Aims. The Exeter V40
Aims. Periprosthetic fracture (PF) after primary total hip arthroplasty
(THA) is an uncommon but potentially devastating complication. This
study aims to investigate the influence of cemented stem designs
on the risk of needing a revision for a PF. Patients and Methods. We analysed data on 257 202 primary THAs with cemented stems
and 390 linked first revisions for PF recorded in the National Joint
Registry (NJR) of England, Wales and Northern Ireland to determine
if a
The removal of well-fixed bone cement from the femoral canal during revision of a total hip replacement (THR) can be difficult and risks the loss of excessive bone stock and perforation or fracture of the femoral shaft. Retaining the cement mantle is attractive, yet the technique of cement-in-cement revision is not widely practised. We have used this procedure at our hospital since 1989. The stems were removed to gain a better exposure for acetabular revision, to alter version or leg length, or for component incompatibility. We studied 136 hips in 134 patients and followed them up for a mean of eight years (5 to 15). A further revision was required in 35 hips (25.7%), for acetabular loosening in 26 (19.1%), sepsis in four, instability in three, femoral fracture in one and stem fracture in one. No femoral stem needed to be re-revised for aseptic loosening. A cement-in-cement revision of the femoral stem is a reliable technique in the medium term. It also reduces the risk of perforation or fracture of the femoral shaft.
We have compared prospectively the incidence of loosening of 20 femoral stems with a matt surface with that of 20 polished stems of an otherwise identical tapered, non-modular design of Exeter hip replacement. The stems were inserted using the same technique at operation and radiographs showed no difference in the adequacy of the cement mantle or of fixation. All the patients were reviewed regularly and none was lost to follow-up. After a minimum follow-up of nine years, four matt but no polished stems had been revised for aseptic loosening. Polished stems subsided slightly within the cement mantle early, but did not loosen.
We performed a case–control study to compare
the rates of further surgery, revision and complications, operating time
and survival in patients who were treated with either an uncemented
hydroxyapatite-coated Corail bipolar femoral stem or a cemented
Exeter stem for a displaced intracapsular fracture of the hip. The
mean age of the patients in the uncemented group was 82.5 years
(53 to 97) and in the cemented group was 82.7 years (51 to 99) We used
propensity score matching, adjusting for age, gender and the presence
or absence of dementia and comorbidities, to produce a matched cohort
receiving an Exeter stem (n = 69) with which to compare the outcome of
patients receiving a Corail stem (n = 69). The Corail had a significantly
lower all-cause rate of further surgery (p = 0.016; odds ratio (OR)
0.18, 95% CI 0.04 to 0.84) and number of hips undergoing major further
surgery (p = 0.029; OR 0.13, 95% CI 0.01 to 1.09). The mean operating
time was significantly less for the Corail group than for the cemented Exeter
group (59 min [12 to 136] Cite this article:
We studied the effect of the surface finish of the stem on the transfer of load in the proximal femur in a sheep model of cemented hip arthroplasty. Strain-gauge analysis and corresponding finite-element (FE) analysis were performed to assess the effect of friction and creep at the cement-stem interface. No difference was seen between the matt and polished stems. FE analysis showed that the effects of cement creep and friction at the stem-cement interface on femoral strain were small compared with the effect of inserting a cemented stem.
