In a prospective study between 2000 and 2005, 22 patients with primary osteoarthritis of the shoulder had a total shoulder arthroplasty with a standard five-pegged glenoid component, 12 with non-offset humeral head and ten with
There are no long-term published results on the survival of a third-generation cemented total shoulder replacement. We describe a clinical and radiological study of the Aequalis total shoulder replacement for a minimum of ten years. Between September 1996 and May 1998, 39 consecutive patients underwent a primary cemented total shoulder replacement using this prosthesis. Data were collected prospectively on all patients each year, for a minimum of ten years, or until death or failure of the prosthesis. At a follow-up of at least ten years, 12 patients had died with the prosthesis intact and two had emigrated, leaving 25 available for clinical review. Of these, 13 had rheumatoid arthritis and 12 osteoarthritis. One refused radiological review leaving 24 with fresh radiographs. Survivorship at ten years was 100% for the humeral component and 92% for the glenoid component. The incidence of lucent lines was low. No humeral component was thought to be at risk and only two glenoid components. The osteoarthritic group gained a mean 65° in forward flexion and their Constant score improved by a mean 41.4 points (13 to 55). The rheumatoid group gained a mean of 24° in flexion and their Constant score improved by 29.4 points. This difference may have been due to failure of the rotator cuff in 75% of the patients with rheumatoid arthritis. Thus a third-generation total shoulder replacement gives an excellent result in patients with osteoarthritis and an intact rotator cuff. Patients with rheumatoid arthritis have a 75% risk of failure of the rotator cuff at ten years.
Reverse total shoulder replacement (RTSR) depends
on adequate deltoid function for a successful outcome. However, the
anterior deltoid and/or axillary nerve may be damaged due to prior
procedures or injury. The purpose of this study was to determine
the compensatory muscle forces required for scapular plane elevation
following RTSR when the anterior deltoid is deficient. The soft
tissues were removed from six cadaver shoulders, except for tendon
attachments. After implantation of the RTSR, the shoulders were
mounted on a custom-made shoulder simulator to determine the mean
force in each muscle required to achieve 30° and 60° of scapular
plane elevation. Two conditions were tested: 1) Control with an
absent supraspinatus and infraspinatus; and 2) Control with anterior
deltoid deficiency. Anterior deltoid deficiency resulted in a mean
increase of 195% in subscapularis force at 30° when compared with
the control (p = 0.02). At 60°, the subscapularis force increased
a mean of 82% (p <
0.001) and the middle deltoid force increased
a mean of 26% (p = 0.04). Scapular plane elevation may still be possible following an RTSR
in the setting of anterior deltoid deficiency. When the anterior
deltoid is deficient, there is a compensatory increase in the force
required by the subscapularis and middle deltoid. Attempts to preserve
the subscapularis, if present, might maximise post-operative function.
This study provides recommendations on the position
of the implant in reverse shoulder replacement in order to minimise
scapular notching and osteophyte formation. Radiographs from 151
patients who underwent primary reverse shoulder replacement with
a single prosthesis were analysed at a mean follow-up of 28.3 months
(24 to 44) for notching, osteophytes, the position of the glenoid
baseplate, the overhang of the glenosphere, and the prosthesis scapular
neck angle (PSNA). A total of 20 patients (13.2%) had a notch (16 Grade 1 and four
Grade 2) and 47 (31.1%) had an osteophyte. In patients without either
notching or an osteophyte the baseplate was found to be positioned
lower on the glenoid, with greater overhang of the glenosphere and
a lower PSNA than those with notching and an osteophyte. Female patients
had a higher rate of notching than males (13.3% Based on these findings we make recommendations on the placement
of the implant in both male and female patients to avoid notching
and osteophyte formation. Cite this article:
Fractures of the proximal humerus can lead to malalignment of the humeral head, necrosis and post-traumatic osteoarthritis. In such cases surface replacement might be a promising option. A total of 28 shoulders with glenohumeral arthritis subsequent to a fracture underwent surface replacement arthroplasty of the humeral head in patients with a mean age of 60 years (35 to 83). On the basis of the inclination of the impacted head, post-traumatic arthritis was divided into three types: type 1, an impacted fracture of the head in an anatomical position (seven cases); type 2, a valgus impacted fracture (13 cases); type 3, a varus impacted fracture (eight cases). The outcome was measured by means of the Constant score. According to the Boileau classification of the sequelae of fractures of the proximal humerus, all 28 patients had a final result of intra-capsular category 1. The mean Constant score for the 28 shoulders increased from 23.2 points (2 to 45) pre-operatively to 55.1 points (20 to 89) at a mean of 31 months (24 to 66) post-operatively. Valgus impacted fractures had significantly better results (p <
0.039). Surface replacement arthroplasty can provide good results for patients with post-traumatic osteoarthritis of the shoulder. Their use avoids post-operative complications of the humeral shaft, such as peri-prosthetic fractures. Further surgery can be undertaken more easily as the bone stock is preserved.
