The August 2015 Shoulder &
Elbow Roundup360 looks at: Clavicular fractures are being fixed – but how?;
We have evaluated the in vivo migration patterns of 164 primary consecutive Charnley-Kerboull total hip replacements which were undertaken in 155 patients. The femoral preparation included removal of diaphyseal cancellous bone to obtain primary rotational stability of the stem before line-to-line cementing. We used the Ein Bild Roentgen Analyse femoral component method to assess the subsidence of the femoral component. At a mean of 17.3 years (15.1 to 18.3) 73 patients were still alive and had not been revised, eight had been revised, 66 had died and eight had been lost to follow-up. The mean subsidence of the entire series was 0.63 mm (0.0 to 1.94). When using a 1.5 mm threshold, only four stems were considered to have subsided. Our study showed that, in most cases, a highly polished double-tapered stem cemented line-to-line does not subside at least up to 18 years after implantation.
We describe 129 consecutive revision total hip replacements using a Charnley-Kerboull femoral component of standard length with impaction allografting. The mean follow-up was 8.2 years (2 to 16). Additionally, extramedullary reinforcement was performed using struts of cortical allograft in 49 hips and cerclage wires in 30. There was one intra-operative fracture of the femur but none later. Two femoral components subsided by 5 mm and 8 mm respectively, and were considered to be radiological failures. No further revision of a femoral component was required. The rate of survival of the femoral component at nine years, using radiological failure as the endpoint, was 98%. Our study showed that impaction grafting in association with a Charnley-Kerboull femoral component has a low rate of subsidence. Reconstruction of deficiencies of distal bone with struts of cortical allograft appeared to be an efficient way of preventing postoperative femoral fracture for up to 16 years.
A total of 219 hips in 192 patients aged between
18 and 65 years were randomised to 28-mm metal-on-metal uncemented
total hip replacements (THRs, 107 hips) or hybrid hip resurfacing
(HR, 112 hips). At a mean follow-up of eight years (6.6 to 9.3)
there was no significant difference between the THR and HR groups
regarding rate of revision (4.0% (4 of 99) Cite this article:
We report the clinical and radiological outcome of total ankle replacement performed in conjunction with hindfoot fusion or in isolation. Between May 2003 and June 2008, 60 ankles were treated with total ankle replacement with either subtalar or triple fusion, and the results were compared with a control group of 288 ankles treated with total ankle replacement alone. After the mean follow-up of 39.5 months (12 to 73), the ankles with hindfoot fusion showed significant improvement in the mean visual analogue score for pain (p <
0.001), the mean American Orthopaedic Foot and Ankle Society score (p <
0.001), and the mean of a modified version of this score (p <
0.001). The mean visual analogue pain score (p = 0.304) and mean modified American Orthopaedic Foot and Ankle Society score (p = 0.119) were not significantly different between the hindfoot fusion and the control groups. However, the hindfoot fusion group had a significantly lower mean range of movement (p = 0.009) and a higher rate of posterior focal osteolysis (p = 0.04). Both groups showed various complications (p = 0.131) and failure occurring at a similar rate (p = 0.685). Subtalar or triple fusion is feasible and has minimal adverse effects on ankles treated with total ankle replacement up to midterm follow-up. The clinical outcome of total ankle replacement when combined with hindfoot fusion is comparable to that of ankle replacement alone. Thus, hindfoot fusion should be performed in conjunction with total ankle replacement when indicated.
We describe the survivorship of the Exeter femoral component in a District General Hospital. Between 1994 and 1996, 230 Exeter Universal cemented femoral components were implanted in 215 patients who were reviewed at a mean of 11.2 years (10 to 13). We used one acetabular implant, the Elite Ogee component, in 218 of the 230 hips. During the period of this study 76 patients (79 hips) died. Of the remaining 139 patients (151 hips), 121 were able to attend for radiological analysis at a minimum of ten years. One patient was lost to follow-up. No femoral component was revised for aseptic loosening. Three hips were revised for deep infection and six acetabular components required revision, four for loosening and two for recurrent dislocation. Taking the ‘worst-case scenario’ including the one patient lost to follow-up, the overall survival rate was 94.4% at 13 years. Our results confirm excellent medium-term results for the Exeter Universal femoral component, implanted in a general setting. The excellent survival of this femoral component, when used in combination with the Ogee acetabular component, suggests that this is a successful pairing.
We investigated whether patients who underwent internal fixation for an isolated acetabular fracture were able to return to their previous sporting activities. We studied 52 consecutive patients with an isolated acetabular fracture who were operated on between January 2001 and December 2002. Their demographic details, fracture type, rehabilitation regime, outcome and complications were documented prospectively as was their level and frequency of participation in sport both before and after surgery. Quality of life was measured using the EuroQol-5D health outcome tool (EQ-5D). There was a significant reduction in level of activity, frequency of participation in sport (both p <
0.001) and EQ-5D scores in patients of all age groups compared to a normal English population (p = 0.001). A total of 22 (42%) were able to return to their previous level of activities: 35 (67%) were able to take part in sport at some level. Of all the parameters analysed, the Matta radiological follow-up criteria were the single best predictor for resumption of sporting activity and frequency of participation.
If a modular convertible total shoulder system
is used as a primary implant for an anatomical total shoulder arthroplasty,
failure of the prosthesis or the rotator cuff can be addressed by
converting it to a reverse shoulder arthroplasty (RSA), with retention
of the humeral stem and glenoid baseplate. This has the potential
to reduce morbidity and improve the results. In a retrospective study of 14 patients (15 shoulders) with a
mean age of 70 years (47 to 83) we reviewed the clinical and radiological
outcome of converting an anatomical shoulder arthroplasty (ASA)
to a RSA using a convertible prosthetic system (SMR system, Lima,
San Daniele, Italy). The mean operating time was 64 minutes (45 to 75). All humeral
stems and glenoid baseplates were found to be well-fixed and could
be retained. There were no intra-operative or early post-operative
complications and no post-operative infection. The mean follow-up was 43 months (21 to 83), by which time the
mean visual analogue scale for pain had decreased from 8 pre-operatively
to 1, the mean American Shoulder and Elbow Surgeons Score from 12
to 76, the mean Oxford shoulder score from 3 to 39, the mean Western
Ontario Osteoarthritis of the Shoulder Score from 1618 to 418 and
the mean Subjective shoulder value from 15 to 61. On radiological review, one patient had a lucency around the
humeral stem, two had stress shielding. There were no fatigue fractures
of the acromion but four cases of grade 1 scapular notching. The use of a convertible prosthetic system to revise a failed
ASA reduces morbidity and minimises the rate of complications. The
mid-term clinical and radiological results of this technique are
promising. Cite this article:
We evaluated clinical and radiographic outcomes of total shoulder
arthroplasty (TSA) using the second-generation Trabecular Metal
(TM) Glenoid component. The first generation component was withdrawn
in 2005 after a series of failures were reported. Between 2009 and
2012, 40 consecutive patients with unilateral TSA using the second-generation
component were enrolled in this clinical study. The mean age of
the patients was 63.8 years (40 to 75) and the mean follow-up was
38 months (24 to 42). Patients were evaluated using the Constant score (CS), the American
Shoulder and Elbow Surgeons (ASES) score and routine radiographs.Aims
Methods
Contracture of the collateral ligaments is considered to be an important factor in post-traumatic stiffness of the elbow. We reviewed the results of isolated release of the medial collateral ligament in a series of 14 patients with post-traumatic loss of elbow flexion treated between 1998 and 2002. There were nine women and five men with a mean age of 45 years (17 to 76). They were reviewed at a mean follow-up of 25 months (9 to 48). The operation was performed through a longitudinal posteromedial incision centred over the ulnar nerve. After decompression of the ulnar nerve, release of the medial collateral ligament was done sequentially starting with the posterior bundle and the transverse component of the ligament, with measurement of the arc of movement after each step. If full flexion was not achieved the posterior half of the anterior bundle of the medial collateral ligament was released. At the latest follow-up, the mean flexion of the elbow improved significantly from 96° (85° to 115°) pre-operatively to 130° (110° to 150°) at final follow-up (p = 0.001). The mean extension improved significantly from 43° (5° to 90°) pre-operatively to 22° (5° to 40°) at final follow-up (p = 0.003). There was a significant improvement in the functional outcome. The mean Broberg and Morrey score increased from a mean of 54 points (29.5 to 85) pre-operatively to 87 points (57 to 99) at final follow-up (p <
0.001). All the patients had normal elbow stability. Our results indicate that partial surgical release of the medial collateral ligament is associated with improved range of movement of the elbow in patients with post-traumatic stiffness, but was less effective in controlling pain.
Most published randomised controlled trials which
compare the rates of wear of conventional and cross-linked (XL) polyethylene
(PE) in total hip arthroplasty (THA) have described their use with
a cementless acetabular component. We conducted a prospective randomised study to assess the rates
of penetration of two distinct types of PE in otherwise identical
cemented all-PE acetabular components. A total of 100 consecutive patients for THA were randomised to
receive an acetabular component which had been either highly XL
then remelted or moderately XL then annealed. After a minimum of eight years follow-up, 38 hips in the XL group
and 30 hips in the annealed group had complete data (mean follow-up
of 9.1 years (7.6 to 10.7) and 8.7 years (7.2 to 10.2), respectively).
In the XL group, the steady state rate of penetration from one year
onwards was -0.0002 mm/year ( These results show that the yearly linear rate of femoral head
penetration can be significantly reduced by using a highly XLPE
cemented acetabular component. Cite this article:
Periprosthetic joint infection (PJI) is a devastating
complication for patients and results in greatly increased costs
of care for both healthcare providers and patients. More than 15
500 revision hip and knee procedures were recorded in England, Wales
and Northern Ireland in 2013, with infection accounting for 13%
of revision hip and 23% of revision knee procedures. We report our experience of using antibiotic eluting absorbable
calcium sulphate beads in 15 patients (eight men and seven women
with a mean age of 64.8 years; 41 to 83) as part of a treatment
protocol for PJI in revision arthroplasty. The mean follow-up was 16 months (12 to 22). We report the outcomes
and complications, highlighting the risk of hypercalcaemia which
occurred in three patients. We recommend that serum levels of calcium be routinely sought
following the implantation of absorbable calcium sulphate beads
in orthopaedic surgery. Cite this article:
In a prospective study, 93 unselected consecutive uncemented hip arthroplasties were performed in 80 patients using the titanium-coated RM acetabular component and the CLS femoral component. The mean age of the patients at operation was 52 years (28 to 81). None were lost to follow-up. In the 23 patients who had died (26 hips) only one acetabular component had been revised. In the 57 living patients (67 hips), 13 such revisions had been performed. Of the 14 revisions, seven were for osteolysis, five for loosening and two for infection. Survival analysis of this implant showed a total probability of survival of 83% (95% confidence interval 73 to 90), with all revisions as the endpoint, and a probability of 94% (95% confidence interval 87 to 98) with revision for aseptic loosening as the endpoint, indicating reliable long-term fixation of the titanium-coated RM acetabular component.
Radiological assessment of total and unicompartmental
knee replacement remains an essential part of routine care and follow-up.
Appreciation of the various measurements that can be identified
radiologically is important. It is likely that routine plain radiographs
will continue to be used, although there has been a trend towards
using newer technologies such as CT, especially in a failing knee,
where it provides more detailed information, albeit with a higher
radiation exposure. The purpose of this paper is to outline the radiological parameters
used to evaluate knee replacements, describe how these are measured
or classified, and review the current literature to determine their
efficacy where possible.
Injuries to the spinal cord may be associated with increased healing of fractures. This can be of benefit, but excessive bone growth can also cause considerable adverse effects. We evaluated two groups of patients with fractures of the spinal column, those with neurological compromise (n = 10) and those without (n = 15), and also a control group with an isolated fracture of a long bone (n = 12). The level of transforming growth factor-beta (TGF-β), was measured at five time points after injury (days 1, 5, 10, 42 and 84). The peak level of 142.79 ng/ml was found at day 84 in the neurology group (p <
0.001 Our findings suggest that TGF-β may have a role in the increased bone turnover and attendant complications seen in patients with acute injuries to the spinal cord.
We present the early clinical and radiological results of Articular Surface Replacement (ASR) resurfacings in 214 hips (192 patients) with a mean follow-up of 43 months (30 to 57). The mean age of the patients was 56 years (28 to 74) and 85 hips (40%) were in 78 women. The mean Harris hip score improved from 52 (11 to 81) to 95 (27 to 100) at two years and the mean University of California, Los Angeles activity score from 3.9 (1 to 10) to 7.4 (2 to 10) in the same period. Narrowing of the neck (to a maximum of 9%) was noted in 124 of 209 hips (60%). There were 12 revisions (5.6%) involving four (1.9%) early fractures of the femoral neck and two (0.9%) episodes of collapse of the femoral head secondary to avascular necrosis. Six patients (2.8%) had failure related to metal wear debris. The overall survival for our series was 93% (95% confidence interval 80 to 98) and 89% (95% confidence interval 82 to 96) for hips with acetabular components smaller than 56 mm in diameter. The ASR implant has a lower diametrical clearance and a subhemispherical acetabular component when compared with other more frequently implanted metal-on-metal hip resurfacings. These changes may contribute to the higher failure rate than in other series, compared with other designs. Given our poor results with the small components we are no longer implanting the smaller size.
Isolated fractures of the anterior column and anterior wall are a relatively rare subgroup of acetabular fractures. We report our experience of 30 consecutive cases treated over ten years. Open reduction and internal fixation through an ilioinguinal approach was performed for most of these cases (76.7%) and percutaneous techniques were used for the remainder. At a mean follow-up of four years (2 to 6), 26 were available for review. The radiological and functional outcomes were good or excellent in 23 of 30 patients (76.7%) and 22 of 26 patients (84.6%) according to Matta’s radiological criteria and the modified Merlé d’Aubigné score, respectively. Complications of minor to moderate severity were seen in six of the 30 cases (20%) and none of the patients underwent secondary surgery or replacement of the hip.
We report our experience of performing an elbow
hemiarthroplasty in the treatment of comminuted distal humeral fractures
in the elderly patients. A cohort of 42 patients (three men and 39 women, mean age 72;
56 to 84) were reviewed at a mean of 34.3 months (24 to 61) after
surgery. Functional outcome was measured with the Mayo Elbow Performance
Score (MEPS) and range of movement. The disabilities of the arm,
shoulder and hand questionnaire (DASH) was used as a patient rated
evaluation. Complications and ulnar nerve function were recorded.
Plain radiographs were obtained to assess prosthetic loosening,
olecranon wear and heterotopic bone formation. The mean extension deficit was 23.5° (0° to 60°) and mean flexion
was 126.8° (90° to 145°) giving a mean arc of 105.5° (60° to 145°).
The mean MEPS was 90 (50 to 100) and a mean DASH score of 20 (0
to 63). Four patients had additional surgery for limited range of
movement and one for partial instability. One elbow was revised
due to loosening, two patients had sensory ulnar nerve symptoms,
and radiographic signs of mild olecranon wear was noted in five
patients. Elbow hemiarthroplasty for comminuted intra-articular distal
humeral fractures produces reliable medium-term results with functional
outcome and complication rates, comparable with open reduction and
internal fixation and total elbow arthroplasty. Cite this article:
Failed internal fixation of a fracture of the
proximal humerus produces many challenges with limited surgical options.
The aim of this study was to evaluate the clinical outcomes after
the use of a reverse shoulder arthroplasty under these circumstances.
Between 2007 and 2012, 19 patients (15 women and four men, mean
age 66 years; 52 to 82) with failed internal fixation after a proximal
humeral fracture, underwent implant removal and reverse shoulder arthroplasty
(RSA). The mean follow-up was 36 months (25 to 60). The mean American
Shoulder and Elbow Score improved from 27.8 to 50.1 (p = 0.019).
The mean Simple Shoulder Test score improved from 0.7 to 3.2 (p
= 0.020), and the mean visual analogue scale for pain improved from
6.8 to 4.3 (p = 0.012). Mean forward flexion improved from 58.7°
to 101.1° (p <
0.001), mean abduction from 58.7° to 89.1° (p
= 0.012), mean external rotation from 10.7° to 23.1° (p = 0.043)
and mean internal rotation from buttocks to L4 (p = 0.034). A major
complication was recorded in five patients (26%) (one intra-operative
fracture, loosening of the humeral component in two and two peri-prosthetic
fractures). A total of 15 patients (79%) rated their outcome as
excellent or good, one (5%) as satisfactory, and three (16%) as
unsatisfactory. An improvement in outcomes and pain can be expected when performing
a RSA as a salvage procedure after failed internal fixation of a
fracture of the proximal humerus. Patients should be cautioned about
the possibility for major complications following this technically
demanding procedure. Cite this article:
We retrospectively reviewed 21 patients (22 shoulders) who presented with deep infection after surgery to the shoulder, 17 having previously undergone hemiarthroplasty and five open repair of the rotator cuff. Nine shoulders had undergone previous surgical attempts to eradicate their infection. The diagnosis of infection was based on a combination of clinical suspicion (16 shoulders), positive frozen sections (>
5 polymorphonuclear leukocytes per high-power field) at the time of revision (15 shoulders), positive intra-operative cultures (18 shoulders) or the pre-operative radiological appearances. The patients were treated by an extensive debridement, intravenous antibiotics, and conversion to a reverse shoulder prosthesis in either a single- (10 shoulders) or a two-stage (12 shoulders) procedure. At a mean follow-up of 43 months (25 to 66) there was no evidence of recurrent infection. All outcome measures showed statistically significant improvements. Mean abduction improved from 36.1° (