The issues surrounding raised levels of metal
ions in the blood following large head metal-on-metal total hip replacement
(THR), such as cobalt and chromium, have been well documented. Despite
the national popularity of uncemented metal-on-polyethylene (MoP)
THR using a large-diameter femoral head, few papers have reported
the levels of metal ions in the blood following this combination.
Following an isolated failure of a 44 mm Trident–Accolade uncemented
THR associated with severe wear between the femoral head and the
trunnion in the presence of markedly elevated levels of cobalt ions
in the blood, we investigated the relationship between modular femoral head
diameter and the levels of cobalt and chromium ions in the blood
following this THR. A total of 69 patients received an uncemented Trident–Accolade
MoP THR in 2009. Of these, 43 patients (23 men and 20 women, mean
age 67.0 years) were recruited and had levels of cobalt and chromium
ions in the blood measured between May and June 2012. The patients
were then divided into three groups according to the diameter of
the femoral head used: 12 patients in the 28 mm group (controls),
18 patients in the 36 mm group and 13 patients in the 40 mm group.
A total of four patients had identical bilateral prostheses in situ
at phlebotomy: one each in the 28 mm and 36 mm groups and two in
the 40 mm group. There was a significant increase in the mean levels of cobalt
ions in the blood in those with a 36 mm diameter femoral head compared
with those with a 28 mm diameter head (p = 0.013). The levels of
cobalt ions in the blood were raised in those with a 40 mm diameter
head but there was no statistically significant difference between
this group and the control group (p = 0.152). The levels of chromium
ions in the blood were normal in all patients. The clinical significance of this finding is unclear, but we
have stopped using femoral heads with a diameter of ≤ 36 mm, and
await further larger studies to clarify whether, for instance, this
issue particularly affects this combination of components. Cite this article:
Many different lengths of stem are available
for use in primary total hip replacement, and the morphology of
the proximal femur varies greatly. The more recently developed shortened
stems provide a distribution of stress which closely mimics that
of the native femur. Shortening the femoral component potentially
comes at the cost of decreased initial stability. Clinical studies
on the performance of shortened cemented and cementless stems are promising,
although long-term follow-up studies are lacking. We provide an
overview of the current literature on the anatomical features of
the proximal femur and the biomechanical aspects and clinical outcomes
associated with the length of the femoral component in primary hip
replacement, and suggest a classification system for the length
of femoral stems. Cite this article:
We performed a case–control study to compare
the rates of further surgery, revision and complications, operating time
and survival in patients who were treated with either an uncemented
hydroxyapatite-coated Corail bipolar femoral stem or a cemented
Exeter stem for a displaced intracapsular fracture of the hip. The
mean age of the patients in the uncemented group was 82.5 years
(53 to 97) and in the cemented group was 82.7 years (51 to 99) We used
propensity score matching, adjusting for age, gender and the presence
or absence of dementia and comorbidities, to produce a matched cohort
receiving an Exeter stem (n = 69) with which to compare the outcome of
patients receiving a Corail stem (n = 69). The Corail had a significantly
lower all-cause rate of further surgery (p = 0.016; odds ratio (OR)
0.18, 95% CI 0.04 to 0.84) and number of hips undergoing major further
surgery (p = 0.029; OR 0.13, 95% CI 0.01 to 1.09). The mean operating
time was significantly less for the Corail group than for the cemented Exeter
group (59 min [12 to 136] Cite this article:
Based on the first implementation of mixing antibiotics
into bone cement in the 1970s, the Endo-Klinik has used one stage
exchange for prosthetic joint infection (PJI) in over 85% of cases.
Looking carefully at current literature and guidelines for PJI treatment,
there is no clear evidence that a two stage procedure has a higher
success rate than a one-stage approach. A cemented one-stage exchange
potentially offers certain advantages, mainly based on the need
for only one operative procedure, reduced antibiotics and hospitalisation time.
