Peri-prosthetic infection remains a leading cause
of revision surgery. Recent publications from the American Musculoskeletal
Infection Society have sought to establish a definition of peri-prosthetic
infection based on clinical findings and laboratory investigations.
The limitations of their approach are discussed and an alternative
definition is proposed, which it is felt may better reflect the
uncertainties encountered in clinical practice.
Symptomatic cobalt toxicity from a failed total
hip replacement is a rare but devastating complication. It has been reported
following revision of fractured ceramic components, as well as in
patients with failed metal-on-metal articulations. Potential clinical
findings include fatigue, weakness, hypothyroidism, cardiomyopathy,
polycythaemia, visual and hearing impairment, cognitive dysfunction,
and neuropathy. We report a case of an otherwise healthy 46-year-old
patient, who developed progressively worsening symptoms of cobalt
toxicity beginning approximately six months following synovectomy
and revision of a fractured ceramic-on-ceramic total hip replacement
to a metal-on-polyethylene bearing. The whole blood cobalt levels
peaked at 6521 µg/l. The patient died from cobalt-induced cardiomyopathy.
Implant retrieval analysis confirmed a loss of 28.3 g mass of the
cobalt–chromium femoral head as a result of severe abrasive wear
by ceramic particles embedded in the revision polyethylene liner.
Autopsy findings were consistent with heavy metal-induced cardiomyopathy. We recommend using new ceramics at revision to minimise the risk
of wear-related cobalt toxicity following breakage of ceramic components. Cite this article:
A consecutive series of 320 patients with an
intracapsular fracture of the hip treated with a dynamic locking
plate (Targon Femoral Neck (TFN)) were reviewed. All surviving patients
were followed for a minimum of two years. During the follow-up period
109 patients died. There were 112 undisplaced fractures, of which three (2.7%) developed
nonunion or re-displacement and five (4.5%) developed avascular
necrosis of the femoral head. Revision to an arthroplasty was required
for five patients (4.5%). A further six patients (5.4%) had elective
removal of the plate and screws. There were 208 displaced fractures, of which 32 (15.4%) developed
nonunion or re-displacement and 23 (11.1%) developed avascular necrosis.
A further four patients (1.9%) developed a secondary fracture around
the TFN. Revision to a hip replacement was required for 43 patients
(20.7%) patients and a further seven (3.3%) had elective removal
of the plate and screws. It is suggested that the stronger distal fixation combined with
rotational stability may lead to a reduced incidence of complications
related to the healing of the fracture when compared with other
contemporary fixation devices but this needs to be confirmed in
further studies. Cite this article:
We prospectively followed 191 consecutive collarless
polished tapered (CPT) femoral stems, implanted in 175 patients
who had a mean age at operation of 64.5 years (21 to 85). At a mean
follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were
still alive. The fate of all original stems is known. The 16-year
survivorship with re-operation for any reason was 80.7% (95% confidence
interval 72 to 89.4). There was no loss to follow-up, with clinical
data available on all 95 hips and radiological assessment performed
on 90 hips (95%). At latest follow-up, the mean Harris hip score
was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48).
Stems subsided within the cement mantle, with a mean subsidence
of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem
(0.5%) has been revised due to aseptic loosening. In total seven
stems were revised for any cause, of which four revisions were required
for infection following revision of the acetabular component. A
total of 21 patients (11%) required some sort of revision procedure;
all except three of these resulted from failure of the acetabular
component. Cemented acetabular components had a significantly lower
revision burden (three hips, 2.7%) than Harris Galante uncemented
components (17 hips, 21.8%) (p <
0.001). The CPT stem continues to provide excellent radiological and
clinical outcomes at 15 years following implantation. Its results
are consistent with other polished tapered stem designs.
Allografts of bone from the femoral head are often used in orthopaedic procedures. Although the donated heads are thoroughly tested microscopically before release by the bone bank, some surgeons take additional cultures in the operating theatre before implantation. There is no consensus about the need to take these cultures. We retrospectively assessed the clinical significance of the implantation of positive-cultured bone allografts. The contamination rate at retrieval of the allografts was 6.4% in our bone bank. Intra-operative cultures were taken from 426 femoral head allografts before implantation; 48 (11.3%) had a positive culture. The most frequently encountered micro-organism was coagulase-negative staphylococcus. Deep infection occurred in two of the 48 patients (4.2%). In only one was it likely that the same micro-organism caused the contamination and the subsequent infection. In our study, the rate of infection in patients receiving positive-cultured allografts at implantation was not higher than the overall rate of infection in allograft surgery suggesting that the positive cultures at implantation probably represent contamination and that the taking of additional cultures is not useful.
There has been a substantial increase in the
number of hip and knee prostheses implanted in recent years, with
a consequent increase in the number of revisions required. Total
femur replacement (TFR) following destruction of the entire femur,
usually after several previous revision operations, is a rare procedure
but is the only way of avoiding amputation. Intramedullary femur
replacement (IFR) with preservation of the femoral diaphysis is
a modification of TFR. Between 1999 and 2010, 27 patients with non-oncological
conditions underwent surgery in our department with either IFR (n
= 15) or TFR (n = 12) and were included in this study retrospectively.
The aim of the study was to assess the indications, complications
and outcomes of IFR and TFR in revision cases. The mean follow-up
period was 31.3 months (6 to 90). Complications developed in 37%
of cases, 33% in the IFR group and 4% in the TFR group. Despite
a trend towards a slightly better functional outcome compared with
TFR, the indication for intramedullary femur replacement should
be established on a very strict basis in view of the procedure’s
much higher complication rate.
This multicentre study analysed 12 alumina ceramic-on-ceramic
components retrieved from squeaking total hip replacements after
a mean of 23 months
Revision surgery of the hip was performed on 114 hips using an extensively porous-coated femoral component. Of these, 95 hips (94 patients) had a mean follow-up of 10.2 years (5 to 17). No cortical struts were used and the cortical index and the femoral cortical width were measured at different levels. There were two revisions for aseptic loosening. Survivorship at 12 years for all causes of failure was 96.9% (95% confidence interval 93.5 to 100) in the best-case scenario. Fibrous or unstable fixation was associated with major bone defects. The cortical index (p = 0.045) and the lateral cortical thickness (p = 0.008) decreased at the proximal level over time while the medial cortex increased (p = 0.001) at the proximal and distal levels. An increase in the proximal medial cortex was found in patients with an extended transtrochanteric osteotomy (p = 0.026) and in those with components shorter than 25 cm (p = 0.008). The use of the extensively porous-coated femoral component can provide a solution for difficult cases in revision surgery. Radiological bony ingrowth is common. Although without clinical relevance at the end of follow-up, the thickness of the medial femoral cortex often increased while that of the lateral cortex decreased. In cases in which a shorter component was used and in those undertaken using an extended trochanteric osteotomy, there was a greater increase in thickness of the femoral cortex over time.
The differential diagnosis of the painful total
hip arthroplasty (resurfacing or total hip) includes infection,
failure of fixation (loosening), tendinitis, bursitis, synovitis,
adverse local tissue reaction (ALTR) to cobalt-chromium alloys,
and non-hip issues, such as spinal disorders, hernia, gynecologic,
and other pelvic pain. Assuming that the hip is the source of pain,
the first level question is prosthetic or non-prosthetic pain generator?
The second level prosthetic question is septic or aseptic? The third
level question (aseptic hips) is well-fixed or loose? ALTR is best
diagnosed by cross-sectional imaging. Successful treatment is dependent
on correct identification and elimination of the pain generator.
Treatment recommendations for ALTR and taper corrosion are evolving.
