Advertisement for orthosearch.org.uk
Results 1 - 20 of 135
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 38 - 46
1 Jan 2010
Langton DJ Jameson SS Joyce TJ Hallab NJ Natu S Nargol AVF

Early failure associated with adverse reactions to metal debris is an emerging problem after hip resurfacing but the exact mechanism is unclear. We analysed our entire series of 660 metal-on-metal resurfacings (Articular Surface Replacement (ASR) and Birmingham Hip Resurfacing (BHR)) and large-bearing ASR total hip replacements, to establish associations with metal debris-related failures. Clinical and radiological outcomes, metal ion levels, explant studies and lymphocyte transformation tests were performed. A total of 17 patients (3.4%) were identified (all ASR bearings) with adverse reactions to metal debris, for which revision was required. This group had significantly smaller components, significantly higher acetabular component anteversion, and significantly higher whole concentrations of blood and joint chromium and cobalt ions than asymptomatic patients did (all p < 0.001). Post-revision lymphocyte transformation tests on this group showed no reactivity to chromium or cobalt ions. Explants from these revisions had greater surface wear than retrievals for uncomplicated fractures. The absence of adverse reactions to metal debris in patients with well-positioned implants usually implies high component wear. Surgeons must consider implant design, expected component size and acetabular component positioning in order to reduce early failures when performing large-bearing metal-on-metal hip resurfacing and replacement


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1501 - 1508
1 Nov 2010
Donell ST Darrah C Nolan JF Wimhurst J Toms A Barker THW Case CP Tucker JK

Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell.

Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%).

At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis.

Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 8 | Pages 1025 - 1030
1 Aug 2009
Ollivere B Darrah C Barker T Nolan J Porteous MJ

The rate and mode of early failure in 463 Birmingham hip resurfacings in a two-centre, multisurgeon series were examined. Of the 463 patients two have died and three were lost to follow-up. The mean radiological and clinical follow-up was for 43 months (6 to 90). We have revised 13 resurfacings (2.8%) including seven for pain, three for fracture, two for dislocation and another for sepsis. Of these, nine had macroscopic and histological evidence of metallosis. The survival at five years was 95.8% (95% confidence interval (CI) 94.1 to 96.8) for revision for all causes and 96.9% (95% CI 95.5 to 98.3) for metallosis. The rate of metallosis related revision was 3.1% at five years. Risk factors for metallosis were female gender, a small femoral component, a high abduction angle and obesity. We do not advocate the use of the Birmingham Hip resurfacing procedure in patients with these risk factors


Bone & Joint Open
Vol. 4, Issue 3 | Pages 182 - 187
14 Mar 2023
Sheridan GA Hanlon M Welch-Phillips A Spratt K Hagan R O'Byrne JM Kenny PJ Kurmis AP Masri BA Garbuz DS Hurson CJ

Aims. Hip resurfacing remains a potentially valuable surgical procedure for appropriately-selected patients with optimised implant choices. However, concern regarding high early failure rates continues to undermine confidence in use. A large contributor to failure is adverse local tissue reactions around metal-on-metal (MoM) bearing surfaces. Such phenomena have been well-explored around MoM total hip arthroplasties, but comparable data in equivalent hip resurfacing procedures is lacking. In order to define genetic predisposition, we performed a case-control study investigating the role of human leucocyte antigen (HLA) genotype in the development of pseudotumours around MoM hip resurfacings. Methods. A matched case-control study was performed using the prospectively-collected database at the host institution. In all, 16 MoM hip resurfacing 'cases' were identified as having symptomatic periprosthetic pseudotumours on preoperative metal artefact reduction sequence (MARS) MRI, and were subsequently histologically confirmed as high-grade aseptic lymphocyte-dominated vasculitis-associated lesions (ALVALs) at revision surgery. ‘Controls’ were matched by implant type in the absence of evidence of pseudotumour. Blood samples from all cases and controls were collected prospectively for high resolution genetic a nalysis targeting 11 separate HLA loci. Statistical significance was set at 0.10 a priori to determine the association between HLA genotype and pseudotumour formation, given the small sample size. Results. Using a previously-reported ALVAL classification, the majority of pseudotumour-positive caseswere found to have intermediate-grade group 2 (n = 10; 63%) or group 3 (n = 4; 25%) histological findings. Two further patients (13%) had high-grade group 4 lesions. HLA-DQB1*05:03:01 (p = 0.0676) and HLA-DRB1*14:54:01 (p = 0.0676) alleles were significantly associated with a higher risk of pseudotumour formation, while HLA-DQA1*03:01:01 (p = 0.0240), HLA-DRB1*04:04:01 (p = 0.0453), HLA-C*01:02:01 (p = 0.0453), and HLA-B*27:05:02 (p = 0.0855) were noted to confer risk reduction. Conclusion. These findings confirm the association between specific HLA genotypes and the risk of pseudotumour development around MoM hip resurfacings. Specifically, the two ‘at risk’ alleles (DQB1*05:03:01 and DRB1*14:54:01) may hold clinical value in preoperative screening and prospective surgical decision-making. Cite this article: Bone Jt Open 2023;4(3):182–187


The Bone & Joint Journal
Vol. 102-B, Issue 6 | Pages 709 - 715
1 Jun 2020
Abdelsamie KR Elhawary I Ali H Ali M EL-Shafie M Dominic Meek RM

Aims. Femoral revision component subsidence has been identified as predicting early failure in revision hip surgery. This comparative cohort study assessed the potential risk factors of subsidence in two commonly used femoral implant designs. Methods. A comparative cohort study was undertaken, analyzing a consecutive series of patients following revision total hip arthroplasties using either a tapered-modular (TM) fluted titanium or a porous-coated cylindrical modular (PCM) titanium femoral component, between April 2006 and May 2018. Clinical and radiological assessment was compared for both treatment cohorts. Risk factors for subsidence were assessed and compared. Results. In total, 65 TM and 35 PCM cases were included. At mean follow-up of seven years (1 to 13), subsidence was noted in both cohorts during the initial three months postoperatively (p < 0.001) then implants stabilized. Subsidence noted in 58.7% (38/65 cases) of the TM cohort (mean 2.3 mm, SD 3.5 mm) compared to 48.8% (17/35) of PCM cohort (mean 1.9 mm, SD 2.6 mm; p = 0.344). Subsidence of PCM cohort were significantly associated with extended trochanteric osteotomy (ETO) (p < 0.041). Although the ETO was used less frequently in PCM stem cohort (7/35), subsidence was noted in 85% (6/7) of them. Significant improvement of the final mean Oxford Hip Score (OHS) was reported in both treatment groups (p < 0.001). Conclusion. Both modular TM and PCM revision femoral components subsided within the femur. TM implants subsided more frequently than PCM components if the femur was intact but with no difference in clinical outcomes. However, if an ETO is performed then a PCM component will subside significantly more and suggests the use of a TM implant may be advisable. Cite this article: Bone Joint J 2020;102-B(6):709–715


Bone & Joint Open
Vol. 4, Issue 11 | Pages 853 - 858
10 Nov 2023
Subbiah Ponniah H Logishetty K Edwards TC Singer GC

Aims

Metal-on-metal hip resurfacing (MoM-HR) has seen decreased usage due to safety and longevity concerns. Joint registries have highlighted the risks in females, smaller hips, and hip dysplasia. This study aimed to identify if reported risk factors are linked to revision in a long-term follow-up of MoM-HR performed by a non-designer surgeon.

Methods

A retrospective review of consecutive MoM hip arthroplasties (MoM-HRAs) using Birmingham Hip Resurfacing was conducted. Data on procedure side, indication, implant sizes and orientation, highest blood cobalt and chromium ion concentrations, and all-cause revision were collected from local and UK National Joint Registry records.


The Bone & Joint Journal
Vol. 105-B, Issue 9 | Pages 946 - 952
1 Sep 2023
Dhawan R Young DA Van Eemeren A Shimmin A

Aims

The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit.

Methods

Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis.


Bone & Joint Open
Vol. 5, Issue 6 | Pages 514 - 523
24 Jun 2024
Fishley W Nandra R Carluke I Partington PF Reed MR Kramer DJ Wilson MJ Hubble MJW Howell JR Whitehouse SL Petheram TG Kassam AM

Aims

In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component.

Methods

Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). Patients underwent follow-up of minimum two-years (mean 4.6 (2.1 to 8.5) years).


The Bone & Joint Journal
Vol. 105-B, Issue 7 | Pages 775 - 782
1 Jul 2023
Koper MC Spek RWA Reijman M van Es EM Baart SJ Verhaar JAN Bos PK

Aims

The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate.

Methods

A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 154 - 158
1 Feb 2008
Calder JD Hine AL Pearse MF Revell PA

Total hip replacement in patients with advanced osteonecrosis of the femoral head is often complicated by early loosening of the femoral component. Recent evidence has suggested that abnormal bone extending into the proximal femur may be responsible for the early failure of the femoral component. We aimed to identify which patients were at high risk of early failure by evaluating gadolinium-enhanced MR images of histologically-confirmed osteonecrotic lesions beyond the femoral head. Although the MR signal intensity has been shown to correlate well with osteonecrosis in the femoral head, it was found to be relatively insensitive at identifying lesions below the head, with a sensitivity of only 51% and a predictive value of a negative result of only 48%. However, the specificity was 90%, with the predictive value of a positive MRI finding being 86%. Only those patients with osteonecrosis of the femoral head secondary to sickle-cell disease, who are known to be at high risk of early loosening, had changes in the MR signal in the greater trochanter and the femoral shaft. This observation suggests that changes in the MR signal beyond the femoral head may represent osteonecrotic lesions in areas essential for the fixation of the femoral component. Pre-operative identification of such lesions in the neck of the femur may be important when considering hip resurfacing for osteonecrosis of the femoral head, following which early loosening of the femoral component and fracture of the neck are possible complications


The early failure and revision of bimodular primary total hip arthroplasty prostheses requires the identification of the risk factors for material loss and wear at the taper junctions through taper wear analysis. Deviations in taper geometries between revised and pristine modular neck tapers were determined using high resolution tactile measurements. A new algorithm was developed and validated to allow the quantitative analysis of material loss, complementing the standard visual inspection currently used. The algorithm was applied to a sample of 27 retrievals (in situ from 2.9 to 38.1 months) of the withdrawn Rejuvenate modular prosthesis. The mean wear volumes on the flat distal neck piece taper was 3.35 mm. 3. (0.55 to 7.57), mainly occurring in a characteristic pattern in areas with high mechanical loading. Wear volume tended to increase with time to revision (r² = 0.423, p = 0.001). Implant and patient specific data (offset, stem size, patient’s mass, age and body mass index) did not correlate with the amount of material loss observed (p >  0.078). Bilaterally revised implants showed higher amounts of combined total material loss and similar wear patterns on both sides. The consistent wear pattern found in this study has not been reported previously, suggesting that the device design and materials are associated with the failure of this prosthesis. Cite this article: Bone Joint J 2015;97-B:1350–7


The Bone & Joint Journal
Vol. 104-B, Issue 5 | Pages 598 - 603
1 May 2022
Siljander MP Gausden EB Wooster BM Karczewski D Sierra RJ Trousdale RT Abdel MP

Aims

The aim of this study was to evaluate the incidence of liner malseating in two commonly used dual-mobility (DM) designs. Secondary aims included determining the risk of dislocation, survival, and clinical outcomes.

Methods

We retrospectively identified 256 primary total hip arthroplasties (THAs) that included a DM component (144 Stryker MDM and 112 Zimmer-Biomet G7) in 233 patients, performed between January 2012 and December 2019. Postoperative radiographs were reviewed independently for malseating of the liner by five reviewers. The mean age of the patients at the time of THA was 66 years (18 to 93), 166 (65%) were female, and the mean BMI was 30 kg/m2 (17 to 57). The mean follow-up was 3.5 years (2.0 to 9.2).


Bone & Joint Open
Vol. 3, Issue 3 | Pages 229 - 235
11 Mar 2022
Syam K Unnikrishnan PN Lokikere NK Wilson-Theaker W Gambhir A Shah N Porter M

Aims

With increasing burden of revision hip arthroplasty (THA), one of the major challenges is the management of proximal femoral bone loss associated with previous multiple surgeries. Proximal femoral arthroplasty (PFA) has already been popularized for tumour surgeries. Our aim was to describe the outcome of using PFA in these demanding non-neoplastic cases.

Methods

A retrospective review of 25 patients who underwent PFA for non-neoplastic indications between January 2009 and December 2015 was undertaken. Their clinical and radiological outcome, complication rates, and survival were recorded. All patients had the Stanmore Implant – Modular Endo-prosthetic Tumour System (METS).


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 19 - 26
1 Jan 2022
Sevaldsen K Schnell Husby O Lian ØB Farran KM Schnell Husby V

Aims

Highly polished stems with force-closed design have shown satisfactory clinical results despite being related to relatively high early migration. It has been suggested that the minimal thickness of cement mantles surrounding the femoral stem should be 2 mm to 4 mm to avoid aseptic loosening. The line-to-line cementing technique of the femoral stem, designed to achieve stem press-fit, challenges this opinion. We compared the migration of a highly polished stem with force-closed design by standard and line-to-line cementing to investigate whether differences in early migration of the stems occur in a clinical study.

Methods

In this single-blind, randomized controlled, clinical radiostereometric analysis (RSA) study, the migration pattern of the cemented Corail hip stem was compared between line-to-line and standard cementing in 48 arthroplasties. The primary outcome measure was femoral stem migration in terms of rotation and translation around and along with the X-, Y-, and Z- axes measured using model-based RSA at three, 12, and 24 months. A linear mixed-effects model was used for statistical analysis.


The Bone & Joint Journal
Vol. 104-B, Issue 3 | Pages 341 - 351
1 Mar 2022
Fowler TJ Aquilina AL Reed MR Blom AW Sayers A Whitehouse MR

Aims

Total hip arthroplasties (THAs) are performed by surgeons at various stages in training with varying levels of supervision, but we do not know if this is safe practice with comparable outcomes to consultant-performed THA. Our aim was to examine the association between surgeon grade, the senior supervision of trainees, and the risk of revision following THA.

Methods

We performed an observational study using National Joint Registry (NJR) data. We included adult patients who underwent primary THA for osteoarthritis, recorded in the NJR between 2003 and 2016. Exposures were operating surgeon grade (consultant or trainee) and whether or not trainees were directly supervised by a scrubbed consultant. Outcomes were all-cause revision and the indication for revision up to ten years. We used methods of survival analysis, adjusted for patient, operation, and healthcare setting factors.


Aims

To establish the survivorship, function, and metal ion levels in an unselected series of metal-on-metal hip resurfacing arthroplasties (HRAs) performed by a non-designer surgeon.

Methods

We reviewed 105 consecutive HRAs in 83 patients, performed by a single surgeon, at a mean follow-up of 14.9 years (9.3 to 19.1). The cohort included 45 male and 38 female patients, with a mean age of 49.5 years (SD 12.5)


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1333 - 1341
1 Oct 2016
van der Voort P Valstar ER Kaptein BL Fiocco M van der Heide HJL Nelissen RGHH

Aims. The widely used and well-proven Palacos R (a.k.a. Refobacin Palacos R) bone cement is no longer commercially available and was superseded by Refobacin bone cement R and Palacos R + G in 2005. However, the performance of these newly introduced bone cements have not been tested in a phased evidence-based manner, including roentgen stereophotogrammetric analysis (RSA). Patients and Methods. In this blinded, randomised, clinical RSA study, the migration of the Stanmore femoral component was compared between Refobacin bone cement R and Palacos R + G in 62 consecutive total hip arthroplasties. The primary outcome measure was femoral component migration measured using RSA and secondary outcomes were Harris hip score (HHS), Hip disability and Osteoarthritis Outcome Score (HOOS), EuroQol 5D (EQ-5D) and Short Form 36 (SF-36). Results. Femoral component migration was comparable between Refobacin bone cement R and Palacos R + G during the two-year follow-up period with an estimated mean difference of 0.06 mm of subsidence (p = 0.56) and 0.08° of retroversion (p = 0.82). Five hips (three Refobacin bone cement R and two Palacos R + G) showed non-stabilising, continuous migration; the femoral cement mantle in these hips, was mean 0.7 mm thicker (p = 0.02) and there were more radiolucencies at the bone-cement interface (p = 0.004) in comparison to hips showing stabilising migration. Post-operative HHS was comparable throughout the follow-up period (p = 0.62). HOOS, EQ5D, and SF-36 scores were also comparable (p-values >  0.05) at the two-year follow-up point. Conclusion. Refobacin bone cement R and Palacos R + G show comparable component migration and clinical outcome during the first two post-operative years. Hips showing continuous migration are at risk for early failure. However, this seems to be unrelated to cement type, but rather to cementing technique. Cite this article: Bone Joint J 2016;98-B:1333–41


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 33 - 37
1 Jul 2021
Ennin KA Elsharkawy KA Dasgupta S Emerson RH

Aims

To achieve the functional benefits of the direct anterior (DA) approach and the fixation benefits of cemented replacement, this study combined the two techniques posing the following questions: does the limited access of the DA approach adversely affect the cement technique?; and does such a cementing technique reduce the incidence of cementless complications?

Methods

A consecutive series of 341 patients (360 hips) receiving the DA approach between 2016 and 2018 were reviewed. There were 203 cementless stems and 157 cemented stems. Mean age was 75 years (70 to 86) in the cementless group and 76 years (52 to 94) in the cemented group, with 239 (70%) females in the whole series. Femoral complications were compared between the two groups. Mean follow-up was 1.5 years (0.1 to 4.4) for patients in the cementless group and 1.3 years (0.0 to 3.9) for patients in the cemented group.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 53 - 58
1 Jul 2021
Lawrie CM Bechtold D Schwabe M Clohisy JC

Aims

The direct anterior approach (DAA) for total hip arthroplasty (THA) has potential advantages over other approaches and is most commonly performed with the patient in the supine position. We describe a technique for DAA THA with the patient in the lateral decubitus position and report the early clinical and radiological outcomes, the characteristics of the learning curve, and perioperative complications.

Methods

All primary DAA THAs performed in the lateral position by a single surgeon over a four-year period from the surgeon’s first case using the technique were identified from a prospectively collected database. Modified Harris Hip Scores (mHHS) were collected to assess clinical outcome, and routine radiological analysis was performed. Retrospective review of the medical records identified perioperative complications, the characteristics of the learning curve, and revisions.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 9 | Pages 1140 - 1143
1 Sep 2007
Itayem R Arndt A McMinn DJW Daniel J Lundberg A

The early designs of hip resurfacing implants suffered high rates of early failure, making it impossible to obtain valuable mid-term radiostereophotogrammetric (RSA) results. The metal-on-metal Birmingham Hip Resurfacing arthroplasty has shown promising mid-term results and we present here the first mid-term RSA analysis of a hip resurfacing implant. The analysis was performed in 19 hips at five years post-operatively. The mean acetabular component translation and rotation, and femoral component translation were compared with the previous RSA measurements at two and six months, and one and two years. There was no statistical significance (t-test, p ≤ 0.05) between these consecutive movements, indicating the mid-term stability of the implant