Objectives. We aimed to examine the characteristics of deep venous flow in
the leg in a cast and the effects of a wearable neuromuscular stimulator
(geko; FirstKind Ltd) and also to explore the participants’ tolerance
of the stimulator. Methods. This is an open-label physiological study on ten healthy volunteers.
Duplex ultrasonography of the superficial femoral vein measured
normal flow and cross-sectional area in the standing and supine
positions (with the lower limb initially horizontal and then elevated).
Flow measurements were repeated during activation of the geko stimulator
placed over the peroneal nerve. The process was repeated after the
application of a below-knee cast. Participants evaluated discomfort
using a questionnaire (verbal rating score) and a scoring index
(visual analogue scale). Results. The geko device was effective in significantly increasing venous
blood flow in the lower limb both with a plaster cast (mean difference
11.5 cm/sec. -1. ; p = 0.001 to 0.13) and without a plaster
cast (mean difference 7.7 cm/sec. -1. ; p = 0.001 to 0.75).
Posture also had a significant effect on peak venous blood flow
when the cast was on and the geko inactive (p = 0.003 to 0.69),
although these differences were less pronounced than the effect
of the geko (mean difference 3.1 cm/sec. -1. (-6.5 to 10)).
The geko device was well tolerated, with participants generally
reporting only mild discomfort using the device. Conclusion. The geko device increases venous blood flow in the lower limb,
offering a potential mechanical thromboprolylaxis for patients in
a cast. Cite this article: Bone Joint Res 2013;2:179–85