We assessed the short- to mid-term survival of
metallic press-fit radial head prostheses in patients with radial
head fractures and acute traumatic instability of the elbow. The medical records of 42 patients (16 males, 26 females) with
a mean age of 56 years (23 to 85) with acute unstable elbow injuries,
including a fracture of the radial head requiring metallic replacement
of the radial head, were reviewed retrospectively. Survival of the
prosthesis was assessed from the radiographs of 37 patients after
a mean follow-up of 50 months (12 to 107). The functional results
of 31 patients were assessed using range-of-movement, Mayo elbow
performance score (MEPS), Disabilities of the Arm, Shoulder and
Hand (DASH) score and the RAND 36-item health survey. At the most recent follow-up 25 prostheses were still well fixed,
nine had been removed because of loosening, and three remained implanted
but were loose. The mean time from implantation to loosening was
11 months (2 to 24). Radiolucent lines that developed around the
prosthesis before removal were mild in three patients, moderate
in one and severe in five. Range of movement parameters and mass
grip strength were significantly lower in the affected elbow than
in the unaffected side. The mean MEPS score was 86 (40 to 100) and
the mean DASH score was 23 (0 to 81). According to RAND-36 scores,
patients had more pain and lower physical function scores than normal population
values. Loosening of press-fit radial head prostheses is common, occurs
early, often leads to severe osteolysis of the proximal radius,
and commonly requires removal of the prosthesis.
We have previously described the short-term outcome of the use of reverse shoulder arthroplasty in the treatment of acute complex proximal humeral fractures in the elderly. We now report the clinical and radiological outcome of 36 fractures at a mean of 6.6 years (1 to 16). Previously, at a mean follow-up of 6 years (1 to 12) the mean Constant score was 58.5; this was reduced to 53 points with the further follow-up. A total of 23 patients (63%) had radiological evidence of loosening of the glenoid component. Nevertheless, only one patient had aseptic loosening of the baseplate at 12 years’ follow-up. The reduction in the mean Constant score with longer follow-up and the further development of scapular notching is worrying. New developments in design, bearing surfaces and surgical technique, and further follow-up, will determine whether reverse shoulder arthroplasty has a place in the management of complex proximal humeral fractures in the elderly.
Replacement of the shoulder in juvenile idiopathic arthritis is not often performed and there have been no published series to date. We present nine glenohumeral hemiarthroplasties in eight patients with systemic or polyarticular juvenile idiopathic arthritis. The mean follow-up was six years (59 to 89 months). The mean age at the time of surgery was 32 years. Surgery took place at a mean of 27 years after diagnosis. The results indicated excellent relief from pain. There was restoration of useful function which deteriorated with time, in part because of progression of the systemic disease in this severely affected group. No patient has required revision to date and there has been no radiological evidence of loosening or osteolysis around the implants. We discuss the pathoanatomical challenges unique to this group. There was very little space for a prosthetic joint and, in some cases, bony deformity and the small size necessitated the use of custom-made implants.
There are theoretical and practical advantages to modular rather than monoblock designs of prostheses for shoulder arthroplasty, but there are no reported studies which specifically compare the clinical and radiological results of their use. We have compared the results of unconstrained total shoulder arthroplasty for osteoarthritis using both types of implant. The monoblock design was used between 1992 and 1995 and the modular design after 1995. Both had cemented all-polyethylene glenoids, the monoblock with matched and the modular with mismatched radii of curvature. There were 34 consecutive shoulders in each group with a mean follow-up of 6.1 years in the first and 5.2 years in the second. There were no significant differences in improvement of pain scores, active elevation, external rotation, internal rotation, patient satisfaction, or the Neer ratings between the two groups. Two of 28 glenoid components in the first group and six of 30 in the second met the criteria for being radiologically at risk for loosening (p = 0.25). There were no significant differences in clinical outcome or radiological changes between the first- and second-generation designs of implant for shoulder arthroplasty.