Aims. National joint registries under-report revisions for periprosthetic joint infection (PJI). We aimed to validate PJI reporting to the
Aims. Several different designs of hemiarthroplasty are used to treat intracapsular fractures of the proximal femur, with large variations in costs. No clinical benefit of modular over monoblock designs has been reported in the literature. Long-term data are lacking. The aim of this study was to report the ten-year implant survival of commonly used designs of hemiarthroplasty. Methods. Patients recorded by the
Aims. The aim of this study was to estimate the 90-day periprosthetic joint infection (PJI) rates following total knee arthroplasty (TKA) and total hip arthroplasty (THA) for osteoarthritis (OA). Methods. This was a data linkage study using the New South Wales (NSW) Admitted Patient Data Collection (APDC) and the
Aims. In recent years, the use of a collared cementless femoral prosthesis has risen in popularity. The design intention of collared components is to transfer some load to the resected femoral calcar and prevent implant subsidence within the cancellous bone of the metaphysis. Conversely, the load transfer for a cemented femoral prosthesis depends on the cement-component and cement-bone interface interaction. The aim of our study was to compare the three most commonly used collared cementless components and the three most commonly used tapered polished cemented components in patients aged ≥ 75 years who have undergone a primary total hip arthroplasty (THA) for osteoarthritis (OA). Methods. Data from the
Aims. Adverse spinal motion or balance (spine mobility) and adverse pelvic mobility, in combination, are often referred to as adverse spinopelvic mobility (SPM). A stiff lumbar spine, large posterior standing pelvic tilt, and severe sagittal spinal deformity have been identified as risk factors for increased hip instability. Adverse SPM can create functional malposition of the acetabular components and hence is an instability risk. Adverse pelvic mobility is often, but not always, associated with abnormal spinal motion parameters. Dislocation rates for dual-mobility articulations (DMAs) have been reported to be between 0% and 1.1%. The aim of this study was to determine the early survivorship from the
Aims. The aim of this study was to investigate the rate of revision for distal femoral arthroplasty (DFA) performed as a primary procedure for native knee fractures using data from the
Aims. The mid-term results of kinematic alignment (KA) for total knee arthroplasty (TKA) using image derived instrumentation (IDI) have not been reported in detail, and questions remain regarding ligamentous stability and revisions. This paper aims to address the following: 1) what is the distribution of alignment of KA TKAs using IDI; 2) is a TKA alignment category associated with increased risk of failure or poor patient outcomes; 3) does extending limb alignment lead to changes in soft-tissue laxity; and 4) what is the five-year survivorship and outcomes of KA TKA using IDI?. Methods. A prospective, multicentre, trial enrolled 100 patients undergoing KA TKA using IDI, with follow-up to five years. Alignment measures were conducted pre- and postoperatively to assess constitutional alignment and final implant position. Patient-reported outcome measures (PROMs) of pain and function were also included. The
Aims. This study aimed to describe the use of revision knee arthroplasty in Australia and examine changes in lifetime risk over a decade. Methods. De-identified individual-level data on all revision knee arthroplasties performed in Australia from 2007 to 2017 were obtained from the
Aims. Displaced femoral neck fractures (FNF) may be treated with partial (hemiarthroplasty, HA) or total hip arthroplasty (THA), with recent recommendations advising that THA be used in community-ambulant patients. This study aims to determine the association between the proportion of FNF treated with THA and year of surgery, day of the week, surgeon practice, and private versus public hospitals, adjusting for known confounders. Patients and Methods. Data from 67 620 patients in the
Aims. The aim of this study was to report the implant survival and patient-reported outcome measures (PROMs) in a consecutive series of patients aged less than 50 years at the time of arthroplasty using the Birmingham Hip Resurfacing system (BHR), with a minimum follow-up of ten years. Patients and Methods. A total of 226 patients with osteoarthritis of the hip, who underwent BHR and presented to a single surgeon, were included in the study. Survival of the implant was confirmed by cross-checking with the