The role of vacuum mixing on the reduction of porosity and on the clinical performance of cemented total hip replacements remains uncertain. We have used paired femoral constructs prepared with either hand-mixed or vacuum-mixed cement in a cadaver model which simulated intra-operative conditions during cementing of the femoral component. After the cement had cured, the distribution of its porosity was determined, as was the strength of the cement-stem and cement-bone interfaces. The overall fraction of the pore area was similar for both hand-mixed and vacuum-mixed cement (hand 6%; vacuum 5.7%; paired
Polished, tapered stems are now widely used for cemented total hip replacement and many such designs have been introduced. However, a change in stem geometry may have a profound influence on stability. Stems with a wide, rectangular proximal section may be more stable than those which are narrower proximally. We examined the influence of proximal geometry on stability by comparing the two-year migration of the Exeter stem with a more recent design, the CPS-Plus, which has a wider shoulder and a more rectangular cross-section. The hypothesis was that these design features would increase rotational stability. Both stems subsided approximately 1 mm relative to the femur during the first two years after implantation. The Exeter stem was found to rotate into valgus (mean 0.2°,
Periprosthetic fractures (PPFs) following hip arthroplasty are complex injuries. This study evaluates patient demographic characteristics, management, outcomes, and risk factors associated with PPF subtypes over a decade. Using a multicentre collaborative study design, independent of registry data, we identified adults from 29 centres with PPFs around the hip between January 2010 and December 2019. Radiographs were assessed for the Unified Classification System (UCS) grade. Patient and injury characteristics, management, and outcomes were compared between UCS grades. A multinomial logistic regression was performed to estimate relative risk ratios (RRR) of variables on UCS grade.Aims
Methods
Cemented titanium stems in hip arthroplasty are associated with proximal cement-stem debonding and early failure. This was well publicised with the 3M Capital hip. However, corrosion in this setting has been reported with only one stem design and is less widely accepted. We present a series of 12 cemented titanium Furlong Straight Stems which required revision at a mean of 78 months for thigh pain. At revision the stems were severely corroded in a pattern which was typical of crevice corrosion. Symptoms were eliminated after revision to an all-stainless steel femoral prosthesis of the same design. We discuss the likely causes for the corrosion. The combination of a titanium stem and cement appears to facilitate crevice corrosion.
The radiological features of the cement mantle around total hip replacements (THRs) have been used to assess aseptic loosening. In this case-control study we investigated the risk of failure of THR as predictable by a range of such features using data from patients recruited to the Trent Regional Arthroplasty Study (TRAS). An independent radiological assessment was undertaken on Charnley THRs with aseptic loosening within five years of surgery and on a control group from the TRAS database. Chi-squared tests were used to test the probability of obtaining the observed data by chance, and odds ratios were calculated to estimate the strength of association for different features. Several features were associated with a clinically important increase (>
twofold) in the risk of loosening, which was statistically significant for four features (p <
0.01). Inadequate cementation (Barrack C and D grades) was the most significant feature, with an estimated odds ratio of 9.5 (95% confidence interval 3.2 to 28.4, p <
0.0001) for failure.
Polished taper-slip (PTS) cemented stems have an excellent clinical track record and are the most common stem type used in primary total hip arthroplasty (THA) in the UK. Due to low rates of aseptic loosening, they have largely replaced more traditional composite beam (CB) cemented stems. However, there is now emerging evidence from multiple joint registries that PTS stems are associated with higher rates of postoperative periprosthetic femoral fracture (PFF) compared to their CB stem counterparts. The risk of both intraoperative and postoperative PFF remains greater with uncemented stems compared to either of these cemented stem subtypes. PFF continues to be a devastating complication following primary THA and is associated with high complication and mortality rates. Recent efforts have focused on identifying implant-related risk factors for PFF in order to guide preventative strategies, and therefore the purpose of this article is to present the current evidence on the effect of
Aims. The aim of this study was to radiologically evaluate the quality of cement mantle and alignment achieved with a polished tapered
Aims. The aim of this large registry-based study was to compare mid-term survival rates of
Aims. Femoral impaction bone grafting was first developed in 1987 using
morselised cancellous bone graft impacted into the femoral canal
in combination with a cemented, tapered, polished stem. We describe
the evolution of this technique and instrumentation since that time. . Patients and Methods. Between 1987 and 2005, 705 revision total hip arthroplasties
(56 bilateral) were performed with femoral impaction grafting using
a
We describe the results at five years of a prospective study of a new tri-tapered polished, cannulated,
The February 2023 Hip & Pelvis Roundup360 looks at: Total hip arthroplasty or internal fixation for hip fracture?; Significant deterioration in quality of life and increased frailty in patients waiting more than six months for total hip or knee arthroplasty: a cross-sectional multicentre study; Long-term cognitive trajectory after total joint arthroplasty; Costal cartilage grafting for a large osteochondral lesion of the femoral head; Foley catheters not a problem in the short term; Revision hips still a mortality burden?; How to position implants with a robotic arm; Uncemented stems in hip fracture?
