mean BMI was 32 kg/m2 (SD 8), and 52% (n = 34) were female. Survival was assessed using Kaplan-Meier methods. Mean follow-up was two years (SD 0.97).Aims
Methods
The number of revision arthroplasties being performed in the elderly is expected to rise, including revision for infection. The primary aim of this study was to measure the treatment success rate for octogenarians undergoing revision total hip arthroplasty (THA) for periprosthetic joint infection (PJI) compared to a younger cohort. Secondary outcomes were complications and mortality. Patients undergoing one- or two-stage revision of a primary THA for PJI between January 2008 and January 2021 were identified. Age, sex, BMI, American Society of Anesthesiologists grade, Charlson Comorbidity Index (CCI), McPherson systemic host grade, and causative organism were collated for all patients. PJI was classified as ‘confirmed’, ‘likely’, or ‘unlikely’ according to the 2021 European Bone and Joint Infection Society criteria. Primary outcomes were complications, reoperation, re-revision, and successful treatment of PJI. A total of 37 patients aged 80 years or older and 120 patients aged under 80 years were identified. The octogenarian group had a significantly lower BMI and significantly higher CCI and McPherson systemic host grades compared to the younger cohort.Aims
Methods
Radiostereometric analysis (RSA) can detect early
micromovement in unstable implant designs which are likely subsequently
to have a high failure rate. In 2010, the Articular Surface Replacement
(ASR) was withdrawn because of a high failure rate. In 19 ASR femoral
components, the mean micromovement over the first two years after implantation
was 0.107 mm (. sd. 0.513) laterally, 0.055 mm (. sd. 0.204)
distally and 0.150 mm (. sd. 0.413) anteriorly. The mean backward
tilt around the x-axis was -0.08° (. sd. 1.088), mean internal
rotation was 0.165° (. sd. 0.924) and mean varus tilt 0.238°
(. sd. 0.420). The baseline to two-year varus tilt was statistically
significant from zero movement, but there was no significant movement
from one year onwards. We conclude that the ASR femoral component achieves initial stability
and that early migration is not the
We retrospectively reviewed 44 consecutive patients
(50 hips) who underwent acetabular re-revision after a failed previous
revision that had been performed using structural or morcellised
allograft bone, with a cage or ring for uncontained defects. Of
the 50 previous revisions, 41 cages and nine rings were used with
allografts for 14 minor-column and 36 major-column defects. We routinely
assessed the size of the acetabular bone defect at the time of revision
and re-revision surgery. This allowed us to assess whether host
bone stock was restored. We also assessed the outcome of re-revision
surgery in these circumstances by means of radiological characteristics,
rates of
Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors. Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti12Mo6Zr2Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker). Uncemented acetabular components (Trident; Stryker) with crosslinked polyethylene liners and CoCr femoral heads of 36 mm diameter were used in all patients. At a mean folllow-up of 10.8 years (SD 1.1), 94 patients (79%) were eligible for follow-up (six patients had already undergone a revision, 15 had died, and six were lost to follow-up). A total of 85 THAs in 80 patients (mean age 61 years (24 to 75); 47 (56%) were female) underwent clinical and radiological evaluation, including the measurement of whole blood levels of cobalt and chrome. Metal artifact reduction sequence MRI scans of the hip were performed in 71 patients.Aims
Methods
Between March 1990 and May 1991 we performed 85 primary total hip replacements in 74 patients using the Landos Atoll hydroxyapatite (HA)-coated cup and the Corail HA-coated stem. The patients were followed up for a mean of ten years. Of the 85 cups, 26 (31%) have already been revised and a further six are radiologically unstable and awaiting revision. Two femoral stems have been revised for infection without loosening. The retrieved acetabular cups were studied by SEM and image-processing techniques to quantify the amount of residual HA on the cup. This was correlated with the clinical variables and
Total hip arthroplasties (THAs) are performed by surgeons at various stages in training with varying levels of supervision, but we do not know if this is safe practice with comparable outcomes to consultant-performed THA. Our aim was to examine the association between surgeon grade, the senior supervision of trainees, and the risk of revision following THA. We performed an observational study using National Joint Registry (NJR) data. We included adult patients who underwent primary THA for osteoarthritis, recorded in the NJR between 2003 and 2016. Exposures were operating surgeon grade (consultant or trainee) and whether or not trainees were directly supervised by a scrubbed consultant. Outcomes were all-cause revision and the indication for revision up to ten years. We used methods of survival analysis, adjusted for patient, operation, and healthcare setting factors.Aims
Methods
We aimed to compare the implant survival, complications, readmissions, and mortality of Vancouver B2 periprosthetic femoral fractures (PFFs) treated with internal fixation with that of B1 PFFs treated with internal fixation and B2 fractures treated with revision arthroplasty. We retrospectively reviewed the data of 112 PFFs, of which 47 (42%) B1 and 27 (24%) B2 PFFs were treated with internal fixation, whereas 38 (34%) B2 fractures underwent revision arthroplasty. Decision to perform internal fixation for B2 PFFs was based on specific radiological (polished femoral components, intact bone-cement interface) and clinical criteria (low-demand patient). Median follow-up was 36.4 months (24 to 60). Implant survival and mortality over time were estimated with the Kaplan-Meier method. Adverse events (measured with a modified Dindo-Clavien classification) and 90-day readmissions were additionally compared between groups.Aims
Methods
Aseptic loosening is a leading cause of uncemented arthroplasty failure, often accompanied by fibrotic tissue at the bone-implant interface. A biological target, neutrophil extracellular traps (NETs), was investigated as a crucial connection between the innate immune system’s response to injury, fibrotic tissue development, and proper bone healing. Prevalence of NETs in peri-implant fibrotic tissue from aseptic loosening patients was assessed. A murine model of osseointegration failure was used to test the hypothesis that inhibition (through Patient peri-implant fibrotic tissue was analyzed for NETs biomarkers. To enhance osseointegration in loose implant conditions, an innate immune system pathway (NETs) was either inhibited (Aims
Methods
The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs). We undertook a registry study combining the National Joint Registry dataset with polyethylene manufacturing characteristics as supplied by the manufacturers. The primary endpoint was revision for any reason. We performed further analyses on other reasons including instability, aseptic loosening, wear, and liner dissociation. The primary analytic approach was Cox proportional hazard regression.Aims
Methods
The aim of this study was to determine whether fixation, as opposed to revision arthroplasty, can be safely used to treat reducible Vancouver B type fractures in association with a cemented collarless polished tapered femoral stem (the Exeter). This retrospective cohort study assessed 152 operatively managed consecutive unilateral Vancouver B fractures involving Exeter stems; 130 were managed with open reduction and internal fixation (ORIF) and 22 with revision arthroplasty. Mean follow-up was 6.5 years (SD 2.6; 3.2 to 12.1). The primary outcome measure was revision of at least one component. Kaplan–Meier survival analysis was performed. Regression analysis was used to identify risk factors for revision following ORIF. Secondary outcomes included any reoperation, complications, blood transfusion, length of hospital stay, and mortality.Aims
Methods
Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components. From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment.Aims
Methods