Dislocation is a common reason for revision following total hip replacement. This study investigated the relationship between the bearing surface and the risk of revision due to dislocation. It was based on 110 239 primary total hip replacements with a diagnosis of osteoarthritis collected by the
Aim. There has been a significant reduction in unicompartmental knee arthroplasty (UKA) procedures recorded in Australia. This follows several national joint registry studies documenting high UKA revision rates when compared to total knee arthroplasty (TKA). With the recent introduction of robotically assisted UKA procedures, it is hoped that outcomes improve. This study examines the cumulative revision rate of UKA procedures implanted with a newly introduced robotic system and compares the results to one of the best performing non-robotically assisted UKA prostheses, as well as all other non-robotically assisted UKA procedures. Methods. Data from the
Economic evaluation provides a framework for assessing the costs and consequences of alternative programmes or interventions. One common vehicle for economic evaluations in the healthcare context is the decision-analytic model, which synthesizes information on parameter inputs (for example, probabilities or costs of clinical events or health states) from multiple sources and requires application of mathematical techniques, usually within a software program. A plethora of decision-analytic modelling-based economic evaluations of orthopaedic interventions have been published in recent years. This annotation outlines a number of issues that can help readers, reviewers, and decision-makers interpret evidence from decision-analytic modelling-based economic evaluations of orthopaedic interventions. Cite this article:
The aim of this study was to describe and compare joint-specific and generic health-related quality of life outcomes of the first versus second knee in patients undergoing staged bilateral total knee arthroplasty (BTKA) for osteoarthritis. This retrospective cohort study used Australian national arthroplasty registry data from January 2013 to January 2021 to identify participants who underwent elective staged BTKA with six to 24 months between procedures. The primary outcome was Oxford Knee Score (OKS) at six months postoperatively for the first TKA compared to the second TKA, adjusted for age and sex. Secondary outcomes compared six-month EuroQol five-dimension five-level (EQ-5D-5L) domain scores, EQ-5D index scores, and the EQ visual analogue scale (EQ-VAS) between knees at six months postoperatively.Aims
Methods
Our aim was to estimate the total costs of all hospitalizations for treating periprosthetic joint infection (PJI) by main management strategy within 24 months post-diagnosis using activity-based costing. Additionally, we investigated the influence of individual PJI treatment pathways on hospital costs within the first 24 months. Using admission and procedure data from a prospective observational cohort in Australia and New Zealand, Australian Refined Diagnosis Related Groups were assigned to each admitted patient episode of care for activity-based costing estimates of 273 hip PJI patients and 377 knee PJI patients. Costs were aggregated at 24 months post-diagnosis, and are presented in Australian dollars.Aims
Methods
This study aimed to compare the effect of antibiotic-loaded bone cement (ALBC) versus plain bone cement (PBC) on revision rates for periprosthetic joint infection (PJI) and all-cause revisions following primary elective total hip arthroplasty (THA) and total knee arthroplasty (TKA). MEDLINE, Embase, Web of Science, and Cochrane databases were systematically searched for studies comparing ALBC versus PBC, reporting on revision rates for PJI or all-cause revision following primary elective THA or TKA. A random-effects meta-analysis was performed. The study was registered in the International Prospective Register of Systematic Reviews (PROSPERO ID CRD42018107691).Aims
Methods
The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit. Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis.Aims
Methods
The February 2024 Hip & Pelvis Roundup360 looks at: Trial of vancomycin and cefazolin as surgical prophylaxis in arthroplasty; Is preoperative posterior femoral neck tilt a risk factor for fixation failure? Cemented versus uncemented hemiarthroplasty for displaced intracapsular fractures of the hip; Periprosthetic fractures in larger hydroxyapatite-coated stems: are collared stems a better alternative for total hip arthroplasty?; Postoperative periprosthetic fracture following hip arthroplasty with a polished taper slip versus composite beam stem; Is oral tranexamic acid as good as intravenous?; Stem design and the risk of early periprosthetic femur fractures following THA in elderly patients; Does powered femoral broaching compromise patient safety in total hip arthroplasty?
