Infection remains a significant and common complication after joint replacement and there is debate about which contributing factors are important. Few studies have investigated the effect of the operating time on infection. We collected data prospectively from 5277 hip and knee replacements which included the type of procedure, the operating time, the use of drains, the operating theatre, surgeon, age and gender. In a subgroup of 3449 knee replacements further analysis was carried out using the tourniquet time in place of the operating time. These variables were assessed by the use of generalised linear modelling against superficial, deep or joint-space post-operative infection as defined by the Australian Surgical-Site Infection criteria. The overall infection rate was 0.98%. In the replacement data set both male gender (z = 3.097, p = 0.00195) and prolonged operating time (z = 4.325, p <
0.001) were predictive of infection. In the knee subgroup male gender (z = 2.250, p = 0.02447), a longer tourniquet time (z = 2.867, p = 0.00414) and total knee replacement ( These findings support the view that a prolonged operating time and male gender are associated with an increased incidence of infection. Steps to minimise intra-operative delay should be instigated, and care should be exercised when introducing measures which prolong the duration of joint replacement.
The strain on clinic and surgeon resources resulting
from a rise in demand for total knee replacement (TKR) requires reconsideration
of when and how often patients need to be seen for follow-up. Surgeons
will otherwise require increased paramedical staff or need to limit
the number of TKRs they undertake. We reviewed the outcome data
of 16 414 primary TKRs undertaken at our centre to determine the
time to re-operation for any reason and for specific failure mechanisms.
Peak risk years for failure were determined by comparing the conditional
probability of failure, the number of failures divided by the total
number of TKRs cases, for each year. The median times to failure
for the most common failure mechanisms were 4.9 years (interquartile
range (IQR) 1.7 to 10.7) for femoral and tibial loosening, 1.9 years
(IQR 0.8 to 3.9) for infection, 3.1 years (IQR 1.6 to 5.5) for tibial
collapse and 5.6 years (IQR 3.4 to 9.3) for instability. The median
time to failure for all revisions was 3.3 years (IQR 1.2 to 8.5),
with an overall revision rate of 1.7% (n = 282). Results from our
patient population suggest that patients be seen for follow-up at
six months, one year, three years, eight years, 12 years, and every
five years thereafter. Patients with higher pain in the early post-operative
period or high body mass index (≥ 41 kg/m2) should be
monitored more closely. Cite this article:
The June 2013 Knee Roundup360 looks at: knee arthroplasty in diabetic patients; whether TKR is a timebomb; the use of antidepressants for knee OA; trochleoplasty; articulated spacers; mental health and joint replacement; and the use of physiotherapy for meniscal tear.
We describe 119 meniscal allograft transplantations performed concurrently with articular cartilage repair in 115 patients with severe articular cartilage damage. In all, 53 (46.1%) of the patients were over the age of 50 at the time of surgery. The mean follow-up was for 5.8 years (2 months to 12.3 years), with 25 procedures (20.1%) failing at a mean of 4.6 years (2 months to 10.4 years). Of these, 18 progressed to knee replacement at a mean of 5.1 years (1.3 to 10.4). The Kaplan-Meier estimated mean survival time for the whole series was 9.9 years ( The survival of the transplant was not affected by gender, the severity of cartilage damage, axial alignment, the degree of narrowing of the joint space or medial
The February 2013 Children’s orthopaedics Roundup360 looks at: the human genome; new RNA; cells, matrix and gene enhancement; the histology of x-rays; THR and VTE in the Danish population; potential therapeutic targets for GCT; optimising vancomycin elution from cement; and how much sleep is enough.
As many as 25% to 40% of unicompartmental knee
replacement (UKR) revisions are performed for pain, a possible cause
of which is proximal tibial strain. The aim of this study was to
examine the effect of UKR implant design and material on cortical
and cancellous proximal tibial strain in a synthetic bone model.
Composite Sawbone tibiae were implanted with cemented UKR components
of different designs, either all-polyethylene or metal-backed. The tibiae
were subsequently loaded in 500 N increments to 2500 N, unloading
between increments. Cortical surface strain was measured using a
digital image correlation technique. Cancellous damage was measured
using acoustic emission, an engineering technique that detects sonic
waves (‘hits’) produced when damage occurs in material. Anteromedial cortical surface strain showed significant differences
between implants at 1500 N and 2500 N in the proximal 10 mm only
(p <
0.001), with relative strain shielding in metal-backed implants.
Acoustic emission showed significant differences in cancellous bone
damage between implants at all loads (p = 0.001). All-polyethylene implants
displayed 16.6 times the total number of cumulative acoustic emission
hits as controls. All-polyethylene implants also displayed more
hits than controls at all loads (p <
0.001), more than metal-backed
implants at loads ≥ 1500 N (p <
0.001), and greater acoustic
emission activity on unloading than controls (p = 0.01), reflecting
a lack of implant stiffness. All-polyethylene implants were associated
with a significant increase in damage at the microscopic level compared
with metal-backed implants, even at low loads. All-polyethylene
implants should be used with caution in patients who are likely
to impose large loads across their knee joint. Cite this article:
The aim of this study was to examine the loading
of the other joints of the lower limb in patients with unilateral osteoarthritis
(OA) of the knee. We recruited 20 patients with no other symptoms
or deformity in the lower limbs from a consecutive cohort of patients
awaiting knee replacement. Gait analysis and electromyographic recordings were
performed to determine moments at both knees and hips, and contraction
patterns in the medial and lateral quadriceps and hamstrings bilaterally.
The speed of gait was reduced in the group with OA compared with
the controls, but there were only minor differences in stance times
between the limbs. Patients with OA of the knee had significant
increases in adduction moment impulse at both knees and the contralateral
hip (adjusted p-values: affected knee: p <
0.01, unaffected knee
p = 0.048, contralateral hip p = 0.03), and significantly increased
muscular co-contraction bilaterally compared with controls (all
comparisons for co-contraction, p <
0.01). The other major weight-bearing joints are at risk from abnormal
biomechanics in patients with unilateral OA of the knee. Cite this article:
We prospectively examined the physical and imaging findings, including MRI, of 23 patients with spontaneous osteonecrosis of the knee after obtaining informed consent to acquire tissue specimens at surgery. There were four men and 19 women, with a mean age of 67.5 years (58 to 77). Plain radiographs were designated as stages 1, 2, 3 or 4 according to the classification of Koshino. Five knees were classified as stage 1, five as stage 2, seven as stage 3 and six as stage 4. The histological specimens were stained with haematoxylin and eosin and tetrachrome. In the early stages of the condition, a subchondral fracture was noted in the absence of any features of osteonecrosis, whereas in advanced stages, osteonecrotic lesions were confined to the area distal to the site of the fracture which showed impaired healing. In such cases, formation of cartilage and fibrous tissue, occurred indicating delayed or nonunion. These findings strongly suggest that the histopathology at each stage of spontaneous osteonecrosis is characterised by different types of repair reaction for subchondral fractures.
The Oxford mobile-bearing unicompartmental knee
replacement (UKR) is an effective and safe treatment for osteoarthritis
of the medial compartment. The results in the lateral compartment
have been disappointing due to a high early rate of dislocation
of the bearing. A series using a newly designed domed tibial component
is reported. The first 50 consecutive domed lateral Oxford UKRs in 50 patients
with a mean follow-up of three years (2.0 to 4.3) were included.
Clinical scores were obtained prospectively and Kaplan-Meier survival
analysis was performed for different endpoints. Radiological variables
related to the position and alignment of the components were measured. One patient died and none was lost to follow-up. The cumulative
incidence of dislocation was 6.2% (95% confidence interval (CI)
2.0 to 17.9) at three years. Survival using revision for any reason
and aseptic revision was 94% (95% CI 82 to 98) and 96% (95% CI 85
to 99) at three years, respectively. Outcome scores, visual analogue
scale for pain and maximum knee flexion showed a significant improvement
(p <
0.001). The mean Oxford knee score was 43 ( Clinical results are excellent and short-term survival has improved
when compared with earlier series. The risk of dislocation remains
higher using a mobile-bearing UKR in the lateral compartment when
compared with the medial compartment. Patients should be informed
about this complication. To avoid dislocations, care must be taken
not to elevate the lateral joint line.
The rotational alignment of the tibia is an unresolved issue in knee replacement. A poor functional outcome may be due to malrotation of the tibial component. Our aim was to find a reliable method for positioning the tibial component in knee replacement. CT scans of 19 knees were reconstructed in three dimensions and orientated vertically. An axial plane was identified 20 mm below the tibial spines. The centre of each tibial condyle was calculated from ten points taken round the condylar cortex. The tibial tubercle centre was also generated as the centre of the circle which best fitted eight points on the outside of the tubercle in an axial plane at the level of its most prominent point. The derived points were identified by three observers with errors of 0.6 mm to 1 mm. The medial and lateral tibial centres were constant features (radius 24 mm ( Alignment of the knee when based on this anatomical axis was more reliable than either the posterior surfaces or any axis involving the tubercle which was the least reliable landmark in the region.
The Cementless Oxford Unicompartmental Knee Replacement
(OUKR) was developed to address problems related to cementation,
and has been demonstrated in a randomised study to have similar
clinical outcomes with fewer radiolucencies than observed with the
cemented device. However, before its widespread use it is necessary
to clarify contraindications and assess the complications. This
requires a larger study than any previously published. We present a prospective multicentre series of 1000 cementless
OUKRs in 881 patients at a minimum follow-up of one year. All patients
had radiological assessment aligned to the bone–implant interfaces
and clinical scores. Analysis was performed at a mean of 38.2 months
(19 to 88) following surgery. A total of 17 patients died (comprising
19 knees (1.9%)), none as a result of surgery; there were no tibial
or femoral loosenings. A total of 19 knees (1.9%) had significant
implant-related complications or required revision. Implant survival
at six years was 97.2%, and there was a partial radiolucency at
the bone–implant interface in 72 knees (8.9%), with no complete radiolucencies.
There was no significant increase in complication rate compared
with cemented fixation (p = 0.87), and no specific contraindications
to cementless fixation were identified. Cementless OUKR appears to be safe and reproducible in patients
with end-stage anteromedial osteoarthritis of the knee, with radiological
evidence of improved fixation compared with previous reports using
cemented fixation. Cite this article:
Current analysis of unicondylar knee replacements
(UKRs) by national registries is based on the pooled results of medial
and lateral implants. Consequently, little is known about the differential
performance of medial and lateral replacements and the influence
of each implant type within these pooled analyses. Using data from
the National Joint Registry for England and Wales (NJR) we aimed
to determine the proportion of UKRs implanted on the lateral side
of the knee, and their survival and reason for failure compared
with medial UKRs. By combining information on the side of operation
with component details held on the NJR, we were able to determine
implant laterality (medial
Using general practitioner records and hospital
notes and through direct telephone conversation with patients, we investigated
the accuracy of nine patient-reported complications gathered from
a self-completed questionnaire after elective joint replacement
surgery of the hip and knee. A total of 402 post-discharge complications
were reported after 8546 elective operations that were undertaken
within a three-year period. These were reported by 136 men and 240
women with a mean age of 71.8 years (34 to 93). A total of 319 reported
complications (79.4%; 95% confidence interval 75.4 to 83.3) were
confirmed to be correct. High rates of correct reporting were demonstrated
for infection (94.5%) and the need for further surgery (100%), whereas
the rates of reporting deep-vein thrombosis (DVT), pulmonary embolism,
myocardial infarction and stroke were lower (75% to 84.2%). Dislocation,
peri-prosthetic fractures and nerve palsy had modest rates of correct
reporting (36% to 57.1%). More patients who had knee surgery delivered
incorrect reports of dislocation (p = 0.001) and DVT (p = 0.013). Despite these variations, it appears that post-operative complications
may form part of a larger patient-reported outcome programme after
elective joint replacement surgery.
We carried out a prospective investigation into
the radiological outcomes of uncemented Oxford medial compartment
unicondylar replacement in 220 consecutive patients (231 knees)
performed in a single centre with a minimum two-year follow-up.
The functional outcomes using the mean Oxford knee score and the
mean high-activity arthroplasty score were significantly improved
over the pre-operative scores (p <
0.001). There were 196 patients
with a two-year radiological examination performed under fluoroscopic
guidance, aiming to provide images acceptable for analysis of the
bone–implant interface. Of the six tibial zones examined on each
knee on the anteroposterior radiograph, only three had a partial
radiolucent line. All were in the medial aspect of the tibial base plate
(zone 1) and all measured <
1 mm. All of these patients were
asymptomatic. There were no radiolucent lines seen around the femoral
component or on the lateral view. There was one revision for loosening
at one year due to initial inadequate seating of the tibial component.
These results confirm that the early uncemented Oxford medial unicompartmental
compartmental knee replacements were reliable and the incidence
of radiolucent lines was significantly decreased compared with the
reported results of cemented versions of this implant. These independent
results confirm those of the designing centre.
This editorial considers the shortcomings of assessing outcome after joint replacement only by the survival of the implant.
Robots have been used in surgery since the late
1980s. Orthopaedic surgery began to incorporate robotic technology
in 1992, with the introduction of ROBODOC, for the planning and
performance of total hip replacement. The use of robotic systems
has subsequently increased, with promising short-term radiological
outcomes when compared with traditional orthopaedic procedures.
Robotic systems can be classified into two categories: autonomous
and haptic (or surgeon-guided). Passive surgery systems, which represent
a third type of technology, have also been adopted recently by orthopaedic
surgeons. While autonomous systems have fallen out of favour, tactile systems
with technological improvements have become widely used. Specifically,
the use of tactile and passive robotic systems in unicompartmental
knee replacement (UKR) has addressed some of the historical mechanisms
of failure of non-robotic UKR. These systems assist with increasing
the accuracy of the alignment of the components and produce more
consistent ligament balance. Short-term improvements in clinical
and radiological outcomes have increased the popularity of robot-assisted
UKR. Robot-assisted orthopaedic surgery has the potential for improving
surgical outcomes. We discuss the different types of robotic systems
available for use in orthopaedics and consider the indication, contraindications
and limitations of these technologies.
Given the growing prevalence of obesity around
the world and its association with osteoarthritis of the knee, orthopaedic
surgeons need to be familiar with the management of the obese patient
with degenerative knee pain. The precise mechanism by which obesity
leads to osteoarthritis remains unknown, but is likely to be due
to a combination of mechanical, humoral and genetic factors. Weight loss has clear medical benefits for the obese patient
and seems to be a logical way of relieving joint pain associated
with degenerative arthritis. There are a variety of ways in which
this may be done including diet and exercise, and treatment with
drugs and bariatric surgery. Whether substantial weight loss can
delay or even reverse the symptoms associated with osteoarthritis
remains to be seen. Surgery for osteoarthritis in the obese patient can be technically
more challenging and carries a risk of additional complications.
Substantial weight loss before undertaking total knee replacement
is advisable. More prospective studies that evaluate the effect
of significant weight loss on the evolution of symptomatic osteoarthritis
of the knee are needed so that orthopaedic surgeons can treat this
patient group appropriately.
The Oxford unicompartmental knee replacement gives good results in patients with symptomatic osteoarthritis of the medial compartment. Previous studies have suggested that the presence of radiolucent lines (RLLs) does not reflect a poor outcome in such patients. However, the reliability and validity of this assessment have not been determined. Our aim was to assess the intra- and interobserver reliability and the sensitivity and specificity of the assessment of RLLs around both tibial and femoral components using standard radiographs. Two reviewers assessed the radiographs of 45 patients who had loosening of the tibial or femoral component confirmed at revision surgery and compared them with those of a series of 45 asymptomatic patients matched for age and gender. The results suggested that, using standard radiographs, tibial RLLs were 63.6% sensitive and 94.4% specific and femoral RLLs 63.9% sensitive and 72.7% specific for loosening. Overall intra- and interobserver reliability was highly variable, but zonal analysis showed that lucency at the tip of the femoral peg was significantly associated with loosening of the femoral component. Fluoroscopically guided radiographs may improve the zonal reliability of the assessment of RLLs, but further independent and comparative studies are required. In the meantime, the innocence of the physiological RLLs detected by standard radiographs should be viewed with caution.
In a systematic review, reports from national registers and clinical studies were identified and analysed with respect to revision rates after joint replacement, which were calculated as revisions per 100 observed component years. After primary hip replacement, a mean of 1.29 revisions per 100 observed component years was seen. The results after primary total knee replacement are 1.26 revisions per 100 observed component years, and 1.53 after medial unicompartmental replacement. After total ankle replacement a mean of 3.29 revisions per 100 observed component years was seen. The outcomes of total hip and knee replacement are almost identical. Revision rates of about 6% after five years and 12% after ten years are to be expected.