Advertisement for orthosearch.org.uk
Results 1 - 20 of 48
Results per page:
Bone & Joint Research
Vol. 8, Issue 7 | Pages 313 - 322
1 Jul 2019
Law GW Wong YR Yew AK Choh ACT Koh JSB Howe TS

Objectives. The paradoxical migration of the femoral neck element (FNE) superomedially against gravity, with respect to the intramedullary component of the cephalomedullary device, is a poorly understood phenomenon increasingly seen in the management of pertrochanteric hip fractures with the intramedullary nail. The aim of this study was to investigate the role of bidirectional loading on the medial migration phenomenon, based on unique wear patterns seen on scanning electron microscopy of retrieved implants suggestive of FNE toggling. Methods. A total of 18 synthetic femurs (Sawbones, Vashon Island, Washington) with comminuted pertrochanteric fractures were divided into three groups (n = 6 per group). Fracture fixation was performed using the Proximal Femoral Nail Antirotation (PFNA) implant (Synthes, Oberdorf, Switzerland; n = 6). Group 1 was subjected to unidirectional compression loading (600 N), with an elastomer (70A durometer) replacing loose fracture fragments to simulate surrounding soft-tissue tensioning. Group 2 was subjected to bidirectional loading (600 N compression loading, 120 N tensile loading), also with the elastomer replacing loose fracture fragments. Group 3 was subjected to bidirectional loading (600 N compression loading, 120 N tensile loading) without the elastomer. All constructs were tested at 2 Hz for 5000 cycles or until cut-out occurred. The medial migration distance (MMD) was recorded at the end of the testing cycles. Results. The MMDs for Groups 1, 2, and 3 were 1.02 mm, 6.27 mm, and 5.44 mm respectively, with reliable reproduction of medial migration seen in all groups. Bidirectional loading groups showed significantly higher MMDs compared with the unidirectional loading group (p < 0.01). Conclusion. Our results demonstrate significant contributions of bidirectional cyclic loading to the medial migration phenomenon in cephalomedullary nail fixation of pertrochanteric hip fractures. Cite this article: G. W. Law, Y. R. Wong, A. K-S. Yew, A. C. T. Choh, J. S. B. Koh, T. S. Howe. Medial migration in cephalomedullary nail fixation of pertrochanteric hip fractures: A biomechanical analysis using a novel bidirectional cyclic loading model. Bone Joint Res 2019;8:313–322. DOI: 10.1302/2046-3758.87.BJR-2018-0271.R1


Aims. There are concerns regarding nail/medullary canal mismatch and initial stability after cephalomedullary nailing in unstable pertrochanteric fractures. This study aimed to investigate the effect of an additional anteroposterior blocking screw on fixation stability in unstable pertrochanteric fracture models with a nail/medullary canal mismatch after short cephalomedullary nail (CMN) fixation. Methods. Eight finite element models (FEMs), comprising four different femoral diameters, with and without blocking screws, were constructed, and unstable intertrochanteric fractures fixed with short CMNs were reproduced in all FEMs. Micromotions of distal shaft fragment related to proximal fragment, and stress concentrations at the nail construct were measured. Results. Micromotions in FEMs without a blocking screw significantly increased as nail/medullary canal mismatch increased, but were similar between FEMs with a blocking screw regardless of mismatch. Stress concentration at the nail construct was observed at the junction of the nail body and lag screw in all FEMs, and increased as nail/medullary canal mismatch increased, regardless of blocking screws. Mean stresses over regions of interest in FEMs with a blocking screw were much lower than regions of interest in those without. Mean stresses in FEMs with a blocking screw were lower than the yield strength, yet mean stresses in FEMs without blocking screws having 8 mm and 10 mm mismatch exceeded the yield strength. All mean stresses at distal locking screws were less than the yield strength. Conclusion. Using an additional anteroposterior blocking screw may be a simple and effective method to enhance fixation stability in unstable pertrochanteric fractures with a large nail/medullary canal mismatch due to osteoporosis. Cite this article: Bone Joint Res 2022;11(3):152–161


Bone & Joint Research
Vol. 6, Issue 8 | Pages 481 - 488
1 Aug 2017
Caruso G Bonomo M Valpiani G Salvatori G Gildone A Lorusso V Massari L

Objectives. Intramedullary fixation is considered the most stable treatment for pertrochanteric fractures of the proximal femur and cut-out is one of the most frequent mechanical complications. In order to determine the role of clinical variables and radiological parameters in predicting the risk of this complication, we analysed the data pertaining to a group of patients recruited over the course of six years. Methods. A total of 571 patients were included in this study, which analysed the incidence of cut-out in relation to several clinical variables: age; gender; the AO Foundation and Orthopaedic Trauma Association classification system (AO/OTA); type of nail; cervical-diaphyseal angle; surgical wait times; anti-osteoporotic medication; complete post-operative weight bearing; and radiological parameters (namely the lag-screw position with respect to the femoral head, the Cleveland system, the tip-apex distance (TAD), and the calcar-referenced tip-apex distance (CalTAD)). Results. The incidence of cut-out across the sample was 5.6%, with a higher incidence in female patients. A significantly higher risk of this complication was correlated with lag-screw tip positioning in the upper part of the femoral head in the anteroposterior radiological view, posterior in the latero-lateral radiological view, and in the Cleveland peripheral zones. The tip-apex distance and the calcar-referenced tip-apex distance were found to be highly significant predictors of the risk of cut-out at cut-offs of 30.7 mm and 37.3 mm, respectively, but the former appeared more reliable than the latter in predicting the occurrence of this complication. Conclusion. The tip-apex distance remains the most accurate predictor of cut-out, which is significantly greater above a cut-off of 30.7 mm. Cite this article: G. Caruso, M. Bonomo, G. Valpiani, G. Salvatori, A. Gildone, V. Lorusso, L. Massari. A six-year retrospective analysis of cut-out risk predictors in cephalomedullary nailing for pertrochanteric fractures: Can the tip-apex distance (TAD) still be considered the best parameter?. Bone Joint Res 2017;6:481–488. DOI: 10.1302/2046-3758.68.BJR-2016-0299.R1


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 3 | Pages 350 - 358
1 Apr 2004
Karachalios T Lyritis GP Kaloudis J Roidis N Katsiri M

We investigated the effect of calcitonin in the prevention of acute bone loss after a pertrochanteric fracture and its ability to reduce the incidence of further fractures in the same patient. Fifty women aged between 70 and 80 years who had a pertrochanteric fracture of the hip were randomly allocated to group A (200 IU of nasal salmon calcitonin daily for three months) or group B (placebo). Patients in group A showed a significantly higher level of total alkaline phosphatase and osteocalcin on the 15th day after injury and a significantly higher level of bone alkaline phosphatase on the 90th day after surgery. These patients also had significantly lower levels of urinary C-telopeptide (CrossLaps) on the 15th, 45th and 90th days after injury and lower levels of urinary hydroxyproline on the 15th and 45th days after injury. Patients in group A had significantly higher bone mineral density at all recorded sites except the greater trochanter at three months and one year after operation. After a four-year period of clinical observation, five patients (24%) in group B sustained a new fracture, in four of whom (20%) it was of the contralateral hip. Our findings show that calcitonin reduces acute bone loss in patients with pertrochanteric fractures and may prevent the occurrence of new fractures of the contralateral hip in the elderly


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 1 | Pages 76 - 81
1 Jan 2005
Pajarinen J Lindahl J Michelsson O Savolainen V Hirvensalo E

We treated 108 patients with a pertrochanteric femoral fracture using either the dynamic hip screw or the proximal femoral nail in this prospective, randomised series. We compared walking ability before fracture, intra-operative variables and return to their residence. Patients treated with the proximal femoral nail (n = 42) had regained their pre-operative walking ability significantly (p = 0.04) more often by the four-month review than those treated with the dynamic hip screw (n = 41). Peri-operative or immediate post-operative measures of outcome did not differ between the groups, with the exception of operation time. The dynamic hip screw allowed a significantly greater compression of the fracture during the four-month follow-up, but consolidation of the fracture was comparable between the two groups. Two major losses of reduction were observed in each group, resulting in a total of four revision operations. Our results suggest that the use of the proximal femoral nail may allow a faster postoperative restoration of walking ability, when compared with the dynamic hip screw