The December 2023 Shoulder & Elbow Roundup360 looks at: Clavicle fractures: is the evidence changing practice?; Humeral shaft fractures, and another meta-analysis…let’s wait for the trials now!; Hemiarthroplasty or total elbow arthroplasty for distal humeral fractures…what does the registry say?; What to do with a first-time shoulder dislocation?; Deprivation indices and minimal clinically important difference for patient-reported outcomes after arthroscopic rotator cuff repair; Prospective randomized clinical trial of arthroscopic repair versus debridement for partial subscapularis tears; Long-term follow-up following closed reduction and early movement for simple dislocation of the elbow; Sternoclavicular joint reconstruction for traumatic acute and chronic anterior and posterior instability.
The aim of this study was to report the long-term follow-up of cemented short Exeter femoral components when used in primary total hip arthroplasty (THA). We included all primary 394 THAs with a cemented short Exeter femoral component (≤ 125 mm) used in our tertiary referral centre between October 1993 and December 2021. A total of 83 patients (21%) were male. The median age of the patients at the time of surgery was 42 years (interquartile range (IQR) 30 to 55). The main indication for THA was a childhood hip disease (202; 51%). The median follow-up was 6.7 years (IQR 3.1 to 11.0). Kaplan-Meier survival analyses were performed to determine the rates of survival with femoral revision for any indication, for septic loosening, for fracture of the femoral component and for aseptic loosening as endpoints. The indications for revision were evaluated. Fractures of the femoral component were described in detail.Aims
Methods
The December 2022 Hip & Pelvis Roundup360 looks at: Fix and replace: simultaneous fracture fixation and hip arthroplasty for acetabular fractures in older patients; Is the revision rate for femoral neck fracture lower for total hip arthroplasty than for hemiarthroplasty?; Femoral periprosthetic fractures: data from the COMPOSE cohort study; Dual-mobility cups and fracture of the femur; What’s the deal with outcomes for hip and knee arthroplasty outcomes internationally?; Osteochondral lesions of the femoral head: is costal cartilage the answer?
Radiostereometric analysis (RSA) is the most accurate radiological method to measure in vivo wear of highly cross-linked polyethylene (XLPE) acetabular components. We have previously reported very low wear rates for a sequentially irradiated and annealed X3 XLPE liner (Stryker Orthopaedics, USA) when used in conjunction with a 32 mm femoral heads at ten-year follow-up. Only two studies have reported the long-term wear rate of X3 liners used in conjunction with larger heads using plain radiographs which have poor sensitivity. The aim of this study was to measure the ten-year wear of thin X3 XLPE liners against larger 36 or 40 mm articulations with RSA. We prospectively reviewed 19 patients who underwent primary cementless THA with the XLPE acetabular liner (X3) and a 36 or 40 mm femoral head with a resultant liner thickness of at least 5.8 mm. RSA radiographs at one week, six months, and one, two, five, and ten years postoperatively and femoral head penetration within the acetabular component were measured with UmRSA software. Of the initial 19 patients, 12 were available at the ten-year time point.Aims
Methods
A novel enhanced cement fixation (EF) tibial implant with deeper cement pockets and a more roughened bonding surface was released to market for an existing total knee arthroplasty (TKA) system.This randomized controlled trial assessed fixation of the both the EF (ATTUNE S+) and standard (Std; ATTUNE S) using radiostereometric analysis. Overall, 50 subjects were randomized (21 EF-TKA and 23 Std-TKA in the final analysis), and had follow-up visits at six weeks, and six, 12, and 24 months to assess migration of the tibial component. Low viscosity bone cement with tobramycin was used in a standardized fashion for all subjects. Patient-reported outcome measure data was captured at preoperative and all postoperative visits.Aims
Methods
The Exeter V40 femoral stem is the most implanted stem in the National Joint Registry (NJR) for primary total hip arthroplasty (THA). In 2004, the 44/00/125 stem was released for use in ‘cement-in-cement’ revision cases. It has, however, been used ‘off-label’ as a primary stem when patient anatomy requires a smaller stem with a 44 mm offset. We aimed to investigate survival of this implant in comparison to others in the range when used in primary THAs recorded in the NJR. We analyzed 328,737 primary THAs using the Exeter V40 stem, comprising 34.3% of the 958,869 from the start of the NJR to December 2018. Our exposure was the stem, and the outcome was all-cause construct revision. We stratified analyses into four groups: constructs using the 44/00/125 stem, those using the 44/0/150 stem, those including a 35.5/125 stem, and constructs using any other Exeter V40 stem.Aims