In order to fulfill a one-stage approach, there are obligatory pre-,
peri- and post-operative details that need to be meticulously respected,
and are described in detail. Essential pre-operative diagnostic
testing is based on the joint aspiration with an exact identification
of any bacteria. The presence of a positive bacterial culture and
respective antibiogram are essential, to specify the antibiotics
to be loaded to the bone cement, which allows a high local antibiotic
elution directly at the surgical side. A specific antibiotic treatment
plan is generated by a microbiologist. The surgical success relies
on the complete removal of all pre-existing hardware, including
cement and restrictors and an aggressive and complete debridement
of any infected soft tissues and bone material. Post-operative systemic
antibiotic administration is usually completed after only ten to
14 days. Cite this article:
We determined the short-term clinical outcome
and migration within the bone of the humeral cementless component
of the Instrumented Bone Preserving (IBP) total elbow replacement
in a series of 16 patients. There were four men and 12 women with
a mean age at operation of 63 years (40 to 81). Migration was calculated
using radiostereometric analysis. There were no intra-operative
complications and no revisions. At two-year follow-up, all patients
showed a significant reduction in pain and functional improvement
of the elbow (both p <
0.001). Although ten components (63%)
showed movement or micromovement during the first six weeks, 14
(88%) were stable at one year post-operatively. Translation was
primarily found in the proximal direction (median 0.3 mm (interquartile range
(IQR) -0.09 to 0.8); the major rotational movement was an anterior
tilt (median 0.7° (IQR 0.4° to 1.6°)). One malaligned component
continued to migrate during the second year, and one component could
not be followed beyond three months because migration had caused
the markers to break off the prosthesis. This study shows promising early results for the cementless humeral
component of the IBP total elbow replacement. All patients had a
good clinical outcome, and most components stabilised within six
months of the operation. Cite this article:
In a time of limited resources, the debate continues
over which types of hip prosthesis are clinically superior and more
cost-effective. Orthopaedic surgeons increasingly need robust economic
evidence to understand the full value of the operation, and to aid
decision making on the ‘package’ of procedures that are available
and to justify their practice beyond traditional clinical preference. In this paper we explore the current economic debate about the
merits of cemented and cementless total hip replacement, an issue
that continues to divide the orthopaedic community. Cite this article:
We investigated the functional outcome in patients
who underwent reverse shoulder replacement (RSR) after removal of
a tumour of the proximal humerus. A total of 16 patients (ten women
and six men) underwent this procedure between 1998 and 2011 in our
hospital. Five patients died and one was lost to follow-up. Ten
patients were available for review at a mean follow-up of 46 months
(12 to 136). Eight patients had a primary and two patients a secondary
bone tumour. At final follow up the mean range of active movement was: abduction
78° (30° to 150°); flexion 98° (45° to 180°); external rotation
32° (10° to 60°); internal rotation 51° (10° to 80°). The mean Musculoskeletal
Tumor Society score was 77% (60% to 90%) and the mean Toronto Extremity
Salvage Score was 70% (30% to 91%). Two patients had a superficial
infection and one had a deep infection and underwent a two-stage
revision procedure. In two patients there was loosening of the RSR;
one dislocated twice. All patients had some degree of atrophy or
pseudo-atrophy of the deltoid muscle. Use of a RSR in patients with a tumour of the proximal humerus
gives acceptable results. Cite this article:
Between January 1990 and December 2000 we carried out 226 SB Charité III disc replacements for lumbar disc degeneration in 160 patients. They were reviewed at a mean follow-up of 79 months (31 to 161) to determine the clinical and radiological outcome. The clinical results were collected by an independent observer, who was not involved in patient selection, treatment or follow-up, using a combination of outcome measures, including the Oswestry Disability Index. Pain was recorded using a visual analogue score, and the most recent radiographs were reviewed. Survival of the device was analysed by the Kaplan-Meier method and showed a cumulative survival of 35% at 156 months when radiological failure was taken as the endpoint. The mean improvement in the Oswestry disability index scores after disc replacement was 14% (6% to 21%) and the mean improvement in the pain score was 1.6 (0.46 to 2.73), both falling below the clinically significant threshold. Removal of the implant was required in 12 patients, four because of implant failure. These poor results indicate that further use of this implant is not justified.
We used an A significant improvement in functional weight-bearing was observed between six and 12 weeks. The significant increase in ground reaction force through the operated limb between six and 12 weeks was greater than that reported previously with morcellised graft augmented reconstructions. Histological appearance and collagen fibre orientation with bone block augmentation more closely resembled that of an intact enthesis compared with the morcellised grafting technique. Bone block augmentation of tendon-implant interfaces results in more reliable functional and histological outcomes, with a return to pre-operative levels of weight-bearing by 24 weeks.
The ideal acetabular component is characterised by reliable, long-term fixation with physiological loading of bone and a low rate of wear. Trabecular metal is a porous construct of tantalum which promotes bony ingrowth, has a modulus of elasticity similar to that of cancellous bone, and should be an excellent material for fixation. Between 2004 and 2006, 55 patients were randomised to receive either a cemented polyethylene or a monobloc trabecular metal acetabular component with a polyethylene articular surface. We measured the peri-prosthetic bone density around the acetabular components for up to two years using dual-energy x-ray absorptiometry. We found evidence that the cemented acetabular component loaded the acetabular bone centromedially whereas the trabecular metal monobloc loaded the lateral rim and behaved like a hemispherical rigid metal component with regard to loading of the acetabular bone. We suspect that this was due to the peripheral titanium rim used for the mechanism of insertion.
The August 2013 Hip &
Pelvis Roundup360 looks at: are we getting it right first time?; tantalum augments in revision hip surgery; lower wear in dual mobility?; changing faces changes outcomes; synovial fluid aspiration in MOM hips; taper disease: the new epidemic of hip surgery; the super-obese and THR; and whether well fixed stems can remain in infected hips
We evaluated the long-term fixation of 64 press-fit cemented stems of constrained total knee prostheses in 32 young patients with primary malignant bone tumours. Initial stable fixation, especially in rotation, was achieved by precise fit of the stem into the reamed endosteum, before cementation. Complementary fixation, especially in migration and rotation, was obtained by pressurised antibiotic-loaded cement. The mean age at operation was 33 years (13 to 61). No patient was lost to follow-up; 13 patients died and the 19 survivors were examined at a mean follow-up of 12.5 years (4 to 21). Standard revision press-fit cemented stems were used on the side of the joint which was not involved with tumour (26 tibial and six femoral), on this side there was no loosening or osteolysis and stem survival was 100%. On the reconstruction side, custom-made press-fit stems were used and the survival rate, with any cause for revision as an end point, was 88%, but 97% for loosening or osteolysis. This longevity is similar to that achieved at 20 years with the Charnley-Kerboull primary total hip replacement with press-fit cemented femoral components. We recommend this type of fixation when extensive reconstruction of the knee is required. It may also be suitable for older patients requiring revision of a total knee replacement or in difficult situations such as severe deformity and complex articular fractures.
Between November 1997 and December 2000 we performed 27 total hip replacements in 22 patients with high congenital dislocation of the hip using porous tantalum monoblock acetabular components implanted in the true acetabular bed. Clinical and radiological evaluation was performed at regular intervals for a mean of 10.2 years (8.5 to 12). The mean Harris Hip Score improved from 48.3 (15 to 65) pre-operatively to 89.5 (56 to 100) at the final follow-up. The mean Oxford Hip Score was 49.5 (35 to 59) pre-operatively and decreased to 21.2 (12 to 48) at one year and 15.2 (10 to 28) at final follow-up. Migration of the acetabular component was assessed with the EBRA software system. There was a mean migration of 0.68 mm (0.49 to 0.8) in the first year and a mean 0.89 mm (0.6 to 0.98) in the second year, after excluding one initial excessive migration. No revision was necessary for any reason, no acetabular component became loose, and no radiolucent lines were observed at the final follow-up. The porous tantalum monoblock acetabular component is an implant offering adequate initial stability in conjunction with a modulus of elasticity and porosity close to that of cancellous bone. It favours bone ingrowth, leading to good mid-term results.
Despite advances in the prevention and treatment of osteoporotic fractures, their prevalence continues to increase. Their operative treatment remains a challenge for the surgeon, often with unpredictable outcomes. This review highlights the current aspects of management of these fractures and focuses on advances in implant design and surgical technique.
We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component. All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting migration comprising subsidence and lift-off from the tibial tray, most of the trabecular metal components showed subsidence only, probably due to the elasticity of the implant. This pattern of subsidence is regarded as being beneficial for uncemented fixation.
Impacted bone allograft is often used in revision joint replacement. Hydroxyapatite granules have been suggested as a substitute or to enhance morcellised bone allograft. We hypothesised that adding osteogenic protein-1 to a composite of bone allograft and non-resorbable hydroxyapatite granules (ProOsteon) would improve the incorporation of bone and implant fixation. We also compared the response to using ProOsteon alone against bone allograft used in isolation. We implanted two non-weight-bearing hydroxyapatite-coated implants into each proximal humerus of six dogs, with each implant surrounded by a concentric 3 mm gap. These gaps were randomly allocated to four different procedures in each dog: 1) bone allograft used on its own; 2) ProOsteon used on its own; 3) allograft and ProOsteon used together; or 4) allograft and ProOsteon with the addition of osteogenic protein-1. After three weeks osteogenic protein-1 increased bone formation and the energy absorption of implants grafted with allograft and ProOsteon. A composite of allograft, ProOsteon and osteogenic protein-1 was comparable, but not superior to, allograft used on its own. ProOsteon alone cannot be recommended as a substitute for allograft around non-cemented implants, but should be used to extend the volume of the graft, preferably with the addition of a growth factor.
We report the long-term survival of a prospective randomised consecutive series of 501 primary knee replacements using the press-fit condylar posterior cruciate ligament-retaining prosthesis. Patients received either cemented (219 patients, 277 implants) or cementless (177 patients, 224 implants) fixation. Altogether, 44 of 501 knees (8.8%) underwent revision surgery (24 cemented This single-surgeon series, with no loss to follow-up, provides reliable data of the revision rates of one of the most commonly-used total knee replacements. The survival of the press-fit condylar total knee replacement remained good at 15 years, irrespective of the method of fixation.
We evaluated the clinical results and complications
after extra-articular resection of the distal femur and/or proximal
tibia and reconstruction with a tumour endoprosthesis (MUTARS) in
59 patients (mean age 33 years (11 to 74)) with malignant bone or
soft-tissue tumours. According to a Kaplan–Meier analysis, limb
survival was 76% (95% confidence interval (CI) 64.1 to 88.5) after
a mean follow-up of 4.7 years (one month to 17 years). Peri-prosthetic infection
was the most common indication for subsequent amputation (eight
patients). Survival of the prosthesis without revision was 48% (95%
CI 34.8 to 62.0) at two years and 25% (95% CI 11.1 to 39.9) at five years
post-operatively. Failure of the prosthesis was due to deep infection
in 22 patients (37%), aseptic loosening in ten patients (17%), and
peri-prosthetic fracture in six patients (10%). Wear of the bearings
made a minor revision necessary in 12 patients (20%). The mean Musculoskeletal
Tumor Society score was 23 (10 to 29). An extensor lag >
10° was
noted in ten patients (17%). These results suggest that limb salvage after extra-articular
resection with a tumour prosthesis can achieve good functional results
in most patients, although the rates of complications and subsequent
amputation are higher than in patients treated with intra-articular
resection. Cite this article:
To date, all surgical techniques used for reconstruction
of the pelvic ring following supra-acetabular tumour resection produce
high complication rates. We evaluated the clinical, oncological
and functional outcomes of a cohort of 35 patients (15 men and 20
women), including 21 Ewing’s sarcomas, six chondrosarcomas, three sarcomas
not otherwise specified, one osteosarcoma, two osseous malignant
fibrous histiocytomas, one synovial cell sarcoma and one metastasis.
The mean age of the patients was 31 years (8 to 79) and the latest
follow-up was carried out at a mean of 46 months (1.9 to 139.5)
post-operatively. We undertook a functional reconstruction of the pelvic ring using
polyaxial screws and titanium rods. In 31 patients (89%) the construct
was encased in antibiotic-impregnated polymethylmethacrylate. Preservation
of the extremities was possible for all patients. The survival rate
at three years was 93.9% (95% confidence interval (CI) 77.9 to 98.4),
at five years it was 82.4% (95% CI 57.6 to 93.4). For the 21 patients
with Ewing’s sarcoma it was 95.2% (95% CI 70.7 to 99.3) and 81.5%
(95% CI 52.0 to 93.8), respectively. Wound healing problems were
observed in eight patients, deep infection in five and clinically
asymptomatic breakage of the screws in six. The five-year implant survival
was 93.3% (95% CI 57.8 to 95.7). Patients were mobilised at a mean
of 3.5 weeks (1 to 7) post-operatively. A post-operative neurological
defect occurred in 12 patients. The mean Musculoskeletal Tumor Society
score at last available follow-up was 21.2 (10 to 27). This reconstruction technique is characterised by simple and
oncologically appropriate applicability, achieving high primary
stability that allows early mobilisation, good functional results
and relatively low complication rates. Cite this article:
We report the clinical and radiological outcomes
of a series of contemporary cementless ceramic-on-ceramic total hip
replacements (THRs) at ten years in patients aged ≤ 55 years of
age. Pre- and post-operative activity levels are described. A total
of 120 consecutive ceramic cementless THRs were performed at a single
centre in 110 patients from 1997 to 1999. The mean age of the patients
at operation was 45 years (20 to 55). At ten years, four patients
had died and six were lost to follow-up, comprising ten hips. The
mean post-operative Harris hip score was 94.7 (55 to 100). Radiological
analysis was undertaken in 90 available THRs of the surviving 106
hips at final review: all had evidence of stable bony ingrowth,
with no cases of osteolysis. Wear was undetectable. There were four
revisions. The survival for both components with revision for any
cause as an endpoint was 96.5% (95% confidence interval 94.5 to
98.7). The mean modified University of California, Los Angeles activity
level rose from a mean of 6.4 (4 to 10) pre-operatively to 9.0 (6
to 10) at the ten-year post-operative period. Alumina ceramic-on-ceramic bearings in cementless primary THR
in this series have resulted in good clinical and radiological outcomes
with undetectable rates of wear and excellent function in the demanding
younger patient group at ten years. Cite this article: