Reverse total shoulder replacement (RTSR) depends
on adequate deltoid function for a successful outcome. However, the
anterior deltoid and/or axillary nerve may be damaged due to prior
procedures or injury. The purpose of this study was to determine
the compensatory muscle forces required for scapular plane elevation
following RTSR when the anterior deltoid is deficient. The soft
tissues were removed from six cadaver shoulders, except for tendon
attachments. After implantation of the RTSR, the shoulders were
mounted on a custom-made shoulder simulator to determine the mean
force in each muscle required to achieve 30° and 60° of scapular
plane elevation. Two conditions were tested: 1) Control with an
absent supraspinatus and infraspinatus; and 2) Control with anterior
deltoid deficiency. Anterior deltoid deficiency resulted in a mean
increase of 195% in subscapularis force at 30° when compared with
the control (p = 0.02). At 60°, the subscapularis force increased
a mean of 82% (p <
0.001) and the middle deltoid force increased
a mean of 26% (p = 0.04). Scapular plane elevation may still be possible following an RTSR
in the setting of anterior deltoid deficiency. When the anterior
deltoid is deficient, there is a compensatory increase in the force
required by the subscapularis and middle deltoid. Attempts to preserve
the subscapularis, if present, might maximise post-operative function.
The February 2013 Foot &
Ankle Roundup360 looks at: replacement in osteonecrosis of the talus; ankle instability in athletes; long-term follow-up of lateral ankle ligament reconstruction; an operation for Stage II TPD; whether you should operate on Achilles tendon ruptures; Weil osteotomies and Freiberg’s disease; MRI scanning not sensitive for intra-articular lesions; and single-stage debridement and reconstruction in Charcot feet.
Endoprosthetic replacement of the proximal femur may be required to treat primary bone tumours or destructive metastases either with impending or established pathological fracture. Modular prostheses are available off the shelf and can be adapted to most reconstructive situations for this purpose. We have assessed the clinical and functional outcome of using the METS (Stanmore Implants Worldwide) modular tumour prosthesis to reconstruct the proximal femur in 100 consecutive patients between 2001 and 2006. We compared the results with the published series for patients managed with modular and custom-made endoprosthetic replacements for the same conditions. There were 52 males and 48 females with a mean age of 56.3 years (16 to 84) and a mean follow-up of 24.6 months (0 to 60). In 65 patients the procedure was undertaken for metastases, in 25 for a primary bone tumour, and in ten for other malignant conditions. A total of 46 patients presented with a pathological fracture, and 19 presented with failed fixation of a previous pathological fracture. The overall patient survival was 63.6% at one year and 23.1% at five years, and was significantly better for patients with a primary bone tumour than for those with metastatic tumour (82.3% vs 53.3%, respectively at one year (p = 0.003)). There were six early dislocations of which five could be treated by closed reduction. No patient needed revision surgery for dislocation. Revision surgery was required by six (6%) patients, five for pain caused by acetabular wear and one for tumour progression. Amputation was needed in four patients for local recurrence or infection. The estimated five-year implant survival with revision as the endpoint was 90.7%. The mean Toronto Extremity Salvage score was 61% (51% to 95%). The implant survival and complications resulting from the use of the modular system were comparable to the published series of both custom-made and other modular proximal femoral implants. We conclude that at intermediate follow-up the modular tumour prosthesis for proximal femur replacement provides versatility, a low incidence of implant-related complications and acceptable function for patients with metastatic tumours, pathological fractures and failed fixation of the proximal femur. It also functions as well as a custom-made endoprosthetic replacement.
The December 2012 Shoulder &
Elbow Roundup360 looks at: whether allograft is biomechanically superior in large Hill-Sachs defects; glenoid bone loss in shoulder dislocators; repairing irreparable cuff tears; acromioclavicular joint injuries; whether more radiographs equals more surgery; whether reverse TSR is cheaper than hemiarthroplasty; autologous chondrocyte implantation in the shoulder; and fracture of the clavicle.
The treatment of bony defects of the tibia at the time of revision total knee replacement is controversial. The place of compacted morsellised bone graft is becoming established, particularly in contained defects. It has previously been shown that the initial stability of impaction-grafted trays in the contained defects is equivalent to that of an uncemented primary knee replacement. However, there is little biomechanical evidence on which to base a decision in the treatment of uncontained defects. We undertook a laboratory-based biomechanical study comparing three methods of graft containment in segmental medial tibial defects and compared them with the use of a modular metal augment to bypass the defect. Using resin models of the proximal tibia with medial defects representing either 46% or 65% of the medial cortical rim, repair of the defect was accomplished using mesh, cement or a novel bag technique, after which impaction bone grafting was used to fill the contained defects and a tibial component was cemented in place. As a control, a cemented tibial component with modular metal augments was used in identical defects. All specimens were submitted to cyclical mechanical loading, during which cyclical and permanent tray displacement were determined. The results showed satisfactory stability with all the techniques except the bone bag method. Using metal augments gave the highest initial stability, but obviously lacked any potential for bone restoration.
This study reports our experience with total elbow replacement for fused elbows. Between 1982 and 2004, 13 patients with spontaneously ankylosed elbows were treated with a linked semi-constrained non-custom total elbow implant. The mean age at operation was 54 years (24 to 80). The stiffness was a result of trauma in ten elbows, juvenile rheumatoid arthritis in one, and rheumatoid arthritis in two. The patients were followed for a mean of 12 years (2 to 26) and were evaluated clinically using the Mayo Elbow Performance Score, as well as radiologically. A mean arc from 37° of extension to 118° of flexion was achieved. Outcomes were good or excellent for seven elbows at final review. Ten patients felt better or much better after total elbow replacement. However, there was a high complication rate and re-operation was required in over half of patients. Two developed peri-operative soft-tissue breakdown requiring debridement. A muscle flap with skin grafting was used for soft-tissue cover in one. Revision was undertaken in one elbow following fracture of the ulnar component. Three patients developed a deep infection. Three elbows were manipulated under anaesthesia for post-operative stiffness. Prophylactic measures for heterotopic ossification were unsuccessful. Total elbow replacement for the ankylosed elbow should be performed with caution. However, the outcome can be reliable in the long term and have a markedly positive impact on patient function and satisfaction. The high potential for complications must be considered. We consider total elbow replacement to be an acceptable procedure in selected patients with reasonable expectations.
Restoration of leg length and offset is an important
goal in total hip replacement. This paper reports a calliper-based technique
to help achieve these goals by restoring the location of the centre
of the femoral head. This was validated first by using a co-ordinate
measuring machine to see how closely the calliper technique could
record and restore the centre of the femoral head when simulating
hip replacement on Sawbone femur, and secondly by using CT in patients
undergoing hip replacement. Results from the co-ordinate measuring machine showed that the
centre of the femoral head was predicted by the calliper to within
4.3 mm for offset (mean 1.6 (95% confidence interval (CI) 0.4 to
2.8)) and 2.4 mm for vertical height (mean -0.6 (95% CI -1.4 to
0.2)).
The CT scans showed that offset and vertical height were restored
to within 8 mm
(mean -1 (95% CI -2.1 to 0.6)) and -14 mm (mean 4 (95% CI 1.8 to
4.3)), respectively. Accurate assessment and restoration of the centre of the femoral
head is feasible with a calliper. It is quick, inexpensive, simple
to use and can be applied to any design of femoral component.
Advanced MRI cartilage imaging such as T1-rho
(T1ρ) for the diagnosis of early cartilage degradation prior to morpholgic
radiological changes may provide prognostic information in the management
of joint disease. This study aimed first to determine the normal
T1ρ profile of cartilage within the hip, and secondly to identify
any differences in T1ρ profile between the normal and symptomatic
femoroacetabular impingement (FAI) hip. Ten patients with cam-type
FAI (seven male and three female, mean age 35.9 years (28 to 48))
and ten control patients (four male and six female, mean age 30.6
years (22 to 35)) underwent 1.5T T1ρ MRI of a single hip. Mean T1ρ relaxation
times for full thickness and each of the three equal cartilage thickness
layers were calculated and compared between the groups. The mean
T1ρ relaxation times for full cartilage thickness of control and
FAI hips were similar (37.17 ms ( These results suggest that 1.5T T1ρ MRI can detect acetabular
hyaline cartilage changes in patients with FAI.
Numerous complications following total knee replacement (TKR)
relate to the patellofemoral (PF) joint, including pain and patellar
maltracking, yet the options for A total of three knees with end-stage osteoarthritis and three
knees that had undergone TKR at more than one year’s follow-up were
investigated. In each knee, sequential biplane radiological images
were acquired from the sagittal direction (i.e. horizontal X-ray
source and 10° below horizontal) for a sequence of eight flexion
angles. Three-dimensional implant or bone models were matched to
the biplane images to compute the six degrees of freedom of PF tracking
and TF kinematics, and other clinical measures.Objectives
Methods
Revision arthroplasty of the hip is expensive
owing to the increased cost of pre-operative investigations, surgical implants
and instrumentation, protracted hospital stay and drugs. We compared
the costs of performing this surgery for aseptic loosening, dislocation,
deep infection and peri-prosthetic fracture. Clinical, demographic
and economic data were obtained for 305 consecutive revision total
hip replacements in 286 patients performed at a tertiary referral
centre between 1999 and 2008. The mean total costs for revision
surgery in aseptic cases (n = 194) were £11 897 (
The use of two implants to manage concomitant ipsilateral femoral
shaft and proximal femoral fractures has been indicated, but no
studies address the relationship of dynamic hip screw (DHS) side
plate screws and the intramedullary nail where failure might occur
after union. This study compares different implant configurations
in order to investigate bridging the gap between the distal DHS
and tip of the intramedullary nail. A total of 29 left synthetic femora were tested in three groups:
1) gapped short nail (GSN); 2) unicortical short nail (USN), differing
from GSN by the use of two unicortical bridging screws; and 3) bicortical
long nail (BLN), with two angled bicortical and one unicortical bridging
screws. With these findings, five matched-pairs of cadaveric femora
were tested in two groups: 1) unicortical long nail (ULN), with
a longer nail than USN and three bridging unicortical screws; and
2) BLN. Specimens were axially loaded to 22.7 kg (50 lb), and internally
rotated 90°/sec until failure.Objectives
Methods
We reviewed our initial seven-year experience
with a non-invasive extendible prosthesis in 34 children with primary bone
tumours. The distal femur was replaced in 25 cases, total femur
in five, proximal femur in one and proximal tibia in three. The
mean follow-up was 44 months (15 to 86) and 27 patients (79%) remain
alive. The prostheses were lengthened by an electromagnetic induction
mechanism in an outpatient setting and a mean extension of 32 mm
(4 to 80) was achieved without anaesthesia. There were lengthening
complications in two children: failed lengthening in one and the
formation of scar tissue in the other. Deep infection developed
in six patients (18%) and local recurrence in three. A total of
11 patients required further surgery to the leg. Amputation was
necessary in five patients (20%) and a two-stage revision in another.
There were no cases of loosening, but two patients had implant breakage
and required revision. The mean Musculoskeletal Tumor Society functional
score was 85% (60% to 100%) at last known follow-up. These early
results demonstrate that the non-invasive extendible prosthesis
allows successful lengthening without surgical intervention, but
the high incidence of infection is a cause for concern.
Our aim was to evaluate the results of triple arthrodesis, performed without the use of supplementary bone graft. We carried out a retrospective review of 100 consecutive triple arthrodeses. All the operations had been performed by the senior author (TSS) using a standard technique. Only local bone graft from the excised joint surfaces had been used, thereby avoiding complications at the donor site. The mean age of the patients at surgery was 58 years (18 to 84). The mean time to union was 5.1 months (3 to 17). There were 75 good, 20 fair and five poor results. There were four cases of nonunion. Our study has shown that comparable rates of union are achieved without the need for supplementary bone graft from the iliac crest or other donor site.
Critical size defects in ovine tibiae, stabilised with intramedullary interlocking nails, were used to assess whether the addition of carboxymethylcellulose to the standard osteogenic protein-1 (OP-1/BMP-7) implant would affect the implant’s efficacy for bone regeneration. The biomaterial carriers were a ‘putty’ carrier of carboxymethylcellulose and bovine-derived type-I collagen (OPP) or the standard with collagen alone (OPC). These two treatments were also compared to “ungrafted” negative controls. Efficacy of regeneration was determined using radiological, biomechanical and histological evaluations after four months of healing. The defects, filled with OPP and OPC, demonstrated radiodense material spanning the defect after one month of healing, with radiographic evidence of recorticalisation and remodelling by two months. The OPP and OPC treatment groups had equivalent structural and material properties that were significantly greater than those in the ungrafted controls. The structural properties of the OPP- and OPC-treated limbs were equivalent to those of the contralateral untreated limb (p >
0.05), yet material properties were inferior (p <
0.05). Histopathology revealed no residual inflammatory response to the biomaterial carriers or OP-1. The OPP- and OPC-treated animals had 60% to 85% lamellar bone within the defect, and less than 25% of the regenerate was composed of fibrous tissue. The defects in the untreated control animals contained less than 40% lamellar bone and more than 60% was fibrous tissue, creating full cortical thickness defects. In our studies carboxymethylcellulose did not adversely affect the capacity of the standard OP-1 implant for regenerating bone.
The aim of this study was to investigate the
effect of training on the arthroscopic performance of a group of
medical students and to determine whether all students could be
trained to competence. Thirty-three medical students with no previous
experience of arthroscopy were randomised to a ‘Trained’ or an ‘Untrained’
cohort. They were required to carry out 30 episodes of two simulated
arthroscopic tasks (one shoulder and one knee). The primary outcome variable
was task success at each episode. Individuals achieved competence
when their learning curve stabilised. The secondary outcome was
technical dexterity, assessed objectively using a validated motion
analysis system. Six subjects in the ‘Untrained’ cohort failed to
achieve competence in the shoulder task, compared with one in the ‘Trained’
cohort. During the knee task, two subjects in each cohort failed
to achieve competence. Based on the objective motion analysis parameters,
the ‘Trained’ cohort performed better on the shoulder task (p <
0.05) but there was no significant difference for the knee task
(p >
0.05). Although
We undertook this study to determine the minimum
amount of coronoid necessary to stabilise an otherwise intact elbow
joint. Regan–Morrey types II and III, plus medial and lateral oblique
coronoid fractures, collectively termed type IV fractures, were
simulated in nine fresh cadavers. An electromagnetic tracking system
defined the three-dimensional stability of the ulna relative to
the humerus. The coronoid surface area accounts for 59% of the anterior articulation.
Alteration in valgus, internal and external rotation occurred only
with a type III coronoid fracture, accounting for 68% of the coronoid
and 40% of the entire articular surface. A type II fracture removed
42% of the coronoid articulation and 25% of the entire articular
surface but was associated with valgus and external rotational changes
only when the radial head was removed, thereby removing 67% of the
articular surface. We conclude that all type III fractures, as defined here, are
unstable, even with intact ligaments and a radial head. However,
a type II deficiency is stable unless the radial head is removed.
Our study suggests that isolated medial-oblique or lateral-oblique
fractures, and even a type II fracture with intact ligaments and
a functional radial head, can be clinically stable, which is consistent
with clinical observation.
Filling the empty holes in peri-articular locking
plates may improve the fatigue strength of the fixation. The purpose of
this A locking/compression plate was applied to 33 synthetic femurs
and then a 6 cm metaphyseal defect was created (AO Type 33-A3).
The specimens were then divided into three groups: unplugged, plugged
with locking screw only and fully plugged holes. They were then
tested using a stepwise or run-out fatigue protocol, each applying
cyclic physiological multiaxial loads. All specimens in the stepwise group failed at the 770 N load
level. The mean number of cycles to failure for the stepwise specimen
was 25 500 cycles ( In conclusion, filling the empty combination locking/compression
holes in peri-articular distal femur locking plates at the level
of supracondylar comminution does not increase the fatigue life
of the fixation in a comminuted supracondylar femoral fracture model
(AO 33-A3) with a 6 cm gap.
An ongoing prospective study to investigate failing metal-on-metal
hip prostheses was commenced at our centre in 2008. We report on
the results of the analysis of the first consecutive 126 failed
mated total hip prostheses from a single manufacturer. Analysis was carried out using highly accurate coordinate measuring
to calculate volumetric and linear rates of the articular bearing
surfaces and also the surfaces of the taper junctions. The relationship
between taper wear rates and a number of variables, including bearing
diameter and orientation of the acetabular component, was investigated.Objectives
Methods