Methods
The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface. Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up.Aims
Methods
Metal and ceramic humeral head bearing surfaces are available choices in anatomical shoulder arthroplasties. Wear studies have shown superior performance of ceramic heads, however comparison of clinical outcomes according to bearing surface in total shoulder arthroplasty (TSA) and hemiarthroplasty (HA) is limited. This study aimed to compare the rates of revision and reoperation following metal and ceramic humeral head TSA and HA using data from the National Joint Registry (NJR), which collects data from England, Wales, Northern Ireland, Isle of Man and the States of Guernsey. NJR shoulder arthroplasty records were linked to Hospital Episode Statistics and the National Mortality Register. TSA and HA performed for osteoarthritis (OA) in patients with an intact rotator cuff were included. Metal and ceramic humeral head prostheses were matched within separate TSA and HA groups using propensity scores based on 12 and 11 characteristics, respectively. The primary outcome was time to first revision and the secondary outcome was non-revision reoperation.Aims
Methods
Robotic-assisted unicompartmental knee arthroplasty (R-UKA) has been proposed as an approach to improve the results of the conventional manual UKA (C-UKA). The aim of this meta-analysis was to analyze the studies comparing R-UKA and C-UKA in terms of clinical outcomes, radiological results, operating time, complications, and revisions. The literature search was conducted on three databases (PubMed, Cochrane, and Web of Science) on 20 February 2024 according to the guidelines for Preferred Reporting Items for Systematic Reviews and Meta-Analysis (PRISMA). Inclusion criteria were comparative studies, written in the English language, with no time limitations, on the comparison of R-UKA and C-UKA. The quality of each article was assessed using the Downs and Black Checklist for Measuring Quality.Aims
Methods
This study aimed to describe practice variation in the use of total hip arthroplasty (THA) for older patients with femoral neck fracture and to determine the association between patient, surgeon, and institution factors and treatment with THA. We performed a cross-sectional analysis of 49,597 patients aged 60 years and older from Ontario, Canada, who underwent hemiarthroplasty or THA for femoral neck fracture between 2002 and 2017. This population-based study used routinely collected healthcare databases linked through ICES (formerly known as the Institute for Clinical Evaluative Sciences). Multilevel logistic regression modelling was used to quantify the association between patient, surgeon, and institution-level variables and whether patients were treated with THA. Variance partition coefficient and median odds ratios were used to estimate the variation attributable to higher-level variables and the magnitude of effect of higher-level variables, respectively.Aims
Methods
Conflicting clinical results are reported for the ATTUNE Total Knee Arthroplasty (TKA). This randomized controlled trial (RCT) evaluated five-year follow-up results comparing cemented ATTUNE and PFC-Sigma cruciate retaining TKAs, analyzing component migration as measured by radiostereometric analysis (RSA), clinical outcomes, patient-reported outcome measures (PROMs), and radiological outcomes. A total of 74 primary TKAs were included in this single-blind RCT. RSA examinations were performed, and PROMs and clinical outcomes were collected immediate postoperatively, and at three, six, 12, 24, and 60 months’ follow-up. Radiolucent lines (RLLs) were measured in standard anteroposterior radiographs at six weeks, and 12 and 60 months postoperatively.Aims
Methods
The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs). At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays.Aims
Methods
The aim of this study was to explore the functional results in a fitter subgroup of participants in the Hip Fracture Evaluation with Alternatives of Total Hip Arthroplasty versus Hemiarthroplasty (HEALTH) trial to determine whether there was an advantage of total hip arthroplasty (THA) versus hemiarthroplasty (HA) in this population. We performed a post hoc exploratory analysis of a fitter cohort of patients from the HEALTH trial. Participants were aged over 50 years and had sustained a low-energy displaced femoral neck fracture (FNF). The fittest participant cohort was defined as participants aged 70 years or younger, classified as American Society of Anesthesiologists grade I or II, independent walkers prior to fracture, and living at home prior to fracture. Multilevel models were used to estimate the effect of THA versus HA on functional outcomes. In addition, a sensitivity analysis of the definition of the fittest participant cohort was performed.Aims
Methods
This study aims to describe the pre- and postoperative self-reported health and quality of life from a national cohort of patients undergoing elective total conventional hip arthroplasty (THA) and total knee arthroplasty (TKA) in Australia. For context, these data will be compared with patient-reported outcome measures (PROMs) data from other international nation-wide registries. Between 2018 to 2020, and nested within a nationwide arthroplasty registry, preoperative and six-month postoperative PROMs were electronically collected from patients before and after elective THA and TKA. There were 5,228 THA and 8,299 TKA preoperative procedures as well as 3,215 THA and 4,982 TKA postoperative procedures available for analysis. Validated PROMs included the EuroQol five-dimension five-level questionnaire (EQ-5D-5L; range 0 to 100; scored worst-best health), Oxford Hip/Knee Scores (OHS/OKS; range 0 to 48; scored worst-best hip/knee function) and the 12-item Hip/Knee disability and Osteoarthritis Outcome Score (HOOS-12/KOOS-12; range 0 to 100; scored best-worst hip/knee health). Additional items included preoperative expectations, patient-perceived improvement, and postoperative satisfaction. Descriptive analyses were undertaken.Aims
Methods
Total hip arthroplasty (THA) with dual-mobility components (DM-THA) has been shown to decrease the risk of dislocation in the setting of a displaced neck of femur fracture compared to conventional single-bearing THA (SB-THA). This study assesses if the clinical benefit of a reduced dislocation rate can justify the incremental cost increase of DM-THA compared to SB-THA. Costs and benefits were established for patients aged 75 to 79 years over a five-year time period in the base case from the Canadian Health Payer’s perspective. One-way and probabilistic sensitivity analysis assessed the robustness of the base case model conclusions.Aims
Methods
Femoral bone preparation using compaction technique has been shown to preserve bone and improve implant fixation in animal models. No long-term clinical outcomes are available. There are no significant long-term differences between compaction and broaching techniques for primary total hip arthroplasty (THA) in terms of migration, clinical, and radiological outcomes. A total of 28 patients received one-stage bilateral primary THA with cementless femoral stems (56 hips). They were randomized to compaction on one femur and broaching on the contralateral femur. Overall, 13 patients were lost to the ten-year follow-up leaving 30 hips to be evaluated in terms of stem migration (using radiostereometry), radiological changes, Harris Hip Score, Oxford Hip Score, and complications.Aims
Methods
Medial pivot (MP) total knee arthroplasties (TKAs) were designed to mimic native knee kinematics with their deep medial congruent fitting of the tibia to the femur almost like a ball-on-socket, and a flat lateral part. GMK Sphere is a novel MP implant. Our primary aim was to study the migration pattern of the tibial tray of this TKA. A total of 31 patients were recruited to this single-group radiostereometric analysis (RSA) study and received a medial pivot GMK Sphere TKA. The distributions of male patients versus female patients and right versus left knees were 21:10 and 17:14, respectively. Mean BMI was 29 kg/m2 (95% confidence interval (CI) 27 to 30) and mean age at surgery was 63 years (95% CI 61 to 66). Maximum total point motions (MTPMs), medial, proximal, and anterior translations and transversal, internal, and varus rotations were calculated at three, 12, and 24 months. Patient-reported outcome measure data were also retrieved.Aims
Methods
The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs). We undertook a registry study combining the National Joint Registry dataset with polyethylene manufacturing characteristics as supplied by the manufacturers. The primary endpoint was revision for any reason. We performed further analyses on other reasons including instability, aseptic loosening, wear, and liner dissociation. The primary analytic approach was Cox proportional hazard regression.Aims
Methods
To report early (two-year) postoperative findings from a randomized controlled trial (RCT) investigating disease-specific quality of life (QOL), clinical, patient-reported, and radiological outcomes in patients undergoing a total shoulder arthroplasty (TSA) with a second-generation uncemented trabecular metal (TM) glenoid versus a cemented polyethylene glenoid (POLY) component. Five fellowship-trained surgeons from three centres participated. Patients aged between 18 and 79 years with a primary diagnosis of glenohumeral osteoarthritis were screened for eligibility. Patients were randomized intraoperatively to either a TM or POLY glenoid component. Study intervals were: baseline, six weeks, six-, 12-, and 24 months postoperatively. The primary outcome was the Western Ontario Osteoarthritis Shoulder QOL score. Radiological images were reviewed for metal debris. Mixed effects repeated measures analysis of variance for within and between group comparisons were performed.Aims