The February 2024 Hip & Pelvis Roundup360 looks at: Trial of vancomycin and cefazolin as surgical prophylaxis in arthroplasty; Is preoperative posterior femoral neck tilt a risk factor for fixation failure? Cemented versus uncemented hemiarthroplasty for displaced intracapsular fractures of the hip; Periprosthetic fractures in larger hydroxyapatite-coated stems: are collared stems a better alternative for total hip arthroplasty?; Postoperative periprosthetic fracture following hip arthroplasty with a polished taper slip versus composite beam stem; Is oral tranexamic acid as good as intravenous?; Stem design and the risk of early periprosthetic femur fractures following THA in elderly patients; Does powered femoral broaching compromise patient safety in total hip arthroplasty?
Osteogenesis imperfecta is a rare inherited disorder of connective tissue which may present with recurrent fractures which are prone to nonunion and malunion resulting in deformity. Some patients develop osteoarthritis of the hip. Formation of hyperplastic callus after recurrent fractures may deform the shape of the femur and preclude the use of standard implants at joint replacement. Replacement can thus be technically demanding. We present a case of bilateral hip replacement in a patient with osteogenesis imperfecta and hyperplastic callus which was treated by the use of long femoral allografts and
Objectives. Orthopaedic surgeons use stems in revision knee surgery to obtain
stability when metaphyseal bone is missing. No consensus exists
regarding stem size or method of fixation. This in vitro study
investigated the influence of stem length and method of fixation
on the pattern and level of relative motion at the bone–implant
interface at a range of functional flexion angles. Methods. A custom test rig using differential variable reluctance transducers
(DVRTs) was developed to record all translational and rotational
motions at the bone–implant interface. Composite femurs were used.
These were secured to permit variation in flexion angle from 0°
to 90°. Cyclic loads were applied through a tibial component based
on three peaks corresponding to 0°, 10° and 20° flexion from a normal
walking cycle. Three different femoral components were investigated
in this study for cementless and cemented interface conditions. Results. Relative motions were found to increase with flexion angle. Stemmed
implants reduced relative motions in comparison to stemless implants
for uncemented constructs. Relative motions for cemented implants
were reduced to one-third of their equivalent uncemented constructs. Conclusions. Stems are not necessary for cemented implants when the metaphyseal
bone is intact. Short
Accurate diagnosis of chronic periprosthetic joint infection (PJI) presents a significant challenge for hip surgeons. Preoperative diagnosis is not always easy to establish, making the intraoperative decision-making process crucial in deciding between one- and two-stage revision total hip arthroplasty (THA). Calprotectin is a promising point-of-care novel biomarker that has displayed high accuracy in detecting PJI. We aimed to evaluate the utility of intraoperative calprotectin lateral flow immunoassay (LFI) in THA patients with suspected chronic PJI. The study included 48 THAs in 48 patients with a clinical suspicion of PJI, but who did not meet European Bone and Joint Infection Society (EBJIS) PJI criteria preoperatively, out of 105 patients undergoing revision THA at our institution for possible PJI between November 2020 and December 2022. Intraoperatively, synovial fluid calprotectin was measured with LFI. Cases with calprotectin levels ≥ 50 mg/l were considered infected and treated with two-stage revision THA; in negative cases, one-stage revision was performed. At least five tissue cultures were obtained; the implants removed were sent for sonication.Aims
Methods
Cementing in arthroplasty for hip fracture is associated with improved postoperative function, but may have an increased risk of early mortality compared to uncemented fixation. Quantifying this mortality risk is important in providing safe patient care. This study investigated the association between cement use in arthroplasty and mortality at 30 days and one year in patients aged 50 years and over with hip fracture. This retrospective cohort study used linked data from the Australian Hip Fracture Registry and the National Death Index. Descriptive analysis and Kaplan-Meier survival curves tested the unadjusted association of mortality between cemented and uncemented procedures. Multilevel logistic regression, adjusted for covariates, tested the association between cement use and 30-day mortality following arthroplasty. Given the known institutional variation in preference for cemented fixation, an instrumental variable analysis was also performed to minimize the effect of unknown confounders. Adjusted Cox modelling analyzed the association between cement use and mortality at 30 days and one year following surgery.Aims
Methods
