In this paper, we will consider the current role
of metal-on-metal bearings by looking at three subtypes of MoM hip
arthroplasty separately: Hip resurfacing, large head (>
36 mm) MoM
THA and MoM THA with traditional femoral head sizes.
Patient specific cutting guides generated by
preoperative Magnetic Resonance Imaging (MRI) of the patient’s extremity
have been proposed as a method of improving the consistency of Total
Knee Arthroplasty (TKA) alignment and adding efficiency to the operative
procedure. The cost of this option was evaluated by quantifying the
savings from decreased operative time and instrument processing
costs compared to the additional cost of the MRI and the guide.
Coronal plane alignment was measured in an unselected consecutive
series of 200 TKAs, 100 with standard instrumentation and 100 with
custom cutting guides. While the cutting guides had significantly lower
total operative time and instrument processing time, the estimated
$322 savings was overwhelmed by the $1,500 additional cost of the
MRI and the cutting guide. All measures of coronal plane alignment
were equivalent between the two groups. The data does not currently
support the proposition that patient specific guides add value to
TKA.
Most fractures of the radial head are stable
undisplaced or minimally displaced partial fractures without an associated
fracture of the elbow or forearm or ligament injury, where stiffness
following non-operative management is the primary concern. Displaced
unstable fractures of the radial head are usually associated with other
fractures or ligament injuries, and restoration of radiocapitellar
contact by reconstruction or prosthetic replacement of the fractured
head is necessary to prevent subluxation or dislocation of the elbow
and forearm. In fractures with three or fewer fragments (two articular
fragments and the neck) and little or no metaphyseal comminution,
open reduction and internal fixation may give good results. However,
fragmented unstable fractures of the radial head are prone to early
failure of fixation and nonunion when fixed. Excision of the radial
head is associated with good long-term results, but in patients
with instability of the elbow or forearm, prosthetic replacement
is preferred. This review considers the characteristics of stable and unstable
fractures of the radial head, as well as discussing the debatable
aspects of management, in light of the current best evidence. Cite this article:
A moderator and panel of five experts led an
interactive session in discussing five challenging and interesting patient
case presentations involving surgery of the hip. The hip pathologies
reviewed included failed open reduction internal fixation of subcapital
femoral neck fracture, bilateral hip disease, evaluation of pain
after metal-on-metal hip arthroplasty, avascular necrosis, aseptic
loosening secondary to osteolysis and polyethylene wear, and management
of ceramic femoral head fracture.
We analysed whether a high body mass index (BMI)
had a deleterious effect on outcome following autologous chondrocyte
implantation (ACI) or matrix-carried autologous chondrocyte implantation
(MACI) for the treatment of full-thickness chondral defects of the
knee from a subset of patients enrolled in the ACI vs MACI trial
at The Royal National Orthopaedic Hospital. The mean Modified Cincinnati scores (MCS) were significantly
higher (p <
0.001) post-operatively in patients who had an ideal
body weight (n = 53; 20 to 24.9 kg/m2) than in overweight
(n = 63; 25 to 30 kg/m2) and obese patients (n = 22;
>
30 kg/m2). At a follow-up of two years, obese patients
demonstrated no sustained improvement in the MCS. Patients with
an ideal weight experienced significant improvements as early as
six months after surgery (p = 0.007). In total, 82% of patients
(31 of 38) in the ideal group had a good or excellent result, compared
with 49% (22 of 45) of the overweight and 5.5% (one of 18) in the
obese group (p <
0.001). There was a significant negative relationship between
BMI and the MCS 24 months after surgery (r = -0.4, p = 0.001). This study demonstrates that obese patients have worse knee function
before surgery and experience no sustained benefit from ACI or MACI
at two years after surgery. There was a correlation between increasing
BMI and a lower MCS according to a linear regression analysis. On
the basis of our findings patient selection can be more appropriately
targeted.
The success of total knee replacement (TKR) depends
on optimal soft-tissue balancing, among many other factors. The
objective of this study is to correlate post-operative anteroposterior
(AP) translation of a posterior cruciate ligament-retaining TKR
with clinical outcome at two years. In total 100 patients were divided
into three groups based on their AP translation as measured by the
KT-1000 arthrometer. Group 1 patients had AP translation <
5
mm, Group 2 had AP translation from 5 mm to 10 mm, and Group 3 had
AP translation >
10 mm. Outcome assessment included range of movement
of the knee, the presence of flexion contractures, hyperextension,
knee mechanical axes and functional outcome using the Knee Society
score, Oxford knee score and the Short-Form 36 questionnaire. At two years, patients in Group 2 reported significantly better
Oxford knee scores than the other groups (p = 0.045). A positive
correlation between range of movement and AP translation was noted,
with patients in group 3 having the greatest range of movement (mean
flexion: 117.9° (106° to 130°)) (p <
0.001). However, significantly
more patients in Group 3 developed hyperextension >
10° (p = 0.01). In this study, the best outcome for cruciate-ligament retaining
TKR was achieved in patients with an AP translation of 5 mm to 10
mm.
We systematically reviewed the published literature
on the complications of closing wedge high tibial osteotomy for
the treatment of unicompartmental osteoarthritis of the knee. Publications
were identified using the Cochrane Library, MEDLINE, EMBASE and
CINAHL databases up to February 2012. We assessed randomised (RCTs), controlled
group clinical (CCTs) trials, case series in publications associated
with closing wedge osteotomy of the tibia in patients with osteoarthritis
of the knee and finally a Cochrane review. Many of these trials
included comparative studies (opening wedge
Radiostereometric analysis (RSA) can detect early
micromovement in unstable implant designs which are likely subsequently
to have a high failure rate. In 2010, the Articular Surface Replacement
(ASR) was withdrawn because of a high failure rate. In 19 ASR femoral
components, the mean micromovement over the first two years after implantation
was 0.107 mm ( We conclude that the ASR femoral component achieves initial stability
and that early migration is not the mode of failure for this resurfacing
arthroplasty.
The purpose of this study was to measure the
radiological parameters of femoral component alignment of the Oxford
Phase 3 unicompartmental knee replacement (UKR), and evaluate their
effect on clinical outcome. Multiple regression analysis was used
to examine the relative contributions of the radiological assessment
of femoral component alignment in 189 consecutive UKRs performed
by a single surgeon. The American Knee Society scores were compared
between groups, defined as being within or outside recommended tolerances
of the position of the femoral component. For the flexion/extension
position 21 UKRs (11.1%) lay outside the recommended limits, and for
posterior overhang of the femoral component nine (4.8%) lay outside
the range. The pre-operative hip/knee/ankle (HKA) angle, narrowest
canal distance from the distal femoral entry point of the alignment
jig and coronal entry-point position had significant effects on
the flexion/extension position. Pre-operative HKA angle had a significant
influence on posterior overhang of the femoral component. However,
there was no significant difference in American Knee Society scores
relative to the position of the femoral component.
The inherent challenges of total hip replacement
(THR) in children include the choice of implant for the often atypical
anatomical morphology, its fixation to an immature growing skeleton
and the bearing surface employed to achieve a successful long-term
result. We report the medium-term results of 52 consecutive uncemented
THRs undertaken in
35 paediatric patients with juvenile idiopathic arthritis. The mean
age at the time of surgery was 14.4 years (10 to 16). The median
follow-up was 10.5 years (6 to 15). During the study period 13 THRs
underwent revision surgery. With revision as an endpoint, subgroup
analysis revealed 100% survival of the 23 ceramic-on-ceramic THRs
and 55% (16 of 29) of the metal- or ceramic-on-polyethylene. This
resulted in 94% (95% CI 77.8 to 98.4) survivorship of the femoral
component and 62% (95% CI 41.0 to 78.0) of the acetabular component.
Revision of the acetabular component for wear and osteolysis were
the most common reasons for failure accounting for 11 of the 13
revisions. The success seen in patients with a ceramic-on-ceramic articulation
seems to indicate that this implant strategy has the potential to
make a major difference to the long-term outcome in this difficult
group of patients.
We report the effectiveness of revision of total
elbow replacement by re-cementing. Between 1982 and 2004, 53 elbows
in 52 patients were treated with re-cementing of a total elbow replacement
into part or all of the existing cement mantle or into the debrided
host-bone interface, without the use of structural bone augmentation
or a custom prosthesis. The original implant revision was still
We systematically reviewed all the evidence published
in the English language on proximal interphalangeal joint (PIPJ)
replacement, to determine its effectiveness on the function of the
hand and the associated post-operative complications. Original studies were selected if they reported clinical outcome
with a minimum of one year’s follow-up. Quality was assessed using
the Cowley systematic review criteria modified for finger-joint
replacements. Of 319 articles identified, only five were adequately
reported according to our quality criteria; there were no randomised
controlled trials. PIPJ replacements had a substantial effect size
on hand pain of -23.2 (95% confidence interval (CI) -27.3 to -19.1)
and grip strength 1.2 (95% CI -10.7 to 13.1), and a small effect
on range of movement 0.2 (95% CI -0.4 to 0.8). A dorsal approach
was most successful. Post-operative loosening occurred in 10% (95%
CI 3 to 30) of ceramic and 12.5% (95% CI 7 to 21) of pyrocarbon
replacements. Post-operative complications occurred in 27.8% (95%
CI 20 to 37). We conclude that the effectiveness of PIPJ replacement has not
been established. Small observational case studies and short-term
follow-up, together with insufficient reporting of patient data,
functional outcomes and complications, limit the value of current
evidence. We recommend that a defined core set of patients, surgical and
outcome data for this intervention be routinely and systematically
collected within the framework of a joint registry.
Plasma levels of cobalt and chromium ions and
Metal Artefact Reduction Sequence (MARS)-MRI scans were performed
on patients with 209 consecutive, unilateral, symptomatic metal-on-metal
(MoM) hip arthroplasties. There was wide variation in plasma cobalt
and chromium levels, and MARS-MRI scans were positive for adverse reaction
to metal debris (ARMD) in 84 hips (40%). There was a significant
difference in the median plasma cobalt and chromium levels between
those with positive and negative MARS-MRI scans (p <
0.001).
Compared with MARS-MRI as the potential reference standard for the
diagnosis of ARMD, the sensitivity of metal ion analysis for cobalt
or chromium with a cut-off of >
7 µg/l was 57%. The specificity was
65%, positive predictive value was 52% and the negative predictive
value was 69% in symptomatic patients. A lowered threshold of >
3.5 µg/l for cobalt and chromium ion levels improved the sensitivity
and negative predictive value to 86% and 74% but at the expense
of specificity (27%) and positive predictive value (44%). Metal ion analysis is not recommended as a sole indirect screening
test in the surveillance of symptomatic patients with a MoM arthroplasty.
The investigating clinicians should have a low threshold for obtaining
cross-sectional imaging in these patients, even in the presence
of low plasma metal ion levels.
Fracture of a ceramic component in total hip
replacement is a rare but potentially catastrophic complication.
The incidence is likely to increase as the use of ceramics becomes
more widespread. We describe such a case, which illustrates how
inadequate initial management will lead to further morbidity and
require additional surgery. We present the case as a warning that
fracture of a ceramic component should be revised to another ceramic-on-ceramic
articulation in order to minimise the risk of further catastrophic
wear.
We review the history and literature of hip resurfacing arthroplasty. Resurfacing and the science behind it continues to evolve. Recent results, particularly from the national arthroplasty registers, have spread disquiet among both surgeons and patients. A hip resurfacing arthroplasty is not a total hip replacement, but should perhaps be seen as a means of delaying it. The time when hip resurfacing is offered to a patient may be different from that for a total hip replacement. The same logic can apply to the timing of revision surgery. Consequently, the comparison of resurfacing with total hip replacement may be a false one. Nevertheless, the need for innovative solutions for young arthroplasty patients is clear. Total hip replacement can be usefully delayed in many of these patients by the use of hip resurfacing arthroplasty.
We The triple-strand technique significantly increased the tensile
strength (p = 0.0001) and gap resistance (p = 0.01) of bovine tendon
repairs, and might have advantages in human application for accelerated
post-operative rehabilitation.
To quantify and compare peri-acetabular bone mineral density
(BMD) between a monoblock acetabular component using a metal-on-metal
(MoM) bearing and a modular titanium shell with a polyethylene (PE)
insert. The secondary outcome was to measure patient-reported clinical
function. A total of 50 patients (25 per group) were randomised to MoM
or metal-on-polyethlene (MoP). There were 27 women (11 MoM) and
23 men (14 MoM) with a mean age of 61.6 years (47.7 to 73.2). Measurements
of peri-prosthetic acetabular and contralateral hip (covariate)
BMD were performed at baseline and at one and two years’ follow-up.
The Western Ontario and McMaster Universities osteoarthritis index
(WOMAC), University of California, Los Angeles (UCLA) activity score,
Harris hip score, and RAND-36 were also completed at these intervals.Objectives
Methods
Femoroacetabular impingement (FAI) causes pain
and chondrolabral damage via mechanical overload during movement
of the hip. It is caused by many different types of pathoanatomy,
including the cam ‘bump’, decreased head–neck offset, acetabular
retroversion, global acetabular overcoverage, prominent anterior–inferior
iliac spine, slipped capital femoral epiphysis, and the sequelae
of childhood Perthes’ disease. Both evolutionary and developmental factors may cause FAI. Prevalence
studies show that anatomic variations that cause FAI are common
in the asymptomatic population. Young athletes may be predisposed
to FAI because of the stress on the physis during development. Other
factors, including the soft tissues, may also influence symptoms and
chondrolabral damage. FAI and the resultant chondrolabral pathology are often treated
arthroscopically. Although the results are favourable, morphologies
can be complex, patient expectations are high and the surgery is
challenging. The long-term outcomes of hip arthroscopy are still
forthcoming and it is unknown if treatment of FAI will prevent arthrosis.
We sought to establish the incidence of joint failure secondary to adverse reaction to metal debris (ARMD) following metal-on-metal hip resurfacing in a large, three surgeon, multicentre study involving 4226 hips with a follow-up of 10 to 142 months. Three implants were studied: the Articular Surface Replacement; the Birmingham Hip Resurfacing; and the Conserve Plus. Retrieved implants underwent analysis using a co-ordinate measuring machine to determine volumetric wear. There were 58 failures associated with ARMD. The median chromium and cobalt concentrations in the failed group were significantly higher than in the control group (p <
0.001). Survival analysis showed a failure rate in the patients with Articular Surface Replacement of 9.8% at five years, compared with <
1% at five years for the Conserve Plus and 1.5% at ten years for the Birmingham Hip Resurfacing. Two ARMD patients had relatively low wear of the retrieved components. Increased wear from the metal-on-metal bearing surface was associated with an increased rate of failure secondary to ARMD. However, the extent of tissue destruction at revision surgery did not appear to be dose-related to the volumetric wear.
We present the clinical results and survivorship
of consecutive 100 Birmingham Hip Resurfacings in 90 patients at
a minimum follow-up of ten years. All procedures were carried out
by an independent surgeon who commenced a prospective study in 1998.
Patients were assessed clinically using the Western Ontario and
McMaster Universities osteoarthritis index, Short-Form 36, Harris
hip score and University of California, Los Angeles activity score. Radiological
analysis was performed by an independent observer and blood metal
ion levels concentrations were measured at ten years post-operatively
in 62 patients. The median acetabular component inclination was
46.2° (34° to 59°) and anteversion 11.0° (0° to 30°). The median
chromium concentration in the unilateral group was 1.74 (0.41 to
15.23) and for the bilateral group was 2.98 (1.57 to 18.01). The
equivalent values for cobalt were 1.67 (0.54 to 20.4) and 1.88 (1.38
to 19.32). In total there were eight failures giving an overall survival
at ten years of 92% (95% confidence interval (CI) 86.7 to 97.3).
BHR in male patients had an improved survivorship of 94.6% (95%
CI 89.4 to 100) compared with females at 84.6% (95% CI 70.7 to 98.5),
but this did not reach statistical significance (p = 0.119). Four
of the nine BHRs with a bearing diameter of 42 mm failed. The overall results were consistent with data produced from other
centres in that the clinical outcome of large male patients was
extremely encouraging, whereas the survival of the smaller joints
was less satisfactory.