Obesity is an epidemic across both the developed
and developing nations that is possibly the most important current
public health factor affecting the morbidity and mortality of the global
population. Obese patients have the potential to pose several challenges
for arthroplasty surgeons from the standpoint of the influence obesity
has on osteoarthritic symptoms, their peri-operative medical management,
the increased intra-operative technical demands on the surgeon,
the intra- and post-operative complications, the long term outcomes
of total hip and knee arthroplasty. Also, there is no consensus
on the role the arthroplasty surgeon should have in facilitating
weight loss for these patients, nor whether obesity should affect
the access to arthroplasty procedures.
Loss of bone stock is a major problem in revision surgery of the hip. Impaction bone grafting of the femur is frequently used when dealing with deficient bone stock. In this retrospective study a consecutive series of 68 patients (69 hips) who had revision of a hip replacement with femoral impaction grafting were reviewed. Irradiated bone allograft was used in all hips. Radiological measurement of subsidence of the stem, incorporation of the graft and remodelling was carried out and showed incorporation of the graft in 26 of 69 hips (38%). However, there was no evidence of trabecular remodelling. Moderate subsidence of 5 mm to 10 mm occurred in ten hips (14.5%), and massive subsidence of >
10 mm in five (7.2%). The results of this study are less favourable than those of others describing studies of revision of the femoral stem using impaction bone grafting. The absence of the characteristic changes of graft remodelling noted in other series raises the question as to whether irradiated bone graft may be a significant factor influencing the post-operative outcome.
Over an eight-month period we prospectively enrolled 122 patients who underwent arthroscopic surgery of the hip for femoroacetabular impingement and met the inclusion criteria for this study. Patients with bilateral hip arthroscopy, avascular necrosis and previous hip surgery were excluded. Ten patients refused to participate leaving 112 in the study. There were 62 women and 50 men. The mean age of the patients was 40.6 yrs (95% confidence interval (CI) 37.7 to 43.5). At arthroscopy, 23 patients underwent osteoplasty only for cam impingement, three underwent rim trimming only for pincer impingement, and 86 underwent both procedures for mixed-type impingement. The mean follow-up was 2.3 years (2.0 to 2.9). The mean modified Harris hip score (HHS) improved from 58 to 84 (mean difference = 24 (95% CI 19 to 28)) and the median patient satisfaction was 9 (1 to 10). Ten patients underwent total hip replacement at a mean of 16 months (8 to 26) after arthroscopy. The predictors of a better outcome were the pre-operative modified HHS (p = 0.018), joint space narrowing ≥ 2 mm (p = 0.005), and repair of labral pathology instead of debridement (p = 0.032). Hip arthroscopy for femoroacetabular impingement, accompanied by suitable rehabilitation, gives a good short-term outcome and high patient satisfaction.
Peri- and sub-prosthetic fractures, or pathological fractures below an existing well-fixed femoral component, with or without an ipsilateral knee replacement, present a difficult surgical challenge. We describe a simple solution, in which a custom-made prosthesis with a cylindrical design is cemented proximally to the stem of an existing, well-fixed femoral component. This effectively treats the fracture without sacrificing the good hip. We describe five patients with a mean age of 73 years (60 to 81) and a mean follow-up of 47 months (6 to 108). The mean overlap of the prosthesis over the femoral component was 7.5 cm (5.5 to 10). There have been no mechanical failures, no new infections and no re-operations. We suggest that in highly selected cases, in which conventional fixation is not feasible, this technique offers a durable option and avoids the morbidity of a total femoral replacement.
Periprosthetic infection following total hip replacement can be a catastrophic complication for the patient. The treatments available include single-stage exchange, and two-stage exchange. We present a series of 50 consecutive patients with a diagnosis of infected total hip replacement who were assessed according to a standardised protocol. Of these, 11 underwent single-stage revision arthroplasty with no recurrence of infection at a mean of 6.8 years follow-up (5.5 to 8.8). The remaining 39 underwent two-stage revision, with two recurrences of infection successfully treated by a second two-stage procedure. At five years, significant differences were found in the mean Harris Hip Scores (single-stage 87.8; two-stage 75.5; p = 0.0003) and in a visual analogue score for satisfaction (8.6; 6.9; p = 0.001) between the single- and two-stage groups. Single-stage exchange is successful in eradicating periprosthetic infection and results in excellent functional and satisfaction scores. Identification of patients suitable for the single-stage procedure allows individualisation of care and provides as many as possible with the correct strategy in successfully tackling their periprosthetic infection
We report a systematic review and meta-analysis of published randomised controlled trials evaluating the efficacy of tranexamic acid (TXA) in reducing blood loss and transfusion in total hip replacement (THR). The data were evaluated using the generic evaluation tool designed by the Cochrane Bone, Joint and Muscle Trauma Group. We identified 11 clinical trials which were suitable for detailed extraction of data. There were no trials that used TXA in revision THR. A total of seven studies (comprising 350 patients) were eligible for the blood loss outcome data. The use of TXA reduced intra-operative blood loss by a mean of 104 ml (95% confidence interval (CI) −164 to −44, p = 0.0006, heterogeneity I2 0%), postoperative blood loss by a mean of 172 ml (95% CI −263 to −81, p = 0.0002, heterogeneity I2 63%) and total blood loss by a mean of 289 ml (95% CI −440 to −138, p <
0.0002, heterogeneity I2 54%). TXA led to a significant reduction in the proportion of patients requiring allogeneic blood transfusion (risk difference −0.20, 95% CI −0.29 to −0.11, p <
0.00001, I2 15%). There were no significant differences in deep-vein thrombosis, pulmonary embolism, infection rates or other complications among the study groups.
We report the long-term results of revision total
hip replacement using femoral impaction allografting with both uncemented
and cemented Freeman femoral components. A standard design of component
was used in both groups, with additional proximal hydroxyapatite
coating in the uncemented group. A total of 33 hips in 30 patients received
an uncemented component and 31 hips in 30 patients a cemented component.
The mean follow-up was 9.8 years (2 to 17) in the uncemented group
and 6.2 years (1 to 11) in the cemented group. Revision procedures
(for all causes) were required in four patients (four hips) in the
uncemented group and in five patients (five hips) in the cemented
group. Harris hip scores improved significantly in both groups and
were maintained independently of the extent of any migration of
the femoral component within the graft or graft–cement mantle.
Between December 2004 and June 2006, 136 patients (156 total hip replacements), were sent from the waiting list of the Cardiff Vale NHS Trust to the NHS Treatment Centre, Weston-super-Mare, in an attempt to reduce the waiting time for total hip replacement. Because of concerns about their outcome, each patient was contacted and invited to attend a review appointment with a consultant specialising in
The aim of this study was to investigate the
possible benefit of large-head metal-on-metal bearing on a stem
for primary hip replacement compared with a 28 mm diameter conventional
metal-on-polyethylene bearing in a prospective randomised controlled
trial. We investigated cemented stem behaviour between these two
different bearings using Einzel-Bild-Röntgen-Analyse, clinical and
patient reported measures (Harris hip score, Western Ontario and
McMaster Universities osteoarthritis index, Short Form-36 and satisfaction)
and whole blood metal ion levels at two years. A power study indicated
that 50 hips were needed in each group to detect subsidence of >
5 mm at two years with a
p-value of <
0.05. Significant improvement (p <
0.001) was found in the mean
clinical and patient reported outcomes at two years for both groups.
Comparison of outcomes between the groups at two years showed no
statistically significant difference for mean stem migration, clinical
and patient reported outcomes; except overall patient satisfaction which
was higher for metal-on-metal group (p = 0.05). Metal ion levels
were raised above the Medicines and Healthcare products Regulatory
Agency advised safety level (7 µg per litre) in 20% of the metal-on-metal
group and in one patient in metal-on-polyethylene group (who had
a metal-on-metal implant on the contralateral side). Two patients
in the metal-on-metal group were revised, one for pseudotumour and
one for peri-prosthetic fracture. Use of large modular heads is associated with a risk of raised
whole blood metal ion levels despite using a proven bearing from
resurfacing. The head-neck junction or excess stem micromotion are
possibly the weak links warranting further research.