Refobacin Bone Cement R and Palacos Overall, 75 patients were included in the study and 71 were available at two years postoperatively. Prior to surgery, they were randomized to one of the three combinations studied: Palacos cement with use of the Optivac mixing system, Refobacin with use of the Optivac system, and Refobacin with use of the Optipac system. Cemented MS30 stems and cemented Exceed acetabular components were used in all hips. Postoperative radiographs were used to assess the quality of the cement mantle according to Barrack et al, and the position and migration of the femoral stem. Harris Hip Score, Oxford Hip Score, Forgotten Joint Score, and University of California, Los Angeles Activity Scale were collected.Aims
Methods
Professional dancers represent a unique patient population in the setting of hip arthroplasty, given the high degree of hip strength and mobility required by their profession. We sought to determine the clinical outcomes and ability to return to professional dance after total hip arthroplasty (THA) or hip resurfacing arthroplasty (HRA). Active professional dancers who underwent primary THA or HRA at a single institution with minimum one-year follow-up were included in the study. Primary outcomes included the rate of return to professional dance, three patient-reported outcome measures (PROMs) (modified Harris Hip Score (mHHS), Hip disability and Osteoarthritis Outcome Score for Joint Replacement (HOOS-JR), and Lower Extremity Activity Scale (LEAS)), and postoperative complications.Aims
Methods
The Birmingham Hip Resurfacing (BHR) was introduced in 1997 to address the needs of young active patients using a historically proven large-diameter metal-on-metal (MoM) bearing. A single designer surgeon’s consecutive series of 130 patients (144 hips) was previously reported at five and ten years, reporting three and ten failures, respectively. The aim of this study was to extend the follow-up of this original cohort at 25 years. The study extends the reporting on the first consecutive 144 resurfacing procedures in 130 patients for all indications. All operations were undertaken between August 1997 and May 1998. The mean age at operation was 52.1 years (SD 9.93; 17 to 76), and included 37 female patients (28.5%). Failure was defined as revision of either component for any reason. Kaplan-Meier survival analysis was performed. Routine follow-up with serum metal ion levels, radiographs, and Oxford Hip Scores (OHSs) was undertaken.Aims
Methods
The December 2022 Hip & Pelvis Roundup360 looks at: Fix and replace: simultaneous fracture fixation and hip arthroplasty for acetabular fractures in older patients; Is the revision rate for femoral neck fracture lower for total hip arthroplasty than for hemiarthroplasty?; Femoral periprosthetic fractures: data from the COMPOSE cohort study; Dual-mobility cups and fracture of the femur; What’s the deal with outcomes for hip and knee arthroplasty outcomes internationally?; Osteochondral lesions of the femoral head: is costal cartilage the answer?
Early failures have seen titanium fall from favour as a material for
We report a clinical and radiological survival analysis of 148 consecutive revisions of hip arthroplasties. All patients referred were offered revision if it was indicated and they were medically fit. About one-third (32%) had had at least one previous revision, and about one-third had an established deep infection. The revisions were performed in ultra-clean air with body-exhaust suits. The usual method of fixation was by contained and pressurised cement of standard viscosity, to which appropriate antibiotics had been added in infected cases. Some patients had cementless revision. Clinical failure meant that one or both of the implants had been removed; radiological failure was assessed from serial radiographs. The clinical survival at ten years was 95%. The Merle D'Aubigne and Postel rating for pain improved from a mean of 2.9 to 5.2; and in Charnley group A and B cases, walking ability improved from 2.3 to 4.3. In contrast to some reports we also found good radiological survival; this was 90.5% for
The aim of this study was to describe the management and associated outcomes of patients sustaining a femoral hip periprosthetic fracture (PPF) in the UK population. This was a multicentre retrospective cohort study including adult patients who presented to 27 NHS hospitals with 539 new PPFs between 1 January 2018 and 31 December 2018. Data collected included: management strategy (operative and nonoperative), length of stay, discharge destination, and details of post-treatment outcomes (reoperation, readmission, and 30-day and 12-month mortality). Descriptive analysis by fracture type was performed, and predictors of PPF management and outcomes were assessed using mixed-effects logistic regression.Aims
Methods
Femoral cement-in-cement revision is a well described technique to reduce morbidity and complications in hip revision surgery. Traditional techniques for septic revision of hip arthroplasty necessitate removal of all bone cement from the femur. In our two centres, we have been using a cement-in-cement technique, leaving the distal femoral bone cement in selected patients for septic hip revision surgery, both for single and the first of two-stage revision procedures. A prerequisite for adoption of this technique is that the surgeon considers the cement mantle to be intimately fixed to bone without an intervening membrane between cement and host bone. We aim to report our experience for this technique. We have analyzed patients undergoing this cement-in-cement technique for femoral revision in infection, and present a consecutive series of 89 patients. Follow-up was undertaken at a mean of 56.5 months (24.0 to 134.7) for the surviving cases.Aims
Methods
This study evaluates risk factors influencing fracture characteristics for postoperative periprosthetic femoral fractures (PFFs) around cemented stems in total hip arthroplasty. Data were collected for PFF patients admitted to eight UK centres between 25 May 2006 and 1 March 2020. Radiographs were assessed for Unified Classification System (UCS) grade and AO/OTA type. Statistical comparisons investigated relationships by age, gender, and stem fixation philosophy (polished taper-slip (PTS) vs composite beam (CB)). The effect of multiple variables was estimated using multinomial logistic regression to estimate odds ratios (ORs) with 95% confidence intervals (CIs). Surgical treatment (revision vs fixation) was compared by UCS grade and AO/OTA type.Aims
Methods
The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems. A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020.Aims
Methods
This aim of this study was to assess the reliability and validity of the Unified Classification System (UCS) for postoperative periprosthetic femoral fractures (PFFs) around cemented polished taper-slip (PTS) stems. Radiographs of 71 patients with a PFF admitted consecutively at two centres between 25 February 2012 and 19 May 2020 were collated by an independent investigator. Six observers (three hip consultants and three trainees) were familiarized with the UCS. Each PFF was classified on two separate occasions, with a mean time between assessments of 22.7 days (16 to 29). Interobserver reliability for more than two observers was assessed using percentage agreement and Fleiss’ kappa statistic. Intraobserver reliability between two observers was calculated with Cohen kappa statistic. Validity was tested on surgically managed UCS type B PFFs where stem stability was documented in operation notes (n = 50). Validity was assessed using percentage agreement and Cohen kappa statistic between radiological assessment and intraoperative findings. Kappa statistics were interpreted using Landis and Koch criteria. All six observers were blinded to operation notes and postoperative radiographs.Aims
Methods
To achieve the functional benefits of the direct anterior (DA) approach and the fixation benefits of cemented replacement, this study combined the two techniques posing the following questions: does the limited access of the DA approach adversely affect the cement technique?; and does such a cementing technique reduce the incidence of cementless complications? A consecutive series of 341 patients (360 hips) receiving the DA approach between 2016 and 2018 were reviewed. There were 203 cementless stems and 157 cemented stems. Mean age was 75 years (70 to 86) in the cementless group and 76 years (52 to 94) in the cemented group, with 239 (70%) females in the whole series. Femoral complications were compared between the two groups. Mean follow-up was 1.5 years (0.1 to 4.4) for patients in the cementless group and 1.3 years (0.0 to 3.9) for patients in the cemented group.Aims
Methods
The main aims were to identify risk factors predictive of a radiolucent line (RLL) around the acetabular component with an interface bioactive bone cement (IBBC) technique in the first year after THA, and evaluate whether these risk factors influence the development of RLLs at five and ten years after THA. A retrospective review was undertaken of 980 primary cemented THAs in 876 patients using cemented acetabular components with the IBBC technique. The outcome variable was any RLLs that could be observed around the acetabular component at the first year after THA. Univariate analyses with univariate logistic regression and multivariate analyses with exact logistic regression were performed to identify risk factors for any RLLs based on radiological classification of hip osteoarthritis.Aims
Methods
The aim of this study was to determine whether fixation, as opposed to revision arthroplasty, can be safely used to treat reducible Vancouver B type fractures in association with a cemented collarless polished tapered femoral stem (the Exeter). This retrospective cohort study assessed 152 operatively managed consecutive unilateral Vancouver B fractures involving Exeter stems; 130 were managed with open reduction and internal fixation (ORIF) and 22 with revision arthroplasty. Mean follow-up was 6.5 years (SD 2.6; 3.2 to 12.1). The primary outcome measure was revision of at least one component. Kaplan–Meier survival analysis was performed. Regression analysis was used to identify risk factors for revision following ORIF. Secondary outcomes included any reoperation, complications, blood transfusion, length of hospital stay, and mortality.Aims
Methods
To explore the effect of different types of articulating antibiotic-loaded cement spacers in two-stage revision for chronic hip prosthetic joint infection (PJI). A retrospective cohort study was performed involving 36 chronic PJI patients treated with different types of articulating antibiotic-loaded cement spacers between January 2014 and December 2017. The incidence of complications and the therapeutic effects of different types of antibiotic-loaded articulating cement spacers were compared.Aims
Methods
There is continued debate as to whether cemented or cementless implants should be utilized in particular cases based upon chronological age. This debate has been rekindled in the UK and other countries by directives mandating certain forms of acetabular and femoral component fixation based exclusively on the chronological age of the patient. This editorial focuses on the literature-based arguments to support the use of cementless total hip arthroplasty (THA), while addressing potential concerns surrounding safety and cost-effectiveness. Cite this article:
Knee osteonecrosis in advanced stages may lead to joint degeneration. Total knee arthroplasty (TKA) for osteonecrosis has traditionally been associated with suboptimal results. We analyzed outcomes of contemporary TKAs for osteonecrosis, with particular emphasis on: survivorship free from aseptic loosening, any revision, and any reoperation plus the clinical outcomes, complications, and radiological results. In total, 156 patients undergoing 167 primary TKAs performed for osteonecrosis between 2004 and 2014 at a single institution were reviewed. The mean age at index TKA was 61 years (14 to 93) and the mean body mass index (BMI) was 30 kg/m2 (18 to 51) The mean follow-up was six years (2 to 12). A total of 110 TKAs (66%) were performed for primary osteonecrosis and 57 TKAs (34%) for secondary osteonecrosis. Overall, 15 TKAs (9%) had tibial stems, while 12 TKAs (7%) had femoral stems. Posterior-stabilized designs were used in 147 TKAs (88%) of TKAs. Bivariate Cox regression analysis was conducted to identify risk factors for revision and reoperation.Aims
Patients and Methods
Previous studies have evidenced cement-in-cement techniques as reliable in revision arthroplasty. Commonly, the original cement mantle is reshaped, aiding accurate placement of the new stem. Ultrasonic devices selectively remove cement, preserve host bone, and have lower cortical perforation rates than other techniques. As far as the authors are aware, the impact of ultrasonic devices on final cement-in-cement bonds has not been investigated. This study assessed the impact of cement removal using the Orthosonics System for Cemented Arthroplasty Revision (OSCAR; Orthosonics) on final cement-in-cement bonds. A total of 24 specimens were manufactured by pouring cement (Simplex P Bone Cement; Stryker) into stainless steel moulds, with a central rod polished to Stryker Exeter V40 specifications. After cement curing, the rods were removed and eight specimens were allocated to each of three internal surface preparation groups: 1) burr; 2) OSCAR; and 3) no treatment. Internal holes were recemented, and each specimen was cut into 5 mm discs. Shear testing of discs was completed by a technician blinded to the original grouping, recording ultimate shear strengths. Scanning electron microscopy (SEM) was completed, inspecting surfaces of shear-tested specimens.Objectives
Methods
The purpose of this exploratory study was to investigate if the 24-hour activity profile (i.e. waking activities and sleep) objectively measured using wrist-worn accelerometry of patients scheduled for total hip arthroplasty (THA) improves postoperatively. A total of 51 THA patients with a mean age of 64 years (24 to 87) were recruited from a single public hospital. All patients underwent THA using the same surgical approach with the same prosthesis type. The 24-hour activity profiles were captured using wrist-worn accelerometers preoperatively and at 2, 6, 12, and 26 weeks postoperatively. Patient-reported outcomes (Hip Disability and Osteoarthritis Outcome Score (HOOS)) were collected at all timepoints except two weeks postoperatively. Accelerometry data were used to quantify the intensity (sedentary, light, moderate, and vigorous activities) and frequency (bouts) of activity during the day and sleep efficiency. The analysis investigated changes with time and differences between Charnley class.Aims
Patients and Methods
Periprosthetic femoral fractures (PFF) following total hip arthroplasty
(THA) are devastating complications that are associated with functional
limitations and increased overall mortality. Although cementless
implants have been associated with an increased risk of PFF, the
precise contribution of implant geometry and design on the risk
of both intra-operative and post-operative PFF remains poorly investigated.
A systematic review was performed to aggregate all of the PFF literature
with specific attention to the femoral implant used. A systematic search strategy of several journal databases and
recent proceedings from the American Academy of Orthopaedic Surgeons
was performed. Clinical articles were included for analysis if sufficient
implant description was provided. All articles were reviewed by
two reviewers. A review of fundamental investigations of implant
load-to-failure was performed, with the intent of identifying similar
conclusions from the clinical and fundamental literature.Aims
Patients and Methods
Joint replacement of the hip and knee remain
very satisfactory operations. They are, however, expensive. The
actual manufacturing of the implant represents only 30% of the final
cost, while sales and marketing represent 40%. Recently, the patents
on many well established and successful implants have expired. Companies
have started producing and distributing implants that purport to
replicate existing implants with good long-term results. The aims of this paper are to assess the legality, the monitoring
and cost saving implications of such generic implants. We also assess
how this might affect the traditional orthopaedic implant companies. Cite this article:
We describe our technique and rationale using
hybrid fixation for primary total hip arthroplasty (THA) at the Hospital
for Special Surgery. Modern uncemented acetabular components have
few screw holes, or no holes, polished inner surfaces, improved
locking mechanisms, and maximised thickness and shell-liner conformity. Uncemented
sockets can be combined with highly cross-linked polyethylene liners,
which have demonstrated very low wear and osteolysis rates after
ten to 15 years of implantation. The results of cement fixation
with a smooth or polished surface finished stem have been excellent,
virtually eliminating complications seen with cementless fixation
like peri-operative femoral fractures and thigh pain. Although mid-term
results of modern cementless stems are encouraging, the long-term
data do not show reduced revision rates for cementless stems compared
with cemented smooth stems. In this paper we review the conduct
of a hybrid THA, with emphasis on pre-operative planning, surgical
technique, hypotensive epidural anaesthesia, and intra-operative
physiology. Cite this article:
Revision of a cemented hemiarthroplasty of the
hip may be a hazardous procedure with high rates of intra-operative complications.
Removing well-fixed cement is time consuming and risks damaging
already weak bone or perforating the femoral shaft. The cement-in-cement
method avoids removal of intact cement and has shown good results
when used for revision total hip arthroplasty (THA). The use of
this technique for the revision of a hemiarthroplasty to THA has
not been previously reported. A total of 28 consecutive hemiarthroplasties (in 28 patients)
were revised to a THA using an Exeter stem and the cement-in-cement
technique. There were four men and 24 women; their mean age was
80 years (35 to 93). Clinical and radiographic data, as well as
operative notes, were collected prospectively and no patient was
lost to follow-up. Four patients died within two years of surgery. The mean follow
up of the remainder was 70 months (25 to 124). Intra-operatively
there was one proximal perforation, one crack of the
femoral calcar and one acetabular fracture. No femoral components
have required subsequent revision for aseptic loosening or are radiologically loose. Four patients with late complications (14%) have since undergone
surgery (two for a peri-prosthetic fracture, and one each for deep
infection and recurrent dislocation) resulting in an overall major
rate of complication of 35.7%. The cement-in-cement technique provides reliable femoral fixation
in this elderly population and may reduce operating time and rates
of complication. Cite this article:
We report the five-year outcome of a randomised
controlled trial which used radiostereometric analysis (RSA) to assess
the influence of surface oxidised zirconium (OxZr, Oxinium) on polyethylene
wear A total of 120 patients, 85 women and 35 men with a mean age
of 70 years (59 to 80) who were scheduled for primary cemented total
hip arthroplasty were randomly allocated to four study groups. Patients
were blinded to their group assignment and received either a conventional
polyethylene (CPE) or a highly cross-linked (HXL) acetabular component
of identical design. On the femoral side patients received a 28
mm head made of either cobalt-chromium (CoCr) or OxZr. The proximal head penetration (wear) was measured with repeated
RSA examinations over five years. Clinical outcome was measured
using the Harris hip score. There was no difference in polyethylene wear between the two
head materials when used with either of the two types of acetabular
component (p = 0.3 to 0.6). When comparing the two types of polyethylene
there was a significant difference in favour of HXLPE, regardless
of the head material used (p <
0.001). In conclusion, we found no advantage of OxZr over CoCr in terms
of polyethylene wear after five years of follow-up. Our findings
do not support laboratory results which have shown a reduced rate
of wear with OxZr. They do however add to the evidence on the better
resistance to wear of HXLPE over CPE. Cite this article: