The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 146 - 152
1 Jan 2010
Bilen FE Kocaoglu M Eralp L Balci HI

We report the results of using a combination of fixator-assisted nailing with lengthening over an intramedullary nail in patients with tibial deformity and shortening. Between 1997 and 2007, 13 tibiae in nine patients with a mean age of 25.4 years (17 to 34) were treated with a unilateral external fixator for acute correction of deformity, followed by lengthening over an intramedullary nail with a circular external fixator applied at the same operating session. At the end of the distraction period locking screws were inserted through the intramedullary nail and the external fixator was removed.

The mean amount of lengthening was 5.9 cm (2 to 8). The mean time of external fixation was 90 days (38 to 265). The mean external fixation index was 15.8 days/cm (8.9 to 33.1) and the mean bone healing index was 38 days/cm (30 to 60).

One patient developed an equinus deformity which responded to stretching and bracing. Another developed a drop foot due to a compartment syndrome, which was treated by fasciotomy. It recovered in three months. Two patients required bone grafting for poor callus formation.

We conclude that the combination of fixator-assisted nailing with lengthening over an intramedullary nail can reduce the overall external fixation time and prevent fractures and deformity of the regenerated bone.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1281 - 1286
1 Oct 2009
Olsen M Davis ET Chiu M Gamble P Tumia N Boyle RA Schemitsch EH

The computed neck-shaft angle and the size of the femoral component were recorded in 100 consecutive hip resurfacings using imageless computer-navigation and compared with the angle measured before operation and with actual component implanted. The reliability of the registration was further analysed using ten cadaver femora. The mean absolute difference between the measured and navigated neck-shaft angle was 16.3° (0° to 52°). Navigation underestimated the measured neck-shaft angle in 38 patients and the correct implant size in 11. Registration of the cadaver femora tended to overestimate the correct implant size and provided a low level of repeatability in computing the neck-shaft angle.

Prudent pre-operative planning is advisable for use in conjunction with imageless navigation since misleading information may be registered intraoperatively, which could lead to inappropriate sizing and positioning of the femoral component in hip resurfacing.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 10 | Pages 1388 - 1393
1 Oct 2009
Aird JJ Hogg A Rollinson P

In 1937 Blount described a series of 28 patients with ‘Tibia vara’. Since then, a number of deformities in the tibia and the femur have been described in association with this condition.

We analysed 14 children with Blount’s disease who were entered into a cross-sectional study. Their mean age was 10 (2 to 18). They underwent a clinical assessment of the rotational profile of their legs and a CT assessment of the angle of anteversion of their hips (femoral version). We compared our results to previously published controls. A statistically significant increase in femoral anteversion was noted in the affected legs, with on average the femurs in patients with Blount’s disease being 26° more anteverted than those in previously published controls.

We believe this to be a previously unrecognised component of Blount’s disease, and that the marked intoeing seen in the disease may be partly caused by internal femoral version, in addition to the well-recognised internal tibial version.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 952 - 956
1 Jul 2008
Haddad F Chemali R Maalouf G

Dysplasia epiphysealis hemimelica of the left proximal femur was diagnosed in an eight-month-old girl. At the age of 18 months, radiographs of the hip and MRI showed overgrowth and loss of containment of the femoral head. She underwent resection of the superior portion of the head and neck of the femur at the age of 2.5 years. Six months later further radiographs and an MR scan show that the mass has increased in size and that hip containment has been lost. Further plain radiographs have shown that the left knee, ankle and spine were involved.

To the authors’ knowledge, this is the first report of dysplasia epiphysealis hemimelica involving both the lower limb and the spine. A review of the literature is presented.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1612 - 1617
1 Dec 2009
Venkatesh KP Modi HN Devmurari K Yoon JY Anupama BR Song HR

Extensive limb lengthening may be indicated in achondroplastic patients who wish to achieve a height within the normal range for their population. However, increasing the magnitude of lengthening is associated with further complications particularly adjacent joint stiffness and fractures. We studied the relationship between the magnitude of femoral lengthening and callus pattern, adjacent joint stiffness and fracture of the regenerate bone in 40 femoral lengthenings in 20 achondroplastic patients. They were divided into two groups; group A had lengthening of less than 50% and group B of more than 50% of their initial femoral length. The patterns of radiological callus formation were classified according to shape, type and features. The incidence of callus features, knee stiffness and regenerate bone fracture were analysed in the two groups. Group B was associated with an increased incidence of concave, lateral and central callus shapes, adjacent joint and stiffness and fracture. Statistically, the incidence of stiffness in adjacent joints and regenerate bone fracture was significantly associated with the magnitude of lengthening.

We suggest that careful radiological assessment of the patterns of callus formation is a useful method for the evaluation and monitoring of regenerate bone.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1127 - 1133
1 Sep 2009
Kang S Sanghera T Mangwani J Paterson JMH Ramachandran M

We performed a systematic review of the optimal management of septic arthritis in children as recommended in the current English literature using MEDLINE, EMBASE, CINAHL, the Cochrane Library and reference lists of retrieved articles without date restrictions up to 31 January 2009. From 2236 citations, 227 relevant full-text articles were screened in detail; 154 papers fulfilled the inclusion criteria, from which conclusions were drawn on the management of infected joints in children.

Our review showed that no single investigation, including joint aspiration, is sufficiently reliable to diagnose conclusively joint infection. The roles of aspiration, arthrotomy and arthroscopy in treatment are not clear cut, and the ideal duration of antibiotic therapy is not yet fully defined. These issues are discussed. Further large-scale, multi-centre studies are needed to delineate the optimal management of paediatric septic arthritis.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 9 | Pages 1133 - 1134
1 Sep 2007
Haddad FS Ashby E Konangamparambath S

Due to economic constraints, it has been suggested that joint replacement patients can be followed up in primary care. There are clinical, ethical and academic reasons why we must ensure that our joint replacements are appropriately clinically and radiologically followed up to minimise complications. This Editorial discusses this.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1323 - 1327
1 Oct 2008
Kim S Lee D Kim T Choi N

We present the operative technique and clinical results of concomitant reconstruction of the medial collateral ligament (MCL) and the posterior oblique ligament for medial instability of the knee using autogenous semitendinosus tendon with preservation of the tibial attachment. The semitendinosus tendon graft between the screw on the medial epicondyle and the tibial attachment of the graft was overlapped by the MCL, while the graft between the screw and the insertion of the direct head of the semimembranosus tendon was overlapped by the central arm of the posterior oblique ligament. Assessment was by stress radiograph and the Lysholm knee scoring scale. After a mean follow-up of 52.6 months (25 to 92), the medial joint opening of the knee was within 2 mm in 22 of 24 patients. The mean Lysholm score was 91.9 (80 to 100).

Concomitant reconstruction of the MCL and posterior oblique ligament using autogenous semitendinosus tendon provides a good solution to medial instability.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 962 - 967
1 Jul 2009
Aston WJS Calder PR Baker D Hartley J Hill RA

We present a retrospective review of a single-surgeon series of 30 consecutive lengthenings in 27 patients with congenital short femur using the Ilizarov technique performed between 1994 and 2005.

The mean increase in length was 5.8 cm/18.65% (3.3 to 10.4, 9.7% to 48.8%), with a mean time in the frame of 223 days (75 to 363). By changing from a distal to a proximal osteotomy for lengthening, the mean range of knee movement was significantly increased from 98.1° to 124.2° (p = 0.041) and there was a trend towards a reduced requirement for quadricepsplasty, although this was not statistically significant (p = 0.07). The overall incidence of regenerate deformation or fracture requiring open reduction and internal fixation was similar in the distal and proximal osteotomy groups (56.7% and 53.8%, respectively). However, in the proximal osteotomy group, pre-placement of a Rush nail reduced this rate from 100% without a nail to 0% with a nail (p < 0.001). When comparing a distal osteotomy with a proximal one over a Rush nail for lengthening, there was a significant decrease in fracture rate from 58.8% to 0% (p = 0.043).

We recommend that in this group of patients lengthening of the femur with an Ilizarov construct be carried out through a proximal osteotomy over a Rush nail. Lengthening should also be limited to a maximum of 6 cm during one treatment, or 20% of the original length of the femur, in order to reduce the risk of complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 535 - 537
1 Apr 2007
Evangelista GT Fulkerson E Kummer F Di Cesare PE

During open reduction of an irreducible anterior dislocation of a total hip replacement with an Oxinium femoral head, it was observed that the head had been significantly damaged. Gross and scanning electron microscopic examination revealed cracking, gouging, and delamination of the surface. Because of the risk which this poses for damaging the polyethylene acetabular liner, it is strongly recommended that patients with this type of prosthetic head be carefully monitored after a dislocation.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 2 | Pages 235 - 241
1 Feb 2010
van der Lugt JCT Valstar ER Witvoet-Braam SW Nelissen RGHH

Mechanical loosening which begins with early-onset migration of the prosthesis is the major reason for failure of the Souter-Strathclyde elbow replacement. In a prospective study of 18 Souter-Strathclyde replacements we evaluated the patterns of migration using roentgen stereophotogrammetric analysis. We had previously reported the short-term results after a follow-up of two years which we have now extended to a mean follow-up of 8.2 years (1 to 11.3). Migration was assessed along the co-ordinal axes and overall micromovement was expressed as the maximum total point movement. The alignment of the prosthesis and the presence of radiolucent lines were examined on conventional standardised radiographs.

All the humeral components showed increased and variable patterns of migration at the extended follow-up and four humeral components were revised. The maximum total point movement at two years in the revised prostheses was 1.8 mm (sd 1.0) and in the non-revised 0.7 mm (sd 0.5, p = 0.01). Most humeral components migrated into external rotation resulting in an anterior and varus tilt. The ulnar components remained stable.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 154 - 158
1 Feb 2008
Calder JD Hine AL Pearse MF Revell PA

Total hip replacement in patients with advanced osteonecrosis of the femoral head is often complicated by early loosening of the femoral component. Recent evidence has suggested that abnormal bone extending into the proximal femur may be responsible for the early failure of the femoral component. We aimed to identify which patients were at high risk of early failure by evaluating gadolinium-enhanced MR images of histologically-confirmed osteonecrotic lesions beyond the femoral head.

Although the MR signal intensity has been shown to correlate well with osteonecrosis in the femoral head, it was found to be relatively insensitive at identifying lesions below the head, with a sensitivity of only 51% and a predictive value of a negative result of only 48%. However, the specificity was 90%, with the predictive value of a positive MRI finding being 86%. Only those patients with osteonecrosis of the femoral head secondary to sickle-cell disease, who are known to be at high risk of early loosening, had changes in the MR signal in the greater trochanter and the femoral shaft. This observation suggests that changes in the MR signal beyond the femoral head may represent osteonecrotic lesions in areas essential for the fixation of the femoral component. Pre-operative identification of such lesions in the neck of the femur may be important when considering hip resurfacing for osteonecrosis of the femoral head, following which early loosening of the femoral component and fracture of the neck are possible complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 4 | Pages 500 - 507
1 Apr 2010
Lu Y Huang C Chang T Ho F Cheng C Huang C

Components from 73 failed knee replacements (TKRs) consisting of rotating-platform, mobile-bearing and fixed-bearing implants were examined to assess the patterns of wear. The patterns were divided into low-grade (burnishing, abrasion and cold flow) and high-grade (scratching, pitting/metal embedding and delamination) to assess the severity of the wear of polyethylene.

The rotating-platform group had a higher incidence of low-grade wear on the upper surface compared with the fixed-bearing group. By contrast, high-grade wear comprising scratching, pitting and third-body embedding was seen on the lower surface. Linear regression analysis showed a significant correlation of the wear scores between the upper and lower surfaces of the tibial insert (R2 = 0.29, p = 0.04) for the rotating-platform group, but no significant correlation was found for the fixed-bearing counterpart.

This suggests that high-grade wear patterns on the upper surface are reduced with the rotating-platform design. However, the incidence of burnishing, pitting/third-body embedding and scratching wear patterns on the lower surface was higher compared with that in the fixed-bearing knee.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 455 - 459
1 Apr 2008
Bollen S Pease F Ehrenraich A Church S Skinner J Williams A

The potential harm to the growth plate following reconstruction of the anterior cruciate ligament in skeletally-immature patients is well documented, but we are not aware of literature on the subject of the fate of the graft itself. We have reviewed five adolescent males who underwent reconstruction of the ligament with four-strand hamstring grafts using MR images taken at a mean of 34.6 months (18 to 58) from the time of operation. The changes in dimension of the graft were measured and compared with those taken at the original operation. No growth arrest was seen on radiological or clinical measurement of leg-length discrepancy, nor was there any soft-tissue contracture. All the patients regained their pre-injury level of activity, including elite-level sport in three. The patients grew by a mean of 17.3 cm (14 to 24). The diameter of the grafts did not change despite large increases in length (mean 42%; 33% to 57%). Most of the gain in length was on the femoral side. Large changes in the length of the grafts were seen.

There is a considerable increase in the size of the graft, so some neogenesis must occur; the graft must grow.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 12 - 19
1 Jan 2010
Vendittoli P Roy A Mottard S Girard J Lusignan D Lavigne M

We have updated our previous randomised controlled trial comparing release of chromium (Cr) and cobalt (Co) ions and included levels of titanium (Ti) ions. We have compared the findings from 28 mm metal-on-metal total hip replacement, performed using titanium CLS/Spotorno femoral components and titanium AlloFit acetabular components with Metasul bearings, with Durom hip resurfacing using a Metasul articulation or bearing and a titanium plasma-sprayed coating for fixation of the acetabular component.

Although significantly higher blood ion levels of Cr and Co were observed at three months in the resurfaced group than in total hip replacement, no significant difference was found at two years post-operatively for Cr, 1.58 μg/L and 1.62 μg/L respectively (p = 0.819) and for Co, 0.67 μg/L and 0.94 μg/L respectively (p = 0.207). A steady state was reached at one year in the resurfaced group and after three months in the total hip replacement group. Interestingly, Ti, which is not part of the bearing surfaces with its release resulting from metal corrosion, had significantly elevated ion levels after implantation in both groups. The hip resurfacing group had significantly higher Ti levels than the total hip replacement group for all periods of follow-up. At two years the mean blood levels of Ti ions were 1.87 μg/L in hip resurfacing and and 1.30 μg/L in total hip replacement (p = 0.001).

The study confirms even with different bearing diameters and clearances, hip replacement and 28 mm metal-on-metal total hip replacement produced similar Cr and Co metal ion levels in this randomised controlled trial study design, but apart from wear on bearing surfaces, passive corrosion of exposed metallic surfaces is a factor which influences ion concentrations. Ti plasma spray coating the acetabular components for hip resurfacing produces significantly higher release of Ti than Ti grit-blasted surfaces in total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 595 - 598
1 May 2007
May O Matar WY Beaulé PE

Femoroacetabular impingement is recognised as being a cause of labral tears and chondral damage. We report a series of five patients who presented with persistent pain in the hip after arthroscopy for isolated labral debridement. All five had a bony abnormality consistent with cam-type femoroacetabular impingement. They had a further operation to correct the abnormality by chondro-osteoplasty of the femoral head-neck junction. At a mean follow-up of 16.3 months (12 to 24) all had symptomatic improvement.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 5 | Pages 668 - 673
1 May 2008
Karva AR Board TN Porter ML

We present a 25-year-old patient with juvenile rheumatoid arthritis and ankylosis of both hips and both knees treated by staged bilateral hip and knee arthroplasty. She was followed up for 18 months. We discuss the pre-operative planning, surgical details and post-operative rehabilitation.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 392 - 393
1 Mar 2006
Kejariwal D

Heparin is a drug widely used for thromboprophylaxis or treatment in many clinical situations. However, it can cause serious adverse effects, including thrombocytopenia, which is a potentially life-threatening condition. Unfortunately, heparin-induced thrombocytopenia is generally under-recognised and under-diagnosed. A case of heparin-induced thrombocytopenia occurring in a post-operative orthopaedic patient in association with prophylactic low-molecular-weight heparin is described.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1079 - 1083
1 Aug 2008
Moonen AFCM Thomassen BJW Knoors NT van Os JJ Verburg AD Pilot P

This prospective randomised clinical trial evaluated the effect of alternatives for allogeneic blood transfusions after total hip replacement and total knee replacement in patients with pre-operative haemoglobin levels between 10.0 g/dl and 13.0 g/dl. A total of 100 patients were randomly allocated to the Eprex (pre-operative injections of epoetin) or Bellovac groups (post-operative retransfusion of shed blood). Allogeneic blood transfusions were administered according to hospital policy.

In the Eprex group, 4% of the patients (two patients) received at least one allogeneic blood transfusion. In the Bellovac group, where a mean 216 ml (0 to 700) shed blood was retransfused, 28% (14 patients) required the allogeneic transfusion (p = 0.002). When comparing Eprex with Bellovac in total hip replacement, the percentages were 7% (two of 30 patients) and 30% (nine of 30 patients) (p = 0.047) respectively, whereas in total knee replacement, the percentages were 0% (0 of 20 patients) and 25% (five of 20 patients) respectively (p = 0.042).

Pre-operative epoetin injections are more effective but more costly in reducing the need for allogeneic blood transfusions in mildly anaemic patients than post-operative retransfusion of autologous blood.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 58 - 63
1 Jan 2009
Hooper G Rothwell A Frampton C

We have examined the outcome of 400 consecutive patients who underwent total knee replacement with the Low Contact Stress mobile-bearing system between 1993 and 1994 and were followed up for a minimum of ten years. All operations were performed by surgeons in Christchurch, New Zealand, who used no other knee prosthesis during the study time.

At ten years after operation 238 patients (244 knees) remained for independent clinical and radiological assessment. There was a significant improvement (p < 0.001) in the postoperative knee scores at one, three, seven and ten years, although a slight deterioration in the score occurred between seven and ten years which did not reach statistical significance. The survival for polyethylene wear or loosening was 97% (95% CI 96 to 98) and survival using reoperation for any cause was 92% (95% CI 90 to 94) at 12 years. Polyethylene wear was more common in the meniscal-bearing component, with five knees requiring revision and a further eight demonstrating early wear. Osteolysis was not seen in the rotating platform component, but was present in three of the meniscal-bearing knees.

There was no association between the radiological alignment at one year and those knees that subsequently required revision. However, there was an association between the overall limb alignment and the Western Ontario McMasters University score (p < 0.001).

The Low Contact Stress mobile-bearing total knee replacement has proved to be a reliable implant at ten years when used in primary knee replacement irrespective of the deformity and diagnosis.