The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1791 - 1801
1 Dec 2021
Bhalekar RM Nargol ME Shyam N Nargol AVF Wells SR Collier R Pabbruwe M Joyce TJ Langton DJ

Aims. The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs). Methods. At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays. Results. A total of 114 explanted fixed-bearing TKAs were examined. This included 76 used with contemporary PE inserts which were compared with 15 used with older generation PEs. The Attune and NexGen (central locking) trays were found to have significantly less cement cover than Triathlon and PFC trays (peripheral locking group) (p = 0.001). The median planicity values of the PE inserts used with central locking trays were significantly greater than of those with peripheral locking inserts (205 vs 85 microns; p < 0.001). Attune and NexGen inserts had a characteristic pattern of backside deformation, with the outer edges of the PE deviating inferiorly, leaving the PE margins as the primary areas of articulation. Conclusion. Explanted TKAs with central locking mechanisms were significantly more likely to debond from the cement mantle. The PE inserts of these designs showed characteristic patterns of deformation, which appeared to relate to the manufacturing process and may be exacerbated in vivo. This pattern of deformation was associated with PE wear occurring at the outer edges of the articulation, potentially increasing the frictional torque generated at this interface. Cite this article: Bone Joint J 2021;103-B(12):1791–1801


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1313 - 1318
1 Oct 2017
Nakamura R Komatsu N Fujita K Kuroda K Takahashi M Omi R Katsuki Y Tsuchiya H

Aims. Open wedge high tibial osteotomy (OWHTO) for medial-compartment osteoarthritis of the knee can be complicated by intra-operative lateral hinge fracture (LHF). We aimed to establish the relationship between hinge position and fracture types, and suggest an appropriate hinge position to reduce the risk of this complication. Patients and Methods. Consecutive patients undergoing OWHTO were evaluated on coronal multiplanar reconstruction CT images. Hinge positions were divided into five zones in our new classification, by their relationship to the proximal tibiofibular joint (PTFJ). Fractures were classified into types I, II, and III according to the Takeuchi classification. Results. Among 111 patients undergoing OWHTOs, 22 sustained lateral hinge fractures. Of the 89 patients without fractures, 70 had hinges in the zone within the PTFJ and lateral to the medial margin of the PTFJ (zone WL), just above the PTFJ. Among the five zones, the relative risk of unstable fracture was significantly lower in zone WL (relative risk 0.24, confidence interval 0.17 to 0.34). Conclusion. Zone WL appears to offer the safest position for the placement of the osteotomy hinge when trying to avoid a fracture at the osteotomy site. Cite this article: Bone Joint J 2017;99B10:1313–18


The Bone & Joint Journal
Vol. 100-B, Issue 2 | Pages 152 - 160
1 Feb 2018
Subramanyam KN Khanchandani P Tulajaprasad PV Jaipuria J Mundargi AV

Aims. The aims of this study were to compare the efficacy and safety of intra-articular and intravenous (IV) tranexamic acid (TXA) in controlling perioperative blood loss in total knee arthroplasty (TKA) using a randomized, double-blinded equivalence trial. Patients and Methods. A total of 182 patients aged between 45 and 75 years undergoing unilateral TKA at a tertiary centre were randomized to receive TXA, either 1.5 g intra-articularly after closure of the wound (n = 91) or two doses of 10 mg/kg IV (n = 91). The primary outcome measure was the reduction in the level of haemoglobin (Hb) in the blood on the fifth postoperative day. Secondary outcome measures were the total, visible, and hidden blood losses (TBL, VBL, HBL). We assumed equivalence of the primary outcome in both routes with a margin of ± 0.35gm/dl. Block randomization using computer-generated random numbers was used. The patients and the assessor of outcome were blinded. Results. All patients completed the study. The mean difference in the reduction of the level of Hb between the two groups was -0.0055 gm/dl, with two-sided 95% confidence interval (CI) being -0.29 to 0.27, well within the predefined equivalence margin of ± 0.35gm/dl. The groups were comparable with regard to TBL, VBL, HBL, and complications. No patient needed a blood transfusion. Conclusion. A single intra-articular dose and two IV doses of TXA give equivalent efficacy and safety in the management of blood loss at TKA. Cite this article: Bone Joint J 2018;100-B:152–60


The Bone & Joint Journal
Vol. 106-B, Issue 8 | Pages 817 - 825
1 Aug 2024
Borukhov I Ismailidis P Esposito CI LiArno S Lyon J McEwen PJ

Aims

This study aimed to evaluate if total knee arthroplasty (TKA) femoral components aligned in either mechanical alignment (MA) or kinematic alignment (KA) are more biomimetic concerning trochlear sulcus orientation and restoration of trochlear height.

Methods

Bone surfaces from 1,012 CT scans of non-arthritic femora were segmented using a modelling and analytics system. TKA femoral components (Triathlon; Stryker) were virtually implanted in both MA and KA. Trochlear sulcus orientation was assessed by measuring the distal trochlear sulcus angle (DTSA) in native femora and in KA and MA prosthetic femoral components. Trochlear anatomy restoration was evaluated by measuring the differences in medial, lateral, and sulcus trochlear height between native femora and KA and MA prosthetic femoral components.


Bone & Joint Research
Vol. 13, Issue 7 | Pages 342 - 352
9 Jul 2024
Cheng J Jhan S Chen P Hsu S Wang C Moya D Wu Y Huang C Chou W Wu K

Aims

To explore the efficacy of extracorporeal shockwave therapy (ESWT) in the treatment of osteochondral defect (OCD), and its effects on the levels of transforming growth factor (TGF)-β, bone morphogenetic protein (BMP)-2, -3, -4, -5, and -7 in terms of cartilage and bone regeneration.

Methods

The OCD lesion was created on the trochlear groove of left articular cartilage of femur per rat (40 rats in total). The experimental groups were Sham, OCD, and ESWT (0.25 mJ/mm2, 800 impulses, 4 Hz). The animals were euthanized at 2, 4, 8, and 12 weeks post-treatment, and histopathological analysis, micro-CT scanning, and immunohistochemical staining were performed for the specimens.


Aims

Classifying trochlear dysplasia (TD) is useful to determine the treatment options for patients suffering from patellofemoral instability (PFI). There is no consensus on which classification system is more reliable and reproducible for the purpose of guiding clinicians’ management of PFI. There are also concerns about the validity of the Dejour Classification (DJC), which is the most widely used classification for TD, having only a fair reliability score. The Oswestry-Bristol Classification (OBC) is a recently proposed system of classification of TD, and the authors report a fair-to-good interobserver agreement and good-to-excellent intraobserver agreement in the assessment of TD. The aim of this study was to compare the reliability and reproducibility of these two classifications.

Methods

In all, six assessors (four consultants and two registrars) independently evaluated 100 axial MRIs of the patellofemoral joint (PFJ) for TD and classified them according to OBC and DJC. These assessments were again repeated by all raters after four weeks. The inter- and intraobserver reliability scores were calculated using Cohen’s kappa and Cronbach’s α.


The Bone & Joint Journal
Vol. 105-B, Issue 2 | Pages 148 - 157
1 Feb 2023
Koster LA Rassir R Kaptein BL Sierevelt IN Schager M Nelissen RGHH Nolte PA

Aims

The primary aim of this study was to compare the migration of the femoral and tibial components of the cementless rotating platform Attune and Low Contact Stress (LCS) total knee arthroplasty (TKA) designs, two years postoperatively, using radiostereometric analysis (RSA) in order to assess the risk of the development of aseptic loosening. A secondary aim was to compare clinical and patient-reported outcome measures (PROMs) between the designs.

Methods

A total of 61 TKAs were analyzed in this randomized clinical RSA trial. RSA examinations were performed one day and three, six, 12, and 24 months postoperatively. The maximal total point motion (MPTM), translations, and rotations of the components were analyzed. PROMs and clinical data were collected preoperatively and at six weeks and three, six, 12, and 24 months postoperatively. Linear mixed effect modelling was used for statistical analyses.


Bone & Joint Research
Vol. 12, Issue 1 | Pages 58 - 71
17 Jan 2023
Dagneaux L Limberg AK Owen AR Bettencourt JW Dudakovic A Bayram B Gades NM Sanchez-Sotelo J Berry DJ van Wijnen A Morrey ME Abdel MP

Aims

As has been shown in larger animal models, knee immobilization can lead to arthrofibrotic phenotypes. Our study included 168 C57BL/6J female mice, with 24 serving as controls, and 144 undergoing a knee procedure to induce a contracture without osteoarthritis (OA).