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 1 | Pages 23 - 29
1 Jan 2002
Vossinakis IC Badras LS

In a prospective, randomised study we have compared the pertrochanteric external fixator (PF) with the sliding hip screw (SHS) in 100 consecutive patients who were allocated randomly to the two methods of treatment. Details of the patients and the patterns of fracture were similar in both groups. Follow-up was for six months. Use of the PF was associated with significantly less blood loss, a shorter operating time, reduced postoperative pain, shorter hospitalisation (p < 0.001), earlier mobilisation (p < 0.001) and a reduced rate of mechanical complications (p < 0.01). Superficial infection was significantly more common with the PF (p < 0.01), but without long-term adverse consequences. There were no differences in the healing of the fracture, mortality or final functional outcome. Our results indicate that the external fixator is an effective and safe device for treating pertrochanteric fractures and should be considered as a useful alternative to conventional fixation with the sliding hip screw


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 110 - 115
1 Jan 2010
Bosma E de Jongh MAC Verhofstad MHJ

This retrospective cohort study was conducted to investigate whether operative treatment of patients with a pertrochanteric femoral fracture outside working hours is associated with an increased risk of complications and higher mortality. During the study period 165 patients were operated on outside working hours and 123 were operated on during working hours (08.00 to 17.00). There was no difference in the rate of early complications (outside working hours 33% versus working hours 33%, p = 0.91) or total complications during follow-up (outside working hours 40% versus working hours 41%, p = 0.91). Both in-hospital mortality (outside working hours 12% versus working hours 11%, p = 0.97) and mortality after one year (outside working hours 29% versus working hours 27%, p = 0.67) were comparable. Adjustment for possible confounders by multivariate logistic regression analysis revealed no increased risk of complications when patients were operated on outside working hours. On the basis of these data, there is no medical reason to postpone operative reduction and fixation in patients with a proximal femoral fracture until working hours


The Journal of Bone & Joint Surgery British Volume
Vol. 40-B, Issue 4 | Pages 694 - 700
1 Nov 1958
Bremner RA Graham WD

1. The results of treatment of 100 consecutive patients with pertrochanteric and basal fractures of the femur treated by early operative fixation with a McKee two-piece nail and plate are reviewed. 2. Technical failures are analysed and discussed. 3. The pattern of mortality is discussed and contrasted with that in a comparable series of patients treated conservatively. 4. It is concluded that early operative fixation is the method of choice in the management of these fractures, and that the McKee pin and plate is a satisfactory and reliable device for securing internal fixation


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 430 - 434
1 May 2024
Eardley WGP


The Journal of Bone & Joint Surgery British Volume
Vol. 39-B, Issue 3 | Pages 436 - 437
1 Aug 1957
Capener N


The Bone & Joint Journal
Vol. 99-B, Issue 1 | Pages 128 - 133
1 Jan 2017
Socci AR Casemyr NE Leslie MP Baumgaertner MR

Aims

The aim of this paper is to review the evidence relating to the anatomy of the proximal femur, the geometry of the fracture and the characteristics of implants and methods of fixation of intertrochanteric fractures of the hip.

Materials and Methods

Relevant papers were identified from appropriate clinical databases and a narrative review was undertaken.


The Bone & Joint Journal
Vol. 96-B, Issue 9 | Pages 1192 - 1197
1 Sep 2014
Egol KA Marcano AI Lewis L Tejwani NC McLaurin TM Davidovitch RI

In March 2012, an algorithm for the treatment of intertrochanteric fractures of the hip was introduced in our academic department of Orthopaedic Surgery. It included the use of specified implants for particular patterns of fracture. In this cohort study, 102 consecutive patients presenting with an intertrochanteric fracture were followed prospectively (post-algorithm group). Another 117 consecutive patients who had been treated immediately prior to the implementation of the algorithm were identified retrospectively as a control group (pre-algorithm group). The total cost of the implants prior to implementation of the algorithm was $357 457 (mean: $3055 (1947 to 4133)); compared with $255 120 (mean: $2501 (1052 to 4133)) after its implementation. There was a trend toward fewer complications in patients who were treated using the algorithm (33% pre- versus 22.5% post-algorithm; p = 0.088). Application of the algorithm to the pre-algorithm group revealed a potential overall cost saving of $70 295.