Methods
Unicompartmental knee arthroplasty (UKA) is a bone-preserving treatment option for osteoarthritis localized to a single compartment in the knee. The success of the procedure is sensitive to patient selection and alignment errors. Robotic arm-assisted UKA provides technological assistance to intraoperative bony resection accuracy, which is thought to improve ligament balancing. This paper presents the five-year outcomes of a comparison between manual and robotically assisted UKAs. The trial design was a prospective, randomized, parallel, single-centre study comparing surgical alignment in patients undergoing UKA for the treatment of medial compartment osteoarthritis (ISRCTN77119437). Participants underwent surgery using either robotic arm-assisted surgery or conventional manual instrumentation. The primary outcome measure (surgical accuracy) has previously been reported, and, along with secondary outcomes, were collected at one-, two-, and five-year timepoints. Analysis of five-year results and longitudinal analysis for all timepoints was performed to compare the two groups.Aims
Methods
To establish our early clinical results of a new total knee arthroplasty (TKA) tibial component introduced in 2013 and compare it to other designs in use at our hospital during the same period. This is a retrospective study of 166 (154 patients) consecutive cemented, fixed bearing, posterior-stabilized (PS) TKAs (ATTUNE) at one hospital performed by five surgeons. These were compared with a reference cohort of 511 knees (470 patients) of other designs (seven manufacturers) performed at the same hospital by the same surgeons. There were no significant differences in age, sex, BMI, or follow-up times between the two cohorts. The primary outcome was revision performed or pending.Aims
Methods
The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems. A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020.Aims
Methods
Vitamin E-infused highly cross-linked polyethylene (E1) has recently been introduced in total knee arthroplasty (TKA). An in vitro wear simulator study showed that E1 reduced polyethylene wear. However there is no published information regarding in vivo wear. Previous reports suggest that newly introduced materials which reduce in vitro polyethylene wear do not necessarily reduce in vivo polyethylene wear. To assist in the evaluation of the newly introduced material before widespread use, we established an in vivo polyethylene wear particle analysis for TKA. The aim of this study was to compare in vivo polyethylene wear particle generation between E1 and conventional polyethylene (ArCom) in TKA. A total of 34 knees undergoing TKA (17 each with ArCom or E1) were investigated. Except for the polyethylene insert material, the prostheses used for both groups were identical. Synovial fluid was obtained at a mean of 3.4 years (SD 1.3) postoperatively. The in vivo polyethylene wear particles were isolated from the synovial fluid using a previously validated method and examined by scanning electron microscopy.Aims
Methods
The primary objective of this study was to compare migration of the cemented ATTUNE fixed bearing cruciate retaining tibial component with the cemented Press-Fit Condylar (PFC)-sigma fixed bearing cruciate retaining tibial component. The secondary objectives included comparing clinical and radiological outcomes and Patient Reported Outcome Measures (PROMs). A single blinded randomized, non-inferiority study was conducted including 74 patients. Radiostereometry examinations were made after weight bearing, but before hospital discharge, and at three, six, 12, and 24 months postoperatively. PROMS were collected preoperatively and at three, six, 12, and 24 months postoperatively. Radiographs for measuring radiolucencies were collected at two weeks and two years postoperatively.Aims
Methods
To investigate the effect of polyethylene manufacturing characteristics and irradiation dose on the survival of cemented and reverse hybrid total hip arthroplasties (THAs). In this registry study, data from the National Joint Registry of England, Wales, Northern Ireland and the Isle of Man (NJR) were linked with manufacturing data supplied by manufacturers. The primary endpoint was revision of any component. Cox proportional hazard regression was a primary analytic approach adjusting for competing risk of death, patient characteristics, head composition, and stem fixation.Aims
Methods
Short-stemmed femoral implants have been used for total hip arthroplasty (THA) in young and active patients to conserve bone, provide physiological loading, and reduce the incidence of thigh pain. Only short- to mid-term results have been presented and there have been concerns regarding component malalignment, incorrect sizing, and subsidence. This systematic review reports clinical and radiological outcomes, complications, revision rates, and implant survival in THA using short-stemmed femoral components. A literature review was performed using the EMBASE, Medline, and Cochrane databases. Strict inclusion and exclusion criteria were used to identify studies reporting clinical and radiological follow-up for short-stemmed hip arthroplasties.Aims
Materials and Methods
It is not known whether change in patient-reported outcome measures (PROMs) over time can be predicted by factors present at surgery, or early follow-up. The aim of this study was to identify factors associated with changes in PROM status between two-year evaluation and medium-term follow-up. Patients undergoing Birmingham Hip Resurfacing completed the Veteran’s Rand 36 (VR-36), modified Harris Hip Score (mHHS), Tegner Activity Score, and Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) at two years and a minimum of three years. A change in score was assessed against minimal clinically important difference (MCID) and patient-acceptable symptom state (PASS) thresholds. Binary logistic regression was used to assess the relationship between patient factors and deterioration in PASS status between follow-ups.Aims
Patients and Methods
Instability continues to be a troublesome complication after total hip arthroplasty (THA). Patient-related risk factors associated with a higher dislocation risk include the preoperative diagnosis, an age of 75 years or older, high body mass index (BMI), a history of alcohol abuse, and neurodegenerative diseases. The goal of this study was to assess the dislocation rate, radiographic outcomes, and complications of patients stratified as high-risk for dislocation who received a dual mobility (DM) bearing in a primary THA at a minimum follow-up of two years. We performed a retrospective review of a consecutive series of DM THA performed between 2010 and 2014 at our institution (Hospital for Special Surgery, New York, New York) by a single, high-volume orthopaedic surgeon employing a single prosthesis design (Anatomic Dual Mobility (ADM) Stryker, Mahwah, New Jersey). Patient medical records and radiographs were reviewed to confirm the type of implant used, to identify any preoperative risk factors for dislocation, and any complications. Radiographic analysis was performed to assess for signs of osteolysis or remodelling of the acetabulum.Aims
Materials and Methods
Between 15% and 20% of patients remain dissatisfied following total knee arthroplasty (TKA). The SAIPH knee system (MatOrtho, Surrey, United Kingdom) is a medial ball and socket TKA that has been designed to replicate native knee kinematics in order to maximize the range of movement, stability, and function. This system is being progressively introduced in a stepwise fashion, with this study reporting the mid-term clinical and radiological outcomes. A retrospective review was undertaken of the first 100 consecutive patients with five-year follow-up following SAIPH TKA performed by the senior authors. The data that were collected included the demographics of the patients, clinical findings, the rate of intraoperative ligamentous release, patient-reported outcome measures (PROMS), radiological assessment, complications, and all-cause revision. Revision data were cross-checked with a national registry.Aims
Patients and Methods
This study reports the mid-term results of total hip arthroplasty (THA) performed using a monoblock acetabular component with a large-diameter head (LDH) ceramic-on-ceramic (CoC) bearing. Of the 276 hips (246 patients) included in this study, 264 (96%) were reviewed at a mean of 67 months (48 to 79) postoperatively. Procedures were performed with a mini posterior approach. Clinical and radiological outcomes were recorded at regular intervals. A noise assessment questionnaire was completed at last follow-up.Aims
Patients and Methods
As the population ages, there is projected to be an increase in the level of demand for total knee arthroplasty (TKA) in octogenarians. We aimed to explore whether those aged ≥ 80 years achieved similar improvements in physical function to younger patients while also comparing the rates of length of stay (LOS), discharge to rehabilitation, postoperative complications, and mortality following TKA in older and younger patients. Patients from one institution who underwent primary elective TKA between 1 January 2006 and 31 December 2014 were dichotomized into those ≥ 80 years old (n = 359) and those < 80 years old (n = 2479) for comparison. Multivariable regression was used to compare the physical status component of the 12-Item Short-Form Health Survey (SF-12), LOS, discharge to rehabilitation, complications, and mortality between the two groups.Aims