Methods

Experimental knees were immobilized for either four weeks (72 mice) or eight weeks (72 mice), followed by a remobilization period of zero weeks (24 mice), two weeks (24 mice), or four weeks (24 mice) after suture removal. Half of the experimental knees also received an intra-articular injury. Biomechanical data were collected to measure passive extension angle (PEA). Histological data measuring area and thickness of posterior and anterior knee capsules were collected from knee sections.


Bone & Joint Open
Vol. 3, Issue 7 | Pages 589 - 595
1 Jul 2022
Joo PY Chen AF Richards J Law TY Taylor K Marchand K Clark G Collopy D Marchand RC Roche M Mont MA Malkani AL

Aims

The aim of this study was to report patient and clinical outcomes following robotic-assisted total knee arthroplasty (RA-TKA) at multiple institutions with a minimum two-year follow-up.

Methods

This was a multicentre registry study from October 2016 to June 2021 that included 861 primary RA-TKA patients who completed at least one pre- and postoperative patient-reported outcome measure (PROM) questionnaire, including Forgotten Joint Score (FJS), Knee Injury and Osteoarthritis Outcomes Score for Joint Replacement (KOOS JR), and pain out of 100 points. The mean age was 67 years (35 to 86), 452 were male (53%), mean BMI was 31.5 kg/m2 (19 to 58), and 553 (64%) cemented and 308 (36%) cementless implants.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1686 - 1694
1 Nov 2021
Yang H Kwak W Kang SJ Song E Seon J

Aims

To determine the relationship between articular cartilage status and clinical outcomes after medial opening-wedge high tibial osteotomy (MOHTO) for medial compartmental knee osteoarthritis at intermediate follow-up.

Methods

We reviewed 155 patients (155 knees) who underwent MOHTO from January 2008 to December 2016 followed by second-look arthroscopy with a mean 5.3-year follow-up (2.0 to 11.7). Arthroscopic findings were assessed according to the International Cartilage Repair Society (ICRS) Cartilage Repair Assessment (CRA) grading system. Patients were divided into two groups based on the presence of normal or nearly normal quality cartilage in the medial femoral condyle: good (second-look arthroscopic) status (ICRS grade I or II; n = 70), and poor (second-look arthroscopic) status (ICRS grade III or IV; n = 85) groups at the time of second-look arthroscopy. Clinical outcomes were assessed using the International Knee Documentation Committee (IKDC) score, Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and 36-Item Short Form survey.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 6 | Pages 766 - 771
1 Jun 2007
Shannon FJ Cronin JJ Cleary MS Eustace SJ O’Byrne JM

Our aims were to map the tibial footprint of the posterior cruciate ligament (PCL) using MRI in patients undergoing PCL-preserving total knee replacement, and to document the disruption of this footprint as a result of the tibial cut. In 26 consecutive patients plain radiography and MRI of the knee were performed pre-operatively, and plain radiography post-operatively. The lower margin of the PCL footprint was located a mean of 1 mm (−10 to 8) above the upper aspect of the fibular head. The mean surface area was 83 mm. 2. (49 to 142). One-third of patients (8 of 22) had tibial cuts made below the lowest aspect of the PCL footprint (complete removal) and one-third (9 of 22) had cuts extending into the footprint (partial removal). The remaining patients (5 of 22) had footprints unaffected by the cuts, keeping them intact. Our study highlights the wide variation in the location of the tibial PCL footprint when referenced against the fibula. Proximal tibial cuts using conventional jigs resulted in the removal of a significant portion, if not all of the PCL footprint in most of the patients in our study. Our findings suggest that when performing PCL-retaining total knee replacement the tibial attachment of the PCL is often removed


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 81 - 86
1 Jun 2021
Mahfouz MR Abdel Fatah EE Johnson JM Komistek RD

Aims

The objective of this study is to assess the use of ultrasound (US) as a radiation-free imaging modality to reconstruct 3D anatomy of the knee for use in preoperative templating in knee arthroplasty.

Methods

Using an US system, which is fitted with an electromagnetic (EM) tracker that is integrated into the US probe, allows 3D tracking of the probe, femur, and tibia. The raw US radiofrequency (RF) signals are acquired and, using real-time signal processing, bone boundaries are extracted. Bone boundaries and the tracking information are fused in a 3D point cloud for the femur and tibia. Using a statistical shaping model, the patient-specific surface is reconstructed by optimizing bone geometry to match the point clouds. An accuracy analysis was conducted for 17 cadavers by comparing the 3D US models with those created using CT. US scans from 15 users were compared in order to examine the effect of operator variability on the output.


Bone & Joint Open
Vol. 2, Issue 5 | Pages 351 - 358
27 May 2021
Griffiths-Jones W Chen DB Harris IA Bellemans J MacDessi SJ

Aims

Once knee arthritis and deformity have occurred, it is currently not known how to determine a patient’s constitutional (pre-arthritic) limb alignment. The purpose of this study was to describe and validate the arithmetic hip-knee-ankle (aHKA) algorithm as a straightforward method for preoperative planning and intraoperative restoration of the constitutional limb alignment in total knee arthroplasty (TKA).

Methods

A comparative cross-sectional, radiological study was undertaken of 500 normal knees and 500 arthritic knees undergoing TKA. By definition, the aHKA algorithm subtracts the lateral distal femoral angle (LDFA) from the medial proximal tibial angle (MPTA). The mechanical HKA (mHKA) of the normal group was compared to the mHKA of the arthritic group to examine the difference, specifically related to deformity in the latter. The mHKA and aHKA were then compared in the normal group to assess for differences related to joint line convergence. Lastly, the aHKA of both the normal and arthritic groups were compared to test the hypothesis that the aHKA can estimate the constitutional alignment of the limb by sharing a similar centrality and distribution with the normal population.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 119 - 125
1 Jun 2021
Springer BD McInerney J

Aims

There is concern that aggressive target pricing in the new Bundled Payment for Care Improvement Advanced (BPCI-A) penalizes high-performing groups that had achieved low costs through prior experience in bundled payments. We hypothesize that this methodology incorporates unsustainable downward trends on Target Prices and will lead to groups opting out of BPCI Advanced in favour of a traditional fee for service.

Methods

Using the Centers for Medicare and Medicaid Services (CMS) data, we compared the Target Price factors for hospitals and physician groups that participated in both BPCI Classic and BPCI Advanced (legacy groups), with groups that only participated in BPCI Advanced (non-legacy). With rebasing of Target Prices in 2020 and opportunity for participants to drop out, we compared retention rates of hospitals and physician groups enrolled at the onset of BPCI Advanced with current enrolment in 2020.


The Bone & Joint Journal
Vol. 96-B, Issue 4 | Pages 473 - 478
1 Apr 2014
van Jonbergen HPW Scholtes VAB Poolman RW

In the absence of patellar resurfacing, we have previously shown that the use of electrocautery around the margin of the patella improved the one-year clinical outcome of total knee replacement (TKR). In this prospective randomised study we compared the mean 3.7 year (1.1 to 4.2) clinical outcomes of 300 TKRs performed with and without electrocautery of the patellar rim: this is an update of a previous report. The overall prevalence of anterior knee pain was 32% (95% confidence intervals [CI] 26 to 39), and 26% (95% CI 18 to 35) in the intervention group compared with 38% (95% CI 29 to 48) in the control group (chi-squared test; p = 0.06). The overall prevalence of anterior knee pain remained unchanged between the one-year and 3.7 year follow-up (chi-squared test; p = 0.12). The mean total Western Ontario McMasters Universities Osteoarthritis Indices and the American Knee Society knee and function scores at 3.7 years’ follow-up were similar in the intervention and control groups (repeated measures analysis of variance p = 0.43, p = 0.09 and p = 0.59, respectively). There were no complications. A total of ten patients (intervention group three, control group seven) required secondary patellar resurfacing after the first year. Our study suggests that the improved clinical outcome with electrocautery denervation compared with no electrocautery is not maintained at a mean of 3.7 years’ follow-up. Cite this article: Bone Joint J 2014;96-B:473–8


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 3 - 12
1 Jun 2021
Crawford DA Duwelius PJ Sneller MA Morris MJ Hurst JM Berend KR Lombardi AV

Aims

The purpose is to determine the non-inferiority of a smartphone-based exercise educational care management system after primary knee arthroplasty compared with a traditional in-person physiotherapy rehabilitation model.

Methods

A multicentre prospective randomized controlled trial was conducted evaluating the use of a smartphone-based care management system for primary total knee arthroplasty (TKA) and partial knee arthroplasty (PKA). Patients in the control group (n = 244) received the respective institution’s standard of care with formal physiotherapy. The treatment group (n = 208) were provided a smartwatch and smartphone application. Early outcomes assessed included 90-day knee range of movement, EuroQoL five-dimension five-level score, Knee Injury and Osteoarthritis Outcome Score for Joint Replacement (KOOS JR) score, 30-day single leg stance (SLS) time, Time up and Go (TUG) time, and need for manipulation under anaesthesia (MUA).


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1324 - 1330
3 Oct 2020
Herregodts S Verhaeghe M Paridaens R Herregodts J Vermue H Arnout N De Baets P Victor J

Aims

Inadvertent soft tissue damage caused by the oscillating saw during total knee arthroplasty (TKA) occurs when the sawblade passes beyond the bony boundaries into the soft tissue. The primary objective of this study is to assess the risk of inadvertent soft tissue damage during jig-based TKA by evaluating the excursion of the oscillating saw past the bony boundaries. The second objective is the investigation of the relation between this excursion and the surgeon’s experience level.

Methods

A conventional jig-based TKA procedure with medial parapatellar approach was performed on 12 cadaveric knees by three experienced surgeons and three residents. During the proximal tibial resection, the motion of the oscillating saw with respect to the tibia was recorded. The distance of the outer point of this cutting portion to the edge of the bone was defined as the excursion of the oscillating saw. The excursion of the sawblade was evaluated in six zones containing the following structures: medial collateral ligament (MCL), posteromedial corner (PMC), iliotibial band (ITB), lateral collateral ligament (LCL), popliteus tendon (PopT), and neurovascular bundle (NVB).


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 507 - 514
1 Mar 2021
Chang JS Kayani B Wallace C Haddad FS

Aims

Total knee arthroplasty (TKA) using functional alignment aims to implant the components with minimal compromise of the soft-tissue envelope by restoring the plane and obliquity of the non-arthritic joint. The objective of this study was to determine the effect of TKA with functional alignment on mediolateral soft-tissue balance as assessed using intraoperative sensor-guided technology.

Methods

This prospective study included 30 consecutive patients undergoing robotic-assisted TKA using the Stryker PS Triathlon implant with functional alignment. Intraoperative soft-tissue balance was assessed using sensor-guided technology after definitive component implantation; soft-tissue balance was defined as intercompartmental pressure difference (ICPD) of < 15 psi. Medial and lateral compartment pressures were recorded at 10°, 45°, and 90° of knee flexion. This study included 18 females (60%) and 12 males (40%) with a mean age of 65.2 years (SD 9.3). Mean preoperative hip-knee-ankle deformity was 6.3° varus (SD 2.7°).


Bone & Joint Open
Vol. 2, Issue 1 | Pages 48 - 57
19 Jan 2021
Asokan A Plastow R Kayani B Radhakrishnan GT Magan AA Haddad FS

Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes.

Compared to cemented implants, the National Joint Registry (NJR) currently reports higher revision rates in cementless total knee arthroplasty (TKA), but lower in unicompartmental knee arthroplasty (UKA). However, recent studies are beginning to show excellent outcomes with cementless implants, particularly with UKA which has shown superior performance to cemented varieties. Cementless TKA has yet to show long-term benefit, and currently performs equivalently to cemented in short- to medium-term cohort studies. However, with novel concepts including 3D-printed coatings, robotic-assisted surgery, radiostereometric analysis, and kinematic or functional knee alignment principles, it is hoped they may help improve the outcomes of cementless TKA in the long-term. In addition, though cementless implant costs remain higher due to novel implant coatings, it is speculated cost-effectiveness can be achieved through greater surgical efficiency and potential reduction in revision costs. There is paucity of level one data on long-term outcomes between fixation methods and the cost-effectiveness of modern cementless knee arthroplasty.

This review explores recent literature on cementless knee arthroplasty, with regards to clinical outcomes, implant survivorship, complications, and cost-effectiveness; providing a concise update to assist clinicians on implant choice.

Cite this article: Bone Jt Open 2021;2(1):48–57.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 19 - 23
1 Jun 2020
Yayac M Schiller N Austin MS Courtney PM

Aims

The purpose of this study was to determine the impact of the removal of total knee arthroplasty (TKA) from the Medicare Inpatient Only (IPO) list on our Bundled Payments for Care Improvement (BPCI) Initiative in 2018.

Methods

We examined our institutional database to identify all Medicare patients who underwent primary TKA from 2017 to 2018. Hospital inpatient or outpatient status was cross-referenced with Centers for Medicare & Medicaid Services (CMS) claims data. Demographics, comorbidities, and outcomes were compared between patients classified as ‘outpatient’ and ‘inpatient’ TKA. Episode-of-care BPCI costs were then compared from 2017 to 2018.