An 86-year-old male presented with a loose total hip replacement (THR) ten years after implantation. At revision for anticipated aseptic loosening, watery pus was found in the joint and Bacille-Calmette-Guérin (BCG) was seen on culture. The bacterial strain was identified and was identical to the BCG used in the intravesicular treatment of superficial bladder carcinoma in this patient ten months earlier. After revision he received a full course of antituberculous treatment. The clinical and radiological results were excellent after follow-up for 30 months with his uncemented THR showing satisfactory incorporation. His inflammatory markers were normal and his Harris hip score was 95 points. The diagnosis of a tuberculous infection can be easily missed, but must be considered, especially if sterile pus is encountered.
In a series of 450 patients over 70 years of age with displaced fractures of the femoral neck sustained between 1995 and 1997 treatment was randomised either to internal fixation or replacement. Depending on age and level of activity the latter was either a total hip replacement or a hemiarthroplasty. Patients who were confused or bed-ridden were excluded, as were those with rheumatoid arthritis. At ten years there were 99 failures (45.6%) after internal fixation compared with 17 (8.8%) after replacement. The rate of mortality was high at 75% at ten years, and was the same in both groups at all times. Patient-reported pain and function were similar in both groups at five and ten years. Those with successfully healed fractures had more hip pain and reduction of mobility at four months compared with patients with an uncomplicated replacement, and they never attained a better outcome than the latter patients regarding pain or function. Primary replacement gave reliable long-term results in patients with a displaced fracture of the femoral neck.
We evaluated the cost and consequences of proximal femoral fractures requiring further surgery because of complications. The data were collected prospectively in a standard manner from all patients with a proximal femoral fracture presenting to the trauma unit at the John Radcliffe Hospital over a five-year period. The total cost of treatment for each patient was calculated by separating it into its various components. The risk factors for the complications that arose, the location of their discharge and the mortality rates for these patients were compared to those of a matched control group. There were 2360 proximal femoral fractures in 2257 patients, of which 144 (6.1%) required further surgery. The mean cost of treatment in patients with complications was £18 709 (£2606.30 to £60.827.10), compared with £8610 (£918.54 to £45 601.30) for uncomplicated cases (p <
0.01), with a mean length of stay of 62.8 (44.5 to 79.3) and 32.7 (23.8 to 35.0) days, respectively. The probability of mortality after one month in these cases was significantly higher than in the control group, with a mean survival of 209 days, compared with 496 days for the controls. Patients with complications were statistically less likely to return to their own home (p <
0.01). Greater awareness and understanding are required to minimise the complications of proximal femoral fractures and consequently their cost.
The technique of femoral cement-in-cement revision
is well established, but there are no previous series reporting its
use on the acetabular side at the time of revision total hip replacement.
We describe the technique and report the outcome of 60 consecutive
acetabular cement-in-cement revisions in 59 patients at a mean follow-up
of 8.5 years (5 to 12). All had a radiologically and clinically
well-fixed acetabular cement mantle at the time of revision. During
the follow-up 29 patients died, but no hips were lost to follow-up.
The two most common indications for acetabular revision were recurrent
dislocation (46, 77%) and to complement femoral revision (12, 20%). Of the 60 hips, there were two cases of aseptic loosening of
the acetabular component (3.3%) requiring re-revision. No other
hip was clinically or radiologically loose (96.7%) at the latest
follow-up. One hip was re-revised for infection, four for recurrent
dislocation and one for disarticulation of a constrained component.
At five years the Kaplan-Meier survival rate was 100% for aseptic
loosening and 92.2% (95% CI 84.8 to 99.6), with revision for any cause
as the endpoint. These results support the use of cement-in-cement revision on
the acetabular side in appropriate cases. Theoretical advantages
include preservation of bone stock, reduced operating time, reduced
risk of complications and durable fixation.
The February 2013 Hip &
Pelvis Roundup360 looks at: amazing alumina; dual mobility; white cells and periprosthetic infection; cartilage and impingement surgery; acetabulum in combination; cementless ceramic prosthesis; metal-on-metal hips; and whether size matters in failure.
Balancing service provision and surgical training is a challenging issue that affects all healthcare systems. A multicentre prospective study of 1501 total hip replacements was undertaken to investigate whether there is an association between surgical outcome and the grade of the operating surgeon, and whether there is any difference in outcome if surgeons’ assistants assist with the operation, rather than orthopaedic trainees. The primary outcome measure was the change in the Oxford hip score (OHS) at five years. Secondary outcomes included the rate of revision and dislocation, operating time, and length of hospital stay. There was no significant difference in ΔOHS or complication rates between operations undertaken by trainers and trainees, or those at which surgeons’ assistants and trainees were the assistant. However, there was a significant difference in the duration of surgery, with a mean reduction of 28 minutes in those in which a surgeons’ assistant was the assistant. This study provides evidence that total hip replacements can be performed safely and effectively by appropriately trained surgeons in training, and that there are potential benefits of using surgeons’ assistants in orthopaedic surgery.
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
A prospective, multi-centre study was carried out on 1421 total hip replacements between January 1999 and July 2007 to examine if obesity has an effect on clinical outcomes. The patients were categorised into three groups: non-obese (body mass index (BMI) <
30 kg/m2), obese (BMI 30 to 40 kg/m2) and morbidly obese (BMI >
40 kg/m2). The primary outcome measure was the change in Oxford hip score at five years. Secondary outcome measures included dislocation and revision rates, increased haemorrhage, deep infection, deep-vein thrombosis and pulmonary embolism, mean operating time and length of hospital stay. Radiological analysis assessing heterotopic ossification, femoral osteolysis and femoral stem positioning was performed. Data were incomplete for 362 hips (25.5%) There was no difference in the change in the Oxford hip score, complication rates or radiological changes at five years between the groups. The morbidly obese group was significantly younger and required a significantly longer operating time. Obese and morbidly obese patients have as much to gain from total hip replacement as non-obese patients.
We present a series of 35 patients (19 men and
16 women) with a mean age of 64 years (36.7 to 75.9), who underwent
total hip replacement using the ESKA dual-modular short stem with
metal on-polyethylene bearing surfaces. This implant has a modular
neck section in addition to the modular head. Of these patients,
three presented with increasing post-operative pain due to pseudotumour
formation that resulted from corrosion at the modular neck-stem
junction. These patients underwent further surgery and aseptic lymphocytic
vaculitis associated lesions were demonstrated on histological analysis. Retrieval analysis of two modular necks showed corrosion at the
neck-stem taper. Blood cobalt and chromium levels were measured
at a mean of nine months (3 to 28) following surgery. These were
compared with the levels in seven control patients (three men and
four women) with a mean age of 53.4 years (32.1 to 64.1), who had
an identical prosthesis and articulation but with a prosthesis that
had no modularity at neck-stem junction. The mean blood levels of
cobalt in the study group were raised at 50.75 nmol/l (5 to 145)
compared with 5.6 nmol/l (2 to 13) in control patients. Corrosion at neck-stem tapers has been identified as an important
source of metal ion release and pseudotumour formation requiring
revision surgery. Finite element modelling of the dual modular stem
demonstrated high stresses at the modular stem-neck junction. Dual
modular cobalt-chrome hip prostheses should be used with caution
due to these concerns.
Clinical, haematological or economic benefits of post-operative blood salvage with autologous blood re-transfusion have yet to be clearly demonstrated for primary total hip replacement. We performed a prospective randomised study to analyse differences in postoperative haemoglobin levels and homologous blood requirements in two groups of patients undergoing primary total hip replacement. A series of 158 patients was studied. In one group two vacuum drains were used and in the other the ABTrans autologous retransfusion system. A total of 58 patients (76%) in the re-transfusion group received autologous blood. There was no significant difference in the mean post-operative haemoglobin levels in the two groups. There were, however, significantly fewer patients with post-operative haemoglobin values less than 9.0 g/dl and significantly fewer patients who required transfusion of homologous blood in the re-transfusion group. There was also a small overall cost saving in this group.
An international faculty of orthopaedic surgeons
presented their work on the current challenges in hip surgery at
the London Hip Meeting which was attended by over
400 delegates. The topics covered included femoroacetabular impingement, thromboembolic
phenomena associated with hip surgery, bearing surfaces (including metal-on-metal
articulations), outcomes of hip replacement surgery and revision
hip replacement. We present a concise report of the current opinions
on hip surgery from this meeting with appropriate references to
the current literature.
The Morscher-Spotorno (MS-30) femoral stem is a stainless-steel, straight, three-dimensionally tapered, collarless implant for cemented fixation in total hip replacement. We report the results at ten years of a consecutive series of 124 total hip replacements in 121 patients with the matt-surfaced MS-30 stem and an alumina ceramic head of 28-mm diameter. All the stems were fixed with Palacos bone cement with gentamicin using a modern cementing technique. They were combined with an uncemented, press-fit cup. The mean period of observation was 10.2 years (8.3 to 12.1) and no patient was lost to follow-up. Twenty-seven patients (22%) died with the implant None of the stems or cups had been revised. The Harris hip score was excellent or good in 97% (85 hips) and moderate in 3% (three hips). Radiologically, six hips (6.8%) had osteolysis adjacent to the stem, mostly in Gruen zone 7. Twenty (22.7%) showed one or more radiolucent lines. Twenty-two stems (25%) had subsided by 2 mm to 5 mm. In these cases two showed osteolysis (9.1%) with subsidence and four without (6.1%). Radiolucent lines were seen in seven with migration (31.8%) and in 13 without (19.7%). No infections and no acetabular osteolysis were observed. The clinical results were excellent with survivorship after ten years of 100% and only a slightly statistically non-significant higher rate of osteolysis and radiolucency in cases of subsidence.
Using data from the Norwegian Hip Fracture Register,
8639 cemented and 2477 uncemented primary hemiarthroplasties for
displaced fractures of the femoral neck in patients aged > 70 years
were included in a prospective observational study. A total of 218
re-operations were performed after cemented and 128 after uncemented
procedures. Survival of the hemiarthroplasties was calculated using
the Kaplan-Meier method and hazard rate ratios (HRR) for revision
were calculated using Cox regression analyses. At five years the
implant survival was 97% (95% confidence interval (CI) 97 to 97)
for cemented and 91% (95% CI 87 to 94) for uncemented hemiarthroplasties.
Uncemented hemiarthroplasties had a 2.1 times increased risk of
revision compared with cemented prostheses (95% confidence interval
1.7 to 2.6, p < 0.001). The increased risk was mainly caused
by revisions for peri-prosthetic fracture (HRR = 17), aseptic loosening
(HRR = 17), haematoma formation (HRR = 5.3), superficial infection
(HRR = 4.6) and dislocation (HRR = 1.8). More intra-operative complications,
including intra-operative death, were reported for the cemented
hemiarthroplasties. However, in a time-dependent analysis, the HRR
for re-operation in both groups increased as follow-up increased. This study showed that the risk for revision was higher for uncemented
than for cemented hemiarthroplasties.
Prospective data on hip fracture from 3686 patients at a United Kingdom teaching hospital were analysed to investigate the risk factors, financial costs and outcomes associated with deep or superficial wound infections after hip fracture surgery. In 1.2% (41) of patients a deep wound infection developed, and 1.1% (39) had a superficial wound infection. A total of 57 of 80 infections (71.3%) were due to No statistically significant pre-operative risk factors were detected. Length of stay, cost of treatment and pre-discharge mortality all significantly increased with deep wound infection. The one-year mortality was 30%, and this increased to 50% in those who developed an infection (p <
0.001). A deep infection resulted in doubled operative costs, tripled investigation costs and quadrupled ward costs. MRSA infection increased costs, length of stay, and pre-discharge mortality compared with non-MRSA infection.
There remains a lack of data on the reliability of methods to
estimate tibial coverage achieved during total knee replacement.
In order to address this gap, the intra- and interobserver reliability
of a three-dimensional (3D) digital templating method was assessed
with one symmetric and one asymmetric prosthesis design. A total of 120 template procedures were performed according to
specific rotational and over-hang criteria by three observers at
time zero and again two weeks later. Total and sub-region coverage
were calculated and the reliability of the templating and measurement
method was evaluated.Objectives
Methods
Impacted morsellised allografts have been used successfully to address the problem of poor bone stock in revision surgery. However, there are concerns about the transmission of pathogens, the high cost and the shortage of supply of donor bone. Bone-graft extenders, such as tricalcium phosphate (TCP) and hydroxyapatite (HA), have been developed to minimise the use of donor bone. In a human cadaver model we have evaluated the surgical and mechanical feasibility of a TCP/HA bone-graft extender during impaction grafting revision surgery. A TCP/HA allograft mix increased the risk of producing a fissure in the femur during the impaction procedure, but provided a higher initial mechanical stability when compared with bone graft alone. The implications of the use of this type of graft extender in impaction grafting revision surgery are discussed.
We report the long-term results of revision total
hip replacement using femoral impaction allografting with both uncemented
and cemented Freeman femoral components. A standard design of component
was used in both groups, with additional proximal hydroxyapatite
coating in the uncemented group. A total of 33 hips in 30 patients received
an uncemented component and 31 hips in 30 patients a cemented component.
The mean follow-up was 9.8 years (2 to 17) in the uncemented group
and 6.2 years (1 to 11) in the cemented group. Revision procedures
(for all causes) were required in four patients (four hips) in the
uncemented group and in five patients (five hips) in the cemented
group. Harris hip scores improved significantly in both groups and
were maintained independently of the extent of any migration of
the femoral component within the graft or graft–cement mantle.
The aim of this study was to investigate the
possible benefit of large-head metal-on-metal bearing on a stem
for primary hip replacement compared with a 28 mm diameter conventional
metal-on-polyethylene bearing in a prospective randomised controlled
trial. We investigated cemented stem behaviour between these two
different bearings using Einzel-Bild-Röntgen-Analyse, clinical and
patient reported measures (Harris hip score, Western Ontario and
McMaster Universities osteoarthritis index, Short Form-36 and satisfaction)
and whole blood metal ion levels at two years. A power study indicated
that 50 hips were needed in each group to detect subsidence of >
5 mm at two years with a
p-value of <
0.05. Significant improvement (p <
0.001) was found in the mean
clinical and patient reported outcomes at two years for both groups.
Comparison of outcomes between the groups at two years showed no
statistically significant difference for mean stem migration, clinical
and patient reported outcomes; except overall patient satisfaction which
was higher for metal-on-metal group (p = 0.05). Metal ion levels
were raised above the Medicines and Healthcare products Regulatory
Agency advised safety level (7 µg per litre) in 20% of the metal-on-metal
group and in one patient in metal-on-polyethylene group (who had
a metal-on-metal implant on the contralateral side). Two patients
in the metal-on-metal group were revised, one for pseudotumour and
one for peri-prosthetic fracture. Use of large modular heads is associated with a risk of raised
whole blood metal ion levels despite using a proven bearing from
resurfacing. The head-neck junction or excess stem micromotion are
possibly the weak links warranting further research.