The implementation of an evidence-based algorithm for the treatment of intertrochanteric fractures reduced costs while maintaining quality of care with a lower rate of complications and re-admissions.

Cite this article: Bone Joint J 2014;96-B:1192–7.


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 686 - 689
1 May 2016
Griffin XL Parsons N McArthur J Achten J Costa ML

Aims. The aim of this study was to inform a definitive trial which could determine the clinical effectiveness of the X-Bolt Dynamic Hip Plating System compared with the sliding hip screw for patients with complex pertrochanteric fragility fractures of the femur. Patients and Methods. This was a single centre, participant blinded, randomised, standard-of-care controlled pilot trial. Patients aged 60 years and over with AO/ASIF A2 and A3 type femoral pertrochanteric fractures were eligible. Results. The primary outcome was the EuroQoL 5 Dimension Score (EQ-5D-3L) at one year following index fixation. A total of 100 participants were recruited, and primary outcome data were available for 88 patients following losses to follow-up and withdrawals. The mean difference in EQ-5D was 0.03 (95% confidence interval -0.17, 0.120; p = 0.720.) There were no significant differences in any of the secondary outcomes measures. The recruitment and follow-up rates from this feasibility study were as predicted. Conclusion. A definitive trial with 90% power to find a clinically important difference in EQ-5D would require 964 participants based upon the data from this study. We plan to start recruitment to this trial in Spring 2016. Take home message: A definitive trial of X-Bolt Dynamic Hip Plating System is feasible and should be conducted now in order to quantify the clinical effectiveness of this novel implant. Cite this article: Bone Joint J 2016;98-B:686–9


Bone & Joint 360
Vol. 2, Issue 1 | Pages 30 - 32
1 Feb 2013

The February 2013 Trauma Roundup. 360 . looks at: the risk of ankle fractures; absorbable implants; minimally invasive heel fracture fixation; pertrochanteric fractures; arthroplasty and intracapsular hip fractures; and extensor mechanism disruption


Bone & Joint Open
Vol. 5, Issue 4 | Pages 294 - 303
11 Apr 2024
Smolle MA Fischerauer SF Vukic I Leitner L Puchwein P Widhalm H Leithner A Sadoghi P

Aims

Patients with proximal femoral fractures (PFFs) are often multimorbid, thus unplanned readmissions following surgery are common. We therefore aimed to analyze 30-day and one-year readmission rates, reasons for, and factors associated with, readmission risk in a cohort of patients with surgically treated PFFs across Austria.

Methods

Data from 11,270 patients with PFFs, treated surgically (osteosyntheses, n = 6,435; endoprostheses, n = 4,835) at Austrian hospitals within a one-year period (January to December 2021) was retrieved from the Leistungsorientierte Krankenanstaltenfinanzierung (Achievement-Oriented Hospital Financing). The 30-day and one-year readmission rates were reported. Readmission risk for any complication, as well as general medicine-, internal medicine-, and surgery/injury-associated complications, and factors associated with readmissions, were investigated.


Bone & Joint 360
Vol. 12, Issue 3 | Pages 37 - 40
1 Jun 2023

The April 2023 Children’s orthopaedics Roundup360 looks at: CT scan of the ipsilateral femoral neck in paediatric shaft fractures; Meniscal injuries in skeletally immature children with tibial eminence fractures: a systematic literature review; Post-maturity progression in adolescent idiopathic scoliosis curves of 40° to 50°; Prospective, randomized Ponseti treatment for clubfoot: orthopaedic surgeons versus physical therapists; FIFA 11+ Kids: challenges in implementing a prevention programme; The management of developmental dysplasia of the hip in children aged under three months: a consensus study from the British Society for Children's Orthopaedic Surgery; Early investigation and bracing in developmental dysplasia of the hip impacts maternal wellbeing and breastfeeding; Hip arthrodesis in children: a review of 26 cases with a mean of 20 years’ follow-up