Patients and Methods
This study presents the long-term survivorship, risk factors for prosthesis survival, and an assessment of the long-term effects of changes in surgical technique in a large series of patients treated by metal-on-metal (MoM) hip resurfacing arthroplasty (HRA). Between November 1996 and January 2012, 1074 patients (1321 hips) underwent HRA using the Conserve Plus Hip Resurfacing System. There were 787 men (73%) and 287 women (27%) with a mean age of 51 years (14 to 83). The underlying pathology was osteoarthritis (OA) in 1003 (75.9%), developmental dysplasia of the hip (DDH) in 136 (10.3%), avascular necrosis in 98 (7.4%), and other conditions, including inflammatory arthritis, in 84 (6.4%).Aims
Patients and Methods
Recently the National Patient Safety Agency in the United Kingdom published a report entitled “Mitigating surgical risk in patients undergoing hip arthroplasty for fractures of the proximal femur”. A total of 26 deaths had been reported to them when cement was used at hemiarthroplasty between October 2003 and October 2008. This paper considers the evidence for using cement fixation of a hemiarthroplasty in the treatment of hip fractures.
This study evaluated the feasibility of using published data from more than one register to define the performance of different hip implants. In order to obtain estimates of performance for specific types of hip system from different register, we analysed data from the annual reports of five national and one Italian regional register. We extracted the number of implants and rates of implant survival at different periods of follow-up. Our aim was to assess whether estimates of cumulative survival rate were comparable with data from registers from different countries, and our conclusion was that such a comparison could only be performed incompletely.
The survivorship of contemporary resurfacing arthroplasty of the hip using metal-on-metal bearings is better than that of first generation designs, but short-term failures still occur. The most common reasons for failure are fracture of the femoral neck, loosening of the component, osteonecrosis of the femoral head, reaction to metal debris and malpositioning of the component. In 2008 the Australian National Joint Registry reported an inverse relationship between the size of the head component and the risk of revision in resurfacing hip arthroplasty. Hips with a femoral component size of ≤ 44 mm have a fivefold increased risk of revision than those with femoral components of ≥ 55 mm irrespective of gender. We have reviewed the literature to explore this observation and to identify possible reasons including the design of the implant, loading of the femoral neck, the orientation of the component, the production of wear debris and the effects of metal ions, penetration of cement and vascularity of the femoral head. Our conclusion is that although multifactorial, the most important contributors to failure in resurfacing arthroplasty of the hip are likely to be the design and geometry of the component and the orientation of the acetabular component.
A series of 100 consecutive osteoarthritic patients was randomised to undergo total knee replacement using a Miller-Galante II prosthesis, with or without a cemented polyethylene patellar component. Knee function was evaluated using the American Knee Society score, Western Ontario and McMaster University Osteoarthritis index, specific patellofemoral-related questions and radiographic evaluation until the fourth post-operative year, then via questionnaire until ten years post-operatively. A ten-point difference in the American Knee Society score between the two groups was considered a significant change in knee performance, with α and β levels of 0.05. The mean age of the patients in the resurfaced group was 71 years (53 to 88) and in the non-resurfaced group was 73 years (54 to 86). After ten years 22 patients had died, seven were suffering from dementia, three declined further participation and ten were lost to follow-up. Two patients in the non-resurfaced group subsequently had their patellae resurfaced. In the resurfaced group one patient had an arthroscopic lateral release. There was no significant difference between the two treatment groups: both had a similar deterioration of scores with time, and no further patellofemoral complications were observed in either group. We are unable to recommend routine patellar resurfacing in osteoarthritic patients undergoing total knee replacement on the basis of our findings.