Bone & Joint Open
Vol. 5, Issue 10 | Pages 843 - 850
8 Oct 2024
Greve K Ek S Bartha E Modig K Hedström M

Aims

The primary aim of this study was to compare surgical methods (sliding hip screw (SHS) vs intramedullary nailing (IMN)) for trochanteric hip fracture in relation to death within 120 days after surgery and return to independent living. The secondary aim was to assess whether the associations between surgical method and death or ability to return to independent living varied depending on fracture subtype or other patient characteristics.

Methods

A total of 27,530 individuals from the Swedish Hip Fracture Register RIKSHÖFT (SHR) aged ≥ 70 years, admitted to hospital between 1 January 2014 and 31 December 2019 with trochanteric hip fracture, were included. Within this cohort, 12,041 individuals lived independently at baseline, had follow-up information in the SHR, and were thus investigated for return to independent living. Death within 120 days after surgery was analyzed using Cox regression with SHS as reference and adjusted for age and fracture type. Return to independent living was analyzed using logistic regression adjusted for age and fracture type. Analyses were repeated after stratification by fracture type, age, and sex.


Bone & Joint Open
Vol. 4, Issue 10 | Pages 801 - 807
23 Oct 2023
Walter N Szymski D Kurtz SM Lowenberg DW Alt V Lau EC Rupp M

Aims

This work aimed at answering the following research questions: 1) What is the rate of mechanical complications, nonunion and infection for head/neck femoral fractures, intertrochanteric fractures, and subtrochanteric fractures in the elderly USA population? and 2) Which factors influence adverse outcomes?

Methods

Proximal femoral fractures occurred between 1 January 2009 and 31 December 2019 were identified from the Medicare Physician Service Records Data Base. The Kaplan-Meier method with Fine and Gray sub-distribution adaptation was used to determine rates for nonunion, infection, and mechanical complications. Semiparametric Cox regression model was applied incorporating 23 measures as covariates to identify risk factors.


Bone & Joint Open
Vol. 3, Issue 10 | Pages 741 - 745
1 Oct 2022
Baldock TE Dixon JR Koubaesh C Johansen A Eardley WGP

Aims

Patients with A1 and A2 trochanteric hip fractures represent a substantial proportion of trauma caseload, and national guidelines recommend that sliding hip screws (SHS) should be used for these injuries. Despite this, intramedullary nails (IMNs) are routinely implanted in many hospitals, at extra cost and with unproven patient outcome benefit. We have used data from the National Hip Fracture Database (NHFD) to examine the use of SHS and IMN for A1 and A2 hip fractures at a national level, and to define the cost implications of management decisions that run counter to national guidelines.

Methods

We used the NHFD to identify all operations for fixation of trochanteric fractures in England and Wales between 1 January 2021 and 31 December 2021. A uniform price band from each of three hip fracture implant manufacturers was used to set cost implications alongside variation in implant use.


The Bone & Joint Journal
Vol. 104-B, Issue 10 | Pages 1156 - 1167
1 Oct 2022
Holleyman RJ Khan SK Charlett A Inman DS Johansen A Brown C Barnard S Fox S Baker PN Deehan D Burton P Gregson CL

Aims

Hip fracture commonly affects the frailest patients, of whom many are care-dependent, with a disproportionate risk of contracting COVID-19. We examined the impact of COVID-19 infection on hip fracture mortality in England.

Methods

We conducted a cohort study of patients with hip fracture recorded in the National Hip Fracture Database between 1 February 2019 and 31 October 2020 in England. Data were linked to Hospital Episode Statistics to quantify patient characteristics and comorbidities, Office for National Statistics mortality data, and Public Health England’s SARS-CoV-2 testing results. Multivariable Cox regression examined determinants of 90-day mortality. Excess mortality attributable to COVID-19 was quantified using Quasi